Professional Documents
Culture Documents
A Comparison of Methods For Analyzing Drop Jump
A Comparison of Methods For Analyzing Drop Jump
jump performance
BACA ARNOLD
Medicine & Science in Sports & Exercise. 31:p 437-442, March 1999.
Author Information
Abstract
Back to Top
Purpose:
Back to Top
Methods:
The purpose of this study was to find out which technique yields the lowest
errors compared with the results obtained by the double force plate
technique. In this investigation, two force plates were used, one located
under the drop platform. Twenty-five drop jumps were analyzed with eight
different methods. There were large differences between the reference
method and other methods. Using the height of the drop platform (0.39 m)
to estimate the velocity at the end of the free fall, in conjunction with data
from one force plate, resulted in a mean difference of 4.2% (SD: 9.6%) in
the calculated jump height. Using video information to estimate the time
that the velocity of the CM fell to zero after the drop phase, in conjunction
with data from one force plate, resulted in differences in the jump height of
up to 17%.
Back to Top
Results:
Differences between the reference method and video based methods were
comparatively small (mean value of differences in jump height: −0.007 m,
SD: 0.013 m for the best of these methods) but not negligible.
Back to Top
Conclusions:
Back to Top
METHODS
Subjects and recording techniques. Five volunteer male subjects aged 18-
25 yr (height = 1.83 ± 0.034 m, body mass = 75.0 ± 5.5 kg) and in good
physical condition participated in this study. Informed consent was obtained
from the subjects in accordance with the policy statement of the American
College of Sports Medicine. The subjects performed five drop jumps. They
were instructed to keep their hands on their hips and to jump as high as they
could. Drop height was 0.39 m.
perimental setup and position of markers applied to the skin of the subject.
mple of a force-time curve of the vertical ground reaction force during a drop jump. At t0, the subjects stands still on the first force platform. t1 is the instant of take-off
from this platform, t2 that of landing on the second force platform. At t3, the vertical velocity of the total CM is zero. t4 is the instant of leaving the ground, t5 that of touch-
down. At t6, which has been estimated by visual inspection of the force data, the subject stands still again. tecc is the duration of the eccentric phase, tconc that of the
concentric phase. Portions of data which came from each platform are indicated.
Back to Top
Double force plate technique (reference method, method A). From the
ground reaction forces recorded by the first force plate vL (the vertical
landing velocity) was calculated as follows: Equation (1) where Fz(t) is the
vertical ground reaction force, g is the acceleration due to gravity and m is
the body mass. This velocity can be used to determine vOFF, hJUMP, and
hDROP (see Table 1): Equations (2)-(4) tECC and tCONC are calculated from t2
and t4(Fig. 2), which are estimated from the force data of the second force
platform and by calculating the instant t3 fulfilling the condition Equation
(5) ΔzECC and ΔzCONC(Table 1) can be calculated as follows: Equation (6)
and Equation (7) where τ is an integration variable.
Back to Top
Back to Top
Video positions plus force data technique (method G). In this method,
the parameter values were obtained by combining force and video data. t2
and t4(Fig. 2) were estimated from the force data. t3 was determined from
the video data using the lowest vertical position of the total CM. hJUMP was
identified using the heights of the CM at t4 and at the vertex of the flight
phase. hDROP was calculated as the change in height of the total CM from
the vertex of the drop phase to the point of landing assuming a vertical
velocity greater than or equal to zero at t1. vL and vOFF were calculated
from hJUMP and hDROP by Equations 9 and 11. tECC and tCONC were
computed using t2, t3 and t4, ΔzECC and ΔzCONC using the vertical positions
of the total CM at these instants.
Because the accuracy of force plates is very high, very precise parameter
values can be expected from the double force plate method, if an adequate
sampling frequency is chosen. Therefore, this technique has been selected
as the reference method. The variables quantifying the performance of the
drop jumps were calculated for each subject from this method (Table 2).
Percentage change values of these variables applying the other measuring
techniques are presented in Figures 3-7.
Variables quantifying drop jump performance. Values were found using the reference method.
Means ± SD of the percentage change values of the durations of the eccentric and concentric phases.
Means ± SD of the percentage change values of the downward and upward vertical displacements during the eccentric and concentric phases.
There are some errors, which may still be present in the double force plate
method: nonlinearity of the force measuring system; incorrect calibration;
errors due to analog to digital conversion, rounding, numerical integration,
and due to incorrect estimation of the instants of landing after the drop
phase and take-off from the ground. Most of these factors were discussed
by Kibele and Schwirtz (16). The data collection system was checked for
linearity by statically loading the drop platform and force platform 2 (see
Fig. 1) with weights up to 1000 N, and the linearity was better than 0.2%.
Beyond that, these weights have been put in different locations on force
platform 1 both with and without the box/pipe/drop platform construction.
Differences in the measured values were negligible. There are no impact
phases during take-off from the drop platform, which could cause problems
due to reduced resonant frequency of the measuring system. Before each
experimental session, the calibration of the data collection system was
controlled by measuring the body weight of the subject performing the drop
jumps by both force plates. Errors due to analog-digital conversion were
neglected, because the resolution of the converter used was high (16 bit)
and rounding errors were expected to cancel out during the integration. If
the sampling rate of the force data is too low, errors can be introduced by
numerical integration. Because, however, numerical integration based on
every second sampled force value resulted in almost the same results, these
errors were considered negligible for the present investigation. Small errors
were unavoidable in the estimation of t2 and t4. These errors should be,
however, not greater than 0.002 s because of the sampling frequency of
1000 Hz. A timing error of 0.002 s in the estimation of t2 would result in a
mean error of about 2% in the downward vertical displacement and a mean
error of less than 2% in the estimation of the drop height. The same error in
the estimation of t4 would result in mean errors of less than 2% in the
upward vertical displacement and the jump height.
Although the results of the one force platform method (B) were closest to
those from the reference method, this method is not a practical one for
analyzing drop jump performance. It was a difficult coordination task for
the subjects to land on the second force platform with dimensions 0.6 × 0.4
m after the jump again and to come to a standstill as soon as possible. This
means that the subject not only has to concentrate on the primary task of
jumping as high as possible but also on subsequent motions. Nevertheless,
the longer the integration period, the more linearity, calibration, numerical
integration, and rounding errors will affect the results.
General problems inherent in all video-based methods are the inadequacy of
mathematical models of the human body, the reduced sampling rate, and
errors introduced by the optical system. One major error source is the
marker shifting relative to the skeleton. Deviations of the point O1 fixed to
the subject's torso (see Fig. 1) strongly affected the results of the present
investigation.
Additional errors occur when incorrect time-points are pulled from video
data. In particular, estimating t3, the instant that the velocity of the CM was
zero, from the video data was a difficult task. The largest errors (−16.8% in
the jump height and −12.7% in the upward vertical displacement) in method
H, which estimates t3 from video information, were observed in a trial in
which t3 was estimated as being not more than 0.008 s (period of two
frames) later than that found in the reference method, using force data.
Differences of the video position technique (E) and the video position plus
force data technique (G) with regard to the reference method were
comparatively small (E: 0.000 ± 0.015 m, G: −0.007 ± 0.013 m for jump
height). Although the mean of the percentage change values of the jump
heights differed less for method E (see Fig. 3), it can be inferred from the
other results (in particular from the standard deviations, see Figs. 4-7) that
method G is superior to method E. A plausible explanation for the
comparatively high mean difference in the calculated effective drop heights
is apparent when the subjects' jumping technique is considered. The
subjects raised the leg more distant from the camera before dropping down.
At the instant that free fall started, the CM of this leg was still elevated.
Therefore, the assumption of a symmetric motion from the very beginning
is not justified, and the vertical position of the CM was underestimated. As
a consequence, the calculated effective drop height was too small. An
additional problem arises if the subjects are still in contact with the drop
platform at the instant that the CM reaches its maximum height. These
problems can be overcome by estimating vOFF from the jump height ((11))
and calculating the velocity of the CM at the instant of impact after the drop
phase by backward integration of the net vertical forces then. Equation (4)
can then be used to calculate the effective drop height. Large relative
differences were also observed for the vertical displacements during the
eccentric phase (Fig. 6). The corresponding absolute differences were,
however, not tremendously large. Marker shifting might be the main reason
for these differences, which were less than 11 mm on average.
Method D overestimated the jump height as well as the velocity at the end
of the concentric phase, because in most trials the subject was in a more
extended position at the end of the concentric phase than at the end of the
flight phase. From the mean values and standard deviations of the relative
differences of the individual subjects' five jump heights with regard to the
reference method (M.J. :8.4 ± 3.3%; C.S.: 5.2 ± 4.4%; M.M.: 5.6 ± 2.8%;
W.J: −0.9 ± 2.6%; H.K.: 1.8 ± 4.4%), it can be inferred that method D is
better suited to compare trials of the same subject than to compare trials of
different subjects or to calculate exact parameter values.
Back to Top
REFERENCES