CR-IR346/348CL
Service Manual
This manual is intended to assist the service person in the safe and efficient
servicing of the equipment described.
The equipment must be used in accordance with the procedures contained in this
manual and must not be used for purposes other than those which are described
herein.
It is the responsibility of the user to ensure that existing legal regulations regarding
installation of the building with respect to the equipment are observed. Moreover
the legal regulations regarding the operation of the equipment are to be observed.
All rights reserved. No part of this publication may be reproduced, stored in a retrieval
system, or transmitted in any form or by means, electronic, mechanical, photocopying,
recording or otherwise, without the prior written permission of Fuji Photo Film Co., Ltd.
010-054-14
CR-IR348CL Service Manual 07.30.2005 FM4543 (ITC)
WARNING
This manual is written for International distribution (including
distribution in the U.S.) and therefore may contain references to
010-054-14
CR-IR348CL Service Manual 07.30.2005 FM4543 (ITC)
BLANK PAGE
010-054-14
CR-IR348CL Service Manual 07.30.2005 FM4543 (ITC)
Introduction
Disclosing this document, in whole or in part, to a person other than NOTE Indicates items which require particular attention.
the authorized service personnel REFERENCE Explanation of related items and terms.
Lending or transferring this document, in whole or in part, to a third
party
010-054-10
02.20.2004 FM4220 (ITC)
CR-IR346CL Service Manual 27
Introduction
00010003.EPS
010-054-10
02.20.2004 FM4220 (ITC)
CR-IR346CL Service Manual 28
Introduction
The unit is also pasted with precautionary labels. Follow the instructions NEC product label
on the labels when carrying out work.
WEEE label
CL-PC CE label
CL-PC (GX150)
(GX110) Identification
Rating plate label
Identification
label
00010009.EPS
● Identification labels
● Identification labels
CR-IR348CL (Lite) CR-IR348CL Plus CR-IR348CL (Mobile model)
CR-IR346CL CR-IR348CL (Lite)
00010012.EPS
00010008.EPS
● Rating plate
CR-IR346CL CR-IR348CL
CR-IR348CL (Mobile model)
00010013.EPS
00010010.EPS
010-054-14
07.30.2005 FM4543 (ITC)
CR-IR346CL Service Manual 29
Introduction
Safety Precautions
■ Safety Precautions (Control Unit)
Follow the precautions below when installing, using, and servicing the
control unit (PC).
See the User’s Guide provided for details on the control unit.
010-054-14
07.30.2005 FM4543 (ITC)
CR-IR346CL Service Manual 30
Cautions Relating to Electromagnetic Compatibility
010-054-15
11.30.2005 FM4760 (ITC)
CR-IR346CL Service Manual 31
Cautions Relating to Electromagnetic Compatibility
Table 2
Guidance and manufacturer’s declaration - electromagnetic immunity
The CR-IR346CL/348CL is intended for use in the electromagnetic environment specified below.
The customer or the user of the CR-IR346CL/348CL should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
±6kV contact ±2kV contact
±4kV contact
±6kV contact
Electrostatic discharge (ESD) Floors should be wood, concrete or ceramic tile. If floors are covered
IEC61000-4-2 with synthetic material, the relative humidity should be at least 30%.
±8kV air ±2kV air
±4kV air
±8kV air
±2kV for power supply lines ±2kV for power supply lines
Electrical fast transient/burst Mains power quality should be that of a typical commercial or hospital
IEC61000-4-4 environment.
±1kV for input/output lines ±1kV for input/output lines
±1kV differential mode ±1kV differential mode
Surge Mains power quality should be that of a typical commercial or hospital
IEC61000-4-5 environment.
±2kV common mode ±2kV common mode
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 0.5 cycle for 0.5 cycle
40% UT 40% UT
(60% dip in UT) (60% dip in UT) Mains power quality should be that of a typical commercial or hospital
Voltage dips, short interruptions
for 5 cycles for 5 cycles environment. If the user of the CR-IR346CL/348CL requires continued
and voltage variations of power
operation during power mains interruptions, it is recommended that the
supply input lines
70% UT 70% UT CR-IR346CL/348CL be powered from an uninterruptible power supply or
IEC61000-4-11
(30% dip in UT) (30% dip in UT) a battery.
for 25 cycles for 25 cycles
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 5 s for 5 s
Power frequency (50/60Hz)
Power frequency magnetic fields should be at levels characteristic of a
magnetic field 3 A/m 3 A/m
typical location in a typical commercial or hospital environment.
IEC61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
010-054-14
07.30.2005 FM4543 (ITC)
CR-IR346CL Service Manual 32
Cautions Relating to Electromagnetic Compatibility
Table 3
Guidance and manufacturer’s declaration - electromagnetic immunity
The CR-IR346CL/348CL is intended for use in the electromagnetic environment specified below.
The customer or the user of the CR-IR346CL/348CL should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications equipment should be used no closer
IEC61000-4-6 150 kHz to 80 MHz to any part of the CR-IR346CL/348CL, including cables, than the
recommended separation distance calculated from the equation applicable to
the frequency of the transmitter.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the CR-IR346CL/348CL is used exceeds the applicable RF compliance, the CR-IR346CL/348CL should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the CR-IR346CL/348CL.
b Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
010-054-14
07.30.2005 FM4543 (ITC)
CR-IR346CL Service Manual 33
Cautions Relating to Electromagnetic Compatibility
Table 4
Recommended separation distances between
Portable and mobile RF communications equipment and the CR-IR346CL/348CL
The CR-IR346CL/348CL is intended for use in the electromagnetic environment in
which radiated RF disturbances are controlled.
The customer or the user of the CR-IR346CL/348CL can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the CR-IR346CL/348CL as
recommended below, according to the maximum output power of the communications
equipment.
Separation distance according to frequency of transmitter
Rated maximum
m
output power of
transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
W
d = 1.2 P d = 1.2 P d = 2.3 P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
010-054-14
07.30.2005 FM4543 (ITC)
CR-IR346CL Service Manual 34
Abbreviations adopted in
this manual:
PC : Hardware (personal computer) used in the CR-IR346CL controller
CL : CR-IR346CL
RU : CR-IR346RU
AP : Application Software
010-054-14
07.30.2005 FM4543 (ITC)
CR-IR346CL Service Manual 35
BLANK PAGE