Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

Airox Technologies Pvt. Ltd.

D-27, 4th Floor, Prozone Trade Centre,


API Corner, Chikalthana , Aurangabad-431 210 (M.S.)
Ph. : +91 240 6602686 .Mo. : +91 9764634964
Email : sales@airoxtechnologies.com
Web. : www.airoxtechnologies.com

DEALS IN: OXYGEN CONCENTRATOR | PNEUMATIC TUBE SYSTEM | HYPERBARIC CHAMBER | WOUNDHUB

Broad Based Specifications for Microprocessor based Oxygen Concentrator Plant

1. Oxygen Concentrator Plant

Microprocessor based PSA type Oxygen concentrator plant to produce minimum 47 Nominal
cu.m/hr (min. 1795 SCFH or min. 850 LPM) at 93% Purity of medical grade oxygen at 93 % +-
3% purity. The oxygen supply be continuous and supplied through a single unit. The oxygen
should be of medical grade and shall be supplied through pipelines at 65 psig pressure.

I. Features:
• Product Flow: Minimum 850LPM /(min. 1795 SCFH/ 47 Nm3/hr)1
• Filling Capacity: 175 Cylinder (D-Type)
• Outlet pressure should not less than 4.48 kg/ CM² or 65 PSI
• Product Concentration 93% ±3%
• Product Dew Point -100°F (-73°C)
• Electrical Supply – 230 VAC , 1 Phase, 10 Amp , 3pole MCB.
• Oxygen should Produces from an independent compressed air source
• Should be Microprocessor controlled
• Should be Automatic and unattended operation

II. Should have below Certifications:

A. The oxygen generator principal manufacturer must have U.S.FDA & CE or EC


(European certificate) certification in accordance with Appendix II (excluding section 4)
of the Medical Devices Directive 93/42/EC & its amendment 2007/47/EC.(Certificate
should be given by recognized European or American third party inspection authority
like LNE )

B. EN- ISO 13485 ( 2012)( Certificate should be given by recognized European or


American third party inspection authority like GMED )

2. Compressed air System (Air compressor) -Compressed air system comprising of


screw air compressor air cooled with microprocessor based control panel coupled with
motor assembly. The compressor should be capable of delivering medical grade air. The
compressor should be capable of delivering air at minimum 7 kg/cm2 and should be
capable of working on 100% continuous duty with starting pressure as atmospheric
pressure. The compressor shall have to be with all standard accessories compatible with
oxygen generator. To save the electricity screw compressor capacity should not be more
then 55 KW.
Technical feature:
- Compressor 55 KW (Maximum capacity allowed for better electricity saving)
- Motor KW - 55 KW / 3 Phase / 415 VAC / 50 hz.
- Flow rate – 310 SCFM
- Rated power supply – 150 Amp 4 pole MCCB
- Pressure - 8 Bar.
Airox Technologies Pvt. Ltd.
D-27, 4th Floor, Prozone Trade Centre,
API Corner, Chikalthana , Aurangabad-431 210 (M.S.)
Ph. : +91 240 6602686 .Mo. : +91 9764634964
Email : sales@airoxtechnologies.com
Web. : www.airoxtechnologies.com

DEALS IN: OXYGEN CONCENTRATOR | PNEUMATIC TUBE SYSTEM | HYPERBARIC CHAMBER | WOUNDHUB

3. Air Receiver:- Air receiver of 2000 liters capacity -The air receiver tank shall be of
2000 ltrs capacity (pressure tested to two times its normal working pressure) along with
auto drain valve

4. Oxygen Receiver : Oxygen receiver of 2000 liters capacity -The oxygen receiver tank
shall be of 2000 ltrs capacity (pressure tested to two times of its normal pressure) .

5. Refrigerant Air Dryer -Refrigerant type air dryer with inlet pressure equal to outlet
pressure from air compressor, inlet air temperature less than 40deg. C, ambient
temperature +0 deg.C To +40 deg.C, Dew point temperature of maximum +3 deg.C and
inlet air capacity compatible to air delivery capacity of air compressor.
Technical feature:
Capacity – 500 CFM ( 14.16 Cu.mtr/min)
3 Phase / 415 VAC / 50 hz.
Dimensions (L X W X H) 800 mm X 1000 mm X 1350 mm.
Weight - 250 Kg
Rated power supply – 16 Amp 4 pole MCCB

6. Filtration System for Compressed air: The feed air quality to the oxygen concentrator
should be confirming to ISO8573 class 4 and should have a filtration grade of 0.01
Micron .The ambient temperature of compressed air should be 100C to 400 C and should
have maximum dew point of +3 degree centigrade (Two stage Filtration System).
Technical feature:
Pre filter – 420 CFM @7.0 Bar Pressure. Cap – 1 Micron
Post Filter – 420 CFM @7.0 Bar Pressure. Cap – 0.01 Micron
Activated Carbon Filter - 420 CFM @7.0 Bar Pressure. Cap –
<0.003mg/Cu.mtr(<0.003ppm)

7. Main Electrical Panel - The panel should have automatic starters, overload protection,
single phasing preventers, timer assemblies, or successful operation of all the
components of oxygen plant.

8. Automatic Change Over System: Providing and Fitting of 01 No. Automatic Change
Over System / Panel to control the supply of Oxygen at 4.1 bar produced by
Microprocessor based Oxygen Concentrator and supply the oxygen through Oxygen
Manifold (cylinders). If pressure drops in supply of Oxygen from Oxygen Generator
Module, the system should automatically change over supply of Oxygen from Oxygen
Manifold (cylinders) and should also automatically change over from Oxygen Manifold
(cylinders) to Oxygen Concentrator when the pressure increases in Oxygen Supply Tank
of Oxygen Generator.

9. Ultrasonic based Oxygen Analyser:


The oxygen Analyser should have following features,
 Oxygen measure Standard : O2 : 0-100%
 Option : 95-100% O2 (MOQ of 100 pieces is required)
Airox Technologies Pvt. Ltd.
D-27, 4th Floor, Prozone Trade Centre,
API Corner, Chikalthana , Aurangabad-431 210 (M.S.)
Ph. : +91 240 6602686 .Mo. : +91 9764634964
Email : sales@airoxtechnologies.com
Web. : www.airoxtechnologies.com

DEALS IN: OXYGEN CONCENTRATOR | PNEUMATIC TUBE SYSTEM | HYPERBARIC CHAMBER | WOUNDHUB

 Oxygen accuracy ± 1.8% FS


 Flow accuracy ± 0.2L/min
 Flow direction No limitation
 Response time 0.5 sec
 Maximum inlet pressure 150 kPa
 Operating temperature 10-45°C
 Span Lifetime ≥ 5years
 Outputs RS232 / LCD / analogue VDC
 Oxygen concentration output Analogue 0-5VDC linear with 0-100% vol O2
 Flow volume output Analogue 0-5VDC linear with 0-10.0L/min or 0-500 L/m
 Power supply 12 VDC - 10mA

10. Medical Gas Alarm system : Providing and fitting of New Medical Gas Alarm system
in New Medical Oxygen Generation Gas Plant of Hospital.

11. The tenderer should provide 12 months warranty on the entire equipment.

12. The company should provide training for operation and maintenance system to the
hospital staff from the successful date of testing and commissioning of the system
without any extra cost.

13. The tenderer will establish service engineering branch in Andhra Pradesh state to
provide service in the case of failure of the unit within 24 hours on receipt of the
information from competent authority.

14. ELIGIBILITY CRITERIA:

1. The quoting firm's last three year average turnover should be more than 15 Crore. In
last three years their turn over should be more then 14 cr.
2. The quoting company should have local engineer in based in Andhra Pradesh.
3. The quoting firm's should have similar model installation in last 7 years,
Performance certificate with all the details should be attached.
4. The quoting firm should have at lest 05 installations in Andhra Pradesh of same
principal make whose oxygen generator quoted in the tender, Performance certificate
with all the details should be attached.
5. The quoting firm should have atlest 75-100 systems in India of same principal make
whose oxygen generator quoted in the tender, Performance certificate with all the
details should be attached. If Indian firm who has installed some other make oxygen
generator in India, that performance certificate will not be considered.
6. International Installation will not be considered.
7. Oxygen Plant must have US FDA and CE Medical Devices Certifications.

You might also like