Product Information

Software Version History Multiwave 3000 & Synthos 3000

Validity: v.2.52, v2.50, v2.02, v2.01, v2.00, v1.52, v1.51, v1.40, v1.27, v1.26.1, v1.26, v1.24.1,
v1.24, v1.23

Other authentic doc.: B83IB63 Multiwave 3000 Manual CD (Video Manual), description of current software
version (v2.0x)
B83AIG3 Installation Instruction
Latest version of manual available on the Anton Paar website.

Sequence
of Version history: Current versions are always entered at the beginning of the document.

Update/Installation: B83AIG3 Installation Instruction and as described in the reference manual.

Modification: updated information for v2.52

RECOMMENDATION
Anton Paar GmbH recommends to always install the latest software version on the instrument. The software is
available as a free download on the Anton Paar website. Keep the instrument up to date by regular updates. All
updated features assist you to work more easily, safely and convenient with your instrument.

Website: http://www.anton-paar.com

The new software version can be installed on every Multiwave 3000 and Synthos 3000 in the field.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 1 of 23
Product Information

Version v2.52
Improved Sensor Handling for Rotor 8S
This release of the Multiwave/Synthos 3000 software includes an improved sensor handling and is only relevant
for users of Rotor 8S.

Version v2.50
Bug-fix Frequency Detection Error
The error handling routine in case of bigger fluctuations of the power supply has been improved.

New Rotor types for Multiwave 3000


The rotor list has been extended and includes now following additional rotor types:

Rotor type: Max. temperature Max. IR-temperature Max. pressure

64MG5 200°C 200°C 20 bar

4x48MC* 200°C 180 C 20 bar

*Allows operating the rotor types 4x24MG5, 4x20MGC and 4x48MC

Version v2.02
Bug-fix Content of Review Menu
The displayed information in the "Sample Documentation" screen of the Review function of version v2.01 did not
correlate with the used method. Version v2.02 will now correctly display the entered data.

Bug-fix Reagents List


Some CPU cards show strange "ÿÿÿÿÿÿÿÿÿ"-data at the end of the "Reagents List". This bug was caused by an
improper software-routine, handling the memory chip and has been solved within v2.02.

Bug-fix – "Starting temperature lower than 10°C" with pT-Sensor


With some pT-Sensors, the instrument prompted "Starting temperature lower than 10°C, even if they have been
calibrated and found to be OK ". This problem is fixed in version v2.02.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 2 of 23
Product Information

Version v2.01
Bug-fix "Internal Memory Failure" if installation is done with empty memory chip
Installation of v2.00 onto CPUs with completely erased memory (e.g. after replacement of the battery packs, pre-
installation at Anton Paar GmbH) had failed, displaying an "Internal Memory Failure".
An additional routine has been included into the source code to overcome this failure.

Updated Information for v2.00


Software v2.00 now supports the new Rotor 16SOLV (i.e. all 16 IR-temperature graphs will be displayed
and stored).

Safety Check – Inner Temperature vs. IR Temperature


This function is intended to avoid that either a run is done with a defective sensor (prior to run) or continued after
a breakage of the capillary. With this function the software will help to avoid overheating of the vessel in case of
using a defective sensor (function originally introduced with software version v1.40).
Changes done in v2.00 (compared to v1.40):
Limit changed (reduced) to 20°C, i.e. as soon as the IR temperature value of vessel one exceeds the inner
temperature of the reference vessel by more than 20°C, the software will automatically abort the run and a pop-
up message will be displayed.

Rotor 8 – p(T)-Sensor has to be activated in setup menu


For detailed description refer to installation instruction B83AIG3

Safety Cooling Settings have to be verified


For detailed description refer to installation instruction B83AIG3

First Software Update requires a two-step installation


For detailed description refer to installation instruction B83AIG3

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 3 of 23
Product Information

Version v2.00 general


New Requirement for Software Update
All future Anton Paar software products and tools are based upon the new Microsoft.NET (Dot Net) development
platform. This allows best compatibility of our products with current and future versions of the Microsoft Windows
family of operating systems and a state-of-the-art user experience in Anton Paar software products and tools.

In order to provide a consistent and reliable experience across all supported operating system versions, it is
necessary to make major updates to some operating system installations and to limit the list of supported
operating system versions itself.

Based upon Microsoft's Support Lifecycle Policy and technical circumstances, the following software
requirements apply for new Anton Paar software products and tools:

Supported operating system versions:


Windows 2000
Windows XP

Only limited support is provided for these legacy operating systems:


Windows 98 SE (Second Edition)
Windows ME

Additional required Software:


Internet Explorer 5.01 or higher (already installed on Windows ME, 2000, XP and Server 2003)
Note: When installing, please make sure that the Internet Explorer language matches the operating system
language (e.g. only German Internet Explorer on German Windows)
Microsoft.NET Framework 2.0

These are the base requirements. For security reasons, Microsoft recommends keeping the operating system up
to date with the latest critical updates and Service Packs. Please refer to the Windows Update site.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 4 of 23
Product Information

Multilingual User interface


User interface is now additionally available in French, German, and Spanish.
Switching between the different languages is easily done in the setup menu; please refer to the Reference
Manual.

Limitation for MF100 Vessel improved


The operation limits for the MF100 vessel (since software version v1.40) have been adapted to allow a longer
run time if the vessel is not operated at maximum temperature limit.
New limits are:
Vessel operated between 150°C and 190°C:
 30 min maximum (hold time)
 Ramp time is no longer included
Therefore, you might be run the method for more than 30 minutes if you program a longer ramp (e.g. 15min
ramp to 175°C and additionally 30min hold time at this temperature = total run time of 45 minutes).
Valid for T- and P-Programs

Vessel operated below or up to 150°C:


 No time limitation
Only valid for temperature controlled runs.

Rotor EVAP - Protection Lid Detection deactivated


Because the Rotor EVAP is not equipped with a Protection Lid and due to an updated protection lid detection
circuit, the function "detection of rotor lid" has been deactivated for this rotor type.

SampleDoc – Vessel numbers included


The sample documentation field will now show a numbering at the left side, according to the rotor used (i.e. 1 – 8
for Rotor 8; 1 – 16 for Rotor 16…).

S-Control indication suppressed if less than five seconds


The "S-Control" which indicates that the 16A limits has been reached will only be displayed if the control takes
longer than 5 seconds.

Power Calibration Factor – Range extended


The range for the calculated power ratio has been extended so that you are now able to store new calibration
factors in a range of "0.7 to 1.3" (old: 0.8 – 1.2).

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 5 of 23
Product Information
Review (Run Data) & Method Memory
Due to the combined software (Multiwave 3000 and Synthos 3000) the method files take more space therefore
the available memory for storing the run-data is reduced.
With the current method file the remaining memory space will allow you to collect run-data of approx. 6 hours of
work before the oldest data is deleted.
If you want to collect more than these six hours simply delete no longer needed methods.

Synthos 3000 / Multiwave 3000 version combined


Due to the fact that both instruments are almost identical, we have combined both software versions to a single
one. This will make it easier for us to maintain the software and allows our distribution partners to switch easily
between the two versions, so that you can use the same demo equipment for Multiwave 3000 and Synthos 3000
demonstrations without reinstalling the software all the time.
To switch between the two versions, enter the System Constant menu; use the pgU/pgD keys on the external
keyboard to select the desired version.

IR Control Parameters
The IR Control Parameters have been made editable for further improvements of the IR control function.
You must not change any of these parameters without written permission of Anton Paar GmbH.

"Starting pressure out of range" message improved (Rotor 8)


If you close the Rotor 8 too tight (or less tight than necessary) the pop-up error message will now display the
current pressure so that it is easily to recognize if you have to re-open the rotor or to close it more tightly.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 6 of 23
Product Information

Reagents List enhanced


The list of reagents has been extended and now contains a broad range of solvents used for solvent extraction
or synthesis runs.
Following solvents have been added:

Acetone n-Hexane

Chloroform Methanol

Cyclohexane Methyl t-Butyl ether

Dichloromethane Tetrachloromethane

Ethanol Tetrahydrofuran

n-Heptane Toluene

"Usage Info" included


Additionally to the general information provided in the "Info" menu, the customer can now access a usage history
/ info of his system (see picture below).
This usage history includes:
 Firmware version of Power Board
 Overview of Magnetron runtime
 Overview of Fan runtime
 The total number of runs done with this instrument
 List of rotors and assorted number of runs done with this instrument

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 7 of 23
Product Information
Software Update changed
The software update function has been changed and therefore a few of the screens might look different. For
details regarding the software update, please refer to the reference manual.

Bug-fix – Starting temperature of p-Sensor


In some cases, when using the p-Sensor (which is without temperature probe) the instrument has prompted
"Starting temperature lower than 10°C". This problem is fixed in version v2.0.
It is not longer required to enter "fake"-calibration data for the p-Sensor. If you have used the p-Sensor with such
faked calibration data, please delete the data entry in the "Temp. Sensor Settings" menu.

Bug-fix – 70 bar Limit displayed in Review


In some cases, the stored run-data of Rotor 8(S)XQ runs displayed a limit of 70 bar even if the original limit of
80 bar has not been changed.
This problem is fixed in version v2.0.

Power Output adjusted according to Vessel Number


When entering the vessel number used for this method, the software will automatically adjust the presetting for
the power in accordance to the vessel number.

Following presetting will be used:


4–7 Vessels: 600 W
8 – 13 Vessels: 800 W
14 – 16 Vessels: 1400 W
If necessary, the preset value can be overwritten.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 8 of 23
Product Information

Software v2.0 Synthos User Interface only


Rotor 8 Starting Pressure Range extended
The allowed range for the starting pressure of Rotor 8 is between 2 bar and 25 bar.

Extended List of Rotors


Additional to the current available rotors (R48MF50, R16MF100, R16HF100, R8SXF100, R8SXQ80) the list will
now contain the following rotors:
 Rotor 4x48MC Well Plate
 Rotor 64MG5

The related specifications are:


Rotor 4x48MC Well Plate
 No protection lid detection
 4 positions
 IR limit 180°C

Rotor 64MG5
 No protection lid detection
 16 positions
 IR limit 200°C

Method Edit Menu


The "Substance/Scale" field has been shortened to five lines (previously ten).

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 9 of 23
Product Information

Version v1.52
Bug-fix Epson Printer Support
The problem in printing graphics with old EPSON needle printers has been corrected.

Bug-fix Power Board Firmware v11 Support


Older instruments, running a Power Board with firmware v11 have caused a "Missing Safety Module" error even
with normal decomposition rotors. Up to now, the only possibility to run a Rotor 48 on such instruments was to
upgrade the firmware (i.e. replacing the firmware chip) which needed the visit of a service engineer.
Version v1.52 corrects this bug.

Transmission Factor Range extended (IR Sensor)


The range of the transmission factor has been extended to ensure that even a slightly discoloration of the IR
sensor's Fresnel lens does not affect the IR calibration.
The new range is from 0.390 to 0.690

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 10 of 23
Product Information

Version v1.51
New Rotor Type – Rotor 48MF50
The rotor list has been extended and includes now the following additional rotor:

Rotor type: Temperature limit: IR-temperature limit: Operation pressure limit:

MF50 200°C 140°C 20 bar

Please note, run-data will contain only the 16 IR values, from the middle circle of vessels.

Start Sequence enhanced


Start procedure has been rewritten to improve the detection of the Safety Module SOLV and the Exhaust Unit.

Minor Restrictions / Problems


Rotor 8SOLV
When starting a SOLV method the instrument will prompt to check the water level of the condensation bottles
and that all hoses are connected to the hose connector.
Due to the modified safety concept (i.e. no condensation bottles) this message is no longer valid and can be
ignored (i.e. confirmed with yes).

Rotor16SOLV
The new Rotor16SOLV is not fully supported in software v1.51, i.e. only 8 of 16 vessels will be shown in the IR-
temperature diagram and run-data.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 11 of 23
Product Information

Version v1.40
IR Limit – MF100 changed to 160°C
The max. IR Limit of the MF100 vessel is reduced from 180°C to 160°C.
All stored and programmed methods will be updated automatically. This updated value is more accurate (heat
transmission between inside and outside of the vessel) and will improve the life time of the jackets especially if
Rotor 16MF100 is used without pT-Sensor.

Operation Time Limit for MF100 vessel


To ensure safe operation of the MF100 vessel and accounting for the material properties we have limited the
maximum operation time to 30min (cooling time not included).

New Application – O2 Combustion included


Anton Paar extends the versatility and flexibility of its Multiwave 3000 Microwave Sample Preparation System by
introducing accessories for Microwave-induced Oxygen Combustion (MOC)
Software v1.40 now fully supports this kind of application for Rotor 8XQ80, "O2-Combustion" can be additionally
selected instead of T-Program or P-Program.
The power and time settings can be reduced by the customer but not exceeded, a ramp is not programmable.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 12 of 23
Product Information

Security Cooling enhanced


Customer Settings:
The customer can now change the temperature limits for the automatic stop of the security cooling to a lower
value.
Possible range for the temperature limits:
Tinside: 60°C to 40°C
TIR: 50°C to 40°C
Additionally the time setting for the security cooling can be increased to a maximum of "99min 59s".

Solvent Runs:
Due to a volatility of solvents even at low temperatures (compared to acids) the temperature limits for the
security cooling after a solvent application have been changed to the following fixed values (not influenced by
customer settings):
Tinside: 40°C
TIR: 40°C
The default setting for the security cooling is intensity "2" for reasons of noise reduction.

Fan 0 only during first program step


For some special methods it might be necessary to use fan intensity "0" during the first step of the program.
Therefore the software allows to use this low intensity during the first step of the program but will automatically
adjust the fan intensity to "1" for the steps 2 – 4.
If you have equipped the instrument with a Safety Module SOLV, fan 1 will be used even for step 1. Otherwise
the safety module will report an error message.

Power Defaults during program start


To avoid unsteady power consumption during the first few seconds of the run, while the magnetrons are still
cold, the power output is limited to 1200W during the first 30 seconds of an "Power Program" even if the set
point is 1400W.
Additionally the default setting for "Temperature Programs" is 1200W, but this value can be changed by the
customer.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 13 of 23
Product Information

Start Sequence enhanced


The software v1.40 will not continue with the start procedure if an activated sensor does not respond during the
system check (possible reasons can be a failure of the sensor or the sensor is not installed). The customer has
to deactivate the sensor in the setup program if the method should be used without sensor or has to check the
sensor.
The p-Sensor of the Rotor 8 is not affected by this function as this Rotor can't be used without pressure control.
If you try to start a run with a defective p-Sensor the software will pop up the error message "Invalid Rotor Type".

Switch Off Limits


Reaching the following limits the software will automatically interrupt the run:

Control Parameter: Switch Off Condition:


(max. Limit +…)

Pressure: +1 bar

Pressure rate: +0.1 bar/s

Temperature: +5 °C

IR-Temperature: +5 °C

Remote Control via Terminal Program


The main key functions of the Multiwave 3000 can be remote controlled via a serial link cable and a terminal
program.
Following keys are supported:
<Stop>, <Start>, <DEL>, <Backspace>, <ESC>, <Tab>, <Enter>, <Up> (cursor up), <Down> (cursor down),
<Left> (cursor left), <Right> (cursor right), <INS>, <Home>, <End>, <Page Up>, <Paged Down>, <Fn> (function
keys F1 to F12)
Letters can be entered as decimal code (A  65, B  66, C  67…)
Port command: "key.press nnn"
e.g.:
key.press stop  for the <Stop> key
key.press pageup  for Page-Up
key.press 65  for letter "A"

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 14 of 23
Product Information

Safety Surveillance System


This software function checks important parameters prior to the start of a run as well as during the run and will
interrupt (stop) the run immediately if the limit of one of these parameters is exceeded.
These limits are:
General for all Rotor Types:
Start Temperature (T-Probe, pT-Sensor) > 9°C
The start temperature has to be higher than 9°C otherwise the instrument will not start the run but report an error
message. The customer has to check for any signs of damages and correct calibration.

Inner Temperature < 50% of the IR Temperature


This function is intended to avoid that either a run is done with a defective sensor (prior to run) or continued after
a breakage of the capillary. With this function the software will help to avoid overheating of the vessel in case of
using a defective sensor.
If the inside temperature does not increase during the first minutes of the run, the software will interrupt the run
immediately as soon as the IR temperature has reached twice the inner temperature.
During the run, the software will automatically stop the run if the inside temperature drops down to a value less
than 50% of the IR temperature.

Inside temperature increase (instruments without IR Sensor only)


When using instruments without IR Sensor the software will automatically check the temperature increase. If the
temperature is not higher than 35°C within the first 5 min the run will be interrupted.
This software function will avoid that the run can be started with an already broken capillary but will not react if
the capillary breaks during the run!

Especially for Rotor 8:


Starting Pressure Surveillance
During the start routines the software will check if the starting pressure of the rotor is between 2 and 10 bar. If
the pressure is outside of this specification, the software will interrupt the starting procedure showing a pop-up
error message. The customer has to check if the rotor upper plate is screwed down properly.
Please note that for O2 – combustions the permitted starting pressure has to be within the range of 2 to 25 bar
so that the vessels can be preloaded with oxygen.

Pressure dropdown during run


If the pressure of the rotor 8 drops down below 2 bar during the run the software will immediately stop the run as
well as the exhaust unit and magnetron fan. An error message will be displayed with a warning of a possible fire
inside the oven. This function will cover cases where the hydraulic system is affected by a melting or fire. By
switching off the fans the oxygen supply is stopped.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 15 of 23
Product Information

Print Out – Additional Information Included


The graphic print out (review protocols) now includes two additional lines with additional information about the
maximum temperature of the two magnetrons and the sensor (pT-Sensor, R8):

The temperature limit for the Toil/Tgas is 90°C and for Tmagn1/Tmagn2 it is 150°C.
The "Status Word: nnnnnnnn" is for special service cases only.

Minor Changes
Serial Port Setup (Text Compression)
The default parameter for "Text Compression" is now "high".

Numeric Screen during run


With version v1.40 you will have the "p-rate limit" displayed within the "Numeric" window during the run. The
"Date/Time" information has been removed within this window.

Serial Number displayed in Info Screen


The instrument serial number can be seen in the "Info" menu of the software. This facilitates access to the serial
number.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 16 of 23
Product Information

Version v1.27
Exhaust Unit Type EU2.0 fully supported:
The new exhaust unit ("EU2.0") needs a different fan-level adjustment than the old type. Therefore a new entry
in the service menu (Service => "Exhaust Unit Installation") has been created.
For optimal cooling efficiency, it is important to select the correct exhaust unit type. For maintenance purposes
you can check the runtime of each cooling unit. Refer to the instruction manual => software (B83IB01-A or
higher)

IR Sensor Calibration/Check:
To improve the accuracy of the IR temperature measurement, the new software includes an "IR Sensor
Calibration" function as well as an "IR Sensor Check" function.
This functions require an upgraded "Temperature Calibration Accessory (T/IR)” (Part No. 16486)
Caution: Running an IR sensor calibration/check without this special accessory will result in a misadjusted IR
measurement and may result in damage to the vessels and or rotor parts.
Before performing the calibration of the IR-sensor read the instruction manual => software (B83IB01-A or higher)

Changed vessel temperature limits:

vessel type: interior temperature limit: IR-temperature limit:

XQ 280°C 280°C

Method administration:
You can download the methods file to an external computer and upload it to any other Multiwave 3000.
Existing "Paar"-Methods will now be deleted prior to installing the new method file. Updated procedures will not
cause double entries anymore. User methods will remain on the instrument.

Method File Version 2.00


Together with version 1.27 Anton Paar released the new method file version 2.00.
Major changes to version 1.00, released together with software version 1.26 are:
 “Performance” methods are renamed to “PQ” (Performance Qualification) and adapted for the new
exhaust unit. New PQ’s for new rotors and options are available, too!
 Microwave extraction standard procedures, to be used with rotor 8SOLV (e.g. EPA 3546), have been
added.

Power Check control function improved:


The control function of the power check menu was improved to avoid the "Power Limit" error message.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 17 of 23
Product Information

Version v1.26.1
Bug-fix position recognition with Rotor 8SOLV:
With the v1.26 on some instruments the position recognition failed when using a Rotor 8SOLV.
The v1.26.1 includes an improved firmware function to avoid this problem.

Version v1.26
Safety Module SOLV
The software automatically recognizes the installation of the safety module for solvent extraction applications
and includes the new rotor types. The software also supports the optional Solvent Sensor Accessory. This
creates a new menu: Service\Solvent Sensor Settings

New Rotors
The rotor list has been extended to include the following rotors:

Rotor type: Interior temperature limit: IR-temperature limit: interior pressure limit:

SOLV MF 190°C 180°C 18 bar

SOLV HF 240°C 210°C 18 bar

EVAP MF n/a 180°C n/a

EVAP HF n/a 210°C n/a

New Printer Driver


The software supports the following printers:
- Epson Dot Matrix Printer 9pin or 24pin (or compatible)
- HP DeskJet (or compatible)
- HP LaserJet (or compatible)

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 18 of 23
Product Information

p-Sensor
The software supports the new Pressure Sensor Accessory, which measures the pressure in vessel 1 of the
16-position rotors. As soon as the sensor is installed, it is recognized and activated. To improve management of
the individual sensors, the „Sensor Menu“ was redesigned: Setup\Sensor:
 p-Sensor
 p/T-Sensor
 T-Probe
 IR-Sensor (Vessel Temp.)

Faster data transfer


In menu: Setup\Serial Port Setup\Text Compression there are 3 options for data compression:

None: Data compression as previously provided


Medium: 50 % data compression (the data transfer will be twice the speed as with “None”)
High: 80 % data compression (the data transfer will be 5 times the speed as with “None”)

Please note:
When the data have been compressed, it is no longer possible to view them with a text editor but conversion into
Excel (or other) file formats is not restricted.

Various Improvements to the User Interface


- New options (e.g. searching for methods by entering the first letter of the name)
- Automatic sensor recognition
- Review report (e.g. display of end times, firmware versions, etc.)
- Display of IR measuring values (empty spaces are faded out)
- IR control: improved control algorithm
- p-limit can be reduced (p-rate remains unchanged)
- More detailed error descriptions
- Improved hardware monitoring

Method Management
A memory optimization allows you to store up to 300 additional user methods. The method library is given a
version number so you can check the current status of the methods. Methods can also be faded out or
downloaded to a PC as an encrypted file.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 19 of 23
Product Information
Method File Version 1.00
Together with version 1.26 Anton Paar released the new method file version 1.00.
Major changes to previous versions are:
 The file has now a version number, starting with 1.00 showing the status of the methods on the
instrument.
 The file is encrypted, therefore it can not be used for older software versions. To use the latest version
an update at least to version 1.26 is necessary.

Control parameters extended:


Following a listing of all control parameters used during a run:

Indicator: Function: Description:

p Pressure Control The pressure limit was reached

T Temperature Control The temperature limit was reached

I IR-Temperature Control The IR-temperature limit was reached

r Pressure Rate Limitation The microwave power was reduced due to high pressure rates

O Over-range The pressure or temperature or IR-temperature limit was exceeded


and microwave power was switched off

R High Pressure Rate Alarm A pressure rate higher than 0,8bar/s was detected by the pressure
sensor and caused the microwave power to be switched off for at
least 3 sec

U User Break The user pressed the STOP key or opened the oven

S Power Supply Limitation The microwave power was reduced due to high current
consumption of the Multiwave 3000

M Magnetron Over-temperature The microwave power was reduced due to an over-temperature of


the Magnetrons.

o Sensor Over-temperature The microwave power was reduced due to an over-temperature of


the pressure or the temperature sensor (e.g. p-Sensor, pT-Sensor,
T-Probe).

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 20 of 23
Product Information

Version v1.24.1
Rotor signal readout adjusted to Rotor 8XF/XQ:
The software now is able to work with both types of Rotor 8 (XQ80 and XF100).

Hint: You have to install software version v1.24.1 to work with Rotor 8XQ/XF

Version v1.24
Changed vessel temperature limits:

Vessel type: Interior temperature limit: IR-temperature limit:

MF100 190°C 180°C

HF100 240°C 210°C

EF100 190°C 180°C

HEAT 190°C 180°C

Security Cooling:
If any cooling phase is programmed shorter than the security cooling setting the system will automatically add
the needed time for proper cooling with the current fan level.
If no cooling phase (cooling phase means Power = 0W and Fan = stage 2 or higher) is programmed the system
will add a security cooling based on the current security cooling setting.
With activated sensors the security cooling will stop automatically when following limits are reached:
Interior temperature lower than 60°C AND exterior (IR) temperature lower than 50°C
(If the limits are not reached after the costumer programmed cooling phase the system will add the security
cooling automatically)

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 21 of 23
Product Information

Exterior (IR) Temperature Controller:


New parameters for the controller (set @ Anton Paar factory):
P: = 20
I: = 0.001
If the controller overshoots the controller will be reset.
IR-Temperature limit for "OVERRANGE" set to 3°C to avoid overshoot.

Pulsed cooling mode (@Anton Paar factory):


Under "Advanced Service Routines" / "Fan Level Settings" the fan stage "0" can be operated in pulse mode.
(pulse wide in percent starting with 10% up to 100%)
That means:

Pulse [%]: Time ON: Time OFF:

10 1 second(s) 9 second(s)

20 2 second(s) 8 second(s)

30 3 second(s) 7 second(s)

40 4 second(s) 6 second(s)

50 5 second(s) 5 second(s)

60 6 second(s) 4 second(s)

70 7 second(s) 3 second(s)

80 8 second(s) 2 second(s)

90 9 second(s) 1 second(s)

100 (default) continuously ON no pulsing

Alteration corresponding to Power Board firmware V11:


- Power Board firmware version can be verified easily through the user interface software
- The Power Board reset will be performed after the CPU initialization
- During the CPU initialization the fan increase will be set to 100/s
Menu sample:
If there is any entry (under notes) present the "Notes" field on the display flashes.

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 22 of 23
Product Information

Configuration of MULTIWAVE 3000 with software v1.24:


Control Parameter Settings (set @ Anton Paar factory):
Pressure Control: P: 5 I: 0.002 D:
Temperature Control: P: 60 I: 0.003 D:
IR-Temperature Control: P: 20 I: 0.001 D:

Method file upload:


See upload instruction B83V050
Method file:
B83IA10A_Application Library.txt
IR-Sensor settings:
Emissivity: 0.480
Transmission: 1.000
Peak Hold Time: 0.0

Version v1.23
Description of the first delivered version: see other authentic document

Anton Paar High-precision Instruments Tel.: +43 316 257-0 B83IE025EN-J


www.anton-paar.com E-mail: info@anton-paar.com Page 23 of 23