Professional Documents
Culture Documents
Hafizal ICNX09 Full PDF
Hafizal ICNX09 Full PDF
Sahrim Ahmad a, A. Aziz Mohamed b, Jaafar Abdullah b, Hafizal Yazid a, *, M. Dahlan b, Rozaidi Rasid a,
* Corresponding Author: Hafizal Yazid, Faculty of Applied Science and Technology, Universiti
Abstract: X-ray microtomography and ImageJ 1.39u is used as a tool to quantify volume percentage of
B4C as fillers in thermoplastic-natural rubber blend composites. The use of percentage of area occupied
by fillers as obtain from ImageJ from the microtomography sliced images enables the proposed technique
to easily obtain the amount volume percentage of B4C in the composite non-destructively. Comparison
with other technique such as density measurement and chemical analysis proves the proposed technique
Carbide, Composites
1. Introduction
Filler distribution in composite is one of the critical issues once the composite is fabricated. The
information is very important since the resulted properties of the composite are affected by the spatial
distribution of the filler or particles [1-3]. This is true especially in the case of particulate composite.
Therefore a systematic approach is used to characterize the filler distribution in polymer matrix
composite. The approach is to detect the present of the filler or particles and to quantify the particle
distribution through the use of X-ray microtomography and other techniques. Similar approach has been
developed and used to quantify the particle distribution in metal matrix composite with promising results
[4]. Another approach to characterize the filler distribution in thermoplastic is through the use of X-ray
control and testing. Thus brings the proposed technique as an attractive method as it is non-destructive in
nature. The non-intrusive nature of the measurement makes it possible for repetition and as a volume is
mapped there is no need for statistical approximation to the final population size or distribution. It is
apparent that tomography is useful in many fields and much work has been done and reported in materials
applications [6-10].
composites is achieved through micrograph. The micrographs are captured using microscope coupled
with image analysis software. The captured micrographs are taken from slices of prepared sample. The
obtained information is only in 2D and extended to 3D by extrapolation and assumption. One typical
assumption is to treat the particle as point form or having a circular particle shape. These assumptions
may lead to inaccuracy or totally wrong information. Furthermore, sample preparation to turn it into
sliced form is necessary and this may damage the sample. The damage is possibly brought by torn-out
particle or smeared boundaries that result in misrepresentation of size or population [4]. Therefore the
non-intrusive method is required as the obtained data is more accurate and reliable.
volumes by reconstruction from multiple X-ray projections, allowing the direct characterization of the 3D
microstructure of samples [11]. The X-ray images are taken around a single axis of rotation and currently
able to resolve details as small as few microns in size, even when imaging objects are made of high
density materials. As long as the density different between reinforcement and the matrix is comparatively
significant, the resulted image would yield a good image contrast. This is the common criterion that has to
produce a boronated thermoplastic-rubber used as neutron shielding material in nuclear research reactor.
The excellent “ in between property ” of elastic and rigid of thermoplastic natural rubber make it an
attractive materials to be used as the matrix [12] and it could provide wide area coverage that require
radiation shielding. High thermal neutron cross-section material such as B4C is used as filler in order to
provide shielding effect against thermal neutrons [13-15]. The aim of the current work is ultimately to
quantify the distribution of boron carbide in the thermoplastic-natural rubber with the use of X-ray
microtomography and other techniques such as density measurement and thermogravimetry analysis. All
the techniques are compared and used to acquire information pertaining to the filler distribution and
2. Methodology
Thermoplastic-natural rubber used in this study was incorporated with various percentage of
boron carbide with average size of 13.5μm. Boron carbide with natural rubber, thermoplastic and other
additives were compounded in Haake twin blade internal mixer for a predetermined optimum time and
mixing scheme. Finally the composite compound was heated press to form a slab sample with 1mm in
thickness. The sample is depicted in Fig. 1 as observed from the top view of the slab.
1, VOI 2, VOI 3, VOI 4 and VOI 5. The size of each VOI is 5mm in diameter and 1mm in thickness. The
location of VOI on the slab are shown in Figure 1 and numbered accordingly. Each of the VOI is the test
A systematic approach is used to obtain information on reinforcement within the sample. The
samples were subjected to density measurement, X-ray microtomography scanning and chemical analysis.
Archimedes principle is used to obtain the density of each small cut section that contain VOI from which
the volume content of B4C could be calculated, as the value of standard blank sample is known. VOI
samples were subjected to Thermogravimetry analysis (TGA) to quantitatively obtain the weight percent
of boron carbide in the test samples. X-ray microtomography measurements were carried out on VOI
using a Sky Scan 1172 desktop x-ray microscanning. Data were collected at 60kV and 167 A. The image
was detected on a high resolution (1280 x 1024 pixels) CCD camera as the sample was rotated and the
output from the CCD was fed through a reconstruction algorithm [9]. Beam hardening was also
computationally corrected, to produce a binary grey-scale bitmap image of the sample. The datasets
collected for the tomography were 360 angular range with a step size of 0.7 between images. The sliced
binary images are thresholded to obtain the areas occupied by B 4C. These values are used with results
from chemical analysis to establish a parameter named correlation factor of B 4C in the composites.
Chemical analysis gives an accurate value of percentage of filler within the tested volume of
interest. In this case, TGA is used to obtain the amount of filler in the composite sample. The drawback of
this technique is that it is destructive in nature. This technique gives the overall vol% of B 4C but the
location of the particle within the VOI is not obtained. Result of TGA is given in Figure 2.
Figure 2: Average mass % of B4C in VOI’s for each of the slab
TGA thermograms reveal the actual concentration of B 4C in the slabs. The amount of B4C was
found slightly lower than the target composition as the filler lost is common during mixing process. This
is due to the filler adhesion at the surface of hopper or feeding system. From Figure 2, as expected slab A
was absent of filler. This is followed by slab B to F with a gradual increment of filler or B 4C. However,
this technique does not provide any information on the filler distribution in the sample and exhibit only
information on filler concentration. Therefore this technique is coupled with the X-ray microtomography
to utilise both the information on the particle distribution and the actual filler concentration.
X-ray microtomography reveals the 3D volume microstructure of the sample. The density
difference between the filler and the matrix enable the filler to appear darker than the matrix. Therefore
particle distribution could be observed within the slices of VOI. Typical image reconstruction of 3D
Typical horizontal X-ray microtomography slices of VOI’s before and after thresholded is shown in
Figure 5: (a) Typical slice of matrix + 10%wt filler, (b) thresholded of image (a) and typically yield
(a) (b)
Figure 6: (a) Typical slice of matrix + 20%wt filler, (b) thresholded of image (a) and typically yield
Figure 7: (a) Typical slice of matrix + 30%wt filler, (b) thresholded of image (a) and typically yield
(a) (b)
Figure 8: (a) Typical slice of matrix + 40%wt filler, (b) thresholded of image (a) and typically yield
Figure 9: (a) Typical slice of matrix + 50%wt filler, (b) thresholded of image (a) and typically yield
The microtomography slices (Figures 5b-9b) in the thresholded images show a variation in the
percentage of bright areas. The bright areas indicate the present or occupancy of filler. This is possible
since the density difference between the matrix and the filler is high. The matrix density is 0.920g/cm3
and the filler is 2.520g/cm3. Thus, this has eased in the thresholding process to obtain the areas that reflect
the present of B4C. It was found that 8.514%wt by chemical analysis corresponds to 2.515% bright areas
in the thresholded image as typically shown in Figure 5b. Other thresholded images give 5.331%,
8.297%, 10.640%, 13.402% bright areas and correspond to 18.4941%wt, 27.8368%wt, 37.6574%wt,
A correlation factor is established to enable the filler in the sliced image being interpreted
quantitatively. This is achieved by having an average value of area occupied by the filler in the sliced
image. In this work, the sliced image is analysed from top to bottom of the VOI with the condition of in
between 2 analysed image, there would be 9 sliced images that are abandoned from analysis to avoid
misleading interpretation. The misleading is from the filler that appeared in more than 1 slice. Thus, the
percentage of area occupied by the filler is compared against the result of chemical analysis. Both results
The existence of the correlation factor is clearly observed from the graph of average volume percentage of
fillers from chemical analysis against average area percentage of fillers in the sliced image. This is shown
in Figure 10.
Figure 10: Volume percentage of fillers from chemical analysis against average area percentage of fillers
Based on Figure 10, the correlation factor is significantly the slope of the resulted graph. Based on the
graph, the correlation factor was found to be approximately 1.3155. Thus, equation (1) reduced to:
Equation (2) could be used to estimate the volume percentage of B4C for any given sample that was
subjected to X-ray microtomography inspection providing any further work done using the same type of
system.
Density measurements lead to the determination of amount of B4C in the sample as the density
of each of the ingredient is known. Average densities of small size cut sections that contain VOI are
obtained for each slab and these results are compared between the slabs. This comparison is shown in
Figure 11.
From Figure 11, slab A shows the lowest value of density with 0.920g/cm3 due to the absent of filler. As
the filler content is increased from slab B to F, the density also increased accordingly. The maximum
Results of chemical analysis, corrected X-ray microtomography, Archimedes method and X-ray
Overall volume percentage of B4C is given by 2 sets of values from chemical analysis and Archimedes
method. Chemical analysis gives an accurate amount vol.% B4C but Archimedes method suffer
discrepancies if samples contain porosity or inclusions. This is shown in Figure 12 for slab C, D and E
where the vol.% B4C value obtained from density result is significantly lower than that obtained from
chemical analysis. This is due to changes in density not only being affected by fillers but also by porosity
Porosity
Although chemical analysis or specifically TGA is good in term of result accuracy and not affected by
porosity, it is still considered as a time consuming analysis and destructive method. However, this
technique could be used as a reference with the microtomography result in order for us to establish the
correlation factor. This is true providing the relationship between the vol.% B 4C by chemical analysis and
area% B4C by microtomography is valid. As observed from Figure 12, microtomography results itself
does not show the actual filler concentration. Therefore, the introduction of correlation factor is a must
and in this work, the value was found 1.315. Once this value multiplied by area% of fillers obtained by
microtomography, this new value or known as corrected x-ray microtomography is found close or almost
identical to the actual filler concentration by chemical analysis as shown in Figure 12. Thus,
quantification of B4C in TPNR composite is possible by the use of correlation factor or corrected X-ray
microtomography method. This method provide a simple mean to find the volume percent of fillers
present providing the result of area% of fillers by microtomography is available and the work done using
4. Conclusions
The work could be considered as an evidence of the applicability of the X-ray microtomography
to quantitatively obtain the concentration of B4C in the TPNR composites. Results of X-ray
microtomography were successfully calibrated using thermogravimetry analysis or TGA. In this work, the
correlation factor was found to be 1.315. This value once multiplied with the area percentage of fillers in
the sliced image would yield the concentration of fillers or B 4C in the sample. This technique enables the
X-ray microtomography results to reveal the composition of B 4C in the sample in a quantitative manner
with the existing capability exhibited the location or distribution of the B 4C. Further research will be
focused on to establish a correlation factor for different types of filler that are typically used in the
polymer composites and to obtain a set of correlation factor data for use in X-ray microtomography
analysis.
Acknowledgements
The authors wish to thank Hishamuddin Husain for proofreading the paper. The excellent support from
References
[1] T. P. Selvin, J. Kuruvilla and T. Sabu, Mechanical properties of titanium dioxide-filled polystyrene
[3] Laura M. McGrath et al, Investigation of the thermal, mechanical, and fracture properties of alumina-
quantify the distribution of TiB2 particulate in aluminium, Scripta Materialia 46 (2002) pp. 25-29.
[5] Dongyang Wu et al, X-ray ultramicroscopy: A new method for observation and measurement of filler
dispersion in thermoplastic composites, Composite Science and Technology 68 (2008), pp. 178-185.
[6] Mummery PM, Derby B, Anderson P, Davis GR, Elliott JC, J. Microsc 177 (1995), pp. 399-406.
[9] E. Cornelis, A. Kottarand and H. P. Degischer, X-ray computed tomography characterizing carbon
[10] Green WH, Sincebaugh P, Evaluation of complex composite structures using advanced computed
tomography (CT) imaging, In: 53rd meeting of the society for machinery failure prevention technology,
[11] J. Martin –Herrero and Ch. Germain, Microstructure reconstruction of fibrous C/C composites from
[12] A. Ibrahim and M. Dahlan, Thermoplastic natural rubber blends, Prog. Polym. Sci. 23 (1998), pp.
665-706.
[14] J. K. Shultis, R. E. Faw, Radiation Shielding, Prentice-Hall, Englewood Cliffs, NJ, (1996).
[15] B. T. Price, C. C. Horten, K. T. Spinnery, Radiation Shielding, Pergamon, Elmsford, NY, (1957).
Biography
Hafizal Yazid
Hafizal received his first degree in Materials Engineering from University of Malaya, Kuala Lumpur,
Malaysia. He starts his work as a metallurgical engineer and currently working in Industrial Technology
Division, Malaysian Nuclear Agency, Bangi, Kajang Malaysia. His work relates to the application of
Radiography technique in the industry and carry out research pertaining to radiography image
enhancement. His work also covers development and characterization of materials used as radiation
1. Yazid H et al, Circular discontinuities detection in welded joints using Circular Hough
Sahrim Ahmad
Sahrim Ahmad received his doctorate degree from Loughborugh University of Technology, UK in the
area of Materials Engineering in 1988. He is currently a Profesor in Universiti Kebangsaan Malaysia and
Abdul Aziz Mohamed gained his master degree from Surrey University, United Kingdom in 1980. He
then received his doctorate degree from Cranfield University in the area of advanced materials on
Jaafar Abdullah
Jaafar Abdullah is currently a senior research officer in Malaysian Nuclear Agency. His work involves in
M. Dahlan
M. Dahlan is currently a manager and senior research officer in Malaysian Nuclear Agency. His work
Rozaidi Rasid
Rozaidi Rasid is currently a senior lecturer in Universiti Kebangsaan Malaysia. His work involves in
W. Saffiey W. Abdullah
W. Saffiey W. Abdullah is currently a senior research officer in Malaysian Nuclear Agency. His work
Mahathir Mohamad
Mahathir Mohamad is a research officer in Malaysian Nuclear Agency. His work involves in research and
Rafhayudi Jamro
Rafhayudi Jamro is a research officer in Malaysian Nuclear Agency. His work involves in research and
M. Hamzah Harun
M. Hamzah Harun is a research officer in Malaysian Nuclear Agency. His work involves in research and
Mouad A. T.
Mouad A. T. is currently a graduate student in Universiti Kebangsaan Malaysia. His work involves in
Ida Rosnita is currently a graduate student in Universiti Kebangsaan Malaysia. Her work involves in