Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 8

Troubleshooting of DM plant

                                                                                                                                          
                Go to Top

Defects Causes Remedies


Decrease in capacity a. Increase in ionic load Check by analysis
between two successive
regenerations b. Flow recorder defective Check

c. Insufficient chemicals used Check

d. Resin dirty Give prolonged backwash

e. Plant being used intermittently Avoid this

f. Channelling in bed Check and ensure uniform distribution


/collection
                                                                       
g. Resin fouled If cation, give HCl wash; if anion, resin
give alkaline brine treatment
                                                                        
h. Resin deteriorated Check and replace charge

i. Resin quantity insufficient in unit Check and top up

   
Treated quality not upto the a. Cation exhausted Check
standard
b. Anion exhausted Check

c. Mixed bed exhausted Check

d. MB resin not in uniform mixed state Repeat air mix and rinse

                                                       Chec
e. Some valves like backwash leaking k

f. Na slip from cation high Check raw water analysis; change in Na/TA
and SiO2/TA ratio; use more chemicals

Check raw water analysis; change in Na/TA


g. SiO2 slip from anion high and SiO2/TA ratio; use more chemicals

Check

h. Unit idle Rinse to satisfactory quality

i. Unit not sufficiently rinsed Adjust to between unit min/max flow rate

j. Excessive/low flow rate Check and ensure uniform


collection/distribution

k. Channelling Check resin and give alkaline brine/ HCL


treatment

l. Resin fouled Check resin and replace

                                                                      
m. Resin deteriorated
Mixed bed quality not good a. Resin not separated during backwash properly Give extended backwash after exhausting
the bed

b. Air mix not proper Repeat

c. Final rinse not proper Repeat

d. Some valves may be leaking and Check and examine


contaminating the treated water
 
 
High residual CO2 from a. It can be due to choked suction filter of Check and clean filter
degasser degasser air blower

b. Improper air flow to the degasser Check damper, speed of blower, discharge
pressure

c. Degasser blower not in operation Check and operate blower

d. Air seal not fitted/broken resulting in short Check and replace fitting
circulating of air
 
 
Unit rinse takes long time a. Flow rate too high Increase flow rate

b. Unit exhausted Regenerate unit

c. Backwash valve passing Check and rectify

d. Anion resin organically fouled Give alkaline brine treatment

e. MB air mix not satisfactory Carry out air mix once again

f. Acid/alkali pockets formed in unit Faulty design check and rectify.


Temporarily backwash (followed by air
  scour if MB) and rinse again

 
Flow rate too less a. Choked valve and suction strainer of pump Check

b. Cavitation in the pump Check

c. Low inlet pressure Check-pump

d. Distribution or collecting system choked Check

e. Resin trap at outlet choked Check and clean

f. Control valve shut due to low off-take Increase off-take

   
Pressure drop across the a. Defective valves Check
bed increasing day by day
b. Packed resin bed and resin fines present Give extended backwash with open
manhole and scrap off fines from top
  surface of the resin

c. Collecting system choked Check, repeat backwash

d. Pressure gauge defective Check and rectify/ replace

   
Flooding in degasser a. Very high air flow rate Reduce air flow rate by adjusting damper

Open and check


b. Packed tower chocked due to dirt or broken
packing material  

Resin being lost a. Excessive backwash pressure Check inlet pressure and reduce if
necessary

b. Faulty collecting system Examine same for breakage

c. Inlet strainer damaged Check and replace

   
Ejector not working a. Low power water pressure Check

b. Air lock in the unit Backwash & open air release

c. Choked or defective valves Examine and rectify

d. Ejector nozzle may be choked Check

e. Too much back pressure from the unit Check for chokage of collecting system;
passage of inlet/outlet valves

f. Bulge in pipe lining Check and rectify

   
Incorrect reading rota- a. Chocked orifice lines/orifice Check and clean
meters
b. Dirty glass and float check and clean

 
Improper reading from flow a. Choked impulse lines/orifice Check and clean
recorder integrator
b. DP transmitter requires recalibration Recalibrate

c. Leakage in signal tube between transmitter and Check


panel
                                               Check
d. Low air pressure for DP transmitter or recorder
 
 
Level electrodes system for a. Improper contact between electrodes and Check contact and rectify
measuring and dilution tank control
not functioning properly cabling                                                      Check
and dry the contacts of moisture and dirt
b. Shorting of the two electrodes due to moisture
or Check
any foreign material
 
c. Improper working of the level controllers

 
Leakage from acid a. Improper adjustment of the mechanical seal Check and adjust
injection/unloading/transfe
r pumps b. Low strength of sulphuric and presence of Check concentration and take appropriate
ferrous sulphate action

   
Corrosion in concentrated a. Low concentration of sulphuric acid Check silica gel breather in acid storage
acid tanks and lines tank and replace silica gel charge if
  exhausted

b. Lining of HCl tank/pipe line damaged Rectify

   
Improper opening and a. Defective solenoid valves Check
closing of pneumatically
operated valves b. Leakage in airline from solenoid valve to the Check
respective control valve.
                                               Check
c. Improper contact of micro switch giving false
indication to panel                                                Check

d. Fused mimic lamp giving false indication to the  


panel

 
Improper operation of a a. Defective relays in the control circuit Check and replace relays
certain regeneration cycle
Solid state programme not a. The controller can be kept in "hold" due to the Remove conditions which cause "hold" of
functioning properly reasons explained under operation controller

b. Improper operation of the controls for the Press test switch & check the complete
controller cycle

c. Defect in the inside of the controller Check the instruments thoroughly from
inside. Meanwhile, operation may be
  continued by using bypass toggle switches

You might also like