The Required Mechanical Properties

of Hip and Knee Components

S. M. Kurtz, Ph.D.

Drexel University and Exponent, Inc., Philadelphia PA, USA

Correspondence addresses:
Implant Research Center, School of Biomedical Engineering, Science and health Systems
Drexel University and Exponent, Inc., 2300Chestnut St., Suite 150, Philadelphia PA, 19103,
USA
skurtz@exponent.com
Introduction
Voluntary national and international standards have been relied on by the orthopedic community
as guidelines for the mechanical properties of medical grade UHMWPE. The two most widely used
standards for medical grade UHMWPE, ASTM F-648 and ISO 5834, include specifications for the
properties of the unconsolidated resin powder, as well as the properties of the consolidated stock
material. For many years now, both the ASTM and ISO standards have been developed by the same
the industrial participants and researchers. Thus, the ASTM and ISO standards are considered to be
harmonized and consequently reflect a unified and international perspective on the properties of
medical grade UHMWPE. The goal of this paper is to first review the material property guidelines
for medical grade UHMWPE, and to describe how these properties may be altered by radiation
crosslinking and thermal processing.

Material Property Guidelines for UHMWPE

The material property limits for medical grade UHMWPE are essentially the same in both
ASTM F648 and ISO 5834 (Table 1), with the exception of different types of impact testing being
specified in ASTM F648 and ISO 5434. Currently Ticona (Oberhausen, Germany) produces a Type
1 and Type 2 resin with the trade names of GUR 1020 and 1050, respectively. Prior to 2002, Basell
Polyolefins (Wilmington, Delaware, USA) produced Type 3 resin with the 1900 trade name. This
resin was discontinued in January 2002 and is no longer produced. Two orthopedic manufacturers
have maintained large stockpiles of this resin, so orthopedic implants will continue to be fabricated
from this resin, at least in the near future.
Although these UHMWPE standards apply only to unirradiated stock material, some
investigators have also adopted these guidelines for radiation crosslinked and thermally treated
UHMWPE as well [1]. It should be noted, however, that none of the existing standards for medical
grade UHMWPE material properties were originally conceived for polymer that had been subjected
to sterilization, irradiation, or other processing steps.

52
Table 1. Specifications for Medical Grade UHMWPE Resins in ASTM F648-00

Type 1 Resin Type 2 Resin Type 3 Resin


Density (kg/m3) 927-944 927-944 927-944
Yield Stress (MPa), 21 19 19
minimum
Ultimate Tensile Stress 35 27 27
(MPa), minimum
Elongation to Failure (%), 300 300 250
minimum
Izod Impact Strength, 140 90 30
double notched (kJ/m2),
minimum

Tradeoffs in Material Properties with Irradiation and Thermal Treatment


Three important processing steps are necessary to produce highly crosslinked polyethylene for
hip and knee bearings. These steps are an irradiation step to promote crosslinking, an intra or post-
irradiation thermal processing step to increase the level of crosslinking and remove residual stress,
and a sterilization step. The properties of UHMWPE are influenced by processes of irradiation,
thermal processing, and sterilization.
In the irradiation step, gamma and electron beam radiation produce free radicals (unpaired
electrons) in the polyethylene, which in secondary chemical reactions leads to a combination of
crosslinking and chain scission. Crosslinking is beneficial for reducing wear. Chain scission
produces a decrease in molecular weight, with concomitant reduction of wear resistance and
mechanical properties. When irradiation is conducted in the presence of oxygen, scission
predominates over crosslinking. However, when conducted in an inert environment, such as
nitrogen, crosslinking predominates over scission. Regardless of whether irradiation is conducted in
air or in an inert environment, some of the free radicals will remain entrapped within the crystalline
phase of the UHMWPE. Over time, these entrapped free radicals can migrate to the surface of
crystals. If irradiation is done in air, these free radicals react with available oxygen causing further
time-dependent chemical degradation.

53
Increased crosslinking improves the wear performance of UHMWPE compared to conventional
material. However, the presence of the crosslinks adversely affects uniaxial ductility [2], and the
uniaxial failure strain of UHMWPE decreases linearly with increasing radiation dose [2]. During
irradiation, the loss of ductility depends on the crystalline microstructure of the UHMWPE, because
crosslinking occurs primarily in the amorphous phase, where the molecular chains are in sufficient
proximity such that a covalent bond can be created between adjacent polymer molecules by the
applied energy [3]. Unirradiated UHMWPE typically has a crystallinity in the region of 50% [4], so
some 50% of the material is amorphous content that may be crosslinked during irradiation. If the
temperature of the UHMWPE changes during the crosslinking process, this can influence the
distribution of crosslinking in the polymer and, hence, influence its ability to accommodate large
strains prior to failure.
The first choice an implant designer has to make is the method of crosslinking (e.g., gamma vs.
electron beam). If irradiation is to be carried out using electron beam irradiation, the designer must
consider the additional factor of irradiation temperature, since the rate of energy dissipation
increases the temperature above the melting temperature. Of the six orthopedic manufacturers
currently producing highly crosslinked UHMWPE implants, two have chosen electron beam
irradiation, whereas the other four use gamma radiation crosslinking. In this review, we will restrict
our attention to gamma radiation crosslinking at room temperature, as it is the most widely used
crosslinking modality. For more information about the differences between electron beam and
gamma irradiation of UHMWPE, the reader is referred to a recent review [3].
In the production of a highly crosslinked UHMWPE, the material is subjected to a thermal
treatment step to reduce the level of free radicals via further crosslinking reactions. At higher
temperatures the polymer molecules have increased mobility, thereby increasing the probability of
free radicals on adjacent chains reacting to form crosslinks. For the thermal treatment to be
effective at eliminating all free radicals, it must be conducted at 150°C, above the melt temperature
of the material. Heating above the melting temperature destroys the crystalline regions of the
material thus making the free radicals that were in the crystals available for crosslinking. The
disadvantage of melting is the reduction crystal size and in material yield and the ultimate strength
that ensues. A compromise solution is to heat the material to just below the melting temperature.
This solution preserves the original crystal structure, retains mechanical properties, and makes more
free radicals available for crosslinking than would be available without thermal treatment while still
retaining some free radicals in the crystal domains. When thermal treatment is conducted below the

54
melt transition of 135oC, it is referred to as “annealing,” and above the melt transition, it is called
“remelting.” Typically, annealing is carried out at 130oC and does not eliminate all free radicals,
although the number is substantially reduced by the elevated temperature.
The choice of thermal treatment has a significant impact on the crystallinity and mechanical
properties of highly crosslinked UHMWPE [2]. At a dose of 100 kGy, the elastic modulus, yield
stress, and ultimate stress of a remelted material is significantly lower than the respective properties
for an annealed material (Figure 1, Table 2). Figure 1 compares the uniaxial tensile behavior of
unirradiated UHMWPE material with conventionally sterilized (30 kGy, in N2) polyethylene, and
with both annealed and remelted highly crosslinked polyethylenes (100 kGy).

Table 2. Effect of Post-irradiation thermal treatment on uniaxial mechanical properties [5]. Note that these
irradiation treatments were achieved with a single dose. Properties were determined from treated rods of GUR 1050.

Dose Heat Yield Stress Ultimate Stress Elongation to


(Gamma) Treatment (MPa) (MPa) Failure (%)
100 kGy None 23.2 ± 0.2 47.6 ± 2.0 238 ± 13
100 kGy 110°C Anneal 23.0 ± 0.3 47.3 ± 1.5 230 ± 12
100 kGy 130°C Anneal 22.6 ± 0.2 48.5 ± 1.5 231 ± 13
100 kGy 150°C Remelt 19.5 ± 0.3 43.9 ± 3.9 246 ± 12

55
Figure 1. Stress v. strain curves for conventional, nitrogen-sterilized UHMWPE (30 kGy, N2) and two forms of
highly crosslinked UHMWPE. The highly crosslinked materials were both irradiated with 100 kGy in a single dose and
either annealed (110°C) or remelted (150°C).
For the two highly crosslinked UHMWPEs shown in Figure 1, the annealed material has an average degree of
crystallinity of 60%, whereas the remelted material has a crystallinity of 43%.

Throughout the entire stress-strain curve, the higher crystallinity of the annealed material results
in a greater resistance to plastic deformation when compared to remelted material. Therefore, the
selection of post-irradiation thermal treatment is the second most important decision for an implant
designer, as it will influence not only the free radical content, but also the crystallinity, yield
strength, and ultimate tensile strength of the highly crosslinked polyethylene. These reduced
mechanical properties may not influence wear but will certainly influence the resistance of the
material to damage caused by impingement or bearing lift off.

Functional Fatigue Performance


Based on contemporary hip simulator studies, adhesive/abrasive wear is no longer expected to be
the primary limiting factor in the longevity of hip replacements incorporating highly crosslinked

56
UHMWPE. The improved wear resistance of highly crosslinked UHMWPE has recently prompted
some implant designers to consider thin metal-backed liner designs and larger femoral heads to
reduce the incidence of dislocation [6]. However, with the incorporation of highly crosslinked
UHMWPE into new large-diameter cup designs, other modes of clinical failure, such as component
fracture associated with rim loading and thin liners [7, 8], as well as impingement-related damage
due to component malpositioning [9], may become new limiting factors for the long-term clinical
performance of hip replacements. The clinical introduction of thin acetabular liners incorporating
highly crosslinked UHMWPE raises new questions regarding the ability of these thin liner designs
to withstand structural fatigue loading.
To address functional fatigue loading of acetabular liners, researchers have recently suggested
evaluating contemporary highly crosslinked UHMWPE materials in historical designs, such as the
ACS design [10, 11], which in some circumstances has shown evidence of rim loading and fracture
[8]. Walsh and colleagues [10], for example, found that remelting highly crosslinked UHMWPE
significantly increased the fatigue crack growth rate in the ACS design, as compared with highly
crosslinked cups that were not heat treated. Wang et al. [11] have evaluated conventional gamma
sterilized, remelted, and annealed highly crosslinked UHMWPE in functional fatigue simulations of
the ACS design. In contrast with the study by Walsh [10], the liners in Wang’s study were not
prenotched so that crack initiation (as opposed to propagation) could be evaluated. The radiation
crosslinked and remelted materials studied by Wang et al. [11] exhibited significantly increased risk
for structural fatigue failure, whereas the highly crosslinked and thermally annealed (below melt)
material showed no increased risk.

Conclusion
Increased radiation crosslinking, beyond the 25 to 40 kGy dose needed for sterilization, is now
recognized to substantially improve the wear resistance of UHMWPE. However, there are several
additional decisions to be made when developing a highly crosslinked UHMWPE for joint
replacement, and these decisions may strongly influence the mechanical properties of the bearing.
The effect of UHMWPE mechanical properties on clinical performance is not fully appreciated at
the present time. The proliferation of crosslinking technology into hip replacements, and more
recently knee replacements, has resulted in six new proprietary UHMWPEs, with trade names like
Crossfire™, Durasul™, Longevity™, and Marathon™. It will be many years until sufficient

57
evidence has been collected to test the hypothesis that these UHMWPEs—all with different material
properties—significantly reduce the incidence of revision in hip and knee replacement.

References
[1]. Muratoglu OK, Bragdon CR, O'Connor DO, Jasty M, Harris WH. A novel method
of cross-linking ultra-high-molecular-weight polyethylene to improve wear,
reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP
Paul Award. J Arthroplasty 2001; 16: 149-60.
[2]. Martell J, Verner JJ, Incavo SJ. Clinical performance of a highly crosslinked
polyethylene at two years in total hip arthroplasty: a randomized prospective trial.
Trans. American Association of Hip and Knee Surgeons 2002; 12: 24.
[3]. Muratoglu OK, Kurtz SM. Alternative bearing surfaces in hip replacement. In Hip
Replacement: Current Trends and Controversies. Eds. R. Sinha. New York:
Marcel Dekker, 2002.
[4]. Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing,
sterilization, and crosslinking of ultra- high molecular weight polyethylene for
total joint arthroplasty. Biomaterials 1999; 20: 1659-88.
[5]. Kurtz SM, Cooper C, Siskey R, Hubbard N. Effects of dose rate and thermal
treatment on the physical and mechanical properties of highly crosslinked
UHMWPE used in total joint replacements. Transactions of the 49th Orthopedic
Research Society 2003; 28.
[6]. Muratoglu OK, Bragdon CR, O'Connor D, Perinchief RS, Estok DM, 2nd, Jasty
M, Harris WH. Larger diameter femoral heads used in conjunction with a highly
cross- linked ultra-high molecular weight polyethylene: A new concept. J
Arthroplasty 2001; 16: 24-30.
[7]. Suh KT, Chang JW, Suh YH, Yoo CI. Catastrophic progression of the disassembly
of a modular acetabular component. J Arthroplasty 1998; 13: 950-2.
[8]. Bono JV, Sanford L, Toussaint JT. Severe polyethylene wear in total hip
arthroplasty. Observations from retrieved AML PLUS hip implants with an ACS
polyethylene liner. J Arthroplasty 1994; 9: 119-25.

58
[9]. Barrack RL, Schmalzried TP. Impingement and rim wear associated with early
osteolysis after a total hip replacement : a case report. J Bone Joint Surg Am 2002;
84-A: 1218-20.
[10]. Walsh HA, Furman BD, Li S. The effects of cross-linking on the fracture and
fatigue properties of UHMWPE acetabular cups. Transactions of the 25th Society
for Biomaterials 2001; 24: 592.
[11]. Wang A, Manley M, Serekian P. Wear and structural fatigue simulation of
crosslinked ultra-high molecular weight polyethylene for hip and knee bearing
applications. In Highly Crosslinked and Thermally Treated Ultra-High Molecular
Weight Polyethylene For Joint Replacements. Eds. S. M. Kurtz, R. Gsell and J.
Martell. West Conshohoken: American Society for Testing and Materials, 2003.

59