1.0 INTRODUCTION............................................................................................................................................. 2
1.1 General ........................................................................................................................................................ 2
2.0 CHANGES TO THE CMP 200® X-RAY GENERATOR SERVICE MANUAL 901476 ..................................... 3
2.1 Chapter 1: .................................................................................................................................................... 3
2.2 Chapter 2: .................................................................................................................................................... 3
2.3 X-ray Generator Service Manual Supplement, X-Ray Tube Stator Compatibility Tables 746026:............ 3
2.4 Chapter 8: .................................................................................................................................................... 5
2.5 Chapter 9: .................................................................................................................................................... 5

PROPRIETARY AND CONFIDENTIAL

The design and information contained in this document is proprietary to Communications & Power Industries
Canada Inc. (CPI), has been designed and developed at private expense, and is the exclusive property of CPI.
The information contained herein is loaned in confidence solely for the installation, operation, and maintenance of
CPI equipment. This document may not be duplicated in whole or in part, or be used for re-engineering, reverse
engineering, or otherwise reproducing in any form or creating or attempting to create or permitting, allowing or
assisting others to create or manufacture CPI’s product or products derived therefrom, without the express written
permission of CPI.

Copyright © 2011 by Communications & Power Industries Canada Inc. All rights reserved.

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Service Manual Supplement # 90383800 Rev. B Page 1


CPI Canada Inc. Supplement

1.0 INTRODUCTION

This supplement is not intended for CMP 200® and CMP 200® DR X-ray generators prior to serial
number CPD05500F09.

This supplement must be used in conjunction with the CMP 200® / CMP 200® DR service and
installation manual part number 901471.

The CMP 200® / CMP 200® DR service and installation manual part number 901471 covers the standard
CMP 200® / CMP 200® DR configurations and options. This supplement describes the manual changes
due to the chassis-mounted phase-shift capacitor that is no longer part of the H.V auxiliary board.

1.1 General

The Low Speed Starter (LSS) was integrated into the H.V auxiliary board. However, for this application,
the phase-shift capacitor for the LSS has been removed from the H.V auxiliary board and has been
mounted on the chassis (figure 2).

A revised H.V auxiliary board (figure 1), part numbers 739442-99 / 739445-99 and the chassis-mounted
phase-shift capacitor that is connected to E17 and E18 on the H.V auxiliary board via a cable makes the
LSS more configurable to be compatible with a larger number of X-ray tubes.

PHASE-SHIFT CAPACITORS HAVE


BEEN MOVED FROM THIS AREA

E17

Figure 1: H.V auxiliary board

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Service Manual Supplement # 90383800 Rev. B Page 2


CPI Canada Inc. Supplement

2.0 CHANGES TO THE CMP 200® X-RAY GENERATOR SERVICE MANUAL 901476

2.1 Chapter 1:
• Section 1.4.0:

FAN (INSIDE FAN COVER)


MAIN POWER
GROUND
(INSIDE FAN INVERTER ASSEMBLY
COVER) (2 USED ON 50/65/80 KW UNITS)

MAIN FUSES REAR OF


(INSIDE FAN CABINET
COVER)
(F1, F2: 1 PHASE
F1-F3: 3 PHASE)

MAIN POWER
CONTACTOR

MAINS RECTIFIER
ASSEMBLY

EMC CAPACITOR
BOARD
PHASE-SHIFT CAPACITOR
DC BUS
CAPACITORS
(UNDERNEATH
DC BUS
INVERTER(S))
CAPACITORS
AUXILIARY
TRANSFORMER
HT TANK

CMP200_026B

Figure 2: Major generator subassemblies view 1

2.2 Chapter 2:
• Section 2.7.0:

It is very important to note the value of the phase-shift capacitor before continuing. Refer to section 2.3 in
this supplement for the X-ray tube stator compatibility. If the value of the phase-shift capacitor is not
compatible with the desired X-ray tube stator type, the phase-shift capacitor must be changed to match
the requirements of the desired X-ray tube stator type. Please contact CPI product support for assistance.

2.3 X-ray Generator Service Manual Supplement, X-Ray Tube Stator Compatibility Tables 746026:
• Section 2:

Table 1 in this supplement provides a cross-reference between the H.V. auxiliary board part number
referenced in table 1 of supplement 746026 and the capacitor field kit part number.

The capacitor field kit part number determines the value of the phase-shift capacitor.

The capacitor field kit 903836-XX for the CMP 200® and CMP 200® DR X-ray generators includes the
specified chassis-mounted phase-shift capacitor (no longer part of the H.V auxiliary board) and its
mounting hardware. This allows the X-ray tube stator compatibility to be changed in the field.

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Service Manual Supplement # 90383800 Rev. B Page 3


CPI Canada Inc. Supplement

TABLE 1: Cross-Reference, X-Ray Tube Stator Compatibility Tables


(Supplement 746026) to Capacitor Field Kit Part Number
SUPPLEMENT 746026 CAPACITOR FIELD KIT

SHIFT CAP. HV AUX BRD PART # SHIFT CAP. PART NUMBER

30 μF 739442-00 30 μF 903836-00
739445-00
45 μF 739442-01 40 μF 903836-01
739445-01

Table 2 in this supplement replaces table 1 of supplement 746026, for the following Tube Types only.

TABLE 2 : TUBE TYPE ( CMP 200® / CMP 200® DR LOW SPEED STARTER )

TUBE TYPE TUBE TYPE SHIFT CAPACITOR FIELD KIT


(HOUSING) (INSERT) CAP. PART NUMBER
Kailong RADII KL76 20 μF 903836-02
H1076
(20/50 Ω stator)
Kailong RADII KL83 20 μF 903836-02
H1083
Toshiba Rotanode E7132 20 μF 903836-02
XH-121 E7239
XH-126 E7240
XH-150 E7242
XS-AV stator (See note 1)
XS-RA stator
(27.5/58 Ω)
Toshiba Rotanode E7252 20 μF 903836-02
XH-106V (See note 1)
XH-180
XH-181
XS-RA stator ONLY
(27.5/58 Ω)
Toshiba Rotanode E7299 20 μF 903836-02
XH-121 E7861
XH-126 E7876
XS-AV stator E7886
XS-RA stator (See note 1)
(27.5/58 Ω)

Note 1: Complete Toshiba tube assemblies include the suffix “X, “FX”, “GX”, or “JX”,
which are interchangeable with respect to rotor and anode characteristics.

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Service Manual Supplement # 90383800 Rev. B Page 4


CPI Canada Inc. Supplement

2.4 Chapter 8:
• Section 8.2.0:

Refer to table 3 below for the part number of the H.V. auxiliary board and table 4 below for the part
number of the capacitor field kit.

Table 3: H.V. Auxiliary Board


X-RAY TUBE STATOR TYPE 208 / 230V GENERATORS 400 / 480V GENERATORS
All stator types 739445-99 739442-99

Table 4: Capacitor Field Kit


Capacitor Field Kit part Number Capacitor Value Notes
903836-00 30 μF Includes Mounting Hardware
903836-01 40 μF
903836-02 20 μF

2.5 Chapter 9:
• MD-0935

C12 has been replaced by the chassis-mounted phase-shift capacitor as shown schematically in figure 3.

J7-4
COM
R19 J7-6
CHASSIS-MOUNTED
SHIFT PHASE-SHIFT CAPACITOR
E17 E18
R20 J7-5
MAIN
R31 J7-3
GROUND
R23

R24
R22

2 1

3 4 +5 V +5 V
1
2
R310

R305

U8
4
1 TP7

U7 TP8 J6-8 J11-8

J6-7 J11-7

J6-6 J11-6

Figure 3: Chassis-mounted phase-shift capacitor

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Service Manual Supplement # 90383800 Rev. B Page 5


CPI Canada Inc. Supplement

(This page intentionally left blank)

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Service Manual Supplement # 90383800 Rev. B Page 6

You might also like