Professional Documents
Culture Documents
90383800B CMP 200 DR LSS External Capacitor PDF
90383800B CMP 200 DR LSS External Capacitor PDF
Supplement
Contents:
1.0 INTRODUCTION............................................................................................................................................. 2
1.1 General ........................................................................................................................................................ 2
2.0 CHANGES TO THE CMP 200® X-RAY GENERATOR SERVICE MANUAL 901476 ..................................... 3
2.1 Chapter 1: .................................................................................................................................................... 3
2.2 Chapter 2: .................................................................................................................................................... 3
2.3 X-ray Generator Service Manual Supplement, X-Ray Tube Stator Compatibility Tables 746026:............ 3
2.4 Chapter 8: .................................................................................................................................................... 5
2.5 Chapter 9: .................................................................................................................................................... 5
The design and information contained in this document is proprietary to Communications & Power Industries
Canada Inc. (CPI), has been designed and developed at private expense, and is the exclusive property of CPI.
The information contained herein is loaned in confidence solely for the installation, operation, and maintenance of
CPI equipment. This document may not be duplicated in whole or in part, or be used for re-engineering, reverse
engineering, or otherwise reproducing in any form or creating or attempting to create or permitting, allowing or
assisting others to create or manufacture CPI’s product or products derived therefrom, without the express written
permission of CPI.
Copyright © 2011 by Communications & Power Industries Canada Inc. All rights reserved.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
1.0 INTRODUCTION
This supplement is not intended for CMP 200® and CMP 200® DR X-ray generators prior to serial
number CPD05500F09.
This supplement must be used in conjunction with the CMP 200® / CMP 200® DR service and
installation manual part number 901471.
The CMP 200® / CMP 200® DR service and installation manual part number 901471 covers the standard
CMP 200® / CMP 200® DR configurations and options. This supplement describes the manual changes
due to the chassis-mounted phase-shift capacitor that is no longer part of the H.V auxiliary board.
1.1 General
The Low Speed Starter (LSS) was integrated into the H.V auxiliary board. However, for this application,
the phase-shift capacitor for the LSS has been removed from the H.V auxiliary board and has been
mounted on the chassis (figure 2).
A revised H.V auxiliary board (figure 1), part numbers 739442-99 / 739445-99 and the chassis-mounted
phase-shift capacitor that is connected to E17 and E18 on the H.V auxiliary board via a cable makes the
LSS more configurable to be compatible with a larger number of X-ray tubes.
E17
Use and disclosure is subject to the restrictions on the title page of this CPI document.
2.0 CHANGES TO THE CMP 200® X-RAY GENERATOR SERVICE MANUAL 901476
2.1 Chapter 1:
• Section 1.4.0:
MAIN POWER
CONTACTOR
MAINS RECTIFIER
ASSEMBLY
EMC CAPACITOR
BOARD
PHASE-SHIFT CAPACITOR
DC BUS
CAPACITORS
(UNDERNEATH
DC BUS
INVERTER(S))
CAPACITORS
AUXILIARY
TRANSFORMER
HT TANK
CMP200_026B
2.2 Chapter 2:
• Section 2.7.0:
It is very important to note the value of the phase-shift capacitor before continuing. Refer to section 2.3 in
this supplement for the X-ray tube stator compatibility. If the value of the phase-shift capacitor is not
compatible with the desired X-ray tube stator type, the phase-shift capacitor must be changed to match
the requirements of the desired X-ray tube stator type. Please contact CPI product support for assistance.
2.3 X-ray Generator Service Manual Supplement, X-Ray Tube Stator Compatibility Tables 746026:
• Section 2:
Table 1 in this supplement provides a cross-reference between the H.V. auxiliary board part number
referenced in table 1 of supplement 746026 and the capacitor field kit part number.
The capacitor field kit part number determines the value of the phase-shift capacitor.
The capacitor field kit 903836-XX for the CMP 200® and CMP 200® DR X-ray generators includes the
specified chassis-mounted phase-shift capacitor (no longer part of the H.V auxiliary board) and its
mounting hardware. This allows the X-ray tube stator compatibility to be changed in the field.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
30 μF 739442-00 30 μF 903836-00
739445-00
45 μF 739442-01 40 μF 903836-01
739445-01
Table 2 in this supplement replaces table 1 of supplement 746026, for the following Tube Types only.
TABLE 2 : TUBE TYPE ( CMP 200® / CMP 200® DR LOW SPEED STARTER )
Note 1: Complete Toshiba tube assemblies include the suffix “X, “FX”, “GX”, or “JX”,
which are interchangeable with respect to rotor and anode characteristics.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
2.4 Chapter 8:
• Section 8.2.0:
Refer to table 3 below for the part number of the H.V. auxiliary board and table 4 below for the part
number of the capacitor field kit.
2.5 Chapter 9:
• MD-0935
C12 has been replaced by the chassis-mounted phase-shift capacitor as shown schematically in figure 3.
J7-4
COM
R19 J7-6
CHASSIS-MOUNTED
SHIFT PHASE-SHIFT CAPACITOR
E17 E18
R20 J7-5
MAIN
R31 J7-3
GROUND
R23
R24
R22
2 1
3 4 +5 V +5 V
1
2
R310
R305
U8
4
1 TP7
J6-7 J11-7
J6-6 J11-6
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Use and disclosure is subject to the restrictions on the title page of this CPI document.