D1762 − 84 (2013)
6.2.1 All of the selected sample shall be ground; no part of NOTE 3—Individual nichrome wire baskets to hold the crucibles are
the sample shall be rejected. The sample shall be pulverized convenient.
NOTE 4—If the sparking sample does not check the results of its
rapidly in a mill. Long grinding times shall be avoided because nonsparking duplicate within 60.5 %, the analysis shall be repeated.
of generated heat which will cause loss of volatile material.
Excessive grinding will produce a large amount of fine 7.4 Ash—Place the lids and the uncovered crucible used for
particles smaller than a No. 100 (150-µm) sieve. These fine the volatile matter determination, and containing the sample in
particles contribute to errors of being swept out of the crucible the muffle furnace at 750°C for 6 h. Cool the crucibles with lids
during the rapid evolution of gases in the determination of in place in a desiccator for 1 h and weigh. Repeat burning of
volatile matter. Particles that will be retained on a No. 20 the sample until a succeeding 1-h period of heating results in a
(850-µm) sieve shall not be used. Samples with the following loss of less than 0.0005 g.
sieve analysis have been found satisfactory: 8. Calculation and Report
Passing Sieve Retained on Sieve Sample, %
8.1 Calculate the percentage of moisture in the sample as
... No. 20 (850-µm) 0 follows:
No. 20 (850-µm) No. 40 (425-µm) 14.5
No. 40 (425-µm) No. 60 (250-µm) 18.7 Moisture, % 5 @ ~ A 2 B ! /A # 3 100 (1)
No. 60 (250-µm) No. 80 (180-µm) 7.0
No. 80 (180-µm) No. 100 (150-µm) 3.4 where:
No. 100 (150-µm) ... 56.4 A = grams of air-dry sample used, and
6.2.2 The ground sample shall be well mixed and stored in B = grams of sample after drying at 105°C (7.2).
an airtight container. 8.2 Calculate the percentage of volatile matter in the sample
as follows:
7. Procedure
Volatile matter, % 5 @ ~ B 2 C ! /B # 3 100 (2)
7.1 Make duplicate determinations.
where:
7.2 Moisture—Heat the muffle furnace to 750°C and place
previously ignited porcelain crucibles (Note 1) and covers in C = grams of sample after drying at 950°C (7.3).
the furnace for 10 min. Cool the crucibles in a desiccator for 1 8.3 Calculate the percentage of ash in the sample as follows:
h. Weigh the crucibles and add to each approximately 1 g, Ash, % 5 ~ D/B ! 3 100 (3)
weighed to the nearest 0.1 mg, of the ground sample. Place the
samples in the oven at 105°C for 2 h. Place the dried samples where:
in a desiccator for 1 h and weigh (Note 2). D = grams of residue (7.4).
NOTE 1—In practice, a crucible from a previous determination is used. 8.4 Report all results to the first decimal place. Values for
NOTE 2—The sample shall be considered oven-dry when the decrease in duplicate determinations should agree within the following:
weight of consecutive weighings is 0.0005 g or less. Succeeding drying
Permissible Differences
periods shall be not less than 1 h. Constituent Determined
Between Duplicates, %
7.3 Volatile Matter—Heat the muffle furnace to 950°C.
Moisture 0.1
Preheat the crucibles used for the moisture determination, with Volatile matter 0.5
lids in place and containing the sample, as follows: with the Ash 0.1
furnace door open, for 2 min on the outer ledge of the furnace
(300°C) and then for 3 min on the edge of the furnace (500°C) 9. Precision and Bias
(Note 3). Then move the samples to the rear of the furnace for 9.1 There is currently no data available with which to
6 min with the muffle door closed. Watch the samples through develop a precision and bias statement.
a small peep-hole in the muffle door. If sparking occurs, results
will be in error (Note 4). Cool the samples in a desiccator for 10. Keywords
1 h and weigh. 10.1 ash; charcoal; moisture; volatile matter
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT/).