Professional Documents
Culture Documents
Muhammad Fadli AKA18018 Kalibrasi Pipet
Muhammad Fadli AKA18018 Kalibrasi Pipet
Data Kalibrasi
Berat Kosong Berat Isi Suhu Air Suhu Udara Kelembapan Tekanan Udara
No
mk (g) mb (g) ta(oC) tu(oC) Rh(%) p (hPa)
1 20.69667 22.68656 24.48 25.5 53.0 1000
2 20.69672 22.68669 24.48 25.5 53.0 1000
3 20.68652 22.68652 24.48 25.5 53.0 1000
4 20.69654 22.68638 24.48 25.5 53.0 1000
5 20.69664 22.68646 24.48 25.5 53.0 1000
Average 22.68652 24.48 25.5 53.0 1000
2019
KALIBRASI PIPET
Suhu : 25.5 °C
Kelembaban : 53 % RH
Tekanan udara : 750 mmHg
AIR
DARA
Berat Kosong Berat Isi Suhu Air Suhu Udara Kelembapan Tekanan Udara ra ru γ Volume
No
mk (g) mb (g) ta(oC) tu(oC) Rh(%) p (hPa) g/ml g/ml ml
1 20.69667 22.68656 24.48 25.5 53 1000 0.9972745 0.001544040 9.9E-06 22.7826
2 20.69672 22.68669 24.48 25.5 53 1000 0.9972745 0.001544040 9.9E-06 22.7827
3 20.68652 22.68652 24.48 25.5 53 1000 0.9972745 0.001544040 9.9E-06 22.7826
4 20.69654 22.68638 24.48 25.5 53 1000 0.9972745 0.001544040 9.9E-06 22.7824
5 20.69664 22.68646 24.48 25.5 53 1000 0.9972745 0.001544040 9.9E-06 22.7825
average 22.7826
SD 0.0001
IDENTITAS ALAT
Nama Alat : pipet volume 5mL
Merk / Pabrik : Socorex
Kode Alat : tipe A
Kapasitas : 5 mL
Kondisi Ruangan :
Tgl. Penerimaan : 13 April 2019 Suhu Ruangan : 25.5 oC
Tgl. Kalibrasi : 14 Juni 2019 Kelembaban : 53.0 %RH
Tempat Kalibrasi : lab kimia analisa Tekanan Udara : 750 mmHg
Lokasi Alat : Lab KM
Metoda : ISO 4787
Standar : Timbangan Elektronik Terkalibrasi Internal
HASIL KALIBRASI
Penunjukan Alat Koreksi Ketidakpastian
(mL) (mL) (mL)
5 -17.783 0.005
Laporan ini dilarang digandakan secara tidak lengkap tanpa persetujuan tertulis dari Laboratorium Kendali Mutu PT SASA INTI Gending
afarma/IV/2019