Temperature and Humidity (Installation and Annual PM) The system unit meet GE specification
Rooms Temp (Max) (Celcius) Temp (Min) (Celcius) Hum (Max) % Hum (Min) %
EXAM Room 24 18 65 60 Recommendation
Technical Room
Operator room 22 18 65 60
Safety Check
System Level Safety Test Status (Pass/Fail/NA)
Patient Touch Leakage Test PASS Date : 18 Maret 2020
System Chasis Leakage Test PASS
Emergency Status Test PASS
Ground Continuity PASS
Door Switch PASS
Xray Warning lamp PASS
Deni Firmansyah ……………………………….
Team Leader PT. GEOI Ka.Instalasi RS/Klinik