Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Proceedings

Measurement of Pelvic Orientation Angles during


Sprinting Using a Single Inertial Sensor †
Tomohito Wada 1,*, Ryu Nagahara 1, Sam Gleadhill 1, Tatsuro Ishizuka 2, Hayato Ohnuma 3 and
Yuji Ohgi 2
1 School of Physical Education, National Institute of Fitness and Sports, Kanoya, Kagoshima 891-2393, Japan;
nagahara@nifs-k.ac.jp (R.N.); gleadhill@nifs-k.ac.jp (S.G.)
2 Graduate School of Media and Governance, Keio University, Fujisawa, Kanagawa 252-0882, Japan;

tatsu47@sfc.keio.ac.jp (T.I.); ohgi@sfc.keio.ac.jp (Y.O.)


3 Faculty of Education, Kansai University of Social Welfare, Ako, Hyogo 678-0255, Japan;

h.ohnuma0613@gmail.com
* Correspondence: wada@nifs-k.ac.jp; Tel.: +81-994-46-4941
† Presented at the 13th Conference of the International Sports Engineering Association, Online,
22–26 June 2020.

Published: 15 June 2020

Abstract: The purpose of this study was to elucidate pelvic orientation angles using a single lower
back-mounted inertial sensor during sprinting. A single inertial sensor was attached to each
sprinter’s lower back, used to measure continuous pelvic movements including pelvic obliquity
(roll), anterior-posterior tilt (pitch) and rotation (yaw) during sprinting from a straight to bend
section. The pelvic orientation angles were estimated with the three-dimensional sensor orientation
using a sensor fusion algorithm. Absolute angles derived from the sensor were compared with
angles obtained from an optical motion capture system over a 15 m length. The root mean squared
error between the sensor and motion capture data were 4.1° for roll, 2.8° for pitch and 3.6° for yaw.
Therefore, the sensor was comparable to the motion capture system for tracking pelvic angle
changes. The inertial sensor is now supported as a valid tool to measure movements of the pelvis
during sprinting.

Keywords: inertial sensor; sprint; curve; pelvic tilt; sensor fusion

1. Introduction
Thoracic and pelvic tilt changes step to step during maximal accelerated sprinting [1]. The pelvic
orientation during sprinting influences the power generation applied onto the ground [2], and thus
is important to consider for coaches. Motion capture systems (mocap) are typically considered a gold
standard method for accurate human kinematic monitoring and have been previously utilised for
sprinting research [2]. However, mocap is expensive, has high expertise requirements, is usually
limited to laboratory environments and typically has only been used to measure sprinting over a
short distance [2–5]. In sprint events, sprinters run on both straight and curved tracks, except short
distance events such as the 100 m. Previous studies showed differences in sprinter’s kinematics
between straight and curved sprinting [6,7], but it was difficult for mocap to measure continuous
changes in kinematics over the entire race.
Wearable sensors including inertial measurement units (IMUs) allow in field monitoring and
may overcome the practical limitations of mocap, while remaining comparable in accuracy for
monitoring sprint kinematics. IMUs have been used for sprinting studies to estimate temporal
parameters [8], stance duration [9] and sprint velocity [10], demonstrating that IMUs can be used to
monitor sprinting in the field. Bergamini et al. [11] measured sprinter’s trunk inclination using a
Proceedings 2020, 49, 10; doi:10.3390/proceedings2020049010 www.mdpi.com/journal/proceedings
Proceedings 2020, 49, 10 2 of 6

lower back-mounted IMU and reported valid angular displacements and angular velocity in the
sagittal plane during the sprint start phase. However, it has not been elucidated whether IMUs are
also valid for the measurement of curved sprinting and angular displacements in the coronal or
transverse plane.
The purpose of this study was to elucidate pelvic orientation angles including pelvic obliquity
in the coronal plane (roll), anterior–posterior tilt in the sagittal plane (pitch) and rotation in the
transverse plane (yaw) using a single lower back-mounted inertial sensor during sprinting. To this
end, sprint runs were performed on an indoor track and an outdoor 400 m track by collegiate male
sprinters. The IMU data obtained from the indoor track trials were compared with mocap data to
determine the absolute angle errors of using a single IMU for pelvic orientation monitoring.

2. Materials and Methods


A DSP wireless 9 axis motion sensor (Sports Sensing, Fukuoka, Japan) was used as the IMU in
this study. This unit has a tri-axial accelerometer, a gyroscope and a magnetometer, with dimensions
53 mm × 38 mm × 11 mm (H × W × D) and a weight of 32 g. One sensor was attached with thin double-
sided tape above the participant’s L5 spinous process and overlapped with a plastic surgical tape and
covered with form-fitting sprint pants. The sensor range was ±16 G and ±1500 dps and the sampling
rate was 200 Hz with 16-bit resolutions. The sensor started recording after it was attached to a
participant and continued until all trials were finished. To synchronise with mocap, a DSP wireless
analogue signal logger (Sports Sensing, Fukuoka, Japan) was used to record triggers pulled at the
start of sprint trials. The orientation angles of the sensor were derived with Madgwick’s previously
validated open-sourced AHRS implementation for an IMU [12]. This algorithm is computationally
inexpensive but as accurate as the Kalman-based algorithm. It achieved a static root mean square
(RMS) error of <0.8° and a dynamic RMS error of <1.7° [12]. In this research, only angular velocity
and acceleration were used for angle estimation, and its parameter β was set to 0.03.
A mocap setup with twenty-eight infrared cameras (Raptor-E, Motion Analysis, CA, USA) was
used as a reference system. The sampling frequency was 250 Hz for the recording, and it was
resampled to 200 Hz for further processing. Cameras were arranged around an indoor sprint track
and captured participant’s pelvic movement from approximately 30 to 45 m from the start line. To
observe the difference between straight sprinting and entering to the bend section, a bend track
(radius 38.2 m) was set at a 40 m position for the bend sprint tests. Full body marker sets were used
to observe sprint technique. Coordinates of markers including the right and left anterior superior
(ASIS) and posterior superior (PSIS) iliac spines were used to calculate pelvic movements (Figure 1).

Figure 1. Positions and coordinate system of the inertial measurement unit (IMU) and mocap markers.

Pelvic orientation angles from the mocap were calculated with four markers according to
previous studies [1,13,14]. Pelvic obliquity (roll) and rotation (yaw) were calculated as tilts of left and
right ASIS markers towards the horizontal line and the runway. Anterior–posterior tilt (pitch) was
obtained as an angle formed by a vertical plane and the plane defined by ASIS markers and a middle
Proceedings 2020, 49, 10 3 of 6

point between the two PSIS markers. Analogue triggers at the start of trials were recorded by the
mocap and used for data synchronisation.
Participants were instructed to stand in an upright anatomical position prior to data capture,
used for IMU data calibration for the pitch and roll. In this research, magnetometer data were not
employed to calculate angles, and thus absolute heading angle (yaw) could not be obtained for the
sensor. Therefore, the sensor yaw angle was offset with the first angle obtained from the mocap for
each trial.
Eight collegiate male sprinters (age: 20.9 ± 0.8 year; mass 66.2 ± 4.2 kg; stature: 171.6 ± 3.5 cm)
participated in the indoor trials. Each participant performed two 50 m straight sprints and two 44 m
sprints that started on the straight track and transitioned into a bend sprint. The participants started
from a crouched starting position for each trial and sprinted through a 15 m capture area of mocap.
Additionally, outdoor 200 m maximal sprint trials were carried out by three participants to confirm
the effectiveness of the method in a practical sprint distance and environment including a curved
path.

3. Results

3.1. Comparison with Motion Capture System


Sixteen straight sprint data and sixteen entering the bend sprint data were collected by indoor
tests. Figure 2 shows a representative sample of pelvic tilts (roll, pitch and yaw) from one participant
estimated by the IMU and calculated from mocap data. Support phase durations of each leg are
shown with vertical lines for reference. Support phase durations were manually retrieved from the
observation of full body markers of the mocap.

(a) (b)

Figure 2. Representative sample of pelvic tilts of a sprinter measured with an IMU (blue line) and
mocap (red dotted line) during straight sprinting (a) and entering the bend section (b).

To assess the IMU accuracy compared to mocap, the root mean squared error (RMSE) and a
Pearson’s correlation coefficient between IMU and mocap were computed for each trial and are
shown in Table 1. The absolute angle RMSE between the IMU and mocap was 4.1° for roll, 2.8° for
pitch and 3.6° for yaw for all trials. Pearson’s correlation coefficients were 0.88 for roll, 0.79 for pitch
and 0.97 for yaw, demonstrating strong linear relationships for all pelvic tilt angles.
Proceedings 2020, 49, 10 4 of 6

Table 1. Root mean square error (RMSE) of differences and Pearson’s correlation coefficients for
pelvic angles between the two systems (mean ± SD).

RMSE (deg) Pearson’s Correlation


Trial
Roll Pitch Yaw Roll Pitch Yaw
Straight 4.2 ± 1.2 2.4 ± 1.0 3.7 ± 1.8 0.86 ± 0.09 0.84 ± 0.06 0.97 ± 0.01
Entering to Bend 4.1 ± 1.3 3.1 ± 0.8 3.5 ± 1.6 0.90 ± 0.05 0.75 ± 0.13 0.96 ± 0.02
Total 4.1 ± 1.2 2.8 ± 1.0 3.6 ± 1.7 0.88 ± 0.07 0.79 ± 0.11 0.97 ± 0.02

3.2. Two-Hundred Metre Sprint Measurements with a Single IMU


Figure 3 shows a representative sample of pelvic angles for a 200 m sprint in lane four of an
outdoor athletic track (radius 41.8 m). In this figure, orientation angles were set to zero at the
sprinter’s stable pose after the “on your marks” call. The start time (t = 0) was manually chosen from
the data and the data include both before and after sprinting.

Figure 3. Representative sample of pelvic tilts transition for 200 m sprint.

4. Discussion
The indoor test results demonstrate accuracy of less than five degrees of RMSE compared to
mocap for all pelvic angles. There were small differences between the types of trials (straight and
entering into bend). This level of accuracy is sufficient for monitoring pelvic angles for coaches to
judge the influence of technique during maximal sprinting.
In the test, mocap tracked four markers attached around the pelvis according to previous studies
[1,13,14]. Therefore, the IMU data were compared with quasi-pelvic movement as reported in
previous studies, not with the device’s housing angles measured by mocap with device attached
markers. The notable finding of the presented method is that a single IMU may be used to estimate
detailed pelvic movement which was previously measured with a four maker mocap setup during
sprinting.
Some differences between pelvic angles measured using IMU and mocap were also found. For
example, on pitch angles, small peaks occurred after toe-off in IMU data but not for mocap. This
difference may be derived from the methods of calculating angles and/or the motion artefacts of the
skin-mounted sensor. However, on a broader basis, the angles estimated by an IMU had minimal
differences for assessing pelvic movements during sprinting. It was observed that the roll and yaw
Proceedings 2020, 49, 10 5 of 6

angles periodically change in two steps, and pitch angles periodically change in a step. Roll angles
decreased (to right upward) at the beginning of the right leg support phase, and started to increase
(to left upward) at end of the right leg support phase. Inverse angle changes were observed at the left
leg support phase. Pitch angles decreased (lean forward) during right and left leg stance phases and
increased (lean backward) during the flight phases. Yaw angles changed direction during right and
left support phases. In the tests, the 19th to 25th steps were captured by mocap at around 30 to 45 m
from the start line. The amplitude of these steps for straight sprints almost matched with the
amplitude during the comparable section reported by a previous study [1].
In Figure 2, the difference between straight sprinting and transitioning into the bend section was
observed in roll and pitch angles. In Figure 2b, the bend section started before the last left leg contact
(around 5.9 s). The roll angle started rotating toward the right upward two steps before the bend
section (around 5.5 s). Clear understanding for bend sprinting has yet to be established but the use
of IMUs may contribute toward future research.
During the field measurements (200 m sprints), all pelvic angle transitions showed consistent
results, as demonstrated in Figure 3. The transition of pelvic angles shows reasonable vibration and
it was observed that the amplitude and waveform were similar to the indoor tests. Therefore, it was
assumed that the results during the measured 15 m capture length (indoor trials) of mocap data are
translatable to practical conditions. However, there were several raw acceleration data in excess of
the measurable limits. These happened in a very short time (only one sample at a time) and did not
appear to influence the estimation of angles, but it should be considered in detail in the future.

5. Conclusions
It was confirmed that a single inertial sensor is now supported as a valid tool to measure
movements of the pelvis during sprinting. The strength of IMU devices is that they are commercially
available and affordable, have comparable accuracy to mocap and overcome the practical limitations
of mocap systems. More practical mounting includes wearing a single sensor in worn garments
during sprinting, opposed to being applied with tape on the skin. This may be more practical for field
use and eliminate expertise requirements for anatomical mark up; however, it may present greater
errors due to movement of the worn garments and no expertise for mark up. This may be an area for
future investigation. In addition, this research supports the use of the presented method for future
sprint research designs. Paired with simple user-friendly graphical user interfaces, these devices may
have significant practical applications for coaches to provide real time feedback for athletes.

Acknowledgments: This work was supported by JSPS KAKENHI Grant Numbers JP17K01675 and JP18H03344.

Conflicts of Interest: The authors declare no conflict of interest. The funding sponsors had no role in the design
of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, and in the
decision to publish the results.

References
1. Nagahara, R.; Matsuo, A.; Matsubayashi, T.; Zushi, K. Kinematics of the thorax and pelvis during
accelerated sprinting. J. Sports Med. Phys. Fit. 2018, 58, 1253–1263.
2. Debaere, S.; Delecluse, C.; Aerenhouts, D.; Hagman, F.; Jonkers, I. From block clearance to sprint running:
Characteristics underlying an effective transition. J. Sports Sci. 2013, 31, 137–149.
3. Mayagoitia, R.E.; Nene, A.V.; Veltink, P.H. Accelerometer and rate gyroscope measurement of kinematics:
An inexpensive alternative to optical motion analysis systems. J. Biomech. 2002, 35, 537–542.
4. Papi, E.; Koh, W.S.; McGregor, A.H. Wearable technology for spine movement assessment: A systematic
review. J. Biomech. 2017, 64, 186–197.
5. Schmitz, A.; Ye, M.; Boggess, G.; Shapiro, R.; Yang, R.; Noehren, B. The measurement of in vivo joint angles
during a squat using a single camera markerless motion capture system as compared to a marker based
system. Gait Posture 2015, 41, 694–698.
6. Alt, T.; Heinrich, K.; Funken, J.; Potthast, W. Lower extremity kinematics of athletics curve sprinting. J.
Sports Sci. 2015, 33, 552–560.
Proceedings 2020, 49, 10 6 of 6

7. Ohnuma, H.; Tachi, M.; Kumano, A.; Hirano, Y. How to Maintain Maximal Straight Path Running Speed
on a Curved Path in Sprint Events. J. Hum. Kinet. 2018, 62, 23–31.
8. Bergamini, E.; Picerno, P.; Pillet, H.; Natta, F.; Thoreux, P.; Camomilla, V. Estimation of temporal
parameters during sprint running using a trunk-mounted inertial measurement unit. J. Biomech. 2012, 45,
1123–1126.
9. Schmidt, M.; Rheinländer, C.; Nolte, K.F.; Wille, S.; Wehn, N.; Jaitner, T. IMU-based Determination of
Stance Duration during Sprinting. Procedia Eng. 2016, 147, 747–752.
10. Gurchiek, R.D.; McGinnis, R.S.; Needle, A.R.; McBride, J.M.; van Werkhoven, H. An adaptive filtering
algorithm to estimate sprint velocity using a single inertial sensor. Sports Eng. 2018, 21, 389–399.
11. Bergamini, E.; Guillon, P.; Camomilla, V.; Pillet, H.; Skalli, W.; Cappozzo, A. Trunk inclination estimate
during the sprint start using an inertial measurement unit: A validation study. J. Appl. Biomech. 2013, 29,
622–627.
12. Madgwick, S.O.H.; Harrison, A.J.L.; Vaidyanathan, R. Estimation of IMU and MARG orientation using a
gradient descent algorithm. In Proceedings of the 2011 IEEE International Conference on Rehabilitation
Robotics (ICORR), Zurich, Switzerland, 29 June–1 July 2011; pp. 1–7.
13. Leardini, A.; Berti, L.; Begon, M.; Allard, P. Effect of trunk sagittal attitude on shoulder, thorax and pelvis
three-dimensional kinematics in able-bodied subjects during gait. PLoS ONE 2013, 8, e77168.
14. Begon, M.; Leardini, A.; Belvedere, C.; Farahpour, N.; Allard, P. Effects of frontal and sagittal thorax
attitudes in gait on trunk and pelvis three-dimensional kinematics. Med. Eng. Phys. 2015, 37, 1032–1036.

© 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access
article distributed under the terms and conditions of the Creative Commons Attribution
(CC BY) license (http://creativecommons.org/licenses/by/4.0/).

You might also like