Estimation of Measurement Uncertainty For Total PH in Water: U1 (Relative)
Estimation of Measurement Uncertainty For Total PH in Water: U1 (Relative)
Estimation of Measurement Uncertainty For Total PH in Water: U1 (Relative)
1.Introduction of Test
Estimation of Measurement Uncertainty for pH in Water
2. Equipments Used
pH meter
Observation pH Value
1 7.12 Hence 'A' Uncertainty
2 7.11 U = δn-1
3 7.13 √n
4 7.12
5 7.10 n=6
6 7.12
Mean 7.12
STD.DEV. 0.010
U1 0.004
U1 (Relative) 0.000592464198626 U12
5.2. Uncertainty due to CRM (pH Buffer 4.0 used for calibration)
Uncertainy from Certificate 0.014
Standad Uncertainty 0.007
U3 (Relative) = 0.00175 U32
5.2. Uncertainty due to CRM (pH Buffer 7.0 used for calibration)
Uncertainy from Certificate 0.014
Standad Uncertainty 0.007
U4 (Relative) = 0.001 U42
5.2. Uncertainty due to CRM (pH Buffer 9.0 used for calibration)
Uncertainy from Certificate 0.014
Standad Uncertainty 0.007
U5 (Relative) = 0.000777777777778 U52
6. Combined Uncertainty
___________________
Uc= √U12+U22+U32+U42+U52
Uc
7. Effective Degrees of freedom
[Uc4(y)]
Veff = -----------------------------
{∑N[Ui4(y/Vi)]}
7.729385958483E-07
= -------------------------------------------- 31366535.3357048
2.46421413004597E-14
8. Expanded Uncertainty
Uexp = K X Uc
0.059301577200531
9. Final Result
Measurement Uncertainty for pH in Water at level of
7.12 is 0.059302
3.51013827E-07
1.23403298E-05
3.0625E-06
1E-06
6.04938272E-07
1.73587819E-05
0.004166387157
0.0296507886
7.18