Professional Documents
Culture Documents
Operation and Calibration of UV-VIS Spectrophotometer
Operation and Calibration of UV-VIS Spectrophotometer
Operation and Calibration of UV-VIS Spectrophotometer
PURPOSE :
To provide a procedure to operate and Calibrate the spectrophotometer, Model
Shimadzu, UV1700 UV-VIS
II. SCOPE:
The procedure is applicable to the UV/Visible Spectrophotometer being used for
analytical purpose at the Quality Control laboratory,
III. RESPONSIBILITY:
It is the responsibility of the Quality Control personnel involved in the
UV/Visible Spectrophotometer Calibration, to follow the procedure and schedule as mentioned
in this Standard Operating Procedure.
IV. DOCUMENT REFERENCE:
SOPs : Core procedure, Calibration of Equipment
Operation of UV Visible Spectrophotometer
Forms : UV Visible Spectrophotometer Calibration Record
V. OPERATION:
1. 0 PRELIMINARY CHECK:
1.1 Check and ensure that the instrument is clean and suitable for starting the
operation. If it is not, clean and ensure the suitability.
2.0 BASIC OPERATION:
2.1 Put Power button `ON’ provided on left hand side Instrument.
2.2 Keep LCD unit in vertical position.
2.3 As soon as, power is kept ON, LCD will show, initialisation checking of
different parts will start. Each test `PASS’ will be indicated by .
2.4 After every check is over & passed it will display mode.
2.5 When the power is turned `ON’ the spectrophotometer is checked and initialised.
As each item enters it’s initialisation operation it is highlighted. However, if any
abnormality is detected, the initialisation process is interrupted at that item,
without it’s star being highlighted.
2.6 Wait for 15 minutes after automatic initialisation for optimum illumination of UV
lamp.
3.0 BASIC MODE MENU:
3.3 Quantitation :- Creates a calibration curve from a standard sample and quantitates an
unknown sample.
3.4 Kinetics : - Used to measure the activity from the time dependent changes in absorbance
& % transmittance
3.5 Time scan: - Used to measure the Change in absorbance, transmittances or energy with
time in the fixed wavelength.
3.6 Multi-Component : - Used for multi - component determination.
3.7 Photometric (multi ) : To measured the absorbances or transmittances up to eight
wavelength.
3.8 Optional program pack: Used for optional program pack.
3.9 Utilities: Used for setting and changing the basic operating parameters of the instrument.
D2 lamp off time should be .
On routine basis following modes are used with the given operations.
3.16 When in basic mode menu press ‘2’ to go in the spectrum mode. Then ‘6’ different
parameters will be displayed.
3.17 Press `1’ to select Absorbance / Transmittance / Energy mode.
3.18 Press ‘2’ & then give the higher wavelength of the desired range and press ‘ ENTER’.
Then give the lower wavelength & press ‘ENTER’.
3.19 Then press, ‘3’ and give higher value of Absorbance which is greater than the expected
Absorbance value & then press ‘ENTER’. Lower value should be kept zero only, so
press ‘ENTER’.
3.20 The speed of the scanning can also be changed. Press ‘4’. Then select ‘1’, ‘2’,`3’`4’
and `5’ for scanning with `Very Fast’, `Fast’, ‘Medium’, ‘Slow’ and `Very slow’
speed respectively.
3.21 Press `6’ to select display mode i.e. sequential or overlay.
3.22 Place the cells filled with blank solutions in both the cuvette holders and press ‘F1’ so
that baseline correction will be done.
3.23 Rinse the cuvette with sample solution. Place the sample in the front side cuvette holder.
Then press ‘START’ key.
3.24 The scanning in the given wavelength range will be done.
3.25 To know the Absorbance value at a particular wavelength in the range given, make use of
‘<-----’ and ‘------> ‘ keys.
3.26 Once scanning is over. Press F2 and select 3 for peak detection and F4 for valley. Press
graph to obtain scan with peak detection. Then press print key.
3.28 Then press ‘Return’ & ‘Mode’ to come back to the basic mode menu.
4.0 MAINTENANCE :
4.1 Do not keep cells above the outside body of the spectrophotometer. Clean the cells
perfectly and wipe out sidewalls with the tissue paper in one direction only. Do not hold
the cells with their optical surfaces between the fingers.
4.2 Press the keys and handle the instrument very carefully.
4.3 The extinction of the cells intended to contain the solution to be examined and the
reference liquid when filled with the same solvent, should be identical. If this is not the
case, an appropriate correction must be applied. The tolerance on the path length of the
cells used is +/- 0.005 cm. Cells should be cleaned and handled with great care.
4.4 When not in use, put off the power switch of the instrument and then the mains.
4.5 Check condition of Silica gel for effectiveness and replace once in a week (or) earlier if
required.
solution.
03. Shake the solution well to dissolve Holmium oxide.
04. Make up the volume to the mark with 1.4 M Perchloric acid solution.
Procedure:
01. Take a pair of matched quartz cells having a path length of 1 cm.
02. Perform the base line correction with 1.4 M Perchloric acid solution from 200 to
600 nm.
03. Remove the cell from sample compartment.
04. Rinse the sample cell with Holmium perchlorate solution, fill the cell with the
same, clean the smooth surfaces of the cell with tissue paper and place it in the sample
compartment
Acceptance criteria:
3.5 Procedure:
Measure the absorbances of the above solutions at about 302 nm against water blank.
Check the absorbance of the above solutions at 302 nm using milli Q water as blank.
Record the absorbance at each wavelength in the Calibration Data Sheet.
Check the Absorbance for each solution against the acceptance criteria given below :
Calculate correlation coefficient using mean of the absorbance of respective mean
concentration.
3.6 Acceptance Criteria:
In `Photometry mode ’ Select transmittance mode. Set the wavelength at 200 nm. Then
press `autozero’ to get 100 % transmittance with air blank in both the cell holders. Then
place the cell in the sample compartment filled with milli Q water using air blank and
read the transmittance. Similarly repeat the operation for 220 nm and 240 nm. Repeat
the test for second cell. The transmittance values should meet the requirement as given
in the following table.
Acceptance Criteria :
Wavelength % Transmittance
200 nm 73 to 77 %
220 nm 83 to 87 %
240 nm 85 to 89 %
7.0- IO - LINE FLATNESS :
a. With air blank in the sample and reference compartment, perform the test.
b. Scan in the wavelength range 220 to 700 nm.
c. Obtain the spectrum in the % transmittance mode.
Acceptance Criteria : The Io line obtained should not deviate from a horizontal
line at any point by more than 99.54 - 100.46 % transmittance.