1. Preface
This manual specifies the Bi-directional communication between analyzer and the
(Host) based on ASTM1381-91 and ASTM1394-91
There are the following Sample ID mode and Requast setting for analyzer.
Sample ID mode
By Sample No. mode: This is a method to inquire to Sample tray No.
By Sample ID mode: This is a method to inquire to barcode attached with sample.
Request
All (Batch): enquire by “ALL” orders of host.
Appoint (individually) : In case enquire by barcode.
Note:
The trace file of communication log between the analyzer and host is automatically
created. The file is named “astm_tushin.log” at following place.
C:¥BiOLiS48i¥PC_Prog¥file¥astm_tushin.log.
Do not open the file during communication.
Create a copied file when viewing the communication log. Original file should not be
opened.
1
V1.06
2. Connection Condition
LAN cable (TCP/IP) is used for connection between analyzer and Host.
Port: 5010
Host: Server
50i: Client
*If you use HUB for connection between analyzer and Host, please use a straight
cable. If you connect directly, please use a cross cable.
Please set the communication port manually. The analyzer does not find the port
automatically. Refer to Appendix B.
2
3. Protocol of Data Link Layer
Protocol of Data Link Layer basically depends on TCP protocol.
3.1. Frame
[FN][TEXT]
1H|¥^&|||HOST^P_1|||||BiOLiS NEO^SYSTEM1||P|1|20130910220351[CR]
2P|1|PID0002|||[BB][C4][B3] [C3][D9][D6]^|||F|||||Dr. Miura[CR]
3O|1|01040218001|0^1^1|^^^1^^0|R||||||N||||Serum||||||||||O[CR]
4L|1|N[CR]
[FN] is a frame number. The first frame starts from number 1 which is ASCII code
of 0 ~ 7.
[TEXT] is the text data to be transmitted.
In analyzer, [TEXT] corresponds to the record (ASTM 1394-91).
[TEXT] does not control maximum number of characters.
The message is processed by one telegram from Header Record to Terminal Record.
Enquiry Message
[FN] Message Header Record
[FN] Enquiry Record
[FN] Termination Record
3
V1.06
4. Record
The records are defined by ASTM1394-91.
The records supported by analyzer are as follows.
The relationship between the hierarchical level supported by analyzer and the
hierarchical level of record is as follows.
* Level supported
-- Level not supported
4
4.1. Delimiter
The setting of the delimiter used by the record that is transmitted from analyzer can be
changed.
The character codes, 20h – 7Eh and 80h – FFh, can be used in the records.
When the records include the character codes that cannot be shown, they are likely to
disrupt the screen of analyzer
In case that the records include the same codes as the codes defined for the delimiters,
escapte characters should be used.
For analyzer escape characters may only be used in the following fields
5
V1.06
This chapter explains the definition of each record sent from the analyzer to Host
In the field of each record, the necessity of transmission means as follows.
Example of transmission
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
6
4.4.2. Patient Information Record (From analyzer to Host)
When the patient information is in analyzer, the optional items of patient information
record are set.
7
V1.06
8
4.4.4. Enquiry Record (From analyzer to Host)
The enquiry timing for “ALL” is when press the “Request” button or “Stat start”
button.
The enquiry timing for “Appoint” is when the analyzer reads the sample barcode.
In case of “All” (batch) request, please transmit measurement order enquiry when
order is decided on host side. The order can accept under state of run.
In case of “All” (batch) request, please transmit no measurement order enquiry
when order is unspecified. ( Refer to example of response No.3 : In case of no
measurement order.)
9
V1.06
Example of transmission
C|1|I|1|I<CR>
The qualitative record of Hemolytic, Lipemic and Icteric for serum information
are set the field 4 as comment text in comment record.
Hemolytic : H
Lipemic :L
Icteric :I
Example of transmission when the results are the Hemolytic(2+), Lipemic(-)
and Icteric(+).
The results of three items (96-98) are transferred the order from 97(Hemolytic),
96(Lipemic) and 98(Icteric).
1C|1|I|2+ - +|G<CR>
Note:
When qualitative serum measured, the qualitative results are sent at the end of the
sequences of records of each samples (such as measurememnt result record or
comment record )
10
4.4.6. Measurement Result Record (From analyzer to Host)
Field No. Field name Explanation Necessity of
transmission
1 Record ID “R” Indispensable
2 Sequence No. Sequence number of measurement result record Indispensable
3 Test item ID Indispensable
1 Test ID Blank is set. Not supported
2 Test name Blank is set. Not supported
3 Test type Blank is set. Not supported
4 Test item No. Test item number Indispensable
5 Test item name Test item name Indispensable
6 Dilution rate “0” : No dilution (default) Indispensable
“1” : 3 “7 “ : 40
“2” : 5 “8” : 50
“3” : 6 “9” : 60
“4” : 10 “10” : 70
“5” : 20 “11” : 80
“6” : 30 “12” : 90
“13” : 100
4 Measurement Measurement value is written by ASCII text. Indispensable
value In case the condition of result is “X” or “I”, blank is set.
5 Unit In case unit is registered in analyzer, it is set. Optional
6 Reference range In the form of “the lower limit to the higher limit” Optional
In case that sex is “U” (unknown) or is omitted in
patient information record, reference range for man is
set.
This is set when normal range is registered in analyzer.
7 Abnormal result “L” : lower than the lower limit. Indispensable
flag “H” : higher than the higher limit.
“N” : normal
“PL” : lower then the panic range.
“PH” : higher then the panic range.
In case of measurement failure, “N”is set, too.
9 Condition of “F” : final result Indispensable
result Measurement result is transmitted.
“I” : in instrument, result pending
This test item is for automatic re-measurement and
in the waiting state for measurement, or is in the
measurement waiting state when STAT is cutting in.
(No measurement value)
As measurement is completed, this is retransmitted
as the final result of “F”.
“P” : preliminary result
When measurement ended in failure, it is set.
(With a measurement value.)
“X” : No result
When measurement ended in failure, it is set.
(No measurement value)
The cause of failure (error code) is set to the
comment record, following to this record.
13 Date & time of Date and time when measurement was over. Indispensable
ending of test (In the form of YYYYMMDDHHMMSS)
Example of transmission when measurement succeeded
R|1|^^^1^GOT^0|54.5143|IU/L|8 TO 38|H||F||||20010618145805<CR>
R|1|^^^1^GOT^0|9999999|IU/L|8 TO 38|N||F||||20010618145805<CR>
If the conc. value can not be calculated even measurement was a success, the results are
transmitted as “9999999”.
11
V1.06
Transmission at STAT
STAT sample will cut in between the items of the sample measurement.
The result data will be transmited twice: at the time when all the item
measurement results are obtained before STAT sample cuts in and at the time
when all the measurement results are obtained after STAT sample has cut in.
12
4.5. The Record to be transmitted from Host to analyzer
Necessity of
Field No. Field name Explanation
transmission
The character to discriminate this record
1 Record ID Indispensable
“H”: The field with only one character, H
Analyzer initial values are as follows.
Field delimiter |
Indispensable
2 Delimiter definition Repeat delimiter ¥
Component delimiter ^
Escape delimiter &
5 Sender ID Indispensable
1 Sender’s name The name of Host is set. Indispensable
2 System ID System ID of Host is set. Indispensable
10 Receiver ID Indispensable
1 Receiver’s name “BIOLIS 50i” Indispensable
2 System ID System ID of analyzer is set. Indispensable
12 Process ID “P” : Production Indispensable
13 Version number “1” Indispensable
The date and time that the message was generated.
Date & time of
14 In the form of YYYYMMDDHHMMSS. Indispensable
message
(year/month/day/hour/min./sec.)
There is no discrimination between capital letters and small letters in Sender ID and
Receiver ID.
Example of transmission
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
13
V1.06
In case there is no patient information to transmit, leave the optional field blank.
14
4.5.3. Measurement Order Record (From Host to analyzer)
As test item number and item name are set or modified by each laboratory, they
are not defined in the specifications. Host system should have compatible item
number and item name list registered in analyzer. Please get the list from the
person in the laboratory.
Note:
In case of measurement of urine sample, ISE (Na, K and Cl) and other items
cannot be measured at the same time.
The reason is that the uring samples for ISE are measured after six times dilution
by uring diluent. Therefore, the result is automatically multiplied by six and the
transmitted.
Also, in case of dialysis sample, please select “Dialysis” as a specimen in the order
screen.
16
Example of discrimination of test item
(The following three records mean the same meaning.)
O|1|12345|^1^1|^^^22^^0|R||||||N||||Serum||||||||||O<CR> *Omitted test item name
O|1|12345|^1^1|^^^^PTT^0|R||||||N||||Serum||||||||||O<CR> *Omitted test item No.
O|1|12345|^1^1|^^^22^PTT^0|R||||||N||||Serum||||||||||O<CR>
The following record is null in case that the test item name of test item No. 78 is “Na”.
O|1|12345|^1^1|^^^40^Na^0|R||||||N||||Serum||||||||||O<CR>
In case that Host rejects an enquiry from analyzer, this record is transmitted.
If the condition code of the enquiry record is not “X”, this record is ignored.
17
V1.06
Example of transmission
C|1|L|Not in Service|G<CR>
Example of transmission
L|1|N<CR>
18
5. Message
Message is a aggregation of records from the message header record to the message
terminator record.
The enquiry message is the message that analyzer inquires to Host about the
measurement order.
Analyzer will transmit inquiry message at the timing of following three cases.
1) When analyzer inquiry button is pressed.
When an operator presses inquiry button, analyzer sends an inquiry message to
host.
Note:
Do not transmit all the measurement orders every time analyzer inquires to Host.
Except the case of changing the contents of measurement order, you do not have to
transmit the measurement order which has been once transmitted.
In case of “All” (batch) request, please transmit measurement order enquiry when
order is decided on host side. The order enquiry transmit when press “Request”
button.
In case of “All” (batch) request, please transmit no measurement order enquiry
when order is indetermination. ( Refer to example of response No.3 : In case of no
measurement order.)
19
V1.06
Note:
In this example of transmission, only the first measurement order record is
effective.
The measurement item of sample ID 123456 is only 1 (GOT).
When requesting measurement items 1 (GOT), 11 (LDH) and 42 (Ca), request
them by a single measurement order record using repeat delimiters.
The followings are the timings for Host to transmit the measurement order messages.
When Host transmits a measurement request to analyzer under the same sample ID for
the second time , analyzer will take different registration action according to the action
code specified in the measurement order record.
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|<CR>
O|1|1234|^1^1|^^^1^GOT^0¥^^^79^K^0|R||||||N||||Serum||||||||||O<CR>
L|1|N<CR>
If transmitted under action code “N”, the registration is over written with the second
measurement request transmitted.
The measurement items of the sample ID 1234 transmitted first are changed into “GOT”,
“K” from “GOT”, “GPT” by the measurement request transmitted for the second time.
However if designated sample is under testing or has already been tested, measurement
request is refused.
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|<CR>
O|1|1234|^1^1|^^^1^GOT^0¥^^^79^K^0|R||||||A||||Serum||||||||||O<CR>
L|1|N<CR>
21
V1.06
If transmitted under action code “A”, the measurement request of the second
transmittion is registered additionally.
The measurement items of the sample ID 1234 transmitted for the first time are
changed into “GOT”, “GPT”, “K” from “GOT”, “GPT” by the measurement request
transmitted for the second time. (“GOT” is over written, “K” is new registeration.)
If the measurement requested first has finished when second request is transmitted, the
second request is registered as the rerun order. The measurement request items, that
will be registered as rerun order items, are “GOT” and “K” of the second transmission.
If the measurement request of the first transmission is under testing, the second
measurement request is refused.
Example of transmission No. 3 : the measurement orders not gathered for each patient
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|<CR>
O|1|1234|^1^1|^^^1^GOT^0¥^^^2^GPT^0¥^^^3^ALP^0|R||||||N||||Serum||||||||||O<C
R>
P|2|<CR>
O|1|1235|^1^1|^^^1^GOT^0¥^^^2^GPT^0¥^^^3^ALP^0|R||||||N||||Serum||||||||||O<C
R>
P|3|123|||Olerud^John||20000101|M<CR>
O|1|1236|^1^1|^^^1^GOT^0|R||||||N||||Serum||||||||||O<CR>
P|4|123|||Olerud^John||20000101|M<CR>
O|1|1237|^1^1|^^^1^GOT^0|R||||||N||||Serum||||||||||O<CR>
22
5.2.3. Test order cancellation
Host can cancel registered measurment request in analyzer.
Note :
Cancellation of measurement request means to cancel all the measurement items for
the sample which measurement request is cancelled.
If sample operation mode is sample ID mode, the sample to be cancelled is to be
the sample of the present sample round in analyzer.
Example of transmittion:
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|<CR>
O|1|QC001|^1^C7|^^^1^GOT^0¥^^^2^GPT^0|R||||||Q||||Serum||||||||||O<CR>
O|2|QC002|^1^C8|^^^1^GOT^0¥^^^2^GPT^0|R||||||Q||||Serum||||||||||O<CR>
O|3|QC003|^1^20|^^^1^GOT^0¥^^^2^GPT^0|R||||||Q||||Serum||||||||||O<CR>
L|1|N<CR>
Example of transmittion:
H|¥^&|||Host^P_1|||||BIOLIS NEO^System1||P|1|20000530192631<CR>
P|1|<CR>
O|1|STAT_001|^1^E1|^^^1^GOT^0¥^^^2^GPT^0|S||||||N||||Serum||||||||||O<CR>
O|2|STAT_002|^1^E2|^^^1^GOT^0¥^^^2^GPT^0|S||||||N||||Serum||||||||||O<CR>
O|3|STAT_003|^1^8|^^^1^GOT^0¥^^^2^GPT^0|S||||||N||||Serum||||||||||O<CR>
L|1|N<CR>
23
V1.06
Example of response
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
P|1|<CR>
O|1|SAMPLE_1|^1^1|^^^1^GOT^0|R||||||N||||Serum||||||||||X<CR>
C|1|I|ASTM_9_4_1|I<CR> … SAMPLE_1 message level abnormal
O|2|SAMPLE_3|^1^1|^^^1^GOT^0|R||||||N||||Serum||||||||||X<CR>
C|1|I|ASTM_9_4_2|I<CR> … SAMPLE_3 sequence number abnormal
O|3|SAMPLE_4|^1^C1|^^^1^GOT^0|R||||||N||||Urine||||||||||X<CR>
C|1|I|A3|I<CR> … SAMPLE_4 sample position abnormal
O|4|SAMPLE_5|^1^C1|^^^1^GOT^0|S||||||Q||||Urine||||||||||X<CR>
C|1|I|A3|I<CR> … SAMPLE_5 sample position abnormal
O|5|SAMPLE_6|^1^E1|^^^1^GOT^0|R||||||N||||Urine||||||||||X<CR>
C|1|I|A3|I<CR> … SAMPLE_6 sample position abnormal
O|6|SAMPLE_7|^1^E1|^^^1^GOT^0|S||||||Q||||Urine||||||||||X<CR>
C|1|I|A3|I<CR> … SAMPLE_7 sample position abnormal
L|1|N<CR>
24
5.3. Measurement Result Message (From analyzer to Host)
The timing for analyzer to transmit the measurement result message is the followings.
Example of transmission
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^30|^^^1^GOT^0¥^^^2^GPT^0¥^^^27^TG^0|R||||||N||||Serum||||||||||O<CR>
R|1|^^^1^GOT^0|21.5143|IU/L|8 TO 38|N||F||||20010530192515<CR>
R|2|^^^2^GPT^0|8.5793|IU/L|4 TO 44|N||F||||20010530192517<CR>
C|1|I|Operator's Comment|G<CR>
R|3|^^^27^TG^0|381.596|mg/dl|50 TO 130|H||F||||20010530192520<CR>
L|1|N<CR>
25
V1.06
H|¥^&|||BIOLIS NEO^System1|||||Host^PC1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^30|^^^1^GOT^0¥^^^2^GPT^0¥^^^27^TG^0|R||||||N||||Serum||||||||||F
<CR>
R|1|^^^1^GOT^0|21.5143|IU/L|8 TO 38|N||F||||20010530192515<CR>
R|2|^^^2^GPT^0||IU/L|4 TO 44|N||X||||20010530192517<CR>
R|3|^^^2^GPT^1||IU/L|4 TO 44|N||X||||20010530192931<CR>
R|4|^^^2^GPT^2|8.6212|IU/L|4 TO 44|N||F||||20010530193551<CR>
R|5|^^^27^TG^0|381.596|mg/dl|50 TO 130|H||F||||20010530192520<CR>
L|1|N<CR>
“State of result” of the measurement result record for the item which is in the waiting
state is transmitted as the waiting result of “I”. (Without measurement result)
When the measurement is completed, the measurement result will be retransmitted
with the state of result, “F”.
H|¥^&|||BIOLIS NEO^System1|||||Host^PC1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^30|^^^1^GOT^0¥^^^2^GPT^0¥^^^27^TG^0|R||||||N||||Serum||||||||||F<CR>
R|1|^^^1^GOT^0|21.5143|IU/L|8 TO 38|N||F||||20010530192515<CR>
R|2|^^^2^GPT^1|230.5687|IU/L|4 TO 44|N||F||||20010530192931<CR>
R|3|^^^2^GPT^2||IU/L|4 TO 44|N||I||||<CR>
R|4|^^^27^TG^0|381.596|mg/dl|50 TO 130|H||F||||20010530192520<CR>
L|1|N<CR>
In case of the measuring failure of sample, the following measurement result comes on.
And the cause of the measuring failure is included in the comment record.
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^30|^^^1^GOT^0¥^^^2^GPT^0|R||||||N||||Serum||||||||||F<CR>
R|1|^^^1^GOT^0|0.02|IU/L|8 TO 38|N||P||||20010530192515<CR>
C|1|I|1|I<CR>
R|2|^^^2^GPT^0||IU/L|4 TO 44|N||X||||20010530192517<CR>
C|1|I|2|I<CR>
L|1|N<CR>
26
5.3.4. Measurement by Simple Operation Mode
Sample which measured by simple operation mode is transferred as sample round No.
991-999.
Sample No. is transferred as Sample No. 1-72.
The result is not transferred automatically.
You have to do transfer (Batch) from R&E screen in normal operation mode.
H|¥^&|||BIOLIS NEO^System1|||||Host^P_1||P|1|20000530192631<CR>
P|1|<CR>
O|1||^1^1601|^^^1^GOT^0¥^^^2^GPT^0|R||||||N||||Serum||||||||||F<CR>
R|1|^^^1^GOT^0|30.0224|IU/L|8 TO 38|N||F||||20010530192515<CR>
R|2|^^^2^GPT^0|25.2268|IU/L|4 TO 44|N||F||||20010530192517<CR>
L|1|N<CR>
27
V1.06
28
Appendix B Analyzer side Configuration
Analyzer control PC needs a LAN port for communication.
If there is not, you need to add on the PC.
Click here.
29
V1.06
Click
Delimiter definition
ComponentDelimiter : ComponentDelimiter example : ^
EscapeDelimiter : EscapeDelimiter example :&
FieldDelimiter : FieldDelimiter example :|
RepeatDelimiter : RepeatDelimiter example :¥
Sender ID
HostName : Sender name example: BIOLIS 50i
System ID
HostSystemID : Receive name example: HOST
30
Appendix C Sample Bar Code
Within ± 10°
・Label direction
Not specified
・Label paste position
Refer to the next page.
Note:
This specification is based on barcode reader specification.
Please note that barcode reader may not read if you edit the barcode to be widened or
narrowed. This barcod reader is a fixed type.
When thermal transfer printer is used to print the barcode, barcode reader may not read
affected by the condition of the printing paper and printer head temperature.
31
V1.06
13
64
Lower limit
55
35
20
14
100mm 75mm
Note:
When affixing the barcode label ensure the bottom of the barcode label is
20mm above the base of the tube.
In case of using a 12.4 x 100 mm tube, the bottom of the barcode labe
should be 35mm from the the base of the tube to avoid turning of tube.
32
Appendix D ASTM Batch Transmission
When transmitting the past data to your host computer, you need to change the
system date on analyzer.
Transmit Procedures:
Click System
Click Date/Print
Send to Host “Sample No.” screen Send to Host “Sample ID(Barcode)” screen
33
V1.06
Input the lowest sample number or first sample ID of data that you wish to transmit in
the Start Sample No. or Start Sample ID box.
Input the highest sample number or last sample ID of data that you wish to transmit in
the End Sample No. or Start Sample ID box.
Please note that it takes some time to transmit data to the host computer. It takes about
3minutes to transmit data for ten samples that have ten test items. If the system is
under operation, it might take more time.
Important Note:
・After the transmission of the past date is finished, follow the same procedure and
ensure the system date is changed back to the current date.
・In case you transmit the results which measured by simple mode, input round
No. 991~999. (Input used “round No.”)
34
D-2 Transmission of Current Data.
Transmit procedures
Click R & E button.
Send to Host “by Sample No.” screen Send to Host “bySample ID(Barcode)” screen
Input the lowest sample number or sample ID that you wish to transmit to the Start
Sample No. or Start Sample ID box.
Input the highest sample number or sample ID that you wish to transmit to the End
Sample No. or Start Sample ID box.
Please note that it takes time to transmit data to the host computer. It takes about
3minutes to transmit data for ten samples that have ten test items. If the system is
under operation, it might take more time.
Inportant Note:
・ After the transmission of the past date is finished, repear the same prodcedures
to return the system date to the current date manually.
・ In case you transmit the results which measured by simple mode, input round
No. 991~999. (Input used “round No.”)
35
V1.06
Code Contents
At measurement of sample or
100000 Noise error of Na electrode :
Calibrator 2
At measurement of sample or
200000 Noise error of K electrode :
Calibrator 2
Noise error of Na and K At measurement of sample or
300000 :
electrodes Calibrator 2
At measurement of sample or
400000 Noise error of Cl electrode :
Calibrator 2
Noise error of Na and Cl At measurement of sample or
500000 :
Electrodes Calibrator 2
Noise error of K and Cl At measurement of sample or
600000 :
electrodes Calibrator 2
Noise error of Na, K and Cl At measurement of sample or
700000 :
electrodes Calibrator 2
S00000 Air detected by the sensor : At measurement of sample
B00000 Air detected by the sensor : At measurement Calibrator 2
Defective drainage detected by At measurement of sample or
E00000 :
the flow senosor Calibrator 2
010000 Noise error of Na electrode : At measurement of Calibrator 1
020000 Noise error of K electrode : At measurement of Calibrator 1
Noise error of Na and K
030000 : At measurement of Calibrator 1
electrodes
040000 Noise error of Cl electrode : At measurement of Calibrator 1
Noise error of Na and Cl
050000 : At measurement of Calibrator 1
Electrodes
Noise error of K and Cl
060000 : At measurement of Calibrator 1
electrodes
Noise error of Na, K and Cl
070000 : At measurement of Calibrator 1
electrodes
0A0000 Air detected by the sensor : At measurement of Calibrator 1
Defective drainage detected by the
0E0000 : At measurement of Calibrator 1
flow sensor
Electric potential drift error at
001000 : At measurement of Calibrator 1
Na electrode.
Electric potential drift error at
002000 : At measurement of Calibrator 1
K electrode.
Electric potential drift error at
003000 : At measurement of Calibrator 1
Na and K electrodes.
Electric potential drift error at
004000 : At measurement of Calibrator 1
Cl electrode.
Electric potential drift error at
005000 : At measurement of Calibrator 1
Na and Cl electrodes.
Electric potential drift error at
006000 : At measurement of Calibrator 1
K and Cl electrodes.
Electric potential drift error at
007000 : At measurement of Calibrator 1
Na, K and Cl electrodes.
Electric potential over range of Na
000100 : Out of range
electrode
36
Code Contents
000002 Pulse over error of return movement to home position (nozzle Z axis)
Note:
There is a possibility to occur multiple errors at the same moment.
For example, 00101 is organized by 00100 and 00001.
This means occuring multiple errors “00100 Na Electric potential drift error” and “00001
Na slope error error”.
The following is each error code.
Na: 1 K : 2 Cl: 4
When multiple items occur errors, the error code is displayed as adding each electrode error
code.
In case of all items, Na + K + Cl > 1 + 2 + 4 = 7
37
V1.06
Request designation
Communication classification
38
Analyzer movement with the combination
↓
Detect sample container. If sample cup or tube detected, measurement will
start. If not, error message will be displayed
↓
Resutlts are transmitted to the host each time when all the results for one
sample has been obtained
Note:
Enquiry is ”ALL” inquiry. Please transmit measurement order enquiry when
order is decided on host side. Please transmit no measurement order enquiry
when order is indetermination. ( Refer to example of response No.3 : In case of
no measurement order.)
Read sample barcode. If barcord value exist in analyzer, proceed to next step
Resutlts are transmitted to the host each time when all the results for one
sample has been obtained.
Note:
Enquiry is “ALL” inquiry. Transmit text without request to analyzer.
When the request has been confirmed, transmit to the analyzer. It can be
received at any time.
39
V1.06
Note:
In case of start button pressed after “sampling stop” status , it inquires from the
next sample number If you press start button after “End of run”, it inquires
from the first number of sample tray.
40
41
V1.06
42