MEASUREMENT SYSTEMS

NUFLO™

NOA-332
Net Oil Analyzer
The Model NOA-332 Net Oil Analyzer is an electronic readout
instrument designed for use with a production or test separator
to measure and record quantities of produced oil, water and
gas. The NOA-332 may be used with two-phase or three-phase
separators. A watercut sensor placed in the oil emulsion line and
determines the percentage of water in the oil, often referred
to as percent BS&W. Flow meter and emulsion signals are read
ten times a second and volumes calculated once a second
allowing the NOA-332 to determine volumes more accurately
than RTUs, PLCs, or other computing devices.

The NOA-332 has three selectable operating modes (timed well


test, cycled well test, and continuous well test) and allows the
operator to program up to sixty-four (64) well names in memory.
All well test data is associated to the well name making well
test accounting simple.

Features
• 512 KB static RAM memory with 5 year lithium Well Test Modes
battery backup
Timed Mode:
• Most recent 256 alarm and event logs plus 366
• Single well test using purge time and test time to define
well test logs maintained in memory
the well test cycle. Manual start and automatic stop of well
• Allows entry of up to 64 well names test cycle.
• Four levels of security access
Cycled Mode:
• Three well testing modes available
• Same as Timed mode except system automatically indexes
• Optional output boards permit automatic well test switching, down the Well Test Table list, testing each well in sequence.
alarm control functions, or volumetric pulse output Specific wells can be easily excluded and testing may begin
• Full SCADA compatibility via Modbus protocol with any well. As each test concludes, the test results are
• Front panel keypad & LCD permit on-site calibration and logged into memory.
viewing of all parameters Continuous Mode:
• System measures gas, emulsion, free-water, and gross • Single well testing using purge time and fixed 24-hour test
volume plus two auxiliary inputs time to define the well test cycle. Testing continues until the
• User selectable analog input assignments operator manually ends the test. Well test data is logged
• Optional switch input board permits alarm and on-off into memory at the end of each 24- hour period and when
control functions, including remote start/stop of well test the test is concluded. Counters track total accumulated
• Industry standard calculations for gas and liquids volumes. Manual start and manual stop of well test cycle.
Calculations 6C, 24A, 24B, or 24C. Pressure proportional volumes of oil and
• Gas Compressibility: PAR Research compensation per API MPMS water. If a Gross Volume flow meter
Project NX-19 (3 component), NX-19 Chapter 11.2.1 or Chapter 11.2.2 is not used, the system calculates
(20 component), AGA-8 1985, AGA-8 • All meter inputs are monitored gross volume as the sum of net oil
(20 component), AGA-8 (3 component) continuously and updated as fluid and total water.
• Orifice Meter: AGA Report No. 3 dumping occurs during test cycle. For
greater accuracy, when calculating
Well Test Table
• Turbine/PD Meter: AGA Report No.
net oil and net water volumes, the • The NOA-332 maintains an
7 for gas, API MPMS Chapter 5.3
Watercut signal is not valid until the internal table of well names and a
for liquids
Emulsion meter registers flow. corresponding purge time, test time,
• Liquid Flow Compensation: and output action (if using the
Meter pulses are accumulated and
Temperature Compensation per API optional DIO board). The operator
compared against %BS&W ten
MPMS Chapter 11.1, Table 6A, 6B, may enter from 1 to 64 names
times each second to determine
and associated values in this table.
The well names may be up to 20
characters long.

Internal Timers
Purge Timer:
• User programmable from 0 to
1440 minutes.
• The purge timer disables the NOA-
332 until flow rates stabilize in the
vessel. When the purge timer counts
down to zero, all internal registers
are reset to zero and the well test
timer begins counting down to zero.
Test Timer:
• ‘Timed Test’ mode and ‘Cycled Test’
mode are user programmable from
0 to 72 hours. When operating in
‘Continuous Test’ mode, the well
test timer is automatically fixed at
24 hours. When the test timer reaches
zero, the NOA-332 stores all data
in memory.
Net Oil Analyzer
(NOA-332)

2
MEASUREMENT SYSTEMS

Well Test Data Options Environmental


The NOA-332 maintains a log of well Capacitance Probe Board: Operating Temperature Range:
test data for each well tested. This log • Provides power to and accepts • With external power supply:
consists of the following: frequency signal from the NuFlo™ -40 °C to +85 °C (-40 °F to +185 °F)
• Well name/number Capacitance Probe. Also provides a • LCD & integral power supply:
• Operator location 4 to 20 mA signal representing -20 °C to +70 °C (-4 °F to +160 °F)
• Time/date of test Percent Watercut for customer use.
Humidity:
• Length of well test Digital I/O Board: • 0 - 90% non-condensing
• Actual volume of oil • Four-channel and eight-channel
Hazardous Location Use:
• 24-hour volume of oil boards available. Four-channel board
• UL listed for USA and Canada,
has 1 amp relay outputs. Eight-
• Actual volume of total water Class 1, Division 2 – requires
channel board has 3 amp FET outputs.
• 24-hour volume of total water external power supply and NuFlo™
• Actual volume of gas Switch Inputs Watercut Monitor
• 24-hour volume of gas • User configurable –- used in
• Actual volume of gross fluid conjunction with digital outputs to
• 24-hour volume of gross fluid provide on/off control features.
• Continuous volume of oil • Remote Start & Stop of well test cycle.
• Continuous volume of total water
Switch Outputs
• Continuous volume of gas
• In alarm output mode, can be used
• Continuous volume of gross fluid to indicate hi/low alarm conditions as
• Flowing average of set by operator; used in conjunction
watercut percentage with alarm inputs to acknowledge
• Average flowing value of auxiliary alarms and provide on/off control.
1 and auxiliary 2 analog inputs • In volumetric pulse output mode,
pulse width and volume per pulse
Software
is defined for use with remote
• Windows 95/98/NT compatible incremental counters to register gas
software. volume, water volume, oil volume
• Quick and easy setup utilizing a and gross volume.
standard PC. • In well test output mode, the digital
• Simple calibration procedure. output can be latched or momentary
• Settings may be changed on line. with user defined pulse interval;
• All changes are logged by outputs are user selectable for binary
operator password. indication of the well number in test.

3
MEASUREMENT SYSTEMS

Inputs Communications Ports Packaging


Flow Meter Inputs: • COM1 – RS-232 for laptop computer • 12-in. x 10-in. x 6-in. NEMA
• Frequency: 1 to 3000 Hz, • COM2 – jumper selectable, RS-232 4 fiberglass enclosure
Amplitude: 20 mV to 10 V p-p DTE or RS-485, Modbus slave port • 2-in. pole mount or
• Frequency inputs can be used to • COM3 – RS-485, Modbus slave bulkhead mount
measure gas, emulsion, free-water, or master • optional stainless steel enclosure
and gross volume
Pressure & Temperature
Data Storage Power Supply Options
Transmitter Inputs: • Battery-sustained memory assures • Various options will require different
• 0 to 5 VDC (includes 1 to 5 VDC) storage of all well test records, alarm power supplies. Your local
• 0 to 20 mA (includes 4 to 20 mA) logs, and event logs. If power outage representative or the factory should
occurs during a well test, the test is always be consulted.
• Two 100 ohm RTD input channels
concluded with all data safely stored
and display will report the power External Power Supply
Water in Oil Meter
failure upon restart. All flow and • Accepts either 230 or 115 volts AC
• 0 to 5 VDC (includes 1 to 5 VDC)
configuration data is saved. power and provides both 12 and 24
• 0 to 20 mA (includes 4 to 20 mA)
volts DC.
• Optional board required for the Display & Keypad • Solar power available – consult
NuFlo™ Capacitance Probe • 24-key, membrane switch, tactile factory
response
Communications
• 4-line by 20-character per line Integral Power Supply
• Modbus and Square D RNIM protocols
alphanumeric liquid crystal display • Power module and/or DC power
• Protected from high-voltage transients board mount inside the NOA-332
• Baud rates up to 19200 enclosure and permit the NOA-332
• Pre and post radio delay supported to be powered by 100 to 240 volts
AC or 10 to 30 volts DC.

WARRANTY-LIMITATION OF LIABILITY: Seller warrants only title to the product, supplies, and materials and that, except to software, the same are free from defects in
workmanship and materials for a period of one (1) year from the date of delivery. Seller does not warranty that software is free from error or that software will run in an
uninterrupted fashion. Seller provides all software “as is”. THERE ARE NO WARRANTIES, EXPRESSED OR IMPLIED, OF MERCHANTABILITY, FITNESS OR OTHERWISE WHICH
EXTEND BEYOND THOSE STATED IN THE IMMEDIATELY PRECEDING SENTENCE. Seller’s liability and Buyer’s exclusive remedy in any case of action (whether in contract, tort,
breach of warranty or otherwise) arising out of the sale or use of any product, software, supplies, or materials is expressly limited to the replacement of such products,
software, supplies, or materials on their return to Seller or, at Seller’s option, to the allowance to the customer of credit for the cost of such items. In no event shall Seller
be liable for special, incidental, indirect, punitive or consequential damages. Seller does not warranty in any way products, software, supplies, and materials not
manufactured by Seller, and such will be sold only with the warranties that are given by the manufacturer thereof. Seller will pass only through to its purchaser of such
items the warranty granted to it by the manufacturer.

MEASUREMENT SYSTEMS
Formerly: NuFlo Measurement Systems • Barton Instrument Systems • Caldon, Inc.

NORTH ASIA EUROPE, MIDDLE EAST USA: Houston, TX • Corpus Christi, TX • Kilgore, TX • Odessa, TX • Dallas, TX • Tulsa, OK
AMERICA PACIFIC & AFRICA Duncan, OK • Denver, CO • Bakersfield, CA • Shreveport, LA • Lafayette, LA
Houma, LA • Pittsburgh, PA • Laurel, MS • Dunbar, WV • Casper, WY • Charleston, WV
1.800.654.3760 65.6737.0444 44.1243.826741
CANADA: Calgary, AB • Edmonton, AB
ms-us @ c-a-m.com ms-asiapacific@c-a-m.com ms-uk@c-a-m.com
Aberdeen, Scotland • Beijing, China • Bognor Regis, UK
I N T E R N AT I O N A L :
HOUSTON HEAD OFFICE: 281.582.9500 • www.c-a-m.com/flo Dubai, UAE • Hassi Messaoud, Algeria • Kuala Lumpur, Malaysia • Singapore

EFM-NOA NF00021 0612

You might also like