Professional Documents
Culture Documents
FIR Selection Guide Melexis
FIR Selection Guide Melexis
FIR Selection Guide Melexis
Chemistry Analyzer
Operation Manual
© 2005-2006 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights
Reserved.
For this Operation Manual, the issued Date is 2006-03 (Version: 1.3).
All information contained in this manual is believed to be correct. Mindray shall not
be liable for errors contained herein nor for incidental or consequential damages in
connection with the furnishing, performance, or use of this manual.
Mindray is responsible for safety, reliability and performance of this product only in
the condition that:
I
WARNING:
It is important for the hospital or organization that employs this
equipment to carry out a reasonable service/maintenance plan.
Neglect of this may result in machine breakdown or injury of human
health.
NOTE:
This equipment is to be operated only by medical professionals trained
and authorized by Mindray or Mindray-authorized distributors.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any
transportation or other charges or liability for direct, indirect or consequential
damages or delay resulting from the improper use or application of the product or the
use of parts or accessories not approved by Mindray or repairs by people other than
Mindray authorized personnel.
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to
Mindray, the following procedure should be followed:
II
Company Contact
III
Preface
WARNING:
The BS-300 Chemistry Analyzer is to be operated only by medical
professionals trained and authorized by Mindray or Mindray-authorized
distributors.
1 System Description
2 Installation
3 Basic Operations
4 Advanced Operations
5 Maintenance
6 Troubleshooting
7 Calculation Methods
AppendixA Specifications
AppendixB Supplies
AppendixC Index
1
Preface
Safety Symbols
The labels attached to some BS-300 panels use symbols with the text to clarify the
meaning of the text. The table below explains the symbols on the labels.
Serial Number
Manufacturer
Date of Manufacture
2
Preface
ON (POWER)
OFF (POWER)
Graphics
All graphics, including screens and printout, are for illustration purposes only and
must not be used for any other purpose.
EC Representative
3
Preface
Safety Precautions
Observe the following safety precautions when using the BS-300 Chemistry Analyzer.
Ignoring any of these safety precautions may lead to personal injury or equipment
damage.
WARNING:
If the system is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
WARNING:
When the POWER is on, users must not open the rear or side cover.
Liquid ingression may lead to electric shock or equipment damage. In
case of liquid ingression, shut off the power supplies and contact
Mindray Customer Service Department or your local distributor.
WARNING:
Do not touch such moving parts as sample probe, reagent probe and
mixing bar, when the BS-300 is in operation.
Do not put your finger or hand into any open part when the BS-300 is
in operation.
WARNING:
Light sent by the photometer lamp may hurt your eyes. Do not stare
into the lamp when the BS-300 is in operation.
If you want to replace the photometer lamp, first switch off the POWER
and then wait at least 30 minutes for the lamp to cool down before
touching it. Do not touch the lamp before it cools down, or you may get
burned.
4
Preface
Preventing Infection
BIOHAZARD:
Inappropriately handling samples may lead to biohazardous infection.
Do not touch the sample, mixture or waste with your hands. Wear
gloves and lab coat and, if necessary, goggles.
In case your skin contacts the sample, follow standard laboratory
safety procedure and consult a doctor.
Certain reagents are strong acid or alkaline. Exercise caution when
using the reagents. In case your skin or clothes contacts the reagents,
wash them off with soap and clean water. In case the reagents spill
into your eyes, wash them off with much water and consult an eye
doctor.
Treating Waste
BIOHAZARD:
Dispose of the waste in accordance with your local or national
guidelines for biohazard waste disposal and consult the manufacturer
or distributor of the reagents for details.
WARNING:
Do not use flammable substance around the BS-300.
5
Preface
Precautions on Use
To use the BS-300 Chemistry Analyzer safely and effectively, pay attention to the
following operation notes.
Intended Use
WARNING:
The BS-300 Chemistry Analyzer (hereinafter referred to as the
BS-300) is a chemistry system that is designed for the in vitro
quantitative determination of clinical chemistries in serum, plasma,
urine or CSF samples. Please consult Mindray first if you want to use
the BS-300 for other purposes.
To draw a clinical conclusion, please also refer to the patient’s clinical
symptom and other test results.
Contraindications: none.
Operator
WARNING:
The BS-300 Chemistry Analyzer is to be operated only by personnel
trained and authorized by Mindray or Mindray-authorized distributors.
Environment
CAUTION:
Please install and operate the BS-300 in an environment specified by
this manual. Installing and operating the BS-300 in other environment
may lead to unreliable results and even equipment damage.
To relocate the BS-300, please contact Mindray Customer Service
Department or your local distributor.
CAUTION:
Electromagnetic noise may interfere with operation of the BS-300. Do
not install devices generating excessive electromagnetic noise around
the BS-300. Do not use such devices as mobile phones or radio
transmitters in the room housing the BS-300. Do not use other CRT
displays around the BS-300.
Do not use other medical instruments around the BS-300 that may
generate electromagnetic noise to interfere with their operations.
6
Preface
CAUTION:
Operate the BS-300 strictly as instructed by this manual. Inappropriate
use of the BS-300 may lead to unreliable test results or even
equipment damage or personal injury.
Before using the system for the first time, run the calibration program
and the QC program to make sure it is in a proper state.
Be sure to run the QC program every time you use the system,
otherwise the result man be unreliable.
Do not open the sample disk or reagent disk cover when the BS-300 is
operating.
The RS-232 port on the analyzing unit is to be used for connection
with the operation unit only. Do not use it for other purposes. Only use
the supplied cable for the connection.
The operation unit is a personal computer with the BS-300 control
system installed. Installing other software or hardware on this
computer may interfere with the system operation. Do not run other
software when the system is working.
Do not use this computer for other purposes. Inappropriate use of this
computer may introduce computer virus, which may spread through
floppy disks, software or network, into the system.
Do not touch the display, mouse or keyboard with wet hands or hands
with chemicals on them.
Don’t place the POWER to ON again within 10 seconds since placing
it to OFF, otherwise the system may enter the protection status. If it
does so, place the POWER to OFF and place it to ON again.
CAUTION:
Maintain the BS-300 strictly as instructed by this manual. Inappropriate
maintenance may lead to equipment damage or personal injury.
To wipe off dust from the BS-300 surface, use a soft, clean and wet
(not too wet) cloth, soaked with soap water if necessary, to clean the
surface. Do not use such organic solvents as ethanol for the cleaning.
After cleaning, wipe the surface dry with dry cloth.
Switch off all the powers and disconnect the power plug before
cleaning. Take necessary measures to prevent water ingression into
the system, otherwise it may lead to equipment damage or personal
injury.
Replacements of such major parts as the photometer, sample probe,
reagent probe, mixing bar and syringe plungers must be followed by a
calibration.
7
Preface
Samples
CAUTION:
Use serum samples that are completely separated from blood clots or
urine samples that are free from suspended matter. If fibrin exists in
the serum samples or suspended matter exist in the urine samples,
the sample probe may be blocked.
Medicines, anticoagulants or preservative in the samples may lead to
unreliable test results.
Hemolysis, jaundice or chylomicron in the samples may lead to
unreliable test results, so sample blanks are recommended
CAUTION:
Use proper reagents, calibrators and controls in the BS-300.
Select appropriate reagents according to the performance
characteristics of the BS-300. Consult the reagent supplier or
Mindray-authorized distributor for details when you are not sure about
your reagent choice.
8
Preface
Setting up BS-300
CAUTION:
You need to set up the BS-300 strictly as instructed by this manual
before using it. To define such parameters as sample volume, reagent
volume and wavelength, consult instructions of reagents.
Backing Up Data
NOTE:
The BS-300 automatically backs up the data to the built-in hard disk.
However, data loss is still possible due to mis-deletion or physical
damage of the disk. Mindray recommends you regularly backs up the
data to such medium as CDs.
NOTE:
Refer to their user’s manuals for details.
External Equipment
WARNING:
External equipment connected to the system, such as PC and printer,
shall be consistent with IEC 60950 or EN 60950.
9
Contents
Preface.......................................................................................................................................... 1
Who Should Read This Manual ............................................................................................. 1
What Can You Find in This Manual........................................................................................ 1
Conventions Used in This Manual ......................................................................................... 1
Safety Precautions ................................................................................................................. 4
Precautions on Use................................................................................................................ 6
Contents ........................................................................................................................................ i
ii
Contents
6 Troubleshooting................................................................................................................ 6-1
6.1 Classification of Error Messages ......................................................................... 6-1
6.2 Corrective Measures ........................................................................................... 6-4
iii
1 System Description
This chapter includes the following two sections:
Hardware Introduction
Software Introduction
NOTE:
Not all the parameters the reagents claim capable of testing can be
tested on the BS-300. Consult the reagent suppliers for detailed
information.
Figure 1-1 Overview of the analyzing unit and the operation unit
1-1
System Description
1-2
System Description
Upper Cover
Mounting Hole
Lower Door
Connect the power cord to the power plug shown in Figure 1-3.
DEIONIZED WATER: If you choose to feed deionized water from the back of
the analyzing unit, pull the deionized water tubing into the unit from here;
D-SENSOR: If you choose to feed deionized water from the back of the
analyzing unit, pull the cable of the deionized water level detector into the unit
from here;
1-3
System Description
WASTE: If choose to discharge waste from the back of the analyzing unit, pull
the waste tubing out of the unit from here;
W-SENSOR: If you choose to discharge the waste from the back of the
analyzing unit, pull the cable of the waste level detector out of the unit from
here;
COM: It is the serial port for communication with the operation unit;
RESET: You can press this button to restart the analyzing unit when it stops
responding.
WARNING:
Press the RESET button only when the analyzing unit stops
responding. Do not press this button when the unit is working normally.
Otherwise, it may lead to system failure, unreliable analysis results or
serious unexpected consequences.
The sample disk (as Figure 1-5 shows) holds sample tubes and rotates
(counter-clockwise only) the designated tube to the position for the sample probe to
aspirate.
The sample disk is composed of two circles – an inner circle and an outer circle. It
provides 60 tube positions, 30 on the inner circle and 30 on the outer circle. Every
tube position has an identification number. Positions 1-45 are for samples, E1-E5 for
emergency samples, S1-S6 for calibrators, C1-C3 for controls, and W for distilled
water. All the other positions except the position W can be used for samples, too.
The sample disk can hold six types of tubes – micro sample cup, collection tube, Φ
12mm×68.5mm, Φ12mm ×99mm, Φ12.7mm×75mm and Φ12.7mm×100mm.
To install the sample disk, keep the handle at the vertical position, align the central
hole of the handwheel to the rotor, gently lower the sample disk all the way down
(rotate the disk a bit as necessary) and move the handle back to the horizontal
1-4
System Description
position to secure the disk to the rotor. When you have finished installing the disk, grab
the handwheel and lightly pull it upward. If the disk remains secured, it means it is
well installed.
To remove the sample disk, first shift the handle from the horizontal position to the
vertical position. Then grab the handle or handwheel and pull the disk upward to
remove it from the rotor.
To load sample tubes, insert the tube into the tube holder until the bottom of the tube
contacts the groove on the tube rack.
To remove sample tubes, grab the tube and pull it upward to remove it from the tube
holder.
WARNING:
Before installing/removing the sample disk, make sure the analyzing
unit is turned off and the sample disk has stopped.
Before installing/removing the sample tubes, make sure the analyzing
unit is turned off and the sample disk has stopped.
Do not use sample tubes other than the specified.
Before running the analyzing unit, make sure that the sample disk
cover is closed and the round red mark on the cover is aligned with its
counterpart on the panel. Otherwise the sample probe may be
damaged.
If samples spill in the sample compartment or on the sample disk, be
sure to wipe the spills with cloth soaked with water or disinfector after
turning off the analyzing unit.
The sample dispenser is composed of a sample probe, probe arm, probe rotor
(Figure 1-6), syringe (Figure 1-7) and a fluid system. It aspirates certain amount of
sample from the designated sample tube and dispenses the aspirated sample to the
designated reaction cuvette.
Open the middle door of the analyzing unit and you will see the sample syringe (right)
and the reagent syringe (left). The sample syringe can aspirate/dispense 3-45µl of
sample and its minimum scale is 0.5µl.
1-5
System Description
Sample Syringe
To dispense the sample, the sample probe first moves to the sample tube to aspirate
certain amount of the sample, then to the reaction disk to dispense the aspirated
sample to a reaction cuvette and finally to the wash well for cleaning.
The sample probe can automatically detect the sample level. When the sample
probe contacts obstacles in the vertical direction, its safeguard will automatically
protect the probe from damage.
WARNING:
When the analyzing unit is running, do not place any part of your body
or any obstacle in the route the arm moves. Otherwise, it may lead to
personnel injury or equipment damage.
1-6
System Description
The reagent disk (as Figure 1-8 shows) holds the reagent bottles and rotates
(counter-clockwise only) the designated bottle to the position for the reagent probe to
aspirate.
The refrigerator keeps the reagents in a low temperature environment to keep them
stable and minimize volatilization.
The reagent disk has two circles, inner circle and outer circle. There are 25 reagent
positions in each circle and the reagent disk has totally 50 reagent positions. The
reagent disk can hold the Mindray inner-circle bottles, Mindray outer-circle bottles,
Hitachi 7060 bottles and Hitachi 7170 bottles.
The refrigerator can run 24 hours a day to keep the temperature at 4-10˚С.
NOTE:
The refrigerator has a power supply independent of that of the
analyzing unit. It will be put into service once the POWER is turned on.
To install the reagent disk, keep the handle at the vertical position, align the central
hole of the handwheel to the rotor, gently lower the reagent disk all the way down
(rotate the disk a bit as necessary) and move the handle back to the horizontal
position to secure the disk to the rotor. When you have finished installing the disk, grab
the handwheel and lightly pull it upward. If the disk remains secured, it means it is
well installed.
To remove the reagent disk, first shift the handle from the horizontal position to the
vertical position. Then grab the handle or handwheel and pull the disk upward to
remove it from the rotor.
1-7
System Description
To load reagent bottles, insert the bottle into the bottle holder until the bottom of the
bottle contacts the groove on the bottle rack.
To remove the reagent bottle, grab the bottle and pull it upward to remove it from the
bottle holder.
WARNING:
Before installing/removing the reagent disk, make sure the analyzing
unit is turned off and the sample disk has stopped.
Before installing/removing the reagent bottles, make sure the
analyzing unit is turned off and the sample disk has stopped. When
installing the bottle, make sure the bottle bottom is in full contact with
the bottle rack.
Before running the analyzing unit, make sure that the reagent disk
cover is closed and the round red mark on the cover is aligned with its
counterpart on the panel. Otherwise the sample probe may be
damaged and cooling capacity of the refrigerator will be degraded.
If reagents spill in the reagent compartment or on the reagent disk, be
sure to wipe the spills with cloth soaked with water or disinfector after
turning off the analyzing unit.
The reagent dispenser is composed of a reagent probe, probe arm, probe rotor
(Figure 1-9), syringe (Figure 1-10) and fluid system. It aspirates certain amount of
reagent from the designated reagent bottle and dispenses the aspirated reagent to
the designated reaction cuvette. The reagent probe arm can preheat the aspirated
reagent.
Open the middle door of the analyzing unit and you will see the reagent syringe
(right) and reagent syringe (left). The reagent syringe can aspirate/dispense
30-450µl of reagent and its minimum scale is 1µl.
1-8
System Description
To dispense the reagent, the reagent probe first moves to the reagent bottle to
aspirate certain amount of the reagent, then to the reaction disk to dispense the
aspirated reagent to a reaction cuvette and finally to the wash well for cleaning.
The reagent probe can automatically detect the reagent level. When the reagent
probe contacts obstacles in the vertical direction its safeguard will automatically
function to protect the probe from damage.
WARNING:
When the analyzing unit is running, do not place any part of your body
or any obstacle in the route the arm moves. Otherwise, it may lead to
personnel injury or equipment damage.
1.1.1.5 Mixer
The mixer is composed of a mixing bar, bar arm and rotor, as Figure 1-11 shows.
The mixing bar thoroughly stirs the reaction mixture (reagents and sample) in the
reaction cuvette.
1-9
System Description
For the single-reagent test, the mixer starts to work after the sample is dispensed
into the reaction cuvette. For the double-reagent test, the mixer starts to work after
the sample or the second reagent is dispensed into the reaction cuvette. The mixing
bar moves to a position above the reaction disk, lowers into the cuvette, and rotates
to mix the mixture in the cuvette. After that, the bar moves to the wash well for
cleaning and de-watering.
This loader loads new cuvettes to the reaction disk and dumps used cuvettes into a
bucket. The feeder is located in the cuvette compartment and transfers cuvettes to
the manipulator, which grabs the cuvettes and loads them to the reaction disk and
then unloads the used cuvettes and dumps them into the used-cuvette bucket. Open
the lower door of the analyzing unit and you will see the used-cuvette bucket in the
middle.
NOTE:
Normally, the used-cuvette bucket can hold 80 segments of reaction
cuvettes (10 cuvettes per segment). Be sure to empty the bucket in
time so that excessive cuvettes will not overflow and contaminate the
cabinet.
Disposable cuvettes are adopted as the reaction cuvettes. Ten cuvettes are bound
together as one cuvette segment. The cuvette compartment can house 30 cuvette
segments, which are sufficient for 300 continuous tests.
When all the cuvettes of a cuvette segment have been used, the reaction disk
rotates the segment to the unloading position, and then the manipulator dumps the
used segment into the used-cuvette bucket and load the new segment. When a
segment is taken from the cuvette compartment, the cuvette feeder pushes the next
segment to the loading position.
1-10
System Description
When there are less than 10 segments in the compartment, the analyzing unit,
though keeping working, will give an alarm. Seeing the alarm, you should, if you still
have more tests to do, add more segments to the compartment.
When all the segments in the compartment are used, the analyzing unit will stop first
loading cuvette segments, and then stop dispensing samples when all the segments
on the reaction disk are used. But it will keep working until all the scheduled tests are
done.
Present the cuvette segment to the entry of the cuvette feeder as Figure 1-12 shows
and gently push it toward inside. Load all the segments one by one until they are all
loaded. When you are done, click the Load Cuvette button (refer to 4.1.4.2 Reaction
Disk tab for details) to command the cuvette feeder to press the newly added
segments forward. Be sure not to load the segments as Figure 1-13 shows.
NOTE:
When loading cuvette segments to the entry of the cuvette
compartment, be sure to keep them upright. Remove the inclined
segments immediately upon seeing them and reload them the right
way.
Be sure to load the segments to the cuvette compartment in the right
direction (Figure 1-12). Reversely loaded segments (Figure 1-13) may
jam the loader and halts the analyzing unit.
Do not click the Load Cuvette button before completing adding all the
cuvette segments to the compartment. Otherwise, the loading
operation may fail.
1-11
System Description
BIOHAZARD:
Be sure to load the used-cuvette bucket to the designated position on
the bottom plate. Otherwise, the used-cuvette may not be dumped into
the bucket and the spilled reaction mixture may contaminate and
corrode the cabinet.
Be sure to wear gloves, goggles and protective clothing when loading
or removing the used-cuvette bucket.
Be sure to dispose of the used-cuvette in compliance with the local
regulations.
The reaction disk holds the reaction cuvettes, in which the sample reacts with the
reagents and colorimetric readings are taken.
The reaction disk can hold 8 cuvette segments (80 cuvettes). The cuvettes adopted
are
Disposable;
5mm×6mm×25mm (5mm optical path);
750µl (capable of holding 180-500µl of the reaction mixture).
1-12
System Description
BIOHAZARD:
Be sure to dispose of the used cuvettes in compliance with the local
regulations.
CAUTION:
The reaction cuvettes should not be re-used. Otherwise, the system
performance may be degraded.
During the analyzing process, the reaction disk rotates (counter-clock wise only) to
cuvette-loading position, sample-dispensing position, reagent-dispensing position, or
mixing position as needed. The colorimetric readings are taken when the specified
cuvette passes through the optical axis.
Manipulator
One
Segment
Position to Mixing
(28#)
Position to Dispensing
Position to Dispensing Sample (23#)
Reagent (1#)
The photometric system measures the absorbance of the reacting liquid in the
reaction cuvettes. It consists of a measurement optical system and a reference
optical system, the former providing 9 monochromatic lights and the latter
compensating the former.
1-13
System Description
Specifications
9 optical paths with fixed interference filters.
Wavelengths: 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm,
700nm
Half band-width: ≤12nm
Wavelength accuracy: ±2nm
Measurement range: -0.1-5.0 Abs
Lamp: 12V 50VA tungsten-halogen
Movement
The measurement optical system consists of 9 optical paths. The absorbance is
taken when the reaction cuvette rotates to the correspond path. The BS-300 can
measure the absorbance of 9 cuvettes simultaneously.
1-14
System Description
37
Pull-down menu
System status
Menu area Header area Title bar Shortcut buttons
area
Display area
Header area
Displays the names of the operating software and the manufacturer.
Title bar
Displays the title of the current display area.
Menu area
Displays: [Routine], [Parameters], [Data], [System], [Help]. You can move the
mouse to any of the listed menu and click the left button of the mouse (hereinafter
referred to as click) to display its pull-down menu.
Pull-down menu
You can click the listed item of the pull-down menu to perform certain operation.
Small buttons
Click the small button to view the user guide. You can also click Help → Guide
to view the user guide. Refer to 4.5.2 Guide for details.
Click the small button to exit the operating software. You can also click Routine
→ Exit to exit the software. Refer to 4.1.8 Exit for details.
Shortcut buttons
, , , , , , , , . You
can click any of the buttons to perform certain operation quickly.
Display area
1-15
System Description
Period area
Displays how many periods (every period lasts 12 seconds) of test have passed
since the beginning of the first test after the latest power-on.
Time area
Displays current date and time.
Operator area
Displays the current operator who operates the system.
Tab
See the figure below for examples of the tab. Click the tab and you can access the
screen that it indexes.
1-16
System Description
Tab
Combo box
Click “▼” and a pull-down list will display, as the figure below shows. Click the
desired item to select it.
Combo box
Option button
Click the option button to select the option it represents. Note that for a given group
of option buttons, you can only select one of them.
Option button
Check box
Click the check box to select the option it represents and click it again to de-select it.
Note that for a given group of check boxes, you can choose more than one of them
at one time.
1-17
System Description
Check box
Edit box
You can enter characters from the keyboard in the edit box.
Edit box
Button
Click a button and you can access the function it indexes.
Scroll bar
When the contents to be displayed are too many for one screen, you can use the
scroll bar to see the hidden contents. Move the cursor to the scroll bar, press left
button of the mouse and hold it, then you can drag the scroll bar upward or
downward to see the hidden contents. Or you can press the key such as Page Up,
Page Down,↑or↓on the keyboard to see the hidden contents.
Scroll bar
1-18
System Description
Double-clicking
In this manual, “double-clicking” refers to moving the pointer of the mouse to the
desired item and click the left button of the mouse twice swiftly.
1-19
2 Installation
WARNING:
The BS-300 should be installed by Mindray-authorized personnel only.
To facilitate the installation, you should prepare a place meeting the requirements
specified in 2.2 Installation Requirements. If you need to move the BS-300 to
another site, please contact Mindray Customer Service Department or your local
distributor, who are the appropriate people for the moving job.
2.1 Unpacking
When you receive your BS-300, carefully inspect the package. If you see any signs
of mishandling or damage, file a claim immediately with Mindray Customer Service
Department or your local distributor.
After opening the package, check the delivered goods against the packing list as
well as the appearance of the BS-300. If you find anything missing or damaged, alert
Mindray Customer Service Department or your local distributor immediately.
CAUTION:
Make sure the BS-300 is installed in a place meeting the following
requirements. Otherwise, it will not perform as promised.
CAUTION:
The BS-300 radiates heat from its rear side when operating. A
well-ventilated environment helps keep the room temperature stable.
Use ventilation equipment if necessary. But if so, be sure not to
expose the BS-300 to the direct draft that may lead to unreliable
results.
2-1
Installation
WARNING:
When the forward or backward gradient is more than 8 degrees, the
BS-300 may tip over. Be sure to take proper measures when
transporting and storing it.
WARNING:
Make sure the power socket is grounded correctly. Improper grounding
may lead to electric shock and/or equipment damage.
Be sure to connect the BS-300 to a power socket that meets the
above-mentioned requirements and has a proper fuse installed.
2-2
Installation
CAUTION:
Storing the BS-300 in an environment other than the specified may
damage its fluidic system.
Operating the BS-300 in an environment other than the specified may
lead to unreliable test results.
CAUTION:
When used in conditions other than the specified, the BS-300 may not
provide reliable test results. If the temperature or relative humidity
does not meet the above-mentioned requirements, be sure to use
air-conditioning equipment.
BIOHAZARD:
Be sure to dispose of the waste according to the local regulations.
CAUTION:
The water to be used on the BS-300 must meet the requirements of
the CAP Type II water. Insufficiently purified water may result in
misleading measurement.
2-3
Installation
WALL
Minimum500 Maximum2500
Operation
Unit
710 Analyzing Unit
980
F
Minimum500
R
O
N
T
Minimum500 Minimum500
Unit: mm
BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below checks.
2-4
Installation
7 Plug the green connector to its counterpart on the analyzing unit and turn
the connector clockwise until secure.
Pickup Tube
8 Plug the red connector to its counterpart on the analyzing unit and turn
the connector clockwise until secure.
9 Plug the BNC connector to its counterpart on the analyzing unit and turn
the connector clockwise until secure.
10 Close the lower door.
Pin
Waste Tap
Waste Tube
8 Plug the BNC connector to its counterpart on the analyzing unit and turn
the connector clockwise until secure.
9 Close the lower door.
2-5
3 Basic Operations
This chapter provides step-by-step procedures for operating the BS-300 for basic
tasks.
Power-on
YES
NO
Setting Up the
BS-300 Analyzer
YES
NO
Reagent Blank
YES
Calibration NO
QC
Samples
YES
Printing Results
Shutdown
3-1
Basic Operations
BIOHAZARD:
Wear gloves, goggles and protective clothing when performing the
below checks.
Check the power supply and make sure it can supply proper voltage for the
analyzer.
Check the connections among the analyzing unit, operation unit and printer.
Make sure the connections are right and secure. Check the power cords of the
analyzing unit, operation unit and printer and make sure they are well
connected to power sockets.
Check and make sure sufficient printing paper is prepared for the printer.
Check the sample and reagent syringes for leakage as instructed by 5.2.1
Checking Sample/Reagent Syringes.
Check the sample and reagent probes and make sure there is no fluid hanging
on the probe tips. If there is, please contact Mindray Customer Service
Department or the distributor.
Check the connections of the deionized water and wastewater as instructed by
5.2.6 Checking Connection of Deionized Water and 5.2.7 Checking
Connection of Wastewater.
Check the sample probe and make sure it is neither dirty nor bent. If it is dirty,
wash it as instructed by 5.3.1 Washing Sample Probe; if it is bent, replace it as
instructed by 5.6.3 Replacing Sample Probe.
Check the reagent probe and make sure it is neither dirty nor bent. If it is dirty,
wash it as instructed by 5.3.2 Washing Reagent Probe; if it is bent, replace it
as instructed by 5.6.4 Replacing Reagent Probe.
Check the mixing bar and make sure it is neither dirty nor bent. If it is dirty, wash
it as instructed by 5.3.3 Washing Mixing Bar; if it is bent, replace it as
instructed by 5.6.5 Replacing Mixing Bar.
Check the cuvette compartment and make sure there are sufficient cuvettes in it.
If the cuvettes are not sufficient, add more cuvettes as instructed by 1.1.1.6
Reaction Cuvette Loader.
CAUTION:
You must prepare enough cuvettes for the upcoming analysis. If you
find the cuvettes are not enough, add more immediately.
Sometimes the new cuvettes are not as smooth as they should be.
Check the surface of the cuvettes for bumps and take necessary
measures to remove the found bumps before loading them to the
compartment. Otherwise, they may jam the analyzing unit.
3-2
Basic Operations
Check and make sure there is enough deionized water in the deionized water
tank as instructed by 5.2.8 Checking Remaining Deionized Water.
CAUTION:
You must prepare enough deionized water. If you find the deionized
water is not enough, add more deionized water to the tank.
Check the waste tank and make sure it is empty. If not, empty the tank as
instructed by 5.2.9 Emptying Waste Tank.
Check the used-cuvettes bucket and make sure it is empty. If not, empty the
bucket as instructed by 5.2.10 Emptying Used-Cuvettes Bucket.
3.2.2 Power-on
Power up the BS-300 in the sequence presented below:
NOTE:
Ensure the hibernation function of the display is unavailable:
After entering the display properties screen, select Never for Turn off
monitor, Turn off hard disks and System standby on the Power
Schemes tab and do not select Enable hibernation on the Hibernate
tab.
After you have logged on the WINDOWS operating system, you may
The control system will do the initialization procedure in the sequence presented
below:
The system does the startup and warm-up procedures in the sequence presented
below and the analyzer will then enter the standby state:
NOTE:
You need to enter proper username and password to access the
control system. Use the initial username and password (username:
administrator; password: bs300) if this is the first time the control
system being used and then you may edit your own username and
password as instructed by 4.4.4 User.
CAUTION:
You may start the analysis once the initialization is finished. However,
Mindray recommends you wait until the temperature and photometric
system is fully stabilized, which is when the status box displays Idle
and the analyzer has been turned on for 30 minutes.
You must complete all the following settings if this is the first time the analyzer being
used.
Before requesting the tests, you must finish the following settings:
To set the options regarding the basic parameters of the system, refer to 4.4.2
Setup.
To set the options regarding the hospital information, refer to 4.4.3 Hospital.
3-4
Basic Operations
To set the options regarding test parameters and reference, refer to 4.2.1.1
Parameters tab and 4.2.1.2 Expected tab.
To set the options regarding calibrators and calibration, refer to 4.2.3.1
Calibrator tab and 4.2.3.2 Calibration tab.
To set the options regarding controls, QC rule and QC test, refer to 4.2.4.1
Controls tab, 4.2.4.2 QC Rules tab and 4.2.4.3 QC tab.
To set the options regarding the reagent parameters, refer to 4.2.2.1 Reagent
Disk tab.
To set the options regarding the carryover information among tests, refer to
4.2.1.5 Carryover tab.
To set the options regarding the printing parameters, refer to 4.4.7 Print.
Load reagent bottles to their assigned positions on the reagent disk. Open the
bottles.
Load distilled water to position W on the sample disk. Ensure enough distilled water
is prepared.
Load distilled water to position 49 on the reagent disk. Ensure enough distilled water
is prepared.
Check remaining reagents if necessary. Refer to 4.2.2.2 Leftover tab for details.
3.3.1 Calibration
You must calibrate if this is the first time the analyzer being used.
CAUTION:
You need to run the calibration again when you change reagent lots,
test parameters, lamp or other conditions.
3-5
Basic Operations
NOTE:
Ensure the calibrators are loaded to the corresponding positions on
the sample disk.
To view the results at the Status→Calib. Status, refer to 4.1.4.4 Calib. Status
tab.
To view the results at the Status→Sample Disk, refer to 4.1.4.1 Sample Disk
tab.
To view the results at the Calibration→Calib. Data, refer to 4.2.3.3 Calib. Data
tab.
3.3.2 QC
To request QCs, refer to 4.1.3 QC.
After requesting QCs, you should load corresponding controls to their assigned
positions on the sample disk.
To view the results at the Status→QC Status, refer to 4.1.4.5 QC Status tab.
To view the results at the Status→Sample Disk, refer to 4.1.4.1 Sample Disk
tab.
To view the results at the QC→Real-time/Daily/Day-to-day, refer to 4.2.4.4
Real-time tab, 4.2.4.5 Daily tab or 4.2.4.6 Day-to-day tab.
3.3.3 Samples
To request samples, refer to 4.1.1 Samples.
You can request a STAT sample run in the way similar to an ordinary run, except that
you have to tick the check box to the right of STAT, as Figure3-1 shows. The
requested STAT run will be inserted into the current work list as Waiting and has
higher priority than other tests.
3-6
Basic Operations
After requesting, load the samples to their corresponding positions on the sample
disk.
CAUTION:
Ensure the samples are loaded to correct positions. Otherwise, you
cannot acquire correct results.
To view the results at the Status→Test Status, refer to 4.1.4.3 Test Status tab.
To view the results at the Status→Sample Disk, refer to 4.1.4.1 Sample Disk
tab.
To view the results at the Data→Result→Test Data, refer to 4.3.4.1 Test Data
tab.
CAUTION:
Sample run results can only be edited by authorized personnel.
To edit one or more sample run results, refer to 4.1.4.3 Test Status tab or 4.3.4.1
Test Data tab for details.
To revise one or more sample run results, refer to 4.1.4.3 Test Status tab or 4.3.4.1
Test Data tab for details.
3-7
Basic Operations
3.4.2 Shutdown
After exiting the Windows operating system, switch off the following powers in the
presented order:
NOTE:
The refrigerator still functions after the ANALYZING UNIT POWER is
placed to OFF. To shut down the refrigerator, place the POWER to
OFF.
BIOHAZARD:
Exercise with caution. Wear gloves, goggles and protective clothing.
3-8
4 Advanced Operations
The chapter presents an introduction of the operating software of the BS-300 by
menus.
The main interface of the operating software has five major menu groups:
Routine
Parameters
Data
System
Help
4.1 Routine
Routine Samples
Calibration
QC
Status
Start
Probe Stop
Stop
Exit
Emergency Exit
4.1.1 Samples
Click Routine → Samples, or press F1 to enter the Request Sample Runs screen,
as Figure4-1 shows. You can also enter the Request Sample Runs screen by
4-1
Advanced Operations
The Tests zone lists all available tests. You cannot request the tests followed by a
forbidden mark, such as . The double-reagent tests, whose reagent
positions have already been set, are yellow-framed; the double-reagent tests, whose
reagent positions have not been set yet, are gray-framed. The selected test is
followed by a green tick.
4-2
Advanced Operations
4-3
Advanced Operations
NOTE:
Besides the individual requirements for the entered values of Dilution
and Sample Vol., you must also guarantee their product is between
180-450.
To request samples
You may also request a pre-defined profile or profiles and the selected
profile or profiles will also be followed by a green tick.
4-4
Advanced Operations
NOTE:
Routine sample can be requested on any virtual sample disk.
STAT sample can only be requested on the default virtual sample disk.
After the system is powered on, the sample disk 1 is the default at that
time.
If there are any sample runs in Waiting or Running status on a sample
disk, the disk is the default at that time.
If the system just finishes the runs on a sample disk, the disk is the
default at that time.
If you request a sample whose ID has been used on the current day
(the same ID, but the different index) and you’ve set the Bar Code and
Sample (sample type) for it, the system will consider the Bar Code
and Sample of the former one as those of the current sample. You can
change the two parameters only at the Patient Information screen.
NOTE:
Click to go to the previous sample run.
3 After you have finished browsing samples, click the Browse button again
to quit the browsing mode.
NOTE:
Click to go to the previous sample run.
4-5
Advanced Operations
NOTE:
You can edit routine sample information here only if the
sample is in Requested status.
You cannot edit STAT sample information here because the
STAT sample is in Waiting status after it is requested.
Disk, ID and Replicates can’t be edited.
You can edit Sample and Bar code only at the Patient
Information screen. Refer to 4.3.3 Patient Info. for details.
5 If you do not want to change the tests for the sample, go directly to the
next step.
Otherwise, select tests and profiles in the Tests area or Profiles area.
If you want to add a test or a profile, select it; if you want to delete a test
or a profile, de-select it.
NOTE:
You can change the tests or profiles for the sample in any
status.
You can only delete a test or a profile for the sample in
Requested status.
6 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.
7 Repeat from step 2 to edit another sample run; or click the Browse button
again to quit the browsing mode.
NOTE:
Click to go to the previous sample run.
4-6
Advanced Operations
NOTE:
You can only delete the routine sample that is in Requested
status here.
You cannot delete the STAT sample here because the STAT
sample is in Waiting status after it is requested. If you want
to delete a requested STAT sample, refer to 4.1.4.1 Sample
Disk tab or 4.1.4.3 Test Status tab for details.
4 Repeat from step 2 to delete another sample run; or click the Browse
button again to quit the browsing mode.
4.1.2 Calibration
At the Request Calibration Runs screen, all the available tests are listed. The tests
followed by forbidden marks, such as , are those whose calibration
parameters have not been set yet and cannot be requested for calibrations. The
yellow-framed tests are the double-reagent tests whose reagent positions have
already been set. The gray-famed tests are the double-reagent tests whose reagent
positions have not been set yet. The selected test will be followed by a green tick.
4-7
Advanced Operations
You can select the desired sample disk (virtual disk) from the pull down list of the
Disk filed. The default sample disk is Sample Disk1. The adopted calibration rule
and calibrator will be displayed in the Rule and Calibrator fields respectively.
1 Click “▼” below Sample Disk and select the desired sample disk from the
pull down list.
2 Click the desired test to select it; click it again to de-select it. The select
test will be followed by a green tick. You can select multiple tests for a
calibration run.
CAUTION:
You must run the calibration if this is the first time the analyzer using
Endpoint or Fixed-time method for double-reagent tests. For the
Kinetic method, you can directly input the Factor without running the
calibration first (see 4.2.1.1 Parameters tab).
You need to run the calibration when you have changed the reagents,
reagent lots, test parameters, lamp or other analysis conditions.
Before calibrating a double-reagent test that uses Endpoint method,
the system will run reagent blank test for it automatically.
4.1.3 QC
Click Routine → QC, or press the F3 to enter the Request Reagent QC Runs
screen, as Figure4-3 shows. You can also enter the Request QC Runs screen by
4-8
Advanced Operations
The screen lists all available tests. The tests whose reagent positions have not been
set yet are followed by a forbidden symbol, such as , and you cannot
request to run them. The double-reagent tests, whose reagent positions have
already been set, are yellow-framed; the double-reagent tests, whose reagent
positions have not been set yet, are gray-framed. The selected test is followed by a
green tick.
When any test is selected, its corresponding control information will be displayed in
the Controls filed. The default sample disk (virtual disk) is disk 1 and you should
select the appropriate sample disk to the actual arrangement.
NOTE:
If the auto QC function is enabled (see 4.4.2.1 Routine Settings tab
for details) and the QC interval is not 0 (see 4.2.4.3 QC tab for
details), the analyzer will automatically insert QC runs among sample
runs and you do not have to request additional QC runs.
To request a QC run
1 Click “▼” below Sample Disk and select the desired sample disk from the
pull down list.
4-9
Advanced Operations
2 Click the desired test to select it; click it again to de-select it. The select
test will be followed by a green tick. You may select multiple tests for one
run.
4.1.4 Status
Click Routine → Status or press F4 to enter the Worklist Status screen, as Figure
4-4 shows. You can also enter the Worklist Status screen directly by clicking the
button.
4-10
Advanced Operations
Sample Disk
Reaction Disk
Test Status
Calib. Status
QC Status
The Sample Disk tab displays current status of the sample disk.
As Figure 4-5 shows the Sample Disk tab consists of three major parts, a sample
disk diagram on the left, sample information and tests on the right and operating
buttons on the bottom.
The sample disk diagram illustrates two circles. Every position consists of two
color-coded rings, the inner ring and the outer ring:
4-11
Advanced Operations
1 Select the desired sample disk from the pull down list in the Sample Disk
field to check the corresponding sample and sample run information.
2 Click the desired position on the selected disk to check the corresponding
sample and its included tests.
1 Select the desired sample disk from the pull down list in the Sample Disk
field.
2 Click the desired position on the selected disk. Note that you can only
select a sample which test(s) is (are) all in Requested or Waiting status.
3 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.
4 Click the OK button to confirm the deletion; click the Cancel button, for
any reason, to cancel deletion.
1 Select the desired sample disk from the pull down list in the Sample Disk
field.
2 Click the desired position on the selected disk. Note that you can only
select a sample which test(s) is (are) all in Finished status.
3 Click the Release button and a dialog box will pop up to ask you to
confirm the release.
4 Click the OK button to confirm the release; click the Cancel button, for
any reason, to cancel the release.
1 Select the desired sample disk from the pull down list in the Sample Disk
field.
4-12
Advanced Operations
2 Click the Clear button and a dialog box will pop up to ask you to confirm
the clearing.
3 Click the OK button to confirm the clearing; click the Cancel button, for
any reason, to cancel the clearing.
To re-run a sample
1 Select the desired sample disk from the pull down list in the Sample Disk
field.
2 Click the desired position on the selected disk. Note that you can only
re-run a routine or STAT sample.
3 Select the test or tests need to be included in the re-run. Click the check
box to select it. Click the check box again to de-select it.
4 To re-run the selected sample without dilution, go directly to the next step;
to dilute the selected sample before the re-run, tick the Diluted field, and
then enter the appropriate sample volume and dilution ratio into the
Sample and Dilution fields respectively.
1 Select the desired sample disk from the pull down list in the Sample Disk
field.
2 Click the desired position on the selected disk to check the corresponding
sample and its included tests.
3 Click the desired test. Note that you can only select a Finished test.
4-13
Advanced Operations
4 Click the Reaction Curve to check the reaction curve and data of the
selected blank runs.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
5 If you have no intention to view other results, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
window; or click the Return button and repeat from step 2.
6 Click the Return button to close the Reaction Curve window.
The Reaction Disk tab displays current status of the reaction disk.
4-14
Advanced Operations
As Figure 4-6 shows, the Reaction Disk tab consists of three major parts, a reaction
disk diagram on the left, test information, status and results on the right and
operating buttons on the bottom.
The reaction disk illustrates 80 reaction cuvette positions, all of which are
color-coded.
1 Click the desired cuvette in the reaction disk diagram. Note that you can
only check the reaction curve and data of the Finished runs.
4-15
Advanced Operations
2 Click the Reaction Curve to check the reaction curve and data of the
selected blank runs.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
3 If you have no intention to view other results, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
window; or click the Return button and repeat from step 1.
4 Click the Return button to close the Reaction Curve window.
1 After loading the segments into the cuvette compartment, click the Load
Cuvette button to command the cuvette feeder to press the newly added
segments forward.
NOTE:
The Load Cuvette button is only to command the cuvette feeder to
press the cuvette segments forward.
4-16
Advanced Operations
As Figure 4-7 shows, the Test Status tab consists of the following tabs:
Sample Runs
Finished Patient Lists
Results
Sample Runs
The Sample Runs tab lists all the Requested, Waiting and Running samples of the
current day since the latest power-on.
4-17
Advanced Operations
4-18
Advanced Operations
2 Click the Delete button and the Delete Sample dialog box will pop up.
NOTE:
In the Delete Sample dialog box, you can click the Refresh button to
view current samples in Requested or Waiting status.
3 Click the Patient Info. button to pop up the Patient Information dialog
box, and you can view or modify the patient information in the dialog box
(refer to 4.3.3 Patient Info. for details).
4-19
Advanced Operations
4 Click the Reaction Curve button to pop up the Reaction Curve screen
that shows the reaction curve of the selected test.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
5 If you have no intention to view other results, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
screen; or click the Return button and repeat from step 2.
6 Click the Return button to close the Reaction Curve window.
4-20
Advanced Operations
The Finished Patient Lists tab lists all the Finished samples of the current day
since the latest power-on. You can print patient reports and edit test results through
this tab.
1 Click the Group Selection button. A dialog box will pop up.
2 Enter the number of the start sample into the first dialog box that pops up
and then click the OK button. A second dialog box will then pop up.
3 Enter the number of the end sample into the second dialog box that pops
up and click the OK button to select the two samples and those between
them.
4-21
Advanced Operations
1 Click the Group Deletion button and a dialog box will pop up.
2 Enter the number of the start sample into the dialog box that pops up and
then click the OK button. Another dialog box will then pop up.
3 Enter the number of the end sample into the second dialog box that pops
up and click the OK button to de-select the two samples and those
between them.
4-22
Advanced Operations
4 Click Reaction Curve to check the reaction curve and data of the
selected test.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
5 If you have no intention to view other results, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
screen; or click the Return button and repeat from step 2.
6 Click the Return button to close the Reaction Curve screen.
4-23
Advanced Operations
3 Click the Print by Patient Button and a print preview window will pop up.
At the Print Preview window, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.
NOTE:
You can click (to the first page), (to the
previous page), (to the next page) or (to the
last page) to go to the desired page.
Pr. Current refers to printing the page currently displayed on
the screen.
Pr. All refers to printing all the pages of the selected sample
runs.
4 When the printing is done, click the Close button to exit the Print Preview
window.
NOTE:
Sample results can only be edited by authorized personnel.
4-24
Advanced Operations
6 If you want to save the new results, click the Save button; otherwise, click
the Cancel button.
NOTE:
You can only edit sample results of the current day since the latest
power-on here.
To print daily summary is to print out all the patient results of a certain day. You can
only use an impact printer for this purpose. Before printing, ensure the analyzer is
not doing anything related to an analysis.
1 Click the Print Daily Summary button and a dialog box will pop up to ask
you to confirm the printing.
2 Click the OK button to confirm the printing; click the Cancel button, for
any reason, to abort the printing.
Once you click the OK button, another dialog box will pop up to ask you to
select a printer.
4-25
Advanced Operations
4 Enter the date (the default is the current date) and click the OK button to
print out the patient results; click the Cancel button, for any reason, to
abort the printing.
If there is no result of the entered date, a dialog box will pop up to alert
you. Click the OK button to close the dialog box and abort the printing.
Results
The Results tab lists all the finished sample tests of the current day since the latest
power-on. You can print the test report and revise results through this tab..
4-26
Advanced Operations
1 Click the Group Selection button. A dialog box will pop up.
2 Enter the number of the start test into the first dialog box that pops up and
then click the OK button. A second dialog box will then pop up.
3 Enter the number of the end test into the second dialog box that pops up
and click the OK button to select the two tests and those between them.
1 Click the Group Deletion button and a dialog box will pop up.
2 Enter the number of the start test into the dialog box that pops up and
then click the OK button. Another dialog box will then pop up.
3 Enter the number of the end test into the second dialog box that pops up
and click the OK button to de-select the two tests and those between
them.
4-27
Advanced Operations
3 Click Reaction Curve to check the reaction curve and data of the
selected test.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
4 If you have no intention to view other results on the current virtual sample
disk, go directly to the next step;
Otherwise, click the Previous or Next button at the Reaction Curve
screen; or click the Return button and repeat from step 2.
5 Click the Return button to close the Reaction Curve screen.
4-28
Advanced Operations
At the Print Preview window, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.
4 When the printing is done, click the Close button to exit the Print Preview
window.
NOTE:
Sample results can only be revised by authorized personnel.
If you are not authorized to revise the results, a dialog box will pop up
to ask you to enter the authorized user name and password.
4-29
Advanced Operations
4 Enter values in the edit boxes to the right of Factor a: and Factor b:.
5 If you want to revise the tests with the formula Y=aX+b (Y stands for the
new result and X stands for the old result), click the Save button;
otherwise, click the Cancel button.
After clicking the Save button, you can see the Test Results in the
Finished Tests field have been changed to the new ones.
NOTE:
You can only revise the sample results of the current day since the
latest power-on here.
The Calib. Status tab lists the calibration runts of the current day since the latest
power-on.
4-30
Advanced Operations
1 Select a calibration test in the Tests field. You can view the basic
information and the status of the selected test and the results of the
Finished tests.
1 Select the desired calibration run in Requested or Waiting status. You can
only delete one run at a time.
3 Click the Reaction Curve and a window will pop up to display the
corresponding reaction curve.
4-31
Advanced Operations
There are two tabs in the Reaction Curve screen: the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
4 Go directly to the next step if you have no intention to check other results;
Otherwise, click the Previous or Next button to view other results (not
only the results of calibrations) or click the Return button and then repeat
from step 1.
5 Click the Return button to close the Reaction Curve window.
4-32
Advanced Operations
2 Click the Calib. Curve button to display the calibration curve and the
calibration result.
To print the displayed curve and result, click the Print button.
3 Click the Return button to exit the Curve window.
The QC Status tab lists the QC runs of the current day since the latest power-on.
4-33
Advanced Operations
1 Select a QC run in the Tests field. You can view the parameters and the
status of the selected run and the results of the Finished runs.
3 Click the Reaction Curve and a window will pop up to display the
corresponding reaction curve.
There are two tabs in the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
4-34
Advanced Operations
4 Go directly to the next step if you have no intention to check other results;
Otherwise, click the Previous or Next button to view other results (not
only the results of QCs) or click the Return button and then repeat from
step 1.
5 Click the Return button to close the Reactive Curve window.
1 Select the desired Requested QC run. You can only delete one run at a
time.
4.1.5 Start
Start is used to run the requested tests.
If the current tests are paused, Start will change to Resume that is used to continue
the paused tests.
4-35
Advanced Operations
Start tests
1
Click Routine → Start or press F5, or click the short-cut button to
pop up the following dialog box.
2 In the Sample dialog box, select a virtual sample disk to the right of
Sample Disk.
NOTE:
If there are any requested STAT samples on the current
default virtual sample disk, you can only select the disk.
3 In the Sample dialog box, select samples to be run in the Please select
the samples you need to run field.
If you select All, go directly to the step 5; if you select Partial, go to the
next step.
NOTE:
If you select All, the system will run all the Requested
samples (Waiting STAT samples included), calibrations and
QC tests.
If you select Partial, the system will run some Requested
samples, calibrations and QC tests. When Partial is
selected, the Sample field is available.
4-36
Advanced Operations
4 In the Sample dialog box, select tests to be run in the Sample field.
If you want to run sample tests, select the check box to the left of Sample
ID and enter the sample ID in the edit boxes to the right of it. The system
will run the Requested samples whose IDs are between the two numbers
(including the two numbers).
If you want to run calibration tests, select the check box to the left of
Calibration and select tests below it.
If you want to run QC tests, select the check box to left of QC and select
tests below it.
5 If you want to continue, click the OK button; otherwise, click the Cancel
button.
When you click the OK button, a dialog box will pop up.
6 Select a virtual reagent disk in the Select Reagent Disk dialog box.
7 If you want to run the selected tests, click the OK button; otherwise, click
the Cancel button and the system will not run the selected tests.
When you click the OK button, the status of the selected tests changes to
Waiting and the system will run the tests which use the reagents on the
selected virtual disk immediately.
4-37
Advanced Operations
NOTE:
If the system finishes the tests that use the reagents on the selected
virtual reagent disk, and there are waiting tests which use reagents on
the other disk, a dialog box will pop up.
Resume tests
click the short-cut button to continue the tests. The system will
take different operations according to different conditions. Refer to the
following steps for details.
2 If there is no test in Requested status and no need to replace the reagent
disk, the system will continue the tests directly.
3 If there is no test in Requested status but you need to replace the reagent
disk, the following dialog box will pop up.
4-38
Advanced Operations
4 If there are tests in Requested status, the following dialog box will pop up.
If you want to continue the tests currently being paused, click Cancel
button;
If you want not only to continue the tests currently being paused but also
to run the tests in Requested status, select tests and click the OK button.
After that, if there is no need to replace the reagent disk, the system will
continue the tests; if you want to replace the reagent disk, a dialog box
will pop up.
Click Routine → Probe Stop, or press F6, or click the short-cut button to
stop the sample and reagent probes and the mixing bar without stopping the reaction
disk. You may use this function to add more samples or reagents without stopping
the analysis. This menu is only functional when the system is testing.
NOTE:
Do not put the probes and the bar on hold for too long. Otherwise,
certain analyses may be affected.
4-39
Advanced Operations
4.1.7 Stop
Click Routine → Stop, or press F8, or click the short-cut button to stop the
current run. This menu is only functional when the system is testing.
NOTE:
Mindray recommends this stop function not be used unless it is
necessary (for instance the analyzer is experiencing problems). When
the analyzer is experiencing a serious problem, the analyzer will be
stopped automatically.
After the analyzer is stopped, all the remaining runs will be invalidated
and return to the Requested status, but the finished runs will be
4.1.8 Exit
When you have finished all analyses and the system is in the idle status, click
Routine → Exit, or press F10, or to click the short-cut button to exit the
control system. A dialog box will pop up to ask you to confirm the exiting operation.
Click the OK button and another dialog box will pop up to ask you to load detergents.
NOTE:
Please use the Mindray-recommended detergents: 0.1mol/l
hydrochloric acid, or 0.5% (V/V) Javel water.
Use of other detergents may lead to inaccurate results. Do not use
other detergents.
4-40
Advanced Operations
Ensure acid and alkaline detergents are loaded to positions 45 and 47 respectively
and then click the OK button and the analyzer will start the following shutdown
procedure:
When the shutdown procedure is done, a dialog box will pop up to remind you that
you can shutdown the ANALYZING UNIT POWER to OFF safely. Click the OK
button to exit the control software.
NOTE:
Use this emergency exit function only when the analyzer is
experiencing a serious problem.
4.2 Parameters
Parameters Tests
Reagents
Calibration
QC
4.2.1 Tests
The Tests screen is where you set parameters regarding the tests, calculation tests,
profiles, carryover and manual tests.
To enter the Tests screen, click Parameters → Tests, or press the F11 key on the
keyboard.
4-41
Advanced Operations
NOTE:
If you are not authorized to edit the test parameters, a dialog box will
pop up to ask you to enter the authorized user name and password.
At the Tests screen, shown in Figure 4-13, you can see the following tabs:
Parameters
Expected
Calculation
Profile
Carryover
Manual Test
4-42
Advanced Operations
The Parameters tab consists of three main parts, the upper part being the
parameters to be set for the selected test, the middle part being all the tests whose
parameters are set already, the lower part being the operating buttons.
In the middle part, yellow-framed tests are the double-reagent tests whose reagent
positions are set already; the gray-framed tests are the double-reagent tests whose
reagent positions are not set yet; the non-framed tests are the single-reagent tests.
Follow the instructions given below to set the parameters for the selected test.
4-43
Advanced Operations
4-44
Advanced Operations
4-45
Advanced Operations
Calculation:
4-46
Advanced Operations
4-47
Advanced Operations
NOTE:
Please define the parameters as instructed by this operation manual
and the instructions of the reagent suppliers. Improper settings may
lead to unreliable test results.
1 Go to the middle part of the tab and click the desired test and then the
upper part will show the parameters of the specific test.
4-48
Advanced Operations
NOTE:
If there are any obligatory parameters missing, once you
click the Save button, the analyzer will automatically
red-frame the missing parameters (or pops up a dialog box
to warning you about it) and disables the Save button. You
need to set the missing parameters and try to save the
settings again.
To delete a test
1 Go to the middle part of the tab and click the desired test.
2 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.
3 Click the YES button to delete the test; click the NO button, for any
reason, to ignore the deletion.
NOTE:
Once you have confirmed to delete the selected test, the
analyzer will delete all the parameters, including the reagent
parameters, calibration parameters, and QC parameters,
related to the test. Ensure the right test is selected before
confirming the deletion.
4-49
Advanced Operations
To edit a test
1 Go to the middle part of the tab and click the desired test.
A dialog box will pop up to ask you whether to delete other parameters
related to this test.
If you click the Yes button, the analyzer will save the changed
parameters and clear all other unchanged ones, including the
corresponding reagent parameters, calibration parameters, and QC
parameters; if you click the No button, the analyzer will save the
changed parameters and keep the other related parameters unchanged.
NOTE:
Once you have clicked the Yes button, the analyzer will
delete all the parameters, including the reagent parameters,
calibration parameters, and QC parameters, related to the
test. Ensure the right test is selected before confirming the
deletion.
4-50
Advanced Operations
Click the Expected tab to edit or view the expected means of the test, as Figure
4-15 shows.
4-51
Advanced Operations
1 Click “▼” to the right of Test and select the desired test from the pull
down list.
2 Click “▼” to the right of Sample Type and select the desired type from the
pull down list.
4-52
Advanced Operations
3 Click “▼” to the right of Gender and select the desired gender from the
pull down list.
The pre-set expected means will then be displayed.
1 Click “▼” to the right of Test and select the desired test from the pull
down list.
2 Click “▼” to the right of Sample Type and select the desired type from the
pull down list.
3 Click “▼” to the right of Gender and select the desired gender from the
pull down list.
4 Click the Setup button.
5 Tick the check box of the first patient group and set the Years, Months,
Lower and Upper field as instructed above.
6 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.
NOTE:
In case of inappropriate settings, a dialog box will pop up to
alert you once you have clicked the Save button.
Calculate certain tests to derive certain new tests of clinical purpose, such as A/G,
TBil-DBil and so forth.
The Calculated tab is where you set the parameters related to such calculation.
4-53
Advanced Operations
As Figure 4-18 shows, the Calculation tab consists of four major parts: the
parameters on the upper left, all the available tests on the upper right, calculation
formula on the middle and operating buttons on the bottom.
4-54
Advanced Operations
1 Select the desire test from the Tests field, or click or until
the desired test is reached.
NOTE:
Click to go to the previous test.
3 Select the tests to be involved in the calculation from the Tests field.
4-55
Advanced Operations
4 Click the Add to list button to add the selected tests to the calculation list.
NOTE:
In case you add inappropriate tests to the calculation list by
mistake, click the Clear button to clear the list and then
re-select appropriate tests to add to the list.
5 Edit the calculation formula by clicking a selected test from the calculation
list and click the Add to formula button to add the test to the Edit formula
field. Clicking the digit buttons (0-9) and operator buttons (+, -, and so
forth) to associate the test with appropriated digits and operators as
needed by the desired calculation formula. Repeat this step on every
selected test until the desired calculation formula is completed.
NOTE:
To clear a mistake, click the Clear Formula button to clear
the formula and re-edit it the right way.
6 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.
1 Select the test to delete from the pull down list on the middle left of the
screen.
4-56
Advanced Operations
2 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.
3 Click the Yes button to confirm the deletion; click the No button, for any
reason, to cancel the deletion.
1 Select the test to edit from the pull down list on the middle left of the
screen.
2 Edit the formula as needed.
NOTE:
In case you add inappropriate tests to the calculation list by
mistake, click the Clear button to clear the list and then
re-select appropriate tests to add to the list.
To clear a mistake, click the Clear Formula button to clear
the formula and re-edit it the right way.
3 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.
Tests grouped together for certain clinical purpose (for instance liver function)
constitute a profile.
4-57
Advanced Operations
As the figure shows, the Profile tab consists of three major parts: existing profiles
(groups) on the left, available tests (tests have no assigned reagent positions are
followed by a forbidden mark) on the right and operating buttons on the bottom.
1 Select the desired group from the pull down list of the Group Name field
or from the Group List field. The included tests will be followed a green
tick.
4 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.
4-58
Advanced Operations
To delete a profile
1 Select the desired group from the pull down list of the Group Name field
or from the Group List field.
3 Click the OK button to confirm the deletion; click the Cancel button, for
any reason, to cancel the deletion.
To edit a profile
1 Select the desired group from the pull down list of the Group Name field
or from the Group List field.
2 Click the Edit button.
3 Select or de-select the tests as you need. The selected tests will be
followed by a green tick. Clicking the ticked test again may de-select the
test.
4 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.
Various reagents are sampled by the reagent probe and carryover between the
adjacent reagents, though minimized by the washing process, is still a factor to be
taken into account. This carryover may impose serious effect on certain reagents
and consequently on the related test results. This Carryover tab is designed so that
you can minimize this effect by keeping the tests whose reagents may affect each
other as far away from each other as possible, or, if this is not an option, ordering
extra probe washing before analyzing those tests.
4-59
Advanced Operations
As the figure shows, the Carryover Parameters screen consists of three major parts:
current reagents listed in the Current field on the upper, the reagents that may be
affected (contaminated) by the current reagents listed in the Contaminated field on
the lower, and operating buttons on the bottom.
Note that the yellow reagents are the first reagents of the double-reagent tests and
the blue reagents are the second reagents of the double-reagent tests.
1 Click the desired reagent from the Current field and the selected reagent
will be followed by a green tick. Check the Contaminated field and the
tests followed by a red tick are those that may be affected by the current
reagent.
4-60
Advanced Operations
1 Click the desired reagent from the Current field and the selected reagent
will be followed by a green tick. Check the Contaminated field and the
tests followed by a red tick are those that may be affected by the current
reagent.
2 Click the Clear button and you will see all the red ticks have disappeared.
All the tests that are not run by the BS-300 are referred to as the manual tests, given
you have to manually enter the test results into the analyzer if you want to use it to
print out the results.
You can set basic information of manual tests at the Manual Test tab. Refer to
4.3.4.1 Test Data tab for details regarding entering the test results manually.
As the figure shows, the Manual Test tab consists of mainly three parts: the
parameters on the upper left, the tests on the upper right and the operating buttons
on the bottom.
4-61
Advanced Operations
1 Select the desire test from the test list and its parameters will be displayed
on the screen.
4-62
Advanced Operations
2 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.
3 Click the OK button to confirm the deletion; click the Cancel button, for
any reason, to cancel the deletion.
4.2.2 Reagents
To enter the Reagents screen, click Parameters → Reagents, or press F12.
NOTE:
If you are not authorized to change the reagent parameters, a dialog
box will pop up to ask you to enter the authorized user name and
password.
4-63
Advanced Operations
At the Reagents screen, Figure 4-22, you can see two tabs:
Reagent Disk
Leftover
The Reagents Disk tab is where you set the parameters regarding the reagents.
The Leftover tab is where you check the remaining reagents and print the checking
results.
Click the Reagent Disk tab to edit or view the reagent disk status, as Figure 4-23
shows.
4-64
Advanced Operations
As the reagent disk diagram shows, the reagent disk has 50 reagent positions, 25 on
the inner circle and marked even number and another 25 on the outer circle and
marked odd numbers.
All the reagent positions are color-coded: white for the empty positions; green for the
positions occupied by reagents for the single-reagent tests; red for the positions
occupied by reagents for the double-reagent tests; blue for the positions (49 and 50)
occupied by distilled water. Note that you cannot use the two blue positions for other
purposes.
The list to the right of the reagent disk diagram displays all the tests that have been
set.
4-65
Advanced Operations
1 Click “▼” to the right of Reag. Disk and select the desired virtual reagent
disk from the pull down list.
2 Click the desired test in the test list.
1 Click “▼” to the right of Reag. Disk and select the desired virtual reagent
disk from the pull down list.
2 Click the desired position in the reagent disk diagram.
1 Click “▼” to the right of Reag. Disk and select the desired virtual reagent
disk from the pull down list. The positions of the reagents will be set on
the selected virtual reagent disk.
2 Click the desired test from the test list.
4-66
Advanced Operations
6 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.
NOTE:
When you have set (not changed) the reagent position for a
test, the analyzer will automatically take the reagent bottle
volume as the volume of the remaining reagent.
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have clicked the Save
button.
1 Click “▼” to the right of Reag. Disk and select the desired virtual reagent
disk from the pull down list.
2 Select the desired test from the test list.
3 Click the Clear button to clear the reagent positions set for the selected
test.
NOTE:
Once you have clicked the Clear button, the analyzer will
immediately clear all the reagent positions set for the
selected test without asking you to confirm the clearing.
Ensure the right test is selected before clicking the Clear
button.
1 Click “▼” to the right of Reag. Disk and select the desired virtual reagent
disk from the pull down list.
2 Select a test.
4-67
Advanced Operations
3 If the new positions for the selected test are on the current virtual reagent
disk, go directly to the next step;
Otherwise, click the Clear button and then select the other virtual reagent
disk to the right of Reag. Disk.
4 Click the Setup button.
5 Select the bottle types to the right of R1 Bottle and R2 Bottle for the
reagents. For the single reagent test, R2 Bottle is unavailable.
6 Select the positions to the right of R1 and R2 for the reagents. For the
single reagent test, R2 is unavailable.
7 If you want to save the settings, click the Save button; otherwise, click the
Cancel button.
NOTE:
If some parameters are not set correctly, after you click the
Save button, the system will remind you.
1 Click the Def. Reag. Disk button to pop up the following dialog box.
Click the Leftover tab to check the remaining reagents and print the checking results,
as Figure 4-24 shows.
4-68
Advanced Operations
NOTE:
Ensure you have opened the reagent bottles before checking the
remaining reagents.
1 Click the Reag. Disk tab, and then click the “▼” to the right of Reag. Disk
and select the desired virtual reagent disk from the pull down list. Refer to
4.2.2.1 Reagent Disk tab for details.
2 To select one reagent, click the desired reagent;
to select multiple discontinued reagents, press and hold CTRL and click
the desired reagents one by one;
to select multiple continue reagents, click the first reagent of the desired
ones then press and hold SHIFT and click the last reagent of the desired
ones.
4-69
Advanced Operations
3 Click the Check Leftover button to view how much reagent is left in the
selected bottle and how many tests the remaining reagent is enough for.
1 Click the Print button to print the currently displayed reagent leftover
information.
4.2.3 Calibration
The Calibration screen is where you set the parameters regarding calibrators and
calibrations and view the calibration results.
NOTE:
If you are not authorized to edit the calibration parameters, a dialog
box will pop up to ask you to enter the authorized user name and
password.
At the Calibration screen, Figure 4-25, you can see the following tabs:
Calibrator
Calibration
Calib. Data
Click the Calibrator tab to display its contents, as Figure 4-26 shows. The tab
consists of 4 zones:
Five virtual sample disks (Disk1 - Disk5) are displayed in the Positions zone and
every virtual disk has 6 calibrator positions (S1 - S6). The 0 position (default) means
this virtual disk has no position set for the calibrator yet. You cannot request a
calibration run for a calibrator whose corresponding calibrator positions on all the five
virtual disks are 0s.
4-71
Advanced Operations
1 Click the desired calibrator in the Calibrators zone and check its
positions and tests in the Positions and Tests included zones
respectively.
1 Click the Add button and an Attribute window will pop up.
2 The analyzer will automatically assign a number to the new calibrator and
you can only view in the No. field without altering it;
Enter calibrator name into the edit box to the right of Name;
Enter the lot number into the edit box to the right of Lot No.;
Click “▼” to the right of Conc. Level and select Low, Medium or High
from the pull down list as the concentration level of the calibrator;
Click “▼” to the right of Exp. Date and select the appropriate as the
expiration date of the calibrator.
NOTE:
Ensure the right expiration date is set so that the analyzer
can correctly judge whether the calibrator has expired.
4-72
Advanced Operations
3 Click the Save button to save the settings; Click the Cancel button, for
any reason, to ignore the settings.
NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have click the Save
button.
2 Click the Edit button and an Attribute window will pop up.
3 The analyzer will automatically assign a number to the new calibrator and
you can only view in the No. field without altering it;
Enter calibrator name into the edit box to the right of Name;
Enter the lot number into the edit box to the right of Lot No.;
Click “▼” to the right of Conc. Level and select Low, Medium or High
from the pull down list as the concentration level of the calibrator;
Click “▼” to the right of Exp. Date and select the appropriate as the
expiration date of the calibrator.
4 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.
NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have clicked the Save
button.
4-73
Advanced Operations
2 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.
3 Click the Yes button to confirm the deletion; Click the No button, for any
reason, to abort the deletion.
NOTE:
Once you have confirmed the deletion, the analyzer will
delete all the parameters, including calibrator positions, tests
included, etc., related to the selected calibrator. Ensure the
right calibrator is selected before confirming the deletion.
You cannot delete a calibrator involved in calibrations that
are in process or have been requested.
4-74
Advanced Operations
4-75
Advanced Operations
4-76
Advanced Operations
The Calibration tab is where you set calibration parameters for the tests.
Click the Calibration tab to display its contents, as Figure 4-28 shows. The tab
consists of 4 zones:
Operating buttons zones – lists such buttons as Save, Cancel, Clear, etc.
Tests Included zone – lists all the available tests. Those tests followed by
4-77
Advanced Operations
cannot set the calibration parameters for them. The blue-framed tests are those
whose calibration parameters have already been set.
Request zone – defines the nine parameters regarding calibration requests.
Calibrators zone – lists the calibrators that have already been set. The
selected calibrator will be preceded by a red tick; the expired calibrator will be
preceded by a red block.
4-78
Advanced Operations
4-79
Advanced Operations
NOTE:
You must determine the number of the calibrators to be used
according to the selected calibration rule. You can only
select calibrators whose standard concentrations have
already been set.
4-80
Advanced Operations
6 Click the Save button to save the changes; Click the Cancel button, for
any reason, to ignore the changes.
NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have clicked the Save
button.
2 Select the test or tests whose calibration settings you want to clear.
NOTE:
Once you have clicked the Clear button, the analyzer will
immediately clear all the calibration settings related to the
selected test without asking you to confirm the clearing.
Ensure the right test is selected before clicking the Clear
button.
The Calib. Data tab, as Figure 4-29 shows, is where you search for the calibration
results and edit them.
4-81
Advanced Operations
Click “▼” to the right of Start (or End) and select the desired date to
specify the dates to be searched. The analyzer will search all the
calibration runs taking place during this specified period.
NOTE:
If no calibration run matches the specified conditions, a
dialog box will pop up to remind you no match is found.
Under this circumstance, go directly to step 6.
4-82
Advanced Operations
Click “▼” to the right of Start (or End) and select the desired date to
specify the dates to be searched. The analyzer will search all the
calibration runs taking place during this specified period.
2 Click the Lock button to lock the selected search conditions.
3 Click the Search button to start the search and the matches will be
displayed in the Results zone. You can’t go on unless there are matches.
4 Select the desired calibrations (except for those whose Result Type
displays Calib. Edit) from the Calibration field.
4-83
Advanced Operations
7 Select the desired calibration rule from the pull down list to the left of
Rule.
8 Click the Calculate button to start the calculation; click the Cancel button,
for any reason, to cancel the calculation.
Click “▼” to the right of Start (or End) and select the desired date to
specify the dates to be searched. The analyzer will search all the
calibration runs taking place during this specified period.
2 Click the Lock button to lock the selected search conditions.
3 Click the Search button to start the search and the matches will be
displayed in the Results zone. You can’t go on unless there are matches.
4 Select the desired calibrations from the Calibration field.
NOTE:
You cannot select the calibration whose Default field
displays Yes.
5 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.
4-84
Advanced Operations
6 Click the OK button to confirm the deletion; click the Cancel button, for
any reason, to cancel the deletion.
Click “▼” to the right of Start (or End) and select the desired date to
specify the dates to be searched. The analyzer will search all the
calibration runs taking place during this specified period.
2 Click the Lock button to lock the selected search conditions.
3 Click the Search button to start the search and the matches will be
displayed in the Results zone. You can’t go on unless there are matches.
4 Select the desired calibrations from the Calibration field.
5 Click the Default button to set the current calibration settings as the
default settings of the selected test. The Default field of the calibration will
display Yes.
NOTE:
This analyzer uses the default calibration settings to
calculate the sample concentrations.
This analyzer will automatically set the latest settings
(include the settings of calibration test, edit, calculation) as
the default.
4.2.4 QC
The QC screen is where you set the parameters regarding controls and QC runs and
view QC results.
4-85
Advanced Operations
NOTE:
If you are not authorized to change the QC parameters, a dialog box
will pop up to ask you to enter the authorized user name and
password.
At the QC screen, Figure 4-30, you can see the following tabs:
The Controls tab is where you set parameters for the controls.
4-86
Advanced Operations
Click the Controls tab to display its contents, as Figure 4-31 shows. The tab mainly
consists of 4 zones:
Five virtual sample disks (Disk1 - Disk5) are displayed in the Positions zone and
every virtual disk has 3 control positions (C1 - C3). The 0 position (default) means
this virtual disk has no position set for the control yet. The Other means you can set
other positions for the controls on the disk.
You cannot request a QC run for a control whose corresponding control positions on
all of the five virtual disks are 0s.
4-87
Advanced Operations
1 Click the desired control in the Controls zone and check its positions and
tests in the Positions and Tests included zones respectively.
1 Click the Add button and an Attribute window will pop up.
2 The analyzer will automatically assign a number to the new control and
you can only view in the No. field without altering it;
Enter control name into the edit box to the right of Name;
Enter the lot number into the edit box to the right of Lot No.;
Click “▼” to the right of Conc. Level and select Low, Medium or High
from the pull down list as the concentration level of the control;
Click “▼” to the right of Exp. Date and select the appropriate date as the
expiration date of the control;
Click “▼” to the right of Tube Type and select Large Tube or Small Tube
for the control.
NOTE:
Ensure the right expiration date is set so that the analyzer
can correctly judge whether the control has expired.
4-88
Advanced Operations
3 Click the Save button to save the changes; Click the Cancel button, for
any reason, to ignore the change.
NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have clicked the Save
button.
2 Click the Edit button and an Attribute window will pop up.
3 The analyzer will automatically assign a number to the new control and
you can only view in the No. field without altering it;
Enter control name into the edit box to the right of Name;
Enter the lot number into the edit box to the right of Lot No.;
Click “▼” to the right of Conc. Level and select Low, Medium or High
from the pull down list as the concentration level of the control;
Click “▼” to the right of Exp. Date and select the appropriate date as the
expiration date of the control;
Click “▼” to the right of Tube Type and select Large Tube or Small Tube
for the control.
4-89
Advanced Operations
4 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes
NOTE:
If any field is inappropriately set, the analyzer will warn you
about the wrong settings once you have click the Save
button.
2 Click the Delete button and a dialog box will pop up to ask you to confirm
the deletion.
3 Click the Yes button to confirm the deletion; Click the No button, for any
reason, to abort the deletion.
NOTE:
Once you have confirmed the deletion, the analyzer will
delete all the parameters, including control positions, tests
included, etc., related to the selected control. Ensure the
right control is selected before confirming the deletion.
You cannot delete a control involved in QC runs that are in
process or have been requested.
4-90
Advanced Operations
The control positions that appear gray are occupied by other controls.
4 Click the Save Position button to save the changes.
4-91
Advanced Operations
4-92
Advanced Operations
6 Click the OK button to save the changes; Click the Cancel button, for any
reason, to ignore the change.
7 If you do not want to set or delete standard concentration or SD for
another control, go directly to the next step;
Otherwise, click “▼” button to the right of Controls and select the desired
control from the pull down list or click or until the desired
control is found. Then go back to step 3.
8 Click the Return button to exit the Concentrations window.
The QC Rules tab is where you set QC rules for the tests.
Click the QC Rules tab to display its contents, as Figure 4-32 shows. The tab
consists of 3 zones:
The QC rules can be divided into two categories: independent rules, such as
Cumulative Sum Check and Westgard Multi-rule, and control-related rules (not
related to controls), such as Accumulated Error and X-R. Note that you cannot set
the parameters of the control-related rules unless you have specified the controls to
be used.
4-93
Advanced Operations
As Figure 4-33 shows, all the available tests are listed on the left and the rules,
Cumulative Sum Check and Westgard Multi-rule, are listed on the right. The
Cumulative Sum Check rule has three sub-rules and you can choose one of the
sub-rules. The Westgard Multi-rule rule has six sub-rules and you can choose one
or more sub-rules.
As Figure 4-34 shows, all the available tests are listed on the left and the QC rule
parameters are listed in the Rule Settings field, including Controls, Accumulated
Error and X-R QC Rule.
Note that any test whose standard concentration and SD have not been set yet will
be followed by a forbidden mark, such as .
4-94
Advanced Operations
3 Select the desired test or tests. The selected test will be followed by a
green tick.
5 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.
6 Repeat from step 2 to set the parameters for other tests, or click the check
box to the left of Independent Parameters again to de-select the field.
4-95
Advanced Operations
3 Select the desired test or tests. The selected test will be followed by a
green tick.
4 To set the parameters: click “▼” button to the right of Controls and select
the desired control from the pull down list; enter the desired E (limit of the
accumulated error) into the edit box to the right of Limit E; enter the
desired SDR value in the edit box to the right SDR (for the X-R rule).
5 Click the Save button to save the settings; click the Cancel button, for any
reason, to ignore the settings.
6 Repeat from step 2 to set the parameters for other tests, or click the check
box to the left of Control-related Parameters again to de-select the field.
4.2.4.3 QC tab
4-96
Advanced Operations
The tests, whose QC parameters have already been set, are blue-framed. The tests
whose reagent positions have not been set yet are followed by a forbidden mark,
such as , and you cannot set QC parameters for the test.
All the controls that have been set are listed in the Controls to be used zone. The
selected tests will be preceded by a red tick and the expired test will be preceded by
a red mark.
You can set the interval of the automatic QC runs by entering the desired number
into the edit box to the right of Interval or clicking on the right of the edit box until
the desired number is reached. To disable the auto QC function, enter 0 into the edit
box.
4-97
Advanced Operations
4-98
Advanced Operations
5 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.
NOTE:
If any parameter is inappropriately set, the analyzer will alert
you after you have clicked the Save button.
NOTE:
Once you have clicked the Clear button, the analyzer will
immediately clear the QC settings related to the selected test
without asking you to confirm the clearing. Ensure the right
test is selected before clicking the Clear button.
The Real-time tab is where you view and print the real-time QC results and delete
the real-time QC data.
4-99
Advanced Operations
1 Select the desired test from the Tests zone. The selected test will be
followed by a green mark.
3 Click the Search button to display the information, including the results,
QC graph, statistic data and QC status, of the last 10 control runs of the
selected test using the selected control or controls.
To print the displayed data and graph, click the Print button.
1 Click the Print button to print out the result and QC graph after checking
the result of a control run.
4-100
Advanced Operations
1 Select the desired test in the Test field. The selected test will be marked
with a green block to the right of it.
3 Click the Delete button and a dialog box will pop up to ask to enter the
user name and the password of the authority to delete the real time
control data.
4 Enter the user name and password as required and click the OK button to
enter the Edit QC data window.
4-101
Advanced Operations
The Daily tab is where you view and print the daily QC results and delete the daily
QC data.
4-102
Advanced Operations
1 Click “▼” to the right of Date and select the desired date from the pull
down list to search the QC runs taking place within this date.
2 Click “▼” to the right of Rule and select the desired QC rule from the pull
down list.
3 Select the desired test from the Tests zone. The selected test will be
followed by a green mark.
5 Click the Search button to display the information, including the results,
QC graph, statistic data and QC status, of the last 10 control runs of the
selected test using the selected control or controls.
To print the displayed data and graph, click the Print button.
1 Click the Print button to print out the result and QC graph after checking
the result of a control run.
4-103
Advanced Operations
2 Select the desired test in the Test field. The selected test will be marked
with a green block to the right of it.
4 Click the Delete button and a dialog box will pop up to ask to enter the
user name and the password of the authority to delete the real time
control data.
5 Enter the user name and password as required and click the OK button to
enter the Edit QC data window.
4-104
Advanced Operations
The Day- to-day tab is where you view and print the day-to-day QC results.
4-105
Advanced Operations
1 Click “▼” to the right of Start (or End) and select the desired date to
specify the dates to be searched. The analyzer will search all the QC runs
taking place during this specified period.
2 Click “▼” to the right of Rule and select the desired QC rule from the pull
down list.
3 Select the desired test from the Tests zone. The selected test will be
followed by a green mark.
5 Click the Search button to display the information, including the results,
QC graph, statistic data and QC status, of the last 10 control runs of the
selected test using the selected control or controls.
To print the displayed data and graph, click the Print button.
1 Click the Print button to print out the result and QC graph after checking
the result of a control run.
4-106
Advanced Operations
4.3 Data
Export Data
Patient Info.
Result
NOTE:
This analyzer automatically backs up its settings and test results.
Exporting data refers to saving some data to a file (.SHD file)
temporarily so that the data can be transmitted or updated. It is not the
same as backing up the database.
Mindray recommends this exporting function not be used unless it is
necessary.
As Figure 4-39 shows, the Export Data screen consists of the following fields:
4-107
Advanced Operations
Test
Calibration
QC
Sample
Others
Select tests
Separator
The item in blue font listed in the Test, Calibration, QC and Sample fields is
test-related and you can select test in the Select Tests field for these items.
NOTE:
If the data to be exported includes the data of last year, Mindray
recommends the Start date be not before November 1. Otherwise,
you may not get the complete data.
To export data
4-108
Advanced Operations
Example1:
Follow the steps below to select: a) calculated tests (Calculation of the Test
field) and units (Unit of the Test field) for all the tests; b) test parameters
(Parameters of the Test field) for TP_KH and ALB_KH only.
(1) Tick the check boxes to the left of Parameters, Calculation, and Unit in
the Test field respectively.
(2) Tick the check box to the left of Test in the Apply to field.
(3) Tick the check boxes to the left of TP_KH and ALB_KH in the Select
tests field respectively.
Example2:
Follow the steps below to select: a) calibration parameters (Parameters in
the Calibration field) for all the tests; b) calibration results (Results in the
Calibration) for the tests analyzed on December 1, 2004 and December 2,
2004 only.
(1) Tick the check boxes to the left of Parameters and Results in the
Calibration field respectively.
(2) Select December 1, 2004 from the pull down list to the left of the Start in
the Calibration field.
(3) Select December 2, 2004 from the pull down list to the left of End in the
Calibration field.
4-109
Advanced Operations
Example3:
Follow the steps below to select the TP_KH results of the samples: a)
sample IDs are 1 - 10; b) analyzed on December 1, 2004 and December 2,
2004.
(1) Tick the check boxes to the left of Test Results in the Sample field
(2) Enter 1 and 10 into the edit boxes to the right of Sample ID respectively.
(3) Select December 1, 2004 from the pull down list to the left of the Start in
the Sample field.
(4) Select December 2, 2004 from the pull down list to the left of End in the
Sample field.
(5) Tick the check box to the left of Sample in the Applied to field.
(6) Tick the check box to the left of TP_KH in the Select tests field.
4-110
Advanced Operations
2 Click the Export button and a dialog box will pop up.
Save in:
3 Select the desired directory and file and click the Save button to save the
data. You may also click the Cancel button to cancel the exporting.
4-111
Advanced Operations
NOTE:
The imported data will overwrite the corresponding data in the
analyzer’s database. Before importing data, ensure the corresponding
data is exported to another file. You can only import data from .SHD
files.
Mindray recommends this importing function not be used unless
necessary.
To import data
1 Click the Import File button and a dialog box will pop up.
Look in:
2 Select the desired file and click the Open button to open the selected file.
3 Click the Import button to import the data into the database. A dialog box
will pop up when then importing is done. Click the OK button to close the
box.
4-112
Advanced Operations
As Figure 4-41 shows, you can click (to the first patient), (to the
previous patient), (to the next patient) or (to the last patient) to
select the desired patient.
1 Select the date when the patient sample was requested from the pull
down list to the right of Date. The current date is the default.
2
You can click , , or to select the desired
patient, or you can enter the ID of the desired sample into the edit box to
the right of Move to and click the Move to button to go to the desired
sample.
1 Select the date from the pull down list to the right of Date. The current
date is the default.
4-113
Advanced Operations
2 Click the Add button to add a new manual sample whose ID (greater than
9000) is generated by the system automatically.
3 Enter the patient information of the sample.
The Date and the Sample ID cannot be changed.
The Bar Code can be entered manually or by the bar code reader.
4 If you want to save the new manual sample, click the Save button;
otherwise, click the Cancel button.
1 Select the date when the patient sample was requested from the pull
down list to the right of Date. The current date is the default.
2
You can click , , or to select the desired
patient, or you can enter the ID of the desired sample into the edit box to
the right of Move to and click the Move to button to go to the desired
sample.
3 Click the Edit button.
4 Edit the patient information as needed.
The Date and the Sample ID cannot be changed.
5 Click the Save button to save the changes; click the Cancel button, for
any reason, to ignore the changes.
1 Select the date from the pull down list to the right of Date. The current
date is the default.
2 Click the buttons , , , , or after
entering the sample ID to the right of Move to, click the Move to, to select
a sample.
3 Click the Edit button.
4 Click the Set as Default button to set the Sample (sample type), Sender
(doctor who sends the sample), From Dep. (department from which the
sample is sent), Tester (doctor who analyzes the sample) and Lab. (the
laboratory to analyze the sample) as the default settings.
5 Click the Save or Cancel button.
NOTE:
The default setting is only available at the current screen. If you click
the Exit button to close the Patient Information screen, the default
setting disappears and you need to set it again when you enter the
screen the next time.
4-114
Advanced Operations
If you add a new manual sample, after you clicking the Add button, the Sample,
Sender, From Dep., Tester and Lab. are set as the default.
If you edit the patient information of a sample, you can insert a step between the
step 3 and step 4 of “To edit the patient information”: click the Load Default button.
4.3.4 Result
Click Data → Results or press Ctrl + R to enter the Results & Statistics screen, as
Figure 4-42 shows.
At the Results & Statistics screen, you can see the following tabs: Test Data, Test
Results and Test Status.
The Test Data tab is where you can view, edit, revise and print the sample results
and send them to LIS in batch mode.
4-115
Advanced Operations
1 Click “▼” to the right of Start (or End) and select the desired date to
specify the dates to be searched. The analyzer will search all the sample
runs taking place during this specified period.
NOTE:
In case of a year-to-year search, the start date must not be
earlier than November 1st of the previous year.
4-116
Advanced Operations
3 Click the Search button to start the search and the matches will be
displayed.
NOTE:
If no sample run matches the specified conditions, a dialog
box will pop up to remind you no match is found.
4 Click the By sample, and then select a sample to view its results.
4-117
Advanced Operations
1 Search for the sample results. Refer to To view sample results for
details.
2 Select a sample run.
If you select By sample, select a sample and then a run.
4-118
Advanced Operations
3
NOTE:
You will see the Curve button except that you select LIS in
the Operation field.
Click the Curve button to check the reaction curve and data of the
selected blank runs.
There are two tabs at the Reaction Curve screen, the Reac. Curve tab
and the Reac. Data tab. Click the former to check the curve and the latter
to check the data.
4-119
Advanced Operations
4 If you do not want to check the results of other runs, go directly to the next
step;
Otherwise, click the Previous or Next button to check other runs (blanks,
samples, controls, or calibrations); or click the Return button and repeat
from step 2.
5 Click the Return button to close the Reaction Curve window.
1 Click the option button to the left of Print in the Operation field.
2 Click the Print Setup button to enter the Printing Setup screen. Refer to
4.4.7 Print for details.
1 Search for the sample results. Refer to To view sample results for
details.
2 Select By sample.
3 Click the option button to the left of Print in the Operation field.
4-120
Advanced Operations
4 Select the sample or samples you want to print. You can select them by
clicking the Group Selection button or ticking the check boxes directly.
NOTE:
Click the Group Selection button, and a dialog box will pop
up to ask you to enter the start number of the sample.
At the Print Preview screen, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.
4-121
Advanced Operations
6 When you finish printing, click the Close button to exit the Print Preview
screen.
This function will help you print different sample results of the same patient in one
report.
1 Search for the sample results. Refer to To view sample results for
details.
2 Select By sample.
3 Click the option button to the left of Print in the Operation field.
4 Select the sample or samples (5 is the maximum) you want to print. You
can select them by clicking the Group Selection button or ticking the
check boxes directly.
5 Click the Combined Print button to pop up the Print Preview screen.
4-122
Advanced Operations
6 If you have intended to print out all the results, go directly to the next step;
Otherwise, click the Selected Test button in the preview window to select
the tests to be printed out. Then after clicking the OK button, click the
Hide Selecting Test button.
7 At the Print Preview screen, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.
8 When you finish printing, click the Close button to exit the Print Preview
screen.
1 Search for the sample results. Refer to To view sample results for
details.
2 Select By test.
3 Click the option button to the left of Print in the Operation field.
4-123
Advanced Operations
4 Select the sample run or runs you want to print. You can select them by
clicking the Group Selection button or ticking the check boxes directly.
NOTE:
Click the Group Selection button, and a dialog box will pop
up to ask you to enter the start number of the run.
At the Print Preview screen, you may choose Pr. Series (series printing)
for your impact printer; you may choose Pr. Current (printing current
page) or Pr. All (printing all pages) for other types of printers.
4-124
Advanced Operations
6 When you finish printing, click the Close button to exit the Print Preview
screen.
1 Search for the sample results. Refer to To view sample results for
details.
NOTE:
The End (end date) must be in the current year. Otherwise,
you cannot edit manual results.
2 Select By sample.
3 Click the option button to the left of Edit manual test in the Operation
field.
9 If you want to save the setting, click the Save button; otherwise, click the
Cancel button.
NOTE:
Sample results can only be edited by authorized personnel.
If an unauthorized personnel wants to edit sample results, a dialog box
will pop up to ask you to enter authorized user name and password.
4-125
Advanced Operations
1 Search for the sample results. Refer to To view sample results for details.
NOTE:
The End (end date) must be in the current year. Otherwise,
you cannot edit sample results.
2 Select By sample.
3 Click the option button to the left of Edit sample test in the Operation field.
4 Select a sample.
8 If you want to save the setting, click the Save button; otherwise, click the
Cancel button.
NOTE:
Sample results can only be revised by authorized personnel.
If an unauthorized personnel wants to revise sample results, a dialog
box will pop up to ask you to enter authorized user name and
password.
4-126
Advanced Operations
1 Search for the sample results. Refer to To view sample results for details.
NOTE:
The End (end date) must be in the current year. Otherwise,
you cannot revise sample results.
2 Select By test.
3 Click the option button to the left of Revise in the Operation field.
7 If you want to revise the results, click the Save button; otherwise, click the
Cancel button.
1 Search for the sample results. Refer to To view sample results for details.
NOTE:
The End (end date) must be in the current year. Otherwise,
you cannot send sample results to the LIS in batch mode.
2 Select By sample.
3 Click the option button to the left of LIS in the Operation field.
4 Select a sample or samples. You can select them by clicking the Group
Selection button or ticking the check boxes directly.
4-127
Advanced Operations
5 Click the Send Result button to send the results of the selected samples to
the LIS.
When the results are sent, a dialog box will pop up.
NOTE:
There are two modes to send information to the LIS, the real-time
mode and the batch mode.
You can send information to the LIS in batch mode at any time only
except that when you select the Real Time Mode at the Test Settings
screen and the system is testing.
1 Click the Close button to exit the Result & Statistics screen.
The Test Results tab is where you view the statistic information of sample results.
4-128
Advanced Operations
As Figure 4-44 shows, the search conditions are listed on the upper left of the
screen, such statistic data as SD and Mean on the lower left, expected range on the
upper right and statistic graph on the middle right.
NOTE:
To specify the age range of the patients, enter the lower limit
of the range into the edit box to the right of Youngest and
the upper limit into the edit box to the right of Oldest.
If you enter 0s into both the boxes, it means there is no
specific requirement on the ages.
2 Click the Search button and the analyzer will search for the matches and
present a statistic graph of the results found.
1 Click the Close button to exit the Results & Statistics screen.
4-129
Advanced Operations
The Test Status tab is where you view the status of the tests and the reagents.
1 Click the Close button to exit the Results & Statistics screen.
4-130
Advanced Operations
4.4 System
Setup
Hospital
User
Log
Maintenance
Temperature
Database
4.4.1 Status
Click System→ System Status or press CTRL + P to enter the Status screen, as
Figure 4-46 shows.
4-131
Advanced Operations
Status
Dark Current
As Figure 4-47 shows, you can check the status of the major components of the
analyzer.
Note that the light intensity of the lamp is expressed in Lamp Status. The bar to the
right of the Lamp Status represents the light intensity of the lamp in term of
percentage, light intensity of a new lamp being 100%. When the light intensity is
normal, the bar is green. When the light intensity is more than 50%, the bar is yellow.
When the lamp is to be changed, the bar is red. If the system does not detect the
light of the lamp, the bar is white.
4-132
Advanced Operations
As Figure 4-48 shows, you can not only view the dark currents of various
wavelengths, but also check the movement of the cuvette loader and wash the
sample and reagent probes and mixing bar.
The Background field shows the dark currents and backgrounds of the nine
wavelengths and one reference wavelength. The dark current refers to the AD output
of the wavelength when the lamp of the photometer is not switched on. The
backgrounds refer to the AD outputs generated by the lights when there are no
cuvettes in the optical paths.
1 When the analyzer is on standby, click the Test Dark Cur. button and the
readings will be displayed on the screen.
1 When the analyzer is on standby, select the desired position and click the
Load Cuv. button to load/unload the segment to/from the selected
position.
To wash the sample and reagent probes and the mixing bar
1 When the analyzer is on standby, click the Wash Probe & Bar button and
the analyzer will use acid and alkaline detergents to wash the sample and
reagent probes and the mixing bar.
4-133
Advanced Operations
4.4.2 Setup
The Test Settings screen is where you set options regarding the routine settings,
re-runs, test units and LIS connection.
To enter the Test Settings screen, click System → Setup, or press Ctrl + F3. A
dialog box will then pop up to ask you to enter your user name and password. You
must have the corresponding authority to enter the Test Settings screen.
NOTE:
Only authorized personnel can enter the Test Settings screen.
At the Test Settings screen, shown in Figure 4-49, you can see the following tabs:
Routine Settings
Re-run Settings
Test Units
LIS Setting
4-134
Advanced Operations
As Figure 4-49 shows, there are several fields you can set at the Routine Settings
tab. Follow the instructions given below to do so.
NOTE:
Set a suitable value for the insufficient reagents alarm so that the
analyzer can alert you about insufficient reagents in time.
When you have finished setting the desired fields, click the Save button to save the
changes. A dialog box will then pop up to remind you the settings are saved, as
Figure 4-50 shows. Click the OK button to close the dialog box.
4-135
Advanced Operations
If you entered an improper value for any field, a dialog box will pop up to warn you
about the wrong setting after you clicked the Save button, as Figure 4-51 shows.
Seeing the dialog box, click OK button to close it and re-enter a proper value for the
field in question (insufficient reagents alarms, in case of Figure 4-51) and then save
the settings again.
When you have finished setting, you can click the Close button to exit the Test
Settings screen.
4-136
Advanced Operations
Click the Re-run Settings tab to view the 9 available fields in it, as Figure 4-52
shows. Follow the instructions given below to set them.
4-137
Advanced Operations
NOTE:
For the Dilution and Sample vol. Fields, besides the restrictions
mentioned above, you must also guarantee the diluted sample is no
less than 180µl and no more than 450µl.
When you have finished setting the desired fields, click the Save button to save the
changes. A dialog box will then pop up to remind you the settings are saved, as
Figure 4-53 shows. Clicks OK button to close the dialog box.
If you entered an improper value for any filed, a dialog box will pop up to warn you
about the wrong setting after you clicked the Save button. Seeing the dialog box,
click OK to close it and then re-enter a proper value for the field in question and then
save the settings again.
When you have finished setting , you can click Close to exit the Test Settings
screen.
4-138
Advanced Operations
Click the Test Units tab (Figure 4-54) to set the units that will be used for the tests.
Note that all the units to be selected at other screens must be entered here.
Follow the instructions given below to add, edit or delete test units.
1 Click “▼” to the right of Test Units to display the pull-down list and select
the desired unit.
2 Click the Edit button.
3 Edit the unit in the edit box.
4 Click the Save button to save the changes; Click the Cancel button, for
any reason, to ignore the change.
4-139
Advanced Operations
1 Click “▼” to the right of Test Units to display the pull-down list and select
the desired unit.
2 Click the Delete button to delete the selected unit.
NOTE:
Once you click the Delete button, the software will immediately delete
the selected unit without asking you to confirm the deletion. Ensure the
right unit is selected before clicking the Delete button.
The LIS Setting tab is where you set the LIS connection mode and the network
parameters.
1 If you select the Real Time Mode in the LIS Connection Mode Setting
field, the system will automatically send the sample results to the LIS
when it is testing.
If you do not select the Real Time Mode in the LIS Connection Mode
Setting field, you can only send the sample results to the LIS in batch
4-140
Advanced Operations
mode at the Result & Statistics screen. Refer to 4.3.4.1 Test Data tab
for details.
2 Enter the IP address of the server on the right of Server IP Address in
the Network Setting field.
3 Enter the port on the right of Port in the Network Setting field.
4 Click the Save button to save the settings.
NOTE:
There are two modes to send information to the LIS, the real-time
mode and the batch mode.
You can send information to the LIS in batch mode at any time only
except that when you select the Real Time Mode at the Test Settings
screen and the system is testing.
NOTE:
If the system is not connected to the LIS, the Test and the Connect
buttons are available; otherwise, the Disconnect button is available.
1 If you do not want to check the connection between the system and the
LIS, go directly to the next step;
Otherwise, click the Test button to check the network connection. The
system will pop up a dialog box to tell you whether the connection is
successful.
2 If you want to connect the system to the LIS, click the Connect button;
Otherwise, click the Disconnect button.
4.4.3 Hospital
The Hospital Information screen is where you set the options regarding the
hospital and doctors.
To enter the Hospital Information screen, click System → Hospital, or press CTRL
+ H. A dialog box will then pop up to ask you to enter your user name and password.
You must have the administrator authority to enter the Hospital Information screen.
NOTE:
You need the administrator authority to enter the Hospital
Information screen.
4-141
Advanced Operations
At the Hospital Information screen, shown in Figure 4-56, you can see three tabs:
Hospital
Doctors
All Doctors
As Figure 4-56 shows, you can set hospital name and departments at this tab. The
hospital name you set here will appear on the patients’ assay reports, and all the
departments to be selected at other screens must be entered here.
4-142
Advanced Operations
3 Click the Save button in the Department field to save the entered name;
Click the Cancel button, for any reason, to discard the entered name.
1 Click “▼” to the right of Department to display the pull-down list and
select the desired department.
2 Click the Edit button in the Department field.
3 Edit the selected department name in the edit box to the right of
Department.
4 Click the Save button in the Department field to save the change; Click
the Cancel button, for any reason, to ignore the change.
1 Click “▼” to the right of Department to display the pull-down list and
select the desired department.
2 Click the Delete button in the Department field to delete the selected
department.
NOTE:
Once you click the Delete button, the analyzer will immediately delete
the selected department, as well as all the doctors (see the next
section) belonging to this department without asking you to confirm the
deletion. Ensure the right department is selected before clicking the
Delete button.
4-143
Advanced Operations
Click the Doctors tab to show the available fields, as Figure 4-57 shows. All the
doctors to be selected at other screens must be entered here.
There are four buttons at the middle part of the Doctors tab, as Figure 4-58 shows.
Button Function
Displays the information of the first doctor of the selected
department.
Displays the information of the previous doctor of the selected
department.
Displays the information of the next doctor of the selected
department.
Displays the information of the last doctor of the selected
department.
If there is no existing doctor information in the selected department, the arrows will
appear gray, indicating they are not usable at this moment.
Follow the instructions given below to add, edit or delete doctors to a selected
department.
4-144
Advanced Operations
1 Click “▼” to the right of Department to display the pull-down list and
select the desired department.
2 Click the Add button and the analyzer will automatically assign a number
to the new doctor. Note that you cannot change this number.
3 Enter the doctor’s name in the edit box to the right of Name.
4 Enter the doctor’ s duty into the edit box to the right of Duty.
5 Click the Save button to save the change; Click the Cancel button, for
any reason, to ignore the change.
1 Click “▼” to the right of Department to display the pull-down list and
select the desired department.
2 Click the arrows to find the desired doctor information if necessary.
3 Click the Edit button.
4-145
Advanced Operations
5 Click the Save button to save the change; Click the Cancel button, for
any reason, to ignore the change.
1 Click “▼” to the right of Department to display the pull-down list and
select the desired department.
2 Click the arrows to find the desired doctor information if necessary.
3 Click the Delete button to delete the information of the selected doctor.
NOTE:
Once you click the Delete button, the analyzer will immediately delete
doctor from the selected department without asking you to confirm the
deletion. Ensure the right doctor is selected before clicking the Delete
button.
Click the All Doctors tab to view all the saved doctor information, as Figure 4-59
shows.
To view the information of all the doctors, you can click “ ▼ ” to the right of
Department and select All departments.
To view the information of doctors of a specific department, click “▼” to the right of
Department and select the desired department.
4-146
Advanced Operations
You can click the Close button to exit the Hospital Information screen.
4.4.4 User
Click System → User or press CTRL + B and a dialog box will pop up, as Figure
4-60 shows, to ask you to enter the appropriate user name and password.
Figure 4-60 A dialog box to ask you enter the appropriate user name and password
Enter a user name and a password of the administrator authority to enter the User
Management screen, as Figure 4-61 shows.
NOTE:
Only users of the administrator authority can enter the User
Management screen.
Administrator
As Figure 4-61 shows, you can check the existing users or user groups at the User
Management screen.
4-147
Advanced Operations
1 Click User Management → Add User to enter the Add User window.
4-148
Advanced Operations
3 Click the OK button to save the changes; click the Cancel button, for any
reason, to ignore the changes.
Administrator
2 Click User Management → Delete User and a dialog box will pop up to
ask you to confirm the deletion.
3 Click the Yes button to confirm the deletion; click the No button to abort
the deletion.
4-149
Advanced Operations
Administrator
2 Click User Management → Edit User to enter the Edit User window.
4-150
Advanced Operations
2 Enter the user name and password as required and click the OK button
and another dialog box will pop up to ask you to enter the new user name
and password.
4 Click the OK button to save the changes; click the Cancel button, for any
reason, to ignore the changes.
4-151
Advanced Operations
1 Click on the upper right corner of the screen, or press ALT+ X, or click
User Management → Exit to exit the User Management screen.
4.4.5 Log
Click System → Log or press CTRL + F2 to enter the Log screen, as Figure 4-62
shows.
As Figure 4-62 shows, the Log screen lists as many as 300 alarm messages. Once
the saved alarm messages reach 300, the analyzer will automatically save the
messages into a .log file and the file name is the saving time.
4-152
Advanced Operations
1 Click File → Current to list the latest 300 (or less) alarm messages.
Look in:
Files of Type:
Cancel
2 Select the desired log file from the “Log” folder under the directory the
BS-300 control system is installed and click the Open button to open the
selected log file.
1 Click File → Clear and a dialog box will pop up to ask you to save the
current log to a .log file.
4-153
Advanced Operations
2 Click the Yes button to save the log and then clear it; click the No button
to clear the log without saving it.
1 Click File → Exit or on the upper right corner of the screen to exit the
Log screen.
4.4.6 Maintenance
Click System → Maintenance or press CTRL + Q and a dialog box will pop up, as
Figure 4-63 shows, to ask whether you are maintenance personnel.
Figure 4-63 A dialog box to ask you whether you are maintenance personnel
Click the No button and another dialog box will pop up to ask you to enter the
appropriate username and password, as Figure 4-64 shows.
Figure 4-64 A dialog box to ask you to enter username and password
Enter a username and password of the administrator authority to enter the System
Maintenance screen, as Figure 4-65 shows.
4-154
Advanced Operations
At the System Maintenance screen, the Sent Instruction field and the Received
Data field respectively display the instruction codes sent to the Analyzing Unit and
received by the Operation Unit.
NOTE:
Do not perform any maintenance action unless you are certain the
analyzer is on standby.
To lock data
1 If you feel the codes displayed in the Sent Instructions and Received
Data fields refresh too fast, click the Lock button, which then will become
the Unlock button, and the displayed data will stop changing. You can
click the Unlock button to resume the data refreshing.
To clear data
1 Click the Clear button to clear the Sent Instructions and Received Data
fields.
4-155
Advanced Operations
4-156
Advanced Operations
NOTE:
Ensure the sample disk is placed in the home position or stopped at
certain tube position before clicking Into Inner Circle or Into Outer
Circle button. Otherwise, it may lead to probe collision.
Ensure the reaction disk is placed in the home position or stopped at
certain cuvette position before clicking Into Reaction Cuvette button.
Otherwise, it may lead to probe collision.
4-157
Advanced Operations
NOTE:
Ensure the reagent disk is placed in the home position or stopped at
certain reagent position before clicking Into Inner Circle or Into Outer
Circle button. Otherwise, it may lead to probe collision.
Ensure the reaction disk is placed in the home position or stopped at
certain cuvette position before clicking Into Reaction Cuvette button.
Otherwise, it may lead to probe collision.
4-158
Advanced Operations
NOTE:
Ensure the reaction disk is placed in the home position or stopped at
certain cuvette position before clicking Into Reaction Cuvette or Mix
button. Otherwise, it may lead to bar collision.
4-159
Advanced Operations
NOTE:
Before operating the reaction disk, ensure the sample and reagent
probes and mixing bar are away from it. Otherwise, the moving disk
may bend the probes or bar.
4-160
Advanced Operations
NOTE:
Before operating the reagent disk, ensure the reagent probe is away
from it. Otherwise, the moving disk may bend the probes.
4-161
Advanced Operations
NOTE:
Before operating the sample disk, ensure the sample probe is away
from it. Otherwise, the moving disk may bend the probe.
4-162
Advanced Operations
NOTE:
When you have clicked the Sam. Syringe Dispense button, if there is
no enough liquid inside the sample syringe, the plunger will stop at the
home position.
When you have clicked the Reag. Syringe Dispense button, if there is
no enough liquid inside the reagent syringe, the plunger will stop at the
home position.
4-163
Advanced Operations
4-164
Advanced Operations
NOTE:
If nothing is to be run for a while and you have no intention to exit the
control system, you can turn off the lamp to maximize its service life.
4-165
Advanced Operations
4-166
Advanced Operations
4.4.7 Print
The Printing Setup screen is where you set the parameters regarding printing
reports.
To enter the Printing Setup screen, click System → Print, or press CTRL + O.
4-167
Advanced Operations
At the Printing Setup screen, Figure 4-77, there are four tabs:
As Figure 4-78 shows, the items listed in the Required field are required to appear
on the patient report; the items listed in the Optional field are selectable items and
you can select any interested items to print on the patient report.
4-168
Advanced Operations
1 Tick the check boxes to the left of the desired items. Clicking the ticked
check box again can cancel the selection.
2 Click the Setup button and a dialog box will pop up to tell you settings are
saved. Click the OK button to close the box.
4-169
Advanced Operations
As Figure 4-79 shows, the items listed in the Required field are required to appear
on the test report; the items listed in the Optional field are selectable items and you
can select any interested items to print on the report.
1 Tick the check boxes to the left of the desired items. Clicking the ticked
check box again can cancel the selection.
2 Click the Setup button and a dialog box will pop up to tell you settings are
saved. Click the OK button to close the box.
4-170
Advanced Operations
2 Select the desired sequence number from the pull down list above.
1 Select the desired printer from the pull down list below Select.
4-171
Advanced Operations
2 Click the Default Printer button to set the selected printer as the default
printer.
NOTE:
The format of the printout may not be exactly the same as that
displayed in the print preview window.
1 Enter the desired header into the edit box to the right of Header.
2 Select the desired font from the pull down list to the right of Font.
3 Click the Setup button and a dialog box will pop up to tell you the settings
that have been saved. Click the OK button to close the box.
4-172
Advanced Operations
4.4.8 Temperature
Click System→ Temperature or press CTRL + T and a dialog box will pop up, as
Figure 4-82 shows, to ask you whether you want to debug the temperature control
unit or check the temperature curve.
Figure 4-82 A dialog box to ask you whether to debug or to view curves
Click the No button and another dialog box will pop up to ask you to enter
appropriate username and password, as Figure 4-83 shows.
CAUTION:
Only service engineers are authorized to debug the temperature
control unit.
Figure 4-83 A dialog box to ask you to enter username and password
4-173
Advanced Operations
As Figure 4-84 shows, you can see two tabs at the Temperature Control Curve
screen:
For every temperature curve, its x-axis represents the time and y-axis the
temperature.
4-174
Advanced Operations
NOTE:
Click to check the previous curve;
To lock a curve
1 Click the Lock Curve button to lock the currently displayed curve.
To unlock a curve
1 Click the Unlock Curve button to unlock the currently locked curve.
1 Click the Close button to exit the Temperature Control Curve screen.
4.4.9 Database
Click System → Database or press CTRL + L and a dialog box will pop up, as
Figure 4-85 shows, to ask you to enter the appropriate user name and password.
Figure 4-85 A dialog box to ask you enter the user name and password
Enter the user name and password of the administrator authority to enter the
Database Maintenance screen as Figure 4-86 shows.
4-175
Advanced Operations
NOTE:
You can only maintain the database when the analyzer is on standby.
This analyzer automatically backs up data. You only need to maintain
the database when the system alerts you that the database is
damaged.
Usually you do not need to delete the reaction curves. Deleting the
curves (only curves, not the results) only when you find insufficient
space left on the hard disk.
1 When the analyzer warns you about database damage, click the Repair
and Compress button to repair and compress current database.
1 Select the desired start date from the pull down list to the right of the
Start.
2 Select the desired end date from the pull down list to the right of the End.
3 Click the Delete certain reaction curve data button to delete the curve of
the selected period.
1 Click the Delete all reaction curve data button to delete all the curves.
1 Click the Close button or the button at the upper right corner to exit
the Database Maintenance screen.
4-176
Advanced Operations
4.5 Help
Help Version
Guide
4.5.1 Version
Click Help→ Version or press CTRL + S to enter the Software version screen, as
the following figure shows.
4.5.2 Guide
Click Help→ Guide or press CTRL + F11 to view the on-line help document.
4-177
5 Maintenance
To ensure reliability, good performance and long service life of the system, regular
maintenance is required. Be sure to follow the instructions given below to maintain
the system. In case of problems beyond your ability or not covered in this chapter, be
sure to call Mindray Customer Service Department or the distributor in time.
WARNING:
Do not perform any maintenance procedures that are not described
in this chapter.
Performing unauthorized maintenance procedures can damage your
system, void any applicable warranty or service contract and even
cause personnel injuries.
After performing any maintenance actions or procedures, ensure
that the system runs normally.
For some maintenance actions or procedures, be sure to place the
POWER to OFF first.
Do not spill water or detergent on the mechanical or electrical
components of the system.
BIOHAZARD:
Wear gloves, goggles and protective clothing.
5.1 Preparation
The following things may facilitate your maintenance.
Tools
5-1
Maintenance
Detergents
NOTE:
Do not mix the acid detergent with the alkali one.
Be sure to use the detergents recommended by Mindray. Otherwise,
proper result may not be obtained.
Others
Ethanol
Disinfector
T-piece
5-2
Maintenance
During the washing process, check whether the flow from the interior of the sample
probe is continuous and in the direction of the probe; check the exterior of the
sample probe to see whether the flow is continuous and normal.
If the flow appears abnormal, wash the probe as instructed by 5.3.1 Washing
Sample Probe.
If the flow remains abnormal after washing, clean the probe as instructed by 5.6.1
Unclogging Sample Probe.
If the flow remains abnormal after cleaning, contact Mindray Customer Service
Department or the distributor.
During the washing process, check whether the flow from the interior of the reagent
probe is continuous and in the direction of the probe; check the exterior of the
reagent probe to see whether the flow is continuous and normal.
If the flow appears abnormal, wash the probe as instructed by 5.3.2 Washing
Reagent Probe.
If the flow remains abnormal after washing, clean the probe as instructed by 5.6.2
Unclogging Reagent Probe.
If the flow remains abnormal after cleaning, contact Mindray Customer Service
Department or the distributor.
5-3
Maintenance
During the washing process, check whether the bar rotates normally. If not, contact
Mindray Customer Service Department or the distributor.
WARNING:
Always wear gloves when checking the detergents.
1 Make sure the system is on standby status, and ensure no part of the
system will motion.
2 Remove the cover from the reagent disk.
3 Check how much detergent left in the bottle on position 45. If there is not
much left, add more detergent or install a new bottle of detergent.
4 Check how much detergent left in the bottle on position 47. If there is not
much left, add more detergent or install a new bottle of detergent.
5 Replace the cover.
Every wash consumes about 450µl detergent when the system uses detergent to
clean the probes and the mixing bar.
Deionized
Waste Tank
Water Tank
Used-Cuvettes
Bucket
Extractable
Plate
5-4
Maintenance
3 Check the connections between the two deionized water connectors (green
and red) and their counterparts on the analyzer.
Pickup Tube
CAUTION:
Make sure the deionized water pickup tubes are neither blocked nor
bent.
NOTE:
If leaking remains, please contact Mindray Customer Service
Department or the distributor.
5-5
Maintenance
BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the following checks.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
Deionized
Waste Tank
Water Tank
Used-Cuvettes
Bucket
Extractable
Plate
3 Check the connections between waste connector and the waste tap of the
analyzing unit.
Pin
Waste Tap
Waste Tube
5-6
Maintenance
6 Check the connection between the waste tube and the tank cap.
If no leaks, go directly to the next step.
If you see leaks, wipe them off with clean gauze and tighten the
connection and proceed to the next step.
7 Push the extractable plate back in.
8 Close the lower door.
CAUTION:
Ensure the waste tube is neither blocked nor bent. A blocked or bent
waste tube may lead to waste overflow that may damage the analyzing
unit.
NOTE:
If leaking remains, please contact Mindray Customer Service
Department or the distributor.
CAUTION:
The deionized water to be used on the BS-300 must meet the CAP
Type II water requirements and the resistance rate must be no less
than 0.5MΩ·cm.
Deionized
Waste Tank
Water Tank
Used-Cuvettes
Bucket
Extractable
Plate
5-7
Maintenance
5
CAUTION:
After removing the cap of the deionized water tank (together
with the pickup tube and level sensor), place it on a clean
table.
Unscrew (counter-clockwise) the tank cap and remove the cap together
with the pickup tube and the level sensor.
6 Add deionized water to the tank.
7 Screw (clockwise) the cap together with the pickup tube and the level
sensor back on to the tank until secure.
8 Push the extractable plate back in.
9 Close the lower door.
CAUTION:
Ensure the deionized water pickup tube is neither blocked nor bent.
BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below checks.
Dispose of the wastewater in accordance with your local or national
guidelines for biohazard waste disposal and consult the manufacturer
or distributor of the reagents for details.
Deionized
Waste Tank
Water Tank
Used-Cuvettes
Bucket
Extractable
Plate
5-8
Maintenance
4
BIOHAZARD:
After removing the cap of the waste tank (together with the
tube and level sensor), place it on an appropriate place to
avoid biohazard contamination.
Unscrew (counter-clockwise) the tank cap and remove it together with the
waste tube and the level sensor from the tank.
5 Take out the tank and empty it.
6 Replace the tank back into the cabinet.
7 Screw (clockwise) the cap (together with the waste tube and the level
sensor) back onto the tank until secure.
8 Push the extractable plate back in.
9 Close the lower door.
CAUTION:
Ensure the waste tube is neither blocked nor bent. A blocked or bent
waste tube may lead to wastewater overflow that may damage the
analyzer.
BIOHAZARD:
To prevent biohazard contamination, always wear gloves, goggles and
protective clothing when doing the below checks.
Dispose of the used cuvettes in accordance with your local or national
guidelines for biohazard waste disposal.
Deionized
Waste Tank
Water Tank
Used-Cuvettes
Bucket
Extractable
Plate
5-9
Maintenance
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when washing the probe.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
3 Pull the sample probe arm to the highest point by hand. Rotate the probe
arm to a position above the sample compartment and convenient to
operate.
5-10
Maintenance
4
CAUTION:
The tweezers may scratch the probe. Exercise caution when
using it to wash the probe. Avoid direct contact between the
tweezers and the probe. Do not use excessive force when
washing the probe. Otherwise it may bend.
NOTE:
Mindray recommends the acid and alkaline detergents be
used alternately for this purpose. For instance, if the acid
detergent was used for last cleaning, the alkaline detergent
would better be used for this time.
CAUTION:
After washing the probe, be sure to rotate the probe to a
position above the wash well.
7 Keep the sample disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the sample disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor.
To ensure the disk is secured, you may grab the handwheel and gently
pull it upward. If the disk does not loose, it means it is secured and you
can replace the cover; otherwise, re-install the disk.
8 Place the ANALYZING UNIT POWER to ON and the system will
automatically reset the sample probe and wash it with deionized water.
5-11
Maintenance
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when washing the probe.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
3 Pull the reagent probe arm to the highest point by hand. Rotate the probe
arm to a position above the reagent compartment and convenient to
operate.
5-12
Maintenance
4
CAUTION:
The tweezers can scratch the probe. Exercise caution when
using it to wash the probe. Avoid direct contact between the
tweezers and the probe. Do not use excessive force when
washing the probe. Otherwise it may bend.
NOTE:
Mindray recommends the acid and alkaline detergents be
used alternately for this purpose. For instance, if the acid
detergent was used for last cleaning, the alkaline detergent
would better be used for this time.
CAUTION:
After washing the probe, be sure to rotate the probe to a
position above the wash well.
7 Keep the reagent disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the reagent disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor.
To ensure the disk is secured, you may grab the handwheel and gently
pull it upward. If the disk does not loose, it means it is secured and you
can replace the cover; otherwise, re-install the disk.
8 Place the ANALYZING UNIT POWER to ON and the system will
automatically reset the reagent probe and wash it with deionized water.
5-13
Maintenance
BIOHAZARD:
Wear gloves, goggles and protective clothing when washing the probe.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
3
CAUTION:
The tweezers can scratch the bar. Exercise caution when
using the tweezers to wash the bar. Avoid direct contact
between the tweezers and the bar. Do not use excessive
force when washing the bar. Otherwise it may bend.
NOTE:
Mindray recommends the acid and alkaline detergents be
used alternately for this purpose. For instance, if the acid
detergent was used for last cleaning, the alkaline detergent
would better be used for this time.
5-14
Maintenance
CAUTION:
After washing the bar, be sure to rotate the bar to a position
above the wash well.
CAUTION:
The mixing bar is precisely fabricated. Replace the scratched or bent
bar once finding it. Refer to 5.6.5 Replacing Mixing Bar for details. .
CAUTION:
The deionized water to be used on the BS-300 must meet the CAP
Type II water requirements and the resistance rate must be no less
than 0.5MΩ·cm.
Deionized
Waste Tank
Water Tank
Used-Cuvettes
Bucket
Extractable
Plate
5-15
Maintenance
4
CAUTION:
After removing the cap of the deionized water tank (together
with the pickup tube and level sensor), place it on a clean
table.
CAUTION:
Ensure the deionized water pickup tube is neither blocked nor bent.
BIOHAZARD:
To prevent biohazard contamination, wear gloves, goggles and
protective clothing when washing the tank.
Exercise caution and do not spill the waste onto other people or things.
Dispose of the waste in accordance with your local or national
guidelines for biohazard waste disposal and consult the manufacturer
or distributor of the reagents for details.
Dispose of used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
5-16
Maintenance
2 Open the lower door of the analyzing unit and you will see the waste tank
on the left.
Deionized
Waste Tank
Water Tank
Used-Cuvettes
Bucket
Extractable
Plate
Unscrew (counter-clockwise) the cap (together with the waste tube and the
level sensor).
5 Take out the waste tank.
6 Empty the waste tank.
7 Wash the tank interior with clean water. Soak the tank with disinfector if
necessary.
8 Wash the waste tube and the level sensor with clean water.
9 Wipe water off the tank exterior, waste tube and sensor cable with clean
gauze.
10 Place the deionized water tank back onto the extractable plate.
11 Screw (clockwise) the cap (together with the waste tube and level sensor)
back onto the tank until secure.
12 Push the extractable plate back in.
13 Close the lower door.
CAUTION:
Ensure the waste tube is neither blocked nor bent. A blocked or bent
waste tube will lead to waste overflow that may damage the analyzing
unit.
5-17
Maintenance
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the
sample disk/compartment.
Dispose of used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
3 Wash the disk with clean water and wipe it dry with clean gauze.
4 Use clean gauze (water or disinfector-dipped gauze if necessary) to clean
the inside of the compartment.
5 Keep the sample disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the sample disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor. To ensure the disk is secured, you may grab the handwheel
and gently pull it upward. If the disk does not loose, it means it is secured
and you can replace the cover; otherwise, re-install the disk.
5-18
Maintenance
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the
reagent disk/compartment.
Dispose of used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
3 Wash the disk with clean water and wipe it dry with clean gauze.
4 Use clean gauze (water or disinfector-dipped gauze if necessary) to clean
the inside of the compartment.
5 Keep the reagent disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the reagent disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor. To ensure the disk is secured, you may grab the handwheel
and gently pull it upward. If the disk does not loose, it means it is secured
and you can replace the cover; otherwise, re-install the disk.
5-19
Maintenance
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the panel
of analyzing unit.
Dispose of used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the wash
well.
Dispose of used cotton sticks in accordance with your local or national
guidelines for biohazard waste disposal.
5-20
Maintenance
3 Clean the inside of and the place around the wash well with cotton sticks.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the wash
well.
Dispose of used cotton sticks in accordance with your local or national
guidelines for biohazard waste disposal.
5-21
Maintenance
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the wash
well.
Dispose of used cotton sticks in accordance with your local or national
guidelines for biohazard waste disposal.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the rotor.
5-22
Maintenance
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the rotor.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
5-23
Maintenance
BIOHAZARD:
Wear gloves, goggles and protective clothing when cleaning the rotor.
5-24
Maintenance
5 Use a screwdriver to unscrew the four screws fixing the screens on the
back cover and remove the screens.
6 Wash the screens with clean water.
7 Wipe the screens dry with clean gauze.
8 Install the screens back to the back cover and the middle door with
screws.
9 Close the middle door.
10 Close the back cover and fix it with three screws.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when removing the
probe.
5-25
Maintenance
3 Pull the sample probe arm to its highest point. Rotate the probe to a
position above the sample disk and convenient to operate.
4 Pinch the clips below the base of the sample probe arm and push it
upward to remove the arm cover.
5 Hold the probe’s fluid connector of probe with one hand and the tubing
connector with the other. Rotate tubing connector counter-clockwise until it
disconnects from the probe. Remove the tubing from the probe.
5-26
Maintenance
6 Press the circuit board with one hand and disconnect the probe’s circuit
connector from the board with the other hand.
CAUTION:
The probe arm is delicate. Exercise caution when
disconnecting the connector. Excessive force may damage
the connector and/or the circuit board.
7 Use a screwdriver to remove the retaining screw and take out the spring.
5-27
Maintenance
8
WARNING:
Save the removed probe at a safe place where it will neither
endanger people working around the area nor be damaged.
NOTE:
Exercise caution when pulling the probe away from the arm
so that the probe tip will not contact or even damage the
probe arm.
Slowly pull the probe away from the probe arm. Exercise caution so that
the gasket inside the probe will not drop out and if it does drop out, save it
at a clean place for later installation.
NOTE:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.3 Replacing Sample Probe for details.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when unclogging the
probe.
Dispose of the used needle in accordance with your local or national
guidelines for biohazard waste disposal.
CAUTION:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.3 Replacing Sample Probe for details.
5-28
Maintenance
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when installing the probe.
CAUTION:
Exercise caution when connecting the probe. Excessive
force may bend the probe.
8 Add deionized water to a clean cup. Immerse the probe tip into the water
by 2-3mm and the LED on the circuit board should be lighted. Take the
probe tip out of water, and the LED should go out. If the test succeeds,
proceed to the next step; if not, please contact Mindray Customer Service
Department or the distributor.
5-29
Maintenance
9 Replace the arm cover as follows: align the hole of pin of the arm cover to
the bump on the rear side (near the rotor) of the base; align the two clasps
of the arm cover to the two holes on the base; gently push the cover back
onto the base.
10 Pinch the probe by the part near the probe arm. Gently push the probe
upward and then release the probe to see if it can move freely. If so,
proceed to the next step; if not, the cover is not properly installed. Remove
the cover and re-install it and try this step again.
11 Pull the sample probe arm to its highest point. Rotate the probe to a
position above the wash well.
CAUTION:
After installing the sample probe, be sure to rotate the probe
to a position above the wash well.
12 Keep the sample disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the sample disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor. To ensure the disk is secured, you may grab the handwheel
and gently pull it upward. If the disk does not loose, it means it is secured
and you can replace the cover; otherwise, re-install the disk.
13 Place the ANALYZING UNIT POWER to ON. The system will automatically
reset the probe and wash it with deionized water.
5-30
Maintenance
CAUTION:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when removing the
probe.
3 Pull the reagent probe arm to its highest point. Rotate the probe to a
position above the reagent disk and convenient to operate.
5-31
Maintenance
4 Pinch the clips below the base of the reagent probe arm and push it
upward to remove the arm cover.
5 Hold the probe’s fluid connector with one hand and the tubing connector
with the other. Rotate tubing connector counter-clockwise until it
disconnects from the probe. Remove the tubing from the probe.
6 Press the circuit board with one hand and disconnect the probe’s circuit
connector from the board with the other hand.
CAUTION:
The probe arm is delicate. Exercise caution when
disconnecting the connector. Excessive force may damage
the connector and/or the circuit board.
5-32
Maintenance
7 Use a screwdriver to remove the retaining screw and take out the spring.
8
WARNING:
Save the removed probe at a safe place where it will neither
endanger people working around the area nor be damaged.
NOTE:
Exercise caution when pulling the probe away from the arm
so that the probe tip will not contact or even damage the
probe arm.
Slowly pull the probe away from the probe arm. Exercise caution so that
the gasket inside the probe will not drop out and if it does drop out, save it
at a clean place for later installation.
NOTE:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.4 Replacing Reagent Probe for details.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when unclogging the
probe.
Dispose of the used needle in accordance with your local or national
guidelines for biohazard waste disposal.
5-33
Maintenance
NOTE:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.4 Replacing Reagent Probe for details.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when installing the probe.
5-34
Maintenance
7
CAUTION:
You must shape the tubing inside the reagent probe arm into
a ring.
5-35
Maintenance
9 Replace the arm cover as follows: align the hole of pin of the arm cover to
the bump on the rear side (near the rotor) of the base; align the two clasps
of the arm cover to the two holes on the base; gently push the cover back
onto the base.
10 Pinch the probe by the part near the probe arm. Gently push the probe
upward and then release the probe to see if it can move freely.
If so, proceed to the next step.
If not, the cover is not properly installed. Remove the cover and re-install it
and try this step again.
5-36
Maintenance
11 Pull the reagent probe arm to its highest point. Rotate the probe to a
position above the wash well.
CAUTION:
After installing the reagent probe, be sure to rotate the probe
to a position above the wash well.
12 Keep the reagent disk handle in the vertical position. Align the central hole
of the handwheel to the rotor and lower the reagent disk all the way down.
Rotate the disk a bit as necessary during the lowering process. When the
disk is in place, shift the handle back to horizontal position to lock the disk
to the rotor. To ensure the disk is secured, you may grab the handwheel
and gently pull it upward. If the disk does not loose, it means it is secured
and you can replace the cover; otherwise, re-install the disk.
13 Place the ANALYZING UNIT POWER to ON. The system will automatically
reset the probe and wash it with deionized water.
NOTE:
The probe tip is precisely fabricated. A bent or damaged probe will lead
to unreliable test results and should be replaced immediately upon
seeing. Refer to 5.6.4 Replacing Reagent Probe for details.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when replacing the
probe.
CAUTION:
Please use Mindray-recommended consumables. Other consumables
may degrade the system performance.
BIOHAZARD:
Dispose of the bent or damaged sample probe in accordance
with your local or national guidelines for biohazard waste
disposal.
5-37
Maintenance
CAUTION:
After installing the sample probe, be sure to rotate the probe
to a position above the wash well before installing the
sample disk.
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury to your hands or arms, exercise caution when working around
the probe.
BIOHAZARD:
Wear gloves, goggles and protective clothing when replacing the
probe.
CAUTION:
Please use Mindray-recommended consumables. Other consumables
may degrade the system performance.
BIOHAZARD:
Dispose of the bent or damaged reagent probe in
accordance with your local or national guidelines for
biohazard waste disposal.
5-38
Maintenance
CAUTION:
You must shape the tubing inside the reagent probe arm into
a ring.
WARNING:
The sample probe tip is sharp and can cause puncture wounds. To
prevent injury to your hands or arms, exercise caution when working
around the probe.
When replacing the bar, pinch the bar only by the knurled part and do
not touch the other part of the bar. Protect the flat par of the mixing bar
from scratches.
BIOHAZARD:
Wear gloves, goggles and protective clothing when replacing the
mixing bar.
Dispose of the damaged mixing bar in accordance with your local or
national guidelines for biohazard waste disposal.
CAUTION:
Please use Mindray-recommended consumables. Other consumables
may degrade the system performance.
5-39
Maintenance
3 Gently pull the bar to its highest point and rotate it to a position convenient
to operate.
4 Pinch the bar with one hand and unscrew (counter-clockwise) the retaining
nut with the other until the mixing bar looses. Pull the bar downward to
remove it and remove the nut.
CAUTION:
When trying to pull out the bar, concentrate your force in the
direction of the axis on the bar arm. Biased force may
damage the bar and/or the axis.
5-40
Maintenance
5 Align the mixing bar to the bigger hole of the retaining nut and gently
screw it into the nut until the end of the bar is in line with the smaller hole
of the nut.
6 Pinch the mixing bar by the knurled part and align the hole of the nut to the
axis of the mixing bar and push the nut onto the bar until it reaches the
end of the bar. Tighten the nut (screw clock-wise) with the other hand.
CAUTION:
When trying to push the bar into the arm, concentrate your
force in the direction of the axis on the bar arm. Biased force
may damage the bar and/or the axis.
Ensure the bar is all the way pushed to the end.
7 After replacing the bar, visually check whether the bar is vertical to the bar
arm.
If yes, proceed to the next step.
If not, repeat step 5 to remove the bar and re-install it.
8 Pull the bar arm to its highest point and rotate it back to a position above
its wash well.
CAUTION:
After installing the mixing bar, be sure to rotate the mixing
bar to a position above its wash well before turning on the
analyzing unit.
9 Place the ANALYZING UNIT POWER to ON. The system will automatically
reset the bar and wash it with deionized water.
10 Follow the steps given below to do the position checks:
Enter the Maintenance screen and select the Mix. Bar tab as instructed
by 4.4.6.3 Mix. Bar tab.
5-41
Maintenance
Click Into Wash Well and observe whether the mixing bar is aligned to the
center of the wash well. If not, please contact Mindray Customer Service
Department or the distributor.
Set the Time to 100 and click Mix. The bar should start to rotate after
enters the reaction cuvette. Observe whether the rotation is normal and
pay attention to abnormal sound. If the rotation is not normal or the
abnormal noise is heart, please contact Mindray Customer Service
Department or the distributor.
CAUTION:
Please use Mindray-recommended consumables. Other consumables
may degrade the system performance.
Exercise caution when installing the syringe. Excessive force may
crack the syringe.
Follow the procedure given below to replace the plunger assembly of sample or
reagent syringe.
5-42
Maintenance
T-piece
Syringe Connector
Space Bar
Syringe
Holder
Plunger Button
3 Prepare a new plunger assembly (shown in the figure below) and soak the
plunger tip in deionized water to eliminate bubbles.
Grab the T-piece with one hand and the syringe connector with the other
hand and unscrew (counter-clockwise) the syringe.
5-43
Maintenance
7
CAUTION:
There may be residual water in the syringe. Exercise caution
when pulling the plunger assembly out. Do not drop water onto
the analyzing unit.
The plunger rod of the sample syringe is slender. Exercise
caution when working on it. Excessive force may bend it.
Unscrew (counter-clockwise) the plunger guide cap and pinch the plunger
button to gently pull the plunger assembly from the syringe.
8
CAUTION:
The plunger rod of the sample syringe is slender. Exercise
caution when working on it. Excessive force may bend it.
Pinch the new plunger assembly by the plunger button and carefully insert
the plunger tip into the syringe and push it all the way to the end. Screw
(clockwise) the plunger guide cap until secure.
9 Immerse the syringe connector into deionized water. Pinch the plunger
button, pull it to aspirate half syringe of deionized water and then push it to
expel the deionized water and the air from the syringe.
10 Grab the T-piece with one hand and the syringe connector with the other
hand. Screw (clockwise) the syringe into the T-piece until secure.
11 Place the syringe on the holder.
12 Install the space bar and two upper retaining screws. Note that do not
tighten the screws now.
13 Align the plunger button to the lower retaining screw of the plunger and
screw (clockwise) the screw until secure.
14 Pinch the plunger guide cap to adjust the syringe. Mindray recommends the
upper edge of the space bar reach the fourth line (the 4µl line of the sample
reagent, or the 40µl line of the reagent syringe) of the scale on the syringe.
15 Tighten the two upper retaining screws.
16 Place the ANALYZING UNIT POWER back to ON.
5-44
Maintenance
CAUTION:
Please use Mindray-recommended consumables. Other consumables
may degrade the system performance.
NOTE:
Do not touch either the light entrance of the lamp housing or the lens in
front of the lamp. In case the entrance is dirty, clean it with
ethanol-soaked defatted cotton.
1 Place the POWER to OFF. Wait at least 30 minutes for the lamp and its
housing to cool down.
WARNING:
After working for a while, the lamp and its housing are
usually hot enough to burn you. Do not proceed with this
procedure until they have cooled down.
5-45
Maintenance
3 Unscrew the two lower screws on the back of the lamp housing and
remove the radiator from the housing.
4 Pinch the lamp base with one hand and press the lever on the left of the
base.
5 Pinch the new lamp by its base and insert the lamp pins into the mounting
holes, the larger pin into the larger hole and the smaller pin into the
smaller hole, on the base.
6 Replace the radiator and tighten the screws.
5-46
Maintenance
7 Close the upper part of the back of the analyzing unit and tighten the
screws.
5-47
Maintenance
5-48
Maintenance
5-49
6 Troubleshooting
This chapter presents all error messages and recommended measures, which
should be taken in time once any error occurs.
If the recommended measures fail to solve the problems, contact Mindray Customer
Service Department or your local distributor.
When an error occurs, its corresponding error message will be displayed on the
lower left side of the screen, as Figure 6-1 shows.
The log will record the time, level, code and message of every error (except for the
level-11 error) to facilitate you to troubleshoot the analyzer. Refer to 4.4.5 Log for
details about the log.
Figure 6-1 Error messages are displayed on the lower left side of the screen
6-1
Troubleshooting
Severity: Warning
6-2
Troubleshooting
6-3
Troubleshooting
NOTE:
All errors except for those of level 11 will be recorded in the log. When
a level-11 error occurs, the operating system will display the error
message and exit the system software after the error is confirmed by
the operator.
WARNING:
When troubleshooting the analyzer, first find out whether it is
necessary to switch off the POWER or ANALYZING UNIT POWER.
Wear gloves, goggles and lab coat if necessary.
6-4
Troubleshooting
100000002 MUI Response Error: Instruction too long. 9 Contact Mindray Customer Service Department or your local distributor.
100000003 MUI Response Error: Incomplete instruction. 9 Contact Mindray Customer Service Department or your local distributor.
100000004 MUI Response Error: Check sum error. 9 Contact Mindray Customer Service Department or your local distributor.
100000005 MUI Response Error: Downloading parameter. Can't 9 Contact Mindray Customer Service Department or your local distributor.
respond to other instructions.
100000006 MUI Response Error: Data at byte 00-0x7F exceed 9 Contact Mindray Customer Service Department or your local distributor.
0x7F.
100960100 MU Execution Error: Wrong command. Illegal frame 9 Contact Mindray Customer Service Department or your local distributor.
type.
100960101 MU Execution Error: Wrong command. No this unit's 9 Contact Mindray Customer Service Department or your local distributor.
command.
100960102 MU Execution Error: Wrong command. Illegal 9 Contact Mindray Customer Service Department or your local distributor.
command identity.
100960103 MU Execution Error: Wrong command. Parameter 9 Contact Mindray Customer Service Department or your local distributor.
significant bits wrong
100960104 MU Execution Error: Wrong command. Instruction 9 Contact Mindray Customer Service Department or your local distributor.
length wrong.
100960105 MU Execution Error: Wrong command. Sample 9 Contact Mindray Customer Service Department or your local distributor.
position not within 1-60
100960106 MU Execution Error: Wrong command. Reagent 9 Contact Mindray Customer Service Department or your local distributor.
position not within 1-50
100960107 MU Execution Error: Wrong command. Other 9 Contact Mindray Customer Service Department or your local distributor.
parameter error.
6-5
Troubleshooting
100960300 MU Execution Error: MU handshaking with sub-unit. 9 Contact Mindray Customer Service Department or your local distributor.
Can't respond to other instructions.
100960500 MU Execution Error: MU calling sub-unit to reset and 9 Contact Mindray Customer Service Department or your local distributor.
self-test. Can't respond to other instructions.
100960900 MU Execution Error: MU work status not supporting 9 Contact Mindray Customer Service Department or your local distributor.
this command. Self-test.
100960901 MU Execution Error: MU work status not supporting 9 Contact Mindray Customer Service Department or your local distributor.
this command. Failed.
100960902 MU Execution Error: MU work status not supporting 9 Contact Mindray Customer Service Department or your local distributor.
this command. Handshake.
100960903 MU Execution Error: MU work status not supporting 9 Contact Mindray Customer Service Department or your local distributor.
this command. Reset.
100960904 MU Execution Error: MU work status not supporting 9 Contact Mindray Customer Service Department or your local distributor.
this command. Shutdown.
100960905 MU Execution Error: MU work status not supporting 9 Contact Mindray Customer Service Department or your local distributor.
this command. Free.
100960906 MU Execution Error: MU work status not supporting 9 Contact Mindray Customer Service Department or your local distributor.
this command. Mechanical reset.
100961000 MU Execution Error: MU busy. Can't respond to current 9 Contact Mindray Customer Service Department or your local distributor.
instruction.
100961100 MU Execution Error: Disassemble instruction wrong. 9 Contact Mindray Customer Service Department or your local distributor.
100961200 MU Execution Error: Attempted undefined system 9 Contact Mindray Customer Service Department or your local distributor.
operation.
6-6
Troubleshooting
100961500 MU Execution Error: Wrong parameter for inquiry. 0 Contact Mindray Customer Service Department or your local distributor.
100961600 MU Execution Error: Attempted to configure undefined 0 Contact Mindray Customer Service Department or your local distributor.
parameter.
100961800 MU Execution Error: Attempted to configure undefined 0 Contact Mindray Customer Service Department or your local distributor.
procedure.
100961900 MU Execution Error: Wrong parameter for procedure 0 Contact Mindray Customer Service Department or your local distributor.
configuration.
100962000 MU Execution Error: E2PROM R/W wrong. Read error. 10 Contact Mindray Customer Service Department or your local distributor.
100962001 MU Execution Error: E2PROM R/W wrong. Write error. 10 Contact Mindray Customer Service Department or your local distributor.
100962002 MU Execution Error: E2PROM R/W wrong. Check sum 10 Contact Mindray Customer Service Department or your local distributor.
error.
100962003 MU Execution Error: E2PROM R/W wrong. Download 10 Contact Mindray Customer Service Department or your local distributor.
error.
100969000 MU Execution Error: Execution result not received 9 Contact Mindray Customer Service Department or your local distributor.
within specified time.
100969100 MU Execution Error: Command response wrong or no 9 Contact Mindray Customer Service Department or your local distributor.
response
101280100 Sample probe Unit Error: Wrong command. Illegal 31 Contact Mindray Customer Service Department or your local distributor.
frame type.
101280101 Sample probe Unit Error: Wrong command. Not this 31 Contact Mindray Customer Service Department or your local distributor.
unit's instruction.
101280102 Sample probe Unit Error: Wrong command. Illegal 31 Contact Mindray Customer Service Department or your local distributor.
command identity.
6-7
Troubleshooting
101280104 Sample probe Unit Error: Wrong command. Parameter 31 Contact Mindray Customer Service Department or your local distributor.
significant bits wrong
101280105 Sample probe Unit Error: Wrong command. Other 31 Contact Mindray Customer Service Department or your local distributor.
parameter wrong.
101280200 Sample probe Unit Error: Control unit self-test wrong 10 Contact Mindray Customer Service Department or your local distributor.
101280300 Sample probe Unit Error: Mechanical device reset 2 Contact Mindray Customer Service Department or your local distributor.
wrong.
101280400 Sample probe Unit Error: Control unit not supporting 2 Contact Mindray Customer Service Department or your local distributor.
this command. Self-test.
101280401 Sample probe Unit Error: Control unit not supporting 4 Contact Mindray Customer Service Department or your local distributor.
this command. Malfunction.
101280402 Sample probe Unit Error: Control unit not supporting 2 Contact Mindray Customer Service Department or your local distributor.
this command. Wait for handshake.
101280403 Sample probe Unit Error: Control unit not supporting 2 Contact Mindray Customer Service Department or your local distributor.
this command. Shutdown.
101280404 Sample probe Unit Error: Control unit not supporting 2 Contact Mindray Customer Service Department or your local distributor.
this command. Processing command.
101280405 Sample probe Unit Error: Control unit not supporting 2 Contact Mindray Customer Service Department or your local distributor.
this command. Free.
101280406 Sample probe Unit Error: Control unit not supporting 2 Contact Mindray Customer Service Department or your local distributor.
this command. Mechanical reset.
101280500 Sample probe Unit Error: Control unit busy. Can't 2 Contact Mindray Customer Service Department or your local distributor.
respond to current instruction.
6-8
Troubleshooting
101280700 Sample probe Unit Error: Speed configuration 0 Contact Mindray Customer Service Department or your local distributor.
parameter beyond limit or unreasonable.
101280800 Sample probe Unit Error: Attempted to configure 0 Contact Mindray Customer Service Department or your local distributor.
undefined parameter.
101280900 Sample probe Unit Error: Parameter value beyond limit 0 Contact Mindray Customer Service Department or your local distributor.
or unreasonable.
101281000 Sample probe Unit Error: Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.
101281100 Sample probe Unit Error: Inquiry parameter beyond 0 Contact Mindray Customer Service Department or your local distributor.
limit or unreasonable.
101281200 Sample probe Unit Error: Attempted undefined system 0 Contact Mindray Customer Service Department or your local distributor.
instruction.
101281300 Sample probe Unit Error: System instruction parameter 0 Contact Mindray Customer Service Department or your local distributor.
beyond limit or unreasonable.
101281400 Sample probe Unit Error: Attempted to commission 0 Contact Mindray Customer Service Department or your local distributor.
undefined position.
101281500 Sample probe Unit Error: Commissioning parameter 0 Contact Mindray Customer Service Department or your local distributor.
beyond limit or unreasonable.
101281600 Sample probe Unit Error: System not commissioned or 0 Contact Mindray Customer Service Department or your local distributor.
confirmation not consistent with instruction.
6-9
Troubleshooting
101281800 Sample probe Unit Error: Sample disk rotation failed. 4 Ensure the sample disk rotates well as instructed by 4.4.6.6 Sam. Disk
Can't reach initial point. tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101281801 Sample probe Unit Error: Sample disk rotation failed. 4 Ensure the sample disk rotates well as instructed by 4.4.6.6 Sam. Disk
Can't leave initial point. tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101281802 Sample probe Unit Error: Sample disk rotation failed. 4 Ensure the sample disk rotates well as instructed by 4.4.6.6 Sam. Disk
Can't detect weight disk signal. tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101281803 Sample probe Unit Error: Sample disk rotation failed. 4 Turn off the analyzing unit and check if the sample tubes are placed too
Sample probe left in Sample disk. high on the sample disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-10
Troubleshooting
101281902 Sample probe Unit Error: Syringe A/D error. Can't 4 Check the sample syringe as instructed by 4.4.6.7 Fluid Sys. tab.
reach initial point.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101281903 Sample probe Unit Error: Syringe A/D error. Can't leave 4 Check the sample syringe as instructed by 4.4.6.7 Fluid Sys. tab.
initial point.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282000 Sample probe Unit Error: Sample probe rotation error. 4 (1) Turn off the analyzing unit and clean the sample disk rotor.
Can't reach initial point.
(2) Turn on the analyzing unit and check the sample probe as instructed
by 4.4.6.1 Sam. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282001 Sample probe Unit Error: Sample probe rotation error. 4 (1) Turn off the analyzing unit and clean the sample disk rotor.
Not at v-position for rotation.
(2) Turn on the analyzing unit and check the sample probe as instructed
by 4.4.6.1 Sam. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282002 Sample probe Unit Error: Sample probe v-move wrong. 4 (1) Turn off the analyzing unit. Clean the sample disk rotor.
Can't reach initial (Sample disk/wash) point.
(2) Turn on the analyzing unit and check the sample probe as instructed
by 4.4.6.1 Sam. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-11
Troubleshooting
101282101 Sample probe Unit Error: Sample probe v-moved 4 Check the sample probe as instructed by 4.4.6.1 Sam. Probe tab.
wrong. Can't leave initial (Sample disk/wash) point.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282102 Sample probe Unit Error: Sample probe v-move wrong. 4 Check the sample probe as instructed by 4.4.6.1 Sam. Probe tab.
Not at initial point (Sample disk/wash) for d-move start.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282107 Sample probe Unit Error: Sample probe v-move wrong. 31 Turn off the analyzing unit and clear the obstacles from the path of the
Collision. vertical movement of the sample probe.
101282108 Sample probe Unit Error: Sample probe v-move wrong. 9 Turn off the analyzing unit and pinch the sample probe and push it upward
Collide with transducer holder. to check whether the spring can move freely.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282110 Sample probe Unit Error: Sample probe v-move wrong. 4 Check the sample probe as instructed by 4.4.6.1 Sam. Probe tab.
Current position prohibits v-move (Sample disk/wash).
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282200 Sample probe Unit Error: Attempted undefined simple 0 Other errors have higher priority.
operation.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101282300 Sample probe Unit Error: Wrong cuvette position 2 Other errors have higher priority.
parameter in command.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-12
Troubleshooting
101282600 Sample probe Unit Error: Syringe not initialized. 4 Contact Mindray Customer Service Department or your local distributor.
101289000 Sample probe Unit Error: No execution result received 0 Contact Mindray Customer Service Department or your local distributor.
within specified time.
101289100 Sample probe Unit Error: Command response wrong or 0 Contact Mindray Customer Service Department or your local distributor.
no response.
101440100 Reagent probe Unit Error: Wrong command. Illegal 31 Contact Mindray Customer Service Department or your local distributor.
frame type.
101440101 Reagent probe Unit Error: Wrong command. Not this 31 Contact Mindray Customer Service Department or your local distributor.
unit's instruction.
101440102 Reagent probe Unit Error: Wrong command. Illegal 31 Contact Mindray Customer Service Department or your local distributor.
instruction identity.
6-13
Troubleshooting
101440104 Reagent probe Unit Error: Wrong command. 31 Contact Mindray Customer Service Department or your local distributor.
Parameter significant bits wrong.
101440105 Reagent probe Unit Error: Wrong command. Other 31 Contact Mindray Customer Service Department or your local distributor.
parameter wrong.
101440200 Reagent probe Unit Error: Control unit self-test wrong. 10 Contact Mindray Customer Service Department or your local distributor.
101440300 Reagent probe Unit Error: Mechanical device reset 0 Contact Mindray Customer Service Department or your local distributor.
wrong
101440400 Reagent probe Unit Error: Control unit not supporting 3 Contact Mindray Customer Service Department or your local distributor.
this command. Self-test.
101440401 Reagent probe Unit Error: Control unit not supporting 5 Contact Mindray Customer Service Department or your local distributor.
this command. Malfunction.
101440402 Reagent probe Unit Error: Control unit not supporting 3 Contact Mindray Customer Service Department or your local distributor.
this command. Wait for handshake.
101440403 Reagent probe Unit Error: Control unit not supporting 3 Contact Mindray Customer Service Department or your local distributor.
this command. Shutdown.
101440404 Reagent probe Unit Error: Control unit not supporting 3 Contact Mindray Customer Service Department or your local distributor.
this command. Processing instruction.
101440405 Reagent probe Unit Error: Control unit not supporting 3 Contact Mindray Customer Service Department or your local distributor.
this command. Busy.
101440406 Reagent probe Unit Error: Control unit not supporting 3 Contact Mindray Customer Service Department or your local distributor.
this command. Mechanical reset.
101440500 Reagent probe Unit Error: Control unit busy. Can't 3 Contact Mindray Customer Service Department or your local distributor.
respond to current instruction.
6-14
Troubleshooting
101440700 Reagent probe Unit Error: Speed configuration 0 Contact Mindray Customer Service Department or your local distributor.
parameter beyond limit or unreasonable.
101440800 Reagent probe Unit Error: Attempted to configure 0 Contact Mindray Customer Service Department or your local distributor.
undefined parameter.
101440900 Reagent probe Unit Error: Parameter configuration 0 Contact Mindray Customer Service Department or your local distributor.
beyond limit or unreasonable.
101441000 Reagent probe Unit Error: Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.
101441100 Reagent probe Unit Error: Inquiry parameter beyond 0 Contact Mindray Customer Service Department or your local distributor.
limit or unreasonable.
101441200 Reagent probe Unit Error: Attempted undefined system 0 Contact Mindray Customer Service Department or your local distributor.
instruction.
101441300 Reagent probe Unit Error: System instruction 0 Contact Mindray Customer Service Department or your local distributor.
parameter beyond limit or unreasonable.
101441400 Reagent probe Unit Error: Attempted to commission 0 Contact Mindray Customer Service Department or your local distributor.
undefined position.
101441500 Reagent probe Unit Error: Commissioning parameter 0 Contact Mindray Customer Service Department or your local distributor.
beyond limit or unreasonable.
101441600 Reagent probe Unit Error: Not commissioned or 0 Contact Mindray Customer Service Department or your local distributor.
confirmation not consistent with instruction.
6-15
Troubleshooting
101441800 Reagent probe Unit Error: Reagent disk rotation failed. 5 Check the reagent disk as instructed by 4.4.6.5 Reag. Disk tab.
Can't reach initial position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101441801 Reagent probe Unit Error: Reagent disk rotation failed. 5 Check the reagent disk as instructed by 4.4.6.5 Reag. Disk tab.
Can't leave initial position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101441802 Reagent probe Unit Error: Reagent disk rotation failed. 5 Check the reagent disk as instructed by 4.4.6.5 Reag. Disk tab.
Can't detect weight disk signal.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101441803 Reagent probe Unit Error: Reagent disk rotation failed. 5 Turn off the analyzing unit and check the types of the reagent bottles.
Reagent probe left in Reagent disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-16
Troubleshooting
101441902 Reagent probe Unit Error: Syringe A/D error. Can't 5 Check the reagent syringe as instructed by 4.4.6.7 Fluid Sys. tab.
reach initial position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101441903 Reagent probe Unit Error: Syringe A/D error. Can't 5 Check the reagent syringe as instructed by 4.4.6.7 Fluid Sys. tab.
leave initial position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101442000 Reagent probe Unit Error: Reagent probe rotation 5 (1) Turn off the analyzing unit. Clean the sample disk rotor.
error. Can't reach initial position.
(2) Turn on the analyzing unit and check the reagent probe as instructed
by 4.4.6.2 Reag. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101442001 Reagent probe Unit Error: Reagent probe rotation 5 (1) Turn off the analyzing unit. Clean the sample disk rotor.
error. Can't leave initial position.
(2) Turn on the analyzing unit and check the reagent probe as instructed
by 4.4.6.2 Reag. Probe tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-17
Troubleshooting
101442100 Reagent probe Unit Error: Reagent probe v-move error. 5 Check the reagent probe as instructed by 4.4.6.2 Reag. Probe tab.
Can't reach initial (Reagent disk/wash) position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101442101 Reagent probe Unit Error: Reagent probe v-move error. 5 Check the reagent probe as instructed by 4.4.6.2 Reag. Probe tab.
Can't leave initial (Reagent disk/wash) position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101442102 Reagent probe Unit Error: Reagent probe v-move error. 5 Check the reagent probe as instructed by 4.4.6.2 Reag. Probe tab.
Not at v-initial (Reagent disk/wash) position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101442107 Reagent probe Unit Error: Reagent probe v-move error. 31 Turn off the analyzing unit and clear the obstacles in the path of the
Collision. vertical movement of the reagent probe.
101442108 Reagent probe Unit Error: Reagent probe v-move error. 9 Turn off the analyzing unit and pinch the reagent probe and push it
Collision transducer hold. upward to see whether it can move freely.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101442110 Reagent probe Unit Error: Reagent probe v-move error. 5 Check the reagent probe as instructed by 4.4.6.2 Reag. Probe tab.
Current position prohibits v-move (Reagent disk/wash).
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-18
Troubleshooting
101442300 Reagent probe Unit Error: Wrong cuvette position 0 Other errors have higher priority.
parameter in command.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101442400 Reagent probe Unit Error: Illegal surface detection. 31 Turn off the analyzing unit and
(1) Check the types of the reagent bottles.
(2) Check the reagent syringe for leakage as instructed by 5.2.1 Checking
Sample/Reagent Syringes.
(3) Check the sample probe to see if there are drops hanging on the probe
tip. If there are, contact Mindray Customer Service Department or your
local distributor.
(4) Check whether the sample probe is bent or scratched. Replace the
damaged probe as instructed by 5.6.4 Replacing Reagent Probe if
necessary.
(5) Clean the reagent probe as instructed by 5.3.2 Washing Reagent
Probe.
If this error occurs frequently, contact Mindray Customer Service
Department or your local distributor.
101442600 Reagent probe Unit Error: Syringe not initialized 5 Contact Mindray Customer Service Department or your local distributor.
(unknown syringe current position)
101449000 Reagent probe Unit Error: Execution result not 0 Contact Mindray Customer Service Department or your local distributor.
received within specified time.
6-19
Troubleshooting
101600100 Loading/mixing Unit Error: Wrong command. Illegal 31 Contact Mindray Customer Service Department or your local distributor.
frame type
101600101 Loading/mixing Unit Error: Wrong command. Not this 31 Contact Mindray Customer Service Department or your local distributor.
unit's instruction
101600102 Loading/mixing Unit Error: Wrong command. Illegal 31 Contact Mindray Customer Service Department or your local distributor.
command identity.
101600103 Loading/mixing Unit Error: Wrong command. 31 Contact Mindray Customer Service Department or your local distributor.
Instruction length wrong
101600104 Loading/mixing Unit Error: Wrong command. 31 Contact Mindray Customer Service Department or your local distributor.
Parameter significant bits wrong
101600105 Loading/mixing Unit Error: Wrong command. Other 31 Contact Mindray Customer Service Department or your local distributor.
parameter wrong
101600200 Loading/mixing Unit Error: Control unit self-test wrong 10 Contact Mindray Customer Service Department or your local distributor.
101600300 Loading/mixing Unit Error: Mechanical device reset 0 Contact Mindray Customer Service Department or your local distributor.
wrong
101600400 Loading/mixing Unit Error: Control unit not supporting 6 Contact Mindray Customer Service Department or your local distributor.
this instruction. Self-test.
101600401 Loading/mixing Unit Error: Control unit not supporting 6 Contact Mindray Customer Service Department or your local distributor.
this instruction. Malfunction
101600402 Loading/mixing Unit Error: Control unit not supporting 6 Contact Mindray Customer Service Department or your local distributor.
this instruction. Wait for handshake.
101600403 Loading/mixing Unit Error: Control unit not supporting 6 Contact Mindray Customer Service Department or your local distributor.
this instruction. Shutdown.
6-20
Troubleshooting
101600405 Loading/mixing Unit Error: Control unit not supporting 6 Contact Mindray Customer Service Department or your local distributor.
this instruction. Free.
101600406 Loading/mixing Unit Error: Control unit not supporting 6 Contact Mindray Customer Service Department or your local distributor.
this instruction. Mechanical reset.
101600500 Loading/mixing Unit Error: Control unit busy. Can't 6 Contact Mindray Customer Service Department or your local distributor.
respond to current instruction.
101600600 Loading/mixing Unit Error: Attempted to configure 0 Contact Mindray Customer Service Department or your local distributor.
undefined speed.
101600700 Loading/mixing Unit Error: Speed configuration 0 Contact Mindray Customer Service Department or your local distributor.
parameter beyond limit or unreasonable.
101600800 Loading/mixing Unit Error: Attempted to configure 0 Contact Mindray Customer Service Department or your local distributor.
undefined parameter.
101600900 Loading/mixing Unit Error: Parameter value beyond 0 Contact Mindray Customer Service Department or your local distributor.
limit or unreasonable.
101601000 Loading/mixing Unit Error: Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.
101601100 Loading/mixing Unit Error: Inquiry parameter beyond 0 Contact Mindray Customer Service Department or your local distributor.
limit or unreasonable.
101601200 Loading/mixing Unit Error: Attempted undefined system 0 Contact Mindray Customer Service Department or your local distributor.
instruction.
101601300 Loading/mixing Unit Error: System instruction 0 Contact Mindray Customer Service Department or your local distributor.
parameter beyond limit or unreasonable.
101601400 Loading/mixing Unit Error: Attempted to commission 0 Contact Mindray Customer Service Department or your local distributor.
undefined position.
6-21
Troubleshooting
101601600 Loading/mixing Unit Error: System not commissioned 0 Contact Mindray Customer Service Department or your local distributor.
or confirmation not consistent with instruction.
101601700 Loading/mixing Unit Error: Mixing bar v-move error. 7 (1) Turn off the analyzing unit. Clean the mixing bar rotor.
Can't reach initial (wash) position.
(2) Turn on the analyzing unit and check the mixing bar as instructed by
4.4.6.3 Mix. Bar tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101601701 Loading/mixing Unit Error: Mixing bar v-move error. 7 (1) Turn off the analyzing unit. Clean the mixing bar rotor.
Can't leave initial (wash) position.
(2) Turn on the analyzing unit and check the mixing bar as instructed by
4.4.6.3 Mix. Bar tab .
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101601702 Loading/mixing Unit Error: Mixing bar v-move error. Not 7 (1) Turn off the analyzing unit. Clean the mixing bar rotor.
at v-initial position (wash) for d-move.
(2) Turn on the analyzing unit and check the mixing bar as instructed by
4.4.6.3 Mix. Bar tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101601707 Loading/mixing Unit Error: Mixing bar v-move error. 7 Other errors have higher priorities.
Current position prohibits v-move (wash).
Check the mixing bar as instructed by 4.4.6.3 Mix. Bar tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-22
Troubleshooting
101601802 Loading/mixing Unit Error: Mixing bar h-move error. Not 7 (1) Turn off the analyzing unit. Clean the mixing bar rotor.
at h-initial position for h-rotation.
(2) Turn on the analyzing unit and check the mixing bar as instructed by
4.4.6.3 Mix. Bar tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101602000 Loading/mixing Unit Error: Robotic arm v-move error. 9 Turn on the analyzing unit. Check the manipulator as instructed by
Can't reach initial (other) position 4.4.6.10 Manipulator tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101602001 Loading/mixing Unit Error: Robotic arm v-move error. 9 Turn on the analyzing unit and check the manipulator as instructed by
Can't leave initial (other) position. 4.4.6.10 Manipulator tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101602100 Loading/mixing Unit Error: Robotic arm h-move error. 6 Turn on the analyzing unit and check the manipulator as instructed by
Can't reach initial position. 4.4.6.10 Manipulator tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-23
Troubleshooting
101602200 Loading/mixing Unit Error: Robotic arm hold error. 6 Contact Mindray Customer Service Department or your local distributor.
Upper hand can't close when taking cuvette.
101602201 Loading/mixing Unit Error: Robotic arm hold error. 6 Contact Mindray Customer Service Department or your local distributor.
Upper hand can't open when placing cuvette.
101602202 Loading/mixing Unit Error: Robotic arm hold error. 6 Contact Mindray Customer Service Department or your local distributor.
Lower hand can't close when taking cuvette.
101602203 Loading/mixing Unit Error: Robotic arm hold error. 6 Contact Mindray Customer Service Department or your local distributor.
Lower hand can't open when placing cuvette.
101602204 Loading/mixing Unit Error: Robotic arm hold error. 6 Contact Mindray Customer Service Department or your local distributor.
Upper hand accidentally open during movement.
101602205 Loading/mixing Unit Error: Robotic arm hold error. 6 Contact Mindray Customer Service Department or your local distributor.
Upper hand accidentally closed during movement.
101602206 Loading/mixing Unit Error: Robotic arm hold error. 6 Contact Mindray Customer Service Department or your local distributor.
Lower hand accidentally opened during movement.
101602207 Loading/mixing Unit Error: Robotic arm hold error. 6 Contact Mindray Customer Service Department or your local distributor.
Lower hand accidentally closed during movement.
101602208 Loading/mixing Unit Error: Robotic arm hold error. No 0 Check the remaining cuvettes and add more cuvettes as needed.
cuvette in compartment.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-24
Troubleshooting
101602210 Loading/mixing Unit Error: Cuvette compartment trolley 0 If there is no cuvette left in the cuvette compartment, add more cuvettes to
moving. the compartment as instructed by 1.1.1.6 Reaction Cuvette Loader and
101609000 Loading/mixing Unit Error: Execution result not 5 Contact Mindray Customer Service Department or your local distributor.
received within specified time.
101609100 Loading/mixing Unit Error: Command response wrong 5 Contact Mindray Customer Service Department or your local distributor.
or no response.
101760100 Temperature Unit Error: Wrong command. Illegal frame 31 Contact Mindray Customer Service Department or your local distributor.
type
101760101 Temperature Unit Error: Wrong command. Not this 31 Contact Mindray Customer Service Department or your local distributor.
Unit's instruction
101760102 Temperature Unit Error: Wrong command. Illegal 31 Contact Mindray Customer Service Department or your local distributor.
command identity.
6-25
Troubleshooting
101760104 Temperature Unit Error: Wrong command. Parameter 31 Contact Mindray Customer Service Department or your local distributor.
significant bits wrong
101760105 Temperature Unit Error: Wrong command. Other 31 Contact Mindray Customer Service Department or your local distributor.
parameter wrong
101760200 Temperature Unit Error: Control unit self-test wrong 10 Contact Mindray Customer Service Department or your local distributor.
101760400 Temperature Unit Error: Control unit not supporting this 0 Contact Mindray Customer Service Department or your local distributor.
instruction. Self-test.
101760401 Temperature Unit Error: Control unit not supporting this 0 Contact Mindray Customer Service Department or your local distributor.
instruction. Malfunction
101760402 Temperature Unit Error: Control unit not supporting this 0 Contact Mindray Customer Service Department or your local distributor.
instruction. Wait for handshake.
101760403 Temperature Unit Error: Control unit not supporting this 0 Contact Mindray Customer Service Department or your local distributor.
instruction. Shutdown.
101760404 Temperature Unit Error: Control unit not supporting this 0 Contact Mindray Customer Service Department or your local distributor.
instruction. Processing instruction.
101760405 Temperature Unit Error: Control unit not supporting this 0 Contact Mindray Customer Service Department or your local distributor.
instruction. Free.
101760406 Temperature Unit Error: Control unit not supporting this 0 Contact Mindray Customer Service Department or your local distributor.
instruction. Mechanical reset.
101760500 Temperature Unit Error: Control unit busy. Can't 0 Contact Mindray Customer Service Department or your local distributor.
respond to current instruction.
101760600 Temperature Unit Error: Attempted undefined inquiry. 0 Contact Mindray Customer Service Department or your local distributor.
6-26
Troubleshooting
101761100 Temperature Unit Error: Attempted undefined 0 Contact Mindray Customer Service Department or your local distributor.
temperature inquiry.
101761200 Temperature Unit Error: Attempted target temperature 0 Contact Mindray Customer Service Department or your local distributor.
beyond limit or unreasonable.
101761300 Temperature Unit Error: Attempted undefined system 0 Contact Mindray Customer Service Department or your local distributor.
instruction.
101769000 Temperature Unit Error: Execution result not received 0 Contact Mindray Customer Service Department or your local distributor.
within specified time.
101769100 Temperature Unit Error: Command response wrong or 0 Contact Mindray Customer Service Department or your local distributor.
no response.
101920100 Reaction disk and Photoelectrical Error: Wrong 31 Contact Mindray Customer Service Department or your local distributor.
command. Illegal frame type.
101920101 Reaction disk and Photoelectrical Error: Wrong 31 Contact Mindray Customer Service Department or your local distributor.
command. Not this unit's instruction.
101920102 Reaction disk and Photoelectrical Error: Wrong 31 Contact Mindray Customer Service Department or your local distributor.
command. Illegal command identity.
101920103 Reaction disk and Photoelectrical Error: Wrong 31 Contact Mindray Customer Service Department or your local distributor.
command. Instruction length wrong.
101920104 Reaction disk and Photoelectrical Error: Wrong 31 Contact Mindray Customer Service Department or your local distributor.
command. Parameter significant bits wrong.
101920105 Reaction disk and Photoelectrical Error: Wrong 31 Contact Mindray Customer Service Department or your local distributor.
command. Other parameter wrong.
6-27
Troubleshooting
101920300 Reaction disk and Photoelectrical Error: Mechanical 0 Contact Mindray Customer Service Department or your local distributor.
device reset wrong.
101920400 Reaction disk and Photoelectrical Error: Control unit 8 Contact Mindray Customer Service Department or your local distributor.
not supporting this instruction. Self-test.
101920401 Reaction disk and Photoelectrical Error: Control unit 9 Contact Mindray Customer Service Department or your local distributor.
not supporting this instruction. Malfunction.
101920402 Reaction disk and Photoelectrical Error: Control unit 8 Contact Mindray Customer Service Department or your local distributor.
not supporting this instruction. Wait for handshake.
101920403 Reaction disk and Photoelectrical Error: Control unit 8 Contact Mindray Customer Service Department or your local distributor.
not supporting this instruction. Shutdown.
101920500 Reaction disk and Photoelectrical Error: Control unit 8 Contact Mindray Customer Service Department or your local distributor.
busy. Can't respond to current instruction.
101920600 Reaction disk and Photoelectrical Error: Attempted 0 Contact Mindray Customer Service Department or your local distributor.
undefined speed.
101920700 Reaction disk and Photoelectrical Error: Speed 0 Contact Mindray Customer Service Department or your local distributor.
configuration beyond limit or unreasonable.
101920800 Reaction disk and Photoelectrical Error: Attempted to 0 Contact Mindray Customer Service Department or your local distributor.
configure undefined parameter.
101920900 Reaction disk and Photoelectrical Error: Parameter 0 Contact Mindray Customer Service Department or your local distributor.
configuration beyond limit or unreasonable.
101921000 Reaction disk and Photoelectrical Error: Attempted 0 Contact Mindray Customer Service Department or your local distributor.
undefined inquiry.
6-28
Troubleshooting
101921200 Reaction disk and Photoelectrical Error: Attempted 0 Contact Mindray Customer Service Department or your local distributor.
undefined system instruction.
101921300 Reaction disk and Photoelectrical Error: System 0 Contact Mindray Customer Service Department or your local distributor.
instruction parameter beyond limit or unreasonable.
101921400 Reaction disk and Photoelectrical Error: Re-disk 8 Check the reaction disk as instructed by 4.4.6.4 Reac. Disk tab.
rotation error. Can't reach initial position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101921401 Reaction disk and Photoelectrical Error: Re-disk 8 Check the reaction disk as instructed by 4.4.6.4 Reac. Disk tab.
rotation error. Can't leave initial position.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101921402 Reaction disk and Photoelectrical Error: Re-disk 8 Check the reaction disk as instructed by 4.4.6.4 Reac. Disk tab.
rotation error. Can't detect weight disk signal.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101921500 Reaction disk and Photoelectrical Error: Photoelectrical 9 Other errors have higher priorities.
data overflow.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
101929000 Reaction disk and Photoelectrical Error: Execution 0 Contact Mindray Customer Service Department or your local distributor.
result not received within specified time.
101929100 Reaction disk and Photoelectrical Error: Command 0 Contact Mindray Customer Service Department or your local distributor.
response wrong or no response
6-29
Troubleshooting
6-30
Troubleshooting
111442102 Reagent probe Unit Error: Reagent probe v-move error. 9 Check the reagent probe as instructed by 4.4.6.2 Reag. Probe tab.
Not at v-initial position for d-move. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-31
Troubleshooting
111442110 Reagent probe Unit Error: Reagent probe v-move error. 9 Check the reagent probe as instructed by 4.4.6.2 Reag. Probe tab.
Current position prohibits v-move. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-32
Troubleshooting
111601700 Loading/mixing Unit Error: Mixing bar v-move error. 9 (1). Turn off the analyzing unit. Clean the mixing bar rotor
Can't reach initial position. (Re-disk)
(2) Turn on the analyzing unit. Check the mixing bar as instructed by
4.4.6.10 Manipulator tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
111601701 Loading/mixing Unit Error: Mixing bar v-move error. 9 (1) Turn off the analyzing unit. Clean the mixing bar rotor
Can't leave initial position. (Re-disk)
(2) Turn on the analyzing unit. Check the mixing bar as instructed by
4.4.6.10 Manipulator tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-33
Troubleshooting
111601707 Loading/mixing Unit Error: Mixing bar v-move error. 9 Other errors have higher priorities. Check the movement of the mixing bar
Current position prohibits v-move. (Re-disk) at the maintenance screen.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
111602000 Loading/mixing Unit Error: Robotic arm v-move error. 9 Turn on the analyzing unit and check the manipulator as instructed by
Can't reach initial position. (Re-disk) 4.4.6.10 Manipulator tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
111602001 Loading/mixing Unit Error: Robotic arm v-move error. 9 Turn on the analyzing unit and check the manipulator as instructed by
Can't leave initial position. (Re-disk) 4.4.6.10 Manipulator tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
121442100 Reagent probe Unit Error: Reagent probe v-move error. 5 Check the reagent probe as instructed by 4.4.6.2 Reag. Probe tab.
Can't reach initial position. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
121442101 Reagent probe Unit Error: Reagent probe v-move error. 5 Check the reagent probe as instructed by 4.4.6.2 Reag. Probe tab.
Can't leave initial position. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-34
Troubleshooting
121442110 Reagent probe Unit Error: Reagent probe v-move error. 5 Check the reagent probe as instructed by 4.4.6.2 Reag. Probe tab.
Current position prohibits v-move. (Reagent disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
131282100 Sample probe Unit Error: Sample probe v-move error. 4 Check the sample probe as instructed by 4.4.6.1 Sam. Probe tab.
Can't reach initial position. (Sample disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
131282101 Sample probe Unit Error: Sample probe v-move error. 4 Check the sample probe as instructed by 4.4.6.1 Sam. Probe tab.
Can't leave initial position. (Sample disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
131282102 Sample probe Unit Error: Sample probe v-move error. 4 Check the sample probe as instructed by 4.4.6.1 Sam. Probe tab.
Not at v-initial position for d-move. (Sample disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
131282110 Sample probe Unit Error: Sample probe v-move error 4 Check the sample probe as instructed by 4.4.6.1 Sam. Probe tab.
Current position prohibits v-move. (Sample disk)
If the error remains, contact Mindray Customer Service Department or
your local distributor.
200000001 Temperature Unit Warn: Re-disk actual temperature 0 Turn off the analyzing unit and check whether the ambient temperature
exceeds top limit. falls into the specified range.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-35
Troubleshooting
200000003 Temperature Unit Warn: Re-disk did not reach 0 Check whether the ambient temperature falls into the specified range.
expected temperature within specified time.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
200000004 Temperature Unit Warn: Reagent preheat didn't reach 0 Check whether the ambient temperature falls into the specified range.
set temperature within specified time.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
200000005 Temperature Unit Warn: Re-disk temperature fluctuates 0 Check whether the ambient temperature falls into the specified range.
over given range under normal operation
If the error remains, contact Mindray Customer Service Department or
your local distributor.
200000006 Temperature Unit Warn: Reagent preheat temperature 0 Check whether the ambient temperature falls into the specified range.
fluctuate over given range under normal operation.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
200000007 Temperature Unit Warn: Re-disk temperature control 0 Turn off the analyzing unit and check whether the ambient temperature
shutdown for overheating. falls into the specified range.
Contact Mindray Customer Service Department or your local distributor.
200000008 Temperature Unit Warn: Preheat temperature 0 Turn off the analyzing unit and check whether the ambient temperature
shutdown for overheating. falls into the specified range.
Contact Mindray Customer Service Department or your local distributor.
201280005 Sample probe Unit Warn: Sample probe syringe full. 0 If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-36
Troubleshooting
201280007 Sample probe Unit Warn: Sample probe didn't detect 20 Check the sample volumes in the sample tubes.
fluid surface in Sample disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201280009 Sample probe Unit Warn: Sample probe didn't detect 2 Reagent unit errors have higher priority.
fluid surface in Re-disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201280010 Sample probe Unit Warn: Insufficient dispensing of 0 Check the dispensing ability of the sample syringe as instructed by 4.4.6.7
Sample probe. Fluid Sys. tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201440005 Reagent probe Unit Warn: Reagent probe syringe full. 0 If the error remains, contact Mindray Customer Service Department or
your local distributor.
201440006 Reagent probe Unit Warn: Reagent probe syringe 0 If the error remains, contact Mindray Customer Service Department or
empty. your local distributor.
201440007 Reagent probe Unit Warn: Reagent probe didn't detect 30 Check the reagent volumes in the reagent bottles.
fluid surface in Reagent disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-37
Troubleshooting
201440010 Reagent probe Unit Warn: Insufficient dispensing of 0 Check the dispensing ability of the reagent syringe as instructed by 4.4.6.7
Reagent probe. Fluid Sys. tab.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201600009 Loading/mixing Unit Warn: Robotic arm can't release 6 Check the manipulator as instructed by 4.4.6.10 Manipulator tab.
held Re-cuvette. Unloading failed.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201600010 Loading/mixing Unit Warn: Robotic arm can't grab used 6 Check the manipulator as instructed by 4.4.6.10 Manipulator tab.
Re-cuvette from Re-disk.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201600011 Loading/mixing Unit Warn: Re-cuvette dropped from 0 Check the manipulator as instructed by 4.4.6.10 Manipulator tab.
robotic arm.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-38
Troubleshooting
201600016 Loading/mixing Unit Warn: Collision occurred when 6 Check the manipulator as instructed by 4.4.6.10 Manipulator tab.
unloading used cuvette. Operation cancelled.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201600017 Loading/mixing Unit Warn: Collision occurred when 6 Check the manipulator as instructed by 4.4.6.10 Manipulator tab.
loading new cuvette. Operation cancelled.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
201600018 Loading/mixing Unit Warn: Used cuvette compartment 0 Turn off the analyzing unit and empty the used-cuvette bucket as
full. Storing segments may cause problem. instructed by 5.2.9 Emptying Waste Tank.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-39
Troubleshooting
400000001 Operation unit: System initialization error. Can't 10 (1) Turn off the POWER and check the connection between the analyzing
connect main unit. unit and the operation unit.
(2) Use the emergency exit function to exit the control system.
If this error remains, contact Mindray Customer Service Department or
your local distributor.
400000002 Operation unit: System initialization error. Main unit 10 Use the emergency exit function to exit the control system and then restart
self-test abnormal. it.
If this error remains, contact Mindray Customer Service Department or
your local distributor.
400000010 Operation unit: System initialization error. Serial port 10 Use the emergency exit function to exit the control system and then restart
initialization error. it.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.
400000012 Operation unit: System initialization error. Can't 0 (1) Ensure the printer is well connected to the printer.
connect printer.
(2) Use the emergency exit function to exit the control system and then
restart it.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.
6-40
Troubleshooting
400000100 Operation unit: Power-on pretreatment error. Lamp of 1 Other errors have higher priority.
insufficient illumination. Replacement needed.
If the error remains, change the lamp or contact Mindray Customer
Service Department or your local distributor.
400000204 Operation unit: Updating database failed. 0 Contact Mindray Customer Service Department or your local distributor.
400000209 Operation unit: Can't open help file. 0 Contact Mindray Customer Service Department or your local distributor.
400000216 Operation unit: Wrong check sum for received data 0 Contact Mindray Customer Service Department or your local distributor.
400000220 Operation unit: Serial port error. Instruction can't be 10 Check the connection between the analyzing unit and the operation unit,
transmitted properly. or change the serial port. Then run the control system again.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.
400000221 Operation unit: Serial port error. Can't receive test data. 10 Check the connection between the analyzing unit and the operation unit,
or change the serial port. Then run the control system again.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.
400000225 Operation unit: Insufficient remaining reagent. 0 Check the remaining reagents for the test and add reagents to the bottles
as needed.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
6-41
Troubleshooting
400000252 Operation unit: Calibration factor difference exceeds 0 Check the calibration settings and calibrators. Then recalibrate.
limit.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.
400000253 Operation unit: Multi-point linear or non- linear 0 Check the calibration settings and calibrators. Then recalibrate.
correlated calibration factor (R2) exceeds limit.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.
400000254 Operation unit: Standard difference of calibration curve 0 Check the calibration settings and calibrators. Then recalibrate.
exceeds limit.
If this error occurs frequently. Contact Mindray Customer Service
Department or your local distributor.
400000255 Operation unit: Can't calculate calibration parameter by 0 Check the calculation method.
given method
If the error remains, contact Mindray Customer Service Department or
your local distributor.
400000276 Operation unit: Received data overflow. 10 Contact Mindray Customer Service Department or your local distributor.
400000281 Operation unit: Calibration zero concentration 1 Check the calibration settings and calibrators. Then recalibrate.
absorbency exceeds limit.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.
6-42
Troubleshooting
400000287 Operation unit: Excessive dark current. Test prohibited. 10 Other errors have higher priority.
If the error remains, contact Mindray Customer Service Department or
your local distributor.
400000289 Operation unit: Response exceeds calibration range. 0 Dilute the sample and re-run it.
Can't calculate result concentration.
400000290 Operation unit: Main unit can't recover from wrong 8 Contact Mindray Customer Service Department or your local distributor.
status.
400000291 Operation unit: Startup pretreatment error. Wrong blank 10 Use the emergency exit function to exit the control system and then restart
test result. it.
If this error remains, Contact Mindray Customer Service Department or
your local distributor.
6-43
Troubleshooting
400000293 Operation unit: Pretreatment not finished properly. 10 This error is triggered by other errors. Check the causes of the root errors
and take the recommended corrective measures. Use the emergency exit
function to exit the control system and then restart it.
If this error remains, contact Mindray Customer Service Department or
your local distributor.
400000294 Operation unit: Shutdown. Not finished properly. 10 This error is triggered by other errors. Check the causes of the root errors
and take the recommended corrective measures. Use the emergency exit
function to exit the control system and then restart it.
If this error remains, contact Mindray Customer Service Department or
your local distributor.
500000001 LIS: Unconnected 0 Check if the LIS server is power on and can work properly.
Check if the line between the system and the LIS can work properly.
If this error remains, contact Mindray Customer Service Department or
your local distributor.
500000002 LIS: Sending HL7 message failed 0 Check if the LIS server is power on and can work properly.
Check if the line between the system and the LIS can work properly.
If this error remains, contact Mindray Customer Service Department or
your local distributor.
6-44
7 Calculation Methods
Endpoint
Fixed-time
Kinetic
7.1.1 Endpoint
The endpoint reaction is the most ideal. The reaction reaches equilibrium after
certain period. Because the equilibrium constant is very large, it can be considered
that all substrates have changed into products, and the absorbance of the reacting
liquid does not change any more. The absorbance change is proportional to the
amount of the product present.
t1 t2 t3 t
As shown in Figure 7-1, t1 is the time when the reagent is added, and t 2 is the
time when the sample is added. The reaction starts when they are mixed. At t 3 the
reaction reaches the equilibrium and the absorbance reading is taken. The reaction
time is t 3 − t 2 .
7.1.2 Fixed-Time
For the fixed-time reaction, the reaction velocity (v), within a specific period, is
proportional to the substrate concentration [S], namely, v=k[S]. As the substrate is
consumed continuously, the reaction velocity becomes smaller and smaller, and so
7-1
Calculation Methods
does the change rate of the absorbance. It takes much time for such a reaction to
reach the equilibrium. Theoretically, the absorbance reading can be taken at any
time. The reaction can, however, become steady only after a delay because it is
complicated at the beginning and there are miscellaneous reactions due to the
complex serum compositions. For any first order reaction, the substrate
concentration [S] at a specific time (t) during the reaction is:
[S ] = [S 0 ]× e − kt
Where,
k - velocity constant
Within the interval between t1 and t 2 , the relationship between ∆[S] (delta
substrate concentration) and [S0] is:
− ∆[ S ]
[ S 0] = − kt1 − kt 2
e −e
That is, within a fixed time interval, the delta substrate concentration is proportional
to the original substrate concentration, which is the generality of the first order
reactions. Within this interval, absorbance change is proportional to the substrate
concentration. See Figure 7-2.
t1 t2 t3 t4 t
As shown in Figure 7-2, t1 is the time when the reagent is added and t 2 is the
time when the sample is added. From t 3 the reaction becomes steady and t 4 is
the time to stop monitoring the reaction. The absorbance readings are respectively
taken at t 3 and t 4 .
Methodologically, the fixed-time method requires more than the endpoint method
does. All factors that have effects on the reaction velocity, such as pH, temperature
and enzyme concentration, must keep constant in the two-point analysis. In addition,
the timing must be precise and the calibration is necessary.
7-2
Calculation Methods
7.1.3 Kinetic
For the kinetic reaction, the reaction velocity is not related to the substrate
concentration and remains constant in the reaction process. As a result, for a given
wavelength, the absorbance of the reacting liquid changes evenly, and the change
rate (∆A/min) is proportional to the activity or concentration of the subject, which is
usually the enzyme.
In fact, it is impossible for the substrate concentration to be large enough, and the
reaction will no longer a zeroth order reaction when the substrate is consumed to a
certain degree. Therefore, the theory only stands within certain period. In addition,
the reaction can become steady only after a certain period because the reaction is
complicated at the beginning and there are miscellaneous reactions due to the
complex serum compositions. All reagent manufacturers have strict specifications for
these two periods. See Figure 7-3.
t1 t2 t3 tn
t
As shown in Figure 7-3, t1 is the time when the reagent is added. t 2 is the time
when the sample is added. From t 3 the reaction becomes steady. t n is the time to
stop testing the reaction. t 3 − t 2 is the time delay, and t n − t 3 is the monitoring
time.
AD Value
Absorbance
Response
Calibration Parameter
QC Conclusion
7-3
Calculation Methods
I0 It
As shown in Figure 7-4, when a parallel monochromatic light beam whose intensity
is I 0 goes through a flow cell (whose length is L ) containing a solution (whose
concentration is C ), some photons are absorbed, and the intensity is attenuated
from I 0 to I t , so the absorbance A of this solution is:
It
A = − Lg
I0
Where,
It
- transmissivity
I0
The BS-300 measures the light intensity through the photo-electric conversion, linear
amplification and A/D conversion. For the light intensity signal I i of Channel i, the
AD output Di is:
Di = K pe ⋅ K a ⋅ K ad ⋅ I i
Where,
Di - data of Channel i
So,
Ii0 D
Ai = lg = lg i 0
Ii Di
7-4
Calculation Methods
Where,
Ai - absorbance of Channel i
Di 0 - background AD output
In theory, when the lights are off, the AD output of each channel will be zero. In
practice, because of the existence of dark currents, there is still a background
output Dibackground , which should be deducted. Then, the complete absorbance
formula should be:
Di 0 − Dibackground
Ai = lg
Di − Dibackground
For the analyzer, the response (R) is defined as the absorbance change before and
after the reaction, or the absorbance change rate during the reaction process.
The formula for calculating the response (R) is closely related to the reaction type
(kinetic, fixed-time, endpoint), the number of reagents (single, double), and the
number of wavelengths (single or double). They are respectively detailed in the
following sections.
t1 t2 t3 t
As shown in Figure 7-5, t1 is the time when the reagent (volume: V) is added. t2
is the time when the sample (volume: S) is added. The reaction starts when they are
mixed. At t 3 the reaction reaches the equilibrium and the absorbance reading is
taken. t 3 − t 2 is the reaction time. So,
R = Rs − RSMPB
7-5
Calculation Methods
Both of Rs and RSMPB are calculated with the formula R=At3 − K1 × At2 −1 .
Where,
RSMPB - response of sample blank. If the sample blank has not been requested,
RSMPB =0.
At 3 - absorbance at t3
t1 t2 t3 t4 t
As shown in Figure 7-6, t1 is the time when reagent 1 (volume: V1) is added. t 2 is
the time when the sample (volume: S) is added. Then, reagent 1 and the sample are
mixed. t 3 is the time when the reagent 2 (volume: V2) is added. Then, reagent 2 is
mixed into the reaction liquid. At t4 the reaction reaches equilibrium. t3 − t2 is the
incubation period, and t 4 − t 3 is the reaction time. So,
R = Rs − RRB
7-6
Calculation Methods
Where,
At4 - absorbance at t 4
t1 t2 t3 t4 t
As shown in Figure 7-7, t1 is the time when the reagent (volume: V) is added and
t 2 is the time when the sample (volume: S) is added. t 3 is the time from which the
reaction becomes steady. t4 is the time the test on the reaction is stopped.
Absorbance is taken respectively at t 3 , t4 and the response R is:
At4 − At3
R=
t4 − t3
7-7
Calculation Methods
Double-wavelength
The response is:
Where,
t1 t2 t3 tn
t
As shown in the figure, t1 is the time when the reagent (volume: V) is added. t 2 is
the time when the sample (volume: S) is added. t 3 is the time from which the
reaction becomes steady. tn is the time when the test on the reaction is stopped.
t 3 − t 2 is the time delay, and t n − t 3 is the reaction time. The response ( R ) is
equal to the slope of the linear section between t 3 and tn .
Double-wavelength
The calculation method is similar to that for the single-wavelength reaction. The only
difference is that the absorbance at a certain point is equal to the primary
wavelength absorbance minus the secondary wavelength absorbance.
7-8
Calculation Methods
The BS-300 provides two calibration methods: linear calibration and nonlinear
calibration.
The linear calibration includes the single-point linear calibration (also called factor
method), two-point linear calibration (also called linear method), and multi-point (over
3 points) linear calibration (also called linear regression method). They are mainly
used for tests determined by colorimetry.
Where,
R - calibrator response
K - calibration parameter
R
This calibration method adopts only one calibration parameter K , and K= .
C
This calibration method requires only one calibrator. For most enzyme tests, you can
input the factor directly (input value F=1/ K ) without any calibration.
Where,
R - calibrator response
K , b - calibration parameters
R2 − R1
K=
C 2 − C1
7-9
Calculation Methods
R2 − R1
b = R1 − ( ) × C1
C 2 − C1
This calibration method requires two calibrators. C 1 and C 2 are respectively the
concentrations of calibrator 1 and calibrator 2. R 1 and R 2 are respectively the
responses of calibrator 1 and calibrator 2.
Where,
R - calibrator response
K , b - calibration parameters
∑ Ci
i =1
2
− (∑ Ci ) 2 / n
i =1
n n n
n ∑ CiRi − (∑ Ci)(∑ Ri) / n n
b = ( ∑ Ri ) / n − [ i =1 i =1 i =1
]( ∑ Ci ) / n
n n
i =1
∑ Ci 2 − (∑ Ci) 2 / n i =1
i =1 i =1
Logistic-Log 4P
1
Calibration formula: R = R0 + K
1 + exp[−(a + b ln C )]
Where,
R - calibrator response
K , R 0, a , b - calibration parameter
7-10
Calculation Methods
This calibration method requires at least four calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding R is equal to R 0.
This calibration method is applied to the calibration curve that the response becomes
smaller and smaller with the concentration increase. See Figure 7-9.
C1 C2 C3 C4 C
Logistic-Log 5P
1
Calibration formula: R = R0 + K
1 + exp[ −(a + b ln C + cC )]
Where,
R - calibrator response
K , R 0, a , b , c - calibration parameter
This calibration method requires at least five calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding R is equal to R 0.
The applications of the calibration method are the same with that of Logistic-Log 4P,
but this method has a higher fitting.
Exponential 5P
Calibration formula: R = R0 + K exp[a ln C + b(ln C ) + c(ln C ) ]
2 3
Where,
R - calibrator response
K , R 0, a , b , c - calibration parameter
7-11
Calculation Methods
This calibration method requires at least five calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding R is equal to R 0.
This calibration method is applied to the calibration curve that the response
increases sharply when the concentration reaches a specific value. See Figure 7-10.
C1 C2 C3 C4 C5
C
Polynomial 5P
R − R0 R − R0 2 R − R0 3
Calibration formula: ln C = a + b ( ) + c( ) + d( )
100 100 100
Where,
R - calibrator response
R 0, a , b , c , d - calibration parameter
This calibration method requires at least five calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding R is equal to R 0.
This calibration method is applied to the calibration curve that the response
increases sharply when the concentration reaches a specific value.
Parabola
Calibration formula: R = aC 2 + bC + c
Where,
R - calibrator response
a, b, c - calibration parameter
7-12
Calculation Methods
Spline
Calibration formula: R = R0i + ai C + bi C + ci C
2 3
Where,
R - calibrator response
This calibration method requires 2-6 calibrators. Because this method adopts the
segment fitting, its fitting degree is the highest among all calibration types.
Where,
K - calibration parameter
Or,
C = R×F
Where,
F - input factor.
7-13
Calculation Methods
R −b
C=
K
Where,
K, b - calibration parameters
Where,
K, b - calibration parameters
Logistic-Log 4P
K
− a − ln( − 1)
R − R0
C = EXP( )
b
Where,
Logistic-Log 5P
Obtain the positive real root with the dichotomy method.
Exponential 5P
Obtain the positive real root with the dichotomy method.
Polynomial5P
R − R0 R − R0 R − R0
C = exp( a + b ( ) + c( )2 + d ( )3 )
100 100 100
Where,
Parabola
7-14
Calculation Methods
Obtain the positive real root of the following linear quadratic equation:
aC 2 + bC + c − R = 0 .
Spline
Obtain the positive real root with the dichotomy method.
Westgard multi-rule includes ten sub-rules. The clinical meanings of the ten
sub-rules are as follows:
7-15
Calculation Methods
QC conclusion flow:
QC Data
No
12S within control
Yes No
No No No No
22 S R4 S 41 S 10 x
13 S
out-of-control
7.2.5.2 X -R
This rule obtains the difference (R) between two tested values (X1 and X2) of a same
control, and then compares it with the input SDR. If R is larger than 2SDR or smaller
than -2SDR, the rule considers it out-of-control. The value of SDR should be obtained
in the following procedures: After pre-testing the control, test it continuously for 20
days. Test it twice before and after adding the sample everyday. Obtain the SDR of
the difference (R) between two tested values (X1 and X2).
7-16
Calculation Methods
6 When the cumulative sum overruns the control limit h (larger than the
upper limit or smaller than the lower limit), the system considers it
out-of-control.
This rule calculates the difference (error, expressed with the symbol plus or minus)
between the tested value and target value of a test of a control.
When the cumulative error overruns the cumulative error limit E or –E, the system
considers it out-of-control.
7.2.5.5 Twin-plot
+3SD
+2SD
-2SD
-3SD
The QC chart, which can sensitively indicate system errors and random errors, is present to
help you make a QC conclusion.
7-17
Appendix A Specifications
Mixing bar: for single-reagent tests, it functions immediately (within the same
period) after sample dispensing; for double-reagent tests, it functions
immediately (within the same period) dispensing of the second reagent.
A-1
Specifications
A-2
Specifications
A.7 Interface
LIS interface: Mindray HL7 interface
Interface between Analyzing Unit and Operation Unit: RS-232
A-3
Appendix B Supplies
To ensure personal safety and system performance, use only supplies manufactured
or recommended by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.. Contact
the Mindray Service Department or your local distributor for details.
B-1
Supplies
B-2
Appendix C Index
C-1
Index
dialog box ..... 1-16, 3-4, 4-12, 4-13, hydrochloric ............................ 4-40
4-21, 4-22, 4-27, 4-29, 4-40,
4-41, 4-42, 4-49, 4-50, 4-53,
4-63, 4-70, 4-74, 4-83, 4-87, I
4-91, 4-118, 4-135, 4-136, 4-137,
4-139, 4-142, 4-170, 4-171, Incubation ........................ 3-4, 4-45
4-173
inner circle ....................... 1-4, 4-65
Diluents ..................................... 1-4
K
dilution ratio ......... 4-3, 4-13, 4-139
Kinetic. 4-8, 4-43, 4-44, 4-45, 4-47,
Door
4-48, 4-138
middle........................................ 1-5
L
double-reagent .. 1-10, 4-2, 4-7, 4-8,
4-9, 4-43, 4-45, 4-47, 4-48, 4-65 lamp .......................................... 3-5
handle ..................... 5, 1-4, 1-5, 1-7 option button ................ 1-17, 4-138
C-2
Index
probe rotor ..........................1-5, 1-8 Routine...... 1-15, 4-1, 4-7, 4-8, 4-9,
4-10, 4-40, 4-135, 4-136
probe tip..................................... 3-2
Running........... 3-6, 3-7, 4-17, 4-21
profile ........................................ 4-4
S
Q
Sample
QC parameter....... 4-49, 4-50, 4-87,
4-98, 4-99 disk............................................ 1-4
C-3
Index
sample disk . 7, 8, 1-4, 1-5, 1-8, 3-5, temperature1-7, 1-12, 1-16, 2-1, 2-2,
3-6, 3-7, 3-8, 4-3, 4-8, 4-9, 4-11, 2-3, 3-4
4-12, 4-13, 4-71, 4-88, 4-121
test report ................... 4-169, 4-171
sample disk diagram ................ 4-11
Tube .......................................... 1-4
sample dispenser........................ 1-5
short-cut button.. 4-1, 4-7, 4-8, 4-40 used-cuvettes bucket ................. 3-3
C-4
P/N: BA33-20-35244 (1.3)