Operation Manual: Statement
Operation Manual: Statement
SBA-733 Plus
Operation Manual
Statement
Sunostik Company has the right to modify the Operation Manual without prior notice.
Sunostik Company has the right to change the technologies without prior notice.
Sunostik Company has the right to revise the specification of the product without prior notice.
Sunostik Company don’t guarantee anything about this document; including the liability for
guarantee that suggested the marketability and compatibility for same special items
Sunostik Company will be responsible for the instrument’s safety and reliability ,but only when
you make sure the following items you has done correctly:
▲Assembly operations, expansion, re-tune, improve and repair must be done by Sunostik
company’s engineer.
▲The related Electrical equipment must be satisfy with the national standards.
▲You must run the instrument under the correct operational guidance.
In the following situation, Sunostik Company will not be responsible for the instrument’s safety
and reliability.
▲The instrument were Disassembled, stretched, re-commissioned by not Sunostik Company’s
engineer.
▲Repair, modify and dismantle the instrument with non-authorized by Sunostik company.
▲The process of Running the instrument is against to the 《Operation Manual》,and caused the
instrument damage.
After-Service
When your instruments need doing free technical service, your instruments must be satisfy with
the Sunostik company warranty service ordinance.
1
SBA-733Plus Manual
1 Your instrument is not satisfied with the Sunostik company warranty service
ordinance.
2 In the following situation, even your instrument is in warranty, we will not doing
free service to you: Damage by the human, un-correctly using, adopt volt was
un-normal, Irresistible natural disasters, replace the parts and consumables are
not allowed by Sunostik or repair by the person that Non-authorized officer of
Sunostik company.
Contact Us
2
SBA-733Plus Manual
Preface
Thank you for your purchase and use our SBA-733 Plus semi-automatic biochemical analyzer.
This operation manual will guide you how to properly run this instrument .
Before using, please carefully read and understand the contents of this operation manual for
operating the equipment properly, After reading, please safekeeping the operation manual and
placed to the right position for easy getting.
Product name:Semi-Automatic Biochemistry Analyzer
Model:SBA-733 Plus
Product registration number: Kyrgyz armed Food Drug Administration (quasi-) character No.
2007, No. 2400064
Product Standard Number:YZB/JI0185-2006
Production license number:Ji Food Drug Administration Xu 20,051,008 Number of firearms
production
Registered Address:733 Kunshan Road, Economy and Technology Development Zone of
Changchun City, Jilin Province, China.
Production address:733 Kunshan Road, Economy and Technology Development Zone of
Changchun City, Jilin Province, China. The operation manual Release Date: August 2009.
Version:1.0.
File Number:SUNO/SOP-733P-01
Intended Use:
SBA-733 Plus semi-automatic biochemical analyzer is used in clinical diagnosis instrument,the
product can be used for routine biochemical test items:Carbohydrates, enzymes, proteins, lipids,
drug abuse, therapeutic drugs testing, as well as emergency and special biochemical tests, specific
proteins, the immune class category and electrolyte testing.
User
The target audience of this manual is for the professional stuff that working in clinical
laboratory. The following people are included:
The staff who operating the SBA-733 Plus in the daily working.
The staff who doing maintenance of this equipment.
The staff who is learning how to operating this equipment.
Content
This operation manual is for the SBA-733 Plus semi-automatic biochemical analyzer. The purpose
of this operation manual is mainly for helping users understand the SBA-733 Plus semi-automatic
biochemical analyzer’s working principle, structure, operational method, routine maintenance,
problem fixing and so on. Please follow this instructions to operate this instrument.
3
SBA-733Plus Manual
Sign
Symbol signs and terminology
Based on the different content, we using some symbol signs for marking, please notice each
symbol sign’s meaning correctly when you read thie user manual.
These signs are as follow:
Sign Description
1.Remind the users following the operation manual, otherwise, may cause personal
injury.
2.Remind the users following the operation manual, otherwise, may damage the
system or affect the test results.
3.Explain the important information of operation by step. or other contents needs the
users attention.
Remind the users following the operation manual, otherwise, there is the
dangerous of biological contamination.
Sign Description
CE Logo
Manufacturing Date
Warning: high-temperature
4
SBA-733Plus Manual
Picture
The pictures in this operation manual are only as the description or give a demonstration, not for
other purposes.
Safety Precautions
In order to use this system safely, please read the following safety precautions. Any breach of the
right operate process may destroy the system damage or personal injury.
Warning
If the user does not run the instrument as the operation manual told you, this may
destroy the equipment and affect the test results.
Warning
When the instrument’s power was on, if you found liquid went inside the instrument
or the instrument was leaking by itself, if the liquid into the system or the systems
internal leakage, Please cut off the power supply immediately, and contact Sunstik's
customer service department or distributors. Liquid may cause electric shock and
destroy the system.
Incorrect use of the sample may lead to infection. Do not touch the sample,
mixtures and liquid waste directly use hand, please wear gloves, work clothes to
avoid being infected when operation, if necessary, please wear protective glasses. If
the sample contact with the skin accidentally ,please processing following the user
standard immediately. Some agents have a strong acid or alkali. Please be careful
using , to prevent the hand and close thing to contact the reagents directly .if the
reagents contact with the hands and clothing accidentally, please washing with soap
and water immediately. If the reagent gets in the eyes accidentally ,please washing
with water immediately and consult an ophthalmologist immediately.
5
SBA-733Plus Manual
Attention Items
In order to operate the equipment correctly and effectively, please pay more attention to the
following items.
System Usage
Warning
This equipment applies to serum, plasma, urine and cerebrospinal fluid. Out of this
scope, please consult SUNOSTIK. Please reference other clinical symptom or other
test results, when estimate the clinical result.
Environment
Attention
Please set the equipment in the appointed environment followed by instruction. Out of
the appointed environment, the test result will be incorrect and broke the system.
If want to change the system, please connect SUNOSTIK service hotline or local
distributor.
6
SBA-733Plus Manual
Attention
Please do not put a noise equipment near to the system. Please turn off any
electromagnetic equipments in the room, where the system in. Such as mobile phone,
radio equipment and so on. And do not use any CRT monitor near the system. Noise,
electromagnetic wave can be cause the malfunction and affect the result.
System Use
Attention
(1)Please performance the equipment followed by instruction. Incorrect use will cause
the wrong results; even the system and person will be harmed.
(2)First time, please check the control to confirm the equipment is run.
(3)After turn off the main power 10 seconds, open again. Otherwise, the system will
be protected. If it is, please turn off the main power then turn on again.
System Maintenance
Please maintain the system regularly followed by below notes. Otherwise, lifetime and
characteristics will be affected.
Attention
(1)Please maintain the system followed “System Maintenance” in the instruction.
Incorrect maintenance will cause the wrong results, affect its characteristics and
lifetime, even system and person can be harmed.
(2)The surface will accumulate dust in long time. Use soft cloth to put in the water
and twist it, then rub the surface lightly. Immerge the cloth in to soap liquid if
necessary. Do not immerge the cloth in ethanol or related solvents. At last, use the
dry cloth to clean.Before cleaning, please close the power and pull out the plug.
During cleaning, please prevent water goes into the system, otherwise, system and
person can be harmed.
(3)Do performance the control analysis, after change the main parts, such as source
lamp.
Parameters Set
Attention
Before use this equipment, please program the volumes of sample, reagent, wavelength
and so on, please follow by the instruction sheet and reagent sheet.
7
SBA-733Plus Manual
Content
AFTER-SERVICE .........................................................................................................1
PREFACE ......................................................................................................................3
USER ...........................................................................................................................3
CONTENT .....................................................................................................................3
SIGN ............................................................................................................................4
LABEL AND SCREEN PRINTING INSTRUCTIONS ..............................................................4
SAFETY PRECAUTIONS .................................................................................................5
PREVENT ELECTRIC SHOCK ..........................................................................................5
BIOCHEMICAL RISK PROTECTION .................................................................................5
WASTE FLUID TREATMENT ...........................................................................................6
ATTENTION ITEMS .......................................................................................................6
ENVIRONMENT ............................................................................................................6
ELECTROMAGNETIC WAVE AND INTERFERENCE PREVENTION ......................................7
SYSTEM USE ................................................................................................................7
SYSTEM MAINTENANCE ..............................................................................................7
PARAMETERS SET ........................................................................................................7
CHAPTER 1 INSTRUMENT INSTALLATIONS ......................................................10
1.1 UNPACKING .........................................................................................................10
1.2 INSTALLATION ENVIRONMENT..............................................................................11
1.3 TEMPERATURE AND HUMIDITY REQUIREMENT .....................................................12
1.4 POWER SUPPLY REQUIREMENT .............................................................................12
1.5 WASTE LIQUID BOX ..............................................................................................12
CHAPTER 2 EQUIPMENT OVERVIEW ..................................................................13
2.1 PHYSICAL FORM ..................................................................................................13
2.2 BACK FORM .........................................................................................................13
2.3 MEMBRANE KEYBOARD .......................................................................................14
2.4 DISPLAY SCREEN .................................................................................................15
2.5 THERMAL PRINTER ..............................................................................................15
2.6 RS232 INTERFACE ...............................................................................................15
2.7 PIPETTE AND IMBIBITION SWITCH ........................................................................16
2.8 COLORIMETRY SYSTEM .......................................................................................16
CHAPTER 3 SOFTWARE INTRODUCTION ...........................................................18
3.1 SELF-TEST ...........................................................................................................18
3.2 WINDOW DESCRIPTION ........................................................................................19
3.3 PARAMETERS SETTING .........................................................................................21
3.4 APPLICATION MENU.............................................................................................22
CHAPTER 4 PROJECT PROGRAMMING ...............................................................27
4.1 PROGRAMING INTRODUCTION ...........................................................................27
8
SBA-733Plus Manual
9
SBA-733Plus Manual
Warning
SBA-733 Plus are required to install by SUNOSTIK’s engineer or by the people who
was authorized by Sunostik company.
1.1 Unpacking
On receiving this instrument, check whether the packing is in perfect condition and
that the seals are intact. Open the box and take out its content with care (Figure- 1.1).
Read this chapter carefully before installing the instrument. Do not throw the packing
material away. In case of the re-shipment, use it again to guarantee good transport.
10
SBA-733Plus Manual
Attention
To avoid the un-necessary argument, please check each thing in side the box, make sure
everything that wrote on the packing list is prepared.
Packing List
Description Quantity
Main unit 1
Fuse(2A) 4
Power cable(3 cores) 1
User’ Manual 1
Thermal paper(58mm) 1
Wash solution(100ml) 1
Waste bottle 1
Dust cover 1
Ground connection(1.5m) 1
Silicon tube 1
serial port line(1.5m) 1
Attention
The working environment should be well ventilated, if necessary, ventilation equipment
11
SBA-733Plus Manual
can be used. However, should avoid blowing air directly to the system.
Warning
If the system is idle for a long time do not use, this may speed up the system’s
mechanical and circuit aging process, so run the instrument on a regular time is required.
12
SBA-733Plus Manual
②
⑥
③ ④
① printer cover ② incubation bit cover ③ pipette ④ imbibitions key ⑤ liquid crystal
display ⑥ keyboard
① ② ③ ④ ⑤ ⑥
Attention
The environment should be ventilation to make the emitting system work
normally, if necessary use the ventilation device. But please prevent the air be blew to
the system directly, otherwise the data will be affected.
13
SBA-733Plus Manual
14
SBA-733Plus Manual
Characteristics:
240×128 dot matrix
LED backlight design
Toshiba T6963C controller
On-chip 32K Byte SRAM
Characteristics:
384dots/line
Resolution 8dots/mm
Max print speed 75mm/s
Built-in thermistor to check the needle of the printer.
Built-in sylvatron to check the paper.
15
SBA-733Plus Manual
Attention
When checking the sample, please mark sure the pipette is straight to prevent the
result accuracy is affected.
The system is controlled by CPU, temperature control and wavelength lookup both be
proceeded under the CPU. To be Noted that you’d better take out the silicon tube from
the peristaltic pump when the instrument will not be used in long time to avoid the
tube block.
16
SBA-733Plus Manual
Warning
The colorimetric system is the main part of this equipment, unprofessional
untrained person don’t open it randomly. Otherwise the properties will be affected.
17
SBA-733Plus Manual
3.1 Self-test
After open the power switch, (see ① in the picture of the equipment back form
picture), the display is as below:
The self-test will be started 5 seconds later, to test the memory, printer, filter and
sampling circuit separately and display the status real time. The below picture will be
displayed after self-test completely.
Flash Memory:16.5Mbit
RAM 32Kbit
Printer test OK
Filter test OK
Sample Test OK
Lamp Test OK
Test Self OK
The temperature adjustment will be started after the self-test, refer to below picture.
Main menu will be indicated after the temperature adjustment, or you can enter the
main menu by pressing any key directly. At this time, the temperature adjustment is
kept going without any display.
18
SBA-733Plus Manual
Temperature adjusting…
20%
Attention
You’d better to start the test after complete the temperature adjustment to make the test
accuracy.
Window Format:
All windows include in this system with the same format, such as main menu(below
picture): There are three parts in the window: title bar up the window to display the
subjects and time, middle is the customer area to indicate the parameter list and
submenu information, status bar down the window to indicate shortcut key and cue
information.
19
SBA-733Plus Manual
Sample Test
Attention
Press the shortcut keys【1】~【6】to enter the submenu directly.(From left to right; from
up to down). Shortcut key “R” means to delete the results; “P” means to print; “T”
means set time; “M” means to A/D; “U” means to update the window; “H” means the
basic information.
20
SBA-733Plus Manual
⑹ QC query: Under this menu, you can check the saved QC information for 100 test
items. There are 3 QC type under each item, each QC type can save the QC result 100
days. There are 2 modes for QC indication (QC data list and QC chart), both can be
printed out. You can evaluate the accuracy of the test items through the QC chart, the
instrument also can automatically calculate out the QC data’s average, differentiation
and variance of the certain month.
Sample Test
21
SBA-733Plus Manual
Attention
The absorption volume calibration and air sipping setting should be correct, incorrect
setting will cause the wrong results.
Language
The operation language is available for many languages. After selection, it works
immediately.
22
SBA-733Plus Manual
View Record
Print Record
Delete Record
Input Code |
Record query: To input the code No. (Range of code No. is 1~200). Press key
【Enter】to start query.
Comprehensive print: Input two code numbers and press key 【Enter】to print all the
records between this two codes.
Comprehensive delete: Input two code numbers and press key 【Enter】to delete all
the records between this two codes.
⑵ Print setting: Enter this submenu to set the print parameter, operate the printer and
check the status of the printer, see below picture:
Test Paper
Lab Name: The default is SUNOSTIK, it can be adjusted according to the customer
request, and the name will be printed in the test report. (It is workable with the
standard print type).
Auto-print: The test result can be auto-printed out after complete the test if this
function is in application.
Print Style: There are two print type is optional, default is standard type, there is
23
SBA-733Plus Manual
blank between the 2 test results to make sure all the content can be indicated out of
the paper outlet. You can select economy type to save the paper, but the result content
will be compactness.。
Print Status: The item is for read only to indicate the printer status.
Self-test: The shortcut key【F1】is used for testing the paper and checking the printer.
Paper Moving: The shortcut key【F2】is used for moving paper, this function can be
acted when the printer is installed. Repress it to stop the moving.
⑶ Time System: Using the keys【<】、
【>】to move the cursor to revise the date and
time after enter this submenu. See below picture:
07/08/01 14:36:20
Attention
The time system is very important for working normally, incorrect setting will
mistake the test result and QC data, please take it in serious.
⑷ System Maintenance: Enter in this submenu to check the A/D value in certain
wavelength. See below picture: The default filter is 340nm; A/D value is the last
number. Press key【Enter】to bounce out the wavelength list, you must select a
wavelength to start the sampling, and keep the A/D value continuous and in real
time.
24
SBA-733Plus Manual
⑸ Software Update: You can follow the below picture to apply the latest software to
upgrade the functions. See below picture:
Please restart the power for running the new procedures after update.
Download Guide:
Use the matched connect line to connect the computer through RS-232 interface(Rear
view③postion),Open the updated software that we provide to download and
update .Software interface as below:
① ⑧
⑦
⑥
③ 25 ④
⑨
SBA-733Plus Manual
Help:
③⑤⑥⑦ position : Set up parameters according to the map,⑧ position :According to
the port NO that contacting by equipment and computer for confirmation. After set up
these parameters,Click ① to select the program file ,the extension of the file is .hex,
Such as SBA733.hex。Open the equipment and click ④。After a few seconds,,There
is presentation on ⑨ that show the connecting well or not ,if display "Reading of
PART ID Success" ,it means connect successful,please click ② to download the file
at this time .At the same time there is a blue progress bar showing on ⑨.when the
progress bar run to the right side .download completely .If display "Reading of PART
ID Failed",that means connection failure .please check the line connection carefully,
Whether the line contact with a solid ,to connect again after the check.
Attention
Please shut off the power of the equipment before download the Software ,use the
matched connect line to connect the computer and equipment. It may burn out the
equipment when connection with the electricity.
⑹ Help:Under this interface you can find the basic information of the equipment and
the contact information of our company, see as below.
www.sunostik.com
26
SBA-733Plus Manual
Press【<】、
【>】key to move the cursor to switch the focus parameter;Press the
shortcut key【F1】、
【F2】、
【F3】、
【F4】 to achieve the function that opposite on the
status bar ;Press【Enter】key to edit the focus parameter ,at this time ,the cursor is
moved to the editing window or the sub-window ;you can use【Clear】key to delete the
mistake during the editing;;After the editing ,press【Esc】key to exit the submenu and
give up the content which is edited . press【Enter】key to confirm and exit the editing
window or the sub-window, back to the above interface.
27
SBA-733Plus Manual
28
SBA-733Plus Manual
Edit with the form of sub-window .there are 13 ordinary Units for selection. please set
up this parameter according to the user’s manual of the reagent correctly.
Reference Upper Limit:
Edit with the form of edit-window, This is the upper limit of the reference
range .please set up this parameter according to the user’s manual of the reagent
correctly.
Reference lower Limit:
Edit with the form of edit-window, This is the lower limit of the reference
range .please set up this parameter according to the user’s manual of the reagent
correctly
Correction coefficient:
Edit with the form of edit-window. When the test results is generally low or too high,
you can use this parameter to amend the results.
Attention
Default number is 1,To setting this coefficient should be according to the huge
statistic data, it should be applied in serious.
29
SBA-733Plus Manual
1. Standard Thickness:
If the calibration mode is standard, the standard thickness should be input for
calculation. If the calibration is Factor, this parameter will inefficacy and hided
automatically. If the test method is Kinetics, this parameter will be inefficacy and
hided automatically. The standard thickness will be show X when the test method is
multi-standard.
2. Standard value X:
The standard value X should be input under the multi-standard test method(each
thickness of the standard liquid),The maximum range is 2-6 , it is according to the set
of standard quantity. Under the non multi-standard method, this parameter will be
inefficacy and hided automatically.
Attention
The setting of the standard value has to be followed from small value to
big value , or it will cause the mistake result.
4.4 QC data
Target H、Target M、Target L:
Edit with the form of edit-window, separate as QC high value, QC middle value, QC
low value. please set up this parameter according to the user’s manual of the reagent
correctly
Standard difference H、Standard difference M、Standard difference L:
Edit with the form of edit-window, separate as the standard difference for QC high value, middle
value and low value, to be applied for plot of the QC chart. please set up this parameter according
to the user’s manual of the reagent correctly
30
SBA-733Plus Manual
Sample Test
Press 【Enter】or shortcut key【1】to enter the following menu on above picture,
namely the “sample test”, see below picture:
ITEM 07/08/01
01 AST 06 CL-,MB
02 ALT 07 ALP
03 a-HBDH 08 α-AMYS
04 LDH 09 γ-GT
05 CK 10 GLU
Input:| END/P NEXT/P
Please input the test item No under this menu. if the item is not exist on the interface ,
press shortcut key【F3】
【F4】,according to the next page indication on the screen to
search the item which you need, total are 10 pages cycling, After input the No, press
【Enter】to enter the baseline test interface.
Attention
For example: If you want to test “AST”,please press the key【1】and 【ENTER】enter
to the baseline test interface ,see as below.
31
SBA-733Plus Manual
Baseline
Mean = … CV% = …
32
SBA-733Plus Manual
5.2 End-point
The interface of end-point testing as below(take GLU for example,only for your
reference).
Blank = 0
Factor = 1746
The default parameter will be the last test data of blank and factor. you can directly to
start the sample or QC test if the indicated data is in your requirement. If the
parameter does not match your inquire or you want to reset the parameter for
increasing the test result .If not, you can press the shortcut key according to the
content of the status bar, reread, blank and standard. below is the instruction for the
four interfaces: Press【F1】to enter the test interface for blank test, see below picture:
33
SBA-733Plus Manual
Blank =
Factor =
The work procedure is the same with absorbing the distilled water , we won’t mention
again, it will return the up menu after reading blank, the data in the system is updated,
if the result is not fit for you, you can retest it follow the same procedure .retest the
blank. Press【F2】to enter into the interface of standard, see as below:
Standard =
Factor =
The work procedure is same with reading the baseline procedure, we won’t mention
again. It will return the up menu after the reading standard, the standard and factor
parameter is updated at this time, now it finish setting the standard. If the result is not
fit for you, you can retest it follow the same procedure. Reset the standard. Press【F3】
to enter into the interface of sample test, see as below:
34
SBA-733Plus Manual
Absorbance =
Result=
The default patient No. is 001,the No. can be edited when the cursor in flash. Press
Key 【Clear】to delete, and use numeric key to edit. The range of the No. is 1--999
(the first 200 patient the test result can be saved automatically). Please press the
button to absorb the sample. The procedure is same as absorbing the distilled water .
we won’t mention again. It will come into the below window after absorb the
sample , the result will be sent to the computer through RS-232. If the auto-print
function is on turn(this function is edited in the print editing), the test report will be
automatically printed out .Otherwise, please press【F4】to print the test report.
Absorbance = 0.5547
Test Result = 30.22
You can select the retest or reread function if you are not satisfied with this result.
Retest: Does not to change the patient No. and re-process the above procedure .need
to absorb the sample again
Re-read: Does not to change the patient No, and no need to absorb the sample again,
reread the current sample
If you want to proceed the other sample test under the same item, please press
shortcut key【F1】to enter the test interface and the patient No. will be add 1
35
SBA-733Plus Manual
automatically . it will test the next sample with reprocess the above procedure.
The next test also can be followed with the same procedure. You also can directly
press the button to start the new sample test without press【F1】,the patient No. will be
added 1 automatically and enter to the procedure of absorbance and testing, in order
to simplify the operation.
Attention
Please press 【Wash】 when the test is finished ,to avoid the
cross-contamination .Press【Wash】again to stop washing.
Absorbance =
Result =
Absorbance = 0.5547
Result = 30.22
36
SBA-733Plus Manual
QC, press【F4】to reprocess the above procedure. Under the same item , type and date,
only the latest QC result will be saved.
Blank = 0
Factor = 1746
Black Sample QC
The blank data that display in the data list is default as the latest test result, you can
directly start the sample test or QC if the blank data is in your requirement. If the
parameter is no fit for this test or want to retest the blank in order to increase the test
accuracy. please re-read the blank data base on the indication on the status bar.
Compare with End-point, it is no need to make the standard .so it is no standard
optional .we are only to introduce the interface of sample test, the other procedure
please refer to the End-point test method.
Press 【F3】to inter the interface of sample test .see as below
37
SBA-733Plus Manual
method .Absorbance could real-time display its present value in the whole testing
process of sample .Factor is an assigned numerical value under control of program.
Please press aspirate liquid key to inhale the sample ,then it will enter the phases of
hatching and testing samples. The following window will come out after test.
38
SBA-733Plus Manual
6.2 QC Test
In order to ensure authenticity and validity of QC data ,QC test is programmed and
tested with sample test at the same time .About editing of QC parameters, please see
item programming.
6.3 QC View
Press QC View under main menu or directly press shortcut key【6】to enter into
application control menu ,please see following picture:
Input month: |
39
SBA-733Plus Manual
40
SBA-733Plus Manual
Chapter 7 Maintenance
Incorre correct
When the instrument is on working(when power is turned on ,it the printer processing
power ),at this time, send printing paper to the feeder under the paper delivery
roll ,press【paper feed】,the printing paper will come out automatically.
1
When replacing fuse, firstly twist off the cap ,remove old fuse and install a new
one equipped together with instrument.
41
SBA-733Plus Manual
Warning
If you replaced a new fuse and turned on power ,the fuse was burned at
Once .Under this situation ,please don’t change a higher current fuse any
More ,turn off the instrument and get in touch with our maintenance
engineer at once.
First, twist off the cap, loosen two inner hexangular set screws(2)be used to fix
component parts of lamp base (4) with an inner hexagon spanner,remove the
component parts and remove halogen lamp (1)from lamp base(3).Install a new
halogen lamp on the lamp base ,and install the component parts on the lamp
holder ,fasten the screws with inner hexagon spanner .Please see the following
pictures to check detailed installing position.
3
4
1 2
42
SBA-733Plus Manual
43
SBA-733Plus Manual
After disconnecting the pump ,press the clips on both sides of peristaltic pump ,at
the same time, pull the pump pipe out with hard strength.
Please take the following steps for your reference on how to exchange pump pipe.
44
SBA-733Plus Manual
45
SBA-733Plus Manual
Warning
Most maintenance were carried on under power off state ,so please maintain it after
turning off power for safety.Never spray any liquid like reagent to instrument or
electronic accessories.
Attention
When wiping flow cell ,do choose the soft material like lens paper to prevent cutting
light-hole which will affect the Visible Light Transmission .At the same time ,pay
attentions to installing direction.
46
SBA-733Plus Manual
Warning:
When settling troubles ,please handle it with great care ,check whether deal it after
power is turned off
Examples of
Reasons analysis Eliminate methods
trouble
1 power cable is broken or on a poor
1 Replace or inspect the power cable
Power light don’t contact
2 Replace the fuse
work 2 Fuse blows out
3 Replace the switch
3 Power light fails
1 Have not pluses negative voltage of 1 Inspect the negative voltage of LCD
LCD 2 Check if flat cable and LCD are in good
LCD don’t work
2 Flat cable is off contacts
3 LCD is broken 3 Replace the LCD
1 Control circuit broke down
1 Check the control circuit
Peristaltic pump 2 Aspirating liquid switch is broken or
2 Adjust or replace injector port
don’t work or don’t on a poor contact
3 Replace the peristaltic pump
aspirate liquid 3 Peristaltic pump is broken
4 Replace the pump pipe
4 Inside pump pipe blocks.
1 Clean the flow cell repeatedly with distilled
water and check autosampler system
2 Check whether the reagent is out of date.
1 There are bubbles or dirt in flow cell
3 Check whether your selected filter confirms
2 Reagent turns bad
to test requests.
3 Selected filter don’t confirm to this
4 inspect the whole operating process and
test.
Test abnormity test again.
4 Incorrect operation or large error
5 Check the halogen lamp and test the voltage
5 Halogen lamp is broken or with an
of it(6V more or less)
insufficient light source
6 Check whether couveuse’s temperature
6 Couveuse’s temperature is inaccurate
confirms to the requested test
temperature ,if the error is big ,adjust
temperature to requested test temperature.
1 Control panel is broken 1 Check the control panel and replace it.
Keyboards don’t
2 Fast booting and Shutting down lead 2 Reboot the instrument (turn power off and
work
system halted then reboot it)
47
SBA-733Plus Manual
48
SBA-733Plus Manual
49
SBA-733Plus Manual
50
SBA-733Plus Manual
Working temperature:10~35℃
Environment humidity:≤80%
Volume: 430(length)×320(width)×220(height)
Weight: 10 Kg
51
SBA-733Plus Manual
52