Operation Manual
Before using this software, be sure to read this manual and the separate Setup Guide
thoroughly.
Also, read the Digital Radiography CXDI series User’s Manual.
Keep the manual where it is easily accessible.
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 2 Thursday, April 7, 2016 2:28 PM
To the Customers
Thank you for purchasing the Canon CXDI Control Software NE (hereinafter called this product).
Operating instructions are divided into two volumes: the Operation Manual and the Setup Guide.
Before using this product, be sure to read these manuals thoroughly in order to utilize this product more
effectively.
Disclaimer
1. In no event shall Canon be liable for any damage or loss arising from fire, earthquake, any action by a third
party or other accidents, any intentional or negligent action by the users, any trial usage, or other usage
under abnormal conditions.
2. In no event will Canon be liable for direct or indirect consequential damage arising out of the use or
unavailability of this product.
Canon will not be liable for loss of image data due to any reason.
3. In no event shall Canon be liable for personal physical harm or property damage that is sustained when
the instructions are not followed or the product is misused.
4. It is the responsibility of the attending physicians to provide medical care services. Canon will not be liable
for the faulty diagnosis.
5. Roentgenography, image processing, image reading, and image data storage must be performed in
accordance with the laws of the country or region in which the product is being used.
The user is responsible for maintaining the privacy of image data.
6. Information in this manual may change without prior notice.
7. Screen images in this manual may differ from those displayed on your image capture computer due to
customization, updates and improvements in the software.
8. We’ve taken all possible means to ensure the content accuracy of this manual. If you have any questions
or problems, contact your sales representative or local Canon dealer.
Trademarks
• The Canon name and Canon logo are registered trademarks of Canon Inc.
• Ethernet is a trademark of Xerox Corporation.
• Microsoft and Windows are either registered trademarks or trademarks of Microsoft Corporation in the
United States and/or other countries or regions.
• TrueType is a registered trademark of Apple Inc.
• Intel and Intel Core are trademarks of Intel Corporation in the U.S. and other countries.
• Other systems and product names in this manual are the trademarks of the manufacturers that have
developed them.
Copyright
• All rights reserved.
• Under copyright laws, this manual may not be reproduced, in whole or in part, without the written
permission of Canon.
2
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 3 Thursday, April 7, 2016 2:28 PM
Safety Summary
Before using this product, read this safety summary thoroughly. This information will prevent the users and
persons involved from sustaining physical harm and/or property damage.
Read the separate Setup Guide and the Digital Radiography CXDI series User’s Manual as well.
Safety symbols
The following safety messages are used to emphasize certain safety instructions. This manual uses the
safety alert symbol along with a safety message.
This is used to identify conditions under which improper use of the product may
WARNING
cause death or serious personal injury. Pay attention to the warning message.
This is used to identify conditions under which improper use of the product may
CAUTION
cause minor personal injury. Pay attention to the caution message.
This notice is used to identify conditions under which improper use of the product
CAUTION
may cause property damage. Pay attention to the caution message.
3
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 4 Thursday, April 7, 2016 2:28 PM
Safety Precautions
Follow the safety precautions below and use the software properly to prevent personal injury or equipment/
data damage.
CAUTION
• Be sure to confirm that the entered information (patient name, ID number, date of birth, and sex)
matches that of the patient. If the information is incorrect, the resulting patient mix-up and a
misdiagnosis may cause harm to the patient.
• Be sure to use the [Emergency] button only for an emergency examination. If not heeded, the
resulting patient mix-up and a misdiagnosis may cause harm to the patient.
CAUTION
• Avoid doing the following actions to the detector when [Ready] is displayed in the system status
indicator. Otherwise, the detector may acquire an image without exposure. For details, refer to the
Digital Radiography CXDI series User’s Manual.
- Giving a strong shock and vibration.
- Using the detector in a location where static electricity is easily generated.
- Using the detector in a location where electromagnetic wave noise is strong.
• Depending on the imaging conditions, such as radiation dose, exposure time, irradiated field or
target body positioning, image cannot be acquired even if exposure is performed. When radiation
dose is lacking or exposure was performed without [Ready] display in the system status indicator,
click [Refresh] before performing exposure again. Check the imaging condition is intended value
before starting exposure.
CAUTION
Be sure to click [End Exam] in the examination screen before shutting down the software. Shutting
down the software without ending the examination may cause data loss.
Examination
CAUTION
Instruct the patient not to move during the examination. If necessary, help the patient to maintain
the proper posture. If the patient moves during the examination, it may fail to obtain the appropriate
images.
4
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 5 Thursday, April 7, 2016 2:28 PM
Contents
Safety Summary ............................................................................................................ 3
1 Getting Started
5
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 6 Thursday, April 7, 2016 2:28 PM
Contents
4 Conducting Examinations
5 Reviewing Images
5.3 Deleting studies from the [PAST > Past List] screen....................................... 100
6
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 7 Thursday, April 7, 2016 2:28 PM
Contents
6 Post-Processing Images
7 Other Functions
7
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 8 Thursday, April 7, 2016 2:28 PM
Contents
10 Troubleshooting Remedies
8
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 9 Thursday, April 7, 2016 2:28 PM
1 Getting Started
1.1 Introduction
1.2 Notations
1.3 Workflow diagram for a general radiographic
examination
1.4 Part names of screens (Quick reference)
1.5 List operations
1.6 Detector connection information
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 10 Thursday, April 7, 2016 2:28 PM
1 Getting Started
1.1 Introduction
The CXDI Control Software NE (hereinafter referred to as the “Software”) is designed
to provide a steady and efficient workflow in the field of digital radiography when
linked to an RIS/HIS network. Up to four CXDI detectors including wireless detectors
can be connected simultaneously to the Software. Images captured using the CXDI
detector are automatically processed by the Software to achieve diagnostic image
quality. Furthermore, the Software offers advanced processing features with intuitive
operability.
1.2 Notations
Active tabs
1.2.2 Buttons
In this manual, the key top labels of buttons on the Software screen are written in
square brackets, such as [OK] and [Cancel], or the icon is shown.
NOTE: Six types of screen background color are prepared for the Software: Warm
Dark, Cool Dark, Monochrome Dark, Warm Light, Cool Light, and
Monochrome Light. Refer to 2.3.1 in the Setup Guide for details on setting
the screen background color.
In this manual, the Cool Light screen, that is the default setting, is used for
explanation.
10
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 11 Thursday, April 7, 2016 2:28 PM
1 Getting Started
Example
A
C
D
C A
A
1.2.5 References
References to this manual:
Associated information in this manual is indicated by the section number or by the
section number followed by the title, as seen below:
(See 3.1.5)
See step 2 in 3.1.5 for details on operation.
Proceed to “3.1.5 Configuring protocols.”
11
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 12 Thursday, April 7, 2016 2:28 PM
1 Getting Started
[EXAM] tab
[Log In]
Start Screen
Configuring protocols
[Edit Exam] (see 3.1.6)
[Start Exam]
[EXAM > Worklist] screen
[Manual]
[Worklist]
Screen transition
Screen buttons
[Start Exam]
[EXAM > Manual] screen
Screen tabs
12
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 13 Thursday, April 7, 2016 2:28 PM
1 Getting Started
[PAST] tab
[Back To List]
[PAST > View] screen [Recall Exam] [PAST > Past List] screen
[View]
Post-Processing Images Reviewing images on the
(see chapter 6) [Past List] [PAST > View] screen (see 5.2)
Printing images manually
(see 7.1.4) Local disk of the
[EXAM] image-capture
[PAST] computer
[Worklist]
[Examination]
Conducting
Examinations
(see chapter 4)
Captured
Exposure switch
images
[End Exam]
13
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 14 Thursday, April 7, 2016 2:28 PM
1 Getting Started
C H
14
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 15 Thursday, April 7, 2016 2:28 PM
1 Getting Started
A C
B
A: Study title (Accession number) (➞3.1.3, page 45) (➞“Study Information Display” in
2.4.1 in the Setup Guide)
B: Protocol list (➞Chapter 4, page 65)
C: Information button (➞3.1.3, page 45) (➞3.1.4, page 46)
D: Thumbnail
E: [Edit Exam] (➞3.1.6, page 50)
15
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 16 Thursday, April 7, 2016 2:28 PM
1 Getting Started
A
F
B G
C H
Body part search mode
16
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 17 Thursday, April 7, 2016 2:28 PM
1 Getting Started
L
I
J
K
Category search mode
17
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 18 Thursday, April 7, 2016 2:28 PM
1 Getting Started
C F
18
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 19 Thursday, April 7, 2016 2:28 PM
1 Getting Started
K
L
E M
N
O
P
19
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 20 Thursday, April 7, 2016 2:28 PM
1 Getting Started
D
B
20
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 21 Thursday, April 7, 2016 2:28 PM
1 Getting Started
E
B
F
21
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 22 Thursday, April 7, 2016 2:28 PM
1 Getting Started
1.4.8 Toolbar
22
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 23 Thursday, April 7, 2016 2:28 PM
1 Getting Started
B
C
D
E
F
G
H*
I
J
K
23
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 24 Thursday, April 7, 2016 2:28 PM
1 Getting Started
A F
B G
C
H
24
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 25 Thursday, April 7, 2016 2:28 PM
1 Getting Started
25
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 26 Thursday, April 7, 2016 2:28 PM
1 Getting Started
/ Column header
Study orders
NOTE: The items displayed in the column header can be selected on the system
setup screen. For details, refer to 2.4.1 in the Setup Guide.
26
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 27 Thursday, April 7, 2016 2:28 PM
1 Getting Started
Detector name
Power supply
Battery attached/
Connection Battery charge status
status Power supply
Battery attached/
Battery charge status Connection status
Charged
Attached Charging
External Wired
Not charging
Not attached
27
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 28 Thursday, April 7, 2016 2:28 PM
1 Getting Started
28
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 29 Thursday, April 7, 2016 2:28 PM
IMPORTANT: After the software has launched, do not press the power button on the
computer. It may cause a hardware malfunction or data corruption. This,
however, does not apply to the case where this manual instructs you to
press the power button. In that case, follow the instructions.
IMPORTANT: Do not press the Ctrl + Alt + Delete keys to restart or shut down the
computer. It may cause a data corruption.
[Log In]
Start screen
NOTE: Another start option that does not require user authentication is also
available. For details, consult your service engineer.
30
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 31 Thursday, April 7, 2016 2:28 PM
[Activate]
[OK]
NOTE: You can use this software or optional functions up to 100 times as a trial use,
when clicking [OK] without registration of license key. After using 100 times,
you cannot use the software or optional functions without registration.
31
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 32 Thursday, April 7, 2016 2:28 PM
Be sure to click [End Exam] in the examination screen before shutting down the
CAUTION software. Shutting down the software without ending the examination may
cause data loss.
[Logout]
NOTE: A dialog box related to image arrangement may appear. For details, see
NOTE in step 4 in 7.1.2.
32
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 33 Thursday, April 7, 2016 2:28 PM
[Shutdown]
[Restart Options]
33
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 34 Thursday, April 7, 2016 2:28 PM
IMPORTANT: Do not turn the image-capture computer off before it automatically shuts
down. It may cause a hardware malfunction or data corruption.
[CXDI Restart]
[PC Sleep]
[PC Hibernate]
[PC Restart]
[Shutdown]
34
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 35 Thursday, April 7, 2016 2:28 PM
NOTE: Do not enable sleep or hibernation mode using the power switch on the
image-capture computer.
NOTE: The image-capture computer resumes in recovery mode from sleep or
hibernation.
NOTE: [PC Sleep] and [PC Hibernate] can be hidden. For details, consult your
service engineer.
NOTE: When [Shutdown] in the recovery mode screen, as well as [Shutdown] in the
Shutdown Confirmation Dialog Box, is clicked, the Controller is exited.
NOTE: If an error persists after restarting the Software by clicking [CXDI Restart],
restart the image-capture computer.
35
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 36 Thursday, April 7, 2016 2:28 PM
• To resume connection
Click [Off Line].
[On Line]
[Off Line]
NOTE: When online mode is resumed, the images captured during offline mode are
transferred to the storage server (PACS), storage device, and printer. Also
the study information is sent to the HIS/RIS database.
NOTE: The Software can be configured to automatically detect network
disconnections. For details, consult your service engineer.
36
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 37 Thursday, April 7, 2016 2:28 PM
37
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 38 Thursday, April 7, 2016 2:28 PM
38
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 39 Thursday, April 7, 2016 2:28 PM
Before starting examinations, learn the following basic terms used in the Software.
Study: A request (order) for radiographic examination, consisting of patient information,
schedule, and examination details, issued by a referring physician
Exam: A group of studies for one patient
Protocol: Details of examination procedures, which include the following information:
Workspace, grid, image processing parameters, and DICOM attributes. In this
Software, a protocol is associated with a specific workspace, and that pair is shown
as a “protocol” button. In other words, according to the requirements of an
individual site, multiple buttons are created for one protocol, and all differ from each
other in the workspace. They are organized in protocol lists. Therefore, users can
select a pair including a protocol and workspace with one click of these buttons
(see 3.1.6).
Workspace: Details of a protocol that specify the detector used and the posture of the patient
Study orders
Click to create an
emergency study
order.
40
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 41 Thursday, April 7, 2016 2:28 PM
Free space
Used space
Click to shrink and expand the pane. Title boxes and their keyword
boxes
Enter any narrowing conditions in the Search For Study List pane.
Three of the following information titles can be selected for the narrowing
conditions: Patient ID, Name, Accession No., Referring Physician,
Requesting Physician, Comment, RP ID, RP Description, and SPS
Description. Select an option in each title box, and then enter any text in its
keyword box. Note that these three keywords are used as AND search
conditions.
Study Status: Specifies the search range from among All Exam, New
Exam, Pending Exam, and Restarted Exam.
41
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 42 Thursday, April 7, 2016 2:28 PM
NOTE: The information titles that can be selected for the narrowing conditions are
those currently displayed in the column header. For details on the column
header settings, refer to the Column Headers option under 2.4.1 in the Setup
Guide.
NOTE: Patient Name text box can be divided into five components for
accommodating family name, given name, middle name, prefix, and suffix
input. For details, refer to 2.4.1 in the Setup Guide.
NOTE: When [Refresh List] and [Search Conditions] are disabled, click [Off Line] to
resume connection with the HIS/RIS database.
[Search Conditions]
[Refresh List]
42
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 43 Thursday, April 7, 2016 2:28 PM
Enter any narrowing conditions in the Search Conditions dialog box, and then
click [Refresh List].
ID: Patient ID
Name: Patient name
ACC#: The accession number is automatically given to each study
order at the point of issue. This number can also serve as the
study title. Refer to “Study Information Display” in 2.4.1 in
the Setup Guide.
Requested Procedure ID:
Identifier that identifies the Requested Procedure in the
Imaging Service Request.
Range: Period refers to the period of time in which the study orders
are to be conducted. Relative refers to a period relative to
the current time in which the study orders are to be
conducted, specified by hours in the past/future. To clear the
Range condition, select All.
Modality: Select among DX, RF, CR, or XA. (Two or more can be
selected.)
Be sure to confirm that the entered information (patient name, ID number, date
CAUTION of birth, and sex) matches that of the patient. If the information is incorrect, the
resulting patient mix-up and a misdiagnosis may cause harm to the patient.
3 Start examination.
Proceed to “4 Conducting Examinations.”
43
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 44 Thursday, April 7, 2016 2:28 PM
Study orders
[Delete]
44
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 45 Thursday, April 7, 2016 2:28 PM
Date of birth
Pregnancy status
(none): not pregnant
: pregnant
: possibly
pregnant
Study List
Click to modify study
information (if necessary,
see 3.1.4).
Study order
45
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 46 Thursday, April 7, 2016 2:28 PM
Click to hide or
show the protocols
in the study order.
[Edit]
Referring Physician/
Study Description/
Reading Physician/
Institution Name/
Study Info 1 (Custom
Field Info) text box
[OK]
46
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 47 Thursday, April 7, 2016 2:28 PM
NOTE: Grayed-out information in the Edit Study Information dialog box cannot be
modified. However, ACC# can be configured to be modified. For details,
consult your service engineer.
NOTE: Other study information items, such as Requesting Physician and Study
Instance UID, can be configured to be shown in this dialog box and modified
in this step. For details, refer to 2.4.1 in the Setup Guide.
NOTE: Frequently used descriptions for some study information items can be
preset. For details, refer to 2.4.2 in the Setup Guide.
3 End modification.
Click [OK].
47
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 48 Thursday, April 7, 2016 2:28 PM
Thumbnail
Protocol
48
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 49 Thursday, April 7, 2016 2:28 PM
Detector name
Detector connection information
Patient posture (detector setup)
: Stand
: Table
: Cassette
: Universal
Available workspaces
Workspace name
[Close]
NOTE: A battery indicator and a signal strength indicator are shown on the
workspace for wireless detectors. For the wireless detector, refer to the
Digital Radiography CXDI series User’s Manual.
49
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 50 Thursday, April 7, 2016 2:28 PM
[Edit Exam]
Click to navigate
Click to shrink and the pages on the
expand the protocol list.
list.
Protocols included
in the selected
study order
Frequently-used
protocols list
Body part selector
Recently-used
protocols list
Examination
history list
[∧]/[∨]
[Delete]
50
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 51 Thursday, April 7, 2016 2:28 PM
2 Configure protocols.
NOTE: In body part search mode, up to three types of protocol lists (Frequently-
used protocols, Recently-used protocols, and Examination history) are
available for protocol selection. For details on list operation, see 3.2, and on
list configuration, refer to Protocol Settings in 2.4.1 in the Setup Guide. In
addition, category search mode is also available (see 3.2.2).
NOTE: The Recently-used protocols list and Frequently-used protocols list cannot
be shrunk at the same time.
• To delete a protocol
Click a target protocol in the study information pane, and then click [Delete].
51
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 52 Thursday, April 7, 2016 2:28 PM
[Unspecified]
Click to clear all
conditions
(including the
Search options
conditions) in the
body part selector.
52
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 53 Thursday, April 7, 2016 2:28 PM
[Search Options]
[Search]
53
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 54 Thursday, April 7, 2016 2:28 PM
Select by
Workspace dialog
box
54
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 55 Thursday, April 7, 2016 2:28 PM
NOTE: PrePack protocols (refer to 3.4.3 in the Setup Guide) can be selected only in
category search mode.
[Search by Category]
[∧]/[∨]
Click to switch to
body part search
mode.
55
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 56 Thursday, April 7, 2016 2:28 PM
[Category List]
Category list
56
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 57 Thursday, April 7, 2016 2:28 PM
Patient list
[Delete]
The [EXAM > Manual] screen appears when the user is logged in to the Software that
does not have data exchange and updating services with the HIS/RIS database.
Be sure to confirm that the entered information (patient name, ID number, date
CAUTION of birth, and sex) matches that of the patient. If the information is incorrect, the
resulting patient mix-up and a misdiagnosis may cause harm to the patient.
NOTE: Multiple patient information can be selected. For details, refer to step 2 in
5.3. However, is not displayed in the Manual tab.
NOTE: The Recent Patient List pane can be hidden. For details, consult your service
engineer.
57
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 58 Thursday, April 7, 2016 2:28 PM
Patient information
[Set ➞]
Enter any of the following information in the text boxes and select an option for
the Sex of the patient.
Name: Patient name
DoB/Age: Birth date of a patient (Age will be automatically calculated
based on this entry.)/Age of a patient
*ID: Patient ID
NOTE: Insufficient entry of ID and items marked with an “*” in step 1 disables
[Set ➞] (refer to 2.4.1 in the Setup Guide for details on the setting of
essential input items).
58
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 59 Thursday, April 7, 2016 2:28 PM
[Edit]
[Delete]
[Add Study]
59
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 60 Thursday, April 7, 2016 2:28 PM
NOTE: Frequently used descriptions for some study information items can be
preset. For details, refer to 2.4.2 in the Setup Guide.
6 Configure protocols.
See step 2 in 3.1.6 for details on operation.
7 Start examination.
Proceed to “4 Conducting Examinations.”
60
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 61 Thursday, April 7, 2016 2:28 PM
[Edit]
[OK]
Be sure to confirm that the entered information (patient name, ID number, date
CAUTION of birth, and sex) matches that of the patient. If the information is incorrect, the
resulting patient mix-up and a misdiagnosis may cause harm to the patient.
61
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 62 Thursday, April 7, 2016 2:28 PM
62
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 63 Thursday, April 7, 2016 2:28 PM
[Emergency]
Temporary patient
name
Patient ID
63
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 64 Thursday, April 7, 2016 2:28 PM
64
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 65 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
4 Conducting Examinations
Stitch type
Captures two to four radiographic images and combines
them to form a single long length image
IMPORTANT: Even if an image transfer has been interrupted when using a wireless
detector, it can resume by switching to the wired connection. For the wired
connection, use the Wiring Unit (sold separately). For details, consult your
service engineer.
IMPORTANT: Be sure to select a protocol for which the Prolonged Exposure option has
been selected in cases where an X-ray irradiation lasts more than one
second. Otherwise, it may not be possible to acquire normal images. For
details about the settings for the Prolonged Exposure option, refer to 3.4.1
in the Setup Guide.
NOTE: The appropriate layouts for the Mass Examination (Workflow tool) and the
appropriate display layouts for the XGA standard displays (Simple tool) are
available. For details, consult your service engineer.
NOTE: The Software can be configured to display a warning dialog box when the
REX value is out of the specified range at radiographic examinations. For
details, refer to “REX Check at Examination” in 2.3.1 in the Setup Guide.
NOTE: The Software conducts image feature analysis to get images suitable for
examination. The feature may not be analyzed properly according to the
anatomical part and direction of patient and the direction of the detector.
Refer to the image feature analysis method list and the information to use in
the suitable layout. Refer to chapter 5.2.2 in Appendix in the Setup Guide.
Be sure to confirm that the entered information (patient name, ID number, date
CAUTION of birth, and sex) matches that of the patient. If the information is incorrect, the
resulting patient mix-up and a misdiagnosis may cause harm to the patient.
66
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 67 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
1 Start an examination.
Click [Start Exam].
NOTE: If [Start Exam] does not work even if a study is selected, click [Edit Exam]
and add the protocol to the study order.
[Workspace Sort]/
[Study Sort]
[Unselect]
Click to configure
protocols
(see 3.1.6).
Preview annotations
67
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 68 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
NOTE: A battery indicator and a signal strength indicator are shown on the system
status indicator during an examination using a wireless detector. For the
wireless detector, refer to the Digital Radiography CXDI series User’s
Manual.
Example of the system status indicator while using a wireless detector
NOTE: While image data and radiation dose reports are being transferred in the
background, the data transfer indicator or blinks (see 1.4.1). If [Start
Exam] is clicked to start an examination during transfer, the transfer will be
interrupted. If the transfer has priority over the examination, wait until the
or indicator goes off, and then click [Start Exam].
NOTE: If [Start Exam] is clicked while the items specified as essential inputs on the
study information are left blank, the dialog box is displayed. Enter the
corresponding items, and then click [OK].
68
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 69 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
: Stand
: Table
: Cassette*
: Universal
Ready thumbnail
Workspace name
Protocol name
Ready status protocol
(highlighted)
* When the patient posture (detector setup) is Cassette, the user can change the icon color. For
details, consult your service engineer.
Instruct the patient not to move during the examination. If necessary, help the
CAUTION patient to maintain the proper posture. If the patient moves during the
examination, it may fail to obtain the appropriate images.
69
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 70 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
4 Perform exposure.
Press and release the exposure switch of the X-ray generator. The
indicator changes to , and then in the system status
indicator.
Exposure After processing is complete, a preview image appears in the image view pane,
switch and then the next available protocol is automatically highlighted.
Preview thumbnail
Complete protocol
Series number
[End Exam]
NOTE: The series number that indicates the examination sequence can be changed
to a number that indicates the protocol order in the study information pane in
descending order. For details, consult your service engineer.
Fast-preview
Depending on the sensor that is using, the preview image is displayed in
approximately one second after an exposure. The preview image may be displayed
grainily for approximately two seconds.
NOTE: The fast-preview function can be set to unavailable. For details, consult your
service engineer.
NOTE: The fast-preview function is not available for Non Generator Connection
imaging.
70
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 71 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
71
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 72 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
Corrupted image
icon
• To perform post-processing
Proceed to “6 Post-Processing Images.”
72
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 73 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
NOTE: Once images are transferred to PACS by clicking [Send to Storage], the
corresponding images stored in the image-capture computer cannot be
modified or post-processed. However, it will be enabled by patient
information rebinding.
NOTE: For transferring to storage, the irradiation direction can be detected with
marking the specified part of image. For details, consult your service engineer.
NOTE: To reduce power consumption, the Ready status, shown in the system
status indicator, can be manually canceled. Click [Unselect].
NOTE: While the data is being transferred, the is blinking. After it stops blinking,
detach your storage media. If storage media are detached during data
transfer, the data cannot be saved.
NOTE: Users can finish the examination even if incomplete protocols remain.
73
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 74 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
Incomplete protocol
(highlighted)
Protocol list
[OK]
74
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 75 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
[OK]
75
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 76 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
Complete protocol
(highlighted)
Protocol list
Protocol name
before replacement
box
Series/Image
Information option
(see 4.9)
76
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 77 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
NOTE: Only replaceable protocols (whose workspaces are identical to those of the
original protocol) are displayed in the protocol list. For details on searching
for a protocol, see 3.1.6 and 3.2.
[OK]
77
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 78 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
[EXAM > Examination] Incomplete protocol used for Preview thumbnail after
screen reprocessing reprocessing
78
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 79 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
Rejection Reason
Rejection Reason
[Retake] Duplication of the original
List
[Reject]/ protocol
[Restore]
79
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 80 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
NOTE: After the image of the rejected original protocol is reprocessed using another
protocol (see 4.3), the duplicated protocol remains even if the original image
is restored.
NOTE: After the duplicated protocol is replaced with another protocol (see 4.2), the
duplicated protocol remains even if the original image is restored.
80
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 81 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
NOTE: Examinations listed in red indicate that they were abnormally terminated
during processes. Indications and operations for the listed items are
identical with those for examinations that are suspended by clicking
[Suspend Exam].
NOTE: After rebinding, incomplete protocols in the study order will be deleted.
NOTE: Once images are transferred to PACS by clicking [Send to Storage], the
corresponding images stored in the image-capture computer cannot be
modified or post-processed. However, it will be enabled by patient
information rebinding.
81
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 82 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
NOTE: When the study order acquired from the worklist or the emergency study
NOTE: If two or more study orders are selected using the multi-selection feature,
click unwanted study orders to clear them and leave only one target study
order highlighted.
(rebind)
Study Status
Suspended study
82
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 83 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
NOTE: In offline mode, is disabled. In such cases, click [Off Line] to enable online
mode.
[Search
Conditions]
[Refresh List]
83
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 84 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
[OK]
84
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 85 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
[Refresh List],
[Search Conditions]
[OK]
NOTE: When the data exchange and updating services with the HIS/RIS database
are not provided with the Software, [EXAM > Manual] screen appears
instead of the dialog box above.
85
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 86 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
Be sure to confirm that the entered information (patient name, ID number, date
CAUTION of birth, and sex) matches that of the patient. If the information is incorrect, the
resulting patient mix-up and a misdiagnosis may cause harm to the patient.
[OK]
86
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 87 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
Positioning
(Custom Field Info) box
87
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 88 Thursday, April 7, 2016 2:28 PM
4 Conducting Examinations
NOTE: Frequently used descriptions for some series information items can be
preset. For details, refer to 2.4.2 in the Setup Guide.
NOTE: Frequently used descriptions for images can be preset and selected from the
drop-down list for Custom Field Info. For details, refer to 2.4.5 in the Setup
Guide.
88
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 89 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
5 Reviewing Images
(single view)
The selected view
mode button is Click to switch the
highlighted. display annotations
(preview annotations/
storage annotations/
no annotation).
Preview thumbnail
[Unselect]
Preview
annotations
(refer to 2.5.1 in the
Setup Guide)
90
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 91 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
[Info]
Storage
annotations (refer
to 2.5.5 in the
Setup Guide)
(multi view)
The selected view
mode button is
highlighted.
• To perform post-processing
Proceed to “6 Post-Processing Images.”
91
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 92 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
NOTE: The directory options are displayed when external storage is specified on the
Output to media tab (refer to 2.6.5 in the Setup Guide). If the directory
options do not appear, it means that all of the listed studies are stored in the
image-capture computer.
Directory options
[Restart Exam]
[Recall Exam]
[PAST > Past List] screen
(example of the Local DB selected for the directory option)
92
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 93 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
[Refresh]
NOTE: Only when All Exam is selected for the Target object option during Local DB
search (see 5.3), [Refresh] appears. In such cases, click [Refresh] to start
search after entering necessary keywords in the text boxes.
NOTE: The information titles that can be selected for the narrowing conditions are
those currently displayed in the column header. For details on the column
header settings, refer to the Column Headers option under 2.4.1 in the Setup
Guide.
Study Date: Specify both or either the start and end date of the target
period. This option is enabled only when the Recent week,
Recent month, or All Exam is selected for the Target Object
option.
93
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 94 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
[Restart Exam]
Click to save
changes for the
current image.
[Back To List]
• To perform post-processing
Proceed to “6 Post-Processing Images.”
94
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 95 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
[View] tab
NOTE: The image of a study listed in the [PAST > Past List] screen with
indicated in the attribute icon column is being reviewed in the [PAST > View]
screen (see NOTE in 5.2.3).
2 Move the study to the Study List in the [EXAM > Worklist]
screen.
Click [Restart Exam] and click [OK] in the confirmation dialog box that appears.
The study disappears from the Study List in the [PAST > Past List] screen.
[OK]
NOTE: [Restart Exam] is also available in the [PAST > View] screen.
95
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 96 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
3 Start examination.
In the [EXAM > Worklist] screen, select the restarted study with the Restarted
indication in the Study Status column, add new protocols (see 3.1.6), and then
click [Start Exam]. For details on conducting examinations, see chapter 4.
Study Status
Selected restarted
study
[Start Exam]
NOTE: It is available to transfer only the images captured with the newly added
protocols and the radiation dose report on them to destination storages and
printers.
NOTE: Only panning (see 6.1.3) and zoom in/out (see 6.1.4) operations can be
enabled for images that are carried over from the restarted study.
96
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 97 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
[Back To List]
• To perform post-processing
Proceed to “6 Post-Processing Images.”
97
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 98 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
NOTE: When the copied study is listed in the [PAST > Past List] screen, is
displayed in its attribute icon column. Furthermore, when the copied study is
protected (see 3.1.4), changes to .
Attribute icon
column
Copied study
98
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 99 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
Protocol
ACC# for the
information for each
selected image
image
appears.
Current/total
studies
Current/total
images
/ ,
/
[PAST > View] screen [Back To List]
NOTE: Age in the patient information pane remains blank during review.
99
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 100 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
5.3 Deleting studies from the [PAST > Past List] screen
This Software is configured to automatically delete past study data that is stored in the
image-capture computer in chronological order from old to recent. To prevent useful
reference study data from being deleted, check the HDD free space icon (see 3.1.1)
and delete unnecessary studies before the available HDD space decreases too much.
NOTE: To prevent an image data from being deleted by mistake, can be hidden.
For details, consult your service engineer.
Directory options
NOTE: When the Disk Storage option is selected, the Software automatically enters
study deletion mode. Therefore, users do not need to click in step 2.
100
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 101 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
Selected study
(highlighted)
Reduction target
[Select All Studies] text box
[Delete]
NOTE: Studies with or displayed in the attribute icon column (no title)
cannot be deleted as the Protect Image option is selected. For details on the
Protect Image option, see step 2 in 3.1.4. Also, refer to 2.4.1 in the Setup
Guide for the Column Headers setting.
101
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 102 Thursday, April 7, 2016 2:28 PM
5 Reviewing Images
NOTE: The reduction target text box and [Delete] are available only when the user
who has the System Settings privilege logs in to the Software. For details,
consult your service engineer.
102
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 103 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
6 Post-Processing Images
[Reset]
104
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 105 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[Tool 2] [Tool 1]
NOTE: Refer to 2.4.3 in the Setup Guide for details on toolbar customization.
105
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 106 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: The Software can be configured to enable you to rotate or flip images on the
X-ray generator as well. For details, consult your service engineer.
106
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 107 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: The panning operation does not affect the images that are transferred to a
storage server, printer, or other destination directories.
See 6.1.4.
[OK]
• To finish panning
Click [OK].
107
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 108 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: The change in magnification does not affect the size of the images that are
transferred to a storage server, printer, or other destination directories.
NOTE: In high throughput mode, high resolution image display is available using
[HQ]. For details, consult your service engineer.
108
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 109 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
109
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 110 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
Laterality marker
[L]/[R]
110
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 111 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[ROI]
[Delete]
[Select]
[OK]
Click to exit
processing mode.
Handles
Click [Select], point any handle to change the pointer to (resize pointer),
and then drag the handle.
111
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 112 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: Preset cropping area is specified in the Common Cropping Area options in
the System Settings tab (refer to 2.3.1 in the Setup Guide), or the Cropping
Area options in the Protocol Workspace Settings 1 screen (refer to 3.4.1 in
the Setup Guide).
NOTE: The line width of a frame specifying the cropping area can be changed (refer
to 2.4.1 in the Setup Guide).
112
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 113 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
Cropping area
preset button
,
[Reset]
[Select]
Handles
Arrow buttons
, ,
[Close]
[Renewed]
Point any handle to change the pointer to (resize pointer), and then drag the
handle.
113
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 114 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: Laterality markers can be synchronized with the movement of the crop
frame. For details, refer to 2.5.3 in the Setup Guide.
NOTE: You can crop areas that are outside the range of the imaging area by moving
the crop frame. In such cases, laterality markers and free annotations can be
placed outside the range of the imaging area.
NOTE: When Custom Area is selected for the Cropping Area option in Protocol
Workspace Settings 1 (refer to 3.4.1 in the Setup Guide), when is clicked,
the setting for the Alignment option will be applied to the cropping area.
NOTE: Custom area buttons other than cropping area preset button can be preset.
For details, refer to 2.3.3 in the Setup Guide.
NOTE: When users compare images on the monitor display, the following should be
observed: As the size of Effective Area/Detected Irradiated Field varies
depending on the detector, the magnification may change among images
that are captured using different detectors when images are displayed after
clicking .
NOTE: For partial images captured by Long-length imaging with multiple exposures
(see chapter 8), the cropping area preset button, custom area buttons and,
are hidden. For stitched images, the cropping area preset button,
custom area buttons, and are hidden.
114
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 115 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: This icon is not shown depending on the mask selected in Peripheral Mask
control.
NOTE: This icon is not shown depending on the mask selected in Peripheral Mask
control.
NOTE: Unlike its gray appearance on the screen, the masked area appears black on
the images transferred to a storage server or printer.
115
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 116 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[Mask]
[Preview]
, ,
, ,
Arrow buttons
Boundary points
116
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 117 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: A mask frame appears in the center of the cropping area when is
selected first. The size of the mask frame is the same as the circle inscribed
in the cropping area.
Mask frame
NOTE: A mask frame appears in the same position and size as the irradiated field
when is selected first.
NOTE: Additionally, to specify a rectangular mask frame, click a point on the image
view pane, and then double-click the diagonally opposite point while
selecting .
117
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 118 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
• To cancel changes and resume to the frame when the mask edit
was started
Click [Reset].
118
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 119 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: Changes made to the on/off setting of the masking effect do not affect the
state of the Mask check box (see 6.2.10).
119
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 120 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
1 Copy images.
Click . When the warning dialog box is displayed, click [OK] to start
copying.
The icon is displayed on the thumbnail of the copied image.
120
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 121 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[Save as Default]
[Image Processing]
121
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 122 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: For further information on auto brightness adjustment applied to bright and
dark regions, refer to 5.2.3 “Dynamic Range Analysis” and 5.3.2 “Dynamic
Range Adjustment” under Appendix in the Setup Guide.
NOTE: These adjustments are not available when the REX option is selected for
Brightness Adjustment mode (see 6.2.4).
Histogram
This indicates the pixel value distribution of the raw data generated by the detector’s
sensor, and is used as a rough indicator of the X-ray dose. The x-axis of this histogram
is logarithmic. The more the histogram leans toward the right, the higher the dose. The
adjustment of processing controls will not affect the histogram.
Histogram
[EXAM >
LUT curve
Examination]
screen
122
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 123 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
Category selector
[EXAM >
Examination] screen
Anatomical Part selector
Direction selector
123
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 124 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[EXAM >
Examination] screen Curve Shape selector
Brightness
Adjustment mode
selector REX control
Base Brightness
control
Contrast control
LUT control (Auto mode for LUT control (REX mode for
radiographic images) radiographic images)
124
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 125 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: These adjustments are not available when the REX option is selected for
Brightness Adjustment mode.
NOTE: When adjusting images where negative/positive inversion is applied, the
moving directions of the Base Brightness control are reversed. The same
goes for step 5.
NOTE: The value of Base Brightness is used as a reference value for the Brightness
control.
NOTE: The value of Base Contrast is used as a reference value for the Contrast
control.
125
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 126 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: Fine-adjust the brightness in a range of -10 to +10 relative to the Base
Brightness value.
The supported adjustment range depends on the Base Brightness value. In
some cases, the range may be narrower than -10 to +10.
NOTE: Fine-adjust the contrast in a range of -10 to +10 relative to the Base Contrast
value.
The supported adjustment range depends on the Base Contrast value. In
some cases, the range may be narrower than -10 to +10.
NOTE: By using a mouse, the brightness and contrast can be adjusted more finely
than normal. For details, consult your service engineer.
126
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 127 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
Enhancement control
127
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 128 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[EXAM >
Examination] screen
Whole Region control
Dark Region control
Bright Region control
128
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 129 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
129
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 130 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[EXAM >
Examination] screen
Effect control
130
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 131 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[EXAM >
Examination] screen
Grid Name selector
NOTE: When using a wireless detector, the grid lines are not shown on the image
displayed just after captured even if the grid line suppression is disabled. To
show them, redisplay the image by performing image processing or
displaying a past image.
131
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 132 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[EXAM >
Examination] screen
Effect control
NOTE: When the Effect check box is selected, the sharpness of the preview image
is also adjusted.
132
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 133 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[EXAM >
Examination] screen
Mask option
NOTE: Even if this setting is changed after exposure, the change does not affect the
captured image.
133
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 134 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
Handles
Measured distance
[Measurement]
NOTE: The measurement unit and the ruler shown at the bottom of the image view
pane can be selected on the system setup screen. For details, refer to 2.4.1
and step 5 under 2.6.2 in the Setup Guide.
134
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 135 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: Once distance mode is started, operations other than step 2 below are
disabled. Be sure to complete operations through step 2.
[Correct]
Measurement panel
To correct the measured values, enter or modify the values in the SOD and SID
boxes, and then click [Correct].
NOTE: The measured value in height difference mode (see 6.3.4) will also be
corrected.
135
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 136 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
Handles
Measured angles
[Measurement]
NOTE: The measurement unit and the ruler shown at the bottom of the image view
pane can be selected on the system setup screen. For details, refer to 2.4.1
and step 5 under 2.6.2 in the Setup Guide.
NOTE: Once angle mode is started, operations other than step 2 below are
disabled. Be sure to complete operations through step 2.
136
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 137 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
Handles
Measured Cobb
angle
137
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 138 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: The measurement unit and the ruler shown at the bottom of the image view
pane can be selected on the system setup screen. For details, refer to 2.4.1
and step 5 under 2.6.2 in the Setup Guide.
NOTE: Once Cobb angle mode is started, operations other than step 2 below are
disabled. Be sure to complete operations through step 2.
138
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 139 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
Handles
Measured height
difference
[Measurement]
NOTE: The measurement unit and the ruler shown at the bottom of the image view
pane can be selected on the system setup screen. For details, refer to 2.4.1
and step 5 under 2.6.2 in the Setup Guide.
NOTE: Once height difference mode is started, operations other than step 2 below
are disabled. Be sure to complete operations through step 2.
139
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 140 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
4)
140
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 141 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
Custom annotation
list
[Add/Edit]
[Annotation]
Text box
141
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 142 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
NOTE: Custom annotations can also be created on the system setup screen. For
details, refer to 2.5.2 in the Setup Guide.
NOTE: A maximum of 100 annotations can be registered in the custom annotation
list. If the target annotation cannot be found in the list, click the scroll arrows
to navigate the list.
142
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 143 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
• To select/deselect an annotation
Click the target annotation on the image view pane. An annotation highlighted
with dashed orange border is selected and active. Before performing other
operation, be sure to deselect an active annotation by dragging it to the same
position. Clicking an inactive annotation makes it active.
Selected annotation
[Enlarge]
[Delete]
[Annotation]
[Reduce]
• To delete an annotation
Click an annotation to select, and then click [Delete].
143
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 144 Thursday, April 7, 2016 2:28 PM
6 Post-Processing Images
[EXAM >
Examination] screen
Effect control
144
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 145 Thursday, April 7, 2016 2:28 PM
7 Other Functions
7 Other Functions
NOTE: For printing image and transferring to storage, the irradiation direction can
be detected with marking the specified part of image. For details, consult
your service engineer.
NOTE: Images cannot be printed from the [PAST > Past List] screen.
(Output setting)
Printer options
[Save to System]
Storage options
Click to cancel the
settings.
[Apply]
146
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 147 Thursday, April 7, 2016 2:28 PM
7 Other Functions
(Output setting)
Printer options*
[Send Image]*
147
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 148 Thursday, April 7, 2016 2:28 PM
7 Other Functions
NOTE: The changes are cleared after the study is ended (and its image printing/
transfer is complete) or suspended, and then the default settings resume.
148
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 149 Thursday, April 7, 2016 2:28 PM
7 Other Functions
NOTE: For details on displaying the image arrangement screen and other printer
settings, refer to step 5 and 6 under 2.6.2 in the Setup Guide.
NOTE: If the Film Size and Direction option in the Protocol Workspace Settings 1
screen has been left blank for the protocol used for capturing the image
currently selected for printing, the automatic arrangement feature does not
function. In such cases, click [OK] in the warning dialog box that appears,
and then arrange images manually (see 7.1.3).
[End Exam]
Study tabs
(Accession number)
Arrangement
preview pane
Click to cancel
and exit image
arrangement
mode.
[OK]
[Auto Layout]
[Add Page]
[Insert Page]
Click to delete the current page.
[Delete All Page]
149
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 150 Thursday, April 7, 2016 2:28 PM
7 Other Functions
1 Finish an examination.
Click [End Exam] in the [EXAM > Examination] screen.
All images to be printed in the current Exam are arranged in the arrangement
preview pane.
NOTE: The essential settings for the automatic arrangement feature such as the
Film Size and Direction, Fit option, and other options, are configured in the
Printer tab. For details on operations, refer to step 6 under 2.6.2 in the Setup
Guide.
NOTE: Images captured with detectors that differ from each other in their pixel
pitches will not be arranged on the same film.
NOTE: Automatic arrangement is performed based on the conducted sequence of
the studies. Although blank segments may remain on some pages
depending on the size and shape of images, this is the best result and does
not indicate a malfunction of the Software.
150
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 151 Thursday, April 7, 2016 2:28 PM
7 Other Functions
NOTE: Settings for this feature can be preset for each protocol.
Refer to “Protocol Workspace Settings 1” in step 2 under 3.4.1 in the Setup
Guide for details on the option settings.
Arrow buttons
Printable area
[OK]
NOTE: The ruler of which magnification is corrected (see 6.3.1) can be embedded at
the bottom of an image. For details on the settings, refer to step 5 under
2.6.2 in the Setup Guide.
NOTE: Even if the size of the ruler is corrected based on the SOD and SID values,
minor errors may occur caused by the margins and other output settings of
the printer.
151
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 152 Thursday, April 7, 2016 2:28 PM
7 Other Functions
NOTE: If a dialog box indicating “Arrange all images on the Image List.” appeared,
drag all images to the arrangement preview pane, and click [OK].
For details on automatic image selection for printing (the Select exposure
mode to be printed automatically option), refer to step 5 under 2.6.2 in the
Setup Guide.
Layout templates
152
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 153 Thursday, April 7, 2016 2:28 PM
7 Other Functions
1 Finish an examination.
Click [Edit Exam] in the [EXAM > Examination] screen.
All images to be printed in the current Exam are arranged in the arrangement
preview pane.
[Add Page]
Template pallet
[Next]
153
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 154 Thursday, April 7, 2016 2:28 PM
7 Other Functions
Size option
NOTE: When all images on the Image List do not blink, add a new page before
dragging the images.
6 Continue arrangement.
Repeat step 3 through step 5.
154
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 155 Thursday, April 7, 2016 2:28 PM
7 Other Functions
3 Start printing.
Click [Send Image] to print the current image or click [Send Exam] to print the
images in the current Exam.
155
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 156 Thursday, April 7, 2016 2:28 PM
7 Other Functions
NOTE: When the storage server (PACS) for rejected images is selected as the
destination, images for diagnosis and rejected images in a study can
automatically be distinguished and transferred to each storage server (refer
to 2.6.1 in the Setup Guide).
156
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 157 Thursday, April 7, 2016 2:28 PM
7 Other Functions
4 Transfer images.
In the [PAST > Past List] screen:
Click [Send Exam] to transfer the images in the currently
selected Exam.
In the [PAST > View] screen:
Click [Send Image] to transfer the current image or click
[Send Exam] to transfer the images in the current Exam.
NOTE: The radiation dose report can be transferred to the preset servers using only
[Send Exam].
NOTE: Preset the CD (DVD) drive for recording (refer to 2.6.6 in the Setup Guide).
NOTE: The title of the optical disc is the date and time that the recording is
performed (yyyymmddhhmmss) and it cannot be changed. For example,
when you create the disc at 9: 43:8 on August 6th, 2012, the title is
“20120806094308”.
NOTE: Recording additional data to the optical disc and deleting the recorded data
from the disc is not available.
157
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 158 Thursday, April 7, 2016 2:28 PM
7 Other Functions
CD (DVD)
drive name
CD (DVD)
drive status
Writing Speed
158
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 159 Thursday, April 7, 2016 2:28 PM
7 Other Functions
NOTE: A radiation dose report is sent by the unit of examination. It cannot be sent
using [Send Image].
NOTE: The dose information of rejected images is also sent.
159
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 160 Thursday, April 7, 2016 2:28 PM
7 Other Functions
160
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 161 Thursday, April 7, 2016 2:28 PM
NOTE: This section explains an examination method that uses stitching markers.
For details, refer to 2.3.1 in the Setup Guide.
162
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 163 Thursday, April 7, 2016 2:28 PM
[Edit Exam]
[Start Exam]
[EXAM > Worklist] screen
Stitch type
protocols
[Stitch]
163
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 164 Thursday, April 7, 2016 2:28 PM
System status
indicator
Ready thumbnail
(highlighted in green)
NOTE: If the protocol does not automatically enter ready mode, click to select the
target protocol.
NOTE: Depending on the X-ray generator, the number of partial images less than
the preset value can be specified from the X-ray generator side.
164
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 165 Thursday, April 7, 2016 2:28 PM
5 Perform exposure.
Press and release the exposure switch on the X-ray generator. The
indicator changes to , and then to in the system status
indicator.
After processing is complete, a preview image appears in the image view pane,
Exposure and the next ready thumbnail is automatically highlighted.
switch
NOTE: The number of partial images required for a long-length image is preset for
each stitch type protocol within a range of two to four.
[Retake]
[Suspend Exam]
[Stitch]
165
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 166 Thursday, April 7, 2016 2:28 PM
NOTE: If [Stitch] is clicked before capturing all the necessary partial images, further
image capturing is not available.
NOTE: The partial images of the rejected protocol are not carried over to the
duplicated protocol.
Rejected protocol
Preview thumbnail
(first image) of the
rejected protocol
NOTE: Even if the rejected protocol is restored, the partial images of the rejected
protocol are not carried over to the restored protocol.
• To suspend an examination
After the number of captured partial images reaches the number set for the
stitch type protocol, click [Suspend Exam] (see 4.6). If there are captured
images which have not been stitched yet, see “To stitch partial images for long-
length imaging after the examination is suspended”.
NOTE: Further image capturing is not available using the suspended protocol.
NOTE: Depending on the X-ray generator, if two or more images have already been
captured, the examination can be ended from the X-ray generator.
166
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 167 Thursday, April 7, 2016 2:28 PM
Aligned images.
The area shown in
the image view
pane is indicated
by the orange
borders.
[OK]
Stitch Screen
167
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 168 Thursday, April 7, 2016 2:28 PM
168
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 169 Thursday, April 7, 2016 2:28 PM
(Maximum
(Fit in pane) magnification)
(Zoom out)
[OK]
Stitch Screen
169
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 170 Thursday, April 7, 2016 2:28 PM
(Cropping)
(Free
rotation)
(Rotate
Selected partial clockwise)
image
(Rotate
counterclockwise)
Arrow buttons
[Reset Position]
170
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 171 Thursday, April 7, 2016 2:28 PM
Handles
[Select]
[OK]
Click to exit
cropping mode
without saving any
changes.
171
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 172 Thursday, April 7, 2016 2:28 PM
Outline for
positioning
Pivot point
Edge Enhancement
control
[OK]
NOTE: can be hidden in the Stitch Screen (refer to 3.4.1 in the Setup Guide).
NOTE: The Edge Enhancement effect on the Stitch Screen is temporal and it is not
saved.
172
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 173 Thursday, April 7, 2016 2:28 PM
[Recall Exam]
[Stitch]
Click to save
changes for the
long-length image.
Click to return to
the [PAST > Past
List] screen.
173
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 174 Thursday, April 7, 2016 2:28 PM
[OK]
174
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 175 Thursday, April 7, 2016 2:28 PM
To use Non Generator Connection imaging, the settings require to be enabled. For the
settings, consult your service engineer.
IMPORTANT: Do not press the exposure switch when [Ready] is not displayed in the
system status indicator. Pressing the exposure switch generates X-ray
regardless of the displays on the software.
IMPORTANT: The users are not allowed to calibrate the detector when the Non Generator
Connection imaging setting is available. Consult your service engineer if
calibrating the detector.
IMPORTANT: When modifying or adding conditions of Non Generator Connection imaging,
confirm it before performing exposure to a patient that Non Generator
Connection imaging operates surely. Depending on the imaging conditions,
such as radiation dose, exposure time, irradiated field or target body
positioning, it may occur that image cannot be acquired even if exposure is
performed or that artifacts are generated.
Refer to chapter 8 in Appendix in the Setup Guide.
IMPORTANT: When the image-capture computer uses its built-in wireless LAN function for
wireless connection with both the detector and the hospital network, the
connection with the hospital network automatically goes offline during
examinations. Do not use the built-in wireless LAN function if it is necessary
to be in constant contact with the hospital network. Note that the following
occur while the system is offline.
• When examinations are started continuously or when an examination
takes a long time, data transfer is interrupted because data transfer to the
hospital network, such as output to storages, printers and report servers,
is interrupted.
• The storage commitment function cannot be used.
• Study orders cannot be acquired and MPPS reports cannot be
transferred.
NOTE: Read this manual and "Setup Guide" before reading this chapter. Also, when
the Non Generator Connection imaging setting is set to available, some
screens may be partly different.
176
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 177 Thursday, April 7, 2016 2:28 PM
1 Select a protocol.
Select a [Radiography] type protocol.
2 Perform exposure.
Confirm that [Ready] is displayed in the system status indicator, and then press
the exposure switch.
The detector detects X-ray and shows an image.
177
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 178 Thursday, April 7, 2016 2:28 PM
The announcement mark appears and continues to blink 1 minute before entering
sleep mode. The mark changes 10 seconds before entering sleep mode.
NOTE: Ready Indicator (an optional product) can notify the users of both 1 minute
and 10 seconds before entering sleep mode by its sound and lamp. For
details, refer to the Digital Radiography CXDI series User's Manual.
178
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 179 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
10 Troubleshooting Remedies
NOTE: Usually, appears accompanied with an error dialog box, except when
an examination is in progress. If only appears, click it to show the
error dialog box after the examination is ended.
Error code
Error type
Description
Click to show
additional
information.
180
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 181 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
F030100005
Required COM+ servers are not installed. Contact a service
engineer.
➞ Inform a service engineer of the error code.
Click [OK] to close the dialog box, and then shut down the image-capture
computer (see 2.2).
F03020200E
Memory allocation failed. Restart the system. If the problem is not
resolved after restarting, contact a service engineer.
F03020201B
Failed to upload calibration data. Restart the system. If the problem
is not resolved after restarting, contact a service engineer.
F03020201C
Failed to transfer the image data. Restart the system. If the problem
is not resolved after restarting, contact a service engineer.
F03020201F
A detector error occurred. Restart the system. If the problem is not
resolved after restarting, contact a service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
If this error occurs again, inform a service engineer of the error code.
F030202021
The control software/controller is not compatible with the current
detector firmware. Update the detector firmware.
➞ Click [OK] to close the dialog box, and inform a service engineer of the error
code.
F030202022
The Jumbo packet settings are incorrect or a network card not
compatible with Jumbo packet is used. Contact a service engineer.
181
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 182 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
F03020FFFF
An internal error occurred. Restart the system. If the problem is not
resolved after restarting, contact a service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
If this error occurs again, inform a service engineer of the error code.
F040100066
The IP addresses for the wireless and wired configurations have
been reversed. Contact a service engineer.
➞ Click [OK] to close the dialog box and inform a service engineer of the error
code.
F040300048
The control software/controller is not compatible with the current
detector firmware. Update the detector firmware.
F040300050
The Jumbo packet settings are incorrect or a network card not
compatible with Jumbo packet is used. Contact a service engineer.
➞ Inform a service engineer of the error code.
Click [OK] to close the dialog box, and then shut down the image-capture
computer (see 2.2).
F040300051
The IP addresses for the wireless and wired configurations have
been reversed. Contact a service engineer.
➞ Click [OK] to close the dialog box and inform a service engineer of the error
code.
F040300053
Cannot start the system because Non Generator Connection Mode
is enabled under the presence of workspaces for the detector which
does not support Non Generator Connection. Contact a service
engineer.
➞ Click [OK] to close the dialog box, and inform a service engineer of the error
code.
F040400003
Failed to acquire the system settings. Restart the system. If the
problem is not resolved after restarting, contact a service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
If this error occurs again, inform a service engineer of the error code.
182
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 183 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
F040700012
The graphic card does not meet the required specifications.
Examinations may fail. Contact a service engineer.
➞ Inform a service engineer of the error code.
Click [OK] to close the dialog box and inform a service engineer of the error
code.
F040700100
The system memory settings are invalid. Check the setting
information.
➞ Inform a service engineer of the error code.
Click [OK] to close the dialog box, and then shut down the image-capture
computer (see 2.2).
F040D00024
The event that affects the operation of the system was detected.
Restart the system. If the problem is not resolved after restarting,
contact a service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
If this error occurs again, inform a service engineer of the error code.
F050502009
A detector communication error occurred. Restart the system. If the
problem is not resolved after restarting, contact a service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
If this error occurs again, inform a service engineer of the error code.
183
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 184 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E020201004
Unexpected error has occurred. Gamma adjustment is not available.
Restart the system. If the problem is not resolved after restarting,
contact a service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
Then perform gamma adjustment (refer to 2.3.2 in the Setup Guide).
If this error occurs again, inform a service engineer of the error code.
E020201005
Failed to connect to the barcode reader. Check that the barcode
reader cable is connected properly. If the problem is not resolved,
contact a service engineer.
➞ Click [OK] to close the dialog box, check the connection status of the barcode
reader, and then reconnect it. After that, confirm that reading a bar code
works properly with the barcode reader. If it does not work, inform a service
engineer of the error code.
E020201006
Cannot start specified examination. Return to the Exam screen and
retry the failed examination.
➞ Click [OK] to close the dialog box and retry the previous operation on the
[EXAM > Worklist] screen or the [EXAM > Manual] screen.
If this error occurs again, inform a service engineer of the error code.
E020201007
A detector communication error occurred. Please retake this
protocol. If the problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box and retry the examination for the same
protocol.
If this error occurs again, inform a service engineer of the error code.
E020201008
The temperature of the detector exceeds the operable limit. Stop the
operation for a while with the power turned on.
➞ Stop the examination (see 4.6), wait until the temperature goes down to the
operable value, and then restart the examination again.
184
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 185 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E020201009
Failed to start up the OverwrapSoftware. Restart the system. If the
problem is not resolved after restarting, contact a service engineer.
➞ Inform a service engineer of the error code.
Click [OK] to close the dialog box, and then shut down the image-capture
computer (see 2.2).
E020201010
Detector cannot be used because the battery has lost its charge.
Charge the detector battery.
➞ Click [OK] to close the dialog box, and charge the detector battery.
E020201011
The images cannot be captured because the battery is not attached
to the detector. To capture the images, attach the battery to the
detector.
➞ Click [OK] to close the dialog box, and attach the battery pack to the detector.
E020201012
The images cannot be captured because the battery has lost its
charge. Charge the detector battery.
➞ Click [OK] to close the dialog box, and charge the battery of the wireless
detector. If the optional wired connection is available, switch to the
connection. For the wireless detector, refer to the Digital Radiography CXDI
series User’s Manual.
E020201013
The images cannot be captured because the battery has lost its
charge. Charge the detector battery.
➞ Click [OK] to close the dialog box, and charge the battery of the wireless
detector. If the optional wired connection is available, switch to the
connection. For the wireless detector, refer to the Digital Radiography CXDI
series User’s Manual.
E020201016
The captured image was not saved normally because an abnormality
occurred last time.
➞ Click [OK] to close the dialog box and retry to capture the images.
If this error occurs again, inform a service engineer of the error code.
185
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 186 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E020201018
A large number of data transfer processes have not completed.
Confirm the transfer status from the Process Viewer. To complete the
processes, ensure good wireless communication and wait until the
all processes completes without starting next examination.
➞ Click [OK] to close the dialog box, and then confirm the status of data transfer
on the Process Viewer (refer to 3.3 in the Setup Guide). Do not start next
examination until all the transfer is completed.
E030100001
No calibration data. Perform calibration in the QC Tool.
➞ Click [OK] to close the dialog box, and then calibrate the detector (refer to
3.5.1 in the Setup Guide).
E030201002
Periodic noise detected in acquisition data. The grid must be
removed in calibration.
➞ Click [OK] to close the dialog box. Remove the grid from the detector, and
then calibrate the detector (refer to 3.5.1 in the Setup Guide).
E030201003
The cooling unit has been attached or detached. Processing was
halted.
➞ Click [OK] to close the dialog box and retry the examination.
If this error occurs again, inform a service engineer of the error code.
E030201005
E030201006
Generator is disconnected. Enable the generator connection with
[Connect GEN] on the system setup screen. Restart the system if the
generator connection has failed. If the problem is not resolved after
restarting, contact a service engineer.
➞ Click [Connect GEN] in the system setup screen (refer to 2.1.1 in the Setup
Guide). If the connection with the X-ray generator device continues to be
disabled, restart the image-capture computer and the Software (see 2.2).
E030201007
Failed to enter the image capture ready condition. Restart the
system. If the problem is not resolved after restarting, contact a
service engineer.
➞ Click [OK] to close the dialog box. Exit the QC Tool mode and restart the
image-capture computer and the Software (see 2.2). If this error occurs again,
inform a service engineer of the error code.
186
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 187 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E030201008
Detector cannot be used because the battery has lost its charge.
Processing was halted.
➞ Click [OK] to close the dialog box and suspend the examination (see 4.6).
Then charge the battery of the wireless detector and restart the suspended
study order. For the wireless detector, refer to the Digital Radiography CXDI
series User’s Manual.
E030201009
The temperature of the detector exceeds the operable limit. Stop the
operation for a while with the power turned on.
➞ Stop the examination (see 4.6), wait until the temperature goes down to the
operable value, and then restart the examination again.
E030201010
No battery is attached to the detector. Attach the battery.
➞ Click [OK] to close the dialog box and attach the battery pack to the detector.
E040100008
An invalid detector has been specified. Check the setting
information. If the problem is not resolved, contact a service
engineer.
➞ Click [OK] to close the dialog box and retry the previous operation.
If this error occurs again, inform a service engineer of the error code.
E040100023
Failed to connect the generator. Check the connecting condition,
and retry to connect it by clicking [Connect GEN] button on the
system setup screen or starting examination.
➞ Click [OK] to close the dialog box, and then check that the generator is
connected properly. Then, click [Connect GEN] on the system setup screen
(refer to 2.1.1 in the Setup Guide). Or, start the examination (see 4.1).
E040100025
Generator is disconnected. Enable the generator connection with
[Connect GEN] on the system setup screen. Restart the system if the
generator connection has failed. If the problem is not resolved after
restarting, contact a service engineer.
➞ Click [OK] to close the dialog box, and then click [Connect GEN] (refer to 2.1.1
in the Setup Guide). Or, restart the image-capture computer and the Software
(see 2.2). If this error occurs again, inform a service engineer of the error code.
187
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 188 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E040100053
The detector issued an error. Try again. If the problem is not resolved,
contact a service engineer.
➞ When using a wireless detector, show Additional Information to check the
Error Code number (detector error code) to be shown and then click [OK] to
close the dialog box. Refer to the “Troubleshooting” in the “User’s Manual” of
the Digital Radiography and check the operation corresponding to the Error
Code number. Perform the operation and retry the examination.
If this error occurs again, inform a service engineer of the error code.
When using a detector other than wireless detector, click [OK] to close the
dialog box and retry the previous operation. If this error occurs again, show
Additional Information to check the Error Code number (detector error code)
to be shown and then inform a service engineer of the error code.
E040100054
The detector issued a fatal error. Restart the system. If the problem
is not resolved after restarting, contact a service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2). (However, users can continue
examinations without restarting by switching to another available detector.) If
this error occurs again, inform a service engineer of the error code.
E040100055
Generator communication failed.
➞ Click [OK] to close the dialog box, and then contact the system administrator.
The problem may be resolved by resetting the imaging conditions.
If this error occurs again, inform a service engineer of the error code.
E040100056
E040100062
E040100076
Generator is disconnected. Enable the generator connection with
[Connect GEN] on the system setup screen. Restart the system if the
generator connection has failed. If the problem is not resolved after
restarting, contact a service engineer.
➞ Click [OK] to close the dialog box, and then click [Connect GEN] (refer to 2.1.1
in the Setup Guide). If the generator cannot be connected, restart the image-
capture computer and the Software (see 2.2). If this error occurs again, inform
a service engineer of the error code.
E040101021
Failed to connect the X-ray interface. Check the connecting
condition. If the problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, check the connection status of the X-ray
interface, and then reconnect it. If this error occurs again, inform a service
engineer of the error code.
188
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 189 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E040101038
Detected a signal which seems to be X-ray in the status that an
examination is not ready.
The causes might be the followings. Take measures, if applicable:
(Cause 2) There are some electronic devices that sends out a strong
electromagnetic noise near the detector.
(Measure 2) Keep the electronic devices away from the detector.
E04020001F
An abnormality was detected in the database. If continued to use, an
operating problem may occur. Contact a service engineer.
➞ Inform a service engineer of the error code.
E040200020
Failed to delete the backup folder of the database. Restart the
system. If the problem is not resolved after restarting, contact a
service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
If this error occurs again, inform a service engineer of the error code.
189
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 190 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E040300055
An abnormality occurs while acquiring the signal strength indicator
of wireless LAN. There is no established wireless LAN adapter.
Contact a service engineer.
➞ Click [OK] to close the dialog box, and then inform a service engineer of the
error code.
E040500003
Exceeded the maximum number of the study order acquisitions.
If you want to change the number of maximum study order
acquisitions, contact a service engineer.
➞ Click [OK] to close the dialog box.
When the target study orders have been acquired, start the examination.
Otherwise, acquire the target order by narrowing down the study orders using
[Search Conditions] (see 3.1.1).
To change the maximum number of orders that can be listed, contact a
service engineer.
E040500004
Invalid studies contained in the acquired data are not displayed.
Check the data identical to the acquired ones on the RIS database,
etc.
➞ Show Additional Information to check the number of the invalid orders and
then click [OK] to close the dialog box.
Check the study orders that failed to be shown. The Patient ID, Study
Instance UID and Modality of the orders may be invalid. In that case, correct
the study order information and retry the acquisition (see 3.1.1). If this error
occurs again, inform a service engineer of the error code.
E040500005
Invalid values in the received tag have been replaced with blanks.
Check the data identical to the acquired ones on the RIS database,
etc.
➞ Show Additional Information to check the study orders that failed to be shown
and then click [OK] to close the dialog box. Correct the study order
information and retry the data acquisition (see 3.1.1). If this error occurs again,
inform a service engineer of the error code.
E040500006
Received data includes characters that do not exist in the CXDI
system’s character set. The characters may not be displayed
correctly. If the problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box.
Supplementary Chinese characters contained in the acquired study order
were replaced with “?.”
Do not use the supplementary Chinese characters for study order creation on
RIS database for this control software does not support these characters.
190
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 191 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E040500008
A date entry before 1753 is included in the acquired Exam. Check the date.
➞ Check the data information and retry a data entry.
E04050000C
Failed to read DICOMDIR file. Check the target disk and retry reading.
➞ Click [OK] to close the dialog box. Check the removable disk and read the
files again. If this error occurs again, files in the DICOMDIR directory may be
corrupted. Check the original files.
E04050000D
A transmission error occurred. Locate the failed study in the [PAST >
Past List] and retransmit the study.
➞ Delete the data transfer error log from the Process Viewer (refer to 3.3 in the
Setup Guide), and then resend the image from the [PAST > Past List] screen
(see 7.1.5).
E04050000F
Failed to read DICOM file. Check the target disk and retry reading.
➞ Click [OK] to close the dialog box. Check that there is no problem in the
storage media and the target file exists in the destination directory, and then
retry reading. Errors also occur when the file format does not comply with
DICOM standard or log-in user has not been granted read access to the file.
E040500010
Failed to read DICOM file. Check the target disk and retry reading.
➞ Click [OK] to close the dialog box. Ensure enough HDD’s free space and retry reading.
E040500011
Failed to read the trigger file. There may be a problem in the setting
of the other communication party. Check the RIS. If the problem is
not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and then confirm the RIS settings. If this
error occurs again, inform a service engineer of the error code.
E040500012
Invalid values were contained in the trigger file. Check the data
identical to the acquired ones on the RIS.
➞ Click [OK] to close the dialog box, and then check the transmit data from the
RIS. If this error occurs again, inform a service engineer of the error code.
191
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 192 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E040500013
There is no destination folder for the trigger file. Check the setting
information. If the problem is not resolved, contact a service
engineer.
➞ Click [OK] to close the dialog box, and then confirm the destination folder for
the trigger file. If this error occurs again, inform a service engineer of the error
code.
E040500014
Failed to lock the acquisition notification of the patient information.
The unintended examination might be performed because the
notification from the communication party is not made invalid.
Check again whether the current examination is correct. There may
be a problem in the settings of the system or that of the other
communication party. Contact a service engineer.
➞ Click [OK] to close the dialog box, and then confirm the destination folder for
the trigger file. If this error occurs again, inform a service engineer of the error
code.
E040501011
Cannot continue the processing because a file already exists at the
specified storage destination. Check the setting information.
➞ Click [OK] to close the dialog box. Delete all files in the folder designated as
the storage, and then retry the previous operation. If the destination folder
cannot be identified, consult the system administrator. If this error occurs
again, inform a service engineer of the error code.
E040501012
Cannot continue the processing because access to the specified
storage destination was denied. Check the setting information.
➞ Click [OK] to close the dialog box, right-click the folder designated as the
storage, select Property > Security tab, change the access privileges, and
then retry the operation from the start. If the destination folder cannot be
identified, consult the system administrator. If this error occurs again, inform
a service engineer of the error code.
E040501013
Cannot continue the processing because there is not enough space
at the specified storage destination. Check the setting information.
➞ Change the destination folder. Or ensure enough HDD’s free space and retry
the operation.
192
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 193 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E040501014
DICOM GSPS output was halted because the object image transfer
was incomplete. Try again after object image transfer.
➞ Click [OK] to close the dialog box.
Select the halted transfer task, and then click [Retry] on the process viewer to
retry the task (refer to 3.3 in the Setup Guide).
If the above remedy does not work, try to reconfigure the Storage tab settings
(refer to 2.6.1 in the Setup Guide), and manually try to transfer the image
starting on the [PAST > View] screen (see 7.1.5).
E040501016
Data communication between the hospital network was interrupted
during file transmission. Check the setting information.
➞ Click [OK] to close the dialog box. After confirming that the destination server
and the network are functioning normally, retry the operation. If this error
occurs again, inform a service engineer of the error code.
E040501017
The CXDI system read an unsupported DICOMDIR file. Check the
target disk and retry reading.
➞ Use a disk that was made using the Software.
E040501018
A transmission error occurred. Try again. If the problem is not
resolved, contact a service engineer.
➞ Retry the data transfer from the Process Viewer (refer to 3.3 in the Setup
Guide). Or resend the image from the [PAST > Past List] screen (see 7.1.5). If
this error occurs again, inform a service engineer of the error code.
E040502002-040502004
E04050200D-04050200F
E040502010-040502014
Association negotiation failed. There may be a problem in the
settings of the system or that of the other communication party.
Check the setting information.
E04050201B-04050201F
E040502020
DICOM communication has been interrupted. There may be a
problem in the settings of the system or that of the other
communication party. Check the setting information.
193
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 194 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E040502021
Timeout occurred during association negotiations. There may be a
problem in the settings of the system or that of the other
communication party. Check the setting information.
➞ Click [OK] to close the dialog box and retry acquiring the worklist, or check
the properties for Host Name, Port or Called AE Title in the Connection tab
screens (refer to 2.6 in the Setup Guide).
Check that the destination server and network are functioning normally.
If the problem cannot be resolved, contact the system administrator. The
problem may be resolved by checking the destination server or the DICOM
conformance statement. If this error occurs again, inform a service engineer
of the error code.
E040502058
Association negotiation failed. Network communication with other
communication party failed. Check the network condition and
setting information.
➞ Click [OK] to close the dialog box and check the properties for Host Name,
Port or Called AE Title in the Connection tab screens (refer to 2.6 in the Setup
Guide).
Check that the destination server and network are functioning normally.
If the problem cannot be resolved, contact the system administrator. The
problem may be resolved by checking the destination server or the DICOM
conformance statement. If this error occurs again, inform a service engineer
of the error code.
E04050205E
An undefined code in DICOM communication has been detected.
Check the destination modality or server, etc.
➞ Contact the system administrator. The problem may be resolved by checking
the destination server or the DICOM conformance statement.
If this error occurs again, inform a service engineer of the error code.
E04050300C
Timeout occurred during DICOM communication. There may be a
problem in the settings of the system or that of the other
communication party. Check the setting information.
➞ Click [OK] to close the dialog box and retry transferring images, or check the
properties for Host Name, Port or Called AE Title in the Port or Connection
tab screens (refer to 2.6 in the Setup Guide).
Check that the destination server and network are functioning normally.
If the problem cannot be resolved, contact the system administrator. The
problem may be resolved by checking the destination server or the DICOM
conformance statement. If this error occurs again, inform a service engineer
of the error code.
194
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 195 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E04050301E
Disconnection request was received during DICOM communication.
There may be a problem in the settings of the system or that of the
other communication party. Check the setting information.
➞ Click [OK] to close the dialog box and retry the previous operation.
If the problem cannot be resolved, contact the system administrator. The
problem may be resolved by checking the destination server or the DICOM
conformance statement. If this error occurs again, inform a service engineer
of the error code.
E040505001
E040505002
E040505004-040505009
E04050500B-040505014
An error occurred in DICOM communication. Try again. If the
problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box. Click [Refresh List] to acquire the study
orders (see 3.1.1).
The problem may be resolved by checking the destination server or the
DICOM conformance statement.
If this error occurs again, inform a service engineer of the error code.
E040505015
A response error occurred in DICOM communication. Check the
destination modality or server, etc.
➞ Click [OK] to close the dialog box, and then contact the system administrator.
The problem may be resolved by checking the destination server or the
DICOM conformance statement.
If this error occurs again, inform a service engineer of the error code.
E040505017
An error occurred in DICOM communication. Try again. If the
problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box. Click [Refresh List] to acquire the study
orders (see 3.1.1).
The problem may be resolved by checking the destination server or the
DICOM conformance statement.
If this error occurs again, inform a service engineer of the error code.
E040505018
A response error occurred in DICOM communication. Check the
destination modality or server, etc.
➞ Click [OK] to close the dialog box, and then contact the system administrator.
The problem may be resolved by checking the destination server or the
DICOM conformance statement.
If this error occurs again, inform a service engineer of the error code.
195
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 196 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E04050501B
An error occurred in DICOM communication. Try again. If the
problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box. Click [Refresh List] to acquire the study
orders (see 3.1.1).
The problem may be resolved by checking the destination server or the
DICOM conformance statement.
If this error occurs again, inform a service engineer of the error code.
E04050501F
An error occurred in DICOM communication. Contact the system
administrator.
E040505020
A response error occurred in DICOM communication. Check the
destination modality or server, etc.
E040505027
An error occurred in the DICOM printer output. Contact the system
administrator.
➞ Click [OK] to close the dialog box, and then contact the system administrator.
The problem may be resolved by checking the destination server or the
DICOM conformance statement.
If this error occurs again, inform a service engineer of the error code.
E040505028
An error occurred in the DICOM printer output. Try again later.
➞ Click [OK] to close the dialog box, and retry the data transfer from the Process
Viewer (refer to 3.3 in the Setup Guide).
If the problem cannot be resolved, contact the system administrator. The
problem may be resolved by checking the DICOM conformance statement. If
this error occurs again, inform a service engineer of the error code.
E040505029
An error occurred in the DICOM printer output. Contact the system
administrator.
➞ Click [OK] to close the dialog box, and then contact the system administrator.
The problem may be resolved by checking the destination server or the
DICOM conformance statement.
If this error occurs again, inform a service engineer of the error code.
E040505031
A response error occurred in DICOM communication. Check the
destination modality or server, etc.
➞ Click [OK] to close the dialog box and contact the system administrator. The
problem may be resolved by checking the destination server or the DICOM
conformance statement. If this error occurs again, inform a service engineer
of the error code.
196
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 197 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E04050502A
E04050502E
An error occurred in the DICOM printer output. Contact the system
administrator.
➞ Click [OK] to close the dialog box, and then contact the system administrator.
The problem may be resolved by checking the destination server or the
DICOM conformance statement.
If this error occurs again, inform a service engineer of the error code.
E04050502C
E04050502F
An error occurred in the DICOM printer output. Try again later.
➞ Click [OK] to close the dialog box, and then retry the data transfer from the
Process Viewer (refer to 3.3 in the Setup Guide).
If the problem cannot be resolved, contact the system administrator. The
problem may be resolved by checking the DICOM conformance statement. If
this error occurs again, inform a service engineer of the error code.
E040506001
The transmission was interrupted as the CXDI system received a
FAILURE status from the destination DICOM printer. Check the
DICOM printer.
➞ Check the DICOM printer. For details of Printer Status, refer to Annex C.13.9.1
in the DICOM standard Part3: Information Object Definitions or the DICOM
conformance statement.
E040507001
DICOM StorageCommitment failed. Try again. If the problem is not
resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and then retry the data transfer from the
Process Viewer (refer to 3.3 in the Setup Guide). Or resend the image from the
[PAST > View] or [PAST > Past List] screen (see 7.1.5).
If this error occurs again, inform a service engineer of the error code.
E040507002-040507005
DICOM StorageCommitment failed. If the problem is not resolved,
contact a service engineer.
➞ Click [OK] to close the dialog box, and then retry sending the image from the
[PAST > View] or [PAST > Past List] screen (see 7.1.5).
If this error occurs again, inform a service engineer of the error code.
197
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 198 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E040507006
DICOM StorageCommitment failed. Try again. If the problem is not
resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and then retry the data transfer from the
Process Viewer (refer to 3.3 in the Setup Guide). Or resend the image from the
[PAST > View] or [PAST > Past List] screen (see 7.1.5).
If this error occurs again, inform a service engineer of the error code.
E040507007
DICOM StorageCommitment failed. If the problem is not resolved,
contact a service engineer.
➞ Click [OK] to close the dialog box, and then retry the data transfer from the
Process Viewer (refer to 3.3 in the Setup Guide). Or resend the image from the
[PAST > View] or [PAST > Past List] screen (see 7.1.5).
If this error occurs again, inform a service engineer of the error code.
E040508001
Failed to create DICOM disc. The problem may be resolved by
changing disc or writing speed or restarting the system. If the
problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and then try the operation again by changing
the disc or recording speed or restart the system (see 7.1.6, page 157).
If this error occurs again, inform a service engineer of the error code.
E040508002
Failed to create DICOM disc. The specified disc recorder is not
found. Check the setting information.
➞ Click [OK] to close the dialog box, and then try the operation again. If there is
another drive, specify it, and then try the operation again (refer to 2.6.6 in the
Setup Guide).
E040600008
Failed to access the target disk. Check the disk drive and retry from
the Process Viewer.
➞ Check that the destination storage disk or device is functioning normally.
Then, retry the data transfer from the Process Viewer (refer to 3.3 in the Setup
Guide).
If this error occurs again, inform a service engineer of the error code.
E040600009
The process was interrupted as there was not enough free disk
space. Replace the disk with a new one and retry from the Process
Viewer. If recording is in progress, wait until the process is
completed and replace the disk.
➞ Replace the current disk with one containing sufficient free space. Then, retry
the data transfer from the Process Viewer (refer to 3.3 in the Setup Guide).
198
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 199 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E04060000B
Cannot retry the process as an error occurred. Delete the failed
process in the process viewer, locate the process in the [PAST > Past
List], and then retransmit the process.
➞ Delete the data transfer error log from the Process Viewer (refer to 3.3 in the
Setup Guide). Then, resend the image from the [PAST > Past List] screen (see
7.1.5).
E04060000C
The system is attempting to transfer old study information created
before data rebinding. Delete the corresponding process as it cannot
be executed.
➞ Delete the data transfer error log from the Process Viewer (refer to 3.3 in the
Setup Guide). The image including old study information cannot be
transferred as the rebound image has been successfully transferred.
E04060000D
The request could not be performed because the recording
destination was in progress. Delete the failed process in the process
viewer, locate the process in the [PAST > Past List], and then
retransmit the process.
➞ Delete the data transfer error log from the Process Viewer (refer to 3.3 in the
Setup Guide). Then, resend the image from the [PAST > Past List] screen (see
7.1.5).
E040700006
Failed to load the image file. If the problem is not resolved, contact a
service engineer.
➞ Click [OK] to close the dialog box, and then retry the previous operation.
E040800100
Cannot start the examination as an invalid protocol exists in the
study. Delete the invalid protocol and retry the examination.
➞ Click the suspended examination including the invalid protocol in
[EXAM > Worklist] and then [Edit Exam] to delete it (see 3.1.6).
E040D0001F
Failed to free up the hard disk. If the problem is not resolved after
restarting, contact a service engineer.
➞ Deselect the Protect Image check box of the study order in the [PAST > Past
List] (see 3.1.4). Select the transfer failed studies in the Process Viewer and
click [Delete] or [Retry] (refer to 3.3 in the Setup Guide). If this error occurs
again, inform a service engineer of the error code.
199
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 200 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E040D00021
An abnormality was detected in the HDD. If continued to use, an
operating problem may occur. Perform a backup of the database.
Contact a service engineer.
➞ Inform a service engineer of the error code.
E040D00025
The event that affects the operation of the system was detected.
Restart the system. If the problem is not resolved after restarting,
contact a service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
If this error occurs again, inform a service engineer of the error code.
E040D00201
The specified network interface for observation setting is not found.
Set the observation function to invalid. Contact a service engineer.
➞ Inform a service engineer of the error code.
E040D00202
It is recognized that CCS communicates wirelessly with a detector
with the built-in WLAN adapter in your computer; nevertheless
timeout has occurred during starting up the wireless Hosted
Network. Restart the system. If the problem is not resolved after
restarting, contact a service engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
If this error occurs again, inform a service engineer of the error code.
E040E01001
Failed to load the extra defect correction file. If the problem is not
resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and then calibrate the detector (refer to
3.5.1 in the Setup Guide). If this error occurs again, inform a service engineer
of the error code.
E041000004
Invalid studies are acquired. Check the examination file.
➞ Check if the study orders are valid. Acquire valid study orders again from the
HIS/RIS database.
200
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 201 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E041000005
Entered information has invalid value. Invalid values are replaced
with blanks. Check the examination file.
➞ Check if the study orders are valid. Acquire valid study orders again from the
HIS/RIS database.
E041000006
Specified ID is not found in the examination file. Check the
examination file.
➞ Check if the study orders are valid. Acquire valid study orders again from the
HIS/RIS database.
E041000008
An error was discovered while checking the user input by the script.
If the problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and inform a service engineer of the error
code.
E041000012
An error has occurred in CCRHIS library. Try again. If the problem is
not resolved, contact a service engineer.
➞ Show Additional Information to check the error details and click [OK] to close
the dialog box. Then retry the previous operation according to the error
details.
If this error occurs again, inform a service engineer of the error code.
E041000020
Examination file loading was halted because the acquired exams
reached the maximum number. End the suspended exam and retry
reading.
➞ Click [OK] to close the dialog box. Restart suspended studies from the
[EXAM > Worklist] screen (see 4.6), finish the examinations, and then retry
reading.
E041000021
Invalid studies are loaded and cannot be displayed. Check the
examination file.
➞ Click [OK] to close the dialog box. Check the values for essential information,
and then retry reading.
201
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 202 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E041000022
Entered information has invalid value. Invalid values are replaced
with blanks. Check the examination file.
➞ Click [OK] to close the dialog box. Check the values for information which are
not essential, and then retry reading.
E041000023
No examination files. Check the examination file or the setting
information.
➞ Click [OK] to close the dialog box. Check that the information entered in CSV
format is identical with the actual location and name of a folder which
includes study orders, and retry reading. If the folder is mapped as a network
drive, check the connection status of the folder.
E041000024
An error occurred during automatic examination file updating. Check
the setting information.
➞ Click [OK] to close the dialog box. Check that the information entered in CSV
format is identical with the actual location and name of a folder which
includes study orders, and retry reading. If the folder is mapped as a network
drive, check the connection status of the folder.
E041000025
Failed to read the examination file. Check the setting information.
➞ Click [OK] to close the dialog box. Check that there is no problem in the media
that includes the examination files or whether the file is corrupted or not.
E041100004
Failed to start up CxdiInfo. Check the setting information.
➞ Click [OK] to close the dialog box and inform a service engineer of the error
code.
E050501008
E050501009
A drive circuit error occurred. Contact a service engineer.
E05050100A
E05050100B
An abnormality was detected in the detector structure. Contact a
service engineer.
E05050100C
E05050100D
A control line connection error was detected in the detector. Contact
a service engineer.
➞ Inform a service engineer of the error code.
202
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 203 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
E050501010
Failed to transfer the image data. Try again. If the problem is not
resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and then retry the image transfer.
If this error occurs again, inform a service engineer of the error code.
E050501012
An abnormality was detected in the analog power supply for the
detector. Contact a service engineer.
E050501014
A grid error was detected. Contact a service engineer.
E050501016
The detector firmware is different. Contact a service engineer.
➞ Inform a service engineer of the error code.
E050502026
A detector communication error occurred. The cable may be
disconnected or there may be power discontinuity in the detector.
Check that the cable is correctly connected and restart the detector.
If the problem is not resolved after restarting, contact a service
engineer.
➞ Click [OK] to close the dialog box, check the cable connection to the detector,
and restart the detector.
If this error occurs again, inform a service engineer of the error code.
203
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 204 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
W020202002
The temperature of the detector exceeds the operable limit. Stop the
operation for a while with the power turned on.
➞ Stop the examination (see 4.6), wait until the temperature goes down to the
operable value, and then restart the examination again.
W020202003
The detector is not available. Irradiation has finished.
➞ Click [OK] to close the dialog box and then continue the operation.
W020202004
The detector battery is weakening. Charge the detector battery.
➞ Click [OK] to close the dialog box and charge the battery of the wireless
detector. If the optional wired connection is available, switch to the
connection. For details on the wireless detector, refer to the Digital
Radiography CXDI series User’s Manual.
W020202005
The specified disk space could not be cleaned up.
➞ Check the target data volume to be deleted and then retry deletion (see 5.3).
W030203003
The temperature of the detector exceeds the operable limit. Stop the
operation for a while with the power turned on.
➞ Stop the examination (see 4.6), wait until the temperature goes down to the
operable value, and then restart the examination again.
W040100045
Calibration data is invalid. Update the calibration data.
➞ Click [OK] to close the dialog box and then calibrate the detector (refer to
3.5.1 in the Setup Guide).
W040100052
The detector issued a warning.
➞ Click [OK] to close the dialog box and then continue the operation.
204
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 205 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
W040101014
The detector has changed to the standby state because it had not
been used for a certain period. To use it, select a protocol button
again.
➞ Click [OK] to close the dialog box, and click the protocol.
W040101029
An error has occurred on the detector battery charging. If the
problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and change to the spare battery. Exposure
can be continued as power feeding continues.
If this error occurs again, inform a service engineer of the error code.
W040300058
An abnormality occurs while acquiring the signal strength indicator
of wireless LAN. It is being automatically acquired again. For
wireless connection with the detector, the signal strength indicator
cannot be shown correctly during the restoration work.
➞ When it is not restored, inform a service engineer of the error code.
W040500007
Characters which do not match the current character code are used
in this data. This character code will be properly changed
automatically. Check the setting information.
➞ Click [OK] to close the dialog box.
Characters that do not match the current character code will be replaced with
“?.” Continue the operation when the replacement is acceptable.
To disable the replacement, inform a service engineer of the error code.
W040500009
Failed to copy a tag to C-STORE. Check the tag information and the
C-STORE communication log.
If the problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and check the original DICOM tag and the
C-STORE communication. Then correct the original tag and retry the previous
operation. If this error occurs again, inform a service engineer of the error
code.
W04050000A
Failed to copy tag to the DicomFile. Check the original data. If the
problem is not resolved, contact a service engineer.
➞ Click [OK] to close the dialog box, and then check the original DICOM tags of
the DicomFile and the DICOM communication. Correct the original tags. If
this error occurs again, inform a service engineer of the error code.
205
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 206 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
W040505003
W04050500A
W04050501C-04050501E
An error occurred in DICOM communication.
➞ Contact the system administrator.
The problem may be resolved by checking the destination server or the
DICOM conformance statement. If the problem cannot be resolved, inform a
service engineer of the error code.
W040505030
The transmission was complete normally. However, the CXDI system
received WARNING status from the destination DICOM printer.
Check the DICOM printer.
➞ Check the DICOM printer. For details on Printer Status, refer to Annex
C.13.9.1 in the DICOM standard Part3: Information Object Definitions or the
DICOM conformance statement.
W040D00022
Performed a check of the database because an abnormality was
detected.
➞ Click [OK] to close the dialog box and then continue the operation.
W040D00023
The system has not been shut down for a long time. Shut down the
system and backup the database periodically.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2). Periodically back up the database
manually or automatically.
W040D00026
The event that affects the operation of the system was detected.
➞ Inform a service engineer of the error code.
W050501002-050501004
A Flash ROM error was detected in the detector. Contact a service
engineer.
➞ Click [OK] to close the dialog box, and then restart the image-capture
computer and the Software (see 2.2).
If this error occurs again, inform a service engineer of the error code.
W050501005
A grid error may have caused an image abnormality.
➞ Click [OK] to close the dialog box. There may be a problem in the captured
image. Check the image and retry the examination.
206
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 207 Thursday, April 7, 2016 2:28 PM
10 Troubleshooting Remedies
W05050100E
W05050100F
An error was detected in the detector.
➞ Inform a service engineer of the error code.
W050501011
The transfer of the image data was stopped.
➞ Click [OK] to close the dialog box. There may be a problem in the captured
image. Check the image and retry the examination.
W050501013
A grid ID loading error was detected.
➞ Check if the grid is properly attached.
If the problem cannot be resolved, inform a service engineer of the error code.
W050501015
A grid error may have caused an image abnormality.
➞ Click [OK] to close the dialog box, and check the captured image.
If there is a problem in the image, inform a service engineer of the error code.
207
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 208 Thursday, April 7, 2016 2:28 PM
Index
A E
Access group 121 Edit Exam 50
Accession Number 41 Edit Patient Information 61
Age 58 Edit Study Information 46, 59
Anatomical Part 123 Emergency 63
Annotation Enhancement 127
Creating/editing 141 Error 180
Preview annotation 67, 90 List 184
Showing/hiding 67 EXAM 10, 12
Storage annotation 91 Exam 40
Arranging images Examination
Automatically 149 Ending 73
Manually 153 Recalling 94
Authentication 31 Restarting 81
Auto 125 Starting 67
Suspending 81
Exposure mode 66
B Exposure type 66
Binding data 85
Birth date 58
Body part search mode 52 F
Brightness Fatal error 180
Base Brightness 125 List 181
Gradation 124 Fit 108
Highlights and shadows 122 Flip 106
ROI 110
Brightness Adjustment mode 125
Button 10 G
Grid Suppression 131
C
Category search mode 55 H
Complete protocol HDD free space 41
Reprocessing images 76 High-resolution mode 108
Contrast HIS
Base Contrast 125 Study order 40
Shape enhancement 127 Histogram 122
Create New Annotation 142
Crop 112 I
ID
D Patient information 58
DICOM disc 157 Search Conditions 43
Directory options 92 Image
Dynamic Range Adjustment 128 Reprocessing 76
Image processing 121
Incomplete protocol
Reprocessing 79
Restarting 81
208
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 209 Thursday, April 7, 2016 2:28 PM
Index
209
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 210 Thursday, April 7, 2016 2:28 PM
Index
T
Tab 10
Toolbar 22, 104
Touch-screen 11
V
View mode 90
W
Warning 180
List 204
Wireless detector 49, 68
Worklist
Screen 40
Tab 10
Workspace 40
Change 48, 74
Z
Zoom in/Zoom out 108
210
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 211 Thursday, April 7, 2016 2:28 PM
211
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 212 Thursday, April 7, 2016 2:28 PM
212
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 213 Thursday, April 7, 2016 2:28 PM
213
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 214 Thursday, April 7, 2016 2:28 PM
214
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 215 Thursday, April 7, 2016 2:28 PM
000_L-IE-4247B_CXDI_NE_OM_V2141.book Page 216 Thursday, April 7, 2016 2:28 PM
*L-IE-4247B*
Manufacturer:
CANON INC. Medical Equipment Group
9-1, Imaikami-cho, Nakahara-ku, Kawasaki, Kanagawa 211-8501, Japan
Telephone: (81)-3-3758-2111
Distributed by:
CANON U.S.A., INC. Healthcare Solutions Division, Business Imaging Solutions Group
One Canon Park, Melville, NY 11747, U.S.A.
Telephone: (1)-800-970-7227