Change Port Number 526521
AMX 700 Cannot Configure Port Numbers 1023 and Lower for Network Nodes and
Printers
Issue
Definium AMX 700 The hospital uses port numbers 1023 or lower in their network setup. Since this is poor networking practice, the system
does not allow you to enter these port numbers in the network setup screens.
Summary:
The hospital uses port numbers 1023 or lower in their network setup. Since this is poor networking practice, the system does not allow you to
enter these port numbers in the network setup screens.
Resolution
To set a port number 1023 or below for a network node or printer, first set up all the network nodes and printers on the system using port
number “1024” - WQA
Details:
Remember the name of the network node or printer that needs to have its port number changed. Follow the instructions below to
change the Port number in the config file manually:
1. Press CTRL + SHIFT + F5 to open the Engineering UI (must have the USB Keyboard installed).
3. Type the following into the command line for the type of host you need to change:
4. Once at the correct directory, type the following for the type of host you need to change:
5. Find the name of the host you need to change in the file and change either of the following lines shown below the name (where “XXX” is
the desired port number for that node):
<port>1024<port> to <port>XXX<port>
<dPort>1024</dPort> to <dPort>XXX</dPort>
This Resolution refers to standard best practices, tools and procedures that fall within the scope of past service training curriculum and