Professional Documents
Culture Documents
2.EI 6401 - Nov 15 - TR Key
2.EI 6401 - Nov 15 - TR Key
Precision is the degree of closeness with which a given value may be repeatedly
measured.
When the input to a transducer is increased slowly from some non-zero arbitrary
value, the change in output is not detected at all until a certain input increment is
exceeded. This increment is defined as the resolution.
PART B – (5 x 16 = 80 marks)
11.(a) (i) Discuss the causes and remedies for different classes of errors. (10)
Types of errors:
Gross Error
Systematic Error
Random Error
Gross Error: This class of errors mainly covers human mistakes in reading instruments
and recording and calculating measurement result.
The responsibility of the mistake normally lies with the experimenter.
The experimenter may grossly misread the scale.
For example One may, read the temperature as 31.5°C due to oversight while the actual
reading may be 21.5°C.
Gross errors can be avoided by adopting two means:
Great care should be taken in reading and recording the data.
Prepared by Prof.S.Nagammai, HOD/EIE, KLNCE Page 2
EI 6401 TRANSDUCER ENGINEERING
Two, three or even more readings should be taken for the quantity under
measurement.
Gross errors may be of any amount and therefore their mathematical
analysis is impossible.
Systematic Error:
These type of errors are divided into three categories:-
i.Instrumental Error
ii.Environmental Error
iii.Observational Error
1. Instrumental Error:
These errors arise due to three main reasons:
Due to inherent short comings in the instrument.
Due to measure of the Instruments.
Due to loading effects of Instruments.
Inherent short comings of Instruments:
These errors are inherent in instruments because of their mechanical structure.
They may be due to construction, calibration or operation of the instruments or
measuring devices.
Errors may be caused of friction, hysteresis or even gear backlash loss of
controlling torque.
These errors are reduced using the following methods:
i) The procedure of measurement must be carefully planned.
ii) Correction factors should be applied after determining the instrumental
errors.
iii) The instrument may be re-calibrated carefully.
Misuse of Instruments:
The errors caused in measurements are due to the fault of the operator than that of
instrument.
Improper practices of holding instrument may cause these errors.
Loading effects:
The capability of the system to faithfully measure, record or control the input signal
(measured)in undistorted from is called the “loading effects”.
The loading effect arises the Instrumental errors.
Environmental Errors:
These errors are due to conditions external to the measuring device including
conditions in the area surrounding the instrument.
These may be effects of temperature, pressure, humidity, dust, vibrations or of
external magnetic or electrostatic fields.
The corrective measures employed to reduce these undesirable effects are:
Arrangements should be made to keep the conditions as nearly as constant as
possible.
Employing techniques which eliminate the effect of these disturbances.
In case it is suspected that external magnetic or electrostatic field can affect the
readings of instruments, magnetic (or) electrostatic shields may be provided.
Observational Errors:
There are many sources of observational errors.
As an example, the pointer of a voltmeter rests slightly above the surface of
the scale. It is also known as “PARALLAX ERROR”. [Error due to observer’s eye]
The parallax errors arise on a account of pointer and the scale not being in
the same plane.
Observational errors in measurement involving timing of an event.
One observer may tend to anticipate the signal and read too soon, but
different observers may produce different results.
Modern electrical instruments have digital display of output which
completely eliminates the errors on account of human observational.
Random (Residual) Errors:
The quantity being measured is affected by many happenings throughout the
universe.
We are aware of and some of the factors influencing the measurement, but about
the rest, we are unaware.
The happenings or disturbances about which we are unaware are lumped together
and called “Random (or) Residual.
Hence the errors caused by their happenings are called Random or Residual errors.
11.(a) (ii) Write short notes on the significance of odds and uncertainty. (6)
±0.6745σ 0.5000 1 to 1
±σ 0.6828 2.15 to 1
±2σ 0.9546 21 to 1
±3σ 0.9974 256 to 1
Uncertainty:
“Uncertainty distribution as the error distribution”
Propagation of uncertainties:
The uncertainty analysis in measurements when many variables are involved is
done on the same basis as is done for error analysis when the results are expressed as
standard deviations or probable errors.
Suppose X is a function of several variables.
X f x1 , x2 , x3 ,........xn
Where x1 , x2 , x3 ,........xn are independent variables with the same degree of odds.
Let ωx be the resultant uncertainty and x1 , x2 , x3 ,........ xn by the uncertainties in
1x
2x
nx
Static Calibration:
It refers to a procedure where an input either constant or slowly time varying is
applied to an instrument and corresponding output is measured while other inputs
are kept constant at some value. Instruments are so constructed that the signal
conversion they perform have the property of irreversibility or directionality. This
implies that a change in an input quantity will cause a corresponding change in the
output. The functional relationship between the output quantity and input quantity is
referred to as static calibration and valid under the stated constant conditions of all
other inputs.
Some applications involve the measurement of quantities that are either constant or
vary slowly with time. Under these circumstances it is possible to define a set of
11.b. (ii) The following 10 observations were recorded when measuring a voltage.
41.7, 42.0, 41.8, 42.0, 42.1, 41.9, 42.5, 42.0, 41.9, 41.8. Find: (10)
(1) Mean
(2) Standard Deviation
(3) Probable error
(4) Mode.
Solution:
41.7 42 41.8 42.0 42.1 41.9 42.0 41.9 42.5 41.8
i) mean X
10
419.7
41.97V
10
ii) The standard deviation
d1 0.27 d12 0.0729
d2 0.03 d22 0.0009
d3 0.17 d32 0.0289
d4 0.03 d42 0.0009
d5 0.13 d52 0.0169 .
d6 0.07 d62 0.0049
d7 0.03 d72 0.0009
d8 0.07 d82 0.0049
d9 0.53 d92 0.2809
d10 0.17 d102 0.0289
d2 0.441
S 0.22V
n 1 9
iii) Probable error of one reading r1 0.6745 S 0.6745 0.22 0.15V
r1 r1 0.15
iv) Probable error of mean 0.05V
n 1 9 9
v) Range = 42.5 – 41.7 = 0.8V
12. (a) (i) Compare the Accuracy and Precision of Instruments. (6)
Accuracy:
Precision:
12. (a) (ii) Obtain the step response of first order instrument and explain the
procedure for estimating the static and dynamic parameters that can
be obtained from the above response curve. (10)
C(s) K
R(s) s 1
Q r (t ) 1
1
R(S)=
S
1 A B
C ( s)
s 1 s S 1 s
A 1 B
1
C ( s) K
s 1 s
C (t ) K 1 e
t
for t 0
At t , C 1 e1 0.632
t 2 , C 2 1 e2 0.864
t 5 , C 5 1 e5 0.993 1
Time constant of a first order system is defined as the time taken by the response to
reach 63.2% of its final steady state value.
The time constant is indicative of how test the system tends to reach the final
value.A large time constant corresponds to a slow system response and a small
time constant represents a fast system response.
12.(b) Obtain the mathematical expression for output of the over damped and under
damped second order transducer. [16]
13. (a) (i) Compare the merits, limitations and typical ranges of operation of RTD
with thermocouple. [6]
Thermocouples are inexpensive, rugged, and have a fast response time but are less
accurate and the least stable and sensitive. Thermocouples also read only relative
temperature difference between the tip and the leads while RTD's and thermistors
read absolute temperature. RTD's are the best choice for repeatability, and are the
most stable and accurate. However they have a slow response time and because
they require a current source they do have a low amount of self heating.
13. (a) (ii) Describe the functioning and typical application for piezoresistive type
of sensor. [10]
PIEZORESISTIVE SENSOR:
13. (b) (i) Discuss the construction and signal conditioning circuit for strain gauges.
(8)
Strain gauge in tension increase resistance , in compression decrease resistance.
13.b. (ii) Explain the working principle of anyone type of hot-wire anemometer. (8)
Refer Q.No.13.a.(ii) April 2015 answer key.
14. (a) Explain the operation, Equivalent circuit and transfer function for LVDT.
(16)
OR
14.(b)(i)Discuss single and double element displacement type of capacitive sensor. (8)
Variable Distance Displacement Sensors
A capacitor displacement sensor, made from two flat coplanar plates with a
variable distance x apart, is illustrated in Figure 6.23. Ignoring fringe effects, the
capacitance of this arrangement can be expressed by:
where
ε= the dielectric constant or permittivity
εr= the relative dielectric constant (in air and vacuum εr≈1)
ε0= 8.854188 ×10–12F/m, the dielectric constant of vacuum
x= the distance of the plates in m
A= the effective area of the plates in m2
This type of sensor is often used for measuring small incremental displacements
without making contact with the object.
Variable Area Displacement Sensors:
Alternatively, the displacements may be sensed by varying the surface area of
the electrodes of a flat plate
capacitor, as illustrated in Figure. In this case, the capacitance would be
Then, the transducer output is linear with displacement x . This type of sensor
is normally implemented as a rotating capacitor for measuring angular displacement.
The rotating capacitor structures are also used as an output transducer for measuring
electric voltages as capacitive voltmeters.
Sound waves are in fact, pressure waves or fluctuations of the pressure of air
around the normal pressure of 1 atm.
Capacitive microphone is used to reproduce the pressure fluctuations
faithfully.
The range of frequencies involved is from 20Hz to 15KHz.
The capacitive microphone membranes should be sensitive to the fluctuating
components only and special provision must be made to equalize the pressure
on either side of the membranes. So as to prevent it from bursting due to
atmospheric pressure variations.
The constructional features of the capacitive microphone are shown in figure.
It consists of a very thin metallic membranes kept under radial tension.
The capillary tube or hole connecting the chamber behind the membranes
atmospheric is intended to equalize the pressure on either side of the
diaphragm.
The capillary path and the volume of microphone chamber act as a low pass
filter allowing the slow variations of atmospheric pressure but preventing the
MEMS pressure microsensors typically have a flexible diaphragm that deforms in the
presence of a pressure difference. The deformation is converted to an electrical signal
appearing at the sensor output. A pressure sensor can be used to sense the absolute air
pressure within the intake manifold of an automobile engine, so that the amount of
fuel required for each engine cylinder can be computed. In this example,
piezoresistors are patterned across the edges of a region where a silicon diaphragm
will be micromachined. The substrate is etched to create the diaphragm. The sensor
die is then bonded to a glass substrate, creating a sealed vacuum cavity under the
diaphragm. The die is mounted on a package, where the topside of the diaphragm is
exposed to the environment. The change in ambient pressure forces the downward
deformation of the diaphragm, resulting in a change of resistance of the piezoresistors.
On-chip electronics measure the resistance change, which causes a corresponding
voltage signal to appear at the output pin of the sensor package .
Nano – Sensors: