Calibration Manual of KR-9000
Calibration Manual of KR-9000
model eyes:
2. while holding down the key 7,and press the key 1 for several
seconds, then the screen will show “welcome! Entry pad:”, the
4.Press the key 6 and select the “3.REF Adjustment”, and then
5. Press the key 7 once, and then press the key 1,the screen will
showing as follows:
6. Press the key 6 several times to change the ITEM data to which
model eye -5.00DS for calibration. Later press the key 7 once and
measure.
8.When we have calibration all the degrees, then press the key 1
once, and then holding press key 3 for 3 seconds ,the screen will
9.Then press the key 7 once to save the data. Then restart the
instrument.