Calibration Manual of KR-9000

When we calibration this instrument, we must use the standard

model eyes:

The key number is as follows:


1.Turn on the instrument and enter into menu, then press the key

1 to enter into “time”, the interfere should be as follows:

2. while holding down the key 7,and press the key 1 for several

seconds, then the screen will show “welcome! Entry pad:”, the

interfere should be as follows:


3.Press the key 6 for more then 8 times, and then press key 7 once,

the calibration interface will show as follows:

4.Press the key 6 and select the “3.REF Adjustment”, and then

press the key 7 once, the interface will be as follows:

5. Press the key 7 once, and then press the key 1,the screen will

showing as follows:
6. Press the key 6 several times to change the ITEM data to which

we want to calibration. For example, if we want to calibration the

-5.00DS, then we press the key 6 several times to change the

ITEM:[+0.01DS ] to ITEM:[-5.00DS ], and then put the standard

model eye -5.00DS for calibration. Later press the key 7 once and

the data will be saved. the screen will showing as follows:


Notice:

Don’t calibration the first two ITEM data of ITEM:[+0.01DS ] and

ITEM:[+2.50DS ], or there will be make some problems when

measure.

7.If you want to calibration other degrees, then press key 6 to

change the ITEM data.

8.When we have calibration all the degrees, then press the key 1

once, and then holding press key 3 for 3 seconds ,the screen will

show the following interface:

9.Then press the key 7 once to save the data. Then restart the

instrument.