Portable X-Ray Unit: 1. Manufacture Company Information
Portable X-Ray Unit: 1. Manufacture Company Information
G. exposure time ; Transistor Timer combined with Condenser & Resistor for
fast and precise functioning and long durability in use
selectable at 23 steps.
Time Setting : 0.03, 0.05, 0.07, 0.08, 0.10, 0.12, 0.14, 0.16, 0.18,
0.2, 0.25, 0.30, 0.40, 0.50, 0.80, 1.00, 1.20, 1.50,
1.70, 2.00, 2.50, 3.00, 5.00 sec
1
3. The Specification for X-ray Tube in X-ray unit.
Manufactor TOSHIBA
2
4. Control panel
3
① Uper knob
② Side knob
③ Acrylic cross-hairs
Principles : Knob(①,②) driving type with two control knobs for free field adjustment in
length and width. Full field indication by visible beam. Open area 35×35㎝ at 65㎝
distance from tube focus.
Power supply to this Collimator is from transformer built in the main unit, and this
Collimator can be easily removed from the main unit.
Lamp – bulb 12V 50W
HOW TO OPERATE
4
The X-ray Machine is made for operation on input-power and V
designed for easy operation by means of just setting Time and mA/kVp
according to the radiography conditions : However, the following technical
procedure and characters should be well considered in advance for better
radiography.
2. Procedure
2.1 Line Voltage Control – Adjustment :
After connected the input Power Supply cord to the Power supply
outlet, Turn Line Voltage Control Knob.
Clockwise from 1 position, when the indicating – needle of Line Voltage
Meter feels.
Set Line Voltage Control knob until the indicating needle comes up to
the red-mark line on the line Voltage Meter.: which corresponds to
correct power supply voltage.
5
2.3 Electronic Time – Exposure time selection : Select the most proper
exposure time by means of turning the Electronic Timer knob.
: Available at 23 steps in the range of 0.03 to 5.0 seconds but not
exceeding the safety exposure time as given below:
6
Keep accessories, codes clean, and arrange and then take custody.
2) Clean x-ray machine, main power code and hand switch in order not
to have any trouble to the following use, and put them to the
designated custody box, and fix them in order they may not move
inside
3) Take custody at place where it has good ventilation and dry place
with no vibration, and be careful in order excessive force might not be
added.
4) When defect is found, don’t touch optionally, but mark discrimination,
and then request purchasing distributor or manufacturer.
5) Don’t take custody at place where chemicals are stored or gas
occurs.
6) Don’t remodel machine & tools absolutely.
4.2 Pay attention to the followings prior to use of machine & tools
1) Check connection condition of switch, indication lamp of control
panel, reservation establishment value etc, and confirm, whether
machine and tools operate exactly and safely.
2) Confirm whether connection of all code is exact and safe or not.
3) When using machine & tools not used for a long time, inevitably
confirm machine & tools are normal and have complete operation.
4.3 Pay attention to the following during use of machine & tools
1) Be careful in order not to exceed time, amount required to diagnosis
and treatment.
2) n case buzzer does not sound or lamp does not operate normally,
measure should be taken only after inquiring to manufacturing
company or distributor, and in case any defect is found to machine &
tools and patient, stop the use of machine immediately and let patient
move to the safe place.
3) When over heat lamp turns on during use, then occurrence of x-ray
stops, and when lamp turns on and off after passage of about 30
7
minutes, then use machine & tools again.
TECHNICAL GUIDE
LOCATION DISTANCE(cm) GRID kVp mA TIME SIZE
CHEST(10kg) 130 – 150 + 70 25 0.2 – 0.3 8 × 10
8
A/S MANUAL
① L.V selector
9
2.1.2 Power Trans (circuit Drawing T3) Check voltage like the following figure.
2.1.3 Power Trans (circuit Drawing T3) Check short-circuit by Multi tester like to figure.
2.1.4 Regarding to follow Figure, remove wired line like Figure No.2 and approve electricity like
Figure No.3 and 4
10
2.1.5 Check short-circuit of wired line to Figure No.5 with Multi Tester.
11
3.1.2 Dismantle Connector of Figure and
confirm soldering part.
12
3.3.2 Confirm contacting point of rotary switch of
timer.
① LV METER
D1
R1
13
5. X-RAY is not output, but needle of mA Meter moves to close to 10mA.
② mA METER
R1
5.2 As same phenomenon occurs in the event of Tube’s defect and inquire to
manufacturer or distributor.
14
6.2 To operate to Relay and check voltage of R1 and R2 with Multi Tester
15
8.2 Confirm whether Halogen lamp is defective
or not
16
17