Professional Documents
Culture Documents
CME430 Lab Report
CME430 Lab Report
Mark:
1
XIAMEN UNIVERSITY MALAYSIA
I/We hereby understand my/our work would be checked for plagiarism or other
misconduct, and the softcopy would be saved for future comparison(s).
I/We hereby confirm that all the references or sources of citations have been correctly
listed or presented and I/we clearly understand the serious consequence caused by any
intentional or unintentional misconduct.
This work is not made on any work of other students (past or present), and it has not
been submitted to any other courses or institutions before.
Date: 9-7-2020
2
XIAMEN UNIVERSITY MALAYSIA
1.0 Introduction
3
XIAMEN UNIVERSITY MALAYSIA
A=εcl
A= Absorbance
c= Molar concentration /M
2.0 Objectives
1. To measure the absorbance of the ferrocene analyte from wavelengths 350 to 700
nm .
2. To determine the unknown concentration of the sample from calibration curve.
3. To verify the relationship of absorbance and concentration by applying Beer-
Lambert's Law.
4
XIAMEN UNIVERSITY MALAYSIA
collected in the table 2. It was ensured that total 3 readings were obtained and
the average value was calculated.
5. The absorbance of sample with unknown concentration was measured in
spectrophotometer at λmax.
6. Calibration curve of absorbance vs. concentration plot was generated by using
Excel. A best fit line is draw by fitting the plots. Calibration equation is
obtained from the best fit line.
7. Concentration value for the unknown solution was predicted from the
calibration curve as shown diagram 2.
4.0 Results
λ (nm) Abs
530 0.3520
540 0.1797
550 0.0894
560 0.0441
570 0.0343
580 0.0249
590 0.0308
600 0.0119
610 0.0235
620 0.0170
630 0.0140
640 0.0224
650 0.0182
660 0.0133
670 0.0096
680 0.0162
690 0.0070
700 0.0067
5
XIAMEN UNIVERSITY MALAYSIA
λ (nm) Abs
350 0.9317
360 0.5951
370 0.8204
380 1.4355
390 2.3241
400 3.3495
410 4.1965
420 4.8185
430 5.1722
440 5.2940
450 5.1647
460 4.8403
470 4.2696
480 3.4756
490 2.6342
500 1.8073
510 1.1206
520 0.6501
4
Absorbace/ A
0
350 400 450 500 550 600 650 700 750
Wavelength /nm
Table 2 shows how the changing of concentration of sample effect on the absorbance.
λmax of Ferrocene: 440 nm
6
XIAMEN UNIVERSITY MALAYSIA
Absorbance at 0.04
M 10.5881 10.5881 10.5881 10.5881
Absorbance at 0.06
M 15.8821 15.8821 15.8821 15.8821
Absorbance at 0.08
M 21.1762 21.1762 21.1762 21.1762
Absorbance at 0.1
M 26.4702 26.4702 26.4702 26.4702
25 f(x) = 264.7 x − 0
20
Absorbace
15
10
0
0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 0.1 0.11
Concentration /M
y=mx+c
21.1762−5.2942
m= = 264.7 M −1
0.08−0.0
C = 5 ×10−3 A
y=264.7 x +3 ×10−5
7
XIAMEN UNIVERSITY MALAYSIA
x = 0.08263 M
1. Formula in eq.1
A=εcl
A (0.08−0.02)×1
ε= = =3.778 × 10−3 M-1cm-1
cl (21.1762−5.294 )
Discussion
From the diagram 1, it shows that the λmax is located at 440nm. This peak is
appeared due to the ferrocenyl group. In an experiment conducted by Fenzel et al.,
there is a peak shown in 335nm in ferrocene spectra [3]. Hence, the experiment shows
to be agree well with the other research’s studies because it supports the data
obtained.
For diagram 2, it shows that the absorbance increases linearly with concentration of
solution. The graph obtained from experiment is agree with the theoretical studies
from Beer-Lambert Law. Beer-Lambert law states that there is a linear relationship
between the concentration and the absorbance of the solution. After that, the unknown
concentration of a sample can be obtained from the calibration curve which is
0.08263M.
There are two methods to obtain the molar absorptivity coefficient. One is using Beer-
Lambert equation and another is obtained by best fit line in graph. The values
obtained by equation and graphical method are the same which is 3.778 ×10−3M-1cm-
1
.Therefore, it can be concluded that this experiment can be predicted well with Beer-
Lambert equation.
All three trial readings are the same in virtual experiment and this shows a perfectly
accuracy. However, this situation won’t be happened in the practical experiment.
There are many internal and external factors affecting on result. Firstly, the error
8
XIAMEN UNIVERSITY MALAYSIA
maybe come from the undissolved ferrocene crystal in the solution. The sample will
scatter the light more than absorb the light and the data will be very skewed. To
improve the result, the solution must be well-stirred to form a homogeneous solution.
Other error can be obtained when the volatile ferrocene solution in cuvette is
evaporated under the room temperature. The decreasing of water content in the
solution lead to increase of the concentration. It is important to close the cuvette with
a lid to avoid both contamination and evaporation. Next, the cuvette must be wiped
with clean cloth before it is measured in spectrophotometer.
Conclusion
All objectives were completed successfully. The measurement for the absorbance of
the ferrocene analyte from wavelength 350 nm to 700 nm had been performed by
visual UV spectrophotometry in visual experiment. The wavelength with maximum
absorbance or peak in other word had been obtained. It is measured to be 440 nm.
After this, the calibration curve was obtained by different peak in different
concentration. This calibration curve was important for is to predict the concentration
of unknown sample. The calculated concentration of unknown sample was 0.08263
M. Observing from the graph obtained in diagram 2, It was concluded that the
absorbance and concentration of analyte are directly proportional to each other which
is described by Beer-Lambert.
9
XIAMEN UNIVERSITY MALAYSIA
References
1.Bingham, E., Cohrssen, B. and Powell, C.H.
Encyclopaedia of Occupational Health and Safety. 4th edition,
Geneva, Switzerland: International Labour Office, 1998,14 , 15-63
10
XIAMEN UNIVERSITY MALAYSIA
APPENDIX 1
CME430 LAB REPORT MARKING RUBRICS
Score and Descriptors
Criteria 3-4 1-2 0 Marks
Title Identifies the independent and/or Too long, or does not identify the No title, or too vague. e.g.
dependent variables, not too vague or independent and/or dependent variables. "Lab report"
too long, encapsulates the keywords Contains superfluous phrases, such as
of the report well. "An investigation into..." or "An
experiment on..."
1
XIAMEN UNIVERSITY MALAYSIA
2
XIAMEN UNIVERSITY MALAYSIA