Healthcare
Silastic® BioMedical Grade ETR
Elastomers (Q7-4720, Q7-4735,
Q7-4750, Q7-4765, and Q7-4780)
Parts A & B
FEATURES Raw materials for medical device fabrication in the Healthcare
No peroxide volatiles or residues Industry
generated during cure
Non-blooming APPLICATIONS
No phthalates or other organic Silastic® BioMedical Grade ETR Elastomers (Q7-4720, Q7-4735, Q7-4750,
plasticizers Q7-4765, and Q7-4780) are heat-cured high consistency silicone rubbers
Optional post-cure to stabilize designed for use by customers fabricating medical devices, including those
properties intended for implantation in humans for less than 30 days.
Sterilizable by ETO, autoclave,
and gamma irradiation DESCRIPTION The site has been ISO registered with
High gas permeability compared to Silastic BioMedical Grade ETR BSI since 1990. These elastomers are
most thermoset elastomers and Elastomers (Q7-4720, Q7-4735, strained through a 200-mesh (75
thermoplastics Q7-4750, Q7-4765, and Q7-4780) are micron) or finer screen to remove
Pigmentable a series of two-part, enhanced-tear- particulate contamination.
resistant (ETR) silicone elastomers
COMPOSITION that consist of dimethyl and HOW TO USE
methylvinyl siloxane copolymers and These elastomers are supplied as A
Two-part silicone elastomer
reinforcing silica. The elastomers and B components that must be
exhibit a range of hardness from soft combined in equal portions by weight
(20 Shore A) to firm (80 Shore A). on a two-roll mill prior to use.
Each elastomer is supplied as a two-
component kit (Part A and Part B), Blending
equal portions (by weight) of which The recommended sequence of
must be thoroughly blended together blending the two components is to first
prior to use. The elastomer is then soften Part B on a cooled two-roll
thermally cured via addition-cure mill. Remove from the mill and soften
(platinum-cure) chemistry Part A. Add an equal portion, by
weight, of softened Part B, and cross-
MANUFACTURING blend the components until thoroughly
ENVIRONMENT mixed. The temperature of the blended
Silastic BioMedical Grade ETR material must be kept as low as
Elastomers are manufactured under possible to give maximum table life.
strict quality-control guidelines. The
Dow Corning Healthcare Industries Blend only the amount that will be
Materials Site (HIMS) in Hemlock, used in 3 to 4 hours. If carefully
MI, is dedicated to the production of wrapped, blended material may be
silicone materials for healthcare stored in a freezer (<0°C/32°F) for at
applications. It is registered with the least 7 days. Material stored in this
FDA (CFN 1816403) as a Drug manner should be warmed to room
Establishment. The site quality temperature before unwrapping to
system is based on principles of avoid condensation on the elastomer.
current Good Manufacturing Practices Condensation may cause voids in
for both Bulk Pharmaceutical Products molded or extruded parts.
and Medical Devices.
Cure REGULATORY STATUS local Dow Corning Global
Cure of the blended elastomer is Silastic BioMedical Grade ETR Connection.
accelerated by heat. The Elastomers, when fully cured and
recommended cure conditions for a washed, meet the requirements of USABLE LIFE AND
cross-section up to 1.905mm (0.075 FDA regulation 21CFR177.2600, STORAGE
inch) thicknesses are 10 minutes at covering rubber articles intended When stored at or below ambient
116°C (241°F). Proportionally for repeated food contact. temperature in the original
more time is required to cure
unopened containers, this product
thicker cross-sections. The cure Master Files for Silastic has a usable life of 12 months from
profiles for these products can be BioMedical Grade ETR Elastomers the date of production.
found in Figures 1and 2. (Q7-4735, Q7-4750, Q7-4765, and
Q7-4780) have been filed with the PACKAGING
CAUTION: The cure may be U.S. Food and Drug
inhibited by traces of amines, INFORMATION
Administration. Customers
sulfur, nitrogen oxide, organotin Silastic BioMedical Grade ETR
interested in authorization to
compounds and carbon monoxide. Elastomers (Q7-4720,
reference the file must contact
Because organic rubbers often Q7-4735, Q7-4750, Q7-4765, and
Dow Corning Corporation.
contain these substances, they Q7-4780) are supplied in 13.6kg
should not come in contact with the and 408.2kg (30 lb and 900 lb) kits,
uncured elastomer. Catalyst IMPORTANT each containing equal portions of
residues from silicone RTV INFORMATION Part A and B. Each component is
elastomers and peroxide-cured THE USER'S sealed in a polyethylene bundle.
silicone elastomers may also inhibit ATTENTION IS
the cure. Samples are available in 908g
IN PARTICULAR DRAWN (2 lb) kits.
TO THE FOLLOWING
All equipment should be
thoroughly cleaned at the end of STATEMENT: HEALTH AND
each use to avoid a build-up of It is the User's responsibility to
ensure the safety and efficacy of
ENVIRONMENTAL
cured stock. The residue may result INFORMATION
in crumbs of elastomer being these materials for all intended
uses. While this Material has To support Customers in their
picked up by the next lot, causing product safety needs, Dow Corning
imperfections. passed screening tests that are
applicable to products intended for has an extensive Product
implantation for up to 29 days, Stewardship organization and a
Post-curing Dow Corning makes no end-use team of Product Safety and
Because these materials vulcanize representation based on such Regulatory Compliance (PS&RC)
via addition-cure, no residues are testing. Nor does Dow Corning specialists available in each area.
present and post-cure is not make any representation
required. The user must confirm concerning the suitability of this For further information, please see
that press molding or short oven product for applications of greater our Web site, dowcorning.com or
cures are suitable for any specific than 29 days of implantation in consult your local Dow Corning
application. The principal volatile the human body. representative.
components of post-curing are
short-chained polydimethylsiloxane LIMITED WARRANTY
fluids and water vapor. See the HANDLING
Typical Properties chart for specific PRECAUTIONS INFORMATION –
post-cure information. Product safety information required PLEASE READ
for safe use is not included. Before CAREFULLY
BIOCOMPATIBILITY handling, read product and safety The information contained herein is
The results of selected data sheets and container labels for offered in good faith and is
biocompatibility tests are shown in safe use, physical and health hazard believed to be accurate. However,
Table I. Elastomer samples were information. The material safety because conditions and methods of
sterilized by autoclaving before data sheet is available on the use of our products are beyond our
testing. Toxicological Summaries Dow Corning website at control, this information should not
are available upon request. dowcorning.com. You can also be used in substitution for
obtain a copy from your local customer’s tests to ensure that our
Dow Corning sales representative products are safe, effective, and
or Distributor or by calling your fully satisfactory for the intended
end use. Suggestions of use shall
Silastic® BioMedical Grade ETR Elastomers (Q7-4720, Q7-4735, Q7-4750, Q7-4765, and Q7-4780) Parts A & B Dow Corning and Silastic are a registered trademark of Dow Corning Corporation.
We help you invent the future is a trademark of Dow Corning Corporation.
2011, February 05 XIAMETER is a registered trademark of Dow Corning Corporation.
Form No. 52-1031C-01 ©2002 - 2011 Dow Corning Corporation. All rights reserved.
2
not be taken as inducements to
infringe any patent.
DOW CORNING
SPECIFICALLY DISCLAIMS
ANY OTHER EXPRESS OR
IMPLIED WARRANTY OF
FITNESS FOR A
PARTICULAR PURPOSE OR
MERCHANTABILITY.
dowcorning.com
Silastic® BioMedical Grade ETR Elastomers (Q7-4720, Q7-4735, Q7-4750, Q7-4765, and Q7-4780) Parts A & B Dow Corning and Silastic are a registered trademark of Dow Corning Corporation.
We help you invent the future is a trademark of Dow Corning Corporation.
2011, February 05 XIAMETER is a registered trademark of Dow Corning Corporation.
Form No. 52-1031C-01 ©2002 - 2011 Dow Corning Corporation. All rights reserved.
3
Table 1: Results of selected Biocompatibility Tests for Silastic BioMedical Grade ETR Elastomers
(Q7-4720, Q7-4735, Q7-4750, Q7-4765, and Q7-4780)
Silastic® BioMedical Grade ETR Elastomers (Q7-4720, Q7-4735, Q7-4750, Q7-4765, and Q7-4780) Parts A & B Dow Corning and Silastic are a registered trademark of Dow Corning Corporation.
We help you invent the future is a trademark of Dow Corning Corporation.
2011, February 05 XIAMETER is a registered trademark of Dow Corning Corporation.
Form No. 52-1031C-01 ©2002 - 2011 Dow Corning Corporation. All rights reserved.
4
Table 2: TYPICAL PROPERTIES Continued
Figure 1: MDR 2000 cure testing of Silastic BioMedical Grade ETR Elastomers
Figure 2: ODR 2000 cure testing of Silastic BioMedical Grade ETR Elastomers
Silastic® BioMedical Grade ETR Elastomers (Q7-4720, Q7-4735, Q7-4750, Q7-4765, and Q7-4780) Parts A & B Dow Corning and Silastic are a registered trademark of Dow Corning Corporation.
We help you invent the future is a trademark of Dow Corning Corporation.
2011, February 05 XIAMETER is a registered trademark of Dow Corning Corporation.
Form No. 52-1031C-01 ©2002 - 2011 Dow Corning Corporation. All rights reserved.