SPESIFIKASI TEKNIS

NAMA
NO SPESIFIKASI TEKNIS JUMLAH Franco
INSTRUMENT
1 Furnace a. Volume : min. 9 L 2 Aceh dan
b o pekanbaru
Temperature min. 1300 C
c Chamber furnace with brick or fiber
insulation
d Completely Ready for use
2 X-ray a. Included in delivery: 1 Medan
Diffraction - Radiation Safety Enclosure
(XRD)
- Base cabinet
- NIST Standard Reference Material
SRM1976 (actual version at time of
order)
b. Main Unit:
- Minimum 1 kW generator
- 4-axes motor driver board
- Stability : <0.005% for high voltage
and current
- SNAP LOCK Change of Optics
- Real Time Status Display
- Allignment Guarantee
c. Vertical goniometer, Theta/Theta or
Theta/2Theta geometry:
- Universal detector mount without slit
- Spacer for universal detector mount,
measurement height 150mm
- Angular range :360° (without
accessories)
- Max. Useable angular range :-110° to
168° or better
- Accuracy : 0.005° or better
- Reproducibility : 0.0002° or better
d. Tube:
- The X-Ray Tube produce by Factory
e. High Speed Energy Dispersive
Detector:
- Maximum global count rate:
>100,000,000 cps
- Energy resolution <380 eV for Cu
radiation at 25°C, reducing
background from luorescence
samples
Motorized divergence slit assembly:
Cu absorber 0.1mm
Ni filter for Cu radiation , 0.02mm
f. Dynamic Beam Optimization
1.) Motorized anti-scatter screen:
- The Motorized Anti-Scatter Screen
is a fully software-controlled device
for effective suppression of
instrument background, most
importantly air-scatter at low angles
2q.
2.) Motorized divergence slit assembly:
- Computer-controlled stepper motor
driven slit.
g. Personal Computer:
- Computer, international version,
international operating system
- PC Assembly built in from Factory
SOFTWARE :
- DIFFRACTION MEASUREMENT
CENTER
- DIFFRACTION.EVALUATION (2
licenses)
- WIBU Dongle for DIFFRACTION
SUITE
- ICDD PDF4+ Full File, min 3 years
Master License
- Main Software for XRD System (2
licenses):
- Main Software for XRD System is a
generalized software for analysis of
X-ray and neutron powder and single
crystal diffraction data. Software
integrates all currently employed
profile fitting techniques and related
applications, including
- Single Line Fitting up to Whole
Powder Pattern Fitting
- Whole Powder Pattern Decomposition
- Indexing
- Structure determination and
refinement
- Microstructure analysis
- Quantitative phase analysis
Miscellaneous:
- Powder pattern calculation
- support of crystal structure data via
CIF (IUCr Crystallographic
Information File)
- Support of ICDD PDF and user
defined d-I data via DIF files
- Support of ShelX HKL files
- Fully automated operation possible
COD Database Software for difraction
data processing:
- Limited to inorganics
- Currently 26.522 entries (6.218 non-
ambient)
- All crystallographic constraints taken
into account: Lattice parameters,
special positions (!)
h. Installation:
- Table for PC
- 10 KVA UPS for XRD
- Water chiller if needed
- Comissioning
- Certification calibration of goniometer
alignment (after commisioning of
instrument)
Includes:
a. 10 set specimen holder rings and 1 set
silicon crystal specimen holder ring,
PMMA (Polymethyl methacrylate)
b. Surveymeter (2pcs)
c. TLD (4 pcs)
d. Personal dosimetry (2 pcs)
e. Completely Ready for use

3 Oven a. Volume : min. 55 L 2 Aceh dan


b. pekanbaru
Temperature range min. up to 300 oC
c. Door and internal made from stainless
steel
d. Completely Ready for use

4 Bomb a. Bomb Type & Bucket : Fixed Bomb & 1 Medan


Calorimeter bucket design
b. Precissions : 0,10% RSD or better (1g
benzoic acid NBS39i)
c. Bucket filling : automatic
d. Oxygen filling : automatic
e. Bomb washing : automatic
f. Temperature resolution : 0,0001 oC or
better
g. Touch screen operation or PC
h. Methode : isoperibol
i. Measuring range : min 8000 cal
j. Reproducibility : 0,05% RSD or better
k. Test per hour : min 6 sample
l. Number of Vessel : min 4
Includes:
a. Printer