Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Fundamentals of

Mass Flow Control


Critical Terminology and Operation
Principles for Gas and Liquid MFCs

A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of a particular gas or liquid.
These devices are essential to most thermal and dry etching processes. Advanced Energy®’s Aera® MFCs provide the most
precise flow control at the most cost-effective price.

This paper defines critical flow control terminology, describes basic MFC operation, and illustrates Aera MFC design
solutions to common mass flow control challenges. The last section introduces the liquid mass flow controller (LMFC)
and describes design solutions specific to flow control of liquids.

Elements of an MFC Bypass Control Valve


Base Also known as the flow splitter, the The control valve establishes the
The base provides the platform bypass maintains a constant ratio flow of gas by responding to a signal
on which all other components of gas flow through the sensor and that compares the actual flow to
of the MFC are mounted and main flow path, dividing the gas the set point. Actuators driving
contains the channels that form stream precisely over the entire the control valve in Aera MFCs are
the main flow path of the gas. The calibrated flow range. As a result, either piezoelectric, solenoid, or
base of an Aera MFC is constructed the total flow can be determined by thermal actuators, depending
of type 316L stainless steel and measuring just the portion of gas on the model.
is precision finished. Metal or that passes through the sensor.
elastomer seals, depending on the Printed Circuit Board
application, are provided between The printed circuit board is
the base and other components. designed for optimum stability.
Aera MFCs use the minimum
Sensor number of electronic components
The thermal sensor is designed for and only the highest-reliability
quick response, long-term stability, components available.
and high reliability. The sensor tube
in an Aera mass flow product has
a very small diameter and mass to
ensure the fastest response to any
change in gas flow conditions.

Figure 1. Inside an Aera® MFC


advanc e d e n e r g y

Operating Principle The bypass forces a constant


The heart of a mass flow controller proportion of the incoming gas to
is a thermal sensor. It consists be fed into the sensor. The gas flow
of a small bore tube with two through the sensor tube causes
resistance-thermometer elements heat to be transferred from the
wound around the outside of the upstream resistance-thermometer
tube. The sensor tube is heated by element to the downstream
applying an electric current to the resistance-thermometer element.
elements. A constant proportion of This temperature differential is No Flow
gas flows through the sensor tube, linearized and amplified into a 0 to Condition
and the cooling effect creates a 5 V flow output signal by means of a
temperature differential between bridge circuit. The output signal is
the two elements. The change in the compared with the external set point Gas Flow

resistance due to the temperature signal to the mass flow controller.


differential is measured as an The error signal that results from
electrical signal. comparing the output signal with Figure 3. Sensor temperature profile
the set point signal directs the
The temperature differential created control valve to open or close to
between the elements is dependent maintain a constant flow at the Performance
on the mass flow of the gas and is a set point level. and Reliability
function of its density, specific heat, MFC manufacturers go to great
and flow rate. Mass flow is normally lengths to explain why their
displayed in terms of volume of products should outperform others,
the gas either in standard cubic often focusing on a few design
centimeters per minute (sccm) or features. But it is not enough
in standard liters per minute (slm). to define MFC performance
The electronics of a mass flow on the basis of a few parameters—
controller convert mass flow into today’s critical processes demand
volume flow at standard conditions MFCs that deliver outstanding
of 0°C (32°F) and 1 atmosphere. performance and reliability.
Because the volume of 1 mole of The primary factors include:
an ideal gas at 0° C (32°F) and 1
atmosphere occupies 22.4 liters, a set Accuracy
point of 22.4 slm will cause 1 mole Accuracy refers to the difference
of gas to flow during 1 minute. between the actual flow of an MFC
and that of a primary standard at
Figure 2. Operational diagram any set point.

2
w h i t e p aper

Repeatability Pressure Change Response Some MFC manufacturers,


Another primary factor is the Pressure change response is the particularly those with zero drift
repeatability of actual flow for an time that it takes for actual flow problems, offer an option of
MFC or from one MFC to another to stabilize after a sudden change automatic zeroing. Auto zero does
at any set point. in gas input pressure. indeed correct the shift in the
zero calibration point, but it does
Linearity not correct the shift of the entire
Linearity is the straightness of the It is not enough to define MFC calibration curve. Admittedly,
curve of actual flow vs. set point, in the shift in the remainder of the
performance on the basis of a
other words, accuracy over the entire calibration curve is less severe than
flow range. few parameters—today’s critical that for the zero point, but it still
represents a degradation of accuracy
processes demand MFCs that
Calibration Drift and linearity.
Calibration drift is the change deliver outstanding performance
in the curve of actual flow vs. set and reliability. On the other hand, zero drift has
point due to aging effects of some been essentially eliminated in Aera
of the component parts that make flow products. The typical zero
up an MFC. drift of our MFCs is less than 0.5%
Temperature Change Effects of full-scale flow over a period of
Stability This parameter concerns the one year. This is the result of using
Stability refers to the ability of an stability of flow during ambient the highest-quality sensor wire and
MFC to maintain stable flow levels temperature variations. electrical components on the PC
through short-term effects such as board. Perhaps more importantly,
pressure and temperature changes, Zero Drift the sensors and the assembled MFCs
and through long-term effects such Zero drift is the most common are subjected to extensive burn-in
as aging of the component parts. complaint of MFC users. It is procedures and stringent, multiple
a time-dependent shifting of the QC inspections to screen out all
Response or Settling Time zero calibration point from its marginal components and
The time that it takes for actual flow original zero value to an offset assembled MFCs.
to stabilize after a set point change is value, and is generally caused by
another critical parameter. aging effects of various electrical In this manner, we do not hide the
components on the PC board as well zero drift problem; we prevent it.
Over-shoot and Under-shoot as by aging of the sensor windings.
Over-shoot and under-shoot refer to The aging phenomenon that results
any spike or dip, respectively, in the in zero drift not only causes a shift
response curve of actual flow vs. time. in the zero calibration point but
also causes a shift of the entire
curve of control voltage vs. flow.

3
advanc e d e n e r g y

Liquid Mass Flow Instruments known as vaporizer Calibration drift is common in


Controllers controllers are typically used to vaporizer controllers that use a
Liquid sources have been used in control vapor flow generated using ratio detector, caused by cumulative
semiconductor processing for as long the carrier gas technique. One type contamination of one of the thermal
as the industry has existed. In the of vaporizer controller combines an conductivity cells as a result of
industry’s early days, liquids were MFC with a ratio detector consisting reaction between the source material
often chosen as source materials of thermal conductivity cells that and finite concentrations of oxygen
because the proper techniques measure the ratio of vapor to carrier or moisture in the carrier gas. There
for handling toxic and corrosive gas. The ratio detector provides are also accuracy and repeatability
gases were unfamiliar to many a signal to the MFC to adjust the limitations. For example, the
development engineers. Gas leaks carrier gas flow to deliver a constant variable carrier gas flow rate
were the norm, with frequent amount of source vapor per unit will at least slightly affect the
harm to workers and equipment. time, independent of all variables. reproducibility of the mainstream
Today, handling procedures are flow within a process chamber.
well understood, and leaks are Another type of vaporizer controller
generally not a major problem. is a complex system that consists When vapor is generated by heating
Nevertheless, liquid sources are now of an MFC, a minicomputer, and a liquid source, MFCs are used to
preferred over some gaseous sources devices for measuring the major control the flow of heated vapor. Re-
because of the appropriateness of variables: bubbler temperature and condensation is prevented by heating
liquid reaction characteristics for pressure. All other variables are the MFC and the lines leading to
particular applications. assumed to be of minor importance. and from it. For TEOS, a widely
The minicomputer calculates the used liquid source for medium-
The most common method for vapor pressure of the liquid source temperature oxide deposition,
introducing liquid source materials based on the temperature and re-condensation is prevented by
into reactors and furnace tubes pressure measurements and provides confining a “high-temperature”
is to first vaporize the liquid and a signal to the MFC to adjust the MFC and associated plumbing
then pass the vapors through mass carrier flow to deliver a constant within an isothermal enclosure
flow instruments that have been amount of source vapor per unit heated to near 80°C (176°F). This
developed specifically for controlling time. The degree of saturation of configuration is sometimes known
vapor flow. Liquid sources can be the carrier gas may vary depending as a thermal vaporizer system. To
vaporized either by bubbling a on bubbler design, liquid level, and withstand the elevated temperature,
carrier gas through the liquid or by bubble size. the flow components and PC board
heating the liquid to generate an of the MFC are designed as separate
adequate vapor pressure. One major modules, with the flow component
constraint that is common to all module located in the enclosure
vaporization methods is that some and the PC board module mounted
liquid sources have too low a vapor external to the enclosure.
pressure to generate adequate vapor
pressure at room temperature or
moderately elevated temperatures.

4
A thermal vaporizer system is more An LMFC consists of the same flow
accurate and repeatable than the components found in traditional
carrier gas/bubbler technique, but MFCs. The major difference is that
it requires a very large, expensive each component must be designed to
system to enclose all of the prevent the most prevalent problem
requisite hardware and maintain with LMFCs—bubble formation
an isothermal environment. Re- and retention. In an LMFC, a liquid
condensation of the vapor can be a source can gain heat from the sensor
chronic problem with this technique tube, which operates at an elevated
unless stringent precautions are temperature, and from contact
taken to heat all lines leading from with the internal parts of a control
the thermal vaporizer system to the valve if it is of a type that generates
reactor or furnace tube. heat. If bubbles were consequently
generated and retained, their Figure 4. Aera® LX-1200/1200C
A New Method for Flow Control presence would adversely affect series LMFC
A mass flow control instrument accuracy and repeatability and
for metering very low flows of could cause malfunction of the
liquid source materials into LMFC. The present design uses a
process chambers is now available, vertically configured flow path that
offering considerable advantages prevents retention of bubbles. The
over vaporizer techniques. This inlet fitting is at a lower level than
instrument is a liquid mass flow the outlet fitting, and the sensor is
controller (LMFC) and is an S-shaped rather than U-shaped as in
adaptation of basic mass flow control conventional MFCs. Additionally, the
technology. The LMFC provides control valve is of a piezoelectric/
accuracy and repeatability, as well as diaphragm type, which has a much
size and cost advantages. cooler operating temperature
than other types of control valves
and permits no liquid to enter the
internal region of the valve. To
prevent bubble formation in the
heated sensor tube, the LMFC is
limited to liquids that have a boiling
point of at least 65°C (149°F).

Advanced Energy Industries, Inc. • 1625 Sharp Point Drive • Fort Collins, Colorado 80525 © Advanced Energy Industries, Inc. 2005
All rights reserved. Printed in U.S.A.
T: 800.446.9167 or 970.221.4670 • F: 970.221.5583 • support@aei.com • www.advanced-energy.com SL-MFCFUND-270-02 0M 6/05

United Kingdom Germany Korea Japan Taiwan China


T: 44.1869.320022 T: 49.711.779270 T: 82.31.777.9191 T: 81.3.32351511 T: 886.2.82215599 T: 86.21.58579011
F: 44.1869.325004 F: 49.711.7778700 F: 82.31.777.9195 F: 81.3.32353580 F: 886.2.82215050 F: 86.21.58579003

Advanced Energy ®, AE®, and Aera® are trademarks of Advanced Energy Industries, Inc.

You might also like