Download as pdf or txt
Download as pdf or txt
You are on page 1of 14

Error list

Dräger Medizintechnik

General

The error numbers shown on the EL display refer to a faulty assembly.

The abbreviations shown on the EL display have the following meaning:

Abbreviation on EL Faulty assembly Location


display
p_sens WT2 Sensor PCB Sensor unit
p_cont WT2 Controller PCB Display housing
p_actu WT2 Actuator PCB E-box
p_matr WT2 Mattress PCB E-box
soft Software error
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 1 of 14


Caleo error list
Dräger Medizintechnik

• Click on an error code to jump to relevant description of the error code.



Micro-controller errors cause a direct program abort. Other errors cause a
program abort only after they have occurred several times. All errors and
their occurrence are written in the error log.

Error code numbers shown on the EL display

1 2 3 5 6

7 8 9 11 12

13 14 15 16

22 23 24

25 27 28 29 30

31 32 35 36

37 38 39 40 41 42

43 46 47 48

49 50 51 52 53 54

55 56 57 58 59 60

61 62 64 65 66

67 68 69 70 71 72
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 2 of 14


No. Plain text Cause
Dräger Medizintechnik

Remedy

1 Sensor drift too The measuring • Check the WT2 Sensor PCB using the
high paths (air and skin) DrägerService Mode, see diagnosis
have drifted out of "Sensor PCB".
the tolerance range.
(Note: The error is • If necessary, replace the
detected during the WT2 Sensor PCB, see Repair
10-minute test and Instructions "Replacing the WT2
leads to INOP.) Sensor PCB".

2 Comm. error on Communication • Check cable connection between WT2


CAN bus between WT2 Actuator PCB and WT2 Controller PCB;
Controller PCB, replace cable connection, if necessary.
WT2 Actuator PCB,
and WT2 Sensor • Check cable connection between WT2
PCB is faulty. (Note: Actuator PCB and WT2 sensor PCB;
The error is replace cable connection, if necessary.
detected during • Replace WT2 Controller PCB, if
operation.) necessary.
• Replace WT2 Actuator PCB, if
necessary.
• Replace WT2 Sensor PCB, if necessary.
3 36 degree ref. The "36 degree • Check the WT2 Sensor PCB using the
faulty test" with the DrägerService Mode, see diagnosis
reference "Sensor PCB".
resistance failed.
(Note: The error is • If necessary, replace the
detected during the WT2 Sensor PCB, see Repair
10-minute test and Instructions "Replacing the WT2
leads to INOP.) Sensor PCB".

5 Air excess temp. The excess • Check the WT2 Sensor PCB using the
test failed temperature circuit DrägerService Mode, see diagnosis
test with a reference "Sensor PCB".
resistance failed.
(Note: The error is • If necessary, replace the
detected during the WT2 Sensor PCB, see Repair
10-minute test.) Instructions "Replacing the WT2
Sensor PCB".
6 Invalid The configuration is • Switch device off and on and re-
configuration invalid. (Note: The configure device.
For internal use only. Copyright reserved.

error is detected
during the power-on • Replace the WT2 Controller PCB, see
test and leads to Repair Instructions "Replacing the
INOP.) WT2 Controller PCB".
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 3 of 14


No.
7
Plain text
40 degree ref.
Cause
The "40 degree
Dräger Medizintechnik

Remedy

• Check the WT2 Sensor PCB using the
faulty test" with the DrägerService Mode, see diagnosis
reference "Sensor PCB".
resistance failed.
(Note: The error is • If necessary, replace the
detected during the WT2 Sensor PCB, see Repair
10-minute test and Instructions "Replacing the WT2
leads to INOP.) Sensor PCB".

8 AirSafetyRelay The air heater safety • Check the WT2 Actuator PCB using the
blocked relay or the control DrägerService Mode, see diagnosis
of the air heater "Digital signals".
safety relay is faulty.
(Note: The error is • If necessary, replace the
detected during the WT2 Actuator PCB, see Repair
10-minute test and Instructions "Replacing the
leads to INOP.) WT2 Actuator PCB".

9 AirSafetyRelay The control of the • Check the WT2 Actuator PCB using the
Feedb. error air heater safety DrägerService Mode, see diagnosis
relay sends no "Digital signals".
feedback signal.
(Note: The error is • If necessary, replace the
detected during the WT2 Actuator PCB, see Repair
10-minute test and Instructions "Replacing the
leads to INOP.) WT2 Actuator PCB".

11 Air heater off The air heater is off, • Check fuse on WT2 Actuator PCB and
but should be on. replace fuse, if necessary.
(Note: The error is
detected during • Check air heater thermostat and replace
operation and leads thermostat, if necessary.
to INOP.) • Check air heater and replace air heater,
if necessary.
• Check the WT2 Actuator PCB using the
DrägerService Mode, see diagnosis
"Actuator PCB".
• If necessary, replace the
WT2 Actuator PCB, see Repair
Instructions "Replacing the
WT2 Actuator PCB".
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 4 of 14


No.
12
Plain text
Air heater on
Cause
The air heater is on,
Dräger Medizintechnik

Remedy

• Check the WT2 Actuator PCB using the
but should be off. DrägerService Mode, see diagnosis
(Note: The error is "Actuator PCB".
detected during
operation and leads • If necessary, replace the
to INOP.) WT2 Actuator PCB, see Repair
Instructions "Replacing the
WT2 Actuator PCB".
13 HumiditySafety The humidifier • Check the WT2 Actuator PCB using the
Relay blocked safety relay or the DrägerService Mode, see diagnosis
control of the "Digital signals".
humidifier safety
relay is faulty. (Note: • If necessary, replace the
The error is WT2 Actuator PCB, see Repair
detected during the Instructions "Replacing the
10-minute test.) WT2 Actuator PCB".

14 HumiditySafety The control of the • Check the WT2 Actuator PCB using the
Relay Feedb. humidifier safety DrägerService Mode, see diagnosis
error relay sends no "Digital signals".
feedback signal.
(Note: The error is • If necessary, replace the
detected during the WT2 Actuator PCB, see Repair
10-minute test.) Instructions "Replacing the
WT2 Actuator PCB".
15 Boiler off The humidifier • Check fuse on WT2 Actuator PCB and
heater is off, but replace fuse, if necessary.
should be on. (Note:
The error is • Check humidifier thermostat and replace
detected during thermostat, if necessary.
operation.) • Check humidifier and replace humidifier,
if necessary.
• Check the WT2 Actuator PCB using the
DrägerService Mode, see diagnosis
"Actuator PCB".
• If necessary, replace the
WT2 Actuator PCB, see Repair
Instructions "Replacing the
WT2 Actuator PCB".
16 Boiler on The humidifier • Check the WT2 Actuator PCB using the
For internal use only. Copyright reserved.

heater is on, but DrägerService Mode, see diagnosis


should be off. (Note: "Actuator PCB".
The error is
detected during • If necessary, replace the
operation.) WT2 Actuator PCB, see Repair
Instructions "Replacing the
R6150000FL1.fm 08.02.01

WT2 Actuator PCB".

6150.000 Caleo 10/2000 Repair Instructions Page 5 of 14


No.
22
Plain text
5V actuator
Cause
The 5-V voltage on
Dräger Medizintechnik

Remedy

• Check the WT2 Actuator PCB using the
error the DrägerService Mode, see diagnosis
WT2 Actuator PCB "Operating voltages".
is incorrect. (Note:
The error is • If necessary, replace the
detected during WT2 Actuator PCB, see Repair
operation and leads Instructions "Replacing the
to INOP.) WT2 Actuator PCB".

23 12V actuator The 12-V voltage on • Check the WT2 Actuator PCB using the
error the DrägerService Mode, see diagnosis
WT2 Actuator PCB "Operating voltages".
is incorrect. (Note:
The error is • If necessary, replace the
detected during WT2 Actuator PCB, see Repair
operation and leads Instructions "Replacing the
to INOP.) WT2 Actuator PCB".
• Replace the WT2 Controller PCB, see
Repair Instructions "Replacing the
WT2 Controller PCB".
24 5V actuator The 5-V voltage for • Check the WT2 Actuator PCB using the
(sensor) error the DrägerService Mode, see diagnosis
WT2 Sensor PCB "Operating voltages".
is faulty. (Note: The
error is detected • If necessary, replace the
during operation WT2 Actuator PCB or the
and leads to INOP.) WT2 Sensor PCB, see Repair
Instructions "Replacing the
WT2 Actuator PCB"or "Replacing the
WT2 Sensor PCB".
25 24V actuator The 24-V voltage on • Check the WT2 Actuator PCB using the
error the DrägerService Mode, see diagnosis
WT2 Actuator PCB "Operating voltages".
is incorrect. (Note:
The error is • If necessary, replace the
detected during WT2 Actuator PCB, see Repair
operation and leads Instructions "WT2 Actuator PCB".
to INOP.) • Check DC motor for height adjustment
and replace, if necessary.
27 O2 valve The O2 valve • Check the WT2 Actuator PCB using the
voltage voltage on the DrägerService Mode, see diagnosis
For internal use only. Copyright reserved.

WT2 Actuator PCB "Digital signals".


is incorrect. (Note:
The error is • If necessary, replace the
detected during WT2 Actuator PCB, see Repair
operation.) Instructions "Replacing the
R6150000FL1.fm 08.02.01

WT2 Actuator PCB".

6150.000 Caleo 10/2000 Repair Instructions Page 6 of 14


No.
28
Plain text
5V sensor error
Cause
The 5-V voltage of
Dräger Medizintechnik

Remedy

• Check the WT2 Sensor PCB using the
the sensor unit is DrägerService Mode, see diagnosis
incorrect. (Note: "Operating voltages".
The error is
detected during • If necessary, replace the
operation and leads WT2 Sensor PCB or the
to INOP.) WT2 Actuator PCB, see Repair
Instructions "Replacing the WT2
Sensor PCB" or "WT2 Actuator PCB".
29 Temp ref 1 (Note: The error is • Check the WT2 Actuator PCB using the
actuator detected during the DrägerService Mode, see diagnosis
10-minute test and "Actuator PCB".
leads to INOP.)
• If necessary, replace the
WT2 Actuator PCB, see Repair
Instructions "Replacing the
WT2 Actuator PCB".
30 Watchdog error The watchdog time • Replace the WT2 Controller PCB, see
has been exceeded. Repair Instructions "Replacing the
(Note: The error is WT2 Controller PCB".
detected during the
10-minute test and
leads to INOP.)
31 Short-circuit Error in analog • Check the WT2 Sensor PCB using the
multiplexer electronics. (Note: DrägerService Mode, see diagnosis
The error is "Sensor PCB".
detected during
operation and leads • If necessary, replace the
to INOP.) WT2 Sensor PCB, see Repair
Instructions "Replacing the WT2
Sensor PCB".
32 Short-circuit The evaluation • Check the WT2 Sensor PCB using the
hybrid electronics DrägerService Mode, see diagnosis
detected a short- "Sensor PCB".
circuit in the
measuring hybrids. • If necessary, replace the
(Note: The error is WT2 Sensor PCB, see Repair
detected during the Instructions "Replacing the WT2
10-minute test.) Sensor PCB".

35 Inop feedb. The INOP feedback • Replace the WT2 Controller PCB, see
error signal is incorrect. Repair Instructions "Replacing the
For internal use only. Copyright reserved.

(Note: The error is WT2 Controller PCB".


detected during
operation and leads
to INOP.)
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 7 of 14


No.
36
Plain text
Stop feedb.
Cause
The stop feedback
Dräger Medizintechnik

Remedy

• Replace the WT2 Controller PCB, see
error signal is incorrect. Repair Instructions "Replacing the
(Note: The error is WT2 Controller PCB".
detected during
operation and leads
to INOP.)
37 Buzzer feedb. The buzzer control • Replace the WT2 Controller PCB, see
error feedback signal is Repair Instructions "Replacing the
incorrect. (Note: WT2 Controller PCB".
The error is
detected during
operation and leads
to INOP.)
38 Restoration Communication • Check cable connections between
CAN between WT2 Actuator PCB and
communication WT2 Controller PC WT2 Controller PCB; replace cable
B, WT2 Actuator connections, if necessary.
PCB, and
WT2 Sensor PCB • Check cable connections between
cannot be restored. WT2 Actuator PCB and
(Note: The error is WT2 Sensor PCB; replace cable
detected during connections, if necessary.
operation and leads • Replace WT2 Controller PCB, if
to INOP.) necessary.
• Replace WT2 Actuator PCB, if
necessary.
• Replace WT2 Sensor PCB, if necessary.
39 Motor control No current is • Check motor using DrägerService Mode,
measured when the see calibration "Height adjustment or Tilt
motor (height/tilt adjustment".
adjustment) is
switched on. (Note: • Replace WT2 Actuator PCB, if
The error is necessary.
detected during • Replace motor, if necessary.
operation.)
40 Variables Software problem • Switch device off and on again. If this
conflict or external does not help, proceed as follows:
interference. (Note:
The error is • Replace the WT2 Controller PCB, see
For internal use only. Copyright reserved.

detected during Repair Instructions "Replacing the


operation and leads WT2 Controller PCB".
to INOP.)
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 8 of 14


No.
41
Plain text
Program error
Cause
Software problem
Dräger Medizintechnik

Remedy

• Switch device off and on again. If this
or external does not help, proceed as follows:
interference. (Note:
The error is • Replace the WT2 Controller PCB, see
detected during Repair Instructions "Replacing the
operation and leads WT2 Controller PCB".
to INOP.)
42 EEPROM error The EEPROM is • Check the WT2 Controller PCB using
defective. (Note: the Repair Instructions "Checking the
The error is WT2 Controller PCB during normal
detected during operation".
operation and leads
to INOP.) • If necessary, replace the WT2 Controller
PCB, see Repair Instructions "Replacing
the WT2 Controller PCB".
43 Humidity sensor The humidity sensor • Check the WT2 Sensor PCB using the
error has generated an DrägerService Mode, see diagnosis
error. (Note: The "Sensor PCB".
error is detected
during operation.) • If necessary, replace the
WT2 Sensor PCB, see Repair
Instructions "Replacing the WT2
Sensor PCB".
• Replace humidity sensor, if necessary.
46 Air sensor 1 Short-circuit in the • Check the WT2 Sensor PCB using the
short-circuit air temperature DrägerService Mode, see diagnosis
sensor 1 on the "Sensor PCB".
WT2 Sensor PCB.
(Note: The error is • If necessary, replace the
detected during WT2 Sensor PCB, see Repair
operation.) Instructions "Replacing the WT2
Sensor PCB".
47 Air sensor 2 Short-circuit in the • Check the WT2 Sensor PCB using the
short-circuit air temperature DrägerService Mode, see diagnosis
sensor 2 on the "Sensor PCB".
WT2 Sensor PCB.
(Note: The error is • If necessary, replace the
detected during WT2 Sensor PCB, see Repair
operation.) Instructions "Replacing the WT2
Sensor PCB".
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 9 of 14


No.
48
Plain text
Air sensor 1
Cause
Breakdown in the
Dräger Medizintechnik

Remedy

• Check the WT2 Sensor PCB using the
breakdown air temperature DrägerService Mode, see diagnosis
sensor 1 on the "Sensor PCB".
WT2 Sensor PCB.
(Note: The error is • If necessary, replace the
detected during WT2 Sensor PCB, see Repair
operation.) Instructions "Replacing the WT2
Sensor PCB".
49 Air sensor 2 Breakdown in the • Check the WT2 Sensor PCB using the
breakdown air temperature DrägerService Mode, see diagnosis
sensor 2 on the "Sensor PCB".
WT2 Sensor PCB.
(Note: The error is • If necessary, replace the
detected during WT2 Sensor PCB, see Repair
operation.) Instructions "Replacing the WT2
Sensor PCB".
50 Air sensors The temperature • Check the WT2 Sensor PCB using the
temp. diff. high difference between DrägerService Mode, see diagnosis
the air temperature "Sensor PCB".
sensors (1 and 2) is
greater than 1.0 °C. • If necessary, replace the
(Note: The error is WT2 Sensor PCB, see Repair
detected during Instructions "Replacing the WT2
operation.) Sensor PCB".

51 Key stuck A key is activated • Check membrane keypad using


for more than DrägerService Mode, see "Control keys
30 seconds. (Note: test".
The error is
detected during • If necessary, replace membrane keypad,
operation and leads see Repair Instructions "Replacing the
to INOP.) membrane keypad".
• If necessary, replace the WT2 Controller
PCB, see Repair Instructions "Replacing
the WT2 Controller PCB".
52 Comm. error Communication • Check cable connection between
pba sensor between WT2 WT2 Sensor PCB and WT2 Actuator
Controller PCB and PCB; replace cable connection, if
WT2 Sensor PCB necessary.
is disturbed. (Note:
The error is • If necessary, replace the
detected during WT2 Sensor PCB, see Repair
For internal use only. Copyright reserved.

operation and leads Instructions "Replacing the WT2


to INOP.) Sensor PCB".
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 10 of 14


No.
53
Plain text
Comm. error
Cause
Communication
Dräger Medizintechnik

Remedy
• Check cable connection between

pba actuator between WT2 WT2 Actuator PCB and WT2 Controller
Controller PCB and PCB; replace cable connection, if
WT2 Actuator PCB necessary.
is disturbed. (Note:
The error is • If necessary, replace the
detected during WT2 Actuator PCB, see Repair
operation and leads Instructions "Replacing the
to INOP.) Error 52 WT2 Actuator PCB".
also occurs.
54 CAN interface Software problem • Switch device off and on again. If this
error or external does not help, proceed as follows:
interference. (Note:
The error is • If necessary, replace the WT2 Controller
detected during PCB, see Repair Instructions "Replacing
operation and leads the WT2 Controller PCB".
to INOP.)
55 Skin sensor 1 Short-circuit in skin- • Check skin-temperature sensor 1 and
short-circuit temperature sensor replace, if necessary.
1. (Note: The error
is detected during • Check the WT2 Sensor PCB using the
operation.) DrägerService Mode, see diagnosis
"Sensor PCB".
• If necessary, replace the
WT2 Sensor PCB, see Repair
Instructions "Replacing the WT2
Sensor PCB".
56 Skin sensor 2 Short-circuit in skin- • Check skin-temperature sensor 2 and
short-circuit temperature sensor replace, if necessary.
2. (Note: The error
is detected during • Check the WT2 Sensor PCB using the
operation.) DrägerService Mode, see diagnosis
"Sensor PCB".
• If necessary, replace the
WT2 Sensor PCB, see Repair
Instructions "Replacing the WT2
Sensor PCB".
57 GoldCap The GoldCap • Check the WT2 Actuator PCB using the
voltage low voltage is too low. DrägerService Mode, see diagnosis
(Note: The error is "Actuator PCB".
For internal use only. Copyright reserved.

detected during
operation.) • If necessary, replace the
WT2 Actuator PCB, see Repair
Instructions: Replacing the
WT2 Actuator PCB
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 11 of 14


No.
58
Plain text
EPROM error
Cause
Checksum test
Dräger Medizintechnik

Remedy
• Carry out download.

failed. (Note: The
error is detected • Replace the WT2 Controller PCB, see
during operation.) Repair Instructions "Replacing the
WT2 Controller PCB".
59 Current limiter Current limiter (+5-V • Check the WT2 Actuator PCB using the
error sensor) test failed. DrägerService Mode, see diagnosis
(Note: The error is "Actuator PCB".
detected during the
power-on test.) • If necessary, replace the
WT2 Actuator PCB, see Repair
Instructions "Replacing the
WT2 Actuator PCB".
60 Error in the The content of the • Replace lithium battery. If replacement
battery-backed battery-backed RAM does not help, proceed as follows:
RAM is not plausible.
(Note: The error is • Replace the WT2 Controller PCB, see
detected during Repair Instructions "Replacing the
operation.) WT2 Controller PCB".

61 Error in O2 O2 measurement • Check the WT2 Sensor PCB using the


processor does not function. DrägerService Mode, see diagnosis
(Note: The error is "Sensor PCB".
detected during
operation.) • If necessary, replace the
O2 Sensor PCB, see Repair Instructions
"".
• If necessary, replace the
WT2 Sensor PCB, see Repair
Instructions "Replacing the WT2
Sensor PCB".
62 O2 shut-off The • Replace the WT2 Sensor PCB or the
faulty O2 Sensor PCB O2 Sensor PCB, see Repair Instructions
cannot be switched "Replacing the WT2 Sensor PCB"or
off on the "Replacing the O2 Sensor PCB".
WT2 Sensor PCB.
(Note: The error is
detected during the
power-on test and
leads to INOP.)
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 12 of 14


No.
64
Plain text
Temp ref 2
Cause
The reference value
Dräger Medizintechnik

Remedy
• If necessary, Replacing the

actuator 2 of the WT2 Actuator PCB.
temperature
measurement at the
air heater is not
within the tolerance
range. (Note: The
error is detected
during operation
and leads to INOP.)
65 Temp ref 3 The reference value • If necessary, Replacing the
actuator 3 of the WT2 Actuator PCB.
temperature
measurement at the
air heater is not
within the tolerance
range. (Note: The
error is detected
during operation
and leads to INOP.)
66 Air excess The measured air • No remedy required.
temperature temperature value is
above 41 °C. (Note:
The error is
detected during
operation.)
67 Motor relay on The motor • Check motor using DrägerService Mode,
(height/tilt see calibration "Height adjustment or Tilt
adjustment) is on, adjustment".
although it should
be off. (Note: The • If necessary, Replacing the
error is detected WT2 Actuator PCB.
during operation.) • Replace motor, if necessary.
68 Motor relay off The motor • Check motor using DrägerService Mode,
(height/tilt see calibration "Height adjustment or Tilt
adjustment) is off, adjustment".
although it should
be on. (Note: The • If necessary, Replacing the
error is detected WT2 Actuator PCB.
during operation.)
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 13 of 14


No.
69
Plain text
Oxygen valve on
Cause
The oxygen valve is
Dräger Medizintechnik

Remedy
• If necessary, Replacing the

on, although it WT2 Actuator PCB.
should be off. (Note:
The error is
detected during
operation and leads
to INOP.)
70 Oxygen valve off The oxygen valve is • If necessary, Replacing the
off, although it WT2 Actuator PCB.
should be on. (Note:
The error is • Replace oxygen valve, if necessary.
detected during
operation.)
71 Oxygen valve The current flow • If necessary, Replacing the
circuit through the oxygen WT2 Actuator PCB.
valve is too high.
(Note: The error is • Replace oxygen valve, if necessary.
detected during
operation.)
72 GoldCap GoldCap voltage is • Check cable connections between
disconnected not present at the WT2 Actuator PCB and
WT2 Controller PC WT2 Controller PCB; replace cable
B. (Note: The error connections, if necessary.
is detected during
operation.)
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01

6150.000 Caleo 10/2000 Repair Instructions Page 14 of 14