Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Pumping Type Gas Detector

SKY2000
User Manual

Welcome to visit ATO.com!

Email: sales@ato.com
Tel: +1 800-585-1519 (Toll-free)
Global Shipping Enabled
SKY2000 Gas Detector (Pumping Suction Type)
*** Thanks for purchasing our product. Please read the user manual carefully before using. ***

Keys Operation:
Turn on/off: long press for 3s to turn on/off the device, short press to confirm or save data.

Menu: in detection mode, press this button to enter into parameter setting mode, and in parameter
setting mode, press this button to move the cursor.
Plus: in parameter setting mode, short press it to select items upward or add value. Long press for 3
seconds to turn on/off alarm function quickly.
Minus: in parameter setting mode, short press it to select items downward or decrease value. Long
press for 3 seconds to turn on/off air pump quickly.
Exit: in parameter setting mode, short press it to go back to previous menu. Long press to shift the
gas concentration unit quickly.
(Note: The long press function only work in detection mode. When you change or reset any data, please remember to click
“ON/OFF” button to confirm and save the settings. Remember to turn on the pump when start the detection. Please turn off the
pump when there is no need to detect, which will be helpful for the service life of the pump.)

Power On / Off
Put the detector in clean air, and press the “ON/OFF” for three seconds, the detector will be turned on and the
pump will start to work. The screen displays the following interface successively: Brand and Logo, main
parameter interface, initialing countdown. The countdown would be 60 or 120 seconds, which is to make sure
that the sensors are fully activated. After countdown, the device enter normal detection interface.
Take the detector out of the detected environment and wait for the data to get down to zero, then press the
“ON/OFF” to turn off the detector

Zero Calibration
Zero calibration must be done in clean air. Better to do once after alarming happens.

Press “MENU” to enter parameter setting mode, select the "calibrate zero", and press “ON/OFF” to enter the
mode. Observe the value of “real time concentration”. If the reading is not zero and the drifting is too big when
the reading is stable, then it is necessary to conduct zero point calibration.

Press “ON/OFF”, the real-time concentration will go back to zero, and press ”ESC” to go back to
detection mode, gas concentration values also back to zero.

Note: Zero calibration of oxygen detector, nitrogen detector and carbon dioxide detector (target gases are
component of the air) cannot be done in the air. Only the target point calibration (value of target point
concentration is the standard concentration values in the clean air) can be done.

www.ato.com sales@ato.com Global Shipping +1 800‐585‐1519


SKY2000 Gas Detector (Pumping Suction Type)

Process of Over-range incorrect operation


Users should avoid using gases which exceed the measure range to impact the sensor, because it will affect
the lifespan and sensitivity of the sensor, even, "poison" the sensor. If there is any over-range incorrect
operation makes the detector displaying an concentration at a large reading, the remedy is to take the device
out of the environment immediately and put it in clean air for over half an hour, and then observe the reading, if
it keeps going down, then wait until the reading back to zero before powering it off, and do the zero point
calibration next time before using it. If the reading maintained at full scale, user should send the device back to
manufacturer or agent for repair or replacing the sensor.

①— Air Inlet
② — Air Outlet
③ — air chamber of sensor
④ — LCD
⑤ — PLUS
⑥ — MINUS
⑦ — MENU
⑧— ESC
⑨— ON/OFF

-----------------------------------------------------------------------------------------------------------------------------
Note
*If detector is used for the gas cylinder detection, considering the high pressure in the cylinder, please use a
pressure regulator, and keep it as the below data:
Flow rate: 800mL~1L/minute
Pressure: 0.1MPa or 1Bar
*Please put the detector in clean air for about 5 minutes and waiting for the data falling down to normal before
turning it off after measurement.

www.ato.com sales@ato.com Global Shipping +1 800‐585‐1519


SKY2000 Gas Detector (Pumping Suction Type)

Operation Interface
When the detector is turned on completely, the device will enter the detecting interface. The gas type will be
showed on the top left, and the time and battery power will be at the right. In the middle, there is the real- time
concentration. There will be BEEP-ON at the bottom left, which means the alarm is on. The PUMP-ON at the
bottom right, which means the pump is working.

Alarm Status
When the detected gas concentration is higher than the low alarm value, the alarm sound will be on, yellow
light will be flickered.
When the detected gas concentration is high than the high alarm value, the alarm sound will be on, red light
will be flickered.

Alarm Setting
Press “Menu”, choose the “Alarm point set”, and start to set the alarm value.
There are LOW POINT and HIGH POINT for set, press “MENU” to move the cursor to the value you want to
set, and press the “PLUS” or “MINUS” to change the value. After the setting is done, press “ON/OFF” button to
save the settings.

Low Battery Alarm


When the battery is low, there will be sound, light low battery alarm. There will be sound and light every 5
seconds to remind the user to charge the device in time. When the battery is used up, the device will be turned
off automatically.

Restore Setting
When all the settings are wrong, or users operate the detector in a wrong way, you can restore the detector to
factory settings. But if the detector works normally, please don’t operate this setting. In order to avoid wrong
setting, password is needed. If you want to restore the detector to factory setting, please contact the supplier.

www.ato.com sales@ato.com Global Shipping +1 800‐585‐1519


SKY2000 Gas Detector (Pumping Suction Type)

Target Point Calibration


Note: Calibration needs to be done by professionals and with necessary accessories and standard gas.

Prepare the standard gas before starting the target point calibration. In detection mode, press “MENU”, select
“calibrate span”, press ON/OFF button to enter. Then you will need to enter password. (If you need password,
please contact the manufacturer to get it)

After entering the password successfully, press ON/OFF to enter the target point calibration interface, as
shown in the picture. The interface will display “concentration”, “set value. The concentration is the result got
from the calculation of several internal calibration parameters. The process of calibration is the process to
recalculate the internal parameters of the device. The set value is the value need to be entered manually; it
should be the concentration of the standard gas to be calibrated. Select the parameter under the current
interface, modify its value with “PLUS” and “MINUS” button.

The calibration process is as below:


1. Connect the detector with standard gas cylinder by hose.
2. Enter the target point calibration interface, set the “SET value” as the standard gas concentration.
3. Release the standard gas to the detector in a flow rate of 500mL/min. The value of “concentration” will
become larger gradually, and become stable after 30seconds.
4. When the value of “concentration” is stable, press the “ON/OFF” button, the value of “concentration” will be
same with or almost same with the “Set value”. The calibration is finished.
5. Cut off the standard gas supply. Press “ESC” to go back to detection interface.

www.ato.com sales@ato.com Global Shipping +1 800‐585‐1519

You might also like