(14374331 - Clinical Chemistry and Laboratory Medicine (CCLM) ) Analytical Evaluation of Four Faecal Immunochemistry Tests For Haemoglobin

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Clin Chem Lab Med 2020; aop

Carolyn Piggott*, Magdalen R. R. Carroll, Cerin John, Shane O’Driscoll and Sally C. Benton

Analytical evaluation of four faecal


immunochemistry tests for haemoglobin
https://doi.org/10.1515/cclm-2020-0251 Conclusions: All four systems are fit for purpose and have
Received March 4, 2020; accepted June 29, 2020; published online an analytical performance as documented by their
xxx manufacturers.
Abstract Keywords: analytical evaluation; colorectal cancer; faecal
immunochemical test; FIT.
Background: Faecal immunochemical tests (FIT) for hae-
moglobin (Hb) are being used in the investigation of
colorectal cancer. These tests use antibodies raised to the
globin moiety of human Hb. Here, four automated quan- Introduction
titative FIT systems (HM-JACKarc, NS-Prime, OC-Sensor
PLEDIA and SENTiFIT 270) are evaluated analytically to The quantitative faecal immunochemical test for haemo-
confirm whether the performance of the systems meet the globin (FIT) measures the concentration of human blood in
manufacturers’ claims. faeces using polyclonal antibodies raised against the
Methods: Assessment of the analytical performance of the globin moiety of human haemoglobin (Hb). These tests are
FIT systems was undertaken using Hb lysates, real patient being used worldwide for both screening of asymptomatic
samples and external quality assessment (EQA) samples. individuals [1] and to aid the assessment of patients with
This analytical assessment focused on detection charac- low risk symptoms [2]. Although these tests are not
teristics, imprecision, linearity, prozone effect, recovery currently widely used in the diagnosis and monitoring of
and carryover. inflammatory bowel diseases such as Crohn’s disease and
Results: All four methods demonstrated good analytical ulcerative colitis, there is potential for this use and studies
performance, with acceptable within- and between-run are being carried out [3].
imprecision, good recovery of f-Hb and limited carryover of FIT has superseded the use of guaiac faecal occult
samples. They also all show good linearity across the range blood testing (gFOBT). FIT offers advantages over gFOBT
of concentrations tested. The results of EQA samples which include quantitative examination with the option to
showed different variations from the target values (−52 to choose the cut-off for positive results, with low cut-offs for
45%), due to the absence of standardisation across the the triage of symptomatic patients, and higher cut-offs for
different methods. asymptomatic participants in screening programmes, the
antibodies are specific to human Hb, and the examination
can be carried out on semi-automated instruments. One FIT
collection device requiring a single faecal sample is
acceptable for screening and to determine the need for a
*Corresponding author: Carolyn Piggott, NHS Bowel Cancer
colonoscopy in symptomatic patients [1–3]. FIT uses a
Screening Programme – Southern Hub, Royal Surrey County Hospital,
20 Priestley Road, Surrey Research Park, Guildford, GU2 7YS, UK;
simpler and more attractive sampling technique than that
Berkshire and Surrey Pathology Services, Royal Surrey County used by gFOBT; it allows the positivity rate to be adjusted to
Hospital, Guildford, England, UK, E-mail: carolyn.piggott@nhs.net. meet local colonoscopy resources; and the Hb concentra-
https://orcid.org/0000-0002-6343-6202 tion has the potential to be incorporated into a multivariate
Magdalen R. R. Carroll, Cerin John, Shane O’Driscoll and Sally C.
risk score to enable a higher sensitivity for cancer or
Benton: NHS Bowel Cancer Screening Programme Southern Hub,
Royal Surrey County Hospital, Guildford, England, UK; Berkshire and
advanced adenoma [4, 5].
Surrey Pathology Services, Royal Surrey County Hospital, Guildford, In this study four quantitative FIT laboratory analysers
England, UK, E-mail: romanyrowan@gmail.com (M.R.R. Carroll), and their appropriate materials have been evaluated
cerin.john@nhs.net (C. John), sodriscoll@nhs.net (S. O’Driscoll), analytically to confirm whether the systems’ performance
sally.benton@nhs.net (S.C. Benton). https://orcid.org/0000-0001- meet manufacturers’ claims. These are the HM-JACKarc
6886-5631 (M.R.R. Carroll). https://orcid.org/0000-0002-7577-6503
(Kyowa Medex Co. Ltd/Hitachi Chemical Diagnostics
(C. John). https://orcid.org/0000-0002-8693-3658 (S. O’Driscoll).
https://orcid.org/0000-0001-9230-9088 (S.C. Benton) Systems, Tokyo, Japan); NS-Prime (Alfresa Pharma
2 Piggott et al.: Evaluation of four FIT

Corporation, Osaka, Japan); OC-Sensor PLEDIA (Eiken Materials


Chemical Co. Ltd, Tokyo, Japan) and the SENTiFIT 270
(Sentinel Diagnostics SpA, Milan, Italy). These systems Assessment of the analytical performance across the FIT systems was
were selected as, in our opinion, they were potentially undertaken using Hb lysates, anonymised patient samples and EQA
materials.
suitable for processing samples for screening programmes
Hb lysates were prepared from venous blood collected into
due to their throughput and ease of use. All analysers and
potassium-EDTA blood collection tubes (VACUETTE, supplied by
their materials were provided free of charge. Berkshire and Surrey Pathology Services) and included a final 1-in-2
The four FIT systems studied are bench top analysers dilution in physiological saline. The Hb concentration of each diluted
which utilise gold- (NS-Prime) or latex-particle aggluti- lysate was determined by measurement as part of the full blood count
nation (HM-JACKarc, OC-Sensor PLEDIA and SENTiFIT on an ADVIA 2120 (Siemens Healthcare Ltd. Frimley, UK). The lysate
samples were frozen at −20 °C until used. Further dilutions were made
270/FOB Gold), and turbidimetric analysis to determine
in manufacturer specific diluent.
the concentration of Hb present in faecal samples pro- Samples were aliquoted into analyser cups, with the number of
vided for examination [6]. Three of the analysers cups needed for each assessment depending on the maximum
(NS-Prime, OC-Sensor PLEDIA and SENTiFIT 270) use number of replicates that could be set for each system (number of
reusable cuvettes which are washed between each sample, replicates: HM-JACKarc up to 3, NS-Prime up to 9, OC-Sensor PLEDIA
up to 10; SENTiFIT 270 up to 10). EQA samples were loaded into, and
whereas the HM-JACKarc uses single use cuvettes, dis-
sampled directly from, the FIT sampling devices. Where samples
carding the cuvettes after each test.
were measured multiple times on any single day, they were
In 2013 an evaluation of analysers (HM-JACKarc, measured in a single run. Where samples were measured on more
NS-Plus, OC-Sensor DIANA and Biomajesty) and their than one day, new aliquots were used for each day, and if taken
associated materials and collection devices available at multiple times from the same original sample, this was mixed
the time, and suitable for use in national screening pro- between removal of the aliquots.
grammes, was carried out [7]. Since then the NS-Plus and
OC-SENSOR DIANA analysers have been updated, with
Detectability characteristics
the HM-JACKarc remaining unchanged; to our knowledge
the methodology has remained unchanged for all three.
Detectability characteristics were assessed as recommended by Fraser
The NS-Prime and the OC-Sensor PLEDIA have larger
and Benton [9] according to the CLSI EP17-A2 protocol [10] through
sample loading and reagent capacities and the user determination of limit of blank (LoB), limit of quantitation (LoQ) and
software programmes are simplified. This evaluation has limit of detection (LoD):
examined the SENTiFIT 270 instead of the Biomajesty, LoB: “the highest measurement result that is likely to be observed
which both use the Sentinel Diagnostics FOB Gold (with stated probability) for a blank sample” [10]; 20 blank collection
devices were analysed and the following calculation used to deter-
method, although a different calibrator has been used in
mine the value: LoB=mean [blank] + 1.645(SD [blank]).
this evaluation. LoQ: “the lowest amount of a measurand in a material that can
All four analysers have associated specimen collection be quantitatively determined with stated accuracy, under stated
devices provided by the manufacturers, which are used experimental conditions” [10]. A range of Hb dilutions, around the
by participants to collect either 2 (HM-JACKarc) or 10 mg LoQ provided by the manufacturer (Table 1), was created for each
FIT system by using samples naturally positive and negative for f-
(NS-Prime, OC-Sensor PLEDIA and SENTiFIT 270) of
Hb. The samples were 2–5 days old and collected from participants
faeces [7]. The devices are then sent to laboratories for
in a multi-sampling study (in which each participant was sent four
examination. different collection devices; samples were returned by post and
To our knowledge, there is currently no peer- refrigerated as they arrived). The contents of the appropriate de-
reviewed independent comprehensive analytical evalu- vices, ranging from three to seven samples per system, were pooled
ation of these quantitative FIT systems. This is despite together, and the concentration of the pool determined by exami-
nation on the relevant analyser, within five days of the samples
their increasing use in screening programmes and
being collected. This was used to create a concentration gradient
symptomatic populations. from 0 to 20 µg Hb/g faeces for each FIT system; each sample was
The aim of this study was to carry out an analytical measured 10 times. The SD and CV were determined for each con-
evaluation of four automated quantitative FIT systems. centration. The LoQ was taken to be the concentration at which the
measured CV was below 10%.
LoD: “the lowest concentration of measurand which can
Methods consistently be detected” [10] was determined using a sample from a
participant in the multi-sampling study, just below the LoQ concen-
The study followed the World Endoscopy Organisation (WEO) faecal tration; this sample was analysed 20 times and the following calcu-
immunochemical tests for haemoglobin evaluation reporting (FITTER) lation was used to determine the value: LoD=LoB + 1.645(SD [low
checklist [8]. concentration sample]).
Piggott et al.: Evaluation of four FIT 3

Table : Limit of blank (LoB, n=), limit of detection (LoD, n=) and limit of quantitation (LoQ, n= each concentration) of all four faecal
immunochemical test (FIT) systems. NA: Not available. LoQ range of concentrations tested: HM-JACKarc – µg/g, NS-Prime – µg/g,
OC-Sensor PLEDIA – µg/g, SENTiFIT  – µg/g.

FIT system LoB (μg Hb/g faeces) LoD (μg Hb/g faeces) LoQ (μg Hb/g faeces)

Manufacturer quoted Study results Manufacturer quoted Study results Manufacturer quoted Study results

HM-JACKarc NA  NA   
NS-PRIME NA     
OC-Sensor PLEDIA NA  NA   
SENTiFIT       

Imprecision Hook/prozone

Between-run imprecision was assessed by measuring three replicates The analytical reaction was tested for the presence of the hook/pro-
of each concentration of Hb lysate diluted in manufacturers’ buffer, zone effect, where immunoassays may give erroneously low values for
measured on five consecutive days, using separate aliquots stored at samples of extremely high concentrations of analyte. Eight samples
4–8 °C for each day. The number of concentrations was determined by for each system were prepared using Hb lysate solution diluted in
each manufacturer, where they had carried out a similar study, and we manufacturer-specific collection device diluent to give concentrations
aimed to replicate their concentrations (see Table 2 for concentra- above the upper limit of measurement range of the method (HM-
tions). The F-test was used to determine if the SD of each set of results JACKarc: 6,200–100,000 µg Hb/g, NS Prime: 600–78,000 µg Hb/g OC
was significantly different to the manufacturer’s SD. SENSOR: 1,000–137,000 µg Hb/g, FOB Gold: 664–85,000 µg Hb/g)
Within-run imprecision was assessed by measuring control so- These samples were measured in duplicate to check whether erro-
lutions provided by the manufacturers 10 times each in one run; the neous results were blocked and reported with a clear error message.
mean, SD and CV were calculated.

Carryover
Linearity
Sample carryover was assessed using the protocol described by
The measurement range of the assay was assessed using between 14 Broughton et al. [11]. A Hb lysate solution of high Hb concentration
and 16 dilutions of the Hb lysate in manufacturer-specific collection and a solution of low Hb concentration (b) were prepared in manu-
device diluent. The concentrations studied covered the analytical facturer specific collection device diluent in sample cups
range of each system (measurement ranges: HM-JACKarc 7–400 µg (HM-JACKarc: 210 and 30 µg Hb/g, NS Prime: 236 and 10 µg Hb/g, OC
Hb/g faeces, NS-Prime 10–240 µg Hb/g faeces, OC-Sensor PLEDIA 10– SENSOR: 96 and 15 µg Hb/g, FOB Gold: 53 and 12 µg Hb/g). Three
200 µg Hb/g faeces, SENTiFIT 270 3–170 µg Hb/g faeces). Each sample aliquots of each solution were measured, one set following the other.
was analysed in duplicate, and the R2 value calculated. This process was repeated 10 times. The carryover factor (k) was

Table : Between-run imprecision of haemoglobin (Hb) measured with four faecal immunochemical test (FIT) systems (n= for each
material); p-value calculated using F test.

FIT system Manufacturers’ results Study between-run imprecision p-value

Mean Hb (µg/g) SD Hb (µg/g) CV (%) Mean Hb (µg/g) SD Hb (µg/g) CV (%)

HMJACK-arc  . .  . . .


 . .  . .
 . .  . .
NS-Prime  . .  . . .
 . .  . .
 . .  . .
 . .  . .
OC-Sensor PLEDIA  . .  . . .
 . .  . .
SENTiFIT   . .  . . .
 . .  . .
 . .  . .
4 Piggott et al.: Evaluation of four FIT

calculated from the equation: k=(b1 − b3)/(a3 − b3), where ‘a’ and ‘b’ 2 μg Hb/g faeces; the LoDs were equal to or less than 3 μg Hb/
are the high and low concentration solutions respectively, and ‘1’ and g faeces and the LoQs were 3 μg Hb/g faeces on the NS-Prime
‘3’ are the results for the first and third sample respectively.
and SENTiFIT 270, 4 μg Hb/g faeces on the HM-JACKarc and
Recovery 6 μg Hb/g faeces on the OC-Sensor PLEDIA. All results met,
or were lower than the manufacturers’ claims.
The Hb lysate solution diluted in manufacturer-specific collection device
diluent was prepared at a low but reliably measurable concentration for Imprecision
each FIT system. Two sets of eight samples were prepared: in one set
small volumes were replaced with high concentration solution (the
The between-run imprecision SDs (Table 2) using diluted
relevant FIT system calibrator), and in the other set, identical volumes
were replaced with diluent; no volume replacement exceeded 10% of the Hb lysate for all concentrations on all FIT systems were not
final volume. The samples were measured in duplicate and the recovery significantly different (p>0.05) than the manufacturers’
calculated by comparing the expected concentration of Hb in each claims.
sample with the measured concentration of Hb; recovery (%) = (mean of Control materials were used to assess the within-run
measured values)/(expected concentration) × 100.
imprecision (Table 3); all FIT systems had CVs within the
External quality assessment manufacturers’ claims and were less than 3%.

The UK National External Quality Assessment Services (UK NEQAS, Linearity


Birmingham, England) provided six EQA samples (faecal-like matrix,
0–120 µg Hb/g faeces) which on receipt were loaded into the relevant All FIT systems showed good linearity (R2 between 0.97 and
sample collection devices before analysis. The amount of loaded
0.99) throughout the ranges tested.
sample covered the grooves or dimples of the sampling devices, with a
slight excess. A slight excess of sample has been shown not to
significantly affect the f-Hb using these same FIT systems [12], as it is Prozone
removed by the internal collar.

All FIT systems detected the prozone appropriately and


Results gave adequate error warnings. These either alerted the user
to the need for dilution, or that the sample was too
Throughout this paper all results are presented as μg Hb/g concentrated for an accurate result to be determined (giv-
faeces in accordance with the WEO FITTER protocol. This ing ‘prozone warnings’). Of the four manufacturers, only
enables direct comparison of the results across the Sentinel include any information on the potential of a
different FIT systems [8]. prozone effect, which they state will not occur below
4,845 µg Hb/g faeces.
Detectability characteristics
Carryover
The LoB, LoD and LoQ for the four FIT systems using patient
samples from a multi-sampling study are shown in Table 1. Carryover determined by the Broughton method [11], was
The LoBs for the four FIT systems were equal to or less than insignificant (k<0.05) on all four FIT systems.

Table : Within-run imprecision of haemoglobin (Hb) measured with four faecal immunochemical test (FIT) systems.

FIT system Manufacturers’ claim for within-run Study within-run imprecision


imprecision
n Mean Hb SD CV Consistent with
(µg Hb/g faeces) (µg Hb/g faeces) (%) claim

HM-JACKarc CV≤% (n=)   . . Yes


  . .
NS-PRIME CV≤% (n unknown)   . . Yes
  . .
OC-Sensor PLEDIA CV % (n=)   . . Yes
  . .
  . .
SENTiFIT  ≤ µg/g CV≤% or SD≤; > µg/g   . . Yes
CV≤% (n=)   . .
Piggott et al.: Evaluation of four FIT 5

Recovery reagent method and a different measurement system to this


study.
The HM-JACKarc recovery was 94–100%, the NS-Prime Within-run and between-run imprecision was studied by
97–114%, OC-Sensor PLEDIA 93–109%, and SENTiFIT 270 Ahn et al. [13] (OC-Sensor PLEDIA and NS-Prime), Kusaka
87–99%. et al. [14] (OC-Sensor PLEDIA), Lee et al. [15] (OC-Sensor
Micro and FOB Gold) using QC material or spiked faecal
EQA sample comparison samples. For within-run imprecision the results all met the
manufacturers’ claims. For between-run imprecision the re-
Figure 1 shows the results from the UK NEQAS material. sults for Ahn et al. [13] and Kusaka et al. [14] met the man-
The HM-JACKarc showed positive bias from the target ufacturers’ claims. The results for Lee et al. [15] showed the
values; the OC-Sensor PLEDIA, NS-Prime and the SENTiFIT CV%s to be higher than the manufacturer’s claim although it
270 showed negative bias from the target values. is unclear if a single sample was stored at 4 °C and remeas-
ured over 17 days, and both methods (OC-Sensor Micro and
FOB Gold) showed a slight but gradual decline in f-Hb over
Discussion the same time. The same three studies also included carry-
over, which was found to be insignificant for OC-Sensor
All four methods demonstrated good analytical perfor- PLEDIA and Micro, and the NS-Prime; for FOB-Gold signifi-
mance, with detectability characteristics (LoB, LoD, LoQ) cant carryover was found by Lee et al. [15], although they did
meeting or better than the manufacturers' claims, within- not state which analyser was used in the evaluation. Line-
run and between-run imprecision meeting or not signifi- arity was studied by Ahn et al. [13] and Kusaka et al. [14] and
cantly different than manufacturers' claims, and linearity they both state that linearity was acceptable. LOD, LOQ and
R2 values between 0.97 and 0.99. All the systems detected prozone were included by Kusaka et al. [14] and all met the
prozone samples correctly, carryover was insignificant, manufacturer’s claims.
and recovery was 87–114%. This evaluation does have limitations, most notably to
Each manufacturer provided details of their own assess the systems analytically without matrix effect bias;
assessment of their FIT systems and this evaluation was we used whole blood lysate diluted in manufacturer spe-
compared with these claims, all data were in agreement. cific buffer for some parts of the evaluation. Ideally real
The results of the EQA examination (faecal-like matrix samples would be used and analysed across the four sys-
spiked with Hb and loaded into FIT sample collection de- tems, but there was a limited supply of these and we could
vices) show that the systems all give different Hb concen- not use them for all parts of the evaluation. Additionally we
trations which varied from the expected values (Figure 1, have no way of accurately determining the actual original
median values: HM-JACKarc 45%, OC-Sensor PLEDIA – concentration of Hb within those samples and cannot
29%, NS-Prime –30%, SENTiFIT 270 –52%); this is to be determine any bias present [16]. All faecal samples are
expected because the methods are traceable to different unique in composition and have varying degradation rates,
reference materials. different effects are seen with different samples making
There have been three previous published studies that them unsuitable to assess the analytical performance of
included aspects of the analytical performance included in systems designed to detect human haemoglobin. The use
this study, although some of these have included the same of lysate as a surrogate eliminates these effects and allows

Figure 1: Box and whisker plot showing the


bias (%) of UK National External Quality
Assessment Services (UK NEQAS) samples
(faecal-like matrix loaded into sample
collection devices) measured on four faecal
immunochemical test (FIT) systems, against
the EQA scheme target value.
6 Piggott et al.: Evaluation of four FIT

an analytical assessment and comparison of the four FIT 2. National Institute for Health and Care Excellence.
systems. The target concentrations for the imprecision Quantitative faecal immunochemical tests to guide referral
for colorectal cancer in primary care, 2017 DG30. Available
assessment were chosen to match those quoted by the
at: https://www.nice.org.uk/guidance/dg30.pdf [Accessed 12
manufacturers, although ideally we should have also Nov 2019].
included concentrations above and below the commonly 3. Kato J, Hiraoka S, Nakarai A, Takashima S, Inokuchi T, Ichinose M.
used cut-offs, a negative concentration and the same Fecal immunochemical test as a biomarker for inflammatory
samples analysed across all four systems. bowel diseases: can it rival fecal calprotectin?. Intest Res 2016;
14: 5–14.
Additionally the evaluations for each system were
4. Hippisley-Cox J, Coupland C. Identifying patients with suspected
performed at different times and there were slight differ-
colorectal cancer in primary care: derivation and validation of an
ences in the protocols, particularly for the range of con- algorithm. Br J Gen Pract 2012; 62: e29–37.
centrations assessed. 5. Halloran SP. Intelligent use of the fecal immunochemical test in
Currently, and for future assessments, a reference population-based screening (editorial). Ann Int Med 2018; 169:
standard for faecal Hb is being investigated by the Inter- 496–7.
6. Koivunen ME, Krogsrud RL. Principles of immunochemical
national Federation of Clinical Chemistry and Laboratory
techniques used in clinical laboratories. Lab Med 2006;37:
Medicine which potentially will increase the accuracy of 490–7.
such assessments [17]. 7. Carroll MRR, Piggott C, Pearson S, Seaman HE, Halloran SP.
Evaluation of quantitative faecal immunochemical tests for

Conclusions haemoglobin. Guildford Medical Device Evaluation Centre


(GMEC); 2012. Available at: http://www.worldendo.org/wp-
content/uploads/2018/07/gmec_fit_evaluation_report_update-
This evaluation assessed four laboratory FIT systems that final.pdf [Accessed 28 Feb 2020].
are used for faecal Hb examination in screening and 8. Fraser CG, Allison JE, Young GP, Halloran SP, Seaman HE.
symptomatic populations. The results confirm that all four Improving the reporting of evaluations of faecal immunochemical
tests for heamoglobin: the FITTER standard and checklist. Eur J
FIT systems perform acceptably when compared with
Cancer Prev 2015; 24: 24–6.
manufacturers data, and from an analytical perspective, all
9. Fraser CG, Benton S. Detection capability of quantitative faecal
are suitable for use. immunochemical tests for haemoglobin (FIT) and reporting of low
faecal haemoglobin concentrations. Clin Chem Lab Med 2019; 57:
611–6.
Acknowledgments: We thank FIT system suppliers
10. Clinical and Laboratory Standards Institute. Evaluation of
(HM-JACKarc: Alpha Laboratories, Eastleigh, UK; detection capability for clinical laboratory measurement
NS-Prime: originally supplied by Alere Ltd., Chester, UK, procedures. In: Approved guideline, 2nd ed. Wayne PA, USA: CLSI
with continuing support from Abbott, Maidenhead, UK, document EP17-A2; 2012.
and Alfresa Pharma, Osaka, Japan; OC-Sensor PLEDIA: 11. Broughton PM, Gowenlock AH, McCormack JJ, Neill DW. A revised
scheme for the evaluation of automatic instruments for use in
Mast Diagnostics Division, Bootle, UK; SENTiFIT 270:
clinical chemistry. Ann Clin Biochem 1974; 11: 207–18.
Sysmex UK Ltd., Milton Keynes, UK) for supplying the 12. Piggott C, John C, Bruce H, Benton SC. Does the mass of
analysers and consumables. We also thank Berkshire and sample loaded affect faecal haemoglobin concentration using
Surrey Pathology Services (Royal Surrey Foundation Trust, the faecal immunochemical test. Ann Clin Biochem 2018; 55:
Guildford, UK) for supplying samples used in the assess- 702–5.
13. Ahn A, Kim J, Jin Ko Y, et al. Perofrmance evaluation of two
ment and the United Kingdom National External Quality
automated quantitative fecal occult blood tests. Lab med Online
Assessment Services (UK NEQAS, Birmingham, UK) for the
2016; 6: 233–9.
external quality assessment (EQA) samples they provided. 14. Kusaka T, Nozaki T, Shibata M, et al. Measurement performance
Research funding: None declared. evaluation of fecal occult blood analyzer OC Sensor PLEDIA. J Clin
Author contributions: All the authors have accepted Lab Inst and Reag 2014; 37: 643–8.
responsibility for the entire content of this submitted 15. Lee C, O’Gorman P, Walsh P. Immunochemical faecal occult blood
tests have superior stability and analytical performance
manuscript and approved submission.
characteristics over guaiac-based tests in a controlled in vitro
Competing interests: None. study. J Clin Pathol 2011; 64: 524–8.
16. Rubeca T, Cellai F, Confortini M, Fraser CG, Rapi S. Impact of
preanalytical factors on fecal immunochemical tests: need for
References new strategies in comparison of methods. Int J Biol Markers 2015;
30: e269–74.
1. Schreuders EH, Ruco A, Rabeneck L, Schoen RE, Sung JJ, Young GP, 17. Benton SC. IFCC – FIT Working Group (FIT-WG). IFCC e-news; 2017,
et al. Colorectal cancer screening: a global overview of existing pp. 16–7. Available at: http://www.ifcc.org/media/461890/
programmes. Gut 2015; 10: 1637–49. IFCCeNewsJune2017.pdf [Accessed 6 March 2019].

You might also like