Professional Documents
Culture Documents
(14374331 - Clinical Chemistry and Laboratory Medicine (CCLM) ) Analytical Evaluation of Four Faecal Immunochemistry Tests For Haemoglobin
(14374331 - Clinical Chemistry and Laboratory Medicine (CCLM) ) Analytical Evaluation of Four Faecal Immunochemistry Tests For Haemoglobin
(14374331 - Clinical Chemistry and Laboratory Medicine (CCLM) ) Analytical Evaluation of Four Faecal Immunochemistry Tests For Haemoglobin
Carolyn Piggott*, Magdalen R. R. Carroll, Cerin John, Shane O’Driscoll and Sally C. Benton
Table : Limit of blank (LoB, n=), limit of detection (LoD, n=) and limit of quantitation (LoQ, n= each concentration) of all four faecal
immunochemical test (FIT) systems. NA: Not available. LoQ range of concentrations tested: HM-JACKarc – µg/g, NS-Prime – µg/g,
OC-Sensor PLEDIA – µg/g, SENTiFIT – µg/g.
FIT system LoB (μg Hb/g faeces) LoD (μg Hb/g faeces) LoQ (μg Hb/g faeces)
Manufacturer quoted Study results Manufacturer quoted Study results Manufacturer quoted Study results
HM-JACKarc NA NA
NS-PRIME NA
OC-Sensor PLEDIA NA NA
SENTiFIT
Imprecision Hook/prozone
Between-run imprecision was assessed by measuring three replicates The analytical reaction was tested for the presence of the hook/pro-
of each concentration of Hb lysate diluted in manufacturers’ buffer, zone effect, where immunoassays may give erroneously low values for
measured on five consecutive days, using separate aliquots stored at samples of extremely high concentrations of analyte. Eight samples
4–8 °C for each day. The number of concentrations was determined by for each system were prepared using Hb lysate solution diluted in
each manufacturer, where they had carried out a similar study, and we manufacturer-specific collection device diluent to give concentrations
aimed to replicate their concentrations (see Table 2 for concentra- above the upper limit of measurement range of the method (HM-
tions). The F-test was used to determine if the SD of each set of results JACKarc: 6,200–100,000 µg Hb/g, NS Prime: 600–78,000 µg Hb/g OC
was significantly different to the manufacturer’s SD. SENSOR: 1,000–137,000 µg Hb/g, FOB Gold: 664–85,000 µg Hb/g)
Within-run imprecision was assessed by measuring control so- These samples were measured in duplicate to check whether erro-
lutions provided by the manufacturers 10 times each in one run; the neous results were blocked and reported with a clear error message.
mean, SD and CV were calculated.
Carryover
Linearity
Sample carryover was assessed using the protocol described by
The measurement range of the assay was assessed using between 14 Broughton et al. [11]. A Hb lysate solution of high Hb concentration
and 16 dilutions of the Hb lysate in manufacturer-specific collection and a solution of low Hb concentration (b) were prepared in manu-
device diluent. The concentrations studied covered the analytical facturer specific collection device diluent in sample cups
range of each system (measurement ranges: HM-JACKarc 7–400 µg (HM-JACKarc: 210 and 30 µg Hb/g, NS Prime: 236 and 10 µg Hb/g, OC
Hb/g faeces, NS-Prime 10–240 µg Hb/g faeces, OC-Sensor PLEDIA 10– SENSOR: 96 and 15 µg Hb/g, FOB Gold: 53 and 12 µg Hb/g). Three
200 µg Hb/g faeces, SENTiFIT 270 3–170 µg Hb/g faeces). Each sample aliquots of each solution were measured, one set following the other.
was analysed in duplicate, and the R2 value calculated. This process was repeated 10 times. The carryover factor (k) was
Table : Between-run imprecision of haemoglobin (Hb) measured with four faecal immunochemical test (FIT) systems (n= for each
material); p-value calculated using F test.
calculated from the equation: k=(b1 − b3)/(a3 − b3), where ‘a’ and ‘b’ 2 μg Hb/g faeces; the LoDs were equal to or less than 3 μg Hb/
are the high and low concentration solutions respectively, and ‘1’ and g faeces and the LoQs were 3 μg Hb/g faeces on the NS-Prime
‘3’ are the results for the first and third sample respectively.
and SENTiFIT 270, 4 μg Hb/g faeces on the HM-JACKarc and
Recovery 6 μg Hb/g faeces on the OC-Sensor PLEDIA. All results met,
or were lower than the manufacturers’ claims.
The Hb lysate solution diluted in manufacturer-specific collection device
diluent was prepared at a low but reliably measurable concentration for Imprecision
each FIT system. Two sets of eight samples were prepared: in one set
small volumes were replaced with high concentration solution (the
The between-run imprecision SDs (Table 2) using diluted
relevant FIT system calibrator), and in the other set, identical volumes
were replaced with diluent; no volume replacement exceeded 10% of the Hb lysate for all concentrations on all FIT systems were not
final volume. The samples were measured in duplicate and the recovery significantly different (p>0.05) than the manufacturers’
calculated by comparing the expected concentration of Hb in each claims.
sample with the measured concentration of Hb; recovery (%) = (mean of Control materials were used to assess the within-run
measured values)/(expected concentration) × 100.
imprecision (Table 3); all FIT systems had CVs within the
External quality assessment manufacturers’ claims and were less than 3%.
Table : Within-run imprecision of haemoglobin (Hb) measured with four faecal immunochemical test (FIT) systems.
an analytical assessment and comparison of the four FIT 2. National Institute for Health and Care Excellence.
systems. The target concentrations for the imprecision Quantitative faecal immunochemical tests to guide referral
for colorectal cancer in primary care, 2017 DG30. Available
assessment were chosen to match those quoted by the
at: https://www.nice.org.uk/guidance/dg30.pdf [Accessed 12
manufacturers, although ideally we should have also Nov 2019].
included concentrations above and below the commonly 3. Kato J, Hiraoka S, Nakarai A, Takashima S, Inokuchi T, Ichinose M.
used cut-offs, a negative concentration and the same Fecal immunochemical test as a biomarker for inflammatory
samples analysed across all four systems. bowel diseases: can it rival fecal calprotectin?. Intest Res 2016;
14: 5–14.
Additionally the evaluations for each system were
4. Hippisley-Cox J, Coupland C. Identifying patients with suspected
performed at different times and there were slight differ-
colorectal cancer in primary care: derivation and validation of an
ences in the protocols, particularly for the range of con- algorithm. Br J Gen Pract 2012; 62: e29–37.
centrations assessed. 5. Halloran SP. Intelligent use of the fecal immunochemical test in
Currently, and for future assessments, a reference population-based screening (editorial). Ann Int Med 2018; 169:
standard for faecal Hb is being investigated by the Inter- 496–7.
6. Koivunen ME, Krogsrud RL. Principles of immunochemical
national Federation of Clinical Chemistry and Laboratory
techniques used in clinical laboratories. Lab Med 2006;37:
Medicine which potentially will increase the accuracy of 490–7.
such assessments [17]. 7. Carroll MRR, Piggott C, Pearson S, Seaman HE, Halloran SP.
Evaluation of quantitative faecal immunochemical tests for