Professional Documents
Culture Documents
(2007.10.29) Gyro TV Wall Mount Monitor Support Pre-Installation and Installation M - Pim - 2129003-1-100 - 1 PDF
(2007.10.29) Gyro TV Wall Mount Monitor Support Pre-Installation and Installation M - Pim - 2129003-1-100 - 1 PDF
OPERATING DOCUMENTATION
2129003-1-100
Revision 1
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
2
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Important Information
LANGUAGE
Important Information 3
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
4 Important Information
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Important Information 5
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
6 Important Information
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Important Information 7
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
8 Important Information
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Important Information 9
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
10 Important Information
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Important Information 11
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
12 Important Information
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Revision History
Part / Rev Date Reason for change Pages
2129003-100 rev 0 Feb. 28, 1995 Initial release. -
2129003-100 rev 1 Nov. 14, 2001 HIGH BRITE 17” and 21” 22
monitors added for Innova 2000.
2129003-1-100 rev 1 October 29, 2007 Conversion from Interleaf into 40
GTPS Platform
Revision History 13
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
14 Revision History
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Table of Contents
CHAPTER 1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
1 Description of the Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2 Responsibility of Customer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3 Monitor Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
CHAPTER 2 PRE-INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
1 Product Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2 Structural Requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3 Physical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4 Shipping Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
CHAPTER 3 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
1 Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2 Check on Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3 IST0218 - Wall Anchor Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4 IST0219 - GYRO TV® Assembly for DISTAR Monitors . . . . . . . . . . . . . . . . . 32
5 IST0220 - GYRO TV® Assembly for HIGH BRITE Monitors . . . . . . . . . . . . . . . 33
6 DR0567 - HIGH BRITE Monitor Replacement . . . . . . . . . . . . . . . . . . . . . 36
Table of Contents 15
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
16 Table of Contents
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Chapter 1 Introduction
1 Description of the Product
This package contains a GYRO TV® wall support for TV monitors DISTAR 38 cm (15”)
and 51 cm (20”). HIGH BRITE monitors 43 cm (17”) and 53 cm (21”) can be mounted
on the GYRO TV®.
This package is provided as an option with R & F and Vascular systems. It enhances the
X–Ray systems by adding a monitor mounted on wall and can be oriented as wished by
the operator.
Chapter 1 Introduction 17
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
2 Responsibility of Customer
“PRE–INSTALLATION” refers to work necessary to plan and prepare a site for installation
of GYRO TV® wall support, either included in a complete new system or being added to
an existing X–Ray room.
“INSTALLATION” refers to work necessary to mount the GYRO TV® wall support as requested
by the customer, in compliance with best practices required by the matter of the wall.
18 2 Responsibility of Customer
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
3 Monitor Identification
A GYRO TV® wall support is designed to support four well identified monitors only.
Chapter 1 Introduction 19
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
20 3 Monitor Identification
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Chapter 2 Pre-Installation
1 Product Identification
The GYRO TV® wall mount is supplied under reference number 2127289 for Distar Monitors
and 2305202 for High Brite Monitors comprising a wall plate and a mounting kit.
Chapter 2 Pre-Installation 21
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
2 Structural Requirement
A normal way to mount the GYRO TV ® on wall needs to have a concrete wall or at least a
concrete block or a cast stone wall with a concrete compression strength based on 25
MPa (1725 psi) at 28 days.
In case of the wall is not a concrete matter, an alternative way to mount the GYRO TV®
can be defined locally by an Architect or a Civil Engineer. Illustration 2-1 and Illustration
2-2 provide information needed.
A suggestion could be done to the Architect: a hard wood timber can by supplied by the
Customer. The GYRO TV® will be mounted on the timber by 3 slipper screws. The wood
timber needs to be put on the floor and anchored to the wall. Dimensions will not be less
than 20 cm (8”) width and 5 cm (2”) thick.
Inside the mounting kit are supplied 3 x M610L RAWL BOLTS. These RAWL BOLTS have
been defined with a concrete compression strength based on 25 MPa (1725 psi) at 28 days.
Also, refer to Table 2-1 for appropriate RAWL BOLT following concrete compression strength.
Concrete Compression 16 MPa (1100 psi) 25 MPa (1725 psi) 35 MPa (2415 psi)
Strength
Use the appropriate RAWL BOLT to meet the limits defined by any Control Board in
your area if it differs from the values above.
22 2 Structural Requirement
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
1. Highest and lowest bridle* * Items supplied Inside GYRO TV® box
2. Verrtical guide* ** Worst case
3. Mobile arm*
4. Cross support* G15: Gravity center for DISTAR 38 cm (15”) Weight = 14 kg (31 lbs)
5. Monitor tray* G20: Gravity center for DISTAR 51 cm (20”) Weight = 25 kg (55 lbs)
6. Monitor
7. Wall anchor* Pull out in worst case (safety factor = 4) and with DISTAR 51 cm (20”):
FC = 0 N (0 lbf)
Chapter 2 Pre-Installation 23
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
1. Highest and lowest bridle* * Items supplied Inside GYRO TV® box
2. Vertical guide*
3. Mobile arm* G17: Gravity center for HIGH BRITE 43 cm (17”) Weight = 25 +/– 1
kg (55 +/–2 lbs)
4. Cross support*
5. Plate support
6. Monitor G21: Gravity center for HIGH BRITE 53 cm (21”) Weight = 32 +/– 1
kg (72 +/–2 lbs)
7. Wall anchor*
8. Tilt & swivel base Pull out in worst case (safety factor = 4) and with HIGH BRITE 53
cm (21”):
FC = 0 N (0 lbf)
24 2 Structural Requirement
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
3 Physical Characteristics
Illustration 2-3: Physical Characteristics – DISTAR Monitors
Chapter 2 Pre-Installation 25
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
26 3 Physical Characteristics
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
4 Shipping Information
This section provides GYRO TV® wall mount monitor support shipping information.
Illustration 2-5: GYRO TV® Wall Mount Shipment (15” and 20” DISTAR MONITORS)
Illustration 2-6: GYRO TV® Wall Mount Shipment (17” and 21” HIGH BRITE MONITORS)
Chapter 2 Pre-Installation 27
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
28 4 Shipping Information
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
Chapter 3 Installation
1 Tools Required
You will need the following tools to install a GYRO TV®.
Chapter 3 Installation 29
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
2 Check on Site
• Make sure that the GYRO TV® is correctly supplied with no part missing.
• Make sure that the wall is a concrete wall.
The maximum pull out per anchor was calculated assuming a regular weight
concrete having a minimum compression strength of 25 MPa (1725 psi) at the time
of installation and with the anchors installed to the required hole depth of:
The anchors are designed to be installed in a predrilled hole having the same
diameter as the nominal anchor diameter.
Following depth needed by the type of bolt, a minimum thickness of concrete will
be defined by adding at least 15 mm (0.6 inches) to the hole depth.
The minimum distance between the anchor hole center and any concrete edge
must be:
• For any other case, an Architect or a Civil Engineer must provide a method to mount
the GYRO TV® wall mount.
• Make sure that you have the tools required.
30 2 Check on Site
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
3.2 Procedure
1. Use the wall plate supplied inside the GYRO TV® box. Make sure the mounting kit is
supplied. Adjust to any particular location requested by an operator in conformity with
the layout if it has been done.
2. With a plumb line define a vertical line on wall at the GYRO TV® location.
3. Using the wall plate, mark the bolt locations with a punch and a hammer.
5. Drill the mounting hole with the appropriate carbide–tipped drill for M610L or M810L
RAWL BOLT.
10. Make sure that the wall plate is correctly mounted on the wall.
3.3 Finalization
No finalization steps.
Chapter 3 Installation 31
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
4.2 Procedure
1. Mount the lowest bridle on wall plate using screws provided (4-M5 x 15 mm).
3. Fit the guide through the nylon washer in the lowest bridle.
4. Fit the mobile arm to the vertical guide. Refer to Chapter 2, Section 2, Structural
Requirement, illustration “Wall Structure Requirement - DISTAR Monitors”.
5. Insert the second nylon washer at the top of the vertical guide.
7. Block all fixation screws (8–M5 x 15 mm). Refer to Chapter 2, Section 2, Structural
Requirement, illustration “Wall Structure Requirement - DISTAR Monitors”.
8. Adjust the tray support on the mobile arm and mount it correctly.
9. Put the monitor tray on the tray support and mount it using nuts, screws and washers
provided (4 hex nuts, 4 plate washers and 4–M5 x 14 mm screws).
4.3 Finalization
No finalization steps.
5.2 Procedure
1. Mount the lowest bridle on wall plate using screws provided (4–M5 x 15 mm).
3. Fit the guide through the nylon washer in the lowest bridle.
4. Fit the mobile arm to the vertical guide. Refer to Chapter 2, Section 2, Structural
Requirement, illustration “Wall Structure Requirement - HIGH BRITE Monitors”.
5. Insert the second nylon washer at the top of the vertical guide.
7. Block all fixation screws (8–M5 x 15 mm). Refer to Chapter 2, Section 2, Structural
Requirement, illustration “Wall Structure Requirement - HIGH BRITE Monitors”.
8.
Unpack the HIGH BRITE monitor and put it on its side .
Illustration 3-1:
9.
Remove the support ( , Illustration 3-1 ) of TILT & SWIVEL BASE ( , Illustration 3-1
) by pressure on the four Catches ( , Illustration 3-1 ).
NOTE:
Cross Support ( , Illustration 3-1 ).
Chapter 3 Installation 33
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
10. Unpack the Plate support, the bag of screws and mount the support of TILT & SWIVEL
BASE on the Plate support.
Take care on both front sides – Plate support and Cross support.
Take care on what orientation is the Cross support on the Plate support.
Illustration 3-3:
12. Adjust the equipped Plate support on mobile arm and mount it correctly. Use a part of
cardboard box as tool to support the Plate support on mobile arm.
Illustration 3-4: Gyro TV® for HIGH BRITE Monitor (general view)
13. Adjust the Plate support on mobile arm and block the plate on its Spherical joint.
14. With the help of an other worker, mount and adjust the monitor on its Plate support. Lock
it as expected by the customer.
Illustration 3-5:
5.3 Finalization
No finalization steps.
Chapter 3 Installation 35
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
6.2 Overview
HIGH BRITE Monitor Replacement on GYRO TV® Wall Mount Monitor Support.
6.3 Procedure
1. Unlock the Spherical joint and tilt the Plate support to its horizontal position.
2.
Remove the equipped Plate support with the monitor . Two people are required
for this operation. Put the Plate removed on the floor.
Illustration 3-6:
Protect the front side of monitor before putting it on the floor ( , Illustration 3-6 ).
3.
Remove the Cross support from the Plate support equipped .
Illustration 3-7:
NOTE:
Support of TILT & SWIVEL BASE coming from the old monitor remains on support plate (
, Illustration 3-7 ).
Do not forget to return with the removed Monitor the Tilt & Swivel support supplied
with the new Monitor.
Protect the front side of monitor before putting it on the floor ( , Illustration 3-7 ).
4. Extract the Monitor from the Plate support equipped by pressure on the four Catches (
, Illustration 3-7 ).
6.4 Finalization
No finalization steps.
Chapter 3 Installation 37
GYRO TV® WALL MOUNT MONITOR Pre-Installation & Installation Manual
GE Healthcare Direction 2129003-1-100, Revision 1
www.gehealthcare.com