i1000SR System Quick Troubleshooting Guide
i1000SR System Quick Troubleshooting Guide
Level Sense Errors: 3000 Error Code 5209 No Reaction Vessel (RV)
• Ensure sample cup or tube is present and adequate Detected in Process path after RV Load
sample is in the cup or tube. Attempt
Reagent probe areas only: • If RV loader chute is not in position to load RVs,
reseat the RV loader chute into the wash zone
• Reagent bottle is empty: Replace empty reagents.
manifold.
• WZ probe maintenance water bottle is missing, over-
filled or filled with distilled (DI) water instead of tap
water or saline or bleach: Repeat the maintenance
procedure, ensuring the WZ probe maintenance water
bottle is in position and contains the correct volume
and solution.
Sample Pipettor:
Tighten the probe clamp finger tight.