Professional Documents
Culture Documents
NSCPL 1066 TR Nde Vgo Coil PR 02
NSCPL 1066 TR Nde Vgo Coil PR 02
NSCPL 1066 TR Nde Vgo Coil PR 02
CONTENT
PAGE NO
PARA DISCRIPTION
1 Scope 03
2 Reference Documents 03
3 Personnel Qualification 03
4 Surface Preparation 03
5 Sensitivity of Radiograph 04
6 Image Quality Indicator(IQI) 04
7 Location Markers 11
8 Screen and Film Holder 12
9 Excessive Back scatter 12
10 Film Selection 12
11 Film Identification 12
12 Radiation energy and Direction 13
13 Geometric Unsharpness 13
14 Radiographic Density 15
15 Radiographic Technique 16
16 Processing of Radiograph 17
17 Viewing of Radiograph 17
18 Quality and Evaluation of Radiograph 17
19 Acceptance Standard 18
20 Examination of Repairs 20
21 Reporting System 21
22 Safety 21
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 3 of 94
1.0 Scope
1.1 This procedure provides the general requirements, methods, techniques and
acceptance standards for the radiographic examination of butt welds in plate, formed
products, tubular products, and welds of pressure vessel joints for wall thickness
range 2.5 mm to 70 mm by using x-ray/gamma ray.
2.0 Reference Documents:
ASME Section-V, Article 2, Edition 2019
ASME Section-VIII, Div-1, Edition 2019
ASME Section IX, Edition 2019
ASME B31.3, Edition 2018
SNT – TC – 1A, Edition 2016
B224-CDU-TPL-000-MP-PR-0001 Rev.0
6-17-0003 Rev.5 Standard Specification for Coil Work For Fired Heaters
3.0 Personal Qualification
3.1 Personnel performing examination to this procedure shall be qualified and
certified according to SNT – TC – 1A
3.2 Personnel performing radiography - RT Level I / II as per a board constituted
by Bhabha Atomic Research center( BARC/AERB) government of India
radiological safety division.
3.3 Personnel evaluating radiographs - RT Level II / III as per ASNT
recommended practice SNT-TC-1A shall carry out Radiography film
interpretation & evaluation of test results.
4.3 The Finished surface of all butt welded joints may be flush with the base material
or may have reasonably uniform crowns with reinforcement not to exceed that
specified in the referencing code section.
5.0 Sensitivity of Radiograph
6.1 IQIs shall be either the hole type or the wire type IQI shall be preferred. Hole type
IQIs shall be manufactured and identified in accordance with the requirements or
alternates allowed in SE 1025.Wire type IQIs shall be manufactured and identified in
accordance with the requirements or alternates allowed in SE 747, except that the
largest wire number or the identity number may be omitted. Hole type IQI’s shall be
identified by permanently attached lead numbers, which indicate the thickness of the
IQI (designation) in thousands of an inch and in accordance with TABLE-1 (a).Wire
IQI’s shall be identified by permanently attached lead numbers and letters, which
indicate the Material group and letters (A, B, C, D) shall represent the set identification
as per TABLE-1(b).
6.2 Material. IQIs shall be selected from either the same alloy material group or grade
as identified in SE-1025 for hole type or SE-747 for wire type, or from an alloy material
group or grade with less radiation absorption than the material being radiographed.
6.3 Size. The designated hole IQI or essential wire shall be as specified in TABLE- 2(a).
A thinner or thicker hole-type IQI may be substituted for any section thickness listed in
TABLE-(2a), provided an equivalent IQI sensitivity is maintained.
a) Weld with reinforcement - The thickness on which the IQI is based is the
nominal single-wall material thickness plus the weld reinforcement thickness
estimated to be present on both sides of the weld (I.D. and O.D.). The values
used for the estimated weld reinforcement thicknesses shall be representative of
the weld conditions and shall not exceed the maximum permitted by the
referencing Code Section. Physical measurement of the actual weld
reinforcements is not required. Backing rings or strips shall not be considered as
part of the thickness in IQI selection.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 5 of 94
c) Actual Values. With regard to a) and b) above, when the actual material/weld
thickness is measured, IQI selection may be based on these known values.
6.4 Welds Joining Dissimilar Materials or Welds With Dissimilar Filler Metal:-
When the weld is of an alloy group or grade that has a radiation attenuation that
differs from the base material, the IQI material selection shall be based on the weld
metal and in accordance with Para 6.2. When the density limits of 14.2 cannot be met
with one IQI and the
exception density areas is at the interface of weld metal and base metal, the material
selection for the additional IQI’s shall be based on base material and be in
accordance with para 6.2 of this procedure.
6.5.2 Equivalent Hole-Type IQI Sensitivity. A thinner or thicker hole-type IQI than
the designated IQI may be substituted, provided an equivalent or better IQI
sensitivity, as listed in TABLE-2(b), is achieved and all other requirements for
radiography are met. Equivalent IQI sensitivity is shown in any row of TABLE-2(b),
which contains the designated IQI and hole. Better IQI sensitivity is shown in any
row of TABLE-2(b), which is above the equivalent sensitivity row. If the designated
IQI and hole are not represented in the table, the next thinner IQI row from TABLE-
2(b), may be used to establish equivalent IQI sensitivity.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 6 of 94
(50.8)
Over 2.00 (50.8) through 2.50
40 2T 12 35 2T 11
(63.5)
Over 2.50 (63.5) through 4.00
50 2T 13 40 2T 12
(101.6)
Over 4.00 (101.6)through
60 2T 14 50 2T 13
6.00(152.4)
Over 6.00 (152.4) through 8.00
80 2T 16 60 2T 14
(203.2)
Over 8.00 (203.2) through 10.00
100 2T 17 80 2T 16
(254.0)
** The thickness on which the IQI is based is the nominal single wall thickness plus the
estimated weld reinforcement permitted by referencing code section.
TABLE - 2 (b) (as per table T-283)
EQUIVALENT HOLE – TYPE IQI SENSITIVITY
Hole – Type
Equivalent Hole – Type Designation
Designation
2T Hole 1T Hole 4T Hole
10 15 5
12 17 7
15 20 10
17 25 12
20 30 15
25 35 17
30 40 20
35 50 25
40 60 30
50 70 35
60 80 40
80 120 60
100 140 70
120 160 80
160 240 120
200 280 140
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 9 of 94
6.6.2 Film side IQI(s)-Where inaccessibility prevents hand placing the IQI(s) on
the source side, the IQI(s) shall be placed on the film side in contact with the
part being examined. A lead letter “F” shall be placed adjacent to or on the
IQI(s) but shall not mask the essential hole where hole IQIs are used.
6.6.3 IQI Placement for welds- Hole IQI(s.)The IQI(s) may be placed adjacent to or
on the weld. The IQI identification number(s) and, when used, the lead
letter “F”, shall not be in the area of interest except when geometric
configuration makes it impractical.
6.6.4 IQI Placement for welds-Wire IQI(s).The IQI(s) shall be placed on the weld
so that the length of the wires are transverse to the longitudinal axis of the
weld. The IQI identification and, when used, the lead letter “F”, shall not be
in the area of interest except when geometric configuration makes it
impractical.
6.6.5 IQI Location for Materials other than welds – The IQI(s) with the IQI
identification and, when used, the lead letter “F”, may be placed in the area
of interest.
(1) For cylindrical components where the source is placed on the axis of the
component for a single exposure, at least three IQIs, spaced approximately 120° apart,
are required under the following conditions:
(-a) When the complete circumference is radiographed using one or more film
holders, or;
(-b) When a section or sections of the circumference, where the length between the
ends of the outermost sections span 240 or more degree, is radiographed using one or
more film holders. Additional film locations may be required to obtain necessary IQI
spacing.
(2) For cylindrical components where the source is placed on the axis of the
component for a single exposure, at least three IQIs, with one placed at each end of
the span of the circumference radiographed and one in the approximate center of the
span, are required under the following
Conditions:
(-a) When a section of the circumference, the length of which is greater than 120° and
less than 240° , is radiographed using just one film holder, or;
(-b) When a section or sections of the circumference, where the length between the
ends of the outermost sections span less than 240 °, is radiographed using more than
one film holder.
(3) In (1) and (2) above, where sections of longitudinal welds adjoining the
circumferential weld are radiographed simultaneously with the circumferential weld,
an additional IQI shall be placed on each longitudinal weld at the end of the section
most remote from the junction with the circumferential weld being radiographed.
(4) For spherical components where the source is placed at the center of the
component for a single exposure, at least three IQIs, spaced approximately 120° apart,
are required under the following conditions:
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 11 of 94
(-a) When a complete circumference is radiographed using one or more film holders,
or;
(-b) When a section or sections of a circumference, where the length between the ends
of the outermost sections span 240 or more degree, is radiographed using one or more
film holders. Additional film locations may be required to obtain necessary IQI
spacing.
(5) For spherical components where the source is placed at the center of the
component for a single exposure, at least three IQIs, with one placed at each end of
the span of the circumference radiographed and one in the approximate center of the
span, are required under the following conditions:
(-a) When a section of a circumference, the length of which is greater than 120° deg
and less than 240° , is radiographed using just one film holder, or;
(-b) When a section or sections of a circumference, where the length between the ends
of the outermost sections span less than 240° is radiographed using more than one
film holder.
(6) In (4) and (5) above, where other welds are radiographed simultaneously with the
circumferential weld, one additional IQI shall be placed on each other weld.
(7) For segments of a flat or curved (i.e., ellipsoidal, tori spherical, tori conical,
elliptical, etc.) component where the source is placed perpendicular to the center of a
length of weld for a single exposure when using more than three film holders, at least
three IQIs, one placed at each end of the radiographed span and one in the
approximate center of the span, are required.
(8) When an array of components in a circle is radiographed, at least one IQI shall
show on each component image.
(9) In order to maintain the continuity of records involving subsequent exposures, all
radiographs exhibiting IQIs that qualify the techniques permitted in accordance
With (1) through (7) above shall be retained.
11.1.4 Line Number with revision, Joint No. , thickness (on cover of film), line
class, welder no. XR no.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 13 of 94
11.1.7 Any weld / weld segment that have been repaired shall be marked as
R1, R2 etc. for 1st and 2nd repair etc. respectively on the radiograph.
11.1.8 Subcontractor’s name or symbol
Where
D = distance from source of radiation to weld or object being radiographed
d = distance from source side of weld or object being radiographed to the film
F = source size: the maximum projected dimension of the radiating source (or
effective focal spot) in the plane perpendicular to the distance D from the weld
or object being radiographed
Ug = geometric unsharpness
D and d shall be determined at the approximate center of the area of interest.
14.2 The density of radiograph any where through area of interest shall not vary
by more than minus 15% or plus 30% from the density through the body of
the designated hole-type IQI adjacent to the essential wire of a wire-type IQI
within the minimum or maximum allowable density range, then an additional
IQI shall be used for each exceptional area or areas and the radiograph taken.
14.3 Densitometer shall be calibrated against national standard step tablet or step
wedge calibration film traceable to national standard step tablet, at a density
closest 1.0, 2.0, 3.0 & 4.0 at least every 3 months. The record of such calibration
will be documented.
14.4 The periodic calibration verification checks shall be performed at least the
beginning of the each shift, after 8 hours of continuous use or after change
operation, which comes first.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 15 of 94
(1) DOUBLE WALL ELLIPTICAL TECHNIQUE: For welds, the radiation beam
may be offset from the plane of the weld at an angle sufficient to separate the
images of the source side and film side portions of the weld so that there is no
overlap of the areas to be interpreted. When complete coverage is required, a
minimum of two exposures taken at 90 to each other shall be made for each joint.
16.4 STOP BATH: Film shall be placed in a stop bath for 30 second before
transferring it to the fixer tank.
16.5 FIXING: Fixing time shall normally be at least twice the developing time ( not
to exceed 15 minutes in relatively fresh fixer)
16.6 REPLACEMENT OF DEVELOPER, STOP BATH AND FIXER: Developer,
Stop Bath and Fixer shall be replaced as per manufacturer’s recommendations.
16.7 Washing: Washing after developing and fixing shall be done in running water
as appropriate.
16.8 DRYING: After washing, the films shall be put to dry. A film drying cabinet
incorporating circulating heated air may be used to assist in drying.
18.1.1 Fogging
18.1.2 Process defects such as streaks, watermarks or chemical stains.
18.1.3 Scratches, finger marks, crimps, dirtiness, static marks, smudges, or tears.
18.1.4 False indications due to defective screens
B. Evaluation of Radiograph
NSCPL shall be responsible for the review, interpretation, evaluation and acceptance
of the completed radiographs to assure compliance with this procedure. The
radiographic technique details and the radiograph review form documentation shall
accompany the radiographs. Acceptance shall be completed prior to presentation of
the radiographs and accompanying documentation to the Inspector.
22. Safety:
1. Only certified personnel shall perform radiography.
2. Radioactive isotopes shall be stored in BARC approved pt.
3. The technician shall be responsible for compliance with safety rules
requirement applicable at NSCPL at ‘Sachin.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 22 of 94
Figure No.-1
LOCATION MARKERS
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 23 of 94
Figure No.-2
SINGLE WALL RADIOGRAPHIC TECHNIQUES
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 24 of 94
Figure No.-3
DOUBLE WALL RADIOGRAPHIC TECHNIQUES
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 25 of 94
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 26 of 94
SOD : mm
Placement of Back Scatter “B” on back of Cassette: Kept / Not Kept Densitometer Verified : Yes / No
CONTENTS
Sr. No Description Page No.
1 Scope 29
2 Reference Documents 29
3 Personnel Qualification 29
4 Chemical Used 29
5 Surface Preparation 30
6 Pre Cleaning 30
7 Temperature Limit 30
8 Drying After Preparation 30
9 Application of Penetrant 30
10 Penetrant Dwell Time 31
11 Removal of Excess Penetrant 31
12 Drying after Excess Penetrant Removal 31
13 Application of Developer 32
14 Interpretation 32
15 Evaluation 32
16 Repair & Re-examination 33
17 Post Examination Cleaning 34
18 Performance Demonstration 34
19 Safety 34
20 Reports 34
21 Annexure –I Essential & Non-Essential Variable 35
22 Annexure -II 36
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 29 of 94
1. SCOPE :
This procedure out lines liquid penetrant examination by solvent removable method
using visible dye penetrant for purpose of detection of surface discontinuities which are
open to surface on weldment and non- porous materials and materials of CS, Low alloy
steel, SS in plate tubular and forging form of any shape and size.
2. REFEERENC:
ASME Section V, Article-6. Ed.: 2019
ASME Code Section VIII Division 1, Ed.: 2019
ASME Code Section B31.3 Ed.: 2018
SNT-TC-1A Ed: 2016
B224-CDU-TPL-000-MP-PR-0001 Rev.0
6-17-0003 Rev.5 Standard Specification for Coil Work For Fired Heaters
B224-00-000-19-41-1002 Rev.0 Digital Radiography Requirement.
3. PERSONNEL QUALIFICATION:
Personnel employed for carrying out examination shall be qualified to certified penetrant
operator, NDE-Level I /II & the personnel employed for interpretation / evaluation shall
be qualified to NDE Level II as per SNT-TC -1A.
4. CHEMICAL USED:
4.4 The certificate shall include penetrant manufacturer’s batch no. and test results
obtained in accordance with Appendix II of Article-6, ASME Sec-V. Records of
these certificates shall be maintained
5. SURFACE PREPARATION:
I. Prior to each liquid penetrant examination, the surface to be examined and all
adjacent areas within at least 25 mm shall be dry and free of all dirt, grease,
lint, scale, welding flux, weld spatter, paint, oil and other extraneous matter
that could obscure surface openings or otherwise interfere with the
examination
II. In general, as welded, as-rolled, as-cast, or as-forged condition is acceptable
unless weld irregularities are present that could mask indications due to
discontinuities or otherwise interfere with the effectiveness of the examination,
in which case grinding/wire/wire brushing shall be used.
6. PRE-CLEANING
The examination area shall be thoroughly cleaned with cleaner and then wiped with lint-
free material. The surface shall be allowed to dry by normal evaporation for minimum 2
minutes and maximum time after pre-cleaning shall be 60 minutes. Adequate care shall be
taken to ensure proper surface condition prior to examination.
7. TEMPERATURE
The temperature and temperature-variation of test surface and penetrant shall not be
below 40F (5C) and more than 125F (52 C). If testing is to be done outside this range,
both penetrant material manufacturers acceptance & procedure demonstration is
required.
9. APPLICATION OF PENETRANT
The penetrant shall be applied either by brushing or spraying by can over the area to be
examined.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 31 of 94
Dwell-Times (Note-1)
Materials Form
(Minutes)
Castings and Welds.
Aluminum, Magnesium 5
,Steel, Brass and Bronze,
Wrought Materials:-
Titanium and High
Extrusions, Forging
Temperature Alloys 10
and Plate
Note:
1. For temperature range from 50 F to 125F (10 C to 52C). For temperature
from 40 F to 50F (5 C to 10C), minimum penetrant dwell time shall be 2
times the value listed.
14. INTERPRETATION
14.1 Final interpretation shall be made within 10 to 60 minutes after developer
coating is dry. If the surface to be examined is large enough to preclude
complete examination within the prescribed time, the examination shall be
performed in increments.
14.2 Close observation of the formation of indication (s )during application of
developer may assist in characterizing and determining the extent of indication
(s )
14.3 Surface discontinuities are indicated by bleed-out of the penetrant, which is
normally a deep red colour that stains the developer. Indication with a light
pink colour may indicate excessive cleaning. Inadequate cleaning may leave an
excessive background making interpretation difficult.
14.4 Illumination (natural or supplemental white light) of the examination surface
is required for the evaluation of indications. The minimum light intensity shall
be 100 fc (1076 lx). The light intensity, natural or supplemental white light
source, shall be measured with a white light meter prior to the evaluation of
indications or a verified light source shall be used. Verification of light sources
is required to be demonstrated only one time, documented and maintained on
file.
14.5 Light intensity of 1076 lx shall be obtained from a 23W/220V LED lamp kept at
30 cms from the surface.
15. EVALUATION
a. All indication shall be evaluated in terms of the acceptance standards of the
referring Code Section.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 33 of 94
19. SAFETY
Personnel involved in examination shall take precautions because of toxic and
inflammable nature of chemicals used. Care shall be taken so that these chemicals do
not come in prolong contact with skin.
20. REPORT
Liquid penetrant examination report shall be as per QCS Manual Exhibit No: ND-
O1Rev-0 attached herewith.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 35 of 94
APPENDIX – II
TABLE-2
Minimum and Maximum Time Limits for Steps in Penetrant Examination procedure
CONTENTS
Sr. No Description Page No.
1 Scope 39
2 References 39
3 Personnel Qualification 39
4 Examination Medium 39
5 Equipment 39
6 Surface Cleaning 40
7 Magnetic Particle 40
8 Examination Procedure 40
9 Evaluation of Indication 41
10 Acceptance Criteria 41
11 Treatment of Indication Believed Non Relevant 41
12 Examination of area from which Imperfections have 41
been Removed
13 Re-examination of Repairs areas 41
14 Post Examination Cleaning 43
15 Procedure Re-qualification 43
16 Demagnetization 43
17 Records 43
18 Safety 43
19 Annexure I( Essential & Non-Essential Variables 44
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 39 of 94
1.0 SCOPE
This procedure shall be applied for performing Magnetic Particle Examination to
detect cracks and other discontinuities on or near the surface of ferromagnetic
materials, like Carbon steel, Low alloy steel in the form of Welds, Plate, Pipes,
Forged or Extruded items using yoke and continuous method.
2.0 REFERENCES
ASME Sec-V, Article 7, Ed:2019
ASME Sec-VIII Div-1, Ed:2019
ASME B 31.3 Ed: 2018
SNT-TC-1A Ed : 2016
B224-CDU-TPL-000-MP-PR-0001 Rev.0
6-17-0003 Rev.5 Standard Specification for Coil Work For Fired Heaters
5.0 EQUIPMENT
The Magnetizing power of yokes shall be verified prior to use each day the yoke is
used. The Magnetizing power of yokes shall be verified whenever the yoke has
been damaged or repaired.
Electromagnetic yoke make Magnaflux or equivalents which are calibrated for
lifting:
4.5 Kg (10 lbs) weight AC Yoke Technique at maximum pole spacing, with contact
similar to what will be used during the examination.
Or 18 Kg (40 lb) weight DC Yoke Technique at maximum pole spacing, with contact
similar to what will be used during the examination.
Shall be considered suitable equipment in applicable AC DC mode and pole
spacing.
Magnetic Field Adequacy:
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 40 of 94
Adequacy of magnetic field shall be verified by using Pie Shaped Magnetic Particle
Field Indicator. Pie indicator shall be positioned on the surface to be examined,
such that the copper plated side is away from the inspected surface. A suitable field
strength is indicated when a clearly defined line (or lines) of magnetic particles
form(s) across the copper face of the indicator when the magnetic particles are
applied simultaneously with the magnetizing force. When a clearly defined line of
particles is not formed, the magnetizing technique shall be changed as needed.
Wet particle may be applied by either by spraying or flowing over the area to
be examined during the application of the magnetizing field current.
The minimum light intensity of the inspection surface shall be 1076 Lux at the
time of evaluation of indication. This can be obtained by using100W lamp
kept at 30 cm from the surface. The light intensity, natural or supplemental
white light source, shall be measured with a white light meter prior to
evaluation of the indications or verified light source shall be used.
Light meter shall be calibrated at least once a year or whenever meter has
been repaired. If meter has not been in use of one year or more, calibration
shall be done before being used.
Examination part surface temperature shall not exceed 50°C. Particle shall be
used within the temperature range limitation set by the manufacture of the
particle.
magnetic permeability variations (such as the edge of heat affected zones), etc. may
produce similar indications.
An indication of an imperfection may be larger than the imperfection that causes it;
however, the size of the indication is the basis for acceptance evaluation. Only
indications which have any dimension greater than 1 /16 in. (1.5 mm) shall
considered relevant.
d. A linear indication is one having a length greater than three times the width.
e. A rounded indication is one of circular or elliptical shape with a length equal to
or less than three times its width.
These acceptance standards shall apply unless other more restrictive standards are
specified for specific materials or applications within ASME SEC VIII Div. 1 &
ASME B 31.3
All surfaces to be examined shall be free of:
(a) Relevant linear indications;
(b) Relevant rounded indications greater than 3 /16 in. (5 mm);
(c) Four or more relevant rounded indications in a line separated by 1 /16 in. (1.5
mm) or less, edge to edge.
by magnetic particle method and by all other approved methods that were
originally required for the affected area, except that, when the depth of the repair is
less than the radiographic sensitivity required, radiography may be omitted.
16.0 DEMAGNETIZATION
Demagnetization is required only if specified in purchase order, drawings or
specification. This shall be performed with AC yoke, by placing the poles on
surface, moving them around the area and slowly withdrawing the yoke while it is
still energized. After demagnetization, residual magnetic field should not exceed
240 A/m anywhere in the piece absolute value, unless otherwise agreed upon with
purchaser.
17.0 RECORDS
The results of magnetic particle examination shall be recorded on the magnetic
particle examination report shown as per QCS Manual Exhibit No: ND-04 Rev-0.
18.0 SAFETY
Care shall be taken to ensure that the cables and connections etc. are not loose and
coil of the yoke is not getting overheated. Due care should be taken in handling
kerosene based particle medium to avoid risk of fire.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 44 of 94
NOTE:
3. Any change in above described Essential Variable would require Re Demonstration of NDE Procedures.
4. Any change in Non Essential Variables may be addressed by incorporating required change and revising existing
procedure without Re Demonstration.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 45 of 94
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 46 of 94
CONTENTS
Sr.No Description Page No
1 Scope 47
2 Reference 47
3 Personnel Qualification 47
4 Application 47
5 Equipment 47
6 Search Unit 49
7 Couplant 49
8 Basic Calibration Block 49
9 Identification of weld examination area 50
10 Surface Preparation 50
11 General Calibration Requirements 51
12 System Calibration for Distance Amplitude 51
13 Calibration Conformation 52
14 Examination 52
15 Technique 53
16 Evaluation 54
17 Acceptance Standard 55
18 Examination of Repairs 55
19 Procedure Re-qualification 55
20 Report of the Examination 55
21 Post Examination Cleaning 57
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 47 of 94
1.0 SCOPE
This procedure shall be applicable for Examination of butt weld and base plate by
Ultrasonic method. RT shall be preferred first; UT may be used if RT is difficult to
perform.
2.0 REFERENCE:
The personnel employed for conducting the examinations shall be qualified to Level –
II according to SNT-TC 1A.
Interpretation of the results of ultrasonic test examination shall be performed by either
Level II or Level III individuals qualified and certified as per SNT-TC-1A.
4.0 APPLICATION
4.1 The principle objective of this procedure and technique described herein is the detection,
location and evaluation of discontinuities within weld, heat affected zone and also in the
adjacent base material.
4.2 The weld shall be examined by angle beam method decide the exact job. All welds shall
be examined from both the sides of weld from an outer surface decide the exact
combination of surfaces so that full volume of the weld is covered.
4.3 Where geometry of the assembly does not allow examination from both sides, then any
of the following alternatives may be used:
Combination of angle beam (mode of wave propagation in material -shear wave)
and straight beam, (mode of wave propagation in material-longitudinal wave)
5.0 EQUIPMENT
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 48 of 94
5.1 A pulse echo type of ultrasonic instrument shall be used. The instrument shall be
capable of operation at frequencies over the range of at least 1 MHz to 5 MHz and shall
be equipped with a stepped gain control in units of 2.0dB or less.
5.2 If instrument has a damping control, it may be used if it does not reduce the sensitivity
of the examination. The reject control shall be in the ‘off’ position for all examinations,
unless it can be demonstrated that it does not affect the linearity of the examination.
5.3 Linearity Checking:
5.3.1 Analog Equipment: The requirements of screen height linearity & amplitude
control linearity shall be checked at intervals not to exceeding three months; if
equipment is not in use for3 months then it shall be checked before use and
record shall be maintained.
5.3.2 Digital Equipment: The requirement of screen height linearity & amplitude
control linearity shall be checked at intervals not to exceeding one year; if
equipment is not in use for one year then it shall be checked before use and
record shall be maintained.
5.3.3 The ultrasonic instrument’s screen height linearity shall be evaluated in
accordance with mandatory Appendix I of ASME Sec V Article 4.The ultrasonic
instrument shall provide linear vertical presentation within ±5 of the full screen
height for at least (larger indication is set) 80% of the calibrated screen
height(base line to maximum calibrated screen point(s)
5.3.4 The ultrasonic instrument’s amplitude control linearity shall be evaluated in
accordance with mandatory Appendix II, of ASME Sec V Article 4.
The ultrasonic instrument shall utilize an amplitude control, adjust the
sensitivity (gain) as per given table. The indication shall fall within the specified
limit. The setting and reading must be of 1% full screen.
INDICATION LIMITS %
INDICATION
a SET AT % dB CONTROL
OF FULL SCREEN
cOF FULL SCREEN CHANGE
HEIGHT
c 80 -6dB 35 to 45
u
80 -12dB 15 to 25
r
40 +6dB 65 to 95
a
20 +12dB 65 to 95
5.3.5 T
he horizontal linearity of the test instrument shall be qualified over the full
sound path distance to be used. The horizontal linearity of the test
instrument shall be requalified at two month intervals in each of the distance
ranges that the instrument will be used.
The horizontal sweep shall be adjusted for distance calibration to present the
equivalent of at least two plate thicknesses on the display.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 49 of 94
7.0 COUPLANT
7.1 Starch which permits satisfactory transmission of Ultrasonic Waves shall be used as
couplant However, same couplant shall be used for calibration as well as testing.
7.1 Couplants used on nickel base alloys shall not contain more than 250 PPM of sulfur.
7.2 Couplants used on austenitic stainless steel or titanium shall not contain more than 250
PPM of halides (Chlorides plus Fluorides).
8.1 Reflectors: Known reflectors shall be selected as per job requirement i.e. weld or
material velocity (i.e., side drilled holes, flat bottom holes, notches etc.) shall be used to
establish primary reference, responses of the equipment.
8.2 Material: The material from which the block is fabricated shall be of the same product
form, same material specification or equivalent P Number grouping (heat treatment
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 50 of 94
shall be same material) as one of the materials being examined. For the purposes of this
procedure P Nos. 1, 3, 4 & 5 materials are considered equivalent.
8.3 Quality: Prior to fabrication, the block material shall be completely examined with a
straight beam search unit. Areas that contain an indication exceeding the remaining
back wall reflection shall be excluded from the beam paths required reach various
calibration reflectors.
8.4 Surface Finish: The finish on the scanning surfaces of the block shall be representative
of the scanning surface finishes on the component to be examined.
8.5 Calibration Block:
Annexure1 (shall be as per actual requirement) shows block configuration with
hole size and location each weld thickness on the component must be
represented by a block having a thickness relative to the component weld.
Where the block thickness is ±1 inch (25mm) spans two of the weld thickness
range shown in the figure, the block’s use shall be acceptable in those portion of
each thickness range covered by 1 inch (25mm).The hole diameter shall be in
accordance with the thickness of the block. Where two or more base material
thickness is involved, the calibration block thickness shall be sufficient to
contain the entire examination beam path. The hole shall be drilled to a depth of
1-1/2 Inch (38mm) minimum but where possible the depth shall be 2 inches
(51mm)
8.6 Alternative blocks: Existing calibration blocks may be used for similar types of
examination provided the sound path distance(s) to the block’s reflector(s) is(are) within
¼ in. (6mm) of what is required and the side-drilled hole(s) is(are) the same or a smaller
diameter than what is required
9.1 Weld Locations: Weld locations and their identification shall be recorded on a weld
map or in an identification plan.
9.2 Marking: If welds are to be permanently marked, low stress stamps may be used.
Markings applied after final stress relief of the component shall not be deeper than 3/64
in. (1.2mm).
9.3 Reference System: Each weld shall be located and identified by a system of reference
points. The system shall permit identification of each weld center line and designation
of regular intervals along the length of the weld.
10.1 Base Metal: The base metal on each side of weld shall be free of weld spatters, surface
irregularities, or foreign matters that might interfere with the examination.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 51 of 94
10.2 Weld Metal: Where the weld metal interferes with the examination, the weld shall be
prepared as needed to permit the Ultrasonic Examination.
11.1 Ultrasonic system: Calibration shall include the complete ultrasonic system and shall be
performed prior to use of the system in the thickness range under examination.
Calibrations shall be performed from the surface corresponding to the surface of the
component from which examination will be performed.
11.2 Examination shall be performed with same couplant, same contact wedges, and same
instrument controls which affects the linearity, used for calibrations. The temperature
differential between the calibration block and examination surface shall be within 14 deg
C.
11.3 Calibration surface: Calibrations shall be performed from the surface (clad or unclad;
convex or concave) corresponding to the surface of the component from which the
examination will be performed.
12.1 Calibration blocks: Calibrations shall be performed using the standard calibration
block e.g. IIW V1/V2 shall be used
12.2 Technique:
12.2.1 The angle beam shall be directed towards the calibration reflector that yields
maximum response in area of interest.
12.2.2 Adjust the sensitivity control to provide an 80% ± 5% of FSH echo from the hole.
Mark the peak indication on the screen.
12.2.3 Without changing gain control, manipulate the search unit for maximum
response from other calibration reflectors at their beam paths to generate the
distance amplitude correction (DAC) curve.
12.2.4 These calibrations shall establish both the distance range calibration and the
distance amplitude correction.
12.3 Angle beam calibration:
12.3.1 As applicable, calibration shall provide the following information.
(a) Distance range calibration
(b) Distance amplitude correction in area of interest
(c) Echo amplitude measurement from the surface notch in the basic
calibration block.
12.3.2 When an electronic distance-amplitude correction device is used, the primary
reference responses from the basic calibration block shall be equalized over the
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 52 of 94
13.1 System Changes: When any part of the examination system is changed, a calibration
check shall be made on the basic calibration block to verify that distance range points
and sensitivity settings satisfy the requirement of clauses 13.3 & 13.4
13.2 Periodic Examination Checks: A calibration check on at least one of the basic reflectors
in the basic calibration block or a check using a simulator shall be made at the finish of
each examination or series of similar examinations, and when examination personnel
are changed. Distance range points and sensitivity settings shall satisfy the requirements
of 13.3 & 13.4
13.3 Distance Range Points: If any distance range point has moved on the sweep line by
more than 10% of the distance reading or 5% of full sweep, whichever is grater, correct
the distance range calibration and note the correction in the examination record. All
recorded indications since the last valid calibration or calibration check shall be re–
examined and their values shall be changed on the data sheets or re–recorded.
13.4 Sensitivity Settings: If any sensitivity setting has changed by more than 20% or 2dB of
its amplitude, correct the sensitivity calibration and note the correction in the
examination record. If the sensitivity setting has decreased, all data sheets since the last
valid calibration check shall be marked void and the area covered by the voided data
shall be re–examined. If the sensitivity setting has increased, all recorded indications
since the last valid calibration or calibration check shall be re–examined and their values
shall be changed on the data sheets or re–recorded.
14.0 EXAMINATION
14.1 Scanning of the adjacent base metal by normal beam shall be performed to detect
reflectors that might affect interpretation of angle beam results and is not to be used for
acceptance or rejection of weld. Location and areas of such reflectors shall be recorded.
14.2 Examination Coverage: The volume to be scanned shall be examined by moving the
search unit over the scanning surface so as to scan the entire examination volume for
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 53 of 94
each required search unit. Each pass of the search unit shall overlap a minimum of 10%
of the active transducer dimension (piezoelectric element) parallel to the direction of
scan indexing.
14.3 Oscillation of the search unit is permitted if it can be demonstrated that overlapping
coverage is provided. The rate of search unit movement shall not exceed 150mm/s
(6in/s).
14.4 Scanning Sensitivity Level: Distance amplitude technique the scanning sensitivity level
shall be set a minimum of 6dB higher than the reference level gain setting. The gain shall
be increased to the extent that no calibration reflector indications are less than 40%FSH.
15.0 TECHNIQUE:
15.1 Weld Joint Distance Amplitude Technique: When the referencing Code section
specifies a distance amplitude technique, weld joints shall be scanned with an angle
beam search unit in both parallel and transverse directions (4 scans) to the weld axis.
Before performing the angle beam examinations, a straight beam examination shall be
performed on the volume of base material through which the angle beams will travel to
locate any reflectors that can limit the ability of the angle beam to examine the weld
volume.
15.1.1 Angle Beam Technique
(a). Beam Angle: The search unit and beam angle selected shall be 45 degree or
an angle appropriate for the configuration being examined and shall be
capable of detecting the calibration reflectors, over the required angle beam
path.
(b). Reflectors Parallel to the Weld Seam: The angle beam shall be directed
at approximate right angle to the weld axis from both sides of the weld
on the same surface when possible. The search unit shall be manipulated
so that the ultrasonic energy passes through the required volume of weld
and adjacent base material.
(c). Reflectors Transverse to the Weld Seam: The angle beam shall be
directed essentially parallel to the weld axis. The search unit shall be
manipulated so that the ultrasonic energy passes through the required
volume of weld and adjacent base material. The search unit shall be rotated
180° and the examination repeated. If the weld cap is not machined or
ground flat, the examination shall be performed from base metal on both
side of the weld cap in both weld axis directions
(d). Restricted Access Welds: Welds that cannot be fully examined from two
directions using the angle beam technique (e.g., Corner and Tee joints)
shall also be examined, if possible, with a straight beam technique. These
areas of restricted access shall be noted in the examination report.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 54 of 94
(e). Inaccessible Welds: Welds that cannot be examined from at least one side
(edge) using the angle beam technique shall be noted in the examination
report.
15.1.2 Straight Beam Technique:
Whenever Straight Beam examination is part of the weld testing the scanning
shall be performed at a gain setting of at least two times the primary reference
level. Evaluation shall be performed at primary levels.
15.1.3 Discontinuity sizing shall be done as per 6dB drop method.
16.0 EVALUATION:
maximum feasible volume is examined, and shall be noted in the record of the
examination .
16.2.5 Alternative Evaluations: Reflector dimensions exceeding the referencing Code
Section requirements may be evaluated to any alternative standards provided by
the referencing Code Section.
17.1 Following acceptance criteria meets the requirement of ASME Sec VIII. Div. 1
Appendix-12:Discontinuities are unacceptable if the amplitude exceeds the
reference level and have a length which exceeds the following:
a) 6mm for T up to 19 mm
b) 1/3 T for T from 19mm to 57mm
c) 19mm for T over 57mm
Where T is the thickness (excluding any allowable reinforcement) of the weld
being examined. If the weld joins two members having different thickness at
the weld, T is the thinner of the two thicknesses.
17.2Where discontinuities are interpreted to be cracks, lack of fusion or incomplete
penetration, they are unacceptable regardless of discontinuity size or signal
amplitude.
All repaired areas shall be re–examined by the same procedure as used for the original
detection of the discontinuity.
If there is any change in value of any essential or non essential procedure shall be
revised, if there is any change in essential variables the procedure shall be re-qualified.
20.1Recording indications:
recording examples for angle and straight beam search units. Other techniques
may be used.
20.2 Examination Records: For each ultrasonic examination, the following information shall
be recorded:
(a) Procedure identification and revision;
(b) Ultrasonic instrument identification (including manufacturer’s serial number);
(c) Search unit(s) identification (including manufacturer’s serial number, frequency,
and size);
(d) Beam angle(s) used;
(e) Couplant used, brand name or type;
(f) Search unit cable(s) used, type and length;
(g) Special equipment when used (search units, wedges, shoes, automatic
Scanning equipment, recording equipment etc).
(h) Computerized program identification and revision when used;
(i) Calibration block identification;
(j) Simulation block(s) and electronic simulator(s) identification when used;
(k) Instrument reference level gain and, if used, damping and reject setting(s);
(l) Calibration data [including reference reflector(s), indication amplitude(s) and
distance reading(s)];
(m) Data correlating simulation block(s) and electronic simulator(s), when used, with
initial calibration;
(n) Identification and location of weld or volume scanned;
(o) Surface from which examination was conducted, including surface condition;
(p) Map or record of rejectable indication detected or area cleared;
(q) Areas of restricted access or inaccessible welds;
(r) Examination personnel identity and, when required by referencing Code Section,
qualification level;
(s) Date of examination. Note:- Items (b) through (m) may be included in a separate
calibration record provided the calibration record identification is included in the
examination record.
20.3 Report: A report of the examinations shall be made. The report shall include those
records. The report shall be filed and maintained in accordance with the referencing
Code section. Exhibit No: ND-03 R1 shall be used for reporting the examination results.
20.4 Storage Media: Storage media for computerized scanning data and viewing software
shall be capable of securely storing and retrieving data for the time period specified by
the referencing Code section
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 57 of 94
When Post Examination cleaning is required, it shall be done as soon as practical after
evaluation and documentation using a process that does not adversely affect the part.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 58 of 94
ANNEXURE-I
NON-PIPING CALIBRATION BLOCK
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 59 of 94
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 60 of 94
Contents
Para Description Page no.
1.0 SCOPE 61
6.0 RECORDS 65
7.0 SAFETY 65
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 61 of 94
1. SCOPE:
This procedure covers the visual examination of welds of pressure vessels by direct
visual technique, for examination of weld fitments prior to welding, and
examination of completed welds. The purpose of these visual examinations is to
detect flaws like cracks, slags, undercut, overlaps etc. that may be present on the
surface of weldment.
2. REFERENCING DOCUMENTS:
ASME Section V Article 9, Ed: 2019.
ASME B 31.3 Ed.: 2018.
NSCPL Written Practice- NSCPL-NDT-WP-01 Latest revision
B224-CDU-TPL-000-MP-PR-0001 Rev.0
6-17-0003 Rev.5 Standard Specification for Coil Work For Fired Heaters
3. PERSONNEL QUALIFICATION:
3.1 Certified NDE personnel in VT method in accordance with our Neo Structo
Construction Pvt. Ltd (NSCPL) written practice NSCPL-NDT-WP-01 (latest
revision) shall be used for conducting the examination.
3.2 When referencing document does not refer Article – 9 of ASME Sec. V directly,
qualification of NDE personnel may be through satisfactory demonstration of
competence in use of the established procedure. Such qualified NDE personnel
shall be employed for conducting examinations as per that procedure only.
3.3 Personnel shall have an annual vision test to assure natural or corrected near
distance acuity such that they are capable of reading standard J-1 letters on
standard Jaeger test type charts for near vision. Equivalent near vision tests are
acceptable.
4. REQUIREMENTS OF PROCEDRURE:
than 30 to the surface to be examined. Use of mirrors to improve the angle of
vision is permitted. Magnifying lens may be used to assist the examination
The light source, technique used and light level verification is required to be
demonstrated only one time, documented and maintained on file.
6. RECORDS:
ESSENTIAL VARIABLES
APPENDIX:-I- VT
NOTE:
5. Any change in above described Essential variable would require re-demonstration
of NDE procedures.
6. Any change in Non-Essential variables may be addressed by incorporating
required change and revising existing procedure without re-demonstration.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 67 of 94
REMARKS:
Sign:
Date
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 68 of 94
DOC. NO.-NSCPL-1066-TR-NDE-CDU-VDU-PR-DRT-06CONTENT
PAGE NO
Sr. No. DISCRIPTION
1 Scope 69
2 Reference Documents 69
3 Personnel Qualification 69
4 Surface Preparation 69
5 Sensitivity of Radiograph 70
6 Image Quality Indicator(IQI) 70
7 Location Markers 78
8 Screen and Film Holder 78
9 Excessive Back scatter 78
10 Film Selection(Phosphor Imaging Plate) 78
11 Image Identification 79
12 Radiation energy and Direction 79
13 Geometric Unsharpness 80
14 Radiographic Density 80
15 Radiographic Technique 80
16 Processing 82
17 Viewing Facilities 82
18 Quality and Evaluation of Radiograph 82
19 Acceptance Standard 84
20 Examination of Repairs 87
21 Reporting System 87
22 Safety 88
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 69 of 94
1 Scope
1.1 This procedure provides the general requirements, methods, techniques and
acceptance standards for the radiographic examination of butt welds in plate,
formed products, tubular products, and welds of pressure vessel joints for wall
thickness range 2.5 mm to 70 mm by using x-ray/gamma ray.
1.2 This procedure provides requirements for using phosphor imaging plate
(photostimulbale luminescent phosphor) as an alternative to film radiography.
The term film, as used within Article 2, applicable to performing radiography in
accordance with this procedure, refers to phosphor imaging plate.
2.0 Reference Documents:
ASME Section-V, Article 2, Edition 2019
ASME Section-VIII, Div-1, Edition 2019
ASME Section IX, Edition 2019
ASME B31.3, Edition 2018
SNT – TC – 1A, Edition 2016
B224-CDU-TPL-000-MP-PR-0001 Rev.0
6-17-0003 Rev.5 Standard Specification for Coil Work For Fired Heaters
B224-00-000-19-41-1002 Rev.0 Digital Radiography Requirement.
ISO 17636-2 Non-destructive testing of welds- Radiographic testing- Part 2: X-ray
and gamma- ray techniques with digital detectors.
that images of surface irregularities cannot mask or be confused with the image of
any discontinuity on the resulting radiograph.
4.2 The weld ripples or weld surface irregularities on both the inside (where
accessible) and outside shall be removed by any suitable process to such a degree
that the images of surface irregularities cannot mask or be confused with the
image of any discontinuities on the radiograph.
4.3 The Finished surface of all butt welded joints may be flush with the base
material or may have reasonably uniform crowns with reinforcement not to
exceed that specified in the referencing code section.
5.0 Sensitivity of Radiograph
6.1 IQIs shall be either the hole type or the wire type IQI shall be preferred. Hole type
IQIs shall be manufactured and identified in accordance with the requirements or
alternates allowed in SE 1025.Wire type IQIs shall be manufactured and identified in
accordance with the requirements or alternates allowed in SE 747, except that the
largest wire number or the identity number may be omitted. Hole type IQI’s shall be
identified by permanently attached lead numbers, which indicate the thickness of the
IQI (designation) in thousands of an inch and in accordance with TABLE-1 (a).Wire
IQI’s shall be identified by permanently attached lead numbers and letters, which
indicate the Material group and letters (A, B, C, D) shall represent the set identification
as per TABLE-1(b).
6.2 Material. IQIs shall be selected from either the same alloy material group or grade
as identified in SE-1025 for hole type or SE-747 for wire type, or from an alloy material
group or grade with less radiation absorption than the material being radiographed.
6.3 Size. The designated hole IQI or essential wire shall be as specified in TABLE- 2(a).
A thinner or thicker hole-type IQI may be substituted for any section thickness listed in
TABLE-(2a), provided an equivalent IQI sensitivity is maintained.
d) Weld with reinforcement - The thickness on which the IQI is based is the
nominal single-wall material thickness plus the weld reinforcement thickness
estimated to be present on both sides of the weld (I.D. and O.D.). The values
used for the estimated weld reinforcement thicknesses shall be representative of
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 71 of 94
the weld conditions and shall not exceed the maximum permitted by the
referencing Code Section. Physical measurement of the actual weld
reinforcements is not required. Backing rings or strips shall not be considered as
part of the thickness in IQI selection.
f) Actual Values. With regard to a) and b) above, when the actual material/weld
thickness is measured, IQI selection may be based on these known values.
6.4 Welds Joining Dissimilar Materials or Welds With Dissimilar Filler Metal:-
When the weld is of an alloy group or grade that has a radiation attenuation that
differs from the base material, the IQI material selection shall be based on the weld
metal and in accordance with Para 6.2. When the density limits of 14.2 cannot be met
with one IQI and the
exception density areas is at the interface of weld metal and base metal, the material
selection for the additional IQI’s shall be based on base material and be in
accordance with para 6.2 of this procedure.
6.5.2 Equivalent Hole-Type IQI Sensitivity. A thinner or thicker hole-type IQI than
the designated IQI may be substituted, provided an equivalent or better IQI
sensitivity, as listed in TABLE-2(b), is achieved and all other requirements for
radiography are met. Equivalent IQI sensitivity is shown in any row of TABLE-2(b),
which contains the designated IQI and hole. Better IQI sensitivity is shown in any
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 72 of 94
row of TABLE-2(b), which is above the equivalent sensitivity row. If the designated
IQI and hole are not represented in the table, the next thinner IQI row from TABLE-
2(b), may be used to establish equivalent IQI sensitivity.
** The thickness on which the IQI is based is the nominal single wall thickness plus the
estimated weld reinforcement permitted by referencing code section.
TABLE - 2 (b) (as per table T-283)
EQUIVALENT HOLE – TYPE IQI SENSITIVITY
Hole – Type
Equivalent Hole – Type Designation
Designation
2T Hole 1T Hole 4T Hole
10 15 5
12 17 7
15 20 10
17 25 12
20 30 15
25 35 17
30 40 20
35 50 25
40 60 30
50 70 35
60 80 40
80 120 60
100 140 70
120 160 80
160 240 120
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 75 of 94
6.9.2 Film side IQI(s)-Where inaccessibility prevents hand placing the IQI(s) on
the source side, the IQI(s) shall be placed on the film side in contact with the
part being examined. A lead letter “F” shall be placed adjacent to or on the
IQI(s) but shall not mask the essential hole where hole IQIs are used.
6.9.3 IQI Placement for welds- Hole IQI(s.)The IQI(s) may be placed adjacent to or
on the weld. The IQI identification number(s) and, when used, the lead
letter “F”, shall not be in the area of interest except when geometric
configuration makes it impractical.
6.9.4 IQI Placement for welds-Wire IQI(s).The IQI(s) shall be placed on the weld
so that the length of the wires are transverse to the longitudinal axis of the
weld. The IQI identification and, when used, the lead letter “F”, shall not be
in the area of interest except when geometric configuration makes it
impractical.
6.9.5 IQI Location for Materials other than welds – The IQI(s) with the IQI
identification and, when used, the lead letter “F”, may be placed in the area
of interest.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 76 of 94
(1) For cylindrical components where the source is placed on the axis of the
component for a single exposure, at least three IQIs, spaced approximately 120° apart,
are required under the following conditions:
(-a) When the complete circumference is radiographed using one or more film
holders, or;
(-b) When a section or sections of the circumference, where the length between the
ends of the outermost sections span 240 or more degree, is radiographed using one or
more film holders. Additional film locations may be required to obtain necessary IQI
spacing.
(2) For cylindrical components where the source is placed on the axis of the
component for a single exposure, at least three IQIs, with one placed at each end of
the span of the circumference radiographed and one in the approximate center of the
span, are required under the following
Conditions:
(-a) When a section of the circumference, the length of which is greater than 120° and
less than 240° , is radiographed using just one film holder, or;
(-b) When a section or sections of the circumference, where the length between the
ends of the outermost sections span less than 240 °, is radiographed using more than
one film holder.
(3) In (1) and (2) above, where sections of longitudinal welds adjoining the
circumferential weld are radiographed simultaneously with the circumferential weld,
an additional IQI shall be placed on each longitudinal weld at the end of the section
most remote from the junction with the circumferential weld being radiographed.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 77 of 94
(4) For spherical components where the source is placed at the center of the
component for a single exposure, at least three IQIs, spaced approximately 120° apart,
are required under the following conditions:
(-a) When a complete circumference is radiographed using one or more film holders,
or;
(-b) When a section or sections of a circumference, where the length between the ends
of the outermost sections span 240 or more degree, is radiographed using one or more
film holders. Additional film locations may be required to obtain necessary IQI
spacing.
(5) For spherical components where the source is placed at the center of the
component for a single exposure, at least three IQIs, with one placed at each end of
the span of the circumference radiographed and one in the approximate center of the
span, are required under the following conditions:
(-a) When a section of a circumference, the length of which is greater than 120° deg
and less than 240° , is radiographed using just one film holder, or;
(-b) When a section or sections of a circumference, where the length between the ends
of the outermost sections span less than 240° is radiographed using more than one
film holder.
(6) In (4) and (5) above, where other welds are radiographed simultaneously with the
circumferential weld, one additional IQI shall be placed on each other weld.
(7) For segments of a flat or curved (i.e., ellipsoidal, tori spherical, tori conical,
elliptical, etc.) component where the source is placed perpendicular to the center of a
length of weld for a single exposure when using more than three film holders, at least
three IQIs, one placed at each end of the radiographed span and one in the
approximate center of the span, are required.
(8) When an array of components in a circle is radiographed, at least one IQI shall
show on each component image.
(9) In order to maintain the continuity of records involving subsequent exposures, all
radiographs exhibiting IQIs that qualify the techniques permitted in accordance
With (1) through (7) above shall be retained.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 78 of 94
Where
D = distance from source of radiation to weld or object being radiographed
d = distance from source side of weld or object being radiographed to the film
F = source size: the maximum projected dimension of the radiating source (or
effective focal spot) in the plane perpendicular to the distance D from the weld
or object being radiographed
Ug = geometric unsharpness
D and d shall be determined at the approximate center of the area of interest.
15.3.1 Single Wall Viewing: For material and for welds in components a technique may
be used in which the radiation passes through two walls and only the weld (material)
on the film side wall is viewed for acceptance on the radiograph. When the complete
coverage is required for circumferential welds (materials) a minimum of three
exposures taken 120 to each other shall be made.
15.3.2 Double Wall Viewing: For material and for welds in components 3 ½”(88.9 mm)
or less in nominal outside diameter, a technique may be used in which the radiation
passes through two walls and the weld (material) in both wall is viewed for acceptance
on the same radiograph. For double wall viewing only source side IQI shall be used.
(1) DOUBLE WALL ELLIPTICAL TECHNIQUE: For welds, the radiation beam
may be offset from the plane of the weld at an angle sufficient to separate the
images of the source side and film side portions of the weld so that there is no
overlap of the areas to be interpreted. When complete coverage is required, a
minimum of two exposures taken at 90 to each other shall be made for each joint.
Viewing facilities shall provide subdued background lighting of an intensity that will
not cause reflections, shadows or glare on the monitor that interfere with the
interpretation process.
The display used for image evaluation shall fulfill the following minimum
requirements:
a. Minimum brightness of 250 cd/m2
b. Display of at least 256 shades of grey.
c. Minimum displayable light intensity ration of 1:250
d. Display of at least 1 million pixels of a pixel size <0.3 mm
System-induced artifacts
The digital image shall be free of system-induced artifacts in the area of interest that
could mask or be confused with the image of any discontinuity
B. Evaluation of Radiograph
All the radiographs shall be examined and interpreted to assure the compliance with
the referencing code.
Image Brightness
The image brightness through the body of the hole-type IQI or adjacent to the
designated wire of the wire-type IQI, shall be judged to be equal to or greater than
the image brightness in the area of interest for a negative image format. This image
brightness requirement is reversed for a positive image format. Additionally, the
requirements of ASME Sec V, Article 2, T-282 are not applicable to phosphor imaging
plate radiography.
MEASURING SCALE
The measuring scale used for interpretation shall be capable of providing dimensions of
the projected image.
The measurement scale tool shall be based upon a known dimensional comparator that
is placed on the cassette or a comparator scale placed on the imaging plate during
exposure or prior to processing
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 83 of 94
INTERPRETATION
Final radiography interpretation of the area of interest shall be performed within the
identified IQI image contrast and brightness values or, if multiple IQIs are used, the
overlapping portions of the identified contrast and brightness values for the intervening
thickness ranges, The IQI and the area of interest shall be of the same image format
(positive or negative). Additionally, Where applicable
(a) The contrast and brightness range that demonstrates the required IQI sensitivity shall
be used to qualify multiple thickness, the overlapping portions of each IQI's determined
contrast and brightness range shall be considered valid for interpretation of intervening
thickness. Pixel intensity values may also be used to quantify image brightness
comparisons.
(b) The digital image may be viewed and evaluated in a negative or positive image
format.
(c) Independent areas of interest of the same image may be displayed and evaluated in
differing image formats, provided the IQI and the area of interest are viewed and
evaluated in the same format.
Each computed radio graphed shall be evaluated for its minimum gray values in
accordance with Annex D.1 of ISO 17636-2.
The minimum SNR measured shall be above those from Table D.1 of Annex D.1 of ISO
17636-2 with a SNR of 100 for Iridium 192.
Measuring of the gray scale should be performed in the heat affected zone and be
plotted and reported for each radiographic image.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 84 of 94
NSCPL shall be responsible for the review, interpretation, evaluation and acceptance
of the completed radiographs to assure compliance with this procedure. The
radiographic technique details and the radiograph review form documentation shall
accompany the radiographs. Acceptance shall be completed prior to presentation of
the radiographs and accompanying documentation to the Inspector.
22. Safety:
4. Only certified personnel shall perform radiography.
5. Radioactive isotopes shall be stored in BARC approved pt.
6. The technician shall be responsible for compliance with safety rules
requirement applicable at NSCPL at ‘Sachin.
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 89 of 94
Figure No.-1
LOCATION MARKERS
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 90 of 94
Figure No.-2
SINGLE WALL RADIOGRAPHIC TECHNIQUES
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 91 of 94
Figure No.-3
DOUBLE WALL RADIOGRAPHIC TECHNIQUES
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 92 of 94
NSCPL-1066-TR-NDE-VGO-
NDE PROCEDURE Document No. COIL-PR-02
FOR
Rev. No. 0
VGO-COIL Date 06.10.2020
PRESSURE PART Page 93 of 94
SOD : mm
Placement of Back Scatter “B” on back of Cassette: Kept / Not Kept Densitometer Verified : Yes / No