Professional Documents
Culture Documents
Design of Measurement Device For Impedance Cardiography: Michaela Snajdarova, Stefan Borik, Ivo Cap
Design of Measurement Device For Impedance Cardiography: Michaela Snajdarova, Stefan Borik, Ivo Cap
Cardiography
Vgen Vx
I R1 ,
R1
Vout Vx
I R2 ,
R2
III. RESULTS
D. Impedance Cardiography Technique The measurement system based on the single chip
ICG is a noninvasive method for monitoring the cardiac impedance analyzer is shown in Fig.9. The accuracy of the
output (CO), stroke volume (SV) and other hemodynamic device was tested on five human subjects without heart
parameters. This technique is based on electrical properties of diseases. ICG measurements were carried out both in the sitting
biological tissues which behave in opposition to an externally and resting conditions. The frequency of the excitation signal
applied current flow. This method consists of applying an was set to 70 kHz and the amplitude of the injected current was
alternating current to the subject and sensing the changes of 4.2 mA.
thoracic impedance caused by blood volume changes in the
aorta during the cardiac cycle. The basic parameter SV informs
us about volume of blood pumped by left ventricle during one
contraction cycle [7],[8].
Four spot Ag-AgCl electrodes were used in the ICG
measurement (see Fig.7.). The sinusoidal current from the
VCCS was injected into the patient's body via the application
electrode placed on the fourth cervical vertebra. The second
electrode, located on the ninth thoracic vertebra, was
connected to the ground and the sensing electrodes were
placed at the beginning and at the end of the sternum [9].
These sensing electrodes were connected to the inputs of the
instrumentation amplifier AD8422 which is connected to the
AD5933 via resistor R3 (see Fig. 8.).
The impedance device measures the ICG signal Fig.9. The constructed device prototype.
simultaneously with the electrocardiographic signal (ECG).
The electrical activity of the heart was recorded via two
additional spot electrodes placed near the heart (modified one- Two biological signals – ICG and ECG – are displayed
lead ECG). The sensing potentials were amplified and online in the software application (see Fig. 10.). The
connected to the inputs of the INA122 operational amplifier. measurement starts with the selection of the injected current
The output signal was digitized by the ATmega328P frequency and pressing the START button. Once set, the
microcontroller and the data was sent together with the ICG frequency cannot be changed during the measurement. The
signal to the PC. data is further processed when the measurement is complete
and sample resulting signals are shown in Fig. 11. Feature
extraction allows the computation of several hemodynamic
parameters [10] such as SV by the following equation:
ρ b L2 dZ
SV 2
LVET , (6)
Z0 dt max
REFERENCES
# Hemodynamic parameters
Gender Thoracic
LVET CO
F-Female impedance SV [ml/beat]
[s] [ml/min]
M-Male [Ω]