Mini-GentleLASE SYSTEM PERFORMANCE & USAGE DATA

Customer Name ________________________ Date __________ Laser S/N __________

Service Engineer ____________________

Energy Meter Serial Number _______________ Calibration Due Date _________________

------------------------------------------ BEFORE SERVICING DATA -------------------------------------------

SOFTWARE REV _______ LASER HEAD S/N _________ TOTAL PULS _____________
Head Detector, fiber installed
CANISTER MASS DI TEMP HIGH VOLTAGE @ 60J
_________________ ____________C __________________V

Into Meter Into Meter


DELIVERED ENERGY DELIVERY TRANS. CP CP Transmission
___________J _________% ___________J ____________%

LIST FAULTS:

--------------------------------------------- AFTER SERVICING DATA -------------------------------------------

SOFTWARE REV ________ LASER HEAD S/N _________

LINE VOLTAGE____ 180-250v DI WATER TEMP______ 65±3ºC DCD PRES_____120± 5 PSI

HV REF@1500VDC___________1500±5VDC HV SMP@1500VDC___________1500±5VDC

BUBBLE SENSOR Cryo 7.75±.0.5VDC No Cryo >11VDC


12-18MM H/P
6-10MM H/P
Canister
Head Detector, Fiber installed Into Meter
HIGH VOLTAGE @ 60 J HEAD _________ <1700 (≤25K Head Pulses) DELIVERED ENERGY_________J
<1800 (25K-50K) 12-18MM DELIVERY TRANS. ______% >80%

CP Transmission: ___________% <1900 (>50K) 6-10MM DELIVERY TRANS. ______% >80%

AIMING BEAM VISIBLE AT LEVEL 1 W/ 18 MM SPOT SIZE THROUGH GOGGLES: _____________

VERIFY 20 msec, 3ms DELAY, 0 POST SPRAY DCD PULSE FILLS 12mm CIRCLE: ____________
Must use MGL/GL+ brown ULTEM 12 mm distance gauge, 7121-00-9240 Rev B or higher. Do not use aiming beam.

TOTAL PULSES ______________ HEAD PULSES ______________

CLEAR FAULT LIST AND RESET PULSE COUNTER(S) IF APPLICABLE

10-400-00144 Revision C Candela Corporation PROPRIETARY Page 1 of 1