Professional Documents
Culture Documents
Experiment 3 Determination of Caffein in Soft Drink
Experiment 3 Determination of Caffein in Soft Drink
Experiment 3 Determination of Caffein in Soft Drink
Acknowledgement of Received
-------------------- --------------------------------------------------------
-----------
-----------------------------------------------
Student’s Copy
Lecturer Chop / Sign
Submitted by : ________________________________
Experiment No. : _______________________________
Subject : Physical Chemistry CLB 10703
1.0 SUMMARY
In this experiment, the presence of caffeine and benzoic acid in soft drink sample was
acknowledged and the amount of caffeine in soft drink sample was determined. Compound
separately retained in the stationary phase each gradually reacted to produce a set of peaks
along the timeline. Each component of the mixture reaches the detector at a different time and
produce a signal at a specified time called the retention time. The area under a peak is related
to the amount of the component present the mixture. In this experiment, serial dilution also
will be prepared to be as standard caffeine and to determine if caffeine was present in the
soda sample by use retention time.
Other than that, by using the concentration to peak area relationship, the concentration of
caffeine in the soda sample can be determined. The peak of caffeine appeared after second
and by measuring caffeine peaks of the standards, the amount of caffeine in a sample can be
determined. The area under the caffeine graph increase when the concentration was
increased. This proved that the mixture contained a higher quantity of caffeine.
2.0 OBJECTIVES
The analytical technique used for separating mixtures into their components due to their
molecular composition and structure is called chromatography. In recent years, the
technology used in chromatography analysis has greatly improved. The improvement in
this area have led to the common use High Performance Liquid Chromatography that
allows for the very efficient separation of small amounts of the components of a mixture.
It is essentially the same as basic liquid chromatography but HPLC allows for far better
separation of the components of a mixture. It contained tiny particle of only 5 μ m diameter
in the column of the HPLC ensuring a very large surface area to which molecules may
absorb. These particles comprise the stationary phase of the chromatographic system. The
mobile phase solvent must be forced through the column under very high pressure because
these tiny particles are so tightly packed. The HPLC are connected to a detector and a
computer to give signal when eluents are coming off the column and fractions should be
collected.
Below is an image of the pathway used in HPLC. The mobile phase we used contained
Methanol. The high pressure provided by the pump solvent manager as the sample is
injected. It then passes with the mobile phase down through the HPLC column where the
interaction between both phases separates the solution. The component then passes out
from the column and are detected with the results being passed through to a computer data
station with the results being shown as peaks on a chromatogram.
There are some factors that are being considered when using HPLC and other forms of
chromatography include the retention time, resolution between peaks and column
efficiency. Retention time or Rt can be defined as the time taken for the components to
flow through the column. Long retention times may lead to higher costs and wasted time
while analysing solutions. Resolution can be explained as how well two elution peaks can
be differentiated on a chromatogram. It can be calculated by using the difference in
retention times between two peaks, divided by the combined widths of the elution peaks.
By altering the mobile phase, the flow rate, and the wavelength these factors can be
optimised to give the best separation. Mobile phase composition alters results due to
changes in its polarity. For normal phase chromatography the stationary phase generally
used is polar for example Silica.
Materials:
Isocratic HPLC system with UV detector, C18 column, Vacuum, Funnel, 0.45 μm filter
paper, 0.45 μm filter syringe,100 μL syringe,60 mL syringe,Volumetric flask
Chemicals:
Caffeine 1000 ppm standard (stock solution), Methanol (HPLC grade), Double distilled
water (filtered with 0.45 μm filter paper), Soft drink sample
Methods:
Standard caffeine samples of 20 ppm, 40 ppm, 60 ppm, 80 ppm and 100 ppm
were prepared by diluting portions of the 1000 ppm solution with distilled water.
2000000.00
Area (µV.s)
1500000.00
1000000.00
500000.00
0.00
0 20 40 60 80 100 120
Based on the objectives of the experiment, this experiment is conducted to identify the
present of Benzoic acid or caffeine in soft drink sample. Not just that, this experiment also
conducted to determine the amount of caffeine in soft drink sample. Samples that have
been using in this experiment are standard caffeine with different concentrations, coke,
and Pepsi. To identify caffeine in stock solutions, High-Performance Liquid
Chromatography or known as HPLC was used. HPLC relies on pumps to pass a
pressurized liquid solvent containing the sample mixture through a column filled with a
solid adsorbent material. Each component in the sample interacts slightly differently with
the adsorbent material, causing different flow rates for the different components and
leading to the separation of the components as they flow out of the column.
After all stock solutions have been scanned with HPLC, we have gained the peak area and
the concentration of the stock solutions of every concentration which are 20ppm, 40ppm,
60ppm, 80ppm, and 100ppm. The retention time that were used is the highest peak of the
graph while for the rest, the nearest retention time was used. In this experiment, the
retention time that were used is around 1.8 minutes. Based on the results, a standard
calibration curve can be conducted with a graph of concentration against the peak area.
2000000
1500000
1000000
500000
0
0 20 40 60 80 100 120
Comparing coke to Pepsi, Pepsi has more amount or concentration of caffeine in it.
Caffeine is a central nervous system (CNS) stimulant of the methylxanthine class. Caffeine
is one of the world's most widely consumed psychoactive drug. Different to many other
psychoactive substances, it is legal and unregulated in nearly all parts of the world. There
are several known mechanisms of action to explain the effects of caffeine to our body and
health. The most important is that the caffeine reversibly blocks the action
of adenosine on its receptors and consequently prevents the onset of sleepiness induced by
adenosine. Caffeine also stimulates certain portions of the autonomic nervous system.
For conclusion, we can conclude that there are presence of caffeine or benzoic acid in soft
drinks as in in this experiment, coke and Pepsi has been used. It is just how much different
the amount of caffeine in the soft drinks. Nevertheless, we also can conclude that High-
Performance Liquid Chromatography can be used to determine the amount of caffeine in
the soft drinks as HPLC has a function of separating components from a mixture.
HPLC has been used widely in industries nowadays. HPLC is used in clinical diagnosis and
health industry, in scientific research, in pharmaceutical industry, and many more. The
effectiveness and uses of HPLC applications in recent days are further enhanced due to
coupling with detectors such as Mass Spectrometer, Nuclear Magnetic Resonance
spectrometer, and many more.
The objectives of this experiment are to identify the present Caffein in soft drink sample
and to determine the amount of Caffein in that sample by using High Performance Liquid
Chromatography. Based on the calibration curve plotted, the concentration of Caffein in
Pepsi is 70 ppm while Coke is 45 ppm. This shows that Pepsi contains more Caffein than
Coke. During this experiment, a few precautions need to care about. For example, we must
always wear safety goggles and gloves as caffein may cause irritation to some people.
Next, all waste must be poured into the waste container following their own category. To
improve the accuracy of the results, we can repeat the experiment five more times and get
the average value.
7.0 TUTORIALS
1. Why does the syringe have to be carefully rinsed before each use?
In general, assume that all syringes are dirty. Each syringe should be rinse first with
the sample you plan to inject two or three times to make sure there is no residue left
from the last sample injected.
2. Retention of caffeine in standards. How could you identify a peak in the soda
was caffeine and not another substance by using retention time?
The pump keeps a precise flow rate so that the positions of the peaks in time can be
used to identify the species in a sample. This is done by comparing the
chromatographs of prepared standards of the species to be determined. The common
peak is an indication of the standard. So, from the graph the highest peak is identified.
8.0 REFERENCES
Figure
1(a): 20
ppm
Figure
2: 40
ppm
Figure 2(a):40 ppm
Figure 3: 60 ppm
Figure 3(a): 60 ppm
Figure 4: 80 ppm
Figure 4(a): 80 ppm
Figure 5: 100 ppm
Figure 5(a): 100 ppm
Figure 6: Coke Sample
Figure 6(a): Coke Sample
Figure 7: Pepsi Sample
Figure 7(a): Pepsi Sample
10.0 JOTTER NOTES
OBJECTIVES:
• To Identify the present of Benzoic acid/ Caffeine in soft drink sample
• To determine amount of caffeine in soft drink sample.
MATERIALS:
• Isocratic HPLC system with UV detector
• C18 column, Vacuum
• Funnel
• 0.45 μm filter paper
• 0.45 μm filter syringe
• 100 μL syringe
• 60 mL syringe
• Volumetric flask
CHEMICALS:
• Caffeine 1000 ppm standard (stock solution)
• Methanol (HPLC grade)
• Double distilled water (filtered with 0.45 μm filter paper)
• Soft drink sample
METHODS:
1) Preparation of caffeine standards
-Prepare standard caffeine samples of 20 ppm, 40 ppm,
60 ppm, 80 ppm and 100 ppm by diluting portions of
the 1000 ppm solution with distilled water.
Concentration
Type of Rentation Area
of Sample
Sample Time (min) (µV. s)
(ppm)
20 1.894 454723.31
40 1.872 869653.17
Standard
60 1.881 1378881.14
Caffeine
80 1.872 1788539.99
100 1.874 2234898.49
Coke 45 1.856 1176173.38
Pepsi 70 1.871 1589421.26
PRE-LABORATORY QUESTION
1. Briefly explain how HPLC is used as a separation technique.
-HPLC used as an instrumental technique of analytical chemistry which mainly used
to identify each component and to quantify each component.
2. What is the purpose of the mobile phase and the stationary phase?
-Mobile phase carries the components of the mixture through the medium being used
and the stationary phase acts as constraint on many of the components in a mixture
that slowing them down to move slower than mobile phase.
ID NUMBER: 55217119075
OBJECTIVES
MATERIALS
Chemicals:
Apparatus:
PRE-LABORATORY QUESTIONS
5. What is the purpose of the mobile phase and the stationary phase?
- Mobile phase carries the components of the mixture through the medium being
used and the stationary phase acts as constraint on many of the components in a
mixture that slowing them down to move slower than mobile phase.
OBJECTIVES
MATERIALS
APPARATUS
PRE-LABORATORY QUESTIONS
It uses a mobile phase (mobile solvent) to carry a sample through a tube containing a stationary phase
which is a liquid coated on solid particles such as silica. The molecules in the sample interact differently
with the stationary phase, the more they interact with it the more they are slowed down. This tends to
separate them out with those interacting least emerging first from the HPLC tube. The mobile phase
speed through the tube is kept uniform by a pump that can sustain a constant (non varying) pressure
level
2. What are the purposes of the mobile phase and the stationary phase?
The mobile phase refers to the liquid or gas, which flows through a chromatography system, moving
the materials to be separated at different rates over the stationary phase while stationary phase refers
to the solid or liquid phase of a chromatography system on which the materials are to be separated or
selectively adsorbed
The purpose of the caffeine standards is to determine if caffeine is present in the beverage samples
JOTTER NOTES
OBJECTIVES
Materials:
Isocratic HPLC system with UV detector, C18 column, Vacuum, Funnel, 0.45 μm filter paper
Chemicals:
Caffeine 1000 ppm standard (stock solution), Methanol (HPLC grade), Double distilled water
(filtered with 0.45 μm filter paper), Soft drink sample
PROCEDURE
After preparing the serial dilution and sample, your instructor will
brief on standard operating procedure of HPLC.
DATA AND RESULTS
2. What are the purposes of the mobile phase and the stationary phase?
The stationary phase remains fixed in place while the mobile phase carries the components
of the mixture through the medium being used. The stationary phase acts as a constraint on
many of the components in a mixture, slowing them down to move slower than the mobile
phase.
Materials:
Isocratic HPLC system with UV detector, C18 column, Vacuum, Funnel, 0.45 μm filter paper
Chemicals:
Caffeine 1000 ppm standard (stock solution), Methanol (HPLC grade), Double distilled water
(filtered with 0.45 μm filter paper), soft drink sample
Procedure
Prepare standard caffeine samples of 20 ppm, 40 ppm, 60 ppm, 80 ppm and 100
ppm by diluting portions of the 1000 ppm solution
Degas the sample by placing it in a vacuum flask and connecting the flask to a
vacuum pump or water aspirator. Leave it under vacuum until no more bubbles
appear in the soda sample. (If no vacuum is available, allow the soda to stand
open overnight)
After preparing the serial dilution and sample, your instructor will brief on
standard operating procedure of HPLC.
OBJECTIVES:
1. To Identify the present of Benzoic acid/ Caffeine in soft drink sample
2. To determine amount of caffeine in soft drink sample
Materials:
Isocratic HPLC system with UV detector, C18 column, Vacuum, Funnel, 0.45 μm filter paper
0.45 μm filter syringe,100 μL syringe,60 mL syringe,Volumetric flask
Chemicals:
Caffeine 1000 ppm standard (stock solution), Methanol (HPLC grade), Double distilled water
(filtered with 0.45 μm filter paper), Soft drink sample
PROCEDURE:
1.Preparation of caffeine standards
2. What are the purposes of the mobile phase and the stationary phase?
In very small amounts, the sample mixture to be separated and tested is sent into a
stream of mobile phase percolating via a column. There are different types of
columns available with sorbents of varying particle sizes and surfaces. The
mixture moves through the column at varying velocities and interacts with the
sorbent, also known as the stationary phase. The composition of the mobile phase
is chosen based on the intensity of interactions between several sample
components and the stationary phase. The composition of the mobile phase is
either maintained as a constant or as varied during the chromatographic analysis.
STUDENT ID : 55217119075
1. Rate your team members on the relative contribution that were made in preparing and submitting your lab report. Please refer to
the rubric provided to assess your peers
2. In rating your peers, use to five point scale.
3. Every single group member is to fill in this form and be honest, do not favour anyone. Form is to be submitting along with
the respective submission
Names: HARITH Names: MAISARAH Names: NUR ANNISA Names: SHAIDATUL Names: MUHD
SAIFULLAH BIN BINTI OTHMAN FARAHIN BINTI NAJWA BINTI SYAHMI BIN YUSOP
JOFFERRY MOHD FAUZI MOHAMAD
ID: 55217119091 ID: 55217119116 ID: 55216119077 ID: 55217119083 ID: 55213117108
(1) Participated in group
5 5 5 5 5
discussions
(2) Contribution of useful ideas 5 5 5 5 5
(3) Focus on the task 5 5 5 5 5
(4) Quality of Work 5 5 5 5 5
(5) Working with others 5 5 5 5 5
TOTAL MARK (25%) 25 25 25 25 25
TOTAL MARK (10%)
UNIVERSITI KUALA LUMPUR
MALAYSIAN INSTITUTE OF CHEMICAL
BIOENGINEERING TECHNOLOGY
STUDENT ID : 55217119083
ID: 55217119091 ID: 55213117108 ID: 55217119075 ID: 55217119116 ID: 55216119077
(1) Participated in group discussions 5 5 5 5 5
(2) Contribution of useful ideas 5 5 5 5 5
(3) Focus on the task 5 5 5 5 5
(4) Quality of Work 5 5 5 5 5
(5) Working with others 5 5 5 5 5
TOTAL MARK (25%) 25 25 25 25 25
TOTAL MARK (10%) 10 10 10 10 10
UNIVERSITI KUALA LUMPUR
MALAYSIAN INSTITUTE OF CHEMICAL
BIOENGINEERING TECHNOLOGY
STUDENT ID : 55217119116
ID: 55217119091 ID: 55213117108 ID: 55217119075 ID: 55217119083 ID: 55216119077
(1) Participated in group discussions 5 5 5 5 5
(2) Contribution of useful ideas 5 5 5 5 5
(3) Focus on the task 5 5 5 5 5
(4) Quality of Work 5 5 5 5 5
(5) Working with others 5 5 5 5 5
TOTAL MARK (25%) 25 25 25 25 25
TOTAL MARK (10%) 10 10 10 10 10
UNIVERSITI KUALA LUMPUR
MALAYSIAN INSTITUTE OF CHEMICAL
BIOENGINEERING TECHNOLOGY
STUDENT ID : 55216119077
ID: 55217119091 ID: 55213117108 ID: 55217119075 ID: 55217119083 ID: 55217119116
(1) Participated in group discussions 5 5 5 5 5
(2) Contribution of useful ideas 5 5 5 5 5
(3) Focus on the task 5 5 5 5 5
(4) Quality of Work 5 5 5 5 5
(5) Working with others 5 5 5 5 5
TOTAL MARK (25%) 25 25 25 25 25
TOTAL MARK (10%) 10 10 10 10 10
UNIVERSITI KUALA LUMPUR
MALAYSIAN INSTITUTE OF CHEMICAL
BIOENGINEERING TECHNOLOGY
STUDENT ID : 55217119091
ID: 55216119077 ID: 55213117108 ID: 55217119075 ID: 55217119083 ID: 55217119116
(1) Participated in group discussions 5 5 5 5 5
(2) Contribution of useful ideas 5 5 5 5 5
(3) Focus on the task 5 5 5 5 5
(4) Quality of Work 5 5 5 5 5
(5) Working with others 5 5 5 5 5
TOTAL MARK (25%) 25 25 25 25 25
TOTAL MARK (10%) 10 10 10 10 10
UNIVERSITI KUALA LUMPUR
MALAYSIAN INSTITUTE OF CHEMICAL
BIOENGINEERING TECHNOLOGY
STUDENT ID : 55213117108
4. Rate your team members on the relative contribution that were made in preparing and submitting your lab report. Please refer to
the rubric provided to assess your peers
5. In rating your peers, use to five point scale.
6. Every single group member is to fill in this form and be honest, do not favour anyone. Form is to be submitting along with
the respective submission
Names: HARITH Names: MAISARAH Names: NUR ANNISA Names: SHAIDATUL Names:
SAIFULLAH BIN BINTI OTHMAN FARAHIN BINTI NAJWA BINTI NORYUSHAINA
JOFFERRY MOHD FAUZI MOHAMAD ADLINA BT
YUSMAINI
ID: 55217119091 ID: 55217119116 ID: 55216119077 ID: 55217119083 ID: 55217119075
(1) Participated in group
5 5 5 5 5
discussions
(2) Contribution of useful ideas 5 5 5 5 5
(3) Focus on the task 5 5 5 5 5
(4) Quality of Work 5 5 5 5 5
(5) Working with others 5 5 5 5 5
TOTAL MARK (25%) 25 25 25 25 25
TOTAL MARK (10%) 10 10 10 10 10