Download as pdf or txt
Download as pdf or txt
You are on page 1of 27

British Standard

Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

A single copy of this British Standard is licensed to


lbocvzr lbocvzr
26 June 2003

This is an uncontrolled copy. Ensure use of the most


current version of this document by searching British
Standards Online at bsonline.techindex.co.uk
BRITISH STANDARD BS 5728-3:
1997

Measurement of flow of
cold potable water in
closed conduits —
Part 3: Methods for determining
principal characteristics of single
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

mechanical water meters (including


test equipment)

ICS 17.120.10
BS 5728-3:1997

Committees responsible for this


British Standard

The preparation of this British Standard was entrusted to Technical


Committee CPL/30, Measurement of fluid flow in closed conduits, upon which
the following bodies were represented:

British Compressed Air Society


British Gas plc
National Engineering Laboratory
DTI — Gas and Oil Measuring Branch
Electricity Association
Energy Industries Council
Engineering Equipment and Materials Users Association
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

GAMBICA (BEAMA) Ltd.


Institute of Measurement and Control
Institute of Petroleum
Institute of Trading Standards Administration
Institution of Gas Engineers
Institution of Mechanical Engineers
Petrol Pump Manufacturers Association
Society of British Gas Industries
Water Services Association of England and Wales

The following bodies were also represented in the drafting of the standard
through subcommittees and panels:

Society of British Water Industries


United Kingdom Automatic Meter Reading Association
Water Research Centre

This British Standard, having


been prepared under the
direction of the Electrotechnical
Sector Board, was published
under the authority
of the Standards
Board and comes
into effect on
15 October 1997

© BSI 10-1999 Amendments issued since publication


First published May 1984
Second edition October 1997
Amd. No. Date Comments

The following BSI references


relate to the work on this
standard:
Committee reference CPL/30
Draft for comment 95/210206 DC

ISBN 0 580 28320 8


BS 5728-3:1997

Contents

Page
Committees responsible Inside front cover
Foreword ii
1 Scope 1
2 Normative references 1
3 Definitions 1
4 General considerations affecting all tests 1
5 Measurement error tests 2
6 Pressure tests 5
7 Pressure loss tests 8
8 Accelerated wear tests 16
9 Test report 17
Annex A (informative) Summary of requirements and
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

permitted tolerances in the measurement of physical quantities


associated with water meter test methods and equipment 20
Figure 1 — Pressure test manifold for concentric meters 6
Figure 2 — Example of plug for pressure test of concentric meters 7
Figure 3 — Layout of measuring section 8
Figure 4 — Example of drilled hole type of pressure tapping with
ring chamber, suitable for small/medium diameter test sections 9
Figure 5 — Example of drilled hole type of pressure
tapping with loop mean static pressure interconnections,
suitable for medium/large diameter test sections 10
Figure 6 — Example of slit type of pressure tapping with ring
chamber, suitable for small/medium diameter test sections 11
Figure 7 — Manifold caps 12
Figure 8 — Pressure loss measurement test procedure (in-line meters) 14
Figure 9 — Pressure loss measurement test procedure
(concentric type meters) 15
Table 1 — Number of meters to be tested 1
Table 2 — Test programme for an accelerated wear test 17
Table 3 — Test procedure and test results. Information required 19
Table 4 — Test conclusions. Information required 19
Table A.1 — Summary of requirements and permitted tolerances 20
List of references Inside back cover

© BSI 10-1999 i
BS 5728-3:1997

Foreword

This Part of BS 5728 has been prepared by Technical Committee CPL/30. It


supersedes BS 5728-3:1984 which is withdrawn.
This revision of BS 5728-3 was prepared in order to bring the techniques and
equipment described up to date. Also, the scope has been extended to incorporate
the testing of concentric meters, in addition to in-line meters.
The other Parts of BS 5728 on the measurement of flow of cold potable water, are
as follows:
— Part 1: Specification for single meters;
— Part 2: Specification for installation requirements for single meters;
— Part 4: Specification for combination meters;
— Part 5: Specification for installation requirements for combination meters;
— Part 6: Test methods for combination meters;
— Part 7: Specification for single mechanical type meters.
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

A British Standard does not purport to include all the necessary provisions of a
contract. Users of British Standards are responsible for their correct application.

Compliance with a British Standard does not of itself confer immunity


from legal obligations.

Summary of pages
This document comprises a front cover, and inside front cover, pages i and ii,
pages 1 to 20, an inside back cover and a back cover.
This standard has been updated (see copyright date) and may have had
amendments incorporated. This will be indicated in the amendment table on the
inside front cover.

ii © BSI 10-1999
BS 5728-3:1997

1 Scope 4.1.1 Written test programme


This Part of BS 5728 gives methods of test for single Before starting testing, a written test programme
mechanical water meters conforming to BS 5728-7. shall be compiled, and shall include a description of
the tests for the determination of measurement
2 Normative references error, pressure loss and wear resistance.
This Part of BS 5728 incorporates, by dated or 4.1.2 Pattern approval
undated reference, provisions from other NOTE Pattern approval consists of verifying that the
publications. These normative references are made characteristics of one model of meter conform to the standards
and regulations in force. This consequently requires that samples
at the appropriate places in the text and the cited of the model fulfil the requirements of the test programme.
publications are listed on the inside back cover. For
4.1.2.1 Number of meters to be tested
dated references, only the edition cited applies; any
subsequent amendments to or revisions of the cited The pattern approval tests shall be carried out on a
publication apply to this Part of BS 5728 only when minimum number of water meters of each model as
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

incorporated in the reference by amendment or given in Table 1 as a function of the meter


revision. For undated references, the latest edition designation of the model submitted.
of the cited publication applies, together with any Table 1 — Number of meters to be tested
amendments.
Meter designation Number of meters

3 Definitions N k 100 3
100 < N k 1 000 2
For the purposes of this Part of BS 5728, the
definitions given in BS 5728-7:1997 apply, together 1 000 < N 1
with the following. NOTE The number of water meters in Table 1 may be regarded
3.1 as the minimum to be tested; the authority responsible for
pattern approval may request additional water meters.
measurement error
4.1.2.2 Pattern approval tests
the difference between the indicated value of the
volume and the conventional true value of the Before tests commence, each water meter and
volume of water passed through the meter during a accompanying certificates shall be inspected to
test check that the meter conforms to the regulations in
NOTE 1 Measurement errors are conventionally expressed as
force, and also to the specifications (e.g. model,
relative errors, calculated as a percentage, and equal to: identification, size and material) and drawings
Vi – Vc submitted when approval was requested.
- × 100
----------------
Vc The tests shall include the following, carried out in
where the order indicated:
Vc is the value accepted as true of the volume passed; a) pressure tests (see 6.4);
Vi is the volume indicated by the water meter at the time b) determination of the error curves as a function
of measurement of the same volume, both expressed in
the same units. of flow-rate (see 5.9);
NOTE 2 The values for maximum permissible errors (MPE) are c) pressure loss tests (see 7.5);
specified in 5.1 of BS 5728-7:1997. d) accelerated wear tests (see 8.3).
3.2 4.1.3 Initial verification
test flow-rate
NOTE Initial verification consists in verifying that all the
the mean rate of flow of water through the meter water meters submitted conform to the approved model and to
under test during the period of a test the standards and regulations in force. The initial verification
programme may authorize total or partial statistical control
NOTE The test flow-rate is usually measured and indicated by procedures.
a calibrated reference device (see 5.2 and 5.4).
4.1.3.1 Check for conformity to the approved model
4 General considerations affecting all The meters shall be inspected to check that they
tests have the same technical characteristics (dimensions
4.1 Test programme of the constituent parts, materials, surface
finish, etc.) as the approved model.
NOTE Water meters are subject to a variety of different test
programmes. For the purposes of pattern approval and initial NOTE This check is normally carried out by sampling.
verification, water meters are covered by the requirements of
EC Directive 75/33/EEC of 17 November 1974.

© BSI 10-1999 1
BS 5728-3:1997

4.1.3.2 Initial verification tests NOTE Within this temperature range, temperature correction
of the results is not required.
Initial verification tests shall include at least the
following: 4.3.4 Location
a) in principle, on all meters: The test rig shall be isolated from any other activity
(for example, manufacture, repairs) and from
— pressure tests (see 6.5); environmental disturbances (for example,
— determination of the measurement fluctuations in ambient temperature, vibration).
errors (see 5.10);
b) normally, by sampling, as required by the 5 Measurement error tests
inspector: NOTE The method described in this Part of BS 5728 to
— a check that the meter conforms to the determine measurement errors is the so-called “collection”
method, in which the quantity of water passed through the water
approved pattern (see 4.1.2.2); meter is collected in one or more collecting vessels, and the
— pressure loss measurement (see 7.6). quantity determined volumetrically or by weighing. Other
methods may be used, provided the accuracy levels stated in this
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

4.2 Water quality Part of BS 5728 are attained.


Water meter tests shall be made with water. The 5.1 Outline of test method
water shall be that of the public potable water The checking of the measurement error consists of
supply or shall meet the same requirements. If comparing the indications given by the meter under
water is being recycled, measures shall be taken to test against a calibrated reference device.
prevent residual water in the meter from becoming
harmful to human beings. 5.2 Description of the test rig
The inclusion of air bubbles shall be avoided by the The test rig shall include:
provision of suitably placed air bleeds in the test rigs a) a water supply (mains, non-pressurized tank,
(see 5.3.1 and 5.3.3) and particles of size greater pressurized tank, pump, etc.);
than 100 4m shall be eliminated from the water by b) pipework (see 5.3);
appropriate filtration.
c) a calibrated reference device (calibrated tank,
4.3 Test rigs and their location reference meter, etc.) (see 5.4);
4.3.1 Freedom from spurious influences d) means for measuring the duration of the test.
Test rigs shall be so designed, constructed and used NOTE Devices for automating the testing of water meters are
permissible.
that the performance of the rig itself does not
introduce significant errors into the measurement. 5.3 Pipework
To this end, high standards of rig maintenance and 5.3.1 Description
adequate supports and fittings, preventing
Pipework shall include:
vibration of the meter, the test rig and its
accessories, shall be observed. a) a test section in which the meter(s) is (are)
placed;
It shall be possible to perform tests rapidly and
easily. b) means to establish the desired flow-rate;
4.3.2 Group testing of meters c) one or two isolating devices;
For the purposes of some tests, meters may be tested d) means for determining the flow-rate; and, if
individually or in groups, depending on the necessary:
purposes of the tests. When meters are tested in e) one or more air bleeds;
groups, the individual characteristics of each f) a non-return device;
individual meter shall be precisely determined.
g) an air separator;
Interaction between meters, and between test rigs,
shall be eliminated by the use of suitable h) a filter.
installation practices. During the test, flow leakage, flow input and flow
When meters are tested in series, the pressure at drainage shall not occur either between the meter(s)
the outlet of each shall be sufficient to prevent and the reference device or from the reference
cavitation. device.
4.3.3 Temperature of the water during the tests The pipework shall be such that in the upper part of
the meter a positive pressure exists of at
The temperature of the water in the test section of least 0.05 bar (5 kPa) even at zero flow-rate.
the test rig shall be between 0 °C and 30 °C during
a test. In no part of the test rig shall the
temperature fall below 0 °C during a test.

2 © BSI 10-1999
BS 5728-3:1997

5.3.2 Test section In order to counteract these effects, the meter shall
The test section shall include, in addition to the be installed for test in a straight length of pipe. The
meter(s): connecting pipework shall have the same internal
diameter as the bore of the meter inlet.
a) one or more pressure tappings for the
NOTE 3 It may also be necessary to put a flow straightener
measurement of pressure, of which one pressure upstream of the straight length.
tapping shall be situated upstream of, and close
5.3.4.3 Common causes of flow disturbance
to, the (first) meter;
b) if necessary, means for measuring the Flow can be subject to two types of disturbance:
temperature of the water at the inlet of the (first) velocity profile distortion and swirl, both of which
meter. affect the accuracy of the water meter.
The different devices placed in the measuring Velocity profile distortion is typically caused by an
section shall not cause cavitation or flow obstruction partially blocking the pipe, for instance
disturbances capable of altering the performance of the presence of a partly closed valve or a misaligned
flange joint. This can easily be eliminated by
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

the meters or causing measurement errors.


removal of the obstruction or by opening the valve.
5.3.3 Precautions to be taken during tests
Swirl is caused mainly by two or more bends in
Make sure that there is no leakage in the pipework different planes. This effect can be controlled either
or fittings between the meter under test and the by ensuring an adequate length of straight pipe
calibrated reference device. upstream of the water meter, or by installing a
Check that pipes (for example the swan neck in the straightening device, or by a combination of the two.
outlet pipe) are filled to the same extent at the 5.3.5 Test commencement and termination
beginning and at the end of the test. errors
Bleed any air from the interconnecting pipework 5.3.5.1 Principle
and the meter(s) (see 4.2).
The magnitude of errors occurring at the beginning
Take precautions to avoid the effects of vibration
and end of a test, due to changing flow-rates, meter
and shock (see 4.3.4).
inertia, and timing errors are established and then
5.3.4 Installation of meter on test into the test reduced to a practical minimum.
rig
5.3.5.2 Effects of meter inertia
NOTE Volumetric water meters (that is, involving measuring
chambers with mobile walls), such as oscillating piston meters, Adequate precautions shall be taken to reduce the
are considered insensitive to upstream installation conditions; uncertainties resulting from the operation of the
hence no special recommendations are given for these types of test rig components during the test.
meter.
NOTE Whilst flow is beginning and during the period of
5.3.4.1 Summary of test installation requirements running at the specified constant flow-rate, the measurement
error of the meter varies as a function of the changes in flow-rate
The meter on test shall be installed into the test rig (measurement error curve). Whilst flow is being stopped, the
such that: combination of inertia of the moving parts of the meter and the
a) the hydrodynamic flow characteristics rotational movements of the water inside the meter may cause an
appreciable error to be introduced in certain types of meter and
immediately upstream and downstream of the for certain test flow-rates.
meter cause no discernible difference to the It has not been possible, in this case, to determine a
performance of the meter when compared with its simple empirical rule which lays down conditions so
performance under undisturbed conditions; that this error may be discounted as negligible.
b) the overall error of the method employed does In case of doubt, it is advisable:
not exceed the stipulated value (see 5.4.1).
a) to increase the volume and duration of the test;
5.3.4.2 Need for straight lengths of pipe or a flow
b) to compare the results with those obtained by
straightener
one or more other methods, and in particular the
NOTE 1 The accuracy of the water meter can be affected by method described in 5.5.2, which eliminates the
upstream disturbance caused for example by the presence of
bends, tees, valves or pumps. causes of uncertainty given in the note.
NOTE 2 Certain velocity type water meters are sensitive to flow Errors due to meter inertia may be assessed by
disturbance which can cause significant errors, but the way
installation conditions affect their accuracy has not yet been
procedure b), and may be minimized by
clearly determined. It is simply recommended to avoid, as far as procedure a).
possible, the presence of bends, pumps, taper pieces and changes
in diameter of pipework immediately upstream, and to position
the meter to afford the maximum possible straight length of pipe
upstream and downstream.

© BSI 10-1999 3
BS 5728-3:1997

5.4 Calibrated reference device During measurement error testing, the volume to be
5.4.1 Overall error of the method employed passed shall be chosen to be sufficiently large, so
that the maximum error in the reading of the
For pattern approval and initial verification, the volume indicated by the meter does not
total error in the method used for the determination exceed 0.5 % (see 4.5.4 of BS 5728-7:1997).
of the volume of water passed through the water
In the reading of a meter having circular scales with
meter shall not exceed 1/10 of the relevant
pointers, in order to avoid errors due to eccentricity
maximum permissible error.
or uneven scale graduations, the volume to be
5.4.2 Minimum volume of the calibrated vessel passed shall be chosen such that whole revolutions
Where a calibrated volumetric vessel is used, the of the pointer or scale are used.
minimum volume permitted depends on 5.7 Major factors affecting measurement error
requirements determined by the test start and end checks
effects and the design of the indicating
NOTE Variations in the pressure, flow-rate, and temperature
device (verification scale division) (see BS 5728-7) of in the test rig, and uncertainties in the precision of measurement
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

the meter under test. of these physical quantities, are the principal factors affecting the
measurement error test results.
5.5 Test methods
5.7.1 Pressure
5.5.1 Standing start/stop test method
The pressure shall be maintained at a constant
The flow is established by opening a valve, value throughout the test at the chosen flow-rate.
preferably situated downstream of the meter, and it
is stopped by the closure of this valve. The meter is For testing water meters of N10 or smaller, at test
read whilst completely stationary. flow-rates k 1 m3/h, constant pressure at the inlet
of the meter (or at the inlet of the first meter of a
Time is measured between the beginning of the series being tested) can be achieved if the test rig is
movement of the valve at opening and at the supplied through a pipe from a constant head tank.
beginning of closure.
Any other methods of supply which can be shown
5.5.2 Flow diversion method not to cause pressure pulsations exceeding those of
The measurement is carried out when the flow a constant head tank may be used.
conditions have stabilized. For all other tests the pressure upstream of the
A flow diverter switches the flow into a calibrated meter shall not vary by more than 10 %.
vessel at the beginning of the measurement and The maximum uncertainty in the measurement of
diverts it away at the end. The meter is read in pressure shall be 5 % of the measured value.
motion. Pressure at the inlet to the meter shall not exceed
The reading of the meter is synchronized with the the nominal pressure for the meter.
movement of the flow diverter. 5.7.2 Flow-rate
The volume collected in the vessel is the volume The flow-rate shall be maintained constant
passed. throughout the test at the chosen value.
NOTE The volume collected may be determined by measuring
its mass and converting this to volume from a knowledge of the The relative variation in the flow-rate during each
water density. test (not including starting and stopping) shall not
The uncertainty introduced into the volumes exceed:
measured shall be considered to be negligible if the ± 2.5 % from qmin to qt (not inclusive);
time of motion of the flow diverter in each direction ± 5.0 % from qt (inclusive) to qs.
is identical within 5 % and if each time is not less
than 1/50 of the total duration of the test. The flow-rate value used in plotting the error curve
shall be the volume passed during the test divided
5.6 Precautions to be taken when reading a by the duration of the test.
meter
The relative pressure variation (in flow to free air)
NOTE The maximum reading error of a scale does not exceed or the relative variation in pressure loss (in closed
half a scale division per observation. Thus in the measurement of
a volume of flow delivered by a water meter (consisting of two circuits) shall not exceed:
observations of the water meter) the total reading error can reach ± 5 % from qmin to qt (not inclusive);
one scale division or, in the case of discontinuous movement of
the control element, 2 digits. ± 10 % from qt (inclusive) to qs.
5.7.3 Temperature
During a test the temperature of the water shall not
change by more than 5 °C.

4 © BSI 10-1999
BS 5728-3:1997

The maximum uncertainty in the measurement of If the error determined lies outside the tolerance
temperature shall not exceed 1 °C. band, the test may be continued. It shall then be
5.8 Interpretation of results repeated twice. If the arithmetical mean of the three
test results is within or at the limits of the tolerance
5.8.1 Single test band, the test results shall be declared to be
Where the test programme specifies a single test, satisfactory.
the meter shall be considered to have passed this 5.10 Initial verification tests. Determination of
test if the measured error does not exceed the the measurement errors
maximum permissible error at the chosen flow-rate.
Test methods and equipment shall be as described
5.8.2 Repeated tests in 5.1 to 5.7.
Where the test programme specifies that the test The measurement errors of each water meter shall
may be repeated, the programme shall specify the be determined at not less than three flow-rates.
rules to be applied for combining the errors
The flow-rates shall be chosen with regard to the
obtained.
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

characteristics of the measurement error curve of


The meter shall be considered to have passed this the model of the meter, in such a manner that it can
test if the error resulting from this combination does be established with certainty whether or not the
not exceed the maximum permissible error at the maximum permissible errors are exceeded.
chosen flow-rate.
The requirements concerning the maximum
5.9 Pattern approval tests. Determination of permissible errors, the definition of the error and
the error curve as a function of flow-rate the test flow-rate shall be the same as in 5.9.
Test methods and equipment used to determine the Initial verification tests shall not be repeated. The
error of the meter at a given flow-rate shall be as test result shall only be declared to be satisfactory if,
described in 5.1 to 5.7. for each test flow-rate, the error determined does
It is recommended that the characteristic curve for not exceed the maximum permissible error.
each water meter be plotted in terms of the error
against flow-rate, so that the general performance of 6 Pressure tests
the water meter over the flow-rate range can be 6.1 Principle
evaluated.
The water meter is subjected to a specified hydraulic
Where it is thought that pressure may have a test pressure for a specified time, after which it is
significant effect on the error curve, the effect of examined for leakage or damage.
pressure up to the MAP shall be investigated by
repeated measurement error tests at different 6.2 Precautions to be taken before the tests
pressures. The test rig and the meter shall be cleared of air by
The measurement errors of the water meters shall bleeding.
be determined at a minimum of seven flow-rates, The test rig shall be leakproof.
five of which shall be as follows: Pressurizing shall be carried out gradually without
a) between qmin and 1.1qmin; pressure surges.
b) between qt and 1.1qt; 6.3 Pressure tests of concentric meters
c) between 0.45qp and 0.5qp; The principle given in 6.1 and requirements given
d) between 0.9qp and qp; and in 6.2 apply to the pressure testing of all water
e) between 0.9qs and qs. meters, but concentric meters require additional
special provisions so that the meter itself and the
The test flow-rate shall be the mean flow-rate seals between the meter and manifold may be
calculated from the indications of the calibrated pressure tested. In particular, the inner seal
reference device. situated at the meter outlet, shall be checked to
For pattern approval tests, the test result shall be ensure that undisclosed internal leaks between the
declared to be satisfactory if the error determined at inlet and outlet passages of the meter do not occur.
each flow-rate is within or at the limits of the A special pressure test manifold adaptor as shown
tolerance band. in Figure 1 shall be provided.

© BSI 10-1999 5
BS 5728-3:1997

Before being submitted for pattern approval or 6.4 Pattern approval tests. Pressure tests
initial verification tests, the meter shall be sealed at The requirement for these tests is that each meter
a point above the location of the inner seal by a withstand, without leakage or seepage through the
means suited to the meter design. When the walls, and without damage, a pressure equal to:
concentric meter is fitted to the pressure test
manifold and pressurized, it shall be possible to a) 16 bar, or 1.6 times the nominal pressure if it
see the source of any leak flowing from the pressure exceeds 10 bar, applied for a period of 15 min;
test manifold outlet and distinguish between it and and
that issuing from an incorrectly fitted sealing b) 20 bar, or twice the nominal pressure, applied
device. for a period of 1 min.
Figure 2 shows an example of a design of plug suited
to many meter designs, but any other means of
sealing may be adopted.
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

A B C
19.65
N1 to N2.5 8
15.13 19.60 1.40
15.03 31.95 1.30
N2.5 to N3.5 8
31.90
Dimensions in millimetres.
Figure 1 — Pressure test manifold for concentric meters

6 © BSI 10-1999
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

© BSI 10-1999
Figure 2 — Example of plug for pressure test of concentric meters

7
BS 5728-3:1997
BS 5728-3:1997

6.5 Initial verification tests. Pressure tests 7.2 Pressure loss test equipment
Each meter shall be subjected to a pressure equal NOTE The equipment needed to carry out pressure loss tests
to 16 bar or 1.6 times the nominal consists of a measuring section of pipework containing the water
meter under test and a system for producing the specified
pressure (see 3.21 and 4.12 of BS 5728-7:1997). The constant flow-rate through the meter. The same method of
meter shall then be inspected for leakage, seepage achieving constant flow-rate as that employed for the
through the walls and damage. measurement error tests described in clause 5 may be used for
the pressure loss tests.

7 Pressure loss tests 7.2.1 Measuring section


NOTE The pressure loss of a water meter as defined in 7.2.1.1 General
BS 5728-7 is obtained by the method given in 7.1 to 7.3.
The upstream and downstream pipe lengths, with
This method of pressure loss testing is a reference method. Other
methods may be used on condition that the values of the pressure their end connections and pressure tappings, plus
loss obtained are equal to those obtained by the reference the water meter on test, or in the case of concentric
method. meters, the water meter plus the pressure loss test
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

7.1 Principle manifold, shall constitute the measuring section.


The pressure loss of the water meter is determined 7.2.1.2 Internal diameter of measuring section
from measurements of the static differential (in-line meters only)
pressure across the water meter at a specified The upstream and downstream pipe lengths in
flow-rate. contact with the water meter shall have the same
Pressure tappings situated in the walls of the pipe internal diameter as the meter connection so as to
fitted upstream and downstream of the water meter avoid hydraulic discontinuities. The pipe internal
are used for the measurement of the static diameters shall be as specified by the meter
differential pressure. manufacturer.
The pressure loss tests take into account any NOTE A difference in the diameter of the connecting pipework
pressure recovery downstream of the meter by and that of the meter may result in a measurement uncertainty
incompatible with the precision desired.
suitably locating the downstream pressure tapping
(see 7.2.1.3). The results of the tests take account of 7.2.1.3 Measuring section straight lengths
the pressure recovery, and are also corrected for the Upstream and downstream of the meter, and
lengths of pipe between the pressure upstream and downstream of the pressure tappings,
tappings (see 7.3.2). straight lengths of pipe shall be provided in
accordance with the provisions of Figure 3.

NOTE P1 and P2 are the planes of the pressure tappings and C is the water meter, or for concentric meters, C is the water meter
plus manifold.
L U 15D, L1 U 10D, L2 U 5D D = internal diameter of the pipework
Figure 3 — Layout of the measuring section

8 © BSI 10-1999
BS 5728-3:1997

7.2.1.4 Design of measuring section pressure Slits shall be perpendicular to the pipe axis and
tappings shall have dimensions as follows (see Figure 6).
Pressure tappings of similar design and dimensions a) Width i shall be equal to or less than 0.08D or
shall be fitted to the inlet and outlet pipes of the less than 4 mm whichever is the smaller.
measuring section. b) Depth k shall be greater than 2i.
NOTE 1 Pressure tappings may consist of holes drilled through
the wall of the pipe or be in the form of an annular slit in the pipe
7.2.1.6 Measurement of static differential pressure
wall, in either case perpendicular to the pipe axis. There should Each group of pressure tappings in the same plane
be at least four such pressure tapping holes, equally spaced in
one plane around the pipe circumference. Examples of both types
shall be connected by a leak-free tube to one limb of
of pressure tapping as specified in 7.2.1.5 are shown in Figure 4, a differential pressure measuring device, for
Figure 5 and Figure 6. example, a manometer. Provision shall be made for
NOTE 2 Four or more pressure tapping holes may be bleeding air from the installation.
interconnected by means of tee-shaped connectors which connect
up the pressure tappings, forming an annulus to give a true mean 7.2.2 Design of pipe lengths and test
static pressure at the pipe cross-section (see Figure 5). manifold (concentric meters only)
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

Alternatively, other means, such as a ring or balance chamber,


may be used. In order to reduce the head loss of the measuring
7.2.1.5 Pressure tappings, hole and slit details section to a minimum, a commercially available,
in-line, water meter manifold of a swept, low loss
Holes drilled through the wall of the construction, shall be used for the test manifold. The
pipe (see Figure 4 and Figure 5) shall be non-return valve in the test manifold shall be
perpendicular to the pipe axis, and the diameter of removed before commencement of the tests. Details
the holes shall not exceed 0.08D or shall be less of the manifold caps to be used in conjunction with
than 4 mm, whichever is the smaller. The diameter the two sizes of test manifolds are given in
of the holes shall remain constant for a distance of Figure 7 a) and Figure 7 b).
not less than two diameters before breaking into the
NOTE The manifold, with manifold cap fitted [see Figure 7 a)
pipe bore. The holes drilled through the pipe wall or Figure 7 b] as appropriate) and with non-return valve
shall be free from burrs at the edges where they removed, should have a head loss (corrected for pipe losses) not
break through into the inlet and outlet pipe bores. exceeding 3 velocity heads.
Edges shall be sharp, that is, neither rounded nor For concentric meter sizes N = 1 to 2.5 (see Table 3
chamfered. of BS 5728-7:1997), i.e. meters having G1 1/2
threaded end connections, the measuring section
pipes shall be 28 mm external diameter, and for
N = 2.5 and 3.5, i.e. meters having G2 threaded end
connections, they shall be 35 mm external diameter.

Figure 4 — Example of drilled hole type of pressure tapping with ring chamber,
suitable for small/medium diameter test sections

© BSI 10-1999 9
BS 5728-3:1997
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

Figure 5 — Example of drilled hole type of pressure tapping with loop mean static
pressure interconnections, suitable for medium/large diameter test sections

10 © BSI 10-1999
BS 5728-3:1997
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

Figure 6 — Example of slit type of pressure tapping with ring chamber,


suitable for small/medium diameter test sections

© BSI 10-1999 11
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

12
BS 5728-3:1997

Figure 7 — Manifold caps

© BSI 10-1999
BS 5728-3:1997

7.3 Test procedure 7.3.4 Determination of the pressure loss


7.3.1 Outline of the test method attributable to the pipe lengths and test
manifold (measurement 1) (concentric meters
The method consists in measuring the static only)
differential pressure (%P2) between the pressure
tappings of the measuring section with the meter The pressure loss of the upstream and downstream
present, and deducting from it the pressure loss pipe lengths and that of the test manifold (%P1) shall
(%P1) of the upstream and downstream pipe lengths be determined prior to the tests proper, and checked
measured at the same flow-rate in the absence of the periodically. This shall be done by firstly fitting the
meter (see Figure 8 or Figure 9). manifold cap detailed in Figure 7 a) or Figure 7 b)
and, in the absence of the meter, measuring the
7.3.2 Determination of the pressure loss pressure loss of the measuring section for each
attributable to the pipe lengths flow-rate. (See Figure 9.)
(measurement 1) (in-line meters only)
7.3.5 Measurement and calculation of the
The pressure loss of the upstream and downstream actual pressure loss of the water
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

pipe lengths (%P1) shall be determined prior to the meter (measurement 2) (concentric meters
tests proper, and checked periodically. This shall be only)
done by joining the upstream and downstream pipe
faces together in the absence of the meter (carefully At the same test flow-rates used to determine the
avoiding joint protrusion into the pipe bore or pipe and test manifold pressure losses, in the same
misalignment of the two faces), and measuring the installation, with the same pressure tappings and
pipe pressure loss of the measuring section for each the same manometer, but with the manifold cap
test flow-rate. removed and the water meter in position, the
differential pressure (%P2) across the metering
The absence of the water meter will shorten the section shall be measured. (See Figure 9.)
measuring section. If telescopic sections are not
fitted on the test rig, the gap may be filled by The actual pressure loss, %P, of the water meter at
inserting, at the downstream end of the measuring a given flow-rate shall be calculated from the
section, either a temporary pipe of the same length following equation:
and internal diameter as the water meter, or the %P = %P2 – %P1
water meter itself. The value arrived at shall be converted to the
7.3.3 Measurement and calculation of the pressure loss corresponding to the value of qs of the
actual pressure loss of the water meter water meter using the formula given in 7.6.
(measurement 2) (in-line meters only) NOTE If the test flow-rates with and without the meter present
differ, they may be adjusted to the same value by the square law
At the same test flow-rates used to determine the formula given in 7.6.
pipe pressure losses, in the same installation, with
7.4 Measurement uncertainty
the same pressure tappings and the same
manometer but with the water meter in position, The maximum relative uncertainty in the results of
the differential pressure %P2 across the metering the measurement of pressure loss shall be ± 5 %.
section shall be measured. 7.5 Pattern approval test. Pressure loss tests
The actual pressure loss, %P, of the water meter at Test methods and equipment used to determine
a given flow-rate shall be calculated from the pressure loss shall be as described in 7.2 and 7.3.
following equation:
The value of pressure loss shall be determined at not
%P = %P2 – %P1 less than the maximum flow-rate so that the water
The value arrived at shall be converted to the meter may be classified in accordance with clause 6
pressure loss corresponding to the value of qs of the of BS 5728-7:1997.
water meter using the formula given in 7.6.
NOTE If the flow-rates with and without the meter present
differ, they may be adjusted to the same value by the square law
formula given in 7.6.

© BSI 10-1999 13
BS 5728-3:1997
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

Measurement 1 (see 7.3.2)


¹P1 = Pressure loss of upstream and downstream pipe lengths
¹P1 = (¹PL2 + ¹PL1)
Measurement 2 (see 7.3.3)
¹P2 = Pressure loss of upstream and downstream pipe lengths + water meter
¹P2 = (¹PL2 + ¹PL1 + ¹Pmeter)
Ü ¹P2 – ¹P1 = (¹PL2 + ¹PL1 + ¹Pmeter) – (¹PL2 + ¹PL1)
= ¹Pmeter
Figure 8 — Pressure loss measurement test procedure (in-line meters)

14 © BSI 10-1999
BS 5728-3:1997
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

Measurement 1 (see 7.3.4)


¹P1 = Pressure loss of test manifold and upstream and downstream pipe lengths
¹P1 = (¹PL2 + ¹PL1 + ¹Pmanifold)
Measurement 2 (see 7.3.5)
¹P2 = Pressure loss of upstream and downstream pipe lengths + test manifold + water meter
¹P2 = (¹PL2 + ¹PL1 + ¹Pmanifold + ¹Pmeter)
Ü ¹P2 – ¹P1 = (¹PL2 + ¹PL1 + ¹Pmanifold + ¹Pmeter) – (¹PL2 + ¹PL1 + ¹Pmanifold)
= ¹Pmeter
Figure 9 — Pressure loss measurement test procedure (concentric type meters)
7.6 Initial verification test. Pressure loss tests It is necessary to obtain the pressure loss at qs in
Test methods and equipment for the measurement order to determine the meter’s pressure loss group
of pressure loss shall be as described in 7.2 and 7.3. (see clause 6 of BS 5728-7:1997).
The value of the pressure loss shall be determined at
a flow-rate between qp and qs.

© BSI 10-1999 15
BS 5728-3:1997

If necessary a calculation shall be made to obtain 8.1.5 Test duration


the pressure loss at qs based on the formula: The duration of the test specified in Table 2 shall be
(q )
2 a minimum.
s
pressure loss at q s = ------------------------------------------------
- × measured
2 pressure
8.1.6 Discharged volume
( test flow-rate )
loss The volume indicated by the meter under test shall
The value of the pressure loss at qs is required to not be less than the product of the specified
conform to the pressure loss group of the approved flow-rate and the duration of the test. To satisfy this
model (see clause 6 of BS 5728-7:1997). condition, sufficiently frequent corrections to the
flow-rate shall be made.
NOTE The pressure loss test is normally carried out on only a
selected sample of the meters. 8.2 Discontinuous flow tests
8.2.1 Outline of the test method
8 Accelerated wear tests
The tests consist of subjecting the meter to a
8.1 Continuous flow tests
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

specified number of starting and stopping cycles of


8.1.1 Outline of test method given duration. The constant flow-rate phase of each
The tests consist of subjecting the meter to specified cycle shall be kept at the same flow-rate throughout
constant flow-rates for given periods of time, to the duration of the test.
determine if the meter will subsequently perform The tests are designed to determine if the meter will
within its maximum permissible errors. subsequently perform within its maximum
For the convenience of laboratories, the duration of permissible errors.
the test can be divided up into periods of at least 6 h. For the convenience of laboratories, the duration of
8.1.2 Description of the installation the test can be divided up into periods of at least 6 h.
The installation shall consist of: 8.2.2 Description of the installation
a) a water supply (mains, non-pressurized tank, The installation shall consist of:
pressurized tank, pump, etc.); a) a water supply (mains, non-pressurized tank,
b) pipework. pressurized tank, pump, etc.);
8.1.3 Pipework b) pipework.
8.1.3.1 Description 8.2.3 Pipework
In addition to the meter or meters on test, the 8.2.3.1 Description
pipework shall comprise: NOTE The meters may be arranged in series or in parallel, or
the two systems may be combined.
a) a flow regulating device;
In addition to the meter or meters under test, the
b) one or two isolating devices; piping system shall comprise:
c) a device for measuring the water temperature a) one flow regulating device (per line of meters in
at the meter inlet; series, if necessary);
d) means for measuring the flow-rate; b) one or more isolating devices;
e) means for measuring the duration of the test. c) a device for measuring the temperature of the
If the pipework is open at the end, that end shall be water upstream of the meters;
located at a higher level than the upper part of the d) devices for measuring: the flow-rate, the
meter. duration of cycles, and the number of cycles;
The presence of the devices required in a), b), c) e) one or more flow interrupting devices (one per
and d) shall not cause cavitation in the pipework. line of meters in series).
8.1.3.2 Precautions to be taken If the pipework is open at the end, that end shall be
The test rig and the meter shall be cleared of air by located at a higher level than the upper part of the
bleeding. meter.
8.1.4 Flow-rate tolerance The presence of the devices required in a), b), c)
and e) shall not cause cavitation in the pipework.
The flow-rate shall be kept constant throughout the
test at a predetermined rate.
The relative variation of the flow-rate values during
each test shall not exceed ± 10 % (except when
starting and stopping).

16 © BSI 10-1999
BS 5728-3:1997

8.2.3.2 Precautions to be taken 8.3 Pattern approval tests. Accelerated wear


The test rig and the meter shall be cleared of air by tests
bleeding. The water meter shall be subjected to an endurance
Water hammer during the repeated flow increases test simulating service conditions.
and decreases shall be avoided by increasing, within In addition to periods of extended running at
permissible limits, the duration of these phases of specified flow-rates, water meters having N k 10
the cycle. shall be subjected to intermittent tests in which the
8.2.4 Flow-rate water meter is operated for a short period at a
flow-rate given in the test programme, followed by a
The relative variation of the flow-rate in the period of rest.
constant flow-rate phase of each cycle shall not
The test programme for an accelerated wear test is
exceed ± 10 %.
given in Table 2.
8.2.5 Cycles
Before the first test and after each test, the error
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

8.2.5.1 General curve shall be redetermined as described in 5.9.


A complete cycle shall comprise the following four After each test, the shift in the error curve shall not
phases: exceed 3 % between qmin and qt or 1.5 % between qt
a) a period from zero flow-rate to test flow-rate; and qs.
b) a period at constant test flow-rate; Test methods and equipment shall be as described
in 8.1 and 8.2.
c) a period from test flow-rate to zero flow-rate;
d) a period at zero flow-rate. 9 Test report
The test programme shall specify the number of
9.1 General rules
cycles, the duration of the four phases of a cycle, and
the total volume to be discharged. 9.1.1 General
8.2.5.2 Test timing tolerance The work carried out by the testing laboratory shall
be described in a report which accurately, clearly
The tolerance on the specified duration of each
and unambiguously presents the test results and all
phase shall not exceed ± 10 %.
relevant information.
The tolerance on the total test duration shall not NOTE The period of time for which results and conditions
exceed ± 5 %. concerning initial verification tests are to be kept is legally
specified in each country.
8.2.5.3 Tolerance on the number of cycles
For pattern approval tests, the records of the tests
The number of cycles shall not be less than that shall be kept for the length of time the approval is
specified, but shall not exceed this number by valid.
more than 1 %.
The report on the approval tests for a type of meter
8.2.6 Discharged volume tolerance and the register concerning the initial verification
The volume discharged during the test shall be tests shall contain:
within 5 % of the product of the specified flow-rate a) an identification of the test laboratory and the
and half the duration of the test. meter tested;
This precision can be obtained by sufficiently b) exact details of the conditions under which the
frequent corrections of the instantaneous flows and various tests were carried out;
operating periods.
c) the results and conclusions of the tests.
Table 2 — Test programme for an accelerated wear test
Meter Test Type of test Number of Duration of Period of Duration of start-up and
designation flow-rate interruptions pauses operation at test rundown
flow-rate
N s s
N k10 qp Discontinuous 100 000 15 15 s 0.15 (qp) a

with a minimum of 1 s
qs Continuous 100 h
N > 10 qp Continuous 800 h
qs Continuous 200 h
a (q ) is the number equal to the value of qp expressed in m3/h.
p

© BSI 10-1999 17
BS 5728-3:1997

9.1.2 Identification data to be included in all b) the signature of the officer accepting technical
reports and test registers responsibility for the test report;
The report on pattern approval tests for a particular c) the date of issue of the test report.
type of meter, and the register concerning the initial 9.2.2 Additions to test reports
verification tests, shall include as a minimum:
Additions to a test report after issue shall be made
a) identification of testing laboratory: only by a further document marked: “Supplement to
— name and address; test report. Serial No...”, and shall meet the relevant
b) identification of meter tested: requirements of the preceding subclauses.
— the name and address of the manufacturer 9.2.3 Publication of test reports
or the trademark used; When issued, the test report shall only be
— the metrological class and the rated flow qp; reproduced in its entirety.
— the type or model (only in the case of 9.3 Initial verification register. Required
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

approval tests for a particular type of meter); contents


— the year of manufacture and the individual The following information shall be recorded in the
serial number of the meter tested. initial verification register:
9.2 Pattern approval test report. Required a) date of test;
contents b) serial number of the meter tested;
The pattern approval test report shall contain, as a c) results of the tests;
minimum, in addition to a reference to BS 5728-7,
d) reference to a description of the test rig.
the information given in Table 3 and Table 4.
9.2.1 Administrative requirements
The pattern approval test report shall also include:
a) a statement to the effect that the test report
relates only to the samples tested;

18 © BSI 10-1999
BS 5728-3:1997

Table 3 — Test procedure and test results. Information required


Type of test Clause Information to be given in the pattern approval test report

All tests The date of testing


Measurement error tests 5 For each test flow-rate:
— flow-rate;
— water pressure;
— water temperature;
— characteristics of the calibrated reference device;
— indicated readings of the meter and the calibrated reference
device.
Pressure tests 6 The values of each test pressure applied and the time for which it was
maintained
For each test flow-rate:
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

Pressure loss tests 7


— maximum water temperature;
— flow-rate;
— meter upstream pressure;
— pressure loss.

Accelerated wear tests 8 Timetable of the tests carried out;


— Continuous tests 8.1 at least every 24 h, or once for every shorter period if the test is so
subdivided:
— maximum pressure;
— maximum temperature;
— flow-rate;
— meter reading at start and end of the test.
Timetable of the tests carried out;
— Discontinuous tests 8.2 at least every 24 h, or once for every shorter period if the test is so
subdivided:
— maximum pressure;
— maximum temperature;
— flow-rate;
— duration of the four phases of the cycle of the discontinuous
tests (see Table 2);
— number of cycles;
— meter reading at start and end of the test.

Table 4 — Test conclusions. Information required


Type of test Clause Information to be given in the pattern approval test report

Measurement error tests 5 Measurement error at each test flow-rate


Measurement error curve
Pressure tests 6 State if results satisfactory or not
Pressure loss tests 7 Pressure loss at overload flow-rate qs
Accelerated wear tests 8 Measurement error test values and the measurement error curves
taken before and after each accelerated wear test defined by the test
programme. For each individual meter, the measurement error
curves taken before and after each accelerated wear test shall be
plotted on the same graph in such a manner that the variations in
measurement error with respect to the maximum permissible errors,
are established. The scale of the ordinate of this graph shall be at
least 10 mm/%. The scale of the abscissa shall be logarithmic.

© BSI 10-1999 19
BS 5728-3:1997

Annex A (informative)
Summary of requirements and permitted tolerances in the measurement of
physical quantities associated with water meter test methods and equipment
A summary of requirements and permitted tolerances for water meter test methods and equipment is given
in Table A.1.
Table A.1 — Summary of requirements and permitted tolerances
Clause
Quantity Details Specified tolerance
referencea
Measurement error checks
Time Flow diversion method. Time of motion of flow switch Identical within 5 % 5.5.2
in each direction. in each direction
and < 2 % of total
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

duration of test
Volume Maximum error in the measurement of the volume of k 10 % of relevant 5.4.1
water passed through the meter under test (as maximum
measured by calibrated reference device). permissible error
Meter reading Maximum error in the reading of the volume indicated k 0.5 % 5.6
by meter under test.
Pressure Flow created by means other than a constant head
tank:
— Pressure variation upstream of meter. k 10 % 5.7.1
— Precision of measurement of pressure. k5% 5.7.1
Flow-rate Relative variation in flow-rate during each test:
— qmin to qt (not incl.) ± 2.5 % 5.7.2
— qt to qs ±5% 5.7.2
This is equivalent, in terms of pressure, to a pressure
variation upstream of the meter (flow to free air) or a
pressure loss variation (flow in closed circuits) of:
— qmin to qy (not incl.)
— qt to qs
Temperature Water temperature change during a test. < 5 °C 5.7.3
Uncertainty of measurement of temperature. < 1 °C 5.7.3
Pressure loss tests
Pressure Maximum uncertainty in the results of pressure loss ±5% 7.4
measurement.
Accelerated wear tests
— Continuous flow test
Flow-rate Relative variation in flow-rate during each test. ± 10 % 8.1.4
Time Tolerance on specified duration of test. Value is a min. 8.1.5
Volume Tolerance on volume discharged during test. Value is a min. 8.1.6
— Discontinuous flow test
Flow-rate Relative variation in flow-rate during constant test ± 10 % 8.2.4
flow phase.
Time Tolerance on specified duration of each test phase. ± 10 % 8.2.5.2
Tolerance on total test duration. ±5% 8.2.5.2
Number of Number of cycles of test. + 1 % and – 0 % 8.2.5.3
cycles
Volume Tolerance on volume discharged during test. ±5% 8.2.6
a
The references are to the subclauses in this Part of BS 5728.

20 © BSI 10-1999
BS 5728-3:1997

List of references (see clause 2)

Normative references

BSI publications
BRITISH STANDARDS INSTITUTION, London

BS 5728, Measurement of flow of cold potable water in closed conduits.


BS 5728-7:1997, Specification for single mechanical type meters.
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

© BSI 10-1999
BS 5728-3:
1997
BSI — British Standards Institution
BSI is the independent national body responsible for preparing
British Standards. It presents the UK view on standards in Europe and at the
international level. It is incorporated by Royal Charter.

Revisions

British Standards are updated by amendment or revision. Users of


British Standards should make sure that they possess the latest amendments or
editions.

It is the constant aim of BSI to improve the quality of our products and services.
We would be grateful if anyone finding an inaccuracy or ambiguity while using
this British Standard would inform the Secretary of the technical committee
responsible, the identity of which can be found on the inside front cover.
Tel: 020 8996 9000. Fax: 020 8996 7400.

BSI offers members an individual updating service called PLUS which ensures
that subscribers automatically receive the latest editions of standards.
Licensed Copy: lbocvzr lbocvzr, University of Loughborough, 26 June 2003, Uncontrolled Copy, (c) BSI

Buying standards

Orders for all BSI, international and foreign standards publications should be
addressed to Customer Services. Tel: 020 8996 9001. Fax: 020 8996 7001.

In response to orders for international standards, it is BSI policy to supply the


BSI implementation of those that have been published as British Standards,
unless otherwise requested.

Information on standards

BSI provides a wide range of information on national, European and


international standards through its Library and its Technical Help to Exporters
Service. Various BSI electronic information services are also available which give
details on all its products and services. Contact the Information Centre.
Tel: 020 8996 7111. Fax: 020 8996 7048.

Subscribing members of BSI are kept up to date with standards developments


and receive substantial discounts on the purchase price of standards. For details
of these and other benefits contact Membership Administration.
Tel: 020 8996 7002. Fax: 020 8996 7001.

Copyright

Copyright subsists in all BSI publications. BSI also holds the copyright, in the
UK, of the publications of the international standardization bodies. Except as
permitted under the Copyright, Designs and Patents Act 1988 no extract may be
reproduced, stored in a retrieval system or transmitted in any form or by any
means – electronic, photocopying, recording or otherwise – without prior written
permission from BSI.

This does not preclude the free use, in the course of implementing the standard,
of necessary details such as symbols, and size, type or grade designations. If these
details are to be used for any other purpose than implementation then the prior
written permission of BSI must be obtained.

If permission is granted, the terms may include royalty payments or a licensing


agreement. Details and advice can be obtained from the Copyright Manager.
BSI Tel: 020 8996 7070.
389 Chiswick High Road
London
W4 4AL

You might also like