PQ of Blender (V-shape)
By Pharma Editor - July 31, 2017
V Blender is used for mixing of active ingredients along with excipients and lubricants.
PQ of V Shaped Blender
Prerequisites
Verify all SOPs are available and operator’s is trained in SOPs and Training record available.
Equipment cleaning logbook available.
All the required instruments are calibrated and in place.
Mixing Time
Percentage Occupancy of the Blender
Blender RPM
Performance qualification of blender shall capture the effect of above variables on the blender
performance at various operating conditions. procedure for testing the consistency of RPM in
No load condition as well as in load condition at a given time interval.
Procedure:
https://pharmapathway.com/pq-of-blender-v-shape/ 1/6
12/4/2020 PQ of Blender (V-shape) | Pharma Pathway
Establish minimal and maximal optimum speed shall fix and documented.
Setting of minimal optimum speed with the potentiometer knob shall be done.
Machine shall run as per the standard operating procedure for about 60 minutes.
RPM shall be count manually at 15 mins interval and shall record. By a calibrated clamp-
meter measurement of actual current drawn by the motor shall noted at that speed.
All results should be met acceptance criteria.
Repeat the test for maximal optimum speed.
Validation Plan
No load trials
Minimal
Frequency
optimum Run Acceptance criteria
Occupancy of data
RPM Time
recording
https://pharmapathway.com/pq-of-blender-v-shape/ 2/6
12/4/2020 PQ of Blender (V-shape) | Pharma Pathway
30 min
There shall not be any leakages of oil
from the gear box
Maximal
Frequency
optimum Run Acceptance criteria
Occupancy of data
RPM Time
recording
To be 60 min 0% 0 min
Machine shall run smoothly without
establish
any abnormal noise and vibration
as per
equipment Motor shall not trip or excessively
heated up during operation
15 min
30 min
There shall not be any leakages of oil
from the gear box
https://pharmapathway.com/pq-of-blender-v-shape/ 3/6
12/4/2020 PQ of Blender (V-shape) | Pharma Pathway
Load trials
Minimal Frequency
Run Acceptance criteria
optimum RPM Occupancy of data
Time
recording
To be establish
Approx. 65
as per 60 min There shall not be any leakages of
%
Equipment 30 min oil from the gear box as well as
powder from the bowl
To be establish
Approx. 65
as per 60 min There shall not be any leakages of
%
Equipment 30 min oil from the gear box as well as
powder from the bowl
Requalification Criteria
https://pharmapathway.com/pq-of-blender-v-shape/ 5/6
12/4/2020 PQ of Blender (V-shape) | Pharma Pathway
Any major modification to the existing machine which can affect the performance of the
equipment.
In case of failure to meet acceptance criteria.