General Specifications: Model OR100 EXA100 Series Panel Mount ORP Analyzer
General Specifications: Model OR100 EXA100 Series Panel Mount ORP Analyzer
FEATURES
Converter
• 96 x 96 mm panel mount design for indoor use
• Easy access to routine maintenance mode
• Practical functions packed in a compact housing
• Large 4-digit display
• One analog output and two/four contact outputs
as standard
• Simple one touch automatic calibration
• Self-diagnostics (e.g., calibration failure, measuring
range failure, converter failure)
GS 12C11A01-01E-H
1st Edition
2
■ GENERAL SPECIFICATIONS Contact output
Type: Relay contact output
1. Panel Mounted ORP Converter Number of contacts: 2 or 4 outputs (must be specified
when ordering)
Model: OR100 Contact action: On/Off
Measurement: Oxidation-reduction potential (ORP) of a Contact functions: Selectable: High, low, high-high,
solution low-low, high-high/low-low limit
Measuring range: -1500 to 1500 mV alarms, FAIL
Indication Alarm setting resolution: 10 mV
Display: Digital (LED) Contact output hysteresis: 0 to 1000 mV (configurable)
Range: -1500 to 1500 mV Contact output delay time: 0 to 200 seconds
Resolution: 1 mV (configurable)
Indication items: ORP reading, setting, status Contact rating: In case of 2 contact outputs
Input signal S1: 240 VAC 3A or 30 VDC 3A (resistance load), Form C
ORP input range: -1500 to 1500 mV (NC/NO/COM, 3 terminals)
Transmission signal output S2: 240 VAC 3A or 30 VDC 3A (resistance load), Form A
Number of output points: 1 output, ORP reading only (NO/COM, 2 terminals) In case of 4 contact outputs
Output signal: 4 to 20 mA DC, isolated S1: 240 VAC 3A or 30 VDC 3A (resistance load), Form C
Load resistance: 600Ω or less (NC/NO/COM, 3 terminals)
Transmission signal range: Configurable within S2, S3, S4: 240 VAC 3A or 30 VDC 3A (resistance load),
measuring range (by 100 mV unit), Form A, shared common Maximum load
factory setting: -1500 to 1500 mV, current on common is 3A.
minimum span: 100 mV Contact status:
Maintenance output signal: Output hold “enabled/
disabled” selectable Terminal
Hold output value: Last measured value/preset NC
value (2.0 to 20.8 mA) selectable 1
Fail output signal: Downscale burnout (2 mA) S1 NO
2
“enabled/disabled” selectable
COM 3
S2 4
S3 5
S4 6
COM 7 F011E.eps
Function selected
Contact H, L, HH, LL, HH/LL limit alarms FAIL
Power off Power on Power off Power on
No alarm Alarm No alarm Alarm
S1 NO-COM Open Open Closed Open Closed Open
NC-COM Closed Closed Open Closed Open Closed
S2 Open Open Closed Open Closed Open
S3 when specified Open Open Closed Open Closed Open
S4 when specified Open Open Closed Open Closed Open
T01E.EPS
Note: When a contact is activated, the LED on display panel turns on.
Dimensions
OR100 Panel Mounted ORP Converter
Panel Cutout Dimensions
Unit : mm
Shield cover
Unit : mm
125 Min.
91.6
92
51-60 41
1-10
145 Min.
Terminal cover
11 118
EXA PH100
92 +0.8
0
pH
91.6
111
S1 S3
96
S2 S4
CANCEL
SET CAL MEAS
ENT 92 +0.8
0
Panel depth : 1 to 10
96 100 F003.EPS
YOKOGAWA HEADQUARTERS YOKOGAWA CORPORATION OF AMERICA Yokogawa has an extensive sales and dis-
9-32, Nakacho 2-chome, 2 Dart Road
Musashinoshi Newnan GA 30265 tribution network.
Tokyo 180 United States
Japan Tel. (1)-770-253-7000 Please refer to the European web-site
Tel. (81)-422-52-5535 Fax (1)-770-251-2088
Fax (81)-422-55-1202 E-mail: info@yca.com (www.yokogawa-europe.com) to contact
E-mail: webinfo@mls.yokogawa.co.jp www.yca.com
www.yokogawa.com your nearest representative.
GS 12C11A01-01E-H
Subject to change without notice Printed in the Netherlands, 01-405 (A) Q
Copyright©