General Foundation® Fieldbus transmitter

for Conductivity or Resistivity


Specifications Model SC202G-F

Flexibility, reliability and low maintenance are among the


benefits provided by the EXA SC202G-F conductivity analyzer.
Designed to meet the exacting requirements of measuring
conductivity and resistivity in the modern industrial environment, it
contains many features to ensure the best precision whatever the
application.

This transmitter is housed in a robust IP65 field mountable case.


The Foundation ® Fieldbus digital communication and a clear LCD
make the SC202G-F a complete industrial analyzer

The famous EXA sensor diagnostics are now available through


status reporting which is part of the Foundation® Fieldbus
protocol, utilizing all build-in intelligents. The logbook can be used
to record events like calibrations and diagnostic messages.

A wide variety of temperature compensation possibilities (NaCl


according to IEC 746-3, manual TC, preprogrammed matrices
and a free programmable 5x5 matrix), provide a high-accuracy
measurement with minimum effort.

The addition of a USP23 pure water monitoring feature makes


the SC202G-F ideal for the pharmaceutical industry.

FEATURES
● Universal conductivity/resistivity, software switchable
● On-line sensor checking
● Foundation® Fieldbus interface (DD available)
● Event logbook in software
● Matrix temperature compensation for pure water applications
● Password protection for all levels of software
● Intrinsically safe version and non-incendive versions
● Easy to use EXA control panel
● Pharmaceutical pure water monitoring (USP23)

SYSTEM CONFIGURATION

Sensors Cables Fittings Transmitters Accessories

GS 12D7B3-02E-H
2nd Edition
2
ACCURATE CONDUCTIVITY MEASUREMENT Matrix
Electrode selection The SC202G-F is equipped with a matrix type algorithm for
In order to make precise conductivity measurements, there are a accurate temperature compensation in various applications.
number of prerequisites. Most important is the selection of For pure water applications the following choices can be made:
suitable sensors. Special attention should be paid to the choice • HCl (cation) compensation (0 - 80 °C)
of the sensors to ensure compatibility with both the chemical • Ammonia compensation (0 - 80 °C)
composition and the specific conductivity of the fluid to be • Morpholine compensation (0 - 80 °C)
measured. The installation of the sensor is also very important for For higher conductivity ranges the choices are:
a correct measurement. Other Yokogawa specification sheets • HCl (1 - 5 %, 0 - 60 °C)
cover the choice of sensors and holders, and any Yokogawa • NaOH (1 - 5 %, 0 - 100 °C)
sales office can provide expert advice. • 25 points (5 x 5) user programmable matrix.
Applications using highly conductive fluids, inductive conductivity
measurement should be considered as an alternative because of Example of matrix temperature compensation
lower maintenance requirements. 500
Sodium hydroxide (0 - 5 % at 0 - 100 °C)
x 5%
450
x 4%
Measuring range as function of the cell constant 400

350
x 3%
SC42-EP15(D)

mS/cm
SC42-EP14/8

SC42-EP04/8

300
SC42-SP34

SC42-SP24

SC4A-.-002

SC4A-.-010

ISC40G-PG

ISC40G-TF

250
x 2%
200

150
x 1%
100.000.000,00 100

50
10.000.000,00
0
1.000.000,00 0ºC 25ºC 50ºC 75ºC 100ºC

100.000,00 Temperature (Celsius)

10.000,00 Polarization
uS/cm

1.000,00 Usable range Manual Temperature coefficient


Low
100,00 It is also possible to have a linear compensation with
10,00 programmable coefficient. A second conductivity display with
1,00
independent TC is possible.
0,10
0,01
At start-up a known temperature coefficient may be entered from
C = 0.01/cm

C = 10.0/cm

C = 0.02/cm
C = 0.1/cm

C = 1.0/cm

C = 0.1/cm
C = 1/cm

sensor type
the service mode, or the TC can be adjusted by calibration, using
actual process solutions. The freely programmable reference
With 2-electrode systems some polarization may decrease the temperature also contributes to a high accuracy measurement.
measured conductivity value at higher concentrations.
LOGBOOK
Fault diagnosis This is a history record of events, calibrations and alarms. A total
The measurement is continuously checked for faults, giving of approximately 100 events can be stored in the Logbook.
warning of non-immersion, fouling and cable problems. These Important diagnostic information can also be read on the LCD or
faults are signaled by the FAIL flag, and transmitted via status through Fieldbus communications.
along with the process value. An error code in the message
display appears. CONCENTRATION
When temperature and conductivity values are know,
Transmitter concentration can be determined. EXA SC202G-F holds most
The system is designed to minimize the effect of external relevent concentration curves which provide accurate
influences. Signal processing ensures that only changes in concentration measurements.
conductivity cause changes in reading. That is why EXA CONDUCTIVITY (S/cm) Output in %
instruments offer a wide range of temperature compensation 1,000 100

possibilities. 800 80

600 60
USP23 Monitoring Conductivity
SC202G-F monitors water quality according to the USP23 400 40 Output
directive (United States Pharmacopeia). Both compensated and
200 20
uncompensated conductivity values can be read from the display,
as can the solution temperature. There is an error message (E13) 0 0
0 2 4 6 8 10 12 14 16 18 20 22 24
alarm to indicate that the USP23 limit is exceeded. USP23
CONCENTRATION (% by weight)
determines a level of uncompensated conductivity for each
temperature. The water must be below this level to be
acceptable. This curve is preprogrammed into SC202G-F. THREE LEVEL OPERATION
The EXA SC202G-F converter uses a 3-level operating system
DD specification Advanced functions are separated from conventional operation to
For optimum use of the Foundation® Fieldbus functionality, the avoid confusion. They can be activated as required for each
device description (DD) of the SC202G-F is required. individual application.
MAINTENANCE level : Normal day to day operation
PROCESS TEMPERATURE COMPENSATION COMMISSIONING level : During commissioning
Automatic, according to NaCl tables (IEC-746-3 tables) SERVICE level : Only for specialized functions
From the factory, the SC202G-F is configured for non-linear With this 3-level user-friendly approach, the instrument can be
temperature compensation according NaCl tables which will give operated by anyone. Commissioning is straight-forward. Special
accuracy in most measurements. In this case no site adjustments functions, available via access codes, are invisible during normal
are required. For applications where NaCl compensation is not operation. All three levels can be separately protected against
sufficient, other compensation possibilities are presented below. unauthorized access by a 3-digit password.

GS 12D7B3-02E-H
3

GENERAL SPECIFICATIONS B. Ambient operating temperature


A. Input specifications : Two or four electrodes measurement : -10 to +55 °C (10 to 130 ºF)
with square wave excitation. Cell Excursions to -30 to +70 °C
constants from 0.008 to 50 cm-1 (-20 to 160 ºF) will not damage the
WU40 sensor cable up to 20m. Up to instrument, specification maybe
30m total using BA10 junction box adversely affected
and WF10 extension cable Drift < 500 ppm/°C
B. Detection method : Frequency, read-pulse position and C. Storage temp. : -30 to +70 ºC (-20 to 160 ºF)
reference voltage are dynamically D. Humidity : 10 to 90% RH non-condensing
optimized. E. FF specification : Twisted 2-wire with shield.
C. Input ranges F. Housing : Cast aluminium case with chemically
- Conductivity : 0.000 µS/cm to 1999 mS/cm at resistant coating, cover with flexible
25 °C (77 °F) reference temperature. polycarbonate window. Case color is
Minimum : 0.2 µS x C at process temperature. off-white and cover is moss green.
Maximum : 500 mS x C at process temperature. Cable entry is via two 1/2” polyamide
- Resistivity : 0.000 kΩ - 999 MΩ/C at 25 °C glands. Cable terminals are provided
(77 °F) reference temperature. for up to 2.5 mm2 finished wires.
Minimum : 0.002 kΩ/C at process temperature. Weather resistant to IP65 and NEMA
Maximum : 5 MΩ/C at process temperature. 4X standards. Pipe wall or panel
- Temperature mounting, using optional hardware.
Pt1000 : -20 to +250 °C (0 - 500 °F) G. Data protection : EEPROM for configuration and
Pt100 and Ni100 : -20 to +200 °C (0 - 400 °F) logbook, and lithium battery for clock.
8K55 NTC : -10 to +120 °C (10 - 250 °F) H. Watchdog timer : Checks microprocessor
Pb36 NTC : -20 to +120 °C (0 - 250 °F) I. Automatic safeguard: Return to measuring mode when no
E. Temperature compensation keystroke is made for 10 min.
: Automatic, for temperature ranges J. Operation protection: 3-digit programmable password.
mentioned under C (inputs). K. Regulatory compliance
- Reference temp. : programmable from 0 to 100 °C or - EMC : meets council directive 89/336/EEC
30 to 210 °F (default 25 °C). - Emmission : meets EN 55022 Class A
F. Compensation algorithm - Immunity : meets EN 61000-6-2
-NaCl : According IEC 746-3 NaCl tables L. DD specification : The SC202 Device Description is
(default). available enabling communications.
-T.C. : Two independent user programmable
temperature coefficients, from 0.00%
to 3.50% per °C (°F) by adjustment or MODELCODE
calibration.
- Matrix : : Conductivity as a function of concen- Model Suffix Option Description
tration and temperature. Choice out
of 5 preprogrammed matrices and a SC202G Conductivity Transmitter
25-point user-programmable matrix. General Purpose version
G. Logbook : Software record of important events
SC202S Conductivity Transmitter
and diagnostic data. Available
through FF interface Intrinsic Safe version
H. Display : Custom liquid crystal display, with a Type -A Milli-amp (+HART®) version
main display of 31/2 digits 12.5 mm -F FOUNDATION® Fieldbus version
high. Message display of 6 alpha- -N Non-Incendive Milli-amp
numeric characters, 7 mm high.
(+HART®) version
Warning flags and units (µS/cm,
mS/cm, kΩ•cm and MΩ•cm) as -B Non-Incendive FOUNDATION®
appropriate. Fieldbus version
I. Power supply : 9 to 32 volt -E Always E
J. Input isolation : 1000 VDC Options /H Hood for Sun Protection
K. Shipping Details : Package size w x h x d
/U Pipe & Wall mounting hardware
290 x 225 x 170 mm.
11.5 x 8.9 x 6.7 in. /SCT Stainless steel tagplate
Packed weight approx. 2.5 kg (5lb). /Q Calibration certificate
Operating specifications /T Turck Fieldbus connector
A. Performance : Conductivity/Resistivity
- Accuracy : ≤ 0.5 %
Performance : Pt1000Ω, Ni100Ω and Pb36 NTC
- Accuracy : ≤ 0.3 °C
Performance : PT100Ω and 8k55Ω
- Accuracy : ≤ 0.4 °C
Performance : Temperature compensation
- NaCl table : ≤1%
- Matrix : ≤3%
- Ambient influence : ≤ 0.05 %/°C
- Step response : 90 % (< 2 decades) in ≤ 7 seconds

GS 12D7B3-02E-H
4

DIMENSIONS AND MOUNTING

162 (6.4)

min. 203
(min. 8.0)
180 (7) 154
(6.06) 30 (1.18)

(1.18)
30 (1.2)

30
(min.9.0)
min.229

(6.77)
172
92 (3.6)
115 (4.5)
2x ø4
(0.16)

1/2" SUPPLY SPACING PANEL


CUT-OUT DIMENSION CUT-OUT DIMENSIONS

1/2" INPUT

56±0.2
(2.20) M6 bolts (2x)

Panel cut-out, spacing and mounting Dimensions

wall mounting pipe mounting pipe mounting


(vertical) (horizontal)
56
(2.20)

2x ø6.5
(0.26)
(7.87)
200

4x ø10
(0.4)

145

70 (5.70)
(2.75)
2” ND. pipe
Option /U: Universal pipe/wall mounting kit

Universal pipe/wall mounting

YOKOGAWA HEADQUARTERS YOKOGAWA CORPORATION OF AMERICA Yokogawa has an extensive sales and
9-32, Nakacho 2-chome, 2 Dart Road
Musashinoshi Newnan GA 30265 distribution network.
Tokyo 180 United States
Japan Tel. (1)-770-253-7000 Please refer to the European web-site
Tel. (81)-422-52-5535 Fax (1)-770-251-2088
Fax (81)-422-55-1202 E-mail: info@yca.com (www.yokogawa-europe.com) to contact
E-mail: webinfo@mls.yokogawa.co.jp www.yca.com
www.yokogawa.com your nearest representative.

YOKOGAWA EUROPE B.V. YOKOGAWA ELECTRIC ASIA Pte. Ltd.


Databankweg 20 5 Bedok South Road
3821 AL AMERSFOORT Singapore 469270
The Netherlands Singapore
Tel. +31-33-4641 611 Tel. (65)-241-9933
Fax +31-33-4641 610 Fax (65)-241-2606
E-mail: info@yokogawa.nl E-mail: webinfo@yas.com.sg
www.yokogawa-europe.com www.yokogawa.com.sg

GS 12D7B3-02E-H
Subject to change without notice Printed in The Netherlands, 02-301 (A) Q