General Specifications: Foundation Fieldbus Transmitter For Conductivity or Resistivity Model SC202G-F
General Specifications: Foundation Fieldbus Transmitter For Conductivity or Resistivity Model SC202G-F
FEATURES
● Universal conductivity/resistivity, software switchable
● On-line sensor checking
● Foundation® Fieldbus interface (DD available)
● Event logbook in software
● Matrix temperature compensation for pure water applications
● Password protection for all levels of software
● Intrinsically safe version and non-incendive versions
● Easy to use EXA control panel
● Pharmaceutical pure water monitoring (USP23)
SYSTEM CONFIGURATION
GS 12D7B3-02E-H
2nd Edition
2
ACCURATE CONDUCTIVITY MEASUREMENT Matrix
Electrode selection The SC202G-F is equipped with a matrix type algorithm for
In order to make precise conductivity measurements, there are a accurate temperature compensation in various applications.
number of prerequisites. Most important is the selection of For pure water applications the following choices can be made:
suitable sensors. Special attention should be paid to the choice • HCl (cation) compensation (0 - 80 °C)
of the sensors to ensure compatibility with both the chemical • Ammonia compensation (0 - 80 °C)
composition and the specific conductivity of the fluid to be • Morpholine compensation (0 - 80 °C)
measured. The installation of the sensor is also very important for For higher conductivity ranges the choices are:
a correct measurement. Other Yokogawa specification sheets • HCl (1 - 5 %, 0 - 60 °C)
cover the choice of sensors and holders, and any Yokogawa • NaOH (1 - 5 %, 0 - 100 °C)
sales office can provide expert advice. • 25 points (5 x 5) user programmable matrix.
Applications using highly conductive fluids, inductive conductivity
measurement should be considered as an alternative because of Example of matrix temperature compensation
lower maintenance requirements. 500
Sodium hydroxide (0 - 5 % at 0 - 100 °C)
x 5%
450
x 4%
Measuring range as function of the cell constant 400
350
x 3%
SC42-EP15(D)
mS/cm
SC42-EP14/8
SC42-EP04/8
300
SC42-SP34
SC42-SP24
SC4A-.-002
SC4A-.-010
ISC40G-PG
ISC40G-TF
250
x 2%
200
150
x 1%
100.000.000,00 100
50
10.000.000,00
0
1.000.000,00 0ºC 25ºC 50ºC 75ºC 100ºC
10.000,00 Polarization
uS/cm
C = 10.0/cm
C = 0.02/cm
C = 0.1/cm
C = 1.0/cm
C = 0.1/cm
C = 1/cm
sensor type
the service mode, or the TC can be adjusted by calibration, using
actual process solutions. The freely programmable reference
With 2-electrode systems some polarization may decrease the temperature also contributes to a high accuracy measurement.
measured conductivity value at higher concentrations.
LOGBOOK
Fault diagnosis This is a history record of events, calibrations and alarms. A total
The measurement is continuously checked for faults, giving of approximately 100 events can be stored in the Logbook.
warning of non-immersion, fouling and cable problems. These Important diagnostic information can also be read on the LCD or
faults are signaled by the FAIL flag, and transmitted via status through Fieldbus communications.
along with the process value. An error code in the message
display appears. CONCENTRATION
When temperature and conductivity values are know,
Transmitter concentration can be determined. EXA SC202G-F holds most
The system is designed to minimize the effect of external relevent concentration curves which provide accurate
influences. Signal processing ensures that only changes in concentration measurements.
conductivity cause changes in reading. That is why EXA CONDUCTIVITY (S/cm) Output in %
instruments offer a wide range of temperature compensation 1,000 100
possibilities. 800 80
600 60
USP23 Monitoring Conductivity
SC202G-F monitors water quality according to the USP23 400 40 Output
directive (United States Pharmacopeia). Both compensated and
200 20
uncompensated conductivity values can be read from the display,
as can the solution temperature. There is an error message (E13) 0 0
0 2 4 6 8 10 12 14 16 18 20 22 24
alarm to indicate that the USP23 limit is exceeded. USP23
CONCENTRATION (% by weight)
determines a level of uncompensated conductivity for each
temperature. The water must be below this level to be
acceptable. This curve is preprogrammed into SC202G-F. THREE LEVEL OPERATION
The EXA SC202G-F converter uses a 3-level operating system
DD specification Advanced functions are separated from conventional operation to
For optimum use of the Foundation® Fieldbus functionality, the avoid confusion. They can be activated as required for each
device description (DD) of the SC202G-F is required. individual application.
MAINTENANCE level : Normal day to day operation
PROCESS TEMPERATURE COMPENSATION COMMISSIONING level : During commissioning
Automatic, according to NaCl tables (IEC-746-3 tables) SERVICE level : Only for specialized functions
From the factory, the SC202G-F is configured for non-linear With this 3-level user-friendly approach, the instrument can be
temperature compensation according NaCl tables which will give operated by anyone. Commissioning is straight-forward. Special
accuracy in most measurements. In this case no site adjustments functions, available via access codes, are invisible during normal
are required. For applications where NaCl compensation is not operation. All three levels can be separately protected against
sufficient, other compensation possibilities are presented below. unauthorized access by a 3-digit password.
GS 12D7B3-02E-H
3
GS 12D7B3-02E-H
4
162 (6.4)
min. 203
(min. 8.0)
180 (7) 154
(6.06) 30 (1.18)
(1.18)
30 (1.2)
30
(min.9.0)
min.229
(6.77)
172
92 (3.6)
115 (4.5)
2x ø4
(0.16)
1/2" INPUT
56±0.2
(2.20) M6 bolts (2x)
2x ø6.5
(0.26)
(7.87)
200
4x ø10
(0.4)
145
70 (5.70)
(2.75)
2” ND. pipe
Option /U: Universal pipe/wall mounting kit
YOKOGAWA HEADQUARTERS YOKOGAWA CORPORATION OF AMERICA Yokogawa has an extensive sales and
9-32, Nakacho 2-chome, 2 Dart Road
Musashinoshi Newnan GA 30265 distribution network.
Tokyo 180 United States
Japan Tel. (1)-770-253-7000 Please refer to the European web-site
Tel. (81)-422-52-5535 Fax (1)-770-251-2088
Fax (81)-422-55-1202 E-mail: info@yca.com (www.yokogawa-europe.com) to contact
E-mail: webinfo@mls.yokogawa.co.jp www.yca.com
www.yokogawa.com your nearest representative.