General Specifications: Model SC450 Conductivity / Resistivity Analyzer
General Specifications: Model SC450 Conductivity / Resistivity Analyzer
The EXAxt 450 offers full functionality with PID control on either
mA output(s) or on contact output(s).
The contact outputs can be selected as pulse frequency
controlled or pulse length controlled contact function to control
chemical metering pumps or solenoid valves.
System Configuration
GS 12D7B5-E-E
1st Edition
2
General specifications of EXAxt SC450 F) Contact Input : Remote range switching to 10 times
the programmed range.
A) Input specifications: Two or four electrodes measurement - Contact open : If impedance > 100 kΩ: Range 1
with square wave excitation, using max - Contact closed : If impedance < 10 Ω: Range 2
60m (200ft) cable (WU40/WF10) and (10 x Range 1)
cell constants from 0.005 to 50.0 cm-1
G) Temperature compensation
B) Input ranges : Automatic or manual, for temperature
Conductivity : 0.000 µS/cm - 2000 mS/cm ranges mentioned under C (inputs).
Minimum : 1µS/cm (underrange 0.00 µS x C) - Reference temp. : programmable from 0 to 100 °C or
Maximum : 200 mS/cm (overrange 2000 mS x C) 30 - 210 °F (default 25 °C).
Resistivity : 0.0 Ω x cm - 1000 MΩ x cm
Minimum : 5 Ω/cm (underrange 0.0 Ω/C) H) Compensation algorithm
Maximum : 1 MΩ/cm (overrange 1000 MΩ/C) : According IEC 746-3 NaCl tables
Temperature : Pt1000 -20 to 250ºC (0 - 500ºF) (default).
: Pt100 -20 to 200ºC (0 - 400ºF) Two independent user programmable
: Ni100 -20 to 200ºC (0 - 400ºF) temperature coefficients, from 0% to
: NTC 8k55 -10 to 120ºC (10 - 250ºF) 3.5% per °C (°F) by adjustment or
: Pb36 (JIS NTC 6k) -20 to 120ºC (0 - 250ºF) calibration.
Matrix compensation : With conductivity function of
C) Accuracy concentration and temperature.
Conductivity/resistivity : ≤ 0.5 % of reading Choice out of 13 preprogrammed
Temperature : ≤ 0.3 ºC (≤ 0.4 ºC for Pt100) matrixes and 2 100-points
mA outputs : ≤ 0.02 mA user-programmable matrices.
Ambient temperature influence I) Calibration : Semi-automatic calibration using
: ± 0.05% /ºC pre-configured OIML (KCl) buffer
Step respons : ≤ 4 sec for 90% (for a 2 decade step) tables, with automatic stability check.
Manual adjustment to grab sample.
D) Transmission signals
General : Two isolated outputs of 4-20 mA. DC J) Logbook : Software record of important events
with common negative. Maximum load and diagnostic data readily available
600Ω. Bi-directional HART® in the display or through HART.
digital communication, superimposed
on mA1 (4-20mA) signal. K) Display : Graphical Quarter VGA (320 x 240
Output Function : Linear or Non linear (21-step table) pixels) LCD with LED backlight and
output for pH, temperature, ORP or rH. touchscreen. Plain language
Control function : PID control. messages in English, German,
Burn out function : Burn up (21.0 mA) or burn down French, Spanish, and Italian.
(3.6 mA) to signal failure acc.
NAMUR NE43. L) Shipping details
: Adjustable damping Package size : 293 x 233 x 230 mm (L x W x D)
: Expire time (11.5 x 9.2 x 9.1 inch)
Hold : The mA-outputs are frozen to the Package weight : app. 2.5 kg (5.5lbs)
last/fixed value during calibration/
commissioning M) Housing : Cast aluminium case with chemically
resistant coating, cover with flexible
E) Contact outputs polycarbonate window. The colour
General : Four SPDT relay contacts with display of the case and cover is silvergrey.
indicators. Contact outputs configurat- Cable entry via six M20 polyamide
able for hysteresis and delay time. glands. Cable terminals are provided
Switch capacity : Maximum values 100 VA, for up to 2.5 mm2 finished wires.
250 VAC, 5 Amps. Weather resistant to IP66 and
Maximum values 50 Watts, NEMA4X standards. Note that the
250 VDC, 5 Amps. glands must be installed properly.
Status : High/Low process alarms, selected Pipe, wall or panel mounting, using
from conductivity, resistvity, optional hardware.
concentration or temperature.
Configurable delay time and hysteresis. N) Power supply : 85-265 VAC (±10%). Max 10VA, 47-
PID duty cycle or pulsed frequency 63Hz, 916-30 VDC (±10%), max 10W
control.
FAIL alarm O) Regulatory compliance
Control function : On / Off EMC : Meets directive 89/336/EEC
: Adjustable damping Emission conform EN 55022 class A
: Expire time Immunity conform IEC 61326-1
Hold : Contact can be used to signal the Low Voltage : Meets directive 73/23/EEC Conform
hold situation. IEC 61010-1, UL61010C-1 and CSA
Fail safe : Contact S4 is programmed as fail-safe 22.2 No. 1010.1, Installation category
contact. II, Pollution degree 2
GS 12D7B5-E-E
3
FM Class 1, Div. 2, Group ABCD, T6 Zoom screen
for Ta -20 to 55ºC
Certification for CCSAus, Kema Keur
Status screen
The Status screen gives access to diagnostic information with
Go to trend screen
regards to analyser or sensors.
Go to zoom screen
No malfunction detected.
Go to status screen
Go to maintenance screen Soft alarm detected. Maintenance is recommended for
best accuracy.
The main screen displays: Hard alarm is detected indicating malfunction that is
- The primary variable in large font (user selectable) critical for good analysis. When this key is pressed
- Other process variable(s) in small font details are displayed with regards to detected
- Unit symbols malfunction and troubleshooting guidelines are
- Tagnumber (user programmable) displayed to resolve the malfunction.
- Process description (user programmable)
- Status of contact output(s)
- Status indicator during HOLD and WASH situation
- Main function keys
Trend screen
Maximum
SC 450
Live reading
120.0
Maintenance screen
SC
90.0
The maintenance screen gives access to calibration,
Minimum
commissioning and setup of the instrument. These levels can be
60.0
protected by passwords.
30.0 109.3 µS/cm Example
Maximum
Average
Minimum
mS/cm
mA output for fault detection by controlling the mA output in the x
300
following way: 250
x 2%
4-20 mA scaled to measuring range 200
3,8 -4 mA for underflow detection 150
x 1%
20- 20,5 mA for overflow detection 100
SC450 offers two additional modes for advanced temperature In addition a free programmable matrix can be selected for
compensation in processes where NaCl compensation does not applications not covered by these standard ones. Measurement
offer the required accuracy: outside the range of the matrix is possible, but can result in
inaccuracies dependent on the chemistry of the electrolyte solution.
1. Linear temperature coefficient setting
2. Matrix temperature compensation
3,5
Linear temperature coefficient 3,0
This function is used when the water chemistry is unknown.
2,5
mS/cm
GS 12D7B5-E-E
5
level is monitored by Conductivity Analysis. If the conductivity Output functions
exceeds the limit, then the user must perform laboratory analysis
to verify the quality of the WFI. SC450 has the limits of USP built Current Outputs
in the firmware and as soon as the water quality does not meet The SC450 offers two isolated 4- 20 mA outputs for indication,
the requirement of step 1, a fault message is generated; this recording or control functions. The parameter that is transmitted
information can be transmitted through HART, contact output or through the current output(s) can be selected out of:
High/Low mA output according to NAMUR recommendation 43.
It is also possible to program a contact output to close when • Conductivity or Resistivity
a preset safety margin is exceeded. The USP functionality is • Temperature
independent on the chosen conductivity range for the mA outputs. • Concentration
The functionality is active in both Conductivity and Resistivity mode. • PID control
10 000 000.00
1 000.00
Code mA Conc. Example Cond. Example
Output 4-20 % H2SO4 mS/cm
100.00 0 4.0 0 0
10.00
5 4.8 1.25 60
10 5.6 2.5 113
µS/cm 1.00 15 6.4 3.75 180
0.10
SC402/SC202/SC150/SC450 20 7.2 5 211
DC402
25 8.0 6.25 290
0.01 30 8.8 7.5 335
SC42-SP34 SC4A-002 SC42-SP24 SC4A-010 SC42-EP14 SC42- SC42-EP04
SX42-SX34 (0.02/cm) SX42-SX24 (0.1/cm) (1/cm) EP15(D) SC42-FP04 35 9.6 8.75 383
(0.01/cm) (0.1/cm) (1/cm) SC42-TP04
(10/cm) 40 10.4 10 424
45 11.2 11.25 466
Measuring range as function of the cell constant 50 12.0 12.5 515
55 12.8 13.75 555
Note: With 2-electrode systems, 60 13.6 15 590
polarization may decrease the 65 14.4 16.25 625
conductivity value at higher 70 15.2 17.5 655
concentrations. Applications using 75 16.0 18.75 685
highly conductive fluids, inductive 80 16.8 20 718
conductivity measurement should be 85 17.6 21.25 735
considered as an alternative because 90 18.4 22.5 755
of lower maintenance requirements. 95 19.2 23.75 775
100 20.0 25 791
GS 12D7B5-E-E
6
Contact Outputs Pulse frequency control
The SC450 offers four SPDT contact outputs. All contact outputs This function is used when the chemical is dosed with pulsating
can be configured for Alarm and Control functions. Contact S1, dosing pumps, where each pulse represents a certain volume of
S2 and S3 are in powered condition when the setpoint is reagent. The maximum number of pulses per minutes is selected,
exceeded. These contacts are normally used for HI, LO Alarm but must be between 1 and 70 pulses per minute. The frequency
or Control functions. Contact S4 is in alarm condition when the of pulses is determined by the PID function of the analyser.
power is down. This fail-safe contact is normally used to
signal a fail situation. The Proportional Band, the Integral Time and the Derivative
Function are set in the control menu of the analyser.
Control Function
SC450 offers advanced process control. Each of the current outputs % controller output 0.3 s
and the contact outputs can represent the PID control output.
100
If a contact is used for process control the contact output can be Maximum pulse frequency
pulse length controlled or pulse frequency controlled.
50% 50%
50 No pulses
ton toff 0
Duty cycle Range
0 Duty cycle
Range
*1
If the tagnumber is predefined with the purchase, Yokogawa will inscript the tagplate with the specified tagnumber, and program the tag-
number in the transmitter.
GS 12D7B5-E-E
7
Wiring Diagram
Connection Diagram
31
C
32 21
NC S1
33 Input contact
NO
41 22
C
42 NC
S2
Sensor Inputs
INPUTS
43 11
NO
+
Relay 51
C
Temp
- 12
Contacts 52 v- 13
NC S3
53 i- 14
NO
71
C
Fail Safe
72 NO
S4
73 v+ 15
NC
i+ 16
80
(3.15")
2x ø6.5
(0.26")
200
(7.87")
4x ø10
(0.4")
141.5
(5.57")
70
(2.75")
GS 12D7B5-E-E
8
Wall and pipe mounting diagram
min.195(7.75)
144(5.67")
138
138
24.5(1")
138(5.43)
27
(1.06")
80
80
114.5(4.51")
M6
138 M5
(5.43)
M6
Housing dimensions and layout of glands Option/UM. Universal mounting kit, panel mounting
diagram
YOKOGAWA HEADQUARTERS YOKOGAWA CORPORATION OF AMERICA Yokogawa has an extensive sales and
9-32, Nakacho 2-chome, 2 Dart Road
Musashinoshi Newnan GA 30265 distribution network.
Tokyo 180 United States
Japan Tel. (1)-770-253-7000 Please refer to the European web-site
Tel. (81)-422-52-5535 Fax (1)-770-251-2088
Fax (81)-422-55-1202 E-mail: info@yca.com (www.yokogawa.com/eu) to contact your
E-mail: webinfo@mls.yokogawa.co.jp www.yca.com
www.yokogawa.com nearest representative.
GS 12D7B5-E-E
Subject to change without notice Printed in The Netherlands, 01-501 (A) Q