AGA9 Report NaturalGas
AGA9 Report NaturalGas
Traceable to the
NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
Serial Number:
The calibration identified by the above CEESI DATA FILE(s) was performed in accordance with the
American Gas Association TMC Report Number 9 using calibration standards that
are traceable to the National Institute of Standards and Technology. All natural gas properties were
calculated in compliance with American Gas Association TMC Report Number 8 and TMC Report
Number 10.
Calibration Performed at: CEESI Iowa High Flow Natural Gas Facility
2365 240th St.
Garner, Iowa 50438
NOTE:
This Certificate and accompanying documentation shall not be reproduced, except in full, without the written
consent of Colorado Engineering Experiment Station Inc.
54043 County Road 37 | Nunn, Colorado 80648 | +1 (970) 897-2711 | +1 (970) 897-2710 (fax) | CEESI.com
Calibration Overview
Job Number:
Order Number:
Serial Number:
Model Number:
Calibration Date:
Page 2 of 22
Witness List
Serial Number:
Page 3 of 22
Uni-directional Calibration Configuration Details
Serial Number:
Identifier Installed Manufacturer Serial Number Flow Arrow Top Dead Center
C - Upstream Flow Conditioner Yes Verified Verified
Fitting Identification:
Identifier Type
A - Entrance Fitting Tee
G - Exit Fitting Tee
Meter package and fittings shipped to CEESI as a single unit, and shipped to the customer as a single unit
Page 4 of 22
Velocity of Sound Verification
Serial Number:
Component Value
Press Temp Calc VOS Meter VOS VOS Error
Methane 84.9404 [psia] [°F] [ft/sec] [ft/sec] [%]
Nitrogen 1.6861 1168.8 70.38 1289.2 1287.8 -0.109
Carbon Dioxide 0.8659
Ethane 11.5356
Press Temp Calc VOS Meter VOS VOS Error
Propane 0.7232 [bara] [°C] [m/sec] [m/sec] [%]
Hydrogen 0.1873 80.586 21.32 392.9 392.5 -0.109
Iso-Butane 0.0157
Normal-Butane 0.0224
Iso-Pentane 0.0010
Normal-Pentane 0.0010
C6+ 0.0002
Helium 0.0212
Heating Value 1087.443
(BTU/ft³)
Total UnNormalized (%) 99.7064
Page 5 of 22
Calibration Data -
Serial Number:
The calibration results are summarized in the table below. The column labeled 'Calibration Factor' is the value by
which the volume flowrate is multiplied to achieve the corrected value.
Page 6 of 22
Coefficient Data
Serial Number:
Upon completion of the calibration the coefficients listed below were entered into the meter.
1 5123 1 -0.2200
2 11365 2 -0.2000
3 21115 3 -0.2100
4 40862.01 4 -0.2300
5 80514.01 5 -0.0800
6 119345 6 -0.1200
7 158416 7 -0.0900
8 198004 8 -0.0400
9 0 9 0.0000
10 0 10 0.0000
11 0 11 0.0000
12 0 12 0.0000
Table 3: Coefficient Inputs
Page 7 of 22
Calibration Results -
Serial Number:
Page 8 of 22
Calibration Data Table -
Serial Number:
Press Temp Density Lab Flowrate Lab Flowrate Meter Flowrate Percent Error
[psia] [°F] [lbm/ft³] [ACFH] Uncertainty [ACFH] [%]
[%]
1168.8 70.38 4.5770 198233 0.18 198221 -0.005
Page 9 of 22
Calibration Data Table -
Serial Number:
Press Temp Density Lab Flowrate Lab Flowrate Meter Flowrate Percent Error
[psia] [°F] [lbm/ft³] [ACFH] Uncertainty [ACFH] [%]
[%]
1215.1 72.16 4.7425 11404.3 0.23 11380.0 -0.214
Page 10 of 22
Verification Data -
Serial Number:
Page 11 of 22
Verification Results -
Serial Number:
Page 12 of 22
Verification Data Table -
Serial Number:
Press Temp Density Lab Flowrate Lab Flowrate Meter Flowrate Percent Error
[psia] [°F] [lbm/ft³] [ACFH] Uncertainty [ACFH] [%]
[%]
1222.6 72.04 4.7871 20805.4 0.23 20803.9 -0.007
Page 13 of 22
Metric Calibration Data -
Serial Number:
The calibration results are summarized in the table below. The column labeled 'Calibration Factor' is the value by
which the volume flowrate is multiplied to achieve the corrected value.
Page 14 of 22
Coefficient Data
Serial Number:
Upon completion of the calibration the coefficients listed below were entered into the meter.
1 5123 1 -0.2200
2 11365 2 -0.2000
3 21115 3 -0.2100
4 40862.01 4 -0.2300
5 80514.01 5 -0.0800
6 119345 6 -0.1200
7 158416 7 -0.0900
8 198004 8 -0.0400
9 0 9 0.0000
10 0 10 0.0000
11 0 11 0.0000
12 0 12 0.0000
Table 8: Coefficient Inputs
Page 15 of 22
Metric Calibration Results -
Serial Number:
Page 16 of 22
Metric Calibration Data Table -
Serial Number:
Press Temp Density Lab Flowrate Lab Flowrate Meter Flowrate Percent Error
[bara] [°C] [kg/m³] [ACMH] Uncertainty [ACMH] [%]
[%]
80.586 21.32 73.316 5613.33 0.18 5613.02 -0.005
Page 17 of 22
Metric Calibration Data Table -
Serial Number:
Press Temp Density Lab Flowrate Lab Flowrate Meter Flowrate Percent Error
[bara] [°C] [kg/m³] [ACMH] Uncertainty [ACMH] [%]
[%]
83.780 22.31 75.968 322.935 0.23 322.245 -0.214
Page 18 of 22
Metric Verification Data -
Serial Number:
Page 19 of 22
Metric Verification Results -
Serial Number:
Page 20 of 22
Metric Verification Data Table -
Serial Number:
Press Temp Density Lab Flowrate Lab Flowrate Meter Flowrate Percent Error
[bara] [°C] [kg/m³] [ACMH] Uncertainty [ACMH] [%]
[%]
84.296 22.24 76.682 589.142 0.23 589.101 -0.007
Page 21 of 22
Uncertainty Analysis and Traceability
Serial Number:
Uncertainty Analysis:
The uncertainty analysis for the Iowa facility is performed using methods documented in the ISO GUM and NIST
Tech. Note 1297. Each component of uncertainty is classified as Type A or Type B and assigned a statistical
distribution (normal or rectangular). Numerical values for standard uncertainty, degrees of freedom, and
correlation coefficients are estimated. Standard uncertainty values are propagated by addition in quadrature as
described in the ISO GUM. The effective degrees of freedom are estimated using the Welch-Satterthwaite
approximation.
The uncertainty values in the current report are stated with a coverage factor of k=2 which represents a
confidence interval of approximately 95%.
Traceability:
Traceability of pressure, temperature, and gas composition measurements is maintained based on traditional
instrument calibration. Traceability of flowrate measurement is based on periodic engineering analysis of data
from three sources:
1) Control Charts
2) Turbine Substitution Test
3) Direct Turbine Meter Calibrations
Page 22 of 22