Portal Dosimetry application

Graphical User Interface (GUI) overview

VARIAN
ONCOLOGY
SYSTEMS Physics Instructor EMEA
1 | VARIAN ONCOLOGY SYSTEMS
Objective

• Identify Portal Dosimetry GUI components


• Review Image evaluation options
• Identify Image alignment options

2 | VARIAN ONCOLOGY SYSTEMS


Portal Dosimetry Screen Layout
Title Bar
Assistant
Menu Bar
Tool Bar

Navigation
Panel

Image View Panels

Patient Alert & Image


Approval
Visibility
Configuration

Task Panel
Display Image
Mode Navigation

3 | VARIAN ONCOLOGY SYSTEMS


Navigation Panel

• Course & Plan selection


dropdowns
• Field tab
• Session tab
• Hide/Show Navigation panel
Hide

Show

4 | VARIAN ONCOLOGY SYSTEMS


Tools – Mask Patient

Show / Hide Task Panel

• Show / Hide
Task Panel
• Mask Patient

5 | VARIAN ONCOLOGY SYSTEMS


Image View - Customization

• View Layout

• Visibility Configuration
• Display Mode

6 | VARIAN ONCOLOGY SYSTEMS


Visibility Configurations

• Color wash • Graticule

• Isodose Lines • Field Outline

7 | VARIAN ONCOLOGY SYSTEMS


Display Modes

• Predicted and Portal Dose Image


• Portal Dose only
• Predicted Dose only
• Dose Difference
• Gamma Evaluation

8 | VARIAN ONCOLOGY SYSTEMS


Task Panel

• Divided into three tabs


• Evaluation
• Alignment
• Normalization

9 | VARIAN ONCOLOGY SYSTEMS


Evaluation Options - ROI

• ROI settings None


• None – full image matrix User Defined
• User defined – within
rectangle
• Field + margin
• MLC CIAO + margin
• Threshold of max. fluence

Treshold fluence
MLC CIAO +
mrg.

10 | VARIAN ONCOLOGY SYSTEMS


Evaluation Options

• Gamma Evaluation
• Area gamma < 1.0
• Maximum Gamma
• Average Gamma
• LCA Gamma > 1.0
• Area Gamma > gamma < ROI fraction
• Dose Difference
• Maximum Dose Diff.
• Average Dose Diff.
• Area Dose Diff. > diff. < ROI fraction

11 | VARIAN ONCOLOGY SYSTEMS


Gamma Evaluation*
D
• Available in absolute and relative modes
• Quantitative evaluation of dose images r
• Factors both spatial and dose differences
DTA
γ
Dmax

dose
• User-defined pass/fail parameters
• Dose Difference (D)
2
• Distance-to-Agreement (DTA)
• Gamma index
• 0<<1 → within acceptance
Reference Image
• >1 → beyond acceptance 3
Evaluated Image

●<1 ●> position

1
12 | VARIAN ONCOLOGY SYSTEMS
Improved Gamma Algorithm

• Original algorithm implementation searches for the best


gamma value only at integer pixel positions around the
given source pixel
• Results in gamma overestimation

2

impr.

0
1

13 | VARIAN ONCOLOGY SYSTEMS


Gamma Evaluation Tests
• Area gamma < 1
• Maximum gamma
• Average gamma

• Largest Connected area of pixels with gamma value >1


• Area Gamma > X as % of evaluated area
• Area Gamma > Y % of evaluated area

14 | VARIAN ONCOLOGY SYSTEMS


Histogram Cut-Off / Near Max. Dose
Histogram cut-off
• Image matrix processed first 25

• Pixel values sorted into dose bins 20

• Histogram Cut-Off value [%] 15 Reference dose


removes outliers
10
• Determines Reference dose for
• Global Gamma - ∆D [%] 5

• ROI Threshold [%] 0


100% 99% 98% 97% 96% 95% 94% 93% 92%

15 | VARIAN ONCOLOGY SYSTEMS


Evaluation Options

• Point Dose • Profiles

Right-mouse-click
Right-mouse-click

16 | VARIAN ONCOLOGY SYSTEMS


Evaluation Options
• Leaf Motion Overlay • Show Field outline

Field Outline

RMC menu

17 | VARIAN ONCOLOGY SYSTEMS


Evaluation Options

• Create Composite Image

18 | VARIAN ONCOLOGY SYSTEMS


Evaluation Options

• Compare to

19 | VARIAN ONCOLOGY SYSTEMS


Evaluation Options
• Set as Constancy Check Baseline
• Perform Constancy Check

Baseline Image constancy


check

RMC menu

20 | VARIAN ONCOLOGY SYSTEMS


Evaluation Options

• Gamma/Dose difference
Height Field

21 | VARIAN ONCOLOGY SYSTEMS


Alignment Options

• Auto alignment
• Manual alignment
• Alignment to
• Reference Dose
• Reference Image
• Copy correction to selected
fields
• Same gantry angle

22 | VARIAN ONCOLOGY SYSTEMS


Normalization Options

• Absolute Mode • Relative Mode


• Max. of Predicted Dose
• Max of each dose
• Selected Dose in point
• Dose at Isocenter
• Minimize Difference
• Factor Modification

23 | VARIAN ONCOLOGY SYSTEMS


Tools – Output Factors

• Easy acquisition of
Output Factor
measurements for
PDIP configuration.

24 | VARIAN ONCOLOGY SYSTEMS


Thank you for your interest
and attention

• Questions?

VARIAN
ONCOLOGY
SYSTEMS Physics Instructor EMEA
25 | VARIAN ONCOLOGY SYSTEMS
Disclaimer
Notice

• This information is an educational aid only and is intended solely as a


supplemental document. This provides information concerning the use of the
Eclipse Treatment Planning System. For complete information concerning
components, safety instructions, installation, maintenance, troubleshooting,
etc., refer to the applicable Varian product Operator’s Manual. This
information does not replace the Varian Operator’s Manual.
• Carefully read all instructions prior to use. Observe all contraindications,
warnings and precautions noted in these instructions. Failure to do so may
result in patient and/or user complications.
• This information is subject to change without notice and does not represent a
commitment on the part of Varian. Varian is not liable for errors contained in
this information or for incidental or consequential damages in connection with
furnishing or use of this material. This contains proprietary information
protected by copyright. No part of this may be reproduced, translated, or
transmitted without the express written permission of Varian Medical Systems,
Inc.