VARIAN
ONCOLOGY
SYSTEMS Physics Instructor EMEA
1 | VARIAN ONCOLOGY SYSTEMS
Objective
Navigation
Panel
Task Panel
Display Image
Mode Navigation
Show
• Show / Hide
Task Panel
• Mask Patient
• View Layout
• Visibility Configuration
• Display Mode
Treshold fluence
MLC CIAO +
mrg.
• Gamma Evaluation
• Area gamma < 1.0
• Maximum Gamma
• Average Gamma
• LCA Gamma > 1.0
• Area Gamma > gamma < ROI fraction
• Dose Difference
• Maximum Dose Diff.
• Average Dose Diff.
• Area Dose Diff. > diff. < ROI fraction
dose
• User-defined pass/fail parameters
• Dose Difference (D)
2
• Distance-to-Agreement (DTA)
• Gamma index
• 0<<1 → within acceptance
Reference Image
• >1 → beyond acceptance 3
Evaluated Image
1
12 | VARIAN ONCOLOGY SYSTEMS
Improved Gamma Algorithm
2
impr.
0
1
Right-mouse-click
Right-mouse-click
Field Outline
RMC menu
• Compare to
RMC menu
• Gamma/Dose difference
Height Field
• Auto alignment
• Manual alignment
• Alignment to
• Reference Dose
• Reference Image
• Copy correction to selected
fields
• Same gantry angle
• Easy acquisition of
Output Factor
measurements for
PDIP configuration.
• Questions?
VARIAN
ONCOLOGY
SYSTEMS Physics Instructor EMEA
25 | VARIAN ONCOLOGY SYSTEMS