Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Evaluation of defects in surface detail for

monophase, 2-phase, and 3-phase


impression techniques: An in vitro study
Giuseppe Varvara, DDS, PhD,a
Giovanna Murmura, MD, DMD, PhD,b Bruna Sinjari, DDS,c
Paolo Cardelli, DDS,d and Sergio Caputi, MD, DMDe
Dental School, ‘G. D’Annunzio’ University of Chieti-Pescara, Chieti-
Pescara, Italy
Statement of the problem. Polyvinyl siloxane materials of various consistencies have been tested to reduce defects commonly
associated with different definitive impression techniques.

Purpose. The purpose of this in vitro study was to compare the number of visible defects within the surface of a new 3-phase,
2-step impression injection technique with the number found in the monophase and in the conventional 2-phase, 1-step and
2-phase, 2-step impression techniques.

Material and methods. Two complete crown abutment preparations were impressed 10 times for each of these 4 techniques,
which gave 20 abutment impressions for each group. The 3-phase, 2-step impression–injection technique included extra-
light–body impression material added for the impression and injected after reinsertion. After the removal of all of these
impressions, an examiner counted the number of open voids and bubble-like enclosed voids visible to the naked eye at a
working distance of approximately 150 mm. Only the defects in the area of the prepared abutments were included in the
assessment.

Results. The frequency of defects ranged from 100% of the impressions in the monophase group to 5% with the new 3-phase,
2-step impression injection technique. No statistical differences were seen between the two 2-phase (1-step and 2-step)
impression techniques, although there were numerically fewer impressions with defects with 2 steps (45%) than with 1 step
(55%). The 3-phase, 2-step impression injection technique had the greatest accuracy, with fewer specimens with defects (5%)
than with either of the 2-phase techniques, although these differences did not reach statistical significance.

Conclusions. The 3-phase, 2-step impression injection technique provides improved defect-free reproduction of detail,
showing fewer defects than other impression techniques. (J Prosthet Dent 2014;-:---)

Clinical Implications
In fixed prosthodontics, a lower incidence of impression surface defects
can be obtained with the use of the 3-phase, 2-step impression injection
technique. This technique improves the overall accuracy of the
subsequent restoration, with improved performance and without
additional steps compared with the conventional 2-phase, 2-step
impression technique.

Impression materials are used to function of prosthetic restorations.1 The accurate cast of the prepared tooth and
provide an accurate record of hard and goal of an impression is to provide a the surrounding tissue when filled with
soft tissues. This step is the main deter- void-free negative representation of a the gypsum product.2 Several studies
minant of the definitive fit, form, and prepared tooth, which will produce an have evaluated the quality of the

a
Assistant Professor, Department of Medical, Oral and Biotechnological Sciences.
b
Aggregate Professor, Department of Medical, Oral and Biotechnological Sciences.
c
Research Fellow, Department of Medical, Oral and Biotechnological Sciences.
d
Research Fellow, Department of Medical, Oral and Biotechnological Sciences.
e
Professor and Head of Department of Medical, Oral and Biotechnological Sciences.

Varvara et al
2 Volume - Issue -

reproduction of the surface detail of impression material in the tray. With impressions. The null hypothesis was
partial fixed dental prosthesis impres- the techniques that use 2-phase mate- that no differences would be found in the
sions sent to dental laboratories. In a rials, such as the putty–light-body reproduction of defect-free surface
clinical study, Samet et al3 reported that impression material techniques (both detail of the impressions with these
more than 89% of the impressions the 1-step and 2-step approaches), the various techniques.
investigated had 1 or more observable light-body impression material should
errors; thus, they recommended that cover the entire preparation, although MATERIAL AND METHODS
dentists evaluate impressions more crit- this cannot always be accomplished
ically. Raigrodski et al4 reported that clinically.8,10 The 2-phase, 2-step To evaluate the reproduction of
92% to 96% of the control impression impression technique has been re- defect-free surface detail, stainless-steel
preparation areas that were analyzed by ported to be more accurate than the 2- abutments were impressed with different
a clinician and a dental technician had phase, 1-step impression technique.5,14 materials and techniques by using stock
small defects, such as tears, voids, and The 2-phase, 2-step impression tech- metal trays, and the numbers of open
bubbles. nique allows mostly complete coverage voids and enclosed voids were deter-
Various factors can influence the of the preparation area with the light- mined. A stainless-steel die (the master
quality and accuracy of impressions, body impression material; however, it cast) that contained 2 complete-crown,
including the impression technique5-9 can be associated with the creation of tapered-abutment preparations was
and the impression material.4,10 Re- an occlusal step because some of the fabricated with a lathe according to the
sults of some studies have indicated light-body impression material can American National Standards Institute–
that, as impression materials have spread along the occlusal surface dur- American Dental Association specifica-
improved, the quality of the reproduc- ing the reseating of the putty.11 tions (no. 19; 8.015 mm in height,
tion of the surface detail is influenced When considering the problems that 6.330 mm and 8.450 mm at the base,
more by the technique that is used can be associated with these techniques, and 28.20 mm as the distance between
instead of by the material itself. Howev- although the 2-phase, 2-step impression the centers of the abutments) (Fig. 1).
er, results of other studies have indicated technique has a number of advantages, The device was used as the standardized
that the impression technique does not further improvements in the reproduction master cast to compare the defects
affect the quality of the reproduction of of dimensional accuracy and defect-free associated with each of the impression
the surface detail of impressions.11,12 surface detail can be achieved with a techniques in this study (Fig. 1).
Thus, in spite of the number of previ- novel 3-phase, 2-step impression injec- This master cast was manually
ous studies on the quality and accuracy tion technique. The principles of this 2- incorporated into an autopolymerizing
of impressions as related to impression step impression injection technique were acrylic resin device (Lang Dental) and
materials and/or impression techniques, described in an in vitro study.13 In the fixed onto a base (Zeiser) (Fig. 2). The
controversies still remain. The types of present study, the initial polymerization autopolymerizing acrylic resin device
impression techniques and the different of the putty and the light-body impres- then was modeled to allow reproduc-
protocols used to assess the quality and sion material was as for the traditional 1- ible positioning of the tray on the
accuracy of impressions might explain step impression technique, with this also master cast (Fig. 2). All of the impres-
the contradictory results that have been followed by injection of extra-light-body sions were made in stock perforated
reported in the literature. impression material into the preparation metal trays (size 6; ASA Dental). The
Impression techniques have been through a hole in the metal stock tray. study design required 20 experimental
categorized according to the mono- This new 2-step impression injection impressions and 20 control impressions
phase, 1-step technique, the 2-phase technique allows the displacement of soft for each group. Preexisting literature
(putty–light-body impression material) tissues, for example, the tongue, during data indicated that the probability
techniques that can be accomplished in the first seating with the putty–light-body of impression defects in monophase
1 or 2 steps (as the 1-step and 2-step impression material, whereas, in the sec- impressions is generally 100%. If the
impression techniques), and the 3- ond step, the extra-light-body impression true probability of exposure among the
phase (putty–light-body impression material can record all of the finer details experimental groups is 50%, with this
material plus extra-light-body impres- without being compressed. study design, then the null hypothesis
sion material) 2-step technique (as the The aim of the present in vitro study that the exposure rates for experimental
2-step impression injection tech- was to evaluate the reproduction of the impressions and control impressions
nique).13 The techniques that use surface detail of this new 2-step impres- are equal can be rejected with a power
monophase materials are accomplished sion injection technique compared with of more than 80%. Therefore, the im-
in a single-step procedure that uses an the monophase and conventional 1-step pressions of the master cast were
impression material with a medium and 2-step impression techniques by made 10 times for each of the 4 tech-
viscosity to record the finer details while analyzing the number of visible defects niques: monophase, 2-phase, 1-step
avoiding the slumping of the on the surface of the resulting and 2-step impressions, and 3-phase,
The Journal of Prosthetic Dentistry Varvara et al
- 2014 3

1 Stainless-steel die used as standardized master cast. 2 Master cast in repositioning device.

2-step impression injection. Thus, 20 prefabricated 2-mm-thick acrylic resin master cast for 12 minutes. The pre-
abutment impressions were obtained copings (Duralay; Dental Mfg Co) that liminary impressions then were removed
for each group. were placed on each abutment to create from the master cast, and a hole was
The impressions were made by using a uniform and optimal space for the made in the vicinity of the most occlusal
addition-reaction polyvinyl siloxane light-body impression material. The surface (edge) of each abutment by
impression materials (Aquasil; Dentsply acrylic resin copings were made by drilling through the polymerized mate-
Intl), as detailed in Table I. The soft putty removing material with a carbide bur rial with a carbide bur (187 023; Komet).
material was mixed with the fingertips (187 023; Komet) until a uniform thick- This was designed to coincide with 1 of
for 30 seconds until the color was uni- ness of 2 mm was obtained. The di- the holes in the stock tray. The selected
form, and all of the other materials were mensions were standardized by using hole on the stock tray of each abutment
dispensed by using an automatic mixing calipers (no. 500-181-21; Roder Elec- was labeled with ink to facilitate its
device (AutoMix; Dentsply Intl). The tronics). The preliminary putty impres- identification during the following pro-
polymerization times used for each sions were made first, and, before the cedures. A thin section of interdental
impression material were double those second step, the acrylic resin copings material also was removed from the
recommended by the manufacturer to were removed and the light-body impression with the same carbide bur.
compensate for the impressions being impression material was added. The Extra-light-body impression material
made at room temperature (20 C) impressions were reinserted until firm then was added to the preliminary
instead of at mouth temperature. Thus, contact was made with the border of the impression, which was immediately
the setting time was increased to 12 tray, and the light-body impression ma- reinserted onto the master cast. Further
minutes. No tray adhesive was used. The terial again was allowed to polymerize extra-light-body impression material
monophase impressions were made with on the master cast for 12 minutes. The 2- was placed into a syringe (Elastomer;
the standard impression material, the 1- step impression injection technique ESPE Dental AG) and injected through
step impressions were made with the initially used putty and light-body the holes marked on the abutments. The
putty and light-body impression mate- impression material simultaneously (as trays were held down during this proce-
rial simultaneously, and the 2-step in the 1-step technique), and again, dure to prevent them from lifting off the
impressions were made by using these were left to polymerize on the teeth. These impressions were again
allowed to polymerize for 12 minutes.
After removal, the impression sur-
Table I. Technical characteristics of addition reaction polyvinyl siloxane
faces of the abutments were observed
impression materials used
by an examiner trained in a standard-
Materiala Lot No. ISO 4823 Type Viscosity ized technique for the recognition and
classification of surface defects. This
Soft putty 0405000002, 2007.06 1 Very high examiner counted the number of open
Regular body 0408001734, 2007.11 3 Low voids (approximately 2 to 4 mm) and
Light body 0408001289, 2007.9 3 Low “bubble-like” enclosed voids (<2 mm)
Extra-light body 0405000010, 2007.07 3 Low visible to the naked eye at a working
distance of approximately 150 mm for
ISO, International Organization for Standardization (http://www.iso.org/iso/home/store/catalogue_
tc/catalogue_detail.htm?csnumber=24125).
all of the surfaces of the abutments,
a
Materials were supplied by Dentsply Intl. and each impression was

Varvara et al
4 Volume - Issue -

3 Impression without open voids or 4 Impression that shows bubble-like 5 Impression showing open voids
bubble-like enclosed voids (specimen enclosed voids (specimen from 2-step (specimen from monophase impres-
from 2-step impression injection impression technique). sion technique).
technique).

photographed with a digital camera worst results in comparison with the defect frequency of 4 different
(Fine Pix S2pro [Fuji]; AF Micro Nikkor, the other procedures, whereas the impression techniques was investigated,
105 mm, 1:2.8D [Nikon]). Only the 3-phase, 2-step impression injection with significant differences seen be-
defects in the impression in the area of technique had the best results. No sta- tween these different approaches. The
the prepared abutments were counted. tistically significant difference was seen monophase technique is considered to
The number of defects on each spec- between the 2 dual-phase (1 step versus be the easiest to perform, although
imen were ranked as follows: type 0, no 2 step) techniques (Table II). The com- many in vitro studies reported this to be
defects (Fig. 3); type 1, 1 or 2 enclosed parison of the defect distribution for the worst technique in terms of
voids (Fig. 4); type 2, >2 enclosed each of the 4 techniques gave similar dimensional accuracy13 and surface
voids; and type 3, the presence of open results: the monophasic technique gave defects14 of the impressions thus
voids (Fig. 5). The differences in the the worst results, whereas the 3-phase, formed. These in vitro studies also have
defect frequencies were investigated 2-step impression injection technique been supported in a clinical study.16
according to c2 tests. An extension of gave the best results (Table II). The present investigation also has
c2 tests according to Bonferroni was confirmed these data, with the mono-
used in a post hoc analysis. The me- DISCUSSION phase technique giving the highest
dians were obtained for each experi- number of voids among the 4 experi-
mental group, and these parameters Polyvinyl siloxane has been reported mental groups. The monophase tech-
were analyzed for statistical significance to be the most accurate impression nique also resulted in statistically
in the comparisons of the 4 impression material for dental impressions because significantly greater numbers of open
techniques with the Kruskal-Wallis tests of its surface reproduction and dimen- voids and enclosed voids, especially in
and Mann-Whitney tests after Bonfer- sional stability.15 In the present study, the finish line compared with the other
roni correction (a¼.05).

RESULTS Table II. Numbers and types of defects according to each impression technique
Defect
The frequency of defects ranged from Type
100% in the monophase group to 5% in
Impression No. Specimens Defect (No.)*
the 3-phase, 2-step impression injection
technique. A statistically significant dif- Technique With Defects Frequency (%) 1 2 3 Mediany
ference among the tested groups was
detected (P<.001). Thus, the data ob- Monophase 20 100a 0 0 20 3d
tained in this study supported the rejec- Two-phase, 1-step 11 55b 8 3 0 1e
tion of the null hypothesis. The presence Two-phase, 2-step 9 45b 7 2 0 0e,f
of the defects and the numbers and types Three-phase, 2-step 1 5 c
1 1 0 0f
of defects according to the 4 impression
Values with same superscript letters are not statistically different according to c2 tests with Bonferroni
techniques are listed in Table II. The
post hoc analysis (a-d), and Bonferroni-corrected Kruskal-Wallis tests and Mann-Whitney tests (d-f).
monophasic technique produced the *Type 1 (1-2 enclosed voids), type 2 (>2 enclosed voids), type 3 (presence of open voids).
y
Number of defect ranking.

The Journal of Prosthetic Dentistry Varvara et al


- 2014 5
procedures. This is due to the higher these 2-phase, 1-step and 2-step impres- 5. Hung SH, Purk JH, Tira DE, Eick JD. Accuracy
of one-step versus two-step putty wash
viscosity and lower flow of this mono- sion techniques.
addition silicone impression technique.
phase impression material. Indeed, in The best outcome in the present study J Prosthet Dent 1992;67:583-9.
the present study, the monophase was instead obtained with the 3-phase, 6. Chen SY, Liang WM, Chen FN. Factors
technique had defects in 100% of the 2-step impression injection technique. affecting the accuracy of elastomeric
impression materials. J Dent 2004;32:603-9.
specimens. Here, only 5% of the specimens analyzed 7. Nissan J, Laufer BZ, Brosh T, Assif D. Accu-
The classic 2-phase, 2-step impres- had defects due to the use of the extra- racy of three polyvinyl siloxane putty-wash
sion technique has been reported to be light-body impression material. This is impression techniques. J Prosthet Dent
2000;83:161-5.
more accurate than the 2-phase, 1-step an extremely fluid material, which pro- 8. Morgano SM, Milot P, Ducharme P, Rose L.
impression technique8,17; indeed, in the vides good reproduction of the prepara- Ability of various impression materials to
2-phase, 2-step impression technique, tion finish line and the subgingival produce duplicate dies from successive im-
pressions. J Prosthet Dent 1995;73:333-40.
after the putty has polymerized, further margins. Furthermore, this extra-light-
9. Craig RG. Review of dental impression ma-
contraction of the light-body impression body impression material has a low sur- terials. Adv Dent Res 1988;2:51-64.
material leads to only minimal dimen- face tension, which also improves the 10. Chee WW, Donovan TE. Fine detail repro-
sional changes.7 The 2-phase, 1-step “wettability” of the prepared teeth. The duction of very high viscosity poly(vinyl
siloxane) impression materials. Int J Pros-
impression technique, however, has application of this extra-light-body im- thodont 1989;2:368-70.
been criticized because of the uncon- pression material through the hole in the 11. Idris B, Houston F, Claffey N. Comparison of
trolled bulk of the light-body impression metal stock tray also avoids its compres- the dimensional accuracy of one- and two-
step techniques with the use of putty/wash
material, which can result in dimensional sion and dislocation, which are the main addition silicone impression materials.
changes that are proportional to the disadvantages of the other 2-phase tech- J Prosthet Dent 1995;74:535-41.
thickness of the light-body impression niques. Therefore, in this way, the whole 12. Lee IK, DeLong R, Pintado MR, Malik R.
Evaluation of factors affecting the accuracy
material during setting.18 The putty in the surface of the prepared tooth can be
of impressions using quantitative surface
2-phase, 1-step impression technique covered by the extra-light-body impression analysis. Oper Dent 1995;20:246-52.
also tends to push the light-body material. Analysis of the results of this 13. Caputi S, Varvara G. Dimensional accuracy
impression material off the prepared study indicates that improved defect-free of resultant casts made by a monophase,
one-step and two-step, and a novel two-step
tooth, and critical areas, such as the finish reproduction of detail can be attributed putty/light-body impression technique: an in-
line, can be covered by the putty, which to the 3-phase, 2-step impression injection vitro study. J Prosthet Dent 2008;99:274-81.
cannot record details satisfactorily.10 A technique versus the other impression 14. Millar BJ, Dunne SM, Robinson PB. In-vitro
study of the number of surface defects in
further disadvantage has been reported techniques examined. monophase and two-phase addition silicone
for the 2-phase, 2-step impression tech- impressions. J Prosthet Dent 1998;80:32-5.
nique, whereby some light-body impres- CONCLUSION 15. Ali KS, Shenoy VK, Rodrigues SJ. Compara-
tive evaluation of dimensional accuracy of
sion material can spread along the
casts made by repeated pouring of addition
occlusal surface during the reseating of The 3-phase, 2-step impression in-
silicone impressions using 1) 2-step putty/
the putty, or it can displace the pre- jection technique shows fewer defects in light-body technique using stock tray, and
liminary putty impression during the the reproduction of detail, such as 2) 1-step simultaneous dual viscosity tech-
open voids and enclosed, bubble-like, nique using custom tray: an in-vitro study.
seating; this can generate distortions that J Nepal Dent Assoc 2010;11:32-9.
reduce dimensional accuracy.11 voids than the monophase and con- 16. Luthardt RG, Walter MH, Quaas S, Koch R,
The data from the present study ventional 2-phase, 1-step and 2-step Rudolph H. Comparison of the three-
impression techniques investigated. dimensional correctness of impression
confirmed the improved results of the 2- techniques: a randomized controlled trial.
phase, 2-step impression technique Quintessence Int 2010;41:845-53.
REFERENCES
compared with those with a 1-step pro- 17. Franco EB, da Cunha LF, Herrera FS,
Benetti AR. Accuracy of single-step versus 2-
cedure. For the 2-phase techniques (both 1. Donovan TE, Chee WW. A review of
contemporary impression materials and step double-mix impression technique. ISRN
1-step and 2-step) and in agreement with Dent 2011;2011:341546.
techniques. Dent Clin North Am 2004;48:
the literature, the present study shows a 445-70. 18. Nissan J, Gross M, Shifman A, Assif D. Effect
lower incidence of open voids and 2. Guiraldo RD, Borsato TT, Berger SB, of wash bulk on the accuracy of polyvinyl
Lopes MB, Gonini-Jr A, Sinhoreti MA. Sur- siloxane putty-wash impressions. J Oral
bubble-like enclosed voids; this result is Rehabil 2002;29:357-61.
face detail reproduction and dimensional
linked to the pressure applied by the accuracy of stone models: influence of
major viscosity impression material (the disinfectant solutions and alginate impres- Corresponding author:
putty) on the minor viscosity impression sion materials. Braz Dent J 2012;23:417-21. Dr Giuseppe Varvara
3. Samet N, Shohat M, Livny A, Weiss EI. University ‘G. D’Annunzio’ of ChietiePescara
material (the light-body impression ma- A clinical evaluation of fixed partial denture Via dei Vestini, 31
terial), which increases its flow and aids in impressions. J Prosthet Dent 2005;94:112-7. 66100 Chieti
the more-exact reproduction of the de- 4. Raigrodski AJ, Dogan S, Mancl LA, Heindl H. ITALY
A clinical comparison of two vinyl poly- E-mail: gvarvara@unich.it
tails. In the present study, however, in
siloxane impression materials using the one-
terms of the number of defects, no sig- step technique. J Prosthet Dent 2009;102: Copyright ª 2014 by the Editorial Council for
nificant differences were found between 179-86. The Journal of Prosthetic Dentistry.

Varvara et al

You might also like