1. Summary
Please perform insertion and extraction of connecter for the Optical Unit on the
back side more than 5 times in case of ECC71 (HSYNC detection timeout) or
EEC73 (Parity error) occurrence. These errors may be caused by a poor
contacting of the LD- IF board connecter for the Optical unit because of flux
attached upon soldering.
2. Procedure
Relay box cover
1/2
Internal use only
4. Working Time
0.5 H
Submitted by M.Oguri
Approved by M.Sato