Initial Interrogation: Quick Look II

Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:
History: NYHA Class II
Device Status (Implanted: 12-Oct-2020)
Remaining Longevity 4.2 years (16-Nov-2020)

RRT > 5 years


(based on initial interrogation)

Atrial(5076) RV(6935M)
Pacing Impedance 456 ohms 494 ohms
Defibrillation Impedance RV=65 ohms
Programmed Amplitude/Pulse Width 5.00 V / 0.60 ms 3.50 V / 0.40 ms
Measured P/ R Wave 0.5 mV 4.8 mV
Programmed Sensitivity 0.45 mV 0.60 mV
Parameter Summary
Mode DDDR Lower Rate 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Detection Rates Therapies


AT/AF Monitor >171 bpm All Rx Off
VF On >200 bpm ATP During Charging, 35J x 6
FVT OFF All Rx Off
VT On 167-200 bpm Burst(3), 20J, 35J x 4
Enhancements On: VT Monitor, AF/Afl, Sinus Tach, Wavelet, TWave, Noise(Timeout)
Clinical Status Since 14-Oct-2020 Cardiac Compass Trends (Oct-2020 to Nov-2020)
Treated
P I
VF 0
FVT (Off) Treated >5
VT/VF 4
VT 0 3
AT/AF(Monitor) (#/day) 2
1
0
Monitored
VT(136-167 bpm) 0 AT/AF 60
VT-NS (>4 beats, >167 bpm) 6 (min/day) 50
40
High Rate-NS 0 30
SVT: VT/VF Rx Withheld 0 20
10
V. Oversensing-TWave Rx Withheld 0
0
V. Oversensing-Noise Rx Withheld 0
AT/AF 0 Patient >8
Activity 6
Time in AT/AF 0.0 hr/day (0.0%) (hr/day)
4

Functional Last Week 2


Patient Activity 4.4 hr/day 0
Nov-20 Jan-21 Mar-21 May-21 Jul-21 Sep-21 Nov-21

Therapy Summary VT/VF AT/AF Pacing (% of Time Since 14-Oct-2020)


Pace-Terminated Episodes 0 0 Total VP 97.8 % (MVP Off)
Shock-Terminated Episodes 0 0 AS-VS <0.1 %
Total Shocks 0 0 AS-VP 1.5 %
Aborted Charges 0 0 AP-VS 0.2 %
AP-VP 98.3 %

OBSERVATIONS (1)
· No Wavelet template. Auto Collection is On so a template will be collected; or collect one manually. (In Tests/Wavelet)

SW033 Software Version 8.3 (4.1) Initial Interrogation: Quick Look II


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:09:17
Initial Interrogation: Cardiac Compass Trends
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Nov 2020 Jan 2021 Mar 2021 May 2021 Jul 2021 Sep 2021 Nov 2021

P = Program
I = Interrogate
P I
_ = Remote

One or more A
shocks/day
V

Treated VT/VF >5


episodes/day 4
3
2
1
0

V. rate during VT/VF >250


(bpm)
200
VF
VT 150
<100

Non-sustained VT >10
episodes/day 8
6
4
2
0

AT/AF total minutes/day 60


50
40
30
20
10
0

V. rate during AT/AF >200


(bpm)
max/day 150
avg/day
100

<50

% Pacing/day 100
Atrial 75
Ventricular 50
25
0

Avg V. rate (bpm) >120


Day 100
Night 80
60
<40

Patient activity >8


hours/day 6
4
2
0
Nov 2020 Jan 2021 Mar 2021 May 2021 Jul 2021 Sep 2021 Nov 2021

SW033 Software Version 8.3 (4.1) Initial Interrogation: Cardiac Compass Trends
Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:09:21
Initial Interrogation: Cardiac Compass Trends
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

P I

Heart rate variability >200


(ms) 160
120
80
<40
Nov 2020 Jan 2021 Mar 2021 May 2021 Jul 2021 Sep 2021 Nov 2021

SW033 Software Version 8.3 (4.1) Initial Interrogation: Cardiac Compass Trends
Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:09:21
Initial Interrogation: Rate Histograms
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Prior to Last Session Since Last Session


12-Oct-2020 to 14-Oct-2020 14-Oct-2020 to 16-Nov-2020
45 hours 33 days
% of Time Total VP 97.7 % 97.8 %
AS-VS < 0.1% < 0.1%
AS-VP < 0.1% 1.5 %
AP-VS 0.1 % 0.2 %
AP-VP 99.8 % 98.3 %

Atrial 2% to 5% of AS may be due to FFRW


% of Time 100 100
80 80
AS 60 60
AP
40 40
20 20
0 0
< 40 60 80 100 120 140 160 180 200 220> < 40 60 80 100 120 140 160 180 200 220>
Atrial Rate (bpm) Atrial Rate (bpm)

Ventricular
% of Time 100 100
80 80
VS
60 60
VP
40 40
20 20
0 0
< 40 60 80 100 120 140 160 180 200 220> < 40 60 80 100 120 140 160 180 200 220>
Ventricular Rate (bpm) Ventricular Rate (bpm)

SW033 Software Version 8.3 (4.1) Initial Interrogation: Rate Histograms


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:09:24
Non-sustained VT Episode #4
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 4 31-Oct-2020 16:25 :03 <30/163 Rest

V-V A-A VF = 300 ms VT = 360 ms


Term.
Interval (ms)
1500
1200
900
600

400

200

-60 -55 -50 -45 -40 -35 -30 -25 -20 -15 -10 -5 0
Time (sec)

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #4


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:10:38
Non-sustained VT Episode #4
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #4 Chart speed: 25.0 mm/sec

EGM1: Atip to Aring


(1 mV)

EGM3: RVtip to RVring


(1 mV)

A-A (ms)
1150 1150 1150 1150 1150
A A A A
P P P P
Markers
V V V V V
P P P P P
V-V (ms) 1150 1150 1150 1150

1150 1150 1150 1150


A A A A
P P P P

V V V V V V V V T T
P P P P S S S S S S
1150 1150 1150 1150 9 4 4 3 3 3 3
0 6 0 7 4 4 6
0 0 0 0 0 0 0

>2000 1160 1160 1160 1160


A A A A
P P P P

V V V V T V V V V
S S S S S P P P P
3 4 3 3 3 1160 1160 1160 1160 1160
6 0 7 6 5
0 0 0 0 0
Termination

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:10:43 Page 1
Non-sustained VT Episode #4
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #4 Chart speed: 25.0 mm/sec

1160 1160 1160 1160 1160 1160


A A A A A A A
P P P P P P P

V V V V V V
P P P P P P
1160 1160 1160 1160 1160 1160

1160 1160 1160 1700 1160


A A A A A
P P P P P

V V V V V V
P P P P S P
1160 1160 1160 5 1160
4
0

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:10:44 Page 2
Non-sustained VT Episode #4
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Episode #4: 31-Oct-2020 16:25:29

Episode Summary
Initial Type VT Non-sustained (spontaneous)
Duration 3 sec (6 beats)
V. Median 162 bpm (370 ms)
Activity at onset Rest, Sensor = 58 bpm

Parameter Settings Initial Redetect V. Interval (Rate)


VF On 30/40 24/32 300 ms (200 bpm)
FVT Off
VT On 36 32 360 ms (167 bpm)
Monitor Monitor 40 440 ms (136 bpm)

Polarity RV
Pace Polarity Bipolar
Sense Polarity Bipolar

EGM Source Range Sensitivity


EGM1 Atip to Aring +/- 8 mV Atrial 0.45 mV
EGM3 RVtip to RVring +/- 8 mV RV 0.6 mV

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #4


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:10:58
Non-sustained VT Episode #6
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 6 11-Nov-2020 00:56 :01 30/222 Active

V-V A-A VF = 300 ms VT = 360 ms


Term.
Interval (ms)
1500
1200
900
600

400

200

-40 -35 -30 -25 -20 -15 -10 -5 0


Time (sec)

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #6


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:11:20
Non-sustained VT Episode #6
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #6 Chart speed: 25.0 mm/sec

EGM1: Atip to Aring


(1 mV)

EGM3: RVtip to RVring


(1 mV)

6 6 6 5 4 3 5 5 4 6
A-A (ms) 0 0 0 7 3 0 9 0 9 1
0 0 0 0 0 0 0 0 0 0
A A A A A A A A A
P P P S R P P S S
Markers
V V V V V V V V V
P P S P P P P P P
V-V (ms) 6 5 6 5 6 7 4 4 6
0 6 4 4 3 3 7 9 4
0 0 0 0 0 0 0 0 0

6 9 6 6 6 6 6 6
1 6 1 1 1 1 1 1
0 2000 0 0 0 0 0 0 0
A A A A A A A A A
P P P P P P P P P

V V F F F F F V T V V V V V V
P P S S S S S P S P P P P P P
6 2 2 2 2 2 6 3 6 6 6 6 6 6
1 9 8 7 7 7 1 5 1 1 1 1 1 1
0 0 0 0 0 0 0 0 0 0 0 0 0 0
Termination

6 6 6 6 6 6 6 6 6 6 6 6
1 1 1 1 1 1 1 2 2 2 2 2
0 0 0 0 0 0 0 0 0 0 0 0
A A A A A A A A A A A A
P P P P P P P P P P P P

V V V V V V V V V V V V
P P P P P P P P P P P P
6 6 6 6 6 6 6 6 6 6 6 6
1 1 1 1 1 1 1 1 2 2 2 2
0 0 0 0 0 0 0 0 0 0 0 0

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:11:29 Page 1
Non-sustained VT Episode #6
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #6 Chart speed: 25.0 mm/sec

6 6 6 6 6 6 6 6 6 6 6 6
2 2 2 2 2 2 2 2 2 2 2 2
0 0 0 0 0 0 0 0 0 0 0 0
A A A A A A A A A A A A
P P P P P P P P P P P P

V V V V V V V V V V V
P P P P P P P P P P P
6 6 6 6 6 6 6 6 6 6 6 6
2 2 2 2 2 2 2 2 2 2 2 2
0 0 0 0 0 0 0 0 0 0 0 0

V
P

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:11:31 Page 2
Non-sustained VT Episode #6
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Episode #6: 11-Nov-2020 00:56:35

Episode Summary
Initial Type VT Non-sustained (spontaneous)
Duration 1 sec (5 beats)
V. Median 122 bpm (490 ms)
Activity at onset Active, Sensor = 105 bpm

Parameter Settings Initial Redetect V. Interval (Rate)


VF On 30/40 24/32 300 ms (200 bpm)
FVT Off
VT On 36 32 360 ms (167 bpm)
Monitor Monitor 40 440 ms (136 bpm)

Polarity RV
Pace Polarity Bipolar
Sense Polarity Bipolar

EGM Source Range Sensitivity


EGM1 Atip to Aring +/- 8 mV Atrial 0.45 mV
EGM3 RVtip to RVring +/- 8 mV RV 0.6 mV

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #6


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:11:48
Non-sustained VT Episode #5
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 5 04-Nov-2020 04:35 :01 <30/200 Rest

V-V A-A VF = 300 ms VT = 360 ms


Term.
Interval (ms)
1500
1200
900
600

400

200

-60 -55 -50 -45 -40 -35 -30 -25 -20 -15 -10 -5 0
Time (sec)

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #5


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:12:01
Non-sustained VT Episode #5
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #5 Chart speed: 25.0 mm/sec

EGM1: Atip to Aring


(1 mV)

EGM3: RVtip to RVring


(1 mV)

A-A (ms)
1200 1200 1200 1200
A A A A
P P P P
Markers
V V V V V
P P P P P
V-V (ms) 1200 1200 1200 1200

1200 1200 1200 1200 1200


A A A A A
P P P P P

V V V V V V V T
P P P P S S S S
1200 1200 1200 1200 1160 4 3 3 3
9 9 4 1
0 0 0 0

>2000 1200 1200 1200 1200 1200


A A A A A
P P P P P

T T T V V V V V
S S S P P P P P
3 3 1200 1200 1200 1200 1200
0 0
0 0
Termination

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:12:13 Page 1
Non-sustained VT Episode #5
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #5 Chart speed: 25.0 mm/sec

1200 1200 1200 1200 1200 1200


A A A A A A
P P P P P P

V V V V V V
P P P P P P
1200 1200 1200 1200 1200 1200

1200 1200 1870 1200 1200 1200


A A A A A A
P P P P P P

V V V V V V
P P S P P P
1200 1200 6 1200 1200 1200
7
0

A
P

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:12:14 Page 2
Non-sustained VT Episode #5
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Episode #5: 04-Nov-2020 04:35:32

Episode Summary
Initial Type VT Non-sustained (spontaneous)
Duration 1 sec (5 beats)
V. Median 52 bpm (1160 ms)
Activity at onset Rest, Sensor = 50 bpm

Parameter Settings Initial Redetect V. Interval (Rate)


VF On 30/40 24/32 300 ms (200 bpm)
FVT Off
VT On 36 32 360 ms (167 bpm)
Monitor Monitor 40 440 ms (136 bpm)

Polarity RV
Pace Polarity Bipolar
Sense Polarity Bipolar

EGM Source Range Sensitivity


EGM1 Atip to Aring +/- 8 mV Atrial 0.45 mV
EGM3 RVtip to RVring +/- 8 mV RV 0.6 mV

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #5


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:12:36
Non-sustained VT Episode #4
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 4 31-Oct-2020 16:25 :03 <30/163 Rest

V-V A-A VF = 300 ms VT = 360 ms


Term.
Interval (ms)
1500
1200
900
600

400

200

-60 -55 -50 -45 -40 -35 -30 -25 -20 -15 -10 -5 0
Time (sec)

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #4


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:12:53
Non-sustained VT Episode #4
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #4 Chart speed: 25.0 mm/sec

EGM1: Atip to Aring


(1 mV)

EGM3: RVtip to RVring


(1 mV)

A-A (ms)
1150 1150 1150 1150 1150
A A A A
P P P P
Markers
V V V V V
P P P P P
V-V (ms) 1150 1150 1150 1150

1150 1150 1150 1150


A A A A
P P P P

V V V V V V V V T T
P P P P S S S S S S
1150 1150 1150 1150 9 4 4 3 3 3 3
0 6 0 7 4 4 6
0 0 0 0 0 0 0

>2000 1160 1160 1160 1160


A A A A
P P P P

V V V V T V V V V
S S S S S P P P P
3 4 3 3 3 1160 1160 1160 1160 1160
6 0 7 6 5
0 0 0 0 0
Termination

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:13:10 Page 1
Non-sustained VT Episode #4
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #4 Chart speed: 25.0 mm/sec

1160 1160 1160 1160 1160 1160


A A A A A A A
P P P P P P P

V V V V V V
P P P P P P
1160 1160 1160 1160 1160 1160

1160 1160 1160 1700 1160


A A A A A
P P P P P

V V V V V V
P P P P S P
1160 1160 1160 5 1160
4
0

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:13:12 Page 2
Non-sustained VT Episode #4
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Episode #4: 31-Oct-2020 16:25:29

Episode Summary
Initial Type VT Non-sustained (spontaneous)
Duration 3 sec (6 beats)
V. Median 162 bpm (370 ms)
Activity at onset Rest, Sensor = 58 bpm

Parameter Settings Initial Redetect V. Interval (Rate)


VF On 30/40 24/32 300 ms (200 bpm)
FVT Off
VT On 36 32 360 ms (167 bpm)
Monitor Monitor 40 440 ms (136 bpm)

Polarity RV
Pace Polarity Bipolar
Sense Polarity Bipolar

EGM Source Range Sensitivity


EGM1 Atip to Aring +/- 8 mV Atrial 0.45 mV
EGM3 RVtip to RVring +/- 8 mV RV 0.6 mV

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #4


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:13:38
Non-sustained VT Episode #3
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 3 25-Oct-2020 22:18 :01 <30/176 Rest

V-V A-A VF = 300 ms VT = 360 ms


Term.
Interval (ms)
1500
1200
900
600

400

200

-60 -55 -50 -45 -40 -35 -30 -25 -20 -15 -10 -5 0
Time (sec)

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #3


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:13:59
Non-sustained VT Episode #3
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #3 Chart speed: 25.0 mm/sec

EGM1: Atip to Aring


(1 mV)

EGM3: RVtip to RVring


(1 mV)

A-A (ms)
1200 1200 1200 1200
A A A A
P P P P
Markers
V V V V V
P P P P P
V-V (ms) 1200 1200 1200 1200

1200 1200 1200 1200 1200 1200


A A A A A A
P P P P P P

V V V V V VV
P P P P P SP
1200 1200 1200 1200 1200 1050 4
3
0

>2000 1200 1200 1200 1200


A A A A
P P P P

V V T T T V V V V
S S S S S P P P P
3 3 3 3 1200 1200 1200 1200 1200
7 5 4 4
0 0 0 0
Termination

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:14:21 Page 1
Non-sustained VT Episode #3
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #3 Chart speed: 25.0 mm/sec

1200 1200 1200 1200 1200 1200


A A A A A A
P P P P P P

V V V V V V
P P P P P P
1200 1200 1200 1200 1200 1200

1200 1200 1200 1200 1200 1200


A A A A A A
P P P P P P

V V V V V V
P P P P P P
1200 1200 1200 1200 1200

A
P

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:14:22 Page 2
Non-sustained VT Episode #3
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Episode #3: 25-Oct-2020 22:18:06

Episode Summary
Initial Type VT Non-sustained (spontaneous)
Duration 1 sec (5 beats)
V. Median 50 bpm (1200 ms)
Activity at onset Rest, Sensor = 50 bpm

Parameter Settings Initial Redetect V. Interval (Rate)


VF On 30/40 24/32 300 ms (200 bpm)
FVT Off
VT On 36 32 360 ms (167 bpm)
Monitor Monitor 40 440 ms (136 bpm)

Polarity RV
Pace Polarity Bipolar
Sense Polarity Bipolar

EGM Source Range Sensitivity


EGM1 Atip to Aring +/- 8 mV Atrial 0.45 mV
EGM3 RVtip to RVring +/- 8 mV RV 0.6 mV

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #3


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:14:54
Non-sustained VT Episode #2
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 2 23-Oct-2020 14:36 :02 ---/190 Rest

V-V A-A VF = 300 ms VT = 360 ms


Term.
Interval (ms)
1500
1200
900
600

400

200

-60 -55 -50 -45 -40 -35 -30 -25 -20 -15 -10 -5 0
Time (sec)

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #2


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:15:21
Non-sustained VT Episode #2
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #2 Chart speed: 25.0 mm/sec

EGM1: Atip to Aring


(1 mV)

EGM3: RVtip to RVring


(1 mV)

A-A (ms)
1200 1200 1200 1200
A A A A
P P P P
Markers
V V V V V
P P P P P
V-V (ms) 1200 1200 1200 1200

1200 1200 1200 1200


A A A A
P P P P

V V V V V V V T T
P P P P S S S S S
1200 1200 1200 1200 9 4 3 3 3 3
0 6 7 5 4 0
0 0 0 0 0 0

>2000 1200 1200


A A
P P

T T T T T T T V V V T V V
S S S S S S S S S S S P P
3 3 3 3 3 3 5 3 5 3 1200 1200 1200
2 0 3 3 2 0 2 6 2 5
0 0 0 0 0 0 0 0 0 0
Termination

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:15:48 Page 1
Non-sustained VT Episode #2
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #2 Chart speed: 25.0 mm/sec

1200 1200 1200 1200 1200 1200


A A A A A A
P P P P P P

V V V V V V
P P P P P P
1200 1200 1200 1200 1200 1200

1200 1200 1200 >2000 1200


A A A A A A
P P P P P P

V V V V V V
P P P P S P
1200 1200 1200 9 1200
5
0

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:15:49 Page 2
Non-sustained VT Episode #2
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Episode #2: 23-Oct-2020 14:36:09

Episode Summary
Initial Type VT Non-sustained (spontaneous)
Duration 2 sec (10 beats)
V. Median 182 bpm (330 ms)
Activity at onset Rest, Sensor = 50 bpm

Parameter Settings Initial Redetect V. Interval (Rate)


VF On 30/40 24/32 300 ms (200 bpm)
FVT Off
VT On 36 32 360 ms (167 bpm)
Monitor Monitor 40 440 ms (136 bpm)

Polarity RV
Pace Polarity Bipolar
Sense Polarity Bipolar

EGM Source Range Sensitivity


EGM1 Atip to Aring +/- 8 mV Atrial 0.45 mV
EGM3 RVtip to RVring +/- 8 mV RV 0.6 mV

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #2


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:16:25
Non-sustained VT Episode #2
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 2 23-Oct-2020 14:36 :02 ---/190 Rest

V-V A-A VF = 300 ms VT = 360 ms


Term.
Interval (ms)
1500
1200
900
600

400

200

-60 -55 -50 -45 -40 -35 -30 -25 -20 -15 -10 -5 0
Time (sec)

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #2


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:16:57
Non-sustained VT Episode #2
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #2 Chart speed: 25.0 mm/sec

EGM1: Atip to Aring


(1 mV)

EGM3: RVtip to RVring


(1 mV)

A-A (ms)
1200 1200 1200 1200
A A A A
P P P P
Markers
V V V V V
P P P P P
V-V (ms) 1200 1200 1200 1200

1200 1200 1200 1200


A A A A
P P P P

V V V V V V V T T
P P P P S S S S S
1200 1200 1200 1200 9 4 3 3 3 3
0 6 7 5 4 0
0 0 0 0 0 0

>2000 1200 1200


A A
P P

T T T T T T T V V V T V V
S S S S S S S S S S S P P
3 3 3 3 3 3 5 3 5 3 1200 1200 1200
2 0 3 3 2 0 2 6 2 5
0 0 0 0 0 0 0 0 0 0
Termination

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:17:29 Page 1
Non-sustained VT Episode #2
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Episode #2 Chart speed: 25.0 mm/sec

1200 1200 1200 1200 1200 1200


A A A A A A
P P P P P P

V V V V V V
P P P P P P
1200 1200 1200 1200 1200 1200

1200 1200 1200 >2000 1200


A A A A A A
P P P P P P

V V V V V V
P P P P S P
1200 1200 1200 9 1200
5
0

SW033 Software Version 8.3 (4.1) Copyright © Medtronic, Inc. 2015 16-Nov-2020 10:17:31 Page 2
Non-sustained VT Episode #2
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Episode #2: 23-Oct-2020 14:36:09

Episode Summary
Initial Type VT Non-sustained (spontaneous)
Duration 2 sec (10 beats)
V. Median 182 bpm (330 ms)
Activity at onset Rest, Sensor = 50 bpm

Parameter Settings Initial Redetect V. Interval (Rate)


VF On 30/40 24/32 300 ms (200 bpm)
FVT Off
VT On 36 32 360 ms (167 bpm)
Monitor Monitor 40 440 ms (136 bpm)

Polarity RV
Pace Polarity Bipolar
Sense Polarity Bipolar

EGM Source Range Sensitivity


EGM1 Atip to Aring +/- 8 mV Atrial 0.45 mV
EGM3 RVtip to RVring +/- 8 mV RV 0.6 mV

SW033 Software Version 8.3 (4.1) Non-sustained VT Episode #2


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:18:12
Arrhythmia Episode List
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Arrhythmia Episode List: 14-Oct-2020 06:58:33 to 16-Nov-2020 10:01:30


Only specified episodes shown below.

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 6 11-Nov-2020 00:56 :01 30/222 Active
VT-NS 5 04-Nov-2020 04:35 :01 <30/200 Rest
VT-NS 4 31-Oct-2020 16:25 :03 <30/163 Rest
VT-NS 3 25-Oct-2020 22:18 :01 <30/176 Rest
VT-NS 2 23-Oct-2020 14:36 :02 ---/190 Rest
VT-NS 1 18-Oct-2020 23:28 :02 <30/162 Rest
---------------------- Last Programmer Session 14-Oct-2020 ----------------------------------------------------------------------------------------------
(Data prior to last session has not been interrogated.)

SW033 Software Version 8.3 (4.1) Arrhythmia Episode List


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:18:50
V. Interval Stability List Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Last Interrogation: 16-Nov-2020 10:01:30


Note: V. Interval Stability is calculated for VT and FVT via VT Episodes only.

ID# Date/Time Type V. Cycle V. Interval Stability


---------------------- Last Programmer Session 14-Oct-2020 ----------------------------------------------------------------------------------------------
(Data prior to last session has not been interrogated.)

SW033 Software Version 8.3 (4.1) V. Interval Stability List Report


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:19:28
Final: Session Summary
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Device Information
Device Medtronic Primo MRI DR DDMD3D4 CWN603292S Implanted: 12-Oct-2020
RV 6935M Sprint Quattro MRI
Atrial 5076 CapsureFix Novus MRI
Device Status (Implanted: 12-Oct-2020)
Remaining Longevity 4.2 years (16-Nov-2020)

RRT > 5 years


(based on initial interrogation)

Atrial(5076) RV(6935M)
Pacing Impedance 456 ohms 456 ohms
Defibrillation Impedance RV=64 ohms
Pace Polarity Bipolar Bipolar
In-Office Threshold 0.75 V @ 0.40 ms
Programmed Amplitude/Pulse Width 1.75 V / 0.40 ms 2.00 V / 0.40 ms
Measured P/ R Wave 0.5 mV 4.8 mV
Programmed Sensitivity 0.45 mV 0.60 mV
Parameter Summary
Mode DDDR Lower Rate 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Detection Rates Therapies


AT/AF Monitor >171 bpm All Rx Off
VF On >200 bpm ATP During Charging, 35J x 6
FVT OFF All Rx Off
VT On 167-200 bpm Burst(3), 20J, 35J x 4
Enhancements On: VT Monitor, AF/Afl, Sinus Tach, Wavelet, TWave, Noise(Timeout)
Changes This Session Session Start Current Value
VT Monitor Interval 440 ms (136 bpm) 420 ms (143 bpm)

A. Amplitude 5.00 V 1.75 V


A. Pulse Width 0.60 ms 0.40 ms
RV Amplitude 3.50 V 2.00 V

SW033 Software Version 8.3 (4.1) Final: Session Summary


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:20:03
Final: Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Pacing Summary
Mode Rates AV Intervals
Mode DDDR Lower 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Pacing Details Atrial RV


Amplitude 1.75 V 2.00 V
Pulse Width 0.40 ms 0.40 ms
Sensitivity 0.45 mV 0.60 mV
Pace Polarity Bipolar Bipolar
Sense Polarity Bipolar Bipolar

AV Therapies Arrhythmia Interventions


Rate Adaptive AV Off A. Rate Stabilization Off
A. Preference Pacing Off
Rate Response
Post Mode Switch Off
ADL Rate 95 bpm
Conducted AF Response Off
Optimization On
V. Rate Stabilization Off
ADL Response 3
Exertion Response 3 Post Shock Pacing
Activity Threshold Low Post Shock Pacing*
Activity Acceleration 30 sec A. Amplitude 4V
Activity Deceleration Exercise A. Pulse Width 1.50 ms
ADL Setpoint 14 V. Amplitude 6V
UR Setpoint 55 V. Pulse Width 1.50 ms
* Settings for first 25 beats post shock
Refractory/Blanking
Post VT/VF Shock Pacing On
PVARP Auto
Overdrive Rate 80 bpm
Minimum PVARP 250 ms
Overdrive Duration 0.5 min
PVAB Interval 150 ms
PVAB Method Partial Additional Features
A. Blank Post AP 200 ms Rate Drop Response Off
A. Blank Post AS 100 ms Sleep Off
V. Blank Post VP 200 ms Non-Comp Atrial Pacing On
V. Blank Post VS 120 ms NCAP Interval 300 ms
MRI SureScan Off
PMT Intervention Off
PVC Response On
V. Safety Pacing On
Device Information
Device Medtronic Primo MRI DR DDMD3D4 CWN603292S Implanted: 12-Oct-2020
RV 6935M Sprint Quattro MRI
Atrial 5076 CapsureFix Novus MRI

SW033 Software Version 8.3 (4.1) Final: Parameters


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:20:39
Final: Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

VT/VF Detection
V. Interval (Rate) Initial Redetect
VF On 300 ms (200 bpm) 30/40 24/32 300 ms
FVT OFF
VT On 360 ms (167 bpm) 36 32 360 ms
Monitor Monitor 420 ms (143 bpm) 40 No Rx 420 ms

PR Logic/Wavelet Other Enhancements Sensitivity


AF/Afl On Stability Off Atrial 0.45 mV
Sinus Tach On Onset Off RV 0.60 mV
Other 1:1 SVTs Off High Rate Timeout
Wavelet On VF Zone Only Off
Template None All Zones Off
Match Threshold 70 % TWave On
Auto Collection On RV Lead Noise On+Timeout
SVT V. Limit 260 ms Timeout 0.75 min
VF Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
VF Therapy Status On On On On On On
Energy 35 J 35 J 35 J 35 J 35 J 35 J
Pathway B>AX AX>B B>AX AX>B B>AX AX>B
ATP During Charging
Deliver ATP if last 8 R-R >= 240 ms,Burst, Pulses = 8, R-S1 = 88 %, Decrement = 10 ms
ChargeSaver = On(1 episodes), SmartMode = On
FVT Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
FVT Therapy Status Off Off Off Off Off Off
VT Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
VT Therapy Status On On On On On On
Therapy Type Burst CV CV CV CV CV
Energy 20 J 35 J 35 J 35 J 35 J
Pathway B>AX AX>B B>AX B>AX AX>B
Initial # Pulses 8
R-S1 Interval=(%RR) 88 %
S1S2(Ramp+)=(%RR)
S2SN(Ramp+)=(%RR)
Interval Dec 10 ms
# Sequences 3
Smart Mode On
Shared V. ATP Shared V. Therapies
V-V Minimum ATP Interval 200 ms Active Can/SVC Coil Can+SVC On
V. Amplitude 8V Progressive Episode Therapies Off
V. Pulse Width 1.5 ms Confirmation+ On
V. Pace Blanking 240 ms

SW033 Software Version 8.3 (4.1) Final: Parameters


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:20:39
Final: Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

AT/AF Detection
Detection A. Interval (Rate)
Monitor
AT/AF 350 ms (171 bpm) No Rx 350 ms
AT/AF Therapies

CV therapies have priority over pacing therapies. Patient Activated CV


Patient Activated CV Status Off
AT/AF ATP Rx1 Rx2 Rx3
AT/AF Rx Status Off Off Off AT/AF Automatic CV Rx4 Rx5
Automatic CV Status Off Off
Shared A. ATP
A-A Minimum ATP Interval 150 ms Shared CV
A. Pacing Amplitude 6V Minimum R-R Interval 500 ms (120 bpm)
A. Pacing Pulse Width 1.5 ms Active Can/SVC Coil Can+SVC On
Stop Atrial Therapies After
Disable all A. Rxs if A. lead position is suspect? No
Atrial Lead Position Check
No measurement since reset.

SW033 Software Version 8.3 (4.1) Final: Parameters


Copyright © Medtronic, Inc. 2015 Page 3 Printed: 16-Nov-2020 10:20:39
Final: Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Data Collection Setup

Source Range
LECG Can to SVC +/- 2 mV
EGM 1 Atip to Aring +/- 8 mV
EGM 2 (Wavelet) Can to RVcoil +/- 12 mV
EGM 3 RVtip to RVring +/- 8 mV
Monitored EGM1 and EGM3
Pre-arrhythmia EGM Off

Device Date/Time 16-Nov-2020 10:11


Holter Telemetry Off
Medtronic CareAlert Setup
Patient Home Monitor No
Alert Time 08:00
Enable-Urgency
Alert Conditions Tone Threshold
AT/AF Daily Burden Settings Off 6 hr
Avg. V. Rate During AT/AF Settings Off 6 hr at 100 bpm
Number of Shocks Delivered in an Episode On-Low 1
All Therapies in a Zone Exhausted On-Low
RV Lead Integrity On-High
RV Lead Noise On-High
A. Pacing Lead Impedance Out of Range On-High <200 or >3000 ohms
RV Pacing Lead Impedance Out of Range On-High <200 or >3000 ohms
RV Defibrillation Lead Impedance Out of Range On-High <20 or >200 ohms
SVC Defib Lead Impedance Out of Range On-High <20 or >200 ohms
Low Battery Voltage RRT On-High 2.73 V(RRT)
Excessive Charge Time EOS On-High
VF Detection OFF, 3+ VF or 3+ FVT Rx Off. On-High

SW033 Software Version 8.3 (4.1) Final: Parameters


Copyright © Medtronic, Inc. 2015 Page 4 Printed: 16-Nov-2020 10:20:39
Quick Look II Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:
History: NYHA Class II
Device Status (Implanted: 12-Oct-2020)
Remaining Longevity 4.2 years (16-Nov-2020)

RRT > 5 years


(based on initial interrogation)

Atrial(5076) RV(6935M)
Pacing Impedance 456 ohms 456 ohms
Defibrillation Impedance RV=64 ohms
Programmed Amplitude/Pulse Width 1.75 V / 0.40 ms 2.00 V / 0.40 ms
Measured P/ R Wave 0.5 mV 4.8 mV
Programmed Sensitivity 0.45 mV 0.60 mV
Parameter Summary
Mode DDDR Lower Rate 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Detection Rates Therapies


AT/AF Monitor >171 bpm All Rx Off
VF On >200 bpm ATP During Charging, 35J x 6
FVT OFF All Rx Off
VT On 167-200 bpm Burst(3), 20J, 35J x 4
Enhancements On: VT Monitor, AF/Afl, Sinus Tach, Wavelet, TWave, Noise(Timeout)
Clinical Status Since 14-Oct-2020 Cardiac Compass Trends (Oct-2020 to Nov-2020)
Treated
P I
VF 0
FVT (Off) Treated >5
VT/VF 4
VT 0 3
AT/AF(Monitor) (#/day) 2
1
0
Monitored
VT(143-167 bpm) 0 AT/AF 60
VT-NS (>4 beats, >167 bpm) 6 (min/day) 50
40
High Rate-NS 0 30
SVT: VT/VF Rx Withheld 0 20
10
V. Oversensing-TWave Rx Withheld 0
0
V. Oversensing-Noise Rx Withheld 0
AT/AF 0 Patient >8
Activity 6
Time in AT/AF 0.0 hr/day (0.0%) (hr/day)
4

Functional Last Week 2


Patient Activity 4.4 hr/day 0
Nov-20 Jan-21 Mar-21 May-21 Jul-21 Sep-21 Nov-21

Therapy Summary VT/VF AT/AF Pacing (% of Time Since 14-Oct-2020)


Pace-Terminated Episodes 0 0 Total VP 97.8 % (MVP Off)
Shock-Terminated Episodes 0 0 AS-VS <0.1 %
Total Shocks 0 0 AS-VP 1.5 %
Aborted Charges 0 0 AP-VS 0.2 %
AP-VP 98.3 %

OBSERVATIONS (1)
· No Wavelet template. Auto Collection is On so a template will be collected; or collect one manually. (In Tests/Wavelet)

SW033 Software Version 8.3 (4.1) Quick Look II Report


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:21:17
Session Summary
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Device Information
Device Medtronic Primo MRI DR DDMD3D4 CWN603292S Implanted: 12-Oct-2020
RV 6935M Sprint Quattro MRI
Atrial 5076 CapsureFix Novus MRI
Device Status (Implanted: 12-Oct-2020)
Remaining Longevity 4.2 years (16-Nov-2020)

RRT > 5 years


(based on initial interrogation)

Atrial(5076) RV(6935M)
Pacing Impedance 456 ohms 456 ohms
Defibrillation Impedance RV=64 ohms
Pace Polarity Bipolar Bipolar
In-Office Threshold 0.75 V @ 0.40 ms
Programmed Amplitude/Pulse Width 1.75 V / 0.40 ms 2.00 V / 0.40 ms
Measured P/ R Wave 0.5 mV 4.8 mV
Programmed Sensitivity 0.45 mV 0.60 mV
Parameter Summary
Mode DDDR Lower Rate 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Detection Rates Therapies


AT/AF Monitor >171 bpm All Rx Off
VF On >200 bpm ATP During Charging, 35J x 6
FVT OFF All Rx Off
VT On 167-200 bpm Burst(3), 20J, 35J x 4
Enhancements On: VT Monitor, AF/Afl, Sinus Tach, Wavelet, TWave, Noise(Timeout)
Changes This Session Session Start Current Value
VT Monitor Interval 440 ms (136 bpm) 420 ms (143 bpm)

A. Amplitude 5.00 V 1.75 V


A. Pulse Width 0.60 ms 0.40 ms
RV Amplitude 3.50 V 2.00 V

SW033 Software Version 8.3 (4.1) Session Summary


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:21:53
Rate Histograms
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Prior to Last Session Since Last Session


12-Oct-2020 to 14-Oct-2020 14-Oct-2020 to 16-Nov-2020
45 hours 33 days
% of Time Total VP 97.7 % 97.8 %
AS-VS < 0.1% < 0.1%
AS-VP < 0.1% 1.5 %
AP-VS 0.1 % 0.2 %
AP-VP 99.8 % 98.3 %

Atrial 2% to 5% of AS may be due to FFRW


% of Time 100 100
80 80
AS 60 60
AP
40 40
20 20
0 0
< 40 60 80 100 120 140 160 180 200 220> < 40 60 80 100 120 140 160 180 200 220>
Atrial Rate (bpm) Atrial Rate (bpm)

Ventricular
% of Time 100 100
80 80
VS
60 60
VP
40 40
20 20
0 0
< 40 60 80 100 120 140 160 180 200 220> < 40 60 80 100 120 140 160 180 200 220>
Ventricular Rate (bpm) Ventricular Rate (bpm)

SW033 Software Version 8.3 (4.1) Rate Histograms


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:22:29
Cardiac Compass Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Nov 2020 Jan 2021 Mar 2021 May 2021 Jul 2021 Sep 2021 Nov 2021

P = Program
I = Interrogate
P I
_ = Remote

One or more A
shocks/day
V

Treated VT/VF >5


episodes/day 4
3
2
1
0

V. rate during VT/VF >250


(bpm)
200
VF
VT 150
<100

Non-sustained VT >10
episodes/day 8
6
4
2
0

AT/AF total minutes/day 60


50
40
30
20
10
0

V. rate during AT/AF >200


(bpm)
max/day 150
avg/day
100

<50

% Pacing/day 100
Atrial 75
Ventricular 50
25
0

Avg V. rate (bpm) >120


Day 100
Night 80
60
<40

Patient activity >8


hours/day 6
4
2
0
Nov 2020 Jan 2021 Mar 2021 May 2021 Jul 2021 Sep 2021 Nov 2021

SW033 Software Version 8.3 (4.1) Cardiac Compass Report


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:23:06
Cardiac Compass Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

P I

Heart rate variability >200


(ms) 160
120
80
<40
Nov 2020 Jan 2021 Mar 2021 May 2021 Jul 2021 Sep 2021 Nov 2021

SW033 Software Version 8.3 (4.1) Cardiac Compass Report


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:23:07
VT/VF Counters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Prior Session Last Session Device Lifetime


12-Oct-2020 to 14-Oct-2020 to Total
14-Oct-2020 16-Nov-2020 (Since 12-Oct-2020)
45 hours 33 days 35 days
VT/VF Counters
VF 0 0 0
FVT 0 0 0
VT 0 0 0
Monitored VT (143 - 167 bpm) 0 0
VT-NS (>4 beats, >167 bpm) 0 6
High Rate-NS 0 0
PVC Runs (2-4 beats) 0.5 per hour 1.6 per hour
PVC Singles 26.3 per hour 72.7 per hour
Runs of VRS Paces 0.0 per hour 0.0 per hour
Single VRS Paces 0.0 per hour 0.0 per hour
SVT: VT/VF Rx Withheld
AFib/AFlutter 0 0 0
Sinus Tach 0 0 0
Other 1:1 SVTs 0 0 0
Wavelet 0 0 0
V. Stability 0 0 0
Onset 0 0 0
V. Oversensing: VT/VF Rx Withheld
V. Oversensing-TWave 0 0 0
V. Oversensing-Noise 0 0 0
VT/VF Therapy Summary
Pace-terminated 0 0 0
Shock-terminated 0 0 0
Total VT/VF Shocks 0 0 0
Aborted Charges 0 0 0

VT/VF Therapy Efficacy Since Last Session 14-Oct-2020 to 16-Nov-2020


VF Therapy ATP+Chg 35 J 35 J 35 J 35 J 35 J 35 J
Delivered 0 0 0 0 0 0 0
Successful 0 0 0 0 0 0 0
FVT Therapy Off
VT Therapy Burst 20 J 35 J 35 J 35 J 35 J
Delivered 0 0 0 0 0 0
Successful 0 0 0 0 0 0
Accelerated 0 0 0 0 0 0

SW033 Software Version 8.3 (4.1) VT/VF Counters


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:23:43
Arrhythmia Episode List
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Arrhythmia Episode List: 14-Oct-2020 06:58:33 to 16-Nov-2020 10:01:30


All collected episodes.

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 6 11-Nov-2020 00:56 :01 30/222 Active
VT-NS 5 04-Nov-2020 04:35 :01 <30/200 Rest
VT-NS 4 31-Oct-2020 16:25 :03 <30/163 Rest
VT-NS 3 25-Oct-2020 22:18 :01 <30/176 Rest
VT-NS 2 23-Oct-2020 14:36 :02 ---/190 Rest
VT-NS 1 18-Oct-2020 23:28 :02 <30/162 Rest
---------------------- Last Programmer Session 14-Oct-2020 ----------------------------------------------------------------------------------------------
(Data prior to last session has not been interrogated.)

SW033 Software Version 8.3 (4.1) Arrhythmia Episode List


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:24:19
Lead Trends - Atrial
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Lead Model 5076

A. Impedance (bipolar) Last Session ohms


Last Measured 456 ohms
>3000

2500

2000

1500

1000

500

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

P-Wave Amplitude mV
Sensitivity 0.45 mV >20
Last Measured 0.5 mV

16

12

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

Last 80 weeks (min/max/avg per week) Last 15 days

SW033 Software Version 8.3 (4.1) Lead Trends - Atrial


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:24:56
Lead Trends - RV
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

RV Pace Polarity Bipolar


RV Sense Polarity Bipolar
Lead Model 6935M

RV Impedance (bipolar) Last Session ohms


Last Measured 456 ohms
>3000

2500

2000

1500

1000

500

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

RV Impedance (tip to coil) ohms


Last Measured 342 ohms
>3000

2500

2000

1500

1000

500

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

RV Impedance ohms
Bipolar
Tip to Coil >3000

2500

2000

1500

1000

500

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

Last 80 weeks (min/max/avg per week) Last 15 days

SW033 Software Version 8.3 (4.1) Lead Trends - RV


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:25:32
Lead Trends - RV
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

R-Wave Amplitude mV
Last Session
Sensitivity 0.60 mV >20
Last Measured 4.8 mV

16

12

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

Last 80 weeks (min/max/avg per week) Last 15 days

SW033 Software Version 8.3 (4.1) Lead Trends - RV


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:25:32
Lead Trends - RV
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

RV Defib Impedance Last Session ohms


Last Measured 64 ohms
>200

150

100

50

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

Last 80 weeks (min/max/avg per week) Last 15 days

SW033 Software Version 8.3 (4.1) Lead Trends - RV


Copyright © Medtronic, Inc. 2015 Page 3 Printed: 16-Nov-2020 10:25:32
Battery and Lead Measurements Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Last Interrogation: 16-Nov-2020 10:01:30

Remaining Longevity 16-Nov-2020


Estimated at: 4.2 years
Minimum: 3.8 years
Maximum: 4.7 years

RRT > 5 years


(based on initial interrogation)
Battery Voltage 16-Nov-2020
Voltage 3.03 V
(RRT=2.73V)
Last Charge 20-Oct-2020
Charge Time 3.7 sec
Energy 0.0 - 18 J
Sensing Integrity Counter Since 05-Nov-2020
Short V-V Intervals 1
Atrial Lead Position Check
No measurement since reset.
Lead Impedance
A. Pacing 456 ohms 16-Nov-2020
RV Pacing (Bipolar) 456 ohms 16-Nov-2020
RV Defib 64 ohms 16-Nov-2020
SVC Defib None
Sensing
P-Wave Amplitude 0.5 mV 16-Nov-2020
R-Wave Amplitude 4.8 mV 15-Nov-2020
Last High Voltage Therapy
No measurement since reset.

SW033 Software Version 8.3 (4.1) Battery and Lead Measurements Report
Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:26:08
Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Pacing Summary
Mode Rates AV Intervals
Mode DDDR Lower 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Pacing Details Atrial RV


Amplitude 1.75 V 2.00 V
Pulse Width 0.40 ms 0.40 ms
Sensitivity 0.45 mV 0.60 mV
Pace Polarity Bipolar Bipolar
Sense Polarity Bipolar Bipolar

AV Therapies Arrhythmia Interventions


Rate Adaptive AV Off A. Rate Stabilization Off
A. Preference Pacing Off
Rate Response
Post Mode Switch Off
ADL Rate 95 bpm
Conducted AF Response Off
Optimization On
V. Rate Stabilization Off
ADL Response 3
Exertion Response 3 Post Shock Pacing
Activity Threshold Low Post Shock Pacing*
Activity Acceleration 30 sec A. Amplitude 4V
Activity Deceleration Exercise A. Pulse Width 1.50 ms
ADL Setpoint 14 V. Amplitude 6V
UR Setpoint 55 V. Pulse Width 1.50 ms
* Settings for first 25 beats post shock
Refractory/Blanking
Post VT/VF Shock Pacing On
PVARP Auto
Overdrive Rate 80 bpm
Minimum PVARP 250 ms
Overdrive Duration 0.5 min
PVAB Interval 150 ms
PVAB Method Partial Additional Features
A. Blank Post AP 200 ms Rate Drop Response Off
A. Blank Post AS 100 ms Sleep Off
V. Blank Post VP 200 ms Non-Comp Atrial Pacing On
V. Blank Post VS 120 ms NCAP Interval 300 ms
MRI SureScan Off
PMT Intervention Off
PVC Response On
V. Safety Pacing On
Device Information
Device Medtronic Primo MRI DR DDMD3D4 CWN603292S Implanted: 12-Oct-2020
RV 6935M Sprint Quattro MRI
Atrial 5076 CapsureFix Novus MRI

SW033 Software Version 8.3 (4.1) Parameters


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:26:44
Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

VT/VF Detection
V. Interval (Rate) Initial Redetect
VF On 300 ms (200 bpm) 30/40 24/32 300 ms
FVT OFF
VT On 360 ms (167 bpm) 36 32 360 ms
Monitor Monitor 420 ms (143 bpm) 40 No Rx 420 ms

PR Logic/Wavelet Other Enhancements Sensitivity


AF/Afl On Stability Off Atrial 0.45 mV
Sinus Tach On Onset Off RV 0.60 mV
Other 1:1 SVTs Off High Rate Timeout
Wavelet On VF Zone Only Off
Template None All Zones Off
Match Threshold 70 % TWave On
Auto Collection On RV Lead Noise On+Timeout
SVT V. Limit 260 ms Timeout 0.75 min
VF Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
VF Therapy Status On On On On On On
Energy 35 J 35 J 35 J 35 J 35 J 35 J
Pathway B>AX AX>B B>AX AX>B B>AX AX>B
ATP During Charging
Deliver ATP if last 8 R-R >= 240 ms,Burst, Pulses = 8, R-S1 = 88 %, Decrement = 10 ms
ChargeSaver = On(1 episodes), SmartMode = On
FVT Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
FVT Therapy Status Off Off Off Off Off Off
VT Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
VT Therapy Status On On On On On On
Therapy Type Burst CV CV CV CV CV
Energy 20 J 35 J 35 J 35 J 35 J
Pathway B>AX AX>B B>AX B>AX AX>B
Initial # Pulses 8
R-S1 Interval=(%RR) 88 %
S1S2(Ramp+)=(%RR)
S2SN(Ramp+)=(%RR)
Interval Dec 10 ms
# Sequences 3
Smart Mode On
Shared V. ATP Shared V. Therapies
V-V Minimum ATP Interval 200 ms Active Can/SVC Coil Can+SVC On
V. Amplitude 8V Progressive Episode Therapies Off
V. Pulse Width 1.5 ms Confirmation+ On
V. Pace Blanking 240 ms

SW033 Software Version 8.3 (4.1) Parameters


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:26:44
Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

AT/AF Detection
Detection A. Interval (Rate)
Monitor
AT/AF 350 ms (171 bpm) No Rx 350 ms
AT/AF Therapies

CV therapies have priority over pacing therapies. Patient Activated CV


Patient Activated CV Status Off
AT/AF ATP Rx1 Rx2 Rx3
AT/AF Rx Status Off Off Off AT/AF Automatic CV Rx4 Rx5
Automatic CV Status Off Off
Shared A. ATP
A-A Minimum ATP Interval 150 ms Shared CV
A. Pacing Amplitude 6V Minimum R-R Interval 500 ms (120 bpm)
A. Pacing Pulse Width 1.5 ms Active Can/SVC Coil Can+SVC On
Stop Atrial Therapies After
Disable all A. Rxs if A. lead position is suspect? No
Atrial Lead Position Check
No measurement since reset.

SW033 Software Version 8.3 (4.1) Parameters


Copyright © Medtronic, Inc. 2015 Page 3 Printed: 16-Nov-2020 10:26:45
Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Data Collection Setup

Source Range
LECG Can to SVC +/- 2 mV
EGM 1 Atip to Aring +/- 8 mV
EGM 2 (Wavelet) Can to RVcoil +/- 12 mV
EGM 3 RVtip to RVring +/- 8 mV
Monitored EGM1 and EGM3
Pre-arrhythmia EGM Off

Device Date/Time 16-Nov-2020 10:11


Holter Telemetry Off
Medtronic CareAlert Setup
Patient Home Monitor No
Alert Time 08:00
Enable-Urgency
Alert Conditions Tone Threshold
AT/AF Daily Burden Settings Off 6 hr
Avg. V. Rate During AT/AF Settings Off 6 hr at 100 bpm
Number of Shocks Delivered in an Episode On-Low 1
All Therapies in a Zone Exhausted On-Low
RV Lead Integrity On-High
RV Lead Noise On-High
A. Pacing Lead Impedance Out of Range On-High <200 or >3000 ohms
RV Pacing Lead Impedance Out of Range On-High <200 or >3000 ohms
RV Defibrillation Lead Impedance Out of Range On-High <20 or >200 ohms
SVC Defib Lead Impedance Out of Range On-High <20 or >200 ohms
Low Battery Voltage RRT On-High 2.73 V(RRT)
Excessive Charge Time EOS On-High
VF Detection OFF, 3+ VF or 3+ FVT Rx Off. On-High

SW033 Software Version 8.3 (4.1) Parameters


Copyright © Medtronic, Inc. 2015 Page 4 Printed: 16-Nov-2020 10:26:45
Changes This Session
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Parameter Session Start Current Value


VT Monitor Interval 440 ms (136 bpm) 420 ms (143 bpm)

A. Amplitude 5.00 V 1.75 V


A. Pulse Width 0.60 ms 0.40 ms
RV Amplitude 3.50 V 2.00 V

SW033 Software Version 8.3 (4.1) Changes This Session


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:27:21
Patient Information Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Patient Identification
Patient Gherman Ioan
ID
Date of Birth
History
VT/VF
Slowest VT
Atrial Status
AV Conduction
Heart Failure NYHA Class II
Activity Level
EF, on 30 %
Physician Information
Physician
Phone
Hospital

Last Update 12-Oct-2020


Leads
Lead 1
Lead Model 6935M Sprint Quattro MRI
MR Conditional Yes
Lead Position RV
Lead Length
Manufacturer
Serial Number
Implant Date
Lead 2
Lead Model 5076 CapsureFix Novus MRI
MR Conditional Yes
Lead Position Atrial
Lead Length
Manufacturer
Serial Number
Implant Date
Implant
Implant Date 12-Oct-2020
Defibrillation Energy ---
HV Impedance (ohms)
Atrial RV
P/R Wave Amplitude (mV)
Slew Rate --- ---
Pacing Threshold Amp. --- ---
Pulse Width Setting --- ---
Pacing Impedance (ohms)
Threshold Current (mA)
Notes:
MRI SureScan System/Other Hardware
Last Update 12-Oct-2020

MRI SureScan System


MR Conditional Device Implanted Yes
MR Conditional Lead 1 Implanted Yes
MR Conditional Lead 2 Implanted Yes

SW033 Software Version 8.3 (4.1) Patient Information Report


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:27:58
Patient Information Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

MRI SureScan System/Other Hardware


Other Hardware
Other Devices Unknown
Other Leads Unknown
Lead Extenders/Adaptors Unknown
Other Hardware Notes:

SW033 Software Version 8.3 (4.1) Patient Information Report


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:27:58
Rate Drop Response Episode List
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Rate Drop Response Episodes Summary


Prior to Last Session Since Last Session
12-Oct-2020 to 14-Oct-2020 14-Oct-2020 to 16-Nov-2020
45 hours 33 days
Drop Episodes 0 0
Low Rate Episodes 0 0
Total Episodes 0.0 per day 0.0 per day

Rate Drop Response Episode List: 14-Oct-2020 06:58:33 to 16-Nov-2020 10:01:30


Time Detection Peak
Type ID# Date
hh:mm V. Rate bpm V. Rate bpm
(No data since last session.)
---------------------- Last Programmer Session 14-Oct-2020 ----------------------------------------------------------------------------------------------
(Data prior to last session has not been interrogated.)

SW033 Software Version 8.3 (4.1) Rate Drop Response Episode List
Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:28:35
Quick Look II Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:
History: NYHA Class II
Device Status (Implanted: 12-Oct-2020)
Remaining Longevity 4.2 years (16-Nov-2020)

RRT > 5 years


(based on initial interrogation)

Atrial(5076) RV(6935M)
Pacing Impedance 456 ohms 456 ohms
Defibrillation Impedance RV=64 ohms
Programmed Amplitude/Pulse Width 1.75 V / 0.40 ms 2.00 V / 0.40 ms
Measured P/ R Wave 0.5 mV 4.8 mV
Programmed Sensitivity 0.45 mV 0.60 mV
Parameter Summary
Mode DDDR Lower Rate 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Detection Rates Therapies


AT/AF Monitor >171 bpm All Rx Off
VF On >200 bpm ATP During Charging, 35J x 6
FVT OFF All Rx Off
VT On 167-200 bpm Burst(3), 20J, 35J x 4
Enhancements On: VT Monitor, AF/Afl, Sinus Tach, Wavelet, TWave, Noise(Timeout)
Clinical Status Since 14-Oct-2020 Cardiac Compass Trends (Oct-2020 to Nov-2020)
Treated
P I
VF 0
FVT (Off) Treated >5
VT/VF 4
VT 0 3
AT/AF(Monitor) (#/day) 2
1
0
Monitored
VT(143-167 bpm) 0 AT/AF 60
VT-NS (>4 beats, >167 bpm) 6 (min/day) 50
40
High Rate-NS 0 30
SVT: VT/VF Rx Withheld 0 20
10
V. Oversensing-TWave Rx Withheld 0
0
V. Oversensing-Noise Rx Withheld 0
AT/AF 0 Patient >8
Activity 6
Time in AT/AF 0.0 hr/day (0.0%) (hr/day)
4

Functional Last Week 2


Patient Activity 4.4 hr/day 0
Nov-20 Jan-21 Mar-21 May-21 Jul-21 Sep-21 Nov-21

Therapy Summary VT/VF AT/AF Pacing (% of Time Since 14-Oct-2020)


Pace-Terminated Episodes 0 0 Total VP 97.8 % (MVP Off)
Shock-Terminated Episodes 0 0 AS-VS <0.1 %
Total Shocks 0 0 AS-VP 1.5 %
Aborted Charges 0 0 AP-VS 0.2 %
AP-VP 98.3 %

OBSERVATIONS (1)
· No Wavelet template. Auto Collection is On so a template will be collected; or collect one manually. (In Tests/Wavelet)

SW033 Software Version 8.3 (4.1) Quick Look II Report


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:29:12
Session Summary
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Device Information
Device Medtronic Primo MRI DR DDMD3D4 CWN603292S Implanted: 12-Oct-2020
RV 6935M Sprint Quattro MRI
Atrial 5076 CapsureFix Novus MRI
Device Status (Implanted: 12-Oct-2020)
Remaining Longevity 4.2 years (16-Nov-2020)

RRT > 5 years


(based on initial interrogation)

Atrial(5076) RV(6935M)
Pacing Impedance 456 ohms 456 ohms
Defibrillation Impedance RV=64 ohms
Pace Polarity Bipolar Bipolar
In-Office Threshold 0.75 V @ 0.40 ms
Programmed Amplitude/Pulse Width 1.75 V / 0.40 ms 2.00 V / 0.40 ms
Measured P/ R Wave 0.5 mV 4.8 mV
Programmed Sensitivity 0.45 mV 0.60 mV
Parameter Summary
Mode DDDR Lower Rate 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Detection Rates Therapies


AT/AF Monitor >171 bpm All Rx Off
VF On >200 bpm ATP During Charging, 35J x 6
FVT OFF All Rx Off
VT On 167-200 bpm Burst(3), 20J, 35J x 4
Enhancements On: VT Monitor, AF/Afl, Sinus Tach, Wavelet, TWave, Noise(Timeout)
Changes This Session Session Start Current Value
VT Monitor Interval 440 ms (136 bpm) 420 ms (143 bpm)

A. Amplitude 5.00 V 1.75 V


A. Pulse Width 0.60 ms 0.40 ms
RV Amplitude 3.50 V 2.00 V

SW033 Software Version 8.3 (4.1) Session Summary


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:29:49
Rate Histograms
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Prior to Last Session Since Last Session


12-Oct-2020 to 14-Oct-2020 14-Oct-2020 to 16-Nov-2020
45 hours 33 days
% of Time Total VP 97.7 % 97.8 %
AS-VS < 0.1% < 0.1%
AS-VP < 0.1% 1.5 %
AP-VS 0.1 % 0.2 %
AP-VP 99.8 % 98.3 %

Atrial 2% to 5% of AS may be due to FFRW


% of Time 100 100
80 80
AS 60 60
AP
40 40
20 20
0 0
< 40 60 80 100 120 140 160 180 200 220> < 40 60 80 100 120 140 160 180 200 220>
Atrial Rate (bpm) Atrial Rate (bpm)

Ventricular
% of Time 100 100
80 80
VS
60 60
VP
40 40
20 20
0 0
< 40 60 80 100 120 140 160 180 200 220> < 40 60 80 100 120 140 160 180 200 220>
Ventricular Rate (bpm) Ventricular Rate (bpm)

SW033 Software Version 8.3 (4.1) Rate Histograms


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:30:26
Cardiac Compass Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Nov 2020 Jan 2021 Mar 2021 May 2021 Jul 2021 Sep 2021 Nov 2021

P = Program
I = Interrogate
P I
_ = Remote

One or more A
shocks/day
V

Treated VT/VF >5


episodes/day 4
3
2
1
0

V. rate during VT/VF >250


(bpm)
200
VF
VT 150
<100

Non-sustained VT >10
episodes/day 8
6
4
2
0

AT/AF total minutes/day 60


50
40
30
20
10
0

V. rate during AT/AF >200


(bpm)
max/day 150
avg/day
100

<50

% Pacing/day 100
Atrial 75
Ventricular 50
25
0

Avg V. rate (bpm) >120


Day 100
Night 80
60
<40

Patient activity >8


hours/day 6
4
2
0
Nov 2020 Jan 2021 Mar 2021 May 2021 Jul 2021 Sep 2021 Nov 2021

SW033 Software Version 8.3 (4.1) Cardiac Compass Report


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:31:03
Cardiac Compass Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

P I

Heart rate variability >200


(ms) 160
120
80
<40
Nov 2020 Jan 2021 Mar 2021 May 2021 Jul 2021 Sep 2021 Nov 2021

SW033 Software Version 8.3 (4.1) Cardiac Compass Report


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:31:04
VT/VF Counters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Prior Session Last Session Device Lifetime


12-Oct-2020 to 14-Oct-2020 to Total
14-Oct-2020 16-Nov-2020 (Since 12-Oct-2020)
45 hours 33 days 35 days
VT/VF Counters
VF 0 0 0
FVT 0 0 0
VT 0 0 0
Monitored VT (143 - 167 bpm) 0 0
VT-NS (>4 beats, >167 bpm) 0 6
High Rate-NS 0 0
PVC Runs (2-4 beats) 0.5 per hour 1.6 per hour
PVC Singles 26.3 per hour 72.7 per hour
Runs of VRS Paces 0.0 per hour 0.0 per hour
Single VRS Paces 0.0 per hour 0.0 per hour
SVT: VT/VF Rx Withheld
AFib/AFlutter 0 0 0
Sinus Tach 0 0 0
Other 1:1 SVTs 0 0 0
Wavelet 0 0 0
V. Stability 0 0 0
Onset 0 0 0
V. Oversensing: VT/VF Rx Withheld
V. Oversensing-TWave 0 0 0
V. Oversensing-Noise 0 0 0
VT/VF Therapy Summary
Pace-terminated 0 0 0
Shock-terminated 0 0 0
Total VT/VF Shocks 0 0 0
Aborted Charges 0 0 0

VT/VF Therapy Efficacy Since Last Session 14-Oct-2020 to 16-Nov-2020


VF Therapy ATP+Chg 35 J 35 J 35 J 35 J 35 J 35 J
Delivered 0 0 0 0 0 0 0
Successful 0 0 0 0 0 0 0
FVT Therapy Off
VT Therapy Burst 20 J 35 J 35 J 35 J 35 J
Delivered 0 0 0 0 0 0
Successful 0 0 0 0 0 0
Accelerated 0 0 0 0 0 0

SW033 Software Version 8.3 (4.1) VT/VF Counters


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:31:41
Arrhythmia Episode List
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Arrhythmia Episode List: 14-Oct-2020 06:58:33 to 16-Nov-2020 10:01:30


All collected episodes.

ATP Time Duration Avg bpm Max bpm Activity at


Type Shocks Success ID# Date
Seq hh:mm hh:mm:ss A/V A/V Onset
VT-NS 6 11-Nov-2020 00:56 :01 30/222 Active
VT-NS 5 04-Nov-2020 04:35 :01 <30/200 Rest
VT-NS 4 31-Oct-2020 16:25 :03 <30/163 Rest
VT-NS 3 25-Oct-2020 22:18 :01 <30/176 Rest
VT-NS 2 23-Oct-2020 14:36 :02 ---/190 Rest
VT-NS 1 18-Oct-2020 23:28 :02 <30/162 Rest
---------------------- Last Programmer Session 14-Oct-2020 ----------------------------------------------------------------------------------------------
(Data prior to last session has not been interrogated.)

SW033 Software Version 8.3 (4.1) Arrhythmia Episode List


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:32:19
Lead Trends - Atrial
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Lead Model 5076

A. Impedance (bipolar) Last Session ohms


Last Measured 456 ohms
>3000

2500

2000

1500

1000

500

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

P-Wave Amplitude mV
Sensitivity 0.45 mV >20
Last Measured 0.5 mV

16

12

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

Last 80 weeks (min/max/avg per week) Last 15 days

SW033 Software Version 8.3 (4.1) Lead Trends - Atrial


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:32:57
Lead Trends - RV
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

RV Pace Polarity Bipolar


RV Sense Polarity Bipolar
Lead Model 6935M

RV Impedance (bipolar) Last Session ohms


Last Measured 456 ohms
>3000

2500

2000

1500

1000

500

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

RV Impedance (tip to coil) ohms


Last Measured 342 ohms
>3000

2500

2000

1500

1000

500

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

RV Impedance ohms
Bipolar
Tip to Coil >3000

2500

2000

1500

1000

500

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

Last 80 weeks (min/max/avg per week) Last 15 days

SW033 Software Version 8.3 (4.1) Lead Trends - RV


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:33:35
Lead Trends - RV
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

R-Wave Amplitude mV
Last Session
Sensitivity 0.60 mV >20
Last Measured 4.8 mV

16

12

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

Last 80 weeks (min/max/avg per week) Last 15 days

SW033 Software Version 8.3 (4.1) Lead Trends - RV


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:33:35
Lead Trends - RV
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

RV Defib Impedance Last Session ohms


Last Measured 64 ohms
>200

150

100

50

0
May-19 Sep-19 Feb-20 Jun-20 02-Nov-20 16-Nov-20

Last 80 weeks (min/max/avg per week) Last 15 days

SW033 Software Version 8.3 (4.1) Lead Trends - RV


Copyright © Medtronic, Inc. 2015 Page 3 Printed: 16-Nov-2020 10:33:35
Battery and Lead Measurements Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Last Interrogation: 16-Nov-2020 10:01:30

Remaining Longevity 16-Nov-2020


Estimated at: 4.2 years
Minimum: 3.8 years
Maximum: 4.7 years

RRT > 5 years


(based on initial interrogation)
Battery Voltage 16-Nov-2020
Voltage 3.03 V
(RRT=2.73V)
Last Charge 20-Oct-2020
Charge Time 3.7 sec
Energy 0.0 - 18 J
Sensing Integrity Counter Since 05-Nov-2020
Short V-V Intervals 1
Atrial Lead Position Check
No measurement since reset.
Lead Impedance
A. Pacing 456 ohms 16-Nov-2020
RV Pacing (Bipolar) 456 ohms 16-Nov-2020
RV Defib 64 ohms 16-Nov-2020
SVC Defib None
Sensing
P-Wave Amplitude 0.5 mV 16-Nov-2020
R-Wave Amplitude 4.8 mV 15-Nov-2020
Last High Voltage Therapy
No measurement since reset.

SW033 Software Version 8.3 (4.1) Battery and Lead Measurements Report
Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:34:13
Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Pacing Summary
Mode Rates AV Intervals
Mode DDDR Lower 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Pacing Details Atrial RV


Amplitude 1.75 V 2.00 V
Pulse Width 0.40 ms 0.40 ms
Sensitivity 0.45 mV 0.60 mV
Pace Polarity Bipolar Bipolar
Sense Polarity Bipolar Bipolar

AV Therapies Arrhythmia Interventions


Rate Adaptive AV Off A. Rate Stabilization Off
A. Preference Pacing Off
Rate Response
Post Mode Switch Off
ADL Rate 95 bpm
Conducted AF Response Off
Optimization On
V. Rate Stabilization Off
ADL Response 3
Exertion Response 3 Post Shock Pacing
Activity Threshold Low Post Shock Pacing*
Activity Acceleration 30 sec A. Amplitude 4V
Activity Deceleration Exercise A. Pulse Width 1.50 ms
ADL Setpoint 14 V. Amplitude 6V
UR Setpoint 55 V. Pulse Width 1.50 ms
* Settings for first 25 beats post shock
Refractory/Blanking
Post VT/VF Shock Pacing On
PVARP Auto
Overdrive Rate 80 bpm
Minimum PVARP 250 ms
Overdrive Duration 0.5 min
PVAB Interval 150 ms
PVAB Method Partial Additional Features
A. Blank Post AP 200 ms Rate Drop Response Off
A. Blank Post AS 100 ms Sleep Off
V. Blank Post VP 200 ms Non-Comp Atrial Pacing On
V. Blank Post VS 120 ms NCAP Interval 300 ms
MRI SureScan Off
PMT Intervention Off
PVC Response On
V. Safety Pacing On
Device Information
Device Medtronic Primo MRI DR DDMD3D4 CWN603292S Implanted: 12-Oct-2020
RV 6935M Sprint Quattro MRI
Atrial 5076 CapsureFix Novus MRI

SW033 Software Version 8.3 (4.1) Parameters


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:34:51
Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

VT/VF Detection
V. Interval (Rate) Initial Redetect
VF On 300 ms (200 bpm) 30/40 24/32 300 ms
FVT OFF
VT On 360 ms (167 bpm) 36 32 360 ms
Monitor Monitor 420 ms (143 bpm) 40 No Rx 420 ms

PR Logic/Wavelet Other Enhancements Sensitivity


AF/Afl On Stability Off Atrial 0.45 mV
Sinus Tach On Onset Off RV 0.60 mV
Other 1:1 SVTs Off High Rate Timeout
Wavelet On VF Zone Only Off
Template None All Zones Off
Match Threshold 70 % TWave On
Auto Collection On RV Lead Noise On+Timeout
SVT V. Limit 260 ms Timeout 0.75 min
VF Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
VF Therapy Status On On On On On On
Energy 35 J 35 J 35 J 35 J 35 J 35 J
Pathway B>AX AX>B B>AX AX>B B>AX AX>B
ATP During Charging
Deliver ATP if last 8 R-R >= 240 ms,Burst, Pulses = 8, R-S1 = 88 %, Decrement = 10 ms
ChargeSaver = On(1 episodes), SmartMode = On
FVT Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
FVT Therapy Status Off Off Off Off Off Off
VT Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
VT Therapy Status On On On On On On
Therapy Type Burst CV CV CV CV CV
Energy 20 J 35 J 35 J 35 J 35 J
Pathway B>AX AX>B B>AX B>AX AX>B
Initial # Pulses 8
R-S1 Interval=(%RR) 88 %
S1S2(Ramp+)=(%RR)
S2SN(Ramp+)=(%RR)
Interval Dec 10 ms
# Sequences 3
Smart Mode On
Shared V. ATP Shared V. Therapies
V-V Minimum ATP Interval 200 ms Active Can/SVC Coil Can+SVC On
V. Amplitude 8V Progressive Episode Therapies Off
V. Pulse Width 1.5 ms Confirmation+ On
V. Pace Blanking 240 ms

SW033 Software Version 8.3 (4.1) Parameters


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:34:51
Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

AT/AF Detection
Detection A. Interval (Rate)
Monitor
AT/AF 350 ms (171 bpm) No Rx 350 ms
AT/AF Therapies

CV therapies have priority over pacing therapies. Patient Activated CV


Patient Activated CV Status Off
AT/AF ATP Rx1 Rx2 Rx3
AT/AF Rx Status Off Off Off AT/AF Automatic CV Rx4 Rx5
Automatic CV Status Off Off
Shared A. ATP
A-A Minimum ATP Interval 150 ms Shared CV
A. Pacing Amplitude 6V Minimum R-R Interval 500 ms (120 bpm)
A. Pacing Pulse Width 1.5 ms Active Can/SVC Coil Can+SVC On
Stop Atrial Therapies After
Disable all A. Rxs if A. lead position is suspect? No
Atrial Lead Position Check
No measurement since reset.

SW033 Software Version 8.3 (4.1) Parameters


Copyright © Medtronic, Inc. 2015 Page 3 Printed: 16-Nov-2020 10:34:52
Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Data Collection Setup

Source Range
LECG Can to SVC +/- 2 mV
EGM 1 Atip to Aring +/- 8 mV
EGM 2 (Wavelet) Can to RVcoil +/- 12 mV
EGM 3 RVtip to RVring +/- 8 mV
Monitored EGM1 and EGM3
Pre-arrhythmia EGM Off

Device Date/Time 16-Nov-2020 10:12


Holter Telemetry Off
Medtronic CareAlert Setup
Patient Home Monitor No
Alert Time 08:00
Enable-Urgency
Alert Conditions Tone Threshold
AT/AF Daily Burden Settings Off 6 hr
Avg. V. Rate During AT/AF Settings Off 6 hr at 100 bpm
Number of Shocks Delivered in an Episode On-Low 1
All Therapies in a Zone Exhausted On-Low
RV Lead Integrity On-High
RV Lead Noise On-High
A. Pacing Lead Impedance Out of Range On-High <200 or >3000 ohms
RV Pacing Lead Impedance Out of Range On-High <200 or >3000 ohms
RV Defibrillation Lead Impedance Out of Range On-High <20 or >200 ohms
SVC Defib Lead Impedance Out of Range On-High <20 or >200 ohms
Low Battery Voltage RRT On-High 2.73 V(RRT)
Excessive Charge Time EOS On-High
VF Detection OFF, 3+ VF or 3+ FVT Rx Off. On-High

SW033 Software Version 8.3 (4.1) Parameters


Copyright © Medtronic, Inc. 2015 Page 4 Printed: 16-Nov-2020 10:34:52
Changes This Session
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Parameter Session Start Current Value


VT Monitor Interval 440 ms (136 bpm) 420 ms (143 bpm)

A. Amplitude 5.00 V 1.75 V


A. Pulse Width 0.60 ms 0.40 ms
RV Amplitude 3.50 V 2.00 V

SW033 Software Version 8.3 (4.1) Changes This Session


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:35:30
Patient Information Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Patient Identification
Patient Gherman Ioan
ID
Date of Birth
History
VT/VF
Slowest VT
Atrial Status
AV Conduction
Heart Failure NYHA Class II
Activity Level
EF, on 30 %
Physician Information
Physician
Phone
Hospital

Last Update 12-Oct-2020


Leads
Lead 1
Lead Model 6935M Sprint Quattro MRI
MR Conditional Yes
Lead Position RV
Lead Length
Manufacturer
Serial Number
Implant Date
Lead 2
Lead Model 5076 CapsureFix Novus MRI
MR Conditional Yes
Lead Position Atrial
Lead Length
Manufacturer
Serial Number
Implant Date
Implant
Implant Date 12-Oct-2020
Defibrillation Energy ---
HV Impedance (ohms)
Atrial RV
P/R Wave Amplitude (mV)
Slew Rate --- ---
Pacing Threshold Amp. --- ---
Pulse Width Setting --- ---
Pacing Impedance (ohms)
Threshold Current (mA)
Notes:
MRI SureScan System/Other Hardware
Last Update 12-Oct-2020

MRI SureScan System


MR Conditional Device Implanted Yes
MR Conditional Lead 1 Implanted Yes
MR Conditional Lead 2 Implanted Yes

SW033 Software Version 8.3 (4.1) Patient Information Report


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:36:07
Patient Information Report
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

MRI SureScan System/Other Hardware


Other Hardware
Other Devices Unknown
Other Leads Unknown
Lead Extenders/Adaptors Unknown
Other Hardware Notes:

SW033 Software Version 8.3 (4.1) Patient Information Report


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:36:08
Rate Drop Response Episode List
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Rate Drop Response Episodes Summary


Prior to Last Session Since Last Session
12-Oct-2020 to 14-Oct-2020 14-Oct-2020 to 16-Nov-2020
45 hours 33 days
Drop Episodes 0 0
Low Rate Episodes 0 0
Total Episodes 0.0 per day 0.0 per day

Rate Drop Response Episode List: 14-Oct-2020 06:58:33 to 16-Nov-2020 10:01:30


Time Detection Peak
Type ID# Date
hh:mm V. Rate bpm V. Rate bpm
(No data since last session.)
---------------------- Last Programmer Session 14-Oct-2020 ----------------------------------------------------------------------------------------------
(Data prior to last session has not been interrogated.)

SW033 Software Version 8.3 (4.1) Rate Drop Response Episode List
Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:36:46
Final: Session Summary
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Device Information
Device Medtronic Primo MRI DR DDMD3D4 CWN603292S Implanted: 12-Oct-2020
RV 6935M Sprint Quattro MRI
Atrial 5076 CapsureFix Novus MRI
Device Status (Implanted: 12-Oct-2020)
Remaining Longevity 4.2 years (16-Nov-2020)

RRT > 5 years


(based on initial interrogation)

Atrial(5076) RV(6935M)
Pacing Impedance 456 ohms 456 ohms
Defibrillation Impedance RV=64 ohms
Pace Polarity Bipolar Bipolar
In-Office Threshold 0.75 V @ 0.40 ms
Programmed Amplitude/Pulse Width 1.75 V / 0.40 ms 2.00 V / 0.40 ms
Measured P/ R Wave 0.5 mV 4.8 mV
Programmed Sensitivity 0.45 mV 0.60 mV
Parameter Summary
Mode DDDR Lower Rate 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Detection Rates Therapies


AT/AF Monitor >171 bpm All Rx Off
VF On >200 bpm ATP During Charging, 35J x 6
FVT OFF All Rx Off
VT On 167-200 bpm Burst(3), 20J, 35J x 4
Enhancements On: VT Monitor, AF/Afl, Sinus Tach, Wavelet, TWave, Noise(Timeout)
Changes This Session Session Start Current Value
VT Monitor Interval 440 ms (136 bpm) 420 ms (143 bpm)

A. Amplitude 5.00 V 1.75 V


A. Pulse Width 0.60 ms 0.40 ms
RV Amplitude 3.50 V 2.00 V

SW033 Software Version 8.3 (4.1) Final: Session Summary


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:37:24
Final: Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Pacing Summary
Mode Rates AV Intervals
Mode DDDR Lower 50 bpm Paced AV 180 ms
Mode Switch 171 bpm Upper Track 130 bpm Sensed AV 150 ms
Upper Sensor 130 bpm

Pacing Details Atrial RV


Amplitude 1.75 V 2.00 V
Pulse Width 0.40 ms 0.40 ms
Sensitivity 0.45 mV 0.60 mV
Pace Polarity Bipolar Bipolar
Sense Polarity Bipolar Bipolar

AV Therapies Arrhythmia Interventions


Rate Adaptive AV Off A. Rate Stabilization Off
A. Preference Pacing Off
Rate Response
Post Mode Switch Off
ADL Rate 95 bpm
Conducted AF Response Off
Optimization On
V. Rate Stabilization Off
ADL Response 3
Exertion Response 3 Post Shock Pacing
Activity Threshold Low Post Shock Pacing*
Activity Acceleration 30 sec A. Amplitude 4V
Activity Deceleration Exercise A. Pulse Width 1.50 ms
ADL Setpoint 14 V. Amplitude 6V
UR Setpoint 55 V. Pulse Width 1.50 ms
* Settings for first 25 beats post shock
Refractory/Blanking
Post VT/VF Shock Pacing On
PVARP Auto
Overdrive Rate 80 bpm
Minimum PVARP 250 ms
Overdrive Duration 0.5 min
PVAB Interval 150 ms
PVAB Method Partial Additional Features
A. Blank Post AP 200 ms Rate Drop Response Off
A. Blank Post AS 100 ms Sleep Off
V. Blank Post VP 200 ms Non-Comp Atrial Pacing On
V. Blank Post VS 120 ms NCAP Interval 300 ms
MRI SureScan Off
PMT Intervention Off
PVC Response On
V. Safety Pacing On
Device Information
Device Medtronic Primo MRI DR DDMD3D4 CWN603292S Implanted: 12-Oct-2020
RV 6935M Sprint Quattro MRI
Atrial 5076 CapsureFix Novus MRI

SW033 Software Version 8.3 (4.1) Final: Parameters


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:38:03
Final: Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

VT/VF Detection
V. Interval (Rate) Initial Redetect
VF On 300 ms (200 bpm) 30/40 24/32 300 ms
FVT OFF
VT On 360 ms (167 bpm) 36 32 360 ms
Monitor Monitor 420 ms (143 bpm) 40 No Rx 420 ms

PR Logic/Wavelet Other Enhancements Sensitivity


AF/Afl On Stability Off Atrial 0.45 mV
Sinus Tach On Onset Off RV 0.60 mV
Other 1:1 SVTs Off High Rate Timeout
Wavelet On VF Zone Only Off
Template None All Zones Off
Match Threshold 70 % TWave On
Auto Collection On RV Lead Noise On+Timeout
SVT V. Limit 260 ms Timeout 0.75 min
VF Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
VF Therapy Status On On On On On On
Energy 35 J 35 J 35 J 35 J 35 J 35 J
Pathway B>AX AX>B B>AX AX>B B>AX AX>B
ATP During Charging
Deliver ATP if last 8 R-R >= 240 ms,Burst, Pulses = 8, R-S1 = 88 %, Decrement = 10 ms
ChargeSaver = On(1 episodes), SmartMode = On
FVT Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
FVT Therapy Status Off Off Off Off Off Off
VT Therapies Rx1 Rx2 Rx3 Rx4 Rx5 Rx6
VT Therapy Status On On On On On On
Therapy Type Burst CV CV CV CV CV
Energy 20 J 35 J 35 J 35 J 35 J
Pathway B>AX AX>B B>AX B>AX AX>B
Initial # Pulses 8
R-S1 Interval=(%RR) 88 %
S1S2(Ramp+)=(%RR)
S2SN(Ramp+)=(%RR)
Interval Dec 10 ms
# Sequences 3
Smart Mode On
Shared V. ATP Shared V. Therapies
V-V Minimum ATP Interval 200 ms Active Can/SVC Coil Can+SVC On
V. Amplitude 8V Progressive Episode Therapies Off
V. Pulse Width 1.5 ms Confirmation+ On
V. Pace Blanking 240 ms

SW033 Software Version 8.3 (4.1) Final: Parameters


Copyright © Medtronic, Inc. 2015 Page 2 Printed: 16-Nov-2020 10:38:03
Final: Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

AT/AF Detection
Detection A. Interval (Rate)
Monitor
AT/AF 350 ms (171 bpm) No Rx 350 ms
AT/AF Therapies

CV therapies have priority over pacing therapies. Patient Activated CV


Patient Activated CV Status Off
AT/AF ATP Rx1 Rx2 Rx3
AT/AF Rx Status Off Off Off AT/AF Automatic CV Rx4 Rx5
Automatic CV Status Off Off
Shared A. ATP
A-A Minimum ATP Interval 150 ms Shared CV
A. Pacing Amplitude 6V Minimum R-R Interval 500 ms (120 bpm)
A. Pacing Pulse Width 1.5 ms Active Can/SVC Coil Can+SVC On
Stop Atrial Therapies After
Disable all A. Rxs if A. lead position is suspect? No
Atrial Lead Position Check
No measurement since reset.

SW033 Software Version 8.3 (4.1) Final: Parameters


Copyright © Medtronic, Inc. 2015 Page 3 Printed: 16-Nov-2020 10:38:04
Final: Parameters
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Data Collection Setup

Source Range
LECG Can to SVC +/- 2 mV
EGM 1 Atip to Aring +/- 8 mV
EGM 2 (Wavelet) Can to RVcoil +/- 12 mV
EGM 3 RVtip to RVring +/- 8 mV
Monitored EGM1 and EGM3
Pre-arrhythmia EGM Off

Device Date/Time 16-Nov-2020 10:12


Holter Telemetry Off
Medtronic CareAlert Setup
Patient Home Monitor No
Alert Time 08:00
Enable-Urgency
Alert Conditions Tone Threshold
AT/AF Daily Burden Settings Off 6 hr
Avg. V. Rate During AT/AF Settings Off 6 hr at 100 bpm
Number of Shocks Delivered in an Episode On-Low 1
All Therapies in a Zone Exhausted On-Low
RV Lead Integrity On-High
RV Lead Noise On-High
A. Pacing Lead Impedance Out of Range On-High <200 or >3000 ohms
RV Pacing Lead Impedance Out of Range On-High <200 or >3000 ohms
RV Defibrillation Lead Impedance Out of Range On-High <20 or >200 ohms
SVC Defib Lead Impedance Out of Range On-High <20 or >200 ohms
Low Battery Voltage RRT On-High 2.73 V(RRT)
Excessive Charge Time EOS On-High
VF Detection OFF, 3+ VF or 3+ FVT Rx Off. On-High

SW033 Software Version 8.3 (4.1) Final: Parameters


Copyright © Medtronic, Inc. 2015 Page 4 Printed: 16-Nov-2020 10:38:04
Unsaved Frozen Strip
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Collected: 16-Nov-2020 10:19:47 25.0 mm/sec


(1 mV)
A A A A A A
P P P P P P
LECG
V V V V V V
P P P P P P

EGM3

EGM1

EGM2

Lead II

10:19:53 25.0 mm/sec

A A A A A A
P S S S S P

V V V V V V
P P P P P P

10:20:00 25.0 mm/sec End

A
P

V
P

SW033 Software Version 8.3 (4.1) Unsaved Frozen Strip


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:38:42
Unsaved Frozen Strip
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Collected: 16-Nov-2020 10:19:47 25.0 mm/sec


(1 mV)
A A A A A A
P P P P P P
LECG
V V V V V V
P P P P P P

EGM3

EGM1

EGM2

Lead II

10:19:53 25.0 mm/sec

A A A A A A
P S S S S P

V V V V V V
P P P P P P

10:20:00 25.0 mm/sec End

A
P

V
P

SW033 Software Version 8.3 (4.1) Unsaved Frozen Strip


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:39:24
Unsaved Frozen Strip
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Collected: 16-Nov-2020 10:20:35 25.0 mm/sec


(1 mV)
A A A A A A
P P S S R S
LECG
V V V V V V V V
P P P P S S S P

EGM3

EGM1

EGM2

Lead II

10:20:41 25.0 mm/sec

A A A A A A
S S S S S P

V V V V V V
P P P P P P

10:20:48 25.0 mm/sec End

A
P

V
P

SW033 Software Version 8.3 (4.1) Unsaved Frozen Strip


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:40:06
Unsaved Frozen Strip
Device: Primo MRI DR DDMD3D4 Serial Number: CWN603292S Date of Visit: 16-Nov-2020 10:01:30
Patient: Gherman Ioan ID: Physician:

Collected: 16-Nov-2020 10:20:59 25.0 mm/sec


(1 mV)
A A A A A A
P P S S S S
LECG
V V V V V V
P P P P P P

EGM3

EGM1

EGM2

Lead II

10:21:05 25.0 mm/sec

A A A A A A
S S S S S S

V V V V V V
P P P P P P

10:21:12 25.0 mm/sec End

A
S

V
P

SW033 Software Version 8.3 (4.1) Unsaved Frozen Strip


Copyright © Medtronic, Inc. 2015 Page 1 Printed: 16-Nov-2020 10:40:48