Professional Documents
Culture Documents
ZFY Portable Gas Analyzer
ZFY Portable Gas Analyzer
FEATURES
1. Single-beam NDIR system is used for simple structure,
providing excellent stability without interference of other
gases.
2. The adoption of a unique mass-flow type sensor assures
high sensitivity and long service life.
3. One-touch zero/span calibration, self-diagnosis function 8) Sample flow : About 1R/min (pump, filter and flowme-
and 2-range select function assure easy operation and ter, built-in)
maintenance. 9) Analog output :
4. Sampling devices such as pump, filter, flow checker, etc., 4 to 20mA DC (allowable load; less than
are built-in to allow the user to make measurements sim- 550Ω)
ply by connecting a sampling tube. 10)Indicator : 4th code digital display
5. Compact design offers easy carrying and handling. 11)Calibration : One-touch calibration using canned gas
or compressed standard gas (φ6 tube
connection attachment supplied)
12)Warming time :
About 30min
SPECIFICATIONS 13)Gas connecting parts material :
1) Application :Facilities gardening, measurement of SUS, polyethylene, polypropylene, CaF2,
gases in rooms, combustible exhaust gas, toalon, chloroprene
gases in thermal processing furnaces or 14)Power source :
ceramic industry furnaces, and various 100V or 115V AC ± 10%, 50/60Hz, about
tests 40VA (power cord supplied)
2) Measurement principle : 15)Self-diagnosis function :
NDIR (Non-dispersive Infrared-absorption Calibration range over, analyzer error, etc.
method; single-beam system) 16)Sample gas condition :
3) Measuring components and measurement range : Dust; 0 to 10mg/Nm3(filter should be used
Measuring
at the front for continuous measurement)
Measurement components CO2 CO CH4 Temperature; Less than 50˚C at analyzer
range
× × inlet Pressure; –2.94 to +2.94kPa {–300
0 to 0.2/0.5% ™
×
to +300mmH2O}
0 to 0.5/1.0% ™ ™
17)Installation condition :
0 to 1.0/2.0% ™ ™ ™
Temperature; 0˚C to 40˚C
0 to 2.0/5.0% ™ ™ ™
Humidity; 90%RH or less
0 to 5.0/10.0% ™ ™ ™ Installation place; indoor (Do not install
0 to 10.0/20.0% ™ ™ ×
in a place near combustible exhaust gas.
0 to 20.0/50.0% ™ ™ ×
Be sure to install on a flat stand.)
0 to 50.0/100.0% ™ × × 18)Case : Plastic, portable type (with ventilating
4) Repeatability : ± 0.5%FS holes)
5) Drift : ± 2%FS /day for zero and span 19)Dimensions : (H)170 × (W)260 × (D)375mm
6) Linearity : ± 2%FS 20)Mass : Approx. 5kg
7) Response : 90%, less than 15sec
EDS3-117
Date Dec. 18, 1998
ZFY
1 2 3 4 5 6 7 8 9 code
Z F Y 1 - Description
Measuring components (4th code)
B CO O+ O– E
D CO2
E CH4
Measurement range (5th code)
G 0~0.2 / 0.5%(Not available for CO and CH4) Output signal
(4 to 20mA DC)
H 0~0.5 / 1.0%(Not available for CH4)
J 0~1.0 / 2.0%
K 0~2.0 / 5.0%
L 0~5.0 / 10.0%
M
N
0~10.0 / 20.0%(Not available for CH4)
0~20.0 / 50.0%(Not available for CH4)
Scope of delivery
P 0~50.0 / 100.0%(Not available for CO and CH4)
Main body 1
Output signal (6th code)
1 4-20mA DC Power cord 1
Power source (7th code) Power fuse (tube type 1A, φ6.4 × 30) 1
1 100V AC 50/60Hz Filter paper 25
2 115V AC 50/60Hz Calibration gas pipe joint 1
Application (9th code) Main body vinyl cover 1
A General use combustible exhaust gas,
B
( greenhouse,
atmospheric gas
For heat processing furnace gas,
) φ9/φ5 Toalon tube, 5cm (for connection of φ6 tube)
Hose band (for φ10)
1
2
Power switch
Rubber foot
260 375
about 60
Membrane filter
Fuse holder
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.