Professional Documents
Culture Documents
RISONIC Modular System Manual E22 730 0085221 001 01
RISONIC Modular System Manual E22 730 0085221 001 01
RISONIC Modular System Manual E22 730 0085221 001 01
System Manual
RITTMEYER Headquarters
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Contents
1. Safety.......................................................................................................................... 11
1.1. Safety information in these operating instructions ......................................................................... 11
1.2. For your safety ............................................................................................................................... 11
1.3. Intended use ................................................................................................................................... 12
1.4. Transport ........................................................................................................................................ 12
1.5. Configuration assistance ................................................................................................................ 13
1.6. Safety regulations, protective measures ........................................................................................ 13
1.7. Information on environmental behavior and disposal .................................................................... 13
1.8. Valid standards and directives ....................................................................................................... 14
2. About this guide ........................................................................................................ 15
2.1. Installation requirements ................................................................................................................ 15
2.2. Tasks during installation and configuration .................................................................................... 15
2.3. Finding specific information ............................................................................................................ 17
3. Components .............................................................................................................. 19
3.1. RICTRL........................................................................................................................................... 19
3.2. RIMOUSTT ..................................................................................................................................... 26
3.3. Extensions ...................................................................................................................................... 30
3.4. Accessories .................................................................................................................................... 30
4. Measurement arrangements and setups ................................................................. 31
4.1. Limitations ...................................................................................................................................... 31
4.2. Paths, planes, and measurement layers ........................................................................................ 31
4.3. Path measurement sequence ........................................................................................................ 32
4.4. Path replacement strategy ............................................................................................................. 33
4.5. Filled pipe (PIPE) ........................................................................................................................... 34
4.6. Open channel (CHANNEL) ............................................................................................................ 37
4.7. Partially filled pipe (PIPE-CHANNEL) ............................................................................................ 39
4.8. Clamp On (CLAMPON) .................................................................................................................. 41
5. Ethernet network setup ............................................................................................. 43
5.1. Microsoft Windows 7 ...................................................................................................................... 43
5.2. Microsoft Windows 10 .................................................................................................................... 45
5.3. Apple Mac OS X (e.g. 10.11.x or newer) ....................................................................................... 47
5.4. DHCP ............................................................................................................................................. 48
6. Installation ................................................................................................................. 51
6.1. General notes ................................................................................................................................. 51
6.2. Installation of a RISONIC modular flow measurement .................................................................. 52
6.3. Hardware installation ...................................................................................................................... 52
6.4. Default network configuration ......................................................................................................... 54
6.5. Accessing the RISONIC modular web interface ............................................................................ 55
6.6. Registration .................................................................................................................................... 57
6.7. General guidelines for coaxial / triaxial cabling .............................................................................. 57
7. Configuration in STANDARD Mode ......................................................................... 69
7.1. Introduction ..................................................................................................................................... 69
7.2. Standard application filled pipe (single path, e.g. 1E1P) ............................................................... 70
7.3. Standard application filled pipe (multiple paths, e.g. 2E8P) ........................................................... 73
7.4. Standard application open channel (multiple paths, e.g. 2E8P) .................................................... 76
7.5. Standard application partially-filled pipe (multiple paths, e.g. 2E8P) ............................................. 78
7.6. Standard application Clamp-On (single path, e.g. 1E1P) .............................................................. 78
7.7. Parent Section A, intelligent application ......................................................................................... 80
7.8. Status page .................................................................................................................................... 82
7.9. STANDARD Mode Measurement Parameters ............................................................................... 88
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
List of figures
Figure 1: RICTRL front .................................................................................................................................... 20
Figure 2: RICTRL bottom................................................................................................................................. 20
Figure 3: RICTRL top ....................................................................................................................................... 20
Figure 4: RICTRL back .................................................................................................................................... 21
Figure 5: RIMOUSTT front............................................................................................................................... 26
Figure 6:RIMOUSTT top .................................................................................................................................. 26
Figure 7: RIMOUSTT back .............................................................................................................................. 27
Figure 8: Multi pipe and section setup with a single RICTRL .......................................................................... 31
Figure 9: Path numbers for filled pipe, e.g. circular 2E8P ............................................................................... 36
Figure 10: Path numbers for open channel, e.g. trapezoidal 2E10P .............................................................. 39
Figure 11: Clamp On calculation methods ...................................................................................................... 41
Figure 12: Hardware installation (two RIMOUSTT modules) .......................................................................... 52
Figure 13: Typical hardware wiring diagram.................................................................................................... 53
Figure 14: RICTRL status page ....................................................................................................................... 56
Figure 15: STANDARD Mode Measurement Parameters (empty configuration) ............................................ 70
Figure 16: Typical setup and geometry ........................................................................................................... 71
Figure 17: RISONIC modular system status, pipe 1E1P................................................................................. 73
Figure 18: RISONIC modular system status, pipe 2E8P ................................................................................. 76
Figure 19: RISONIC modular system status, open channel 2E8P .................................................................. 78
Figure 20: RISONIC modular clamp-on parameters ....................................................................................... 79
Figure 21: RISONIC modular with two sections (pipes/penstocks) ................................................................. 81
Figure 22: Total flow and volume application for two pipes/penstocks ........................................................... 81
Figure 23: System status (system information) ............................................................................................... 82
Figure 24: System status, section information ................................................................................................ 84
Figure 25: STANDARD Mode Measurement Parameters (Sections) ............................................................. 89
Figure 26: STANDARD Mode Measurement Parameters (General) .............................................................. 90
Figure 27: STANDARD Mode Measurement Parameters (filled pipe geometries) ......................................... 91
Figure 28: STANDARD Mode Measurement Parameters (Clamp On geometry) ........................................... 91
Figure 29: STANDARD Mode Measurement Parameters (open channel / partially filled pipe geometries) ... 92
Figure 30: STANDARD Mode Measurement Parameters (open channel geometry USER-DEFINED) ......... 93
Figure 31: STANDARD Mode Measurement Parameters (wetted channel wall) ............................................ 93
Figure 32: STANDARD Mode Measurement Parameters (flow cutoff) ........................................................... 95
Figure 33: STANDARD Mode Measurement Parameters (transducer offset, cable length) ........................... 96
Figure 34: STANDARD Mode Measurement Parameters (flow correction) .................................................... 97
Figure 35: STANDARD Mode Measurement Parameters (normal discharge) ................................................ 98
Figure 36: STANDARD Mode Measurement Parameters (Ultrasonic Modules)........................................... 100
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Figure 75: EXPERT Mode Measurement Parameters (Limit values) ............................................................ 169
Figure 76: EXPERT Mode Measurement Parameters (lower/upper flow limit values).................................. 170
Figure 77: EXPERT Mode Measurement Parameters (maximum flow change limit value) .......................... 170
Figure 78: EXPERT Mode Measurement Parameters (Outlier Filter) ........................................................... 172
Figure 79: EXPERT Mode Measurement Parameters (outlier filter path input buffers) ................................ 173
Figure 80: EXPERT Mode Measurement Parameters (outlier filters for each path) ..................................... 173
Figure 81: EXPERT Mode Measurement Parameters (Layer) ...................................................................... 174
Figure 82: EXPERT Mode Measurement Parameters PIPE CIRCLE (Path) ................................................ 175
Figure 83: EXPERT Mode Measurement Parameters PIPE RECTANGLE (Path) ....................................... 175
Figure 84: EXPERT Mode Measurement Parameters (PIPE-)CHANNEL CIRCLE (Path) ........................... 176
Figure 85: EXPERT Mode Measurement Parameters (PIPE-)CHANNEL RECTANGLE (Path) .................. 177
Figure 86: EXPERT Mode Measurement Parameters CHANNEL TRAPEZOID (Path) ............................... 177
Figure 87: EXPERT Mode Measurement Parameters CHANNEL USER-DEFINED (Path) ......................... 178
Figure 88: EXPERT Mode Measurement Parameters (switch on-off levels) ................................................ 180
Figure 89: IEC 41 integration weights (1991-11, third edition) ...................................................................... 180
Figure 90: IEC 41 integration weights (2015, draft) ....................................................................................... 181
Figure 91: Syslog ........................................................................................................................................... 182
Figure 92: Service .......................................................................................................................................... 184
Figure 93: MMI menu map............................................................................................................................. 185
Figure 94: Receive signal with small reflection ............................................................................................. 190
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
1. Safety
1.1. Safety information in these operating instructions
The following safety symbols and notes are used throughout this manual:
Note
Important information
Caution
Caution
1. Before installing the equipment, please carefully read through the following safety
instructions. Failure to observe the following rules can have severe consequences, be
contrary to regulations, or lead to equipment failure.
2. Rittmeyer assumes no liability for unprofessional handling of the equipment or for
consequential damage as a result of ignoring the following precautionary measures.
MANUAL 22.730.0085221.001.01.4.4
Note
Waterproof cases providing mounting space for several modules, power supply, etc. are
available on request.
1.4. Transport
The RISONIC modular components are delivered in special packaging by which the equipment is effectively
protected against electrostatic discharge and vibration as well as the effect of shock. Save this packaging in
a safe and dry place so that you can transport the equipment safely for expansion or repair purposes.
On receipt of the delivery, please check for transport damage to packaging and/or modules. In the event of
transport damage, a functional test should be carried out on the modules.
MANUAL 22.730.0085221.001.01.4.4
The ultrasonic transducers are operated with short pulses of high frequency voltages
exceeding 120VAC. Although it is unlikely that this voltage is generated continuously due to
equipment failure, the ultrasonic transducer terminals must not be touched!
The ultrasonic transducers are operated with short pulses of high frequency voltages
exceeding 120VAC. The ultrasonic transducer terminals must not be touched!
Caution
1. All modules contain highly integrated electronic circuits; shock or vibration must be
prevented during transport, installation and operation.
2. All modules are housed in steel enclosures which may have sharp edges; handle with
care to prevent injury.
MANUAL 22.730.0085221.001.01.4.4
The devices meet the requirements for UL2 certification according to:
QUYX Process Control Equipment, Electrical – Component
QUYX7 Process Control Equipment, Electrical Certified for Canada -
Component
(See also online certificates: QUYX7.E326219 - Process Control Equipment, Electrical Certified for Canada)
MANUAL 22.730.0085221.001.01.4.4
Note
The device shall be supplied with a power supply with double or reinforced insulation and
24 VDC - 20 % / + 25 % (SELV)!
MANUAL 22.730.0085221.001.01.4.4
Note
The shipped RICTRL may contain a more recent software release than this manual is based
on. Therefore, if the software release differs from the information in this manual, follow also
the information provided in the software datasheet and release notes.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
3. Components
A RISONIC modular application needs two different types of electronic modules and transducers:
RICTRL: Rittmeyer Instrumentation Controller
RIMOUSTT: RISONIC Ultrasonic Transit Time
RISONIC transducers: a variety of ultrasonic transducers of type MFATx are available, please refer
to the corresponding datasheets and/or Rittmeyer representative for further information (not covered
in the manual)
3.1. RICTRL
The Rittmeyer Instrumentation Controller RICTRL is the system controller running and supervising the
measurement. It interfaces with other control systems and auxiliary equipment, such as pressure transmitters
and additional display units, data recorders, etc.
The following pictures show the RICTRL module and its interfaces in more detail.
MANUAL 22.730.0085221.001.01.4.4
Legend:
1. Status LED
3 2. Sleep LED
12
3. LCD
MANUAL 22.730.0085221.001.01.4.4
25
3.1.1.1. Displays
MANUAL 22.730.0085221.001.01.4.4
3.1.1.2. Controls
MANUAL 22.730.0085221.001.01.4.4
Caution
Pressing the ”Reset“ button immediately sets the analogue and digital outputs to their
defined watchdog states, sets the STATUS OUTPUT into an error state and reboots the
RICTRL!
Caution
Pressing the ”Factory Reset“ button for at least 5 seconds deletes all user defined data
(confirmation needed, please refer to MMI) and sets the RICTRL back to factory defaults!
3.1.1.3. Inputs/outputs
MANUAL 22.730.0085221.001.01.4.4
Saturation ON:
o Value < 3.8 mA Output 3.8 mA
o Value > 20.5 mA Output 20.5 mA
DO 1 … 5: Relay outputs, switch-over contact, contact load 2 A
/ 30 VDC nominal.
Galvanically isolated
Refresh rate: 1 Hz
User-defined pulse duration (100ms, 200ms, 1s)
Floating average filtering (off, 5, 10, 15, 20, 30, 45,
60, 120s)
Measured value output (per measuring point):
Depending on measurement type
Status output (per measuring point):
Warning: measurement section
Error: measurement section
DI 1: Digital optocoupler input
Galvanically isolated
Refresh rate: 1 Hz
Digital value input (per measuring point):
Depending on measurement type
User-selectable as counter or logic input
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
3.2. RIMOUSTT
The RIMOUSTT module connects to the ultrasonic transducers, carries out the fast ultrasonic transit time
measurements, and interfaces with auxiliary equipment, such as pressure transmitters and additional display
units, data recorders, etc.
The following pictures show the RIMOUSTT module and its interfaces in more detail.
Legend:
2
1. Status LED
2. DO 1 … DO 4 LEDs
1
3. PATH A … PATH D LEDs
7 7. ID switch
8. 50-60Hz switch
9
9. LAN2 Ethernet connector
4 10. Removable cover, R1 … R8 connectors
3 11. EARTH connector (functional earth,
removable cover)
12. 24VDC IN connector (galvanically isolated,
functional earth)
10
11
13. DO 1 … DO 4 connectors (digital outputs,
galvanically isolated)
14. AO connector (analogue output, galvanically
isolated)
15. AI connector (analogue input, galvanically
isolated)
12 13 14 8 15
MANUAL 22.730.0085221.001.01.4.4
16
3.2.1.1. Displays
MANUAL 22.730.0085221.001.01.4.4
3.2.1.2. Controls
Caution
Pressing the “Reset“ button immediately sets the analogue and digital outputs to their
defined watchdog states and reboots the RIMOUSTT!
3.2.1.3. Inputs/outputs
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
3.3. Extensions
Neither the RICTRL nor the RIMOUSTT modules are extendable with additional (internal) components.
3.4. Accessories
For a list of available accessories please refer to the hardware datasheets and / or contact a Rittmeyer AG
representative.
MANUAL 22.730.0085221.001.01.4.4
A typical setup with two penstocks and two measurements on both on top and on the bottom of the
penstocks is easily possible with a single RICTRL:
Note
MANUAL 22.730.0085221.001.01.4.4
1. Logical velocity vectors (axial and perpendicular i.e. in 90° cross to the flow direction) of one or two
crossed paths which are installed at the same height are calculated, i. e. two paths are grouped
together.
2. Subsequent calculations use these logical velocity vectors to calculate the flow and volume in
forward and reverse directions.
3. With these velocity vectors, i.e. horizontal measurement layers, much more simplified and
deterministic path replacement algorithms are possible.
The following figures show this new concept consisting of paths and vertical planes. As an example, the
measurement setup “filled pipe, circular cross section, and 2E8P” is shown below:
These internal details of planes and layer concepts are usually quite well hidden from the user as they are
not important for the typical application.
However, in EXPERT Mode much more detailed information about these horizontal measurement layers are
available (especially for troubleshooting purposes, configuration and diagnostics).
MANUAL 22.730.0085221.001.01.4.4
R8
R16
v
4 R6
R15 8
3
R7 R4 R14
7
R13 2 R12
6
R5 R2
R11 1
5 R10 y
R3
R9
R1
x
z
non-crossed measurements 1ExP: the system mirrors on the middle horizontal layer of the cross
section geometry i.e. the mirrored path will be used.
Example 1E4P: lowest path R1-R2 fails velocity of the top most path R7-R8 is used instead of
the failed path:
R8 v
R6
4
R7 R4
3
2
R5 R2
1 y
R3
R1
x
z
With several path failures, combinations of the strategy mentioned above are used; the system keeps trying
to measure the flow until no working/valid paths are available. With this, e.g. a 2E8P measurement may
change into a 1E4P measurement.
Typically, at least one path velocity must be available per horizontal layer (original or mirrored), otherwise no
measurement is possible. This results in a change of the section status into error state.
Depending on the chosen measurement setup and the effective replacement strategy, the system may
change the section calculation method automatically. Refer to chapter 7.8.2.2 for a list of possible methods.
MANUAL 22.730.0085221.001.01.4.4
For special applications (very much silt or air content), a section replacement strategy is available too: with
this, a 2E8P measurement may change into a 1E2P measurement setup with degraded accuracy, still
delivering velocity and even flow information as long as possible!
MANUAL 22.730.0085221.001.01.4.4
Note
Path arrangements with odd numbers (e.g. 1E5P or 2E10P) are also supported. Because of
the need for special angles of the transducers they are only available on request.
MANUAL 22.730.0085221.001.01.4.4
R7 R15
R16 R8
P4
R5 P8
R13
P3 P7
R14
R6
P2
P6
R3
R11
P1
P5
R12
R4
R1 R9
R10
R2
MANUAL 22.730.0085221.001.01.4.4
Note
Note
MANUAL 22.730.0085221.001.01.4.4
Note
Please note: the path numbering scheme for open channel measurement setups is
different than for filled-pipe measurement setups, due to historic reasons!
The paths are numbered in the following way, examples with 5 and 10 paths:
- One vertical plane / non-crossed setup, e.g. 1E5P:
o Lowermost path: R1 – R2
o Second-lowest path: R3 – R4
o Middle path: R5 – R6
o Second-highest path: R7 – R8
o Topmost path: R9 – R10
- Two vertical planes / crossed setup, e.g. 2E10P:
o First vertical plane:
Lowermost path: R1 – R2
Second-lowest path: R5 – R6
Middle path: R9 – R10
Second-highest path: R13 – R14
Topmost path: R17 – R18
o Second vertical plane:
Lowermost path: R3 – R4
Second-lowest path: R7 – R8
Middle path: R11 – R12
Second-highest path: R15 – R16
Topmost path: R19 – R20
The following figure shows a typical open channel 10-path measurement 2E10P, trapezoidal geometry, and
the path numbers:
MANUAL 22.730.0085221.001.01.4.4
R17 R19
P9
P7 P10
R13 R15
P5
P8
R9 R11
P6
R5 R7
P4
R3
R1
P2
R20 R18
R16 R14
R12 R10
R8 P3 R6
P1
R4 R2
Figure 10: Path numbers for open channel, e.g. trapezoidal 2E10P
MANUAL 22.730.0085221.001.01.4.4
Note
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Select the network connection (i.e. ”LAN Connection“) and click on ”Change adapter settings“:
MANUAL 22.730.0085221.001.01.4.4
Select ”Properties“:
Select ”Use the following IP address“ and enter ”192.168.1.10“ as IP address and ”255.255.255.0“ as subnet
mask:
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Right-click on the corresponding network connection (i.e. ”Local Area Connection“) and select ”Properties“:
Select ”Use the following IP address“ and enter ”192.168.1.10“ as IP address and ”255.255.255.0“ as subnet
mask:
MANUAL 22.730.0085221.001.01.4.4
In the adapter list on the left select the connected Ethernet adapter and select “Manually” in the “Configure
IPv4” dropdown list:
Enter ”192.168.1.10“ as IP address and ”255.255.255.0“ as subnet mask and click ”Apply“.
MANUAL 22.730.0085221.001.01.4.4
5.4. DHCP
Starting from RICTRL-SW release 3.4.4, the RICTRL supports a DHCP client mode: this mode allows a
RICTRL to automatically find an IP address, in case a DHCP service is available in the network or a suitable
computer is directly connected to the Instrumentation Controller (crossed Ethernet cable).
The received IP address is shown on the MMI or can be found in the MMI menu, if measurement values are
shown on the MMI.
The DHCP client mode can be activated with the Mode rotary switch setting 1.
The above mentioned software version supports additional modes, please refer to the following table for a
complete list of all supported modes:
Mode rotary switch (after changing the mode setting 0: Measurement mode, factory default:
a reboot / reset must be done, in order to apply the
changed setting): Static IP address
SW-Update via Web-Interface activated
1: Measurement mode, DHCP client / dynamic IP
address:
DHCP client / dynamic IP address
SW-Update via Web-Interface activated
2: Measurement mode, no remote software update:
Static IP address
SW-Update via Web-Interface deactivated
3: Measurement mode, fixed access parameters:
Static IP address
SW-Update via Web-Interface deactivated
Web interface: no factory reset and/or
password change possible
B: software update by USB memory stick
C: software update by CF card
D: diagnostics mode, allows the firmware update of
connected RIMOUSTT modules
E: Bootrom update
F: FTP server
MANUAL 22.730.0085221.001.01.4.4
Note
The DHCP client mode is very helpful for first contact / commissioning of a RICTRL; it is not
advised to leave the RICTRL in DHCP mode in productive environments (without stable
DHCP services) as the Ethernet connection for Modbus TCP and IEC 60870-5-104 may
become unstable.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
6. Installation
6.1. General notes
Connections marked as “Functional Earth”, symbol , must be connected before all other
connections!
Caution
The wires and cables used for the connection of the RICTRL, RIMOUSTT and RIPDIx.xxx
must be suitable for an ambient temperature of min. 80 °C!
Note
Provide an external switch or circuit breaker to fully disconnect the device from the mains
power supply.
This switch or circuit breaker shall be located near the device, easily accessible and
marked/labeled as a separator from the mains power supply.
The device must be installed in such a way that handling of the switch or circuit breaker is
not hindered.
Note
It is recommended to test the system in-house before setting up the (outdoor) installation,
RISONIC measurement types preferably with a QSIModular (simulation of the ultrasonic
transducers) and RIPRESS measurement types with a 4 … 20 mA current source simulator!
MANUAL 22.730.0085221.001.01.4.4
Connections marked as “Functional Earth”, symbol , must be connected before all other
connections!
MANUAL 22.730.0085221.001.01.4.4
24VDC IN
24VDC IN
Input
24VDC
1. Connect the RICTRL power input (24VDC IN) to a suitable 24VDC power source / power supply
(power supply with double or reinforced insulation and 24VDC - 20% / +25%, SELV)
2. Connect the RICTRL power output (24VDC OUT) to the RIMOUSTT power inputs (24VDC IN,
RIMOUSTT ID0 and ID1)
3. Connect the Ethernet switch power input (24VDC IN) to a suitable 24VDC power source / power
supply
4. Connect the RIMOUSTT PATH connectors to either ultrasonic transducers or a QSIModular
5. Connect the RIMOUSTT EARTH connector with short and low impedance wires to a valid ground
connection (e. g. 16mm2)
6. Connect the RICTRL Ethernet interface LAN1 to a PC or laptop with a crossed Ethernet cable (RJ-
45)
7. Connect the RICTRL Ethernet interface LAN2 with a crossed Ethernet cable (RJ-45) to the Ethernet
switch
8. Connect the RIMOUSTT Ethernet interfaces LAN2 with crossed Ethernet cables (RJ-45) to the
Ethernet switch (RIMOUSTT ID0 and ID1)
9. Power on the 24VDC power supply
10. Check the RICTRL STATUS LED and LCD/MMI to ensure that the module is correctly powered and
starting. The RIMOUSTT modules are not powered until a flow measurement is configured; correct
configuration of their rotary switches ‘ID’ is essential (see next chapter)!
11. After a successful startup, the LCD/MMI on the RICTRL displays “RISONIC modular” and the date
and time of its real-time clock4
12. Check the network LEDs of LAN1 to ensure a basic network connection (i. e. correct network cabling
of the modules) is available
4 The hardware real-time clock is backed up with an internal capacitor; backup time is app. 4 weeks. If power
is disconnected for more than 4 weeks, the clock will reset to its default date+time settings.
201707 Hir, Ges Subject to change Page 53/198
MANUAL 22.730.0085221.001.01.4.4
Note
Caution
Duplicate IP addresses mustn’t be used and cause unpredictable results: they lead to a non-
functioning system setup!
Note
The LAN2 IP address of RIMOUSTT consists of the base address in Table 5 plus the setting
of the ID rotary switch on the front of the RIMOUSTT; the ID rotary switch defines an
additional offset to the base address (0…F +0…+15). Example: ID=1 RIMOUSTT
LAN2 IP address = 10.0.0.100 + 1 = 10.0.0.101
By default, both RIMOUSTT modules are set to the same ID, thus resulting in the same Ethernet IP address.
For a correct network setup change the ID switch on one RIMOUSTT (e. g. the second one) to ‘1’. The
following network settings apply:
MANUAL 22.730.0085221.001.01.4.4
1. Ensure that the Ethernet interface of the PC/Laptop is configured in the same subnet as RICTRL
LAN1, e.g. IP address 192.168.1.10 with a subnet mask of 255.255.255.0 (default setting). If the
physical connection is working properly the LEDs of RICTRL LAN1 should light up and blink
randomly/with network traffic.
2. Launch a web browser (Microsoft Internet Explorer, Mozilla Firefox, Safari or Opera are examples of
commonly used web browsers). In the address line enter the IP address that is assigned to RICTRL
LAN1, for example http://192.168.1.1 (default setting).
3. The RICTRL status page should appear immediately as shown below. The status, datalog and trend
pages are accessible without further authorization; navigation through all pages is possible using the
menu on the left hand side.
4. In order to configure the system please select “Login” on the menu. A new page opens requesting
login authorization. When prompted, enter the default password “1234” (without quotes), and press
the “Login” button.
5. Once logged in, the main menu on the left splits into two main sections: STANDARD and EXPERT.
6. From this point on, it is possible to change all system and measurement parameters and settings.
Usually, the changes take effect immediately after selecting “Save” without rebooting the RICTRL
(except for some system-wide settings like Network and Unit Settings which need a reboot of the
RICTRL).
MANUAL 22.730.0085221.001.01.4.4
Note
If a RICTRL is not configured the system status will display “OK (MEASUREMENT
STOPPED)” – this is not a fault but by design as there is no working configuration available
yet!
Note
Inactive web sessions are logged out automatically - typically after 10 minutes - and
unconfirmed / unsaved settings will be lost!
Do not forget to save your changes!
MANUAL 22.730.0085221.001.01.4.4
6.6. Registration
The web interface of a new device (or when the settings have been reset to factory defaults) shows a small
registration note at the bottom of the browser windows, please refer to Figure 14: RICTRL status page.
The registration is free, of course; but it allows a better support in case of problem or warranty issues and a
customer-oriented improvement of the hard- and software.
For this the installed MTA6 is used for sending an Email with the following information:
Subject: „Product registration of Controller SN: 00:17:96:nn:nn:nn“
Text: „Please attach updated: “
o Contact information
o System report (EXPERT Mode -> Service -> System report -> Generate)
o Parameter export (EXPERT Mode -> Service -> Parameter export/import -> Export)
The customer/user should add the requested contact information, a current system report and a current
parameter export.
Please note that these files should be added at the end of a successful commissioning, as soon as the
installation has been finished and is working properly, such that they contain up-to-date information.
These data are treated confidentially and only in the context of the corresponding customer relationship!
Note
The customer gets one (1) additional year of warranty for the registered RICTRL module!
MANUAL 22.730.0085221.001.01.4.4
A Shield (screen) connecting system is used near the RISONIC modular RIMOUSTT module to
connect the outer shield to protective earth (i.e. in the field cabinet, where the cable comes into the
cabinet)
Overview of the connection of the different cable types, e.g. with the transducer MFATBx:
Matching print
Piezo element
(galvanic isolation)
Coaxial cable
Inner conductor
Shield
Matching print
Piezo element
(galvanic isolation)
Triaxial cable
Inner conductor
Inner shield
Outer shield
(earth connection
on field frame) Earth connection to
transducer body
MANUAL 22.730.0085221.001.01.4.4
Overview of the protective earth connection of the different cable types in the field cabinet:
Triaxial cables
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Caution
For safety reasons, the cover must always be mounted and fastened correctly!
MANUAL 22.730.0085221.001.01.4.4
These Shield (screen) clamping saddles must be mounted near the cable glands of the field cabinet, by
usage of a low-resistance connection to protective earth.
1. Uncover the outer shield in two places (distance depends on the location of the Shield (screen)
clamping saddles):
2. Near the RISONIC modular the outer shield is not used and must be cut shortly (and MUST NOT be
connected to the RIMOUSTT module!!!). At the end of the cable, uncover the inner shield for approx.
13 mm:
MANUAL 22.730.0085221.001.01.4.4
5. To prevent short circuits: isolate the end of the outer shield with a short piece (e.g. 1.5 cm) of heat
shrink tube:
MANUAL 22.730.0085221.001.01.4.4
7. Connect outer shield to protective earth (with the use of Shield (screen) clamping saddles):
Caution
For safety reasons, the cover must always be mounted and fastened correctly!
MANUAL 22.730.0085221.001.01.4.4
Standard cable gland, outer view EMV-type cable gland, outer view
Standard cable gland, inner view EMV-type cable gland, inner view
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Picture of correctly assembled EMC-type cable gland (attention: the inner shield must not get into contact
with the cable gland):
MANUAL 22.730.0085221.001.01.4.4
Bestandteile: Doppelnippel (A), Feder (B), Globe-Dichteinsatz (C), Vorbereitung: Leitung abmanteln, Geflecht mit Isolierband schützen
Druckschraube (D)
Preparation: Dismantle wire, protect braid below
Components: Double nipple (A), spring (B), globe-sealing insert (C),
pressure screw (D)
Kabel mit leichter Drehung einführen Markieren, wenn der Kabelmantel die Feder berührt
Install cable with slight turn Mark when cable sheath touches spring
Kabel gemäß Maß a zurückziehen (siehe Tabelle unten) Druckschraube mit Drehmoment festziehen (siehe Tabelle unten)
Withdraw cable acc. size a (see table below) Fix pressure screw with nominal torque (see table below)
Tabelle
Table
Art.-Nr. a Anzugsdrehmoment E
Art.-No. mm Nominal torque mm
MANUAL 22.730.0085221.001.01.4.4
In STANDARD Mode, the Measurement Parameters are divided into separate groups:
(Measurement) Sections:
o Section A (if license is valid, otherwise not shown): intelligent application section, can be
used for higher-level calculations, e.g.
Total flow / sum of several pipe flows
PLDS, simplified penstock leakage detection system
…
o Section 1 … 4: up to 4 measurement sections, e.g. RISONIC modular
Ultrasonic Modules (RIMOUSTT):
o Search: automatically detects connected RIMOUSTT modules (and paths which are
connected to these modules)
Note: needs at least one RISONIC modular section, otherwise this section is not shown
o RIMOUSTT ID 0 … 4: up to 5 RIMOUSTT modules. Each of them can include up to 4 paths,
resulting in a total path number of 20!
Note
The section A (when shown) cannot be used for RISONIC measurement applications; use
section 1 … 4 instead!
Section A is for superordinate measurement setups (intelligent applications), accessing
measurement and status values/information from section 1…4.
MANUAL 22.730.0085221.001.01.4.4
Note
The Ultrasonic Modules (RIMOUSTT) group is folded (i.e. hidden) as soon as there is at
least one module configured. Unfold the group with the sign to view all modules and
paths!
MANUAL 22.730.0085221.001.01.4.4
L_C1
T2
L_
phi
L
L_T = L_T1 + L_T2
phi
T1
L_
L_C2
Usually, only the following few tasks are needed to configure a working flow measurement:
Note
Text marked with a rectangle such as Add … refer to the corresponding button on the web
interface!
MANUAL 22.730.0085221.001.01.4.4
Note
Because the added section isn’t fully configured yet an error message appears:
PATH 1 : Not mapped to a RIMOUSTT ultrasonic path!
5. Edit the section parameters (at least the following parameters, sample values) and confirm; for
explanation sketches are shown with the most important parameters:
Section 1 General Edit ...
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Note
Because the added section isn’t fully configured yet, error messages will appear due to non-
configured paths:
5. Edit the section parameters (at least the following parameters) and confirm them by selecting “OK”:
Section 1 General Edit...
MANUAL 22.730.0085221.001.01.4.4
o RIMOUSTT PATH (the ID of Ultrasonic Module defined above and one of its
paths, i.e. the mapping between them and this section), e.g. RIMOUSTT PATH
= RIMOUSTT ID 0 – PATH A
o SOUND PATH LENGTH L depending on geometry, e.g. L = 0.83 m
o TRANSDUCER OFFSET L_T in m (enter the sum of both L_T1 and L_T2;
usually L_T is a negative number), e.g. L_T = -0.08 m
o ANGLE PHI phi in degrees, e.g. phi = 45.000 °
o CABLE LENGTH L_C in m (enter the average of both cable lengths L_C1 and
L_C2), e.g. L_C = 20.000 m
- Confirm OK
6. Save all changes:
Save
After saving the changes, the flow measurement will start when all parameters are reasonably configured.
If the flow measurement is working properly, the web interface will show the flow measurement values on the
status page:
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
4. Add a section of a defined type (e. g. OPEN CHANNEL, 2E08P, TRAPEZOIDAL, MFATK2) and
confirm:
Sections Add…
- Measurement Type OPEN CHANNEL Next
- Configuration 2E08P Next
- Shape TRAPEZOIDAL Next
- Transducer Type MFATK2 OK
5. Edit the section parameters (at least the following parameters) and confirm them by selecting “OK”:
Section 1 General Edit...
o RIMOUSTT PATH (the ID of Ultrasonic Module defined above and one of its
paths, i.e. the mapping between them and this section), e.g. RIMOUSTT PATH
= RIMOUSTT ID 0 – PATH A
o SOUND PATH LENGTH L depending on geometry, e.g. L = 0.83 m
o TRANSDUCER OFFSET L_T in m (enter the sum of both L_T1 and L_T2;
usually L_T is a negative number), e.g. L_T = -0.08 m
o ANGLE PHI phi in degrees, e.g. phi = 45.000 °
o CABLE LENGTH L_C in m (enter the average of both cable lengths L_C1 and
L_C2), e.g. L_C = 20.000 m
- Confirm OK
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Optional outer coating with thickness and material parameters (speed of sound), t_l1 and c_l1
Pipe wall thickness and pipe wall material parameters (speed of sound), t_l2 and c_l2
Optional inner coating with thickness and material parameters (speed of sound), t_l3 and c_l3
The web interface shows a sample sketch of a typical pipe including inner and outer coatings:
MANUAL 22.730.0085221.001.01.4.4
Note
Special processing rules are available for the correct handling of large volume counts as the
typically used 32-bit floating point number format is not sufficient (loss of accuracy with very
large volumes).
Several applications notes are available for these enhanced applications, describing the application and
configuration in details:
E22_260_0067760_105_01 Application note RISONIC simplified pipe rupture monitoring (PLDS)
E22_260_0067760_101_01 Application note RISONIC total-flow-volume calculation
E22_260_0067760_106_01 Application note RISONIC silt monitoring
E21_260_0067760_105_01 Application note Winter-Kennedy
E22_260_0067760_107_01 Application note RISONIC volume counter with offset
E21_260_0067760_102_01 Application note RICTRL digital input
E21_260_0067760_103_01 Application note generic Modbus driver
E21_260_0067760_104_01 Application note SMS alarms
MANUAL 22.730.0085221.001.01.4.4
Q1 Q2
P1 P2
Figure 22: Total flow and volume application for two pipes/penstocks
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Value Definition
System Status Reflects the state of the system:
OK: the RICTRL and all configured (RIMOUSTT) modules and devices (e.g.
Modbus, GSM modem etc.) operate without errors
ERROR; possible causes:
o Communication malfunction between RICTRL and one or more
modules and/or devices (e.g. RIMOUSTT module not available)
o Internal device error
o Parameter error
System Time Current system date and time (internal hardware real-time clock)
System Uptime System uptime since last power-on or reboot
Automatic Allows refreshing the status page automatically; default = OFF. Can be set to 5, 10,
Refresh 20, or 30 seconds
Note
The output “STATUS OUT” on a RICTRL directly corresponds to the System Status
(OK / ERROR)!
Note
In case of system errors, it is advised to check the Syslog for additional information!
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Path Min. number of Filled Pipe Open channel and partially filled pipe (in channel mode i.e. not
setup avail. horiz. completely filled)
path layers
NO CALCULATION: no calculation possible (no paths and no
water level available)
NO CALCULATION: no calculation possible section
Any 0 NormalDischarge: open channel and partially filled pipe normal
error
discharge (no paths but water level available, flow measurement
by lookup table)
LogLaw: open channel and partially filled pipe hydraulic
calculation with log law (any cross section geometry)
1E1P PipeHydraulic: filled pipe hydraulic calculation, based on
1 PowLaw: open channel and partially filled pipe hydraulic
2E2P Nikuradse velocity profile
calculation with power law (rectangular and trapezoidal cross
section geometries)
PipeIntegration: filled pipe numeric integration, based on
different velocity profiles:
Gauss-Jacobi: Gauss-Jacobi integration method
(IEC 60041, round pipe)
Gauss-Legendre: Gauss-Legendre integration
method (rectangular pipe)
OWICS: Optimized Weighted Integration for
≥ 1E2P Circular Sections (including correction of sensor MeanSection: open channel and partially filled pipe mean
1
≥ 2E4P positions, round pipe) section calculation method
MANUAL 22.730.0085221.001.01.4.4
Note
During normal operating mode, measurement values are shown as floating point numbers
with the defined unit system.
If a measurement is not possible or the measurement value is not available, the value “NaN”
is shown which means “Not a Number”.
NaN is a numeric data type value representing an undefined or unrepresentable value,
especially in floating-point calculations. Systematic use of NaNs was introduced by the IEEE
754 floating-point standard in 1985, along with the representation of other non-finite
quantities like infinities.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
The following measurement setups are available (depending on the installed software version and license):
Filled pipe (PIPE)
Open channel (CHANNEL)
Partially filled pipe (PIPE-CHANNEL)
Clamp On (CLAMPON)
7.9.1.1. General
Example for pipe rectangular, 2E8P = 8 paths:
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
7.9.1.1.1. Description
Parameter Definition Unit Default
Value
name
Description - Description of this section, e.g. name - -
R7-R8 R7-R8
R5-R6 R5-R6
R3-R4 R3-R4
R1-R2 R1-R2
7.9.1.1.3. Geometry – open channel (CHANNEL) and partially filled pipe (PIPE-CHANNEL)
The open channel and partially filled pipe (CIRCLE, RECTANGLE and TRAPEZOID) geometries are defined
by the following parameters:
MANUAL 22.730.0085221.001.01.4.4
1 2
B
B
Figure 29: STANDARD Mode Measurement Parameters (open channel / partially filled pipe
geometries)
Note
Only the cross sections “CIRCLE” and “RECTANGLE” can be used for partially filled pipe
measurements!
In addition to the three standard geometries (CIRCLE, RECTANGLE and TRAPEZOID), a forth open
channel geometry USER-DEFINED is available for arbitrary cross sections:
By dividing an arbitrary cross section into up to 20 segments an interpolated area can be defined.
Each segment is defined by three parameters: level, A(level), P(level)
Linear approximation is used to calculate the intermediate values.
MANUAL 22.730.0085221.001.01.4.4
R9-R10
R7-R8
R5-R6
... A(level n)
P(level n)
R3-R4
R1-R2 A(level 2)
level 2
level n
level 1
A(level 1) P(level 2)
P(level 1)
Figure 30: STANDARD Mode Measurement Parameters (open channel geometry USER-DEFINED)
The USER-DEFINED geometry is defined by a table which consists of up to 20 value triplets; at least one
value triplet is needed for flow calculation.
A triplet consists of:
Elevation/height of the segment (height above channel floor): level
Cross-sectional area of the segment: A(level)
Length of the wetted channel wall of the segment (without water surface): P(level)
A(level n)
P(level n)
Length of wetted
channel wall P(level):
MANUAL 22.730.0085221.001.01.4.4
Note
MANUAL 22.730.0085221.001.01.4.4
The flow cutoff sets small positive and negative flow values to zero, as shown on the following figure:
Q
lv_Q_cutoff
0 t
-lv_Q_cutoff
MANUAL 22.730.0085221.001.01.4.4
Caution
L_C1
T2
L_
phi
phi
T1
L_
L_C2
Figure 33: STANDARD Mode Measurement Parameters (transducer offset, cable length)
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Caution
MANUAL 22.730.0085221.001.01.4.4
Caution
The normal discharge table is only available for open channel and partially filled pipe
configurations!
Values out of range are set to the minimum value (< lower margin) or extrapolated (> upper
margin)!
MANUAL 22.730.0085221.001.01.4.4
Caution
Some changes made in this section cause an immediate reboot of the system!
Changing the IP address will also interrupt the network connection!
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
7.10.1. Description
Add a description for this installation which will appear on the Status page.
Value Definition Default
Description Description of this system / installation. -
The description string must be a combination of characters (a-zA-Z0-9_-
/+*%(){}<>[]:;,.$?!) and spaces with a maximum length of 32 characters to be valid!
7.10.3. MMI/LCD
Define the MMI/LCD language (refer to the software data sheet and/or release notes for a list of supported
languages), backlight timeout (Off, 1, 2, 5, 10 minutes), default measurement value (fixed measured value
display after restart) and additional MMI permissions (refer to chapter 0):
7The client operating system and web browser must support Eastern Asian languages and fonts for a
correct visualization!
201707 Hir, Ges Subject to change Page 102/198
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Autom. daylight Enable/disable the automatic time change depending on the daylight saving OFF
saving time time (defined in the time zone above):
OFF
ON
Time Enable/disable the time synchronization with the built-in RTC (hardware real ON
synchronization time clock):
with RTC
OFF
ON
MANUAL 22.730.0085221.001.01.4.4
7.10.6.1. Restrictions
Some restrictions apply especially for the MMI as it has a very reduced number of available digits (2 lines, 20
characters on each line):
1. By default, two measurement values are shown with 6 relevant digits on the MMI, one per line and
with complete identifier (up to four characters, including section number) and measurement unit (up
to six characters):
01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20
N N N N = S V V V P V V V U U U U U U
N N N N = S V V V P V V V U U U U U U
Legend:
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Caution
Enabling and/or setting the "Reset volume" clears all section volume counters!
The "Volume forward/reverse reset counter" (Vf_rst / Vr_rst) goes to 0 after 4’294’967’295
(2^32-1) reset counts!
By default:
All volume counters are persistently stored in 5 minute intervals (for an indefinite time span, even
while unpowered).
If the RIMOCTRL module is shut down (or restarted) by web interface, mmi or sleep mode: all
volume counters are stored with their actual value.
If the RIMOCTRL module is abruptly unpowered the actual values can't be stored. In this worst case
the volume counts of the last 5 minutes are lost (limitations of the used real-time operating system).
7.10.10. SNTP client
Define an external SNTP9 server and polling interval/timeout (if available/needed):
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
7.10.12. Password
Define a system password. Limitations:
must be at least 4 characters long
must be at maximum 16 characters long
must not contain other characters than: a-zA-Z0-9_+-$?!%&
default setting (after factory reset): 1234
MANUAL 22.730.0085221.001.01.4.4
Caution
The error message “ERROR: NET eth0” on the MMI indicates that there are duplicate IP
addresses in the local LAN1 network i.e. a second device (personal computer, another
RICTRL module, …) is configured to the same IP address as this RICTRL module; therefore,
no access to the web interface is possible. Resolve the network conflict immediately!
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
7.10.18. SMSD
Enable or change the SMSD10 settings used for SMS alarms or requests.
The SMSD interface can be used to send SMS messages with status and measurement values
automatically, or authorized user can request status and measurement values by sending special formatted
SMS messages to the RICTRL. Additionally, special system events may send SMS alarms, too:
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Note
Supported by default are the GSM modems / terminals Cinterion MC52i (9600 Baud)
and CEP GSM UMTS HT63E (115200 Baud), connected via 1:1 cable to RS232
COM1.
Acknowledgment of an SMS alarm is done with a reply-SMS with the station-
password as content, e.g. “1234”. Several pending SMS alarms must be
acknowledged individually!
A status/value request via SMS is done with the station-password followed by the
key string (without space, as content of the message), e.g. “1234Status”.
7.11. Input/Output
Inputs and outputs are made with so-called input and output rules.
The available physical devices are listed below:
Built-in analog inputs and outputs (RICTRL, RIMOUSTT)
Built-in digital inputs and outputs (RICTRL, RIMOUSTT)
External Modbus Slave devices, connected to the RICTRL and configured as Modbus devices:
o Generic Modbus driver
o MPC/MPJ
o RIPDI
o RICODE-RELAYS
201707 Hir, Ges Subject to change Page 123/198
MANUAL 22.730.0085221.001.01.4.4
Note
The external GSM modem can be used for requesting system and section values, too; for
this so-called SMS-request can be defined.
In addition to the physical inputs and outputs, additional digital communication interfaces are available in this
section:
External Modbus Master device (no additional configuration necessary)
Remote IEC104 partner stations
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Note
On a RISONIC measurement up to two different measuring value inputs (e.g. height levels)
can be assigned to one section. If two height levels are assigned the average value is
automatically calculated and used.
Note
Analogue inputs of RIMOUSTT modules can only be used if at least one path of the
RIMOUSTT module is assigned to the current section.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
- Moving average
Limit value LU (off, 5, 10, 15, 20,
30, 45, 60, 120 s)
- Moving average
Limit value LU INV (off, 5, 10, 15, 20,
30, 45, 60, 120 s)
- Hysteresis
- Moving average
Limit value LUH
(off, 5, 10, 15, 20,
30, 45, 60, 120 s)
- Hysteresis
- Moving average
Limit value LUH INV
(off, 5, 10, 15, 20,
30, 45, 60, 120 s)
- Hysteresis
Flow direction (RISONIC modular - Moving average
only) (off, 5, 10, 15, 20,
30, 45, 60, 120 s)
- Hysteresis
Flow direction INV (RISONIC - Moving average
modular only) (off, 5, 10, 15, 20,
30, 45, 60, 120 s)
MANUAL 22.730.0085221.001.01.4.4
- Scaling
- Pulse Time = 100
Impulse
ms, 200 ms, 1000
ms)
- Scaling
- Pulse Time = 100
Impulse INV
ms, 200 ms, 1000
ms)
MANUAL 22.730.0085221.001.01.4.4
- Threshold
- Time limit
Trend
- Trend (rising,
stable, falling)
- Type (daily,
Switch clock (note: does not need
periodic, daily +
an input value i.e. works
periodic
independently)
- (De)activation time
7.11.4. Add Modbus output devices (e.g. RICODE-RELAYS for parallel code output)
In order to make a measuring value via Modbus output available, you first have to define the according
Modbus device:
1. In the menu on the left select:
STANDARD Input/Output
2. Add a Modbus device:
Modbus devices Add …
Define a unique name as identifier:
Identifier “RICODE-RELAYS” Next
Select a device:
Device name RICODE-RELAYS Next
Check device address:
Address 247 Save
MANUAL 22.730.0085221.001.01.4.4
Note
MANUAL 22.730.0085221.001.01.4.4
Note
A complete list of the available Modbus Slave registers is also available in the system report;
after changing measurement parameters the system report needs to be re-generated in
order to show the changes.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
A detailed overview of all IEC 104 functions can be found in the software datasheet “Rittmeyer
Instrumentation Controller IEC 60870-5-104 E21_220_0067760_004_xx”.
MANUAL 22.730.0085221.001.01.4.4
Identifier: Alarm_SMS
Output device: SMS
Application: Limit value LU
Process value: Water level [m](pv1,1)
Lower value L [m]: 0.4
Upper value U [m]: 0.6
Option: Cutoff time
SMS message: Alarm Limit value LU: Water Level [m] (pv1,S01)
An alarm message is sent as soon as the limit value is activated. This message must be answered with the
station password within the acknowledge time, otherwise the SMS daemon retries and sends another text
message.
An alarm may look as follows on an Android smart phone:
MANUAL 22.730.0085221.001.01.4.4
Alarm
Alarm
1st retry
Confirmation
MANUAL 22.730.0085221.001.01.4.4
Request
Response
MANUAL 22.730.0085221.001.01.4.4
7.11.7.3. Diagnostics
Diagnostics information are available in EXPERT Mode Diagnostics GSM state infos. In this section it
is also possible to query the remaining budget on a prepaid SIM card or sending arbitrary USSD12 codes.
Note
Do not edit these values unless you have a highly accurate (much better than 0.1 %FS) and
calibrated meter available to check your settings.
12USSD = Unstructured Supplementary Service Data is a protocol used by GSM cellular telephones to
communicate with the service provider’s network system
201707 Hir, Ges Subject to change Page 140/198
MANUAL 22.730.0085221.001.01.4.4
Temporarily set the analog output of a RIMOUSTT or RICTRL module (test function)
Temporarily set the digital outputs of a RIMOUSTT or RICTRL module (test function)
7.12. Datalog
7.12.1. Add datalog tasks
Each Instrumentation Controller features advanced data logging functionality; by default, all relevant
measurement values, i.e. characteristic values are logged with their units and saved in a semicolon-
separated text file (extension .txt) on the Compact Flash card.
Available datalog files can be downloaded or copied to a USB memory stick even while a datalog is still in
progress (partially).
The datalog file name is concatenated from the following components (each separated with a hyphen “-“):
File type - Start date and time or date - Identifier - File extension
and sequence number (defined above)
LOG - YYYYMMDDHHMMSS/ - simpledate - .txt
YYYYMMDD-NN
MANUAL 22.730.0085221.001.01.4.4
Examples of datalog tasks, including their status, and some data log files:
MANUAL 22.730.0085221.001.01.4.4
Description Parameter/
measurement
value
Date Date
Time Time
Status of measuring point: state_section
-1: Section Error
0: Section OK
1: Section Warning
2: Section Not-Wet
Adjusted and filtered flow Q
Average flow velocity v_m
Channel: external water level H_ext
Volume (forward) Vf
Volume (reverse) Vr
Average water temperature T_m
Average speed of sound c_m
Adjusted and unfiltered flow Q_adj
Flow correction dQ
Unfiltered and unadjusted flow Q_raw
Channel: calculated bottom coefficient for the MeanSection method K_bot_calc
Channel: calculated wetted surface at current water level A_calc
Channel: calculated wetted circumference at current water level P_calc
Channel: hydraulic radius R_h
External flow speed v_ext
External flow Q_ext
External water temperature T_ext
Calculation method: calculation_method
-1: NO CALCULATION
0: Gauss-Jacobi
1: Gauss-Legendre
2: OWICS
3: OWIRS
4: OWISS
5: PipeHydraulic
10: NormalDischarge
11: PowLaw
12: LogLaw
13: MeanSection
Table 7: Datalog contents, example for measurement type RISONIC modular per section
MANUAL 22.730.0085221.001.01.4.4
Description Parameter/
measurement
value
Layer status: state_layer
-1: Layer Error
0: Layer OK
2: Layer Not-Wet
Average axial layer flow speed v_axial
Average cross layer flow speed v_cross
Table 8: Datalog contents, example for measurement type RISONIC modular per layer
Description Parameter/
measurement.
value
Path status: state_path
-1: Path Error
0: Path OK
2: Path Not-Wet
Path status forward, refer to chapter 10.3.1 state_tt_forward
Path status reverse, refer to chapter 10.3.1 state_tt_reverse
Path amplitude forward tx_data_ampl_f
Path amplitude reverse tx_data_ampl_r
Path gain forward gain_act_f
Path gain reverse gain_act_r
Adjusted and filtered transit time in positive flow direction tf
Adjusted and filtered transit time against of positive flow direction tr
Filtered differential transit time dt
Average path water temperature T
Average filtered path sonic speed c
Average adjusted and filtered path flow velocity v
Average adjusted path flow speed v_adj
Average unadjusted path flow speed v_raw
Factor for protrusion correction K_prot
… …
Table 9: Datalog contents, example for measurement type RISONIC modular per path
Description Parameter/
measurement.
value
Depending on configured devices (RICTRL, RIMOUSTT): …
state and value of DO, DI, AO, AI
Table 10: Datalog contents, example for measurement type RISONIC modular per device
MANUAL 22.730.0085221.001.01.4.4
If an export filter is enabled (refer to chapter 7.10.8) an additional export column is available: “Filtered (ON)”.
By selecting a file in this column, only a subset of all possible datalog columns is exported in the selected
file.
Note
With the help of a typical spreadsheet program (like e.g. Microsoft ® Excel) you can easily
evaluate the datalog files.
MANUAL 22.730.0085221.001.01.4.4
b. Delimiter = Semicolon. This many times does not matter in opening .txt file; Excel many
times already understands what the delimiter is. The ‘Data Preview’ window will show you if
you succeeding.
c. Click Finish.
5. Save As, Save As Type: Excel Workbook.
It is possible to record the measurement values with/without average and to show them with user-definable
labels; an automatic scaling of x- and y-axis in order to show all values simplifies the visualization.
Note
Changing the trend system settings deletes previously recorded trend data!
MANUAL 22.730.0085221.001.01.4.4
Two buttons “Download chart” and “Download data” allow to save the visualized trend data as SVG graphic
or CSV text files.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
8.1. Diagnostics
8.1.1. Section popup
The section popup opens a new browser window which gives you a more detailed live view of the
measurement section, besides status information and system timestamp:
Current calculation mode, refer to chapter 7.8.2.2 “Section calculation method” for a list of all
possible calculation methods
Detailed volume counter information incl. reset counts
For each horizontal layer:
o Axial velocity
o Cross velocity – helpful for difficult hydraulic situations
o Path state(s)
Possibility to generate a Service report (send the generated ZIP file to Rittmeyer AG for
troubleshooting purposes)
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
8.1.2.1. Status
Please refer to chapter 7.8.2.4 “Path Status“ for a list of all possible states.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
The much more detailed characteristic values counters are usually folded, unfold them with the sign. They
are extremely helpful while diagnosing a RISONIC modular setup as they may indicate problems with the
fundamental signal acquisition:
Each characteristic value counter corresponds to an internal processing step and is incremented if
an appropriate error arises.
Some, but not all, characteristic values counters are summed into the total number of bad
measurements.
All counters are reset on power-up or by selecting “Status counter clear” on the service page.
Note
All characteristic value counters are reset on power-up or by selecting “Status counter clear”
on the service page.
Usually it’s a good idea to reset these counters after configuration changes, in order to begin
with fresh starting conditions!
The last timestamp of clearage is shown on the bottom of the receive signal graph popup
window!
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
TOTAL OUTLIER n_low_outlier Not enough values in the outlier filter, refer to chapter
FILTER EMPTY 8.2.3.4, n_value_min
NO RECEIVE SIGNAL n_ns_f Number of measurements with no receive signal in
FORWARD / REVERSE n_ns_r forward / reverse direction
RECEIVE SIGNAL TOO n_sl_f Number of measurements with a “too low” receive
LOW FORWARD / n_sl_r signal in forward / reverse direction
REVERSE
RECEIVE SIGNAL TOO n_sh_f Number of measurements with a “too high” receive
HIGH FORWARD / n_sh_r signal in forward / reverse direction
REVERSE
CUT BORDER n_cb_f Number of measurements with a cut border problem in
PROBLEM FORWARD / n_cb_r forward / reverse direction
REVERSE
MINIMUM OUTSIDE n_mm_f Number of measurements with a global minimum found
WINDOW FORWARD / n_mm_r outside the receive window in forward / reverse
REVERSE direction
DC-OFFSET TOO HIGH n_dch_f Number of measurements with a dc offset too high in
FORWARD / REVERSE n_dch_r forward / reverse direction
PREPROCESS n_calc_preproc_f Number of measurements with an invalid preprocessing
INVALID FORWARD / n_calc_preproc_r in forward / reverse direction
REVERSE
WINDOW SELECTION n_calc_win Number of measurements with an invalid receive
INVALID window selection
CORRELATION n_calc_corr Number of measurements with an invalid correlation
INVALID
INTERPOLATION n_calc_interp Number of measurements with an invalid interpolation
INVALID
TRANSIT TIMES n_calc_t_raw Number of measurements with an invalid absolute
INVALID transit time
ADJUSTMENT INVALID n_adjust Number of measurements with an invalid adjustment
VALIDATION n_valid_abstime Number of measurements with an incorrect validation of
ABSOLUTE TIME the absolute transit time
VALIDATION DELTA n_valid_deltatime Number of measurements with an incorrect validation of
TIME the delta time
MANUAL 22.730.0085221.001.01.4.4
Figure 69: States and values of inputs and outputs, transducer test
Note
Modbus slave devices must be powered during the startup of the RICTRL module. This
means that these devices may not be powered using 24VDC OUT of RICTRL if their startup
time is too long!
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Caution
Some of the changes made in the EXPERT mode may get overwritten when basic
parameters are changed afterwards in STANDARD mode!
MANUAL 22.730.0085221.001.01.4.4
8.2.2.1. General
8.2.2.1.1. General
Value Parameter Definition Unit Default
name
RIMOUSTT enable Enable/disable the ultrasonic module: - TRUE
ENABLE
TRUE
FALSE
REMOTE remote RIMOUSTT is remotely powered: - FALSE
POWERED
TRUE (powered by external 24VDC
power supply)
FALSE (powered by RICTRL 24VDC
OUT)
8.2.2.1.2. Watchdog
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
8.2.2.2.1. General
Value Parameter Definition Unit Default
name
PATH ENABLE enable Enable/disable the ultrasonic module path: - TRUE
TRUE
FALSE
NUMBER OF n_burst Number of bursts 1
-
BURST
8.2.3. Sections
Parameters can be changed separately for each configured section. Only the most important parameters are
explained in more details, other parameters should not be changed.
MANUAL 22.730.0085221.001.01.4.4
8.2.3.1. General
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Hydraulic calculation:
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Hydraulic calculation:
Parameter Definition Unit Default
Value
name
CHANNEL channel_law Select the single layer channel calculation [] POW_LAW
SINGLE LAYER method:
LAW
POW_LAW
LOG_LAW
MANUAL 22.730.0085221.001.01.4.4
Note
Values lower or higher than the default settings may lead to section warnings/errors and/or
system errors - do not change any values if not advised by Rittmeyer AG!
MANUAL 22.730.0085221.001.01.4.4
Some figures to explain these limit values, vertical axis shows output of flow Q (if valid, otherwise a section
error is shown):
MANUAL 22.730.0085221.001.01.4.4
lv_Q_raw_max
lv_Q_raw_min
0 t
-lv_Q_raw_min
-lv_Q_raw_max
Figure 76: EXPERT Mode Measurement Parameters (lower/upper flow limit values)
Q
dQ/dt
lv_dQ_raw
0 t
-lv_dQ_raw
Figure 77: EXPERT Mode Measurement Parameters (maximum flow change limit value)
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Each configured path has an input buffer of configurable length buffer_size, organized with the following
contents / columns:
Timestamp time
Transit time forward tf_filter
Transit time reverse tr_filter
201707 Hir, Ges Subject to change Page 172/198
MANUAL 22.730.0085221.001.01.4.4
...
Path 2
Path 1
time tf_filter tr_filter dt_filter state_tt
time tf_filter tr_filter dt_filter state_tt
16:32:05::920 714.53ms 714.52ms 1.3453us nok
16:32:05::920 714.53ms 714.52ms 1.3453us ok
16:32:05::941 714.54ms 714.53ms 1.7445us ok
16:32:05::941 714.54ms 714.53ms 1.7445us ok
16:32:05::959 714.54ms 714.54ms 1.2199us ok
buffer_
16:32:05::959 714.54ms 714.54ms 1.2199us nok
size
1...256 : : : : :
: :: :: :: :: :
Default: : : : : :
256
16:32:06::907 714.94ms 714.44ms 9.754us nok
16:32:06::907 714.94ms 714.44ms 9.754us ok
16:32:06::931 714.92ms 714.41ms 9.810us nok
16:32:06::931 714.92ms 714.41ms 9.810us nok
16:32:06::955 714.89ms 714.49ms 9.698us ok
16:32:06::955 714.89ms 714.49ms 9.698us ok
Figure 79: EXPERT Mode Measurement Parameters (outlier filter path input buffers)
Once per second, the RISONIC modular filters valid measurement values (values older than valid_time
are ignored) with a configurable maximum count (filter_size) out of these path input buffers into three
separate outlier filters. The outlier filters are sorted in a descending order in order to concentrate the
measurement values like a histogram:
tf_filter tr_filter dt_filter
715.01ms 714.88ms 10.619us
Figure 80: EXPERT Mode Measurement Parameters (outlier filters for each path)
In an additional step a mean filter is applied on these buffers (filter_size) to get filtered measurement
values for further processing:
If zero or less than n_value_min valid values are found the path state is set to error.
For open channels and partially filled pipe: validation includes the water level H_ext. If a path is not
available, its state is set to not wet.
MANUAL 22.730.0085221.001.01.4.4
Depending on the measurement setup (short or long paths, number of paths, number of sections etc.), the
behavior of the RISONIC modular calculation may be fine-tuned: changing these outlier filter parameters
heavily influences the responsiveness of the system regarding flow changes and sensitivity to erroneous
situations:
After power up: response time until valid values are shown (filter is initially empty and must be filled
with valid measurements)
Response time until flow changes are shown (large values of filter_size and valid_time give
slow responsiveness to changes)
Delay time until measurement errors are shown; sporadic measurement errors may be filtered
completely
8.2.3.5. Section x, Layer y
MANUAL 22.730.0085221.001.01.4.4
b4
R7-R8
d4
b3
R5-R6
d3
H
d2
b2
R3-R4
d1
b1
R1-R2
b4
R7-R8
d4
b3
R5-R6
d3
d2
b2
R3-R4
d1
b1
R1-R2
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
b5
R9-R10
z5
b4
R7-R8
z4
b3
R5-R6
z3
b2
R3-R4
z2
b1 z z_L z_U
R1-R2
z1
b5
R9-R10
z5
b4
R7-R8
z4
b3
R5-R6
z3
b2
R3-R4
z2
b1 z z_L z_U
R1-R2
z1 s1 tan( 1)
1 2 s2 tan( 2)
MANUAL 22.730.0085221.001.01.4.4
b5
R9-R10
b4 z5
R7-R8
b3 z4
R5-R6
b2 z3
R3-R4
b1 z2
R1-R2 z z_L z_U
z1
R9-R10
R7-R8
R5-R6
... A(level n)
P(level n)
R3-R4
R1-R2 A(level 2)
level 2
level n
level 1
A(level 1) P(level 2)
P(level 1)
MANUAL 22.730.0085221.001.01.4.4
𝐿𝑖
𝑧𝑖,𝑈 = 𝑧𝑖 + 60 ∙ √ + 0.2
𝑓
𝐿𝑖
𝑧𝑖,𝐿 = 𝑧𝑖 + 60 ∙ √
𝑓
The parameters z_U and z_U define the levels at which the ultrasonic path switches on and off, depending
on the water level z. The following figure shows the threshold area:
MANUAL 22.730.0085221.001.01.4.4
Water level/height
zU
z1,U
z1,L
zL
z1
Off On
0,597566
For reference and path arrangements with more than 4 horizontal layers (e.g. 2E16P), an excerpt from the
draft standard is shown below. Please note: this table is provided “as is”, with no guarantee or assurance
that it will be released in this way! Please refer to the official standard committee for additional information.
13 d is the distance from the centerline of the conduit to the acoustic path
201707 Hir, Ges Subject to change Page 180/198
MANUAL 22.730.0085221.001.01.4.4
0.064413
0.150878
0.218183
0.269783
0.295355
8.3. Syslog
In addition to the Diagnostics which shall help to solve problems with the flow measurement itself, the
RISONIC modular features a Syslog that can solve problems with the controller / installation i.e. the system.
The Syslog stores all events in text files (located on the Compact Flash Card) for later analysis, each event is
logged with a time stamp:
MANUAL 22.730.0085221.001.01.4.4
The syslog is saved onto the Compact Flash Card; by default, 10% of the total space is reserved – amongst
other things – for storing syslog events.
8.4. Service
The instrumentation components (RICTRL, RIMOUSTT) do not need to be serviced i.e. are maintenance-
free.
Some software service functions are available on the service page:
MANUAL 22.730.0085221.001.01.4.4
System: Show status information, CPU load, memory consumption. Shutdown/restart the system,
start/stop the measurement, reset volume and status counters, factory reset
System report
Parameter backup/restore (internal backup)
Parameter export/import (external backup)
Software update via web interface / remotely (only when measurement stopped and enabled by the
rotary switch Mode)
Versions: Show version information per module (software and hardware versions)
8.4.1. Reboot of the controller
Reboot of the controller is possible by pressing the reboot button: the datalog will be properly terminated, the
syslog and the persistent data (e.g. volume counters) will be saved.
Pressing the Reset button on the MMI will immediately restart the controller, without ability to terminate or
successfully save syslog and/or datalog files.
Caution
Do not power-off the controller abruptly as this may result in faulty datalog and syslog
entries!
Avoid extremely short power-losses, use an uninterruptable power supply!
Caution
If the power is not removed within 5 minutes, the controller will restart and start-up again; this
in case the shutdown button has been accidently pressed!
8.5. License
Detailed information according to the current license on your RICTRL is displayed on this page together with
maintenance functions for saving and restoring license information and generating license requests.
For license upgrades please contact a Rittmeyer AG representative.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
9. MMI
The MMI on the front of the RICTRL is generally a read-only interface with the option of selected additional
service tasks, such as copying log files to a USB memory stick or shutting down the system. Pressing any
button will turn the backlight on; this actuation will not change a setting or activate a sub-menu. The backlight
turns off after a defined time to save power (see system settings in STANDARD Mode).
Usually, the MMI starts in Measurement mode, i.e. it displays measured values which can be selected by
pressing the or button. When the “OK” button is pressed the menu tree is entered as shown in the
following figure:
Power-
up / reset
Error
Measure
Error OK values
OK
OK
Menu: Info
Info
Info
Info Exit ...
...
...
OK
Menu: Measure
Measure
Measure
Measure Exit ...
...
...
OK
OK
Menu: Service
Service
Service
Service Exit ...
...
...
Exit
MMI permissions can be set in the system settings, refer to chapter 7.10.3. For a complete listing of the
menu structure refer to the RICTRL software datasheet.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
10. Troubleshooting
10.1. System errors
Typical causes:
Configured devices are not available or not working correctly (missing RIMOUSTT modules or
Modbus Slave devices)
No compact flash card inserted
System parameter error
Internal device error
Checking the status page and syslog will probably help to identify the root cause.
10.2. Section warnings and errors
Typical causes:
Path failure(s)
Limit values violated
Open channel and partially filled pipe: missing level/height information
10.3. Path errors
Typical causes:
Bad signal quality:
o Receive signal too high/low
o Receive signal reflections, resulting in pre-processing and/or correlation errors
Air bubbles and/or too much silt, resulting in intermittent path failures
Transducer and/or cabling failures
Geometrical issues i.e. wrong parameters
…
10.3.1. Datalog status counters
Setting up a (diagnostics) datalog task may help troubleshooting path errors.
A typical datalog file includes detailed status information for each configured path (refer to chapter 7.12.2),
namely the following columns; each data column must be checked in case of path errors:
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
max
Reflection
max
n_loc_mm = 2
- cut-off the reflection with n_zero_cross, e.g. set it to 7. With this at the given number=’zero cross’
after the global peak, the signal will be set to zero.
MANUAL 22.730.0085221.001.01.4.4
Disadvantage: the correlation will use a smaller signal and is more sensitive to correlation errors.
max
n_zero_cross = 7
Most probably, these changes will lower the characteristic value „PREPROCESS INVALID“ significantly.
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
11. Maintenance
11.1. Cleaning
Note
The surface of the electronic modules can be cleaned with a dry cloth. To not use any
liquids, solvents etc.
11.2. Service
Caution
Neither RICTRL nor RIMOUSTT contain serviceable components; do not open their
housings as these modules contain highly integrated and sensitive electronic circuits
susceptible to electronic discharge!
Warranty is void if the housing of a module has been opened by unqualified
personnel!
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4
Q V
Q 87 v 88, 188
Q_damp ...................................................................... 95, 164 V 153
v_axial...........................................................................151
v_cross...........................................................................151
R v_damp .............................................................................164
R 92, 167 V_m ......................................................................................87
valid_time ...................................................................174
Vf ........................................................................................87
S Vr ........................................................................................87
s_1.............................................................................. 92, 167
s_2.............................................................................. 92, 167 W
state_path ................................................................... 188
w 176
w_axial...........................................................................175
T wd_aout...........................................................................161
T 153, 188 wd_dout...........................................................................161
t_damp ............................................................................ 164 wd_timer ........................................................................161
T_m...................................................................................... 87 weightings ...................................................................166
tf ...................................................................................... 153
tx_data_ampl_avg_f ................................................ 154 Z
tx_data_ampl_r.......................................................... 188
tx_pwr .............................................................................. 94 z 97, 179
z_axial...........................................................................175
z_L ....................................................................................179
U z_U ....................................................................................179
uncorr .............................................................................. 98
MANUAL 22.730.0085221.001.01.4.4
13. Index
Integration weight coefficients 180
# Intended use 12
IOA 135
OK buttons 20 IP address 55
IP-address 54
2
24VDC IN connector 26 L
LAN1 54, 55
5 LAN1 connector LEDs 20
LAN1 port for management access 20
50-60Hz switch 26 LAN2 54, 55
LAN2 connector LEDs 20, 26
LAN2 Ethernet connector 26
A LAN2 port for US-Module(s) 20
AI connector 20, 26 LCD 20
ANGLE PHI phi 72, 75, 77 Level trigger 140
AO connector 20, 26 login authorization 55
C M
CABLE LENGTH L_C 72, 75, 77 Measurement Type 71, 74, 77
Change network settings for LAN1 115 METRIC/SI 106
Clamps for standard DIN rail 27 MMI/LCD backlight timeout 102
Compact Flash slot 20 Mode switch 20, 26
Configuration 71, 74, 77
N
D Net mask 54, 55
Default gateway 54, 55
default password 55 P
DESCRIPTION 72, 74, 77
DHCP client 48 Parent Section A, intelligent application 80
diagnostic datalog 141 PATH A … PATH D LEDs 26
digital impulse counter 140 PIPE HEIGHT 72, 75, 77
DIN rail 52 Power input connector 20
DO 1 … DO 4 connectors 20, 26
DO 1 … DO 4 LEDs 26
R
E R1 … R8 connectors 26
Reset button 20, 26
EARTH connector 26, 53 RIMOUSTT PATH 72, 75, 77
Environmental conditions 12 RS232 interface connector 20
EXPERT mode 149 RS485 interface connector 20
RS485 termination switch 20
F
S
Factory reset button 20
Fine tuning 140 Safety 11
sequence datalog 141
Shape 71, 74, 77
H Shutdown of the instrument 183
http://192.168.1.1 55 simple date datalog 141
Sleep LED 20
SOUND PATH LENGTH L 72, 75, 77
I STANDARD Mode 69
Status LED 20
ID rotary switch 53, 54 Status LED 26
ID switch 26 Status output connector 20
Identifier 126, 128, 133 Storage 12
IMPERIAL 106, 110
201707 Hir, Ges Subject to change Page 197/198
MANUAL 22.730.0085221.001.01.4.4
MANUAL 22.730.0085221.001.01.4.4