Name/Function Diathermy: Pairt Number Description Qty Note
Name/Function Diathermy: Pairt Number Description Qty Note
Name/Function Diathermy: Pairt Number Description Qty Note
Technical data
Power requirement 110 / 220 V AC +/- reconnectable
Current consumption max 440 VA
Nominal frequency 500 kc/s
Impulse frequency of modulation 33 kc/s
CREST-FACTOR
Current type spray-coagulation 7.6
Current type contact- coagulation 3.0
Current type pure cut 1.6
Current type blend I 2.2
Current type blend II 2.6
Current type blend III 3.0
Current type bipolar coagulation 1.8
HF-OUTPUT
Current type spray-coagulation 80 Watts +/- 20 % on 300 ohm
Current type contact- coagulation 175 Watts +/- 20 % on 200 ohm
Current type pure cut 175 Watts +/- 20 % on 300 ohm
Current type blend I 175 Watts +/- 20 % on 300 ohm
Current type blend II 175 Watts +/- 20 % on 300 ohm
Current type blend III 175 Watts +/- 20 % on 300 ohm
Current type bipolar coagulation 50 Watts +/- 20 % on 100 ohm
Weight 8,8 kg
Dimensions 130 mm height
400 mm width
380 mm depth