Professional Documents
Culture Documents
Difference Between Dispersive and Fourier Transform IR Spectrophotometer
Difference Between Dispersive and Fourier Transform IR Spectrophotometer
Instrument Emergence:
Components:
The components of dispersive IR includes are: radiation source, mirrors, reference cell,
sample cell, attenuator, beam chopper, slit, grating monochromator and detector, on the
contrary in the FT-IR it contain the interferometer, source, beam splitter, fixed mirrors,
moveable mirrors, laser and detector.
1
The components of dispersive IR have many moving parts and it results in the mechanical
slippage whereas in FTIR only mirror moves. so less the wearing out of components,
greater will be the reliability of spectrometer.
In an FTIR spectrophotometer few mirror surfaces are present and as a result there are less
reflection losses than in dispersive IR spectrophotometer.
The FT-IR doesn’t use slit as used by dispersive IR spectrometer. Because of no slit the
energy production is a lot greater than in a dispersive instrument.
Also because of presence of slit in dispersive IR it limits the frequency reaching the
sample.
The detector in dispersive IR spectrophotometer can receive only few percent of energy of
light source as compared to FT IR spectrophotometer in which 50% of original light source
is detected.
Spectrum:
Sample:
2
The sample is placed between the source and the monochromator in the dispersive IR
spectrophotometry and in the case of FT-IR spectrophotometer the sample is typically
between the interferometer and the detector.
The sample compartment in dispersive IR is cluttered whereas in FT-IR it is less messy.
In FT-IR spectrophotometer the detail in sample spectrum will be easily differentiate as
compared to dispersive IR of the same sample.
The IR radiation from the sample spectrum can easily detected in dispersive IR detector
whereas it is opposite in FT IR, no emission from sample is detected.
Calibration:
Resolution:
In dispersive IR only small infrared beams are allowed to pass from the slit which can
improve the resolution of the instrument. While large IR beam is used in FT-IR which
provide the higher energy throughout.
The resolution of dispersive IR spectrophotometer is less as compared to FT-IR
spectrophotometer.
The resolution for dispersive IR Spectrophotometer is 0.5-10 cm-1 whereas for FT-IR it is
very high that is 0.1 -0.005 cm-1.
The same is the case for accuracy and sensitivity in the dispersive and Fourier transform
IR spectrometer. As due to sensitivity FT IR is at advantage to identify even small
contaminants.
Beam Optics:
3
The double beam instruments are mostly used as compared to single beam instruments in
dispersive spectrophotometer. While in FT-IR spectrophotometer single beam optics are
used.
The double beam optics provides continuous real-time background subtraction of
dispersive IR spectrophotometer, on the contrary in FT IR the background spectrum is
collected separately in time for sample spectrum, if not collected it result in error.
The single beam has advantage over double beam as it provides greater energy output
whereas double beam provide better baseline flatness and stability for long term stability.
The stray light are involved in dispersive IR spectrophotometer which causes false
readings whereas in FT-IR spectrophotometer no such involvement take place because it
cannot affect the detector as all signals are modulated.
Frequency:
Scan Speed:
Scan speed in dispersive IR instrument is very slow for monitoring samples to undergo
rapid changes. Whereas in the FT-IR fast scan speed is present and it permit monitoring
sample to show rapid changes.
4
Due to slow scan speed in dispersive IR spectrophotometer the sample is contacted with
the thermal effect from the source whereas in FT-IR spectrophotometer no such heat or
thermal effect is subjected to sample from the source because of rapid speed scan.
Only 1 scan of sample measurement is completed in a dispersive instrument
characteristically, however, in the FT-IR many scans can be completed and combined as
compared to those we can see in dispersive IR spectrophotometer.
Usage:
Dispersive IR spectrophotometer:
Use in identification of all types of organic, and many types of inorganic compounds.
Determination of functional groups in organic materials.
Determination of molecular composition of surfaces.
Use in identification of chromatographic effluents and quantitative determination of
compounds in mixture.
Determination of structural isomers and geometric isomers and also in molecular
orientation.
5
Fourier Transform IR spectrophotometer:
6
References
file:///C:/Users/ASUS/Downloads/Documents/productPDF_21615.pdf
https://chem.libretexts.org/Bookshelves/Physical_and_Theoretical_Chemistry_Textboo
k_Maps/Supplemental_Modules_(Physical_and_Theoretical_Chemistry)
https://sg.inflibnet.ac.in/bitstream/10603/134021/9/09_chapter%203.pdf
https://www.slideshare.net/bhavanavedantam/dispersive-ftir
https://www.chem.uci.edu/~dmitryf/manuals/Fundamentals/FTIR%20principles.pdf
https://sci-hub.st/https://doi.org/10.1016/0022-2860(82)85022-9
https://sci-hub.st/https://doi.org/10.1016/B978-0-444-42880-6.50023-4
https://books.google.com.pk/books?principleofinstrumentanalysis
http://www.authorstream.com/Presentation/vvkvjoshi-1948442-dispersive-ir-
spectrophotometer/